EPX-SERIES

Diagnostic X-ray System

Service Manual

Ecotron Co., Ltd


Made by Korea
4F&5F, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul 08511, Korea
Obelis S.A
Bd. General Wahis 53, 1030 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03

D/N: ECO-SM-EPX-001 Rev. G


EPX-SERIES SERVICE MANUAL

(This page intentionally left blank)

Rev. F 2/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

CONTENTS
1. CALIBRATION ________________________________________________________ 8

1.1 EPX-F1600 ____________________________________________________________________________________ 8

1.1.1 LINE POWER SWITCH ___________________________________________________________________ 8

1.1.2 CALIBRATION PROCEDURE ____________________________________________________________ 8

1.2 EPX-F2400 __________________________________________________________________________________ 12

1.2.1 LINE POWER SWITCH _________________________________________________________________ 12

1.2.2 CALIBRATION PROCEDURE __________________________________________________________ 12

1.3 EPX-F2800 __________________________________________________________________________________ 16

1.3.1 LINE POWER SWITCH _________________________________________________________________ 16

1.3.2 CALIBRATION PROCEDURE __________________________________________________________ 16

1.4 EPX-F3200 __________________________________________________________________________________ 20

1.4.1 LINE POWER SWITCH _________________________________________________________________ 20

1.4.2 CALIBRATION PROCEDURE __________________________________________________________ 20

1.5 EPX-F4000 __________________________________________________________________________________ 24

1.5.1 LINE POWER SWITCH _________________________________________________________________ 24

1.5.2 CALIBRATION PROCEDURE __________________________________________________________ 24

1.6 EPX-F5000 __________________________________________________________________________________ 28

1.6.1 LINE POWER SWITCH _________________________________________________________________ 28

1.6.2 CALIBRATION PROCEDURE __________________________________________________________ 28

2. DIP SWITCH SETTING _________________________________________________ 32


3. HARD-WIRED HAND SWITCH CONTROL _________________________________ 33

3.1 HAND SWITCH INTRODUCTION___________________________________________________________ 33

3.2 EXPOSURE PROCESS _____________________________________________________________________ 33

4. COLLIMATOR ________________________________________________________ 34

Rev. F 3/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5. TROUBLE SHOOTING _________________________________________________ 36

5.1 EPX-F1600 / F2400 / F2800 / F3200 / F4000 / F5000 ERROR CODE TABLE ______________ 36

5.2 ACTION TO CORRECT THE ERROR OF EPX-SERIES ____________________________________ 37

6. CALIBRATION _______________________________________________________ 38

6.1 EPX-F1600 __________________________________________________________________________________ 38

6.1.1 VOLTAGE CHECK ______________________________________________________________________ 38

6.1.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 39

6.1.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 41

6.1.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 42

6.1.5 mA CALIBRATION DATA CHECK _____________________________________________________ 43

6.1.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 44

6.2 EPX-F2400 __________________________________________________________________________________ 46

6.2.1 VOLTAGE CHECK ______________________________________________________________________ 46

6.2.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 47

6.2.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 49

6.2.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 50

6.2.5 mA CALIBRATION DATA CHECK _____________________________________________________ 51

6.2.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 52

6.3 EPX-F2800 __________________________________________________________________________________ 54

6.3.1 VOLTAGE CHECK ______________________________________________________________________ 54

6.3.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 55

6.3.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 57

6.3.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 58

6.3.5 mA CALIBRATION DATA CHECK _____________________________________________________ 59

6.3.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 60

Rev. F 4/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4 EPX-F3200 __________________________________________________________________________________ 62

