Sunteți pe pagina 1din 1

Introduction 9

2  INTRODUCTION

2.1  SERVICE MODE


This section describes the possibility to check individual modules and functions
of the HumaClot Junior operated by authorized staff using the service code.

2.1.1  ACCESS TO THE SERVICE MENU


The service mode is part of the installed software. Special number code grants
accesses for the service technician, to access the hidden service menu do the
following steps:
-- Switch on the Analyzer by plugging the power supply, the below message
will be displayed:

-- While the line is scrolling then it is the only possible way to enter the service
menu by entering the service code 22926.

-- If the service code was entered correctly this message below will be dis-
played:

! Note: Each keystroke is confirmed by a beep tone.

It is not possible to correct the service code.

To re-enter the service code the Analyzer should be turned off


then turned on.
By pressing the left and the right cursor keys you can select be-
tween test functions, calibration functions, and parameter func-
tions by confirm it with Enter Key