6.4.1 VOLTAGE CHECK ______________________________________________________________________ 62

6.4.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 63

6.4.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 65

6.4.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 66

6.4.5 mA CALIBRATION DATA CHECK _____________________________________________________ 67

6.4.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 68

6.5 EPX-F4000 __________________________________________________________________________________ 70

6.5.1 VOLTAGE CHECK ______________________________________________________________________ 70

6.5.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 71

6.5.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 73

6.5.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 74

6.5.5 mA CALIBRATION DATA CHECK _____________________________________________________ 75

6.5.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 77

6.6 EPX-F5000 __________________________________________________________________________________ 79

6.6.1 VOLTAGE CHECK ______________________________________________________________________ 79

6.6.2 kV, mA WAVEFORM CHECK ___________________________________________________________ 80

6.6.3 WAVEFORM OF POOR AC POWER CONDITION _____________________________________ 82

6.6.4 kV, mA PWM FREQUENCY CHECK ___________________________________________________ 83

6.6.5 mA CALIBRATION DATA CHECK _____________________________________________________ 84

6.6.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK ____________________ 86

Rev. F 5/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

REVSON HISTORY

Revision Number Date Description

A SEP 11, 2013 First Edition


B SEP 30, 2014 European Authorized Representative Change
C JAN 12, 2015 Manual Updates
D MAR 23, 2015 Manual Updates
E JUN 11, 2015 Manual Updates

F AUG 23, 2018 Manual Update


G June 17, 2019 Modify misprinted contents

ADVISORY SYMBOLS

The following advisory symbols are used throughout this manual.

Their application and meaning are described below.

WARNING

Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death
or radiation exposure.

CAUTION

Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of
equipment.

NOTE

Note symbol used to indicate important information need to proper use and right operation of equipment.

NOTE

KEEP THIS SERVICE MANUAL WITH THE EQUIPMENT AT ALL TIMES, AND REVIEW THE
IMPORTANT INFORMATION WHENEVER REQUIRED.

THE STATISTICS AND SPECIFICATIONS OF THIS UNIT AND MANUAL CAN BE MODIFIED
WITH OR WITHOUT NOTIFICATION FOR THE IMPROVEMENT OF PERFORMANCE AND
SAFETY.

Rev. F 6/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

SERVICE CHAPTER
This manual is designed to ensure correct use and operation of EPX model diagnostic x-ray units. Please
read all the lines thoroughly before you use the equipment.

Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.

This equipment should be used only by the legally qualified persons and practitioners.

Regarding the AC cable and software, only the ones developed and supplied by Ecotron Co., Ltd should be
used.

Only trained maintenance staff may remove the covers of the x-ray system.

Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask the
help of Ecotron or its authorized dealer for the service.

Information for repair will be provided by Ecotron. (Circuit diagram, Component part list, descriptions,
Calibration instructions, or other information)

No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties which
many results from the use of this manual.

Always keep the manual at hand for your reference.

NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following rules
and methods described in this manual.

CAUTION

THE IONIZING RADIATION IS DANGEROUS FOR THE OPERATOR IF THE FOLLOWING


SAFETY MEASURES ARE NOT STRICTLY OBSERVED.

NOTE

ASSEMBLY INFORMATION: NO ASSEMBLY IS REQUIRED.

Rev. F 7/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1. CALIBRATION

1.1 EPX-F1600

1.1.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.1.2 CALIBRATION PROCEDURE

1.1.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 KV and 5 mAs (30mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 90 KV and 5 mAs (16 mA), respectively.
k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 4.5 V, adjust KV by turning VR3 on the OP board.

Rev. F 8/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.1.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 kV and 5 mAs (30 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C09 are calibration code of tube current at AC110V model and C10~C19 are calibration code of
tube current at AC220V model.

C00 or C10 is calibration code of tube current (30mA at 40kV~50kV) at AC110V or AC220V and 30.0 is
calibration data of tube current (30mA at 40kV~50kV).

Rev. F 9/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec (see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C11 is calibration code of tube current (25mA at 40kV~50kV) and 25.0 is calibration data of
tube current (25mA at 40kV~50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec (see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 10/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Where, P00~P07 are calibration code of filament preheat current at AC110V, P08~P15 are calibration code of
filament preheat current at AC220V, P00 or P08 is Filament preheat calibration code (Preheat calibration code
of 30mA), and 1.50 or 1.40 is Filament preheat calibration data (Preheat calibration data of 30mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P09 is Filament preheat calibration code (Preheat calibration code of 25mA) and 1.45 or 1.35
is Filament preheat calibration data (Preheat calibration data of 25mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 11/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.2 EPX-F2400

1.2.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.2.2 CALIBRATION PROCEDURE

1.2.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 KV and 5 mAs (40mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100 KV and 5 mAs (20 mA), respectively.
k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5 V, adjust KV by turning VR3 on the OP board.

Rev. F 12/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.2.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 kV and 5 mAs (40 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of
tube current at AC220V model.
C00 or C08 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0 is
calibration data of tube current (40mA at 40kV~60kV).

Rev. F 13/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec (see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (35mA at 40kV~60kV) and 35.0 is calibration data of
tube current (35mA at 40kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec (see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.25V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 14/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Where, P00~P04 are calibration code of filament preheat current at AC110V, P05~P09 are calibration code of
filament preheat current at AC220V, P00 or P05 is Filament preheat calibration code (Preheat calibration code
of 40mA), and 0.70 or 0.72 is Filament preheat calibration data (Preheat calibration data of 40mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec (see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P06 is Filament preheat calibration code (Preheat calibration code of 35mA) and 0.67 or 0.69
is Filament preheat calibration data (Preheat calibration data of 35mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec (see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 15/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.3 EPX-F2800

1.3.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.3.2 CALIBRATION PROCEDURE

1.3.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs (40mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120 KV and 5 mAs (16 mA), respectively.
k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6 V, adjust KV by turning VR3 on the OP board.

Rev. F 16/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.3.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 kV and 5 mAs (40 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C15 are calibration code of tube current at AC110V model and C16~C31 are calibration code of
tube current at AC220V model.
C00 or C16 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0 is
calibration data of tube current (40mA at 40kV~60kV).

Rev. F 17/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec (see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C17 is calibration code of tube current (30mA at 61kV~70kV) and 30.0 is calibration data of
tube current (30mA at 61kV~70kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec (see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.15V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 18/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

where, P00~P06 are calibration code of filament preheat current at AC110V, P07~P13 are calibration code of
filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat calibration code
of 35mA), and 0.70 or 0.70 is Filament preheat calibration data (Preheat calibration data of 35mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA) and 0.67 or 0.67
is Filament preheat calibration data (Preheat calibration data of 30mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 19/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.4 EPX-F3200

1.4.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.4.2 CALIBRATION PROCEDURE

1.4.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs (60mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(30mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5V, adjust KV by turning VR3 on the OP board.

Rev. F 20/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.4.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40 kV and 5 mAs (60 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of
tube current at AC220V model. C00 or C08 is calibration code of tube current (60mA at 40kV~50kV) at
AC110V and 60.0 is calibration data of tube current (60mA at 40kV~50kV).

Rev. F 21/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec (see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6.0V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (50mA at 51kV~60kV) and 50.0 is calibration data of
tube current (50mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec (see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.0V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 22/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Where, P00~P05 are calibration code of filament preheat current at AC110V, P06~P11 are calibration code of
filament preheat current at AC220V, P00 or P06 is Filament preheat calibration code (Preheat calibration code
of 60mA), and 1.65 or 1.65 is Filament preheat calibration data (Preheat calibration data of 60mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P07 is Filament preheat calibration code (Preheat calibration code of 50mA) and 1.60 or 1.60
is Filament preheat calibration data (Preheat calibration data of 50mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 23/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.5 EPX-F4000

1.5.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.5.2 CALIBRATION PROCEDURE

1.5.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs (80mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2.5V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs (25mA), respectively.
k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

Rev. F 24/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.5.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 50 kV and 5 mAs (80 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.44V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C14 are calibration code of tube current at AC110V model and C15~C29 are calibration code
of tube current at AC220V model. C00 or C15 is calibration code of tube current (80mA at 50kV) at 80.0 is
calibration data of tube current (80mA at 50kV).

Rev. F 25/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 4V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C16 is calibration code of tube current (70mA at 50kV) and 70.0 is calibration data of tube
current (70mA at 50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec (see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 3.0V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 26/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Where, P00~P09 are calibration code of filament preheat current at AC110V, P10~P19 are calibration code of
filament preheat current at AC220V, P00 or P10 is Filament preheat calibration code (Preheat calibration code
of 80mA), and 2.50 or 2.50 is Filament preheat calibration data (Preheat calibration data of 80mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec (see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P11 is Filament preheat calibration code (Preheat calibration code of 70mA) and 2.20 or 2.20
is Filament preheat calibration data (Preheat calibration data of 70mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 27/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.6 EPX-F5000

1.6.1 LINE POWER SWITCH

- Line switch should be at off position.


- Connect power line cord to the power outlet.
- Turn on the line switch.
- The FND on the control panel should illuminate.

1.6.2 CALIBRATION PROCEDURE

1.6.2.1 TUBE VOLATAGE CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 50KV and 5mAs (100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2.0V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs (40mA), respectively.
k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

Rev. F 28/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

1.6.2.2 TUBE CURRENT CONFIRMATION AND CALIBRATION

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 50 kV and 5 mAs (100 mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C20 are calibration code of tube current at AC110V model and C21~C41 are calibration code of
tube current at AC220V model. C00 or C21 is calibration code of tube current (100mA at 50kV) at A0.0(=100)
is calibration data of tube current (100mA at 50kV).

Rev. F 29/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec (see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C22 is calibration code of tube current (80mA at 50kV) and 80.0 is calibration data of tube
current (80mA at 50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

Filament preheat calibration

This is preheating filament for making tube current. The overshooting or the undershooting may occur from
the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in
accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

Rev. F 30/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Where, P00~P08 are calibration code of filament preheat current at AC110V, P09~P15 are calibration code of
filament preheat current at AC220V, P00 or P09 is Filament preheat calibration code (Preheat calibration code
of 100mA), and 2.40 or 2.40 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.25 or 2.25
is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec (see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

Rev. F 31/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

2. DIP SWITCH SETTING

DIP Switches are used for following setting.

SW #1: ERROR Disable (ON) / Enable (OFF)


SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to OFF position.

Rev. F 32/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

3. HARD-WIRED HAND SWITCH CONTROL

3.1 HAND SWITCH INTRODUCTION

The Exposure Hand Switch consists of a two step hand switch.


The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage.
The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.

3.2 EXPOSURE PROCESS

There are two ways to use the hand switch to get the exposure.

a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or Ready” status on, then
press Switch #2 to get the “Exposure”.

b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second. This action will result
in “Prep or Ready” status and “Exposure” at the same time.

Rev. F 33/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

4. COLLIMATOR

a. Halogen Lamp * 1
b. Laser pointer * 2
c. Horizontal and vertical lighting range control

I. Block Diagram

II. Layout
a. External Layout

Rev. F 34/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

b. Internal Layout

c. Knob Layout

Rev. F 35/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5. TROUBLE SHOOTING

If you have any error message when you are using portable x-ray, please check below service check list

5.1 EPX-F1600 / F2400 / F2800 / F3200 / F4000 / F5000 ERROR CODE TABLE

Error
DESCRIPTION INDICATION
Code

2 THERMAL ERROR (INTERLOCK) Excess from 60℃

7 READY SW ERR Ready Switch Malfunction

8 EXPOSURE SW ERROR Exposure Switch Malfunction

10 STANDBY FIL FB ERROR Excess from 200mA (10% of Max. Current, 2A)

11 STANDBY EP FB ERROR Excess from 12kV (10% of Max. Voltage, 120kV)

12 STANDBY IP FB ERROR Excess from 3.5mA (10% of Max. Current, 35mA)

14 RUN FIL FB ERROR Out of Range from Reference ±10%

RUN EP FB or AC POWER LOW Out of Range from Reference -10% or AC Power Line
15
ERROR poor

16 RUN IP FB HIGH ERROR Out of Range from Reference +10%

17 RUN EP FB HIGH ERROR Out of Range from Reference +10%

18 RUN IP FB LOW ERROR Out of Range from Reference -10%

28 PULSE BY PULSE ERROR FET Board Fault (Hardware)

29 FET-OTP ERROR FET board overheat

30 KV_OVP ERROR kV Over Voltage (Hardware)

31 FIL_OC ERROR FIL Over Current (Hardware)

32 IP_OCP ERROR mA Over Current (Hardware)

Rev. F 36/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5.2 ACTION TO CORRECT THE ERROR OF EPX-SERIES

Error Codes Check points and Action

7 Ready SW ERR
a. Replace the hand switch
8 Exposure SW ERR

10 Standby Fil FB Error


a. If error repeats at the condition of 40kV 0.4mAs EXP, replace the OP
11 Standby EP FB Error
board

12 Standby IP FB Error

14 Run Fil FB Error


a. If error repeats at the condition of 40kV 0.4mAs EXP, check the followings
15 Run EP FB low ERR
by sequence.
1. Check TP in the power input of the OP board
16 Run IP FB High ERR
2. Check the Reference voltage value
17 Run EP FB High ERR 3. If the reference voltage is normal, replace the FET board
4. If replacing FET board does not solve the problem, the replace the
18 Run IP FB low ERR
HV tank.

28 Pulse by Pulse ERR

a. If error pops up right after the unit power is turned on, then replace the
30 KV_OVP ERR
OP board.

31 FIL_OC ERR FIL b. If error repeats at the condition of 40kV 0.4mAs EXP,
1. Replace the FET board
2. If replacing FET board does not solve the problem, the replace the
32 IP_OCP ERR
HV tank.

Warning Codes
Thermal ERR excess Wait until the temperature inside the monotank comes down to normal
2
from 60°C (Once the temperature drop to normal, the Warning Message will disappear)
FET-OTP FET board Wait until the temperature inside the FET board comes down to normal
29
overheat (Once the temperature drop to normal, the Warning Message will disappear)

Rev. F 37/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6. CALIBRATION

6.1 EPX-F1600

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.1.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 38/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.1.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 5.1V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 39/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.06V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 2.72V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 40/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.1.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 41/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.1.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 42/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.1.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C09 (AC110V), C10~C19 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 43/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current Parameter Tube Voltage mAs Tube Current INPUT DATA

AC110V AC220V KV MIN MAX mA mA

C00 C10 40~50 (1) 0.4 20 30.0 30.0

C01 C11 40~50 (2) 25 32 25.0 25.0

C02 C12 51~60 (1) 0.4 20 25.0 25.0

C03 C13 51~60 (2) 25 32 20.0 20.0

C04 C14 61~70 (1) 0.4 20 20.0 20.0

C05 C15 61~70 (2) 25 32 15.0 15.0

C06 C16 71~80 (1) 0.4 20 18 18

C07 C17 71~80 (2) 25 32 14 14

C08 C18 81~90 (1) 0.4 20 16 16

C09 C19 81~90 (2) 25 32 12 12

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.1.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

Rev. F 44/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P07 (AC110V), P08~P15 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode (110V) P_Mode (220V) P_DATA

30 P00 P08

25 P01 P09

20 P02 P10

18 P03 P11
1.6kW
16 P04 P12

15 P05 P13

14 P06 P14

12 P07 P15

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 45/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.2 EPX-F2400

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.2.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 46/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.2.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 47/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 100kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 100 kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 48/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.2.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 49/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.2.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 50/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.2.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C07 (AC110V), C08~C15 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 51/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current Parameter Tube Voltage mAs Tube Current INPUT DATA

AC110V AC220V KV MIN MAX mA mA

C00 C08 40~60(1) 0.4 64 40.0 40.0

C01 C09 40~60(2) 80 100 35.0 30.0

C02 C10 61~70(1) 0.4 100 30.0 25.0

C03 C11 71~80(1) 0.4 32 30.0 30.0

C04 C12 71~80(2) 40 80 25.0 25.0

C05 C13 81~90(1) 0.4 40 25.0 25.0

C06 C14 81~90(2) 50 80 20.0 20.0

C07 C15 91~100(1) 0.4 80 20.0 20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.2.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON )/ Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

Rev. F 52/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P04 (AC110V), P05~P09 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode (110V) P_Mode (220V) P_DATA

40 P00 P05

35 P01 P06

30 P02 P07

25 P03 P08
2.4 kW
20 P04 P09

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 53/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.3 EPX-F2800

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.3.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 54/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.3.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 55/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 5.25V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 120kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 120kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 120 kV 6V 2.4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 56/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.3.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 57/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.3.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 58/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.3.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C15 (AC110V), C16~C31 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 59/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current Parameter Tube Voltage mAs Tube Current INPUT DATA

AC110V AC220V KV MIN MAX mA mA

C00 C16 40~60 0.4 100 40.0 40.0

C01 C17 61~70(1) 0.4 64 30.0 30.0

C02 C18 61~70(2) 80 100 25.0 25.0

C03 C19 71~80(1) 0.4 10 35.0 35.0

C04 C20 71~80(2) 13 32 30.0 30.0

C05 C21 71~80(3) 40 80 25.0 25.0

C06 C22 81~90(1) 0.4 10 30.0 30.0

C07 C23 81~90(2) 13 40 25.0 25.0

C08 C24 81~90(3) 50 80 20.0 20.0

C09 C25 91~100(1) 0.4 13 25.0 25.0

C10 C26 91~100(2) 16 50 20.0 20.0

C11 C27 91~100(3) 64 80 16.0 16.0

C12 C28 101~110(1) 0.4 32 18.0 18.0

C13 C29 101~110(2) 40 64 16.0 16.0

C14 C30 111~120(1) 0.4 20 16.0 16.0

C15 C31 111~120(2) 25 50 14.0 14.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.3.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

Rev. F 60/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON )/ Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P06 (AC110V), P07~P13 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode (110V) P_Mode (220V) P_DATA

40 P00 P08

35 P01 P09

30 P02 P10

25 P03 P11
2.8 kW
20 P04 P12

18 P05 P13

16 P06 P14

14 P07 P15

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 61/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4 EPX-F3200

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.4.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 62/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 63/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 100kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 100 kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 64/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 65/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 66/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.4.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C07 (AC110V), C08~C15 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 67/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current Parameter Tube Voltage mAs Tube Current INPUT DATA

AC110V AC220V KV MIN MAX mA mA

C00 C08 40~50 0.4 100 60.0 60.0

C01 C09 51~60 0.4 80 50.0 50.0

C02 C10 61~70 0.4 100 40.0 40.0

C03 C11 71~80(1) 0.4 32 40.0 40.0

C04 C12 71~80(2) 40 80 35.0 35.0

C05 C13 81~90 0.4 80 30.0 30.0

C06 C14 91~100(1) 0.4 50 30.0 30.0

C07 C15 91~100(2) 64 80 25.0 25.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.4.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON )/ Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

Rev. F 68/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P05 (AC110V), P06~P11 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode (110V) P_Mode (220V) P_DATA

60 P00 P06

50 P01 P07

40 P02 P08

35 P03 P09
3.2 kW
30 P04 P10

25 P05 P11

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 69/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5 EPX-F4000

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.5.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 70/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 5.44V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 71/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 110kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 110kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 110 kV 5.5V 2.04V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 72/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 73/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 74/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C14 (AC110V), C15~C29 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 75/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter

AC 110V AC 220V KV MIN MAX mA mA

C00 C15 50 (1) 0.4 16 80.0 80.0

C01 C16 50 (2) 20 32 70.0 70.0

C02 C17 50 (3) 40 100 60.0 60.0

C03 C18 51~60 (1) 0.4 32 60.0 60.0

C04 C19 51~60 (2) 40 100 50.0 50.0

C05 C20 61~70 (1) 0.4 32 50.0 50.0

C06 C21 61~70 (2) 40 100 40.0 40.0

C07 C22 71~80 (1) 0.4 10 45.0 45.0

C08 C23 71~80 (2) 13 80 40.0 40.0

C09 C24 81~90 (1) 0.4 32 35.0 35.0

C10 C25 81~90 (2) 40 80 30.0 30.0

C11 C26 91~100 (1) 0.4 32 30.0 30.0

C12 C27 91~100 (2) 40 80 25.0 25.0

C13 C28 101~110 (1) 0.4 32 25.0 25.0

C14 C29 101~110 (2) 40 80 20.0 20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

Rev. F 76/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.5.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON )/ Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P09 (AC110V), P10~P19 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Rev. F 77/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA

80 P00 P10

70 P01 P11

60 P02 P12

50 P03 P13

45 P04 P14
4.0KW
40 P05 P15

35 P06 P16

30 P07 P17

25 P08 P18

20 P09 P19

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 78/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6 EPX-F5000

Please fill in the blank.

Kv / mAs condition
Product S/N Error code
(When error is occurred)

※ Open upper cover of portable x-ray below picture.

6.6.1 VOLTAGE CHECK

1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V

Pass Fail Pass Fail Pass Fail Pass Fail


Pass/Fail

Measured Voltage
If it is fail case

Rev. F 79/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6.2 kV, mA WAVEFORM CHECK

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 5.1V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

Rev. F 80/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

5) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.06V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 110kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 110kV/5mAs
CH 1: kV 2V/div CH 2: mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 110 kV 5.5V 1.53V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

Rev. F 81/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6.3 WAVEFORM OF POOR AC POWER CONDITION

If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

Rev. F 82/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6.4 kV, mA PWM FREQUENCY CHECK

1) Turn off AC swtich


2) Please connect probe to 80 kHz (TP15), 140 kHz (TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM

Frequency 80kHz 140kHz

Pass/Fail Pass Fail Pass Fail

Measured frequency

Rev. F 83/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6.5 mA CALIBRATION DATA CHECK

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON) / Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

1) Turn on AC switch.

Up/Down switches increase and decrease C00~C20 (AC110V), C21~C41 (AC220V).


Please check Input data.
Input data and Tube current should be the same with below table.

Rev. F 84/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Tube Current Parameter Tube Voltage mAs Tube Current INPUT DATA

AC110V AC220V KV MIN MAX mA mA

C00 C21 50(1) 0.4 10 100 100

C01 C22 50(2) 13 32 80 80

C02 C23 50(3) 40 100 70 70

C03 C24 51~60(1) 0.4 10 80 80

C04 C25 51~60(2) 13 32 70 70

C05 C26 51~60(3) 40 100 60 60

C06 C27 61~70(1) 0.4 10 70 70

C07 C28 61~70(2) 13 32 60 60

C08 C29 61~70(3) 40 100 50 50

C09 C30 71~80(1) 0.4 10 60 60

C10 C31 71~80(2) 13 32 50 50

C11 C32 71~80(3) 40 100 40 40

C12 C33 81~90(1) 0.4 10 50 50

C13 C34 81~90(2) 13 32 40 40

C14 C35 81~90(3) 40 100 30 30

C15 C36 91~100(1) 0.4 10 45 45

C16 C37 91~100(2) 13 32 40 40

C17 C38 91~100(3) 40 100 30 30

C18 C39 101~110(1) 0.4 10 40 40

C19 C40 101~110 (2) 13 32 30 30

C20 C41 101~110 (3) 40 100 20 20

4) If mA calibration is done, Please dip switch 4 set to OFF position.

Rev. F 85/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

6.6.6 FILAMENT PREHEAT CALIBRATION CODE AND DATA CHECK

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting:


SW #1: ERROR Enable (ON )/ Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON) / Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

Up/Down switches increase and decrease P00~P08 (AC110V), P09~P17 (AC220V)


Please check Input data.
Input data and Tube current should be the same.

Rev. F 86/87 ECOTRON Co., Ltd.


EPX-SERIES SERVICE MANUAL

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)

100 P00 P09

80 P01 P10

70 P02 P11

60 P03 P12

5kW 50 P04 P13

45 P05 P14

40 P06 P15

30 P07 P16

20 P08 P17

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

Rev. F 87/87 ECOTRON Co., Ltd.