Documente Academic
Documente Profesional
Documente Cultură
CERTIFICATE OF ANALYSIS
ERM®- FA013bk
Steel Charpy V-notch test pieces
Certified value 3) Uncertainty 4)
[J] [J]
1) The absorbed energy (KV) is a method defined measurand. KV is the impact energy required to break a V-notched bar
of standardised dimensions, as defined in ISO 148-1. The certified value is valid only for strikers with a 2 mm tip radius, and
in the temperature range of (20 ± 2) °C.
2) The absorbed energy (KV) is a method defined measurand. KV is the impact energy required to break a V-notched bar
of standardised dimensions, as defined in ISO 148-1. The certified value is valid only for strikers with a 2 mm tip radius, and
in the temperature range of (0 ± 2) °C.
®
3) The certified values of ERM -FA013bk, and the associated uncertainties, are traceable to the International System
of Units (SI), via the master batches ERM®-FA013ba and ERM®-FA013ay respectively, of the same nominal absorbed
® ® ®
energy (30 J) by testing samples of ERM -FA013ba and ERM -FA013bk at 20 °C and samples of ERM -FA013ay and
®
ERM -FA013bk at 0 °C, under repeatability conditions on an impact pendulum verified and calibrated with SI-traceable
calibrated tools.
4) Estimated expanded uncertainty of the mean KV of the 5 specimens (delivered as 1 set), with a coverage factor k = 2,
corresponding to a level of confidence of about 95 %, as defined in ISO/IEC Guide 98-3, Guide to the expression of
uncertainty in measurement (GUM:1995). The number of degrees of freedom of the certified uncertainty is RM = 70 (20 °C)
and 44 (0 °C).
NOTE
European Reference Material ERM®-FA013bk was produced and certified under the responsibility of the IRMM
according to the principles laid down in the technical guidelines of the European Reference Materials® co-
operation agreement between BAM-IRMM-LGC. Information on these guidelines is available on the internet
(http://www.erm-crm.org).
Accepted as an ERM®, Geel, March 2013
Signed:
PARTICIPANTS
- Aubert&Duval, Les Ancizes and Gennevillers (FR), ISO 9001 certification by AFNOR No.
1993/1378h
- European Commission, Joint Research Centre, Institute for Reference Materials and Measurements
(IRMM), Geel (BE), ISO/IEC 17025 accreditation by BELAC No. 268-TEST and ISO Guide 34
accreditation by BELAC No. 268-RM for the production of reference materials
- Laboratoire National de Métrologie et d’Essais, Trappes (FR), ISO/IEC 17025 accreditation by
COFRAC No. 2.1287.
SAFETY INFORMATION
Precautions need to be taken to avoid injury of the operator by broken specimens when operating the
Charpy impact pendulum.
Page 2 of 3
4a. Before testing at 20 °C, bring the specimens to the test temperature (20 ± 2) °C. To assure thermal
equilibrium is reached, move the specimens to the test laboratory at least 3 h before the tests.
4b. Before testing at 0 °C, bring the specimens to the test temperature (0 ± 2 °C). To assure thermal
equilibrium, the specimens shall be at least 10 minutes in a liquid cooling medium or at least 30
minutes in a gaseous cooling medium, the temperature of which is measured and monitored. The
test piece shall be broken within 5 seconds of the time of removal from the cooling medium.
After cleaning and equilibration, the 5 samples constituting a CRM-unit need to be broken with a pendulum
impact test machine operated in accordance with ISO 148-1 standards. Prior to the tests, the anvils must be
cleaned. It must be noted that Charpy test pieces sometimes leave debris on the Charpy pendulum anvils.
Therefore, the anvils must be checked regularly and if debris is found, it must be removed.
The comparison of the indirect verification results with the certified value and uncertainty must be based on
the mean of the 5 measured KV values, because this is the sample size for which the uncertainty of the
certified value has been calculated. For some pendulums and for some samples, post-fracture interaction
between broken samples and pendulum hammer can affect the measured KV values. The resulting
excessively high values are easily related to indentations and deformations of the broken samples. Users of
ERM-FA013bk reference materials encountering outlier values that can be related to post-fracture
indentation marks on the broken samples are requested to eliminate the corresponding data from the
analysis of their results. The corresponding expanded uncertainty for tests at 20 °C for n = 4 remaining
results is U = 0.9 J (RM = 70), for n = 3 remaining results U = 1.0 J (RM = 67). The corresponding expanded
uncertainty for tests at 0 °C for n = 4 remaining results is U = 1.1 J (RM = 47), for n = 3 remaining results U
= 1.2 J (RM = 50).
STORAGE
Specimens should be kept at room temperature (18 ± 5) °C in their original packing until used. The European
Commission cannot be held responsible for changes that happen during storage of the material at the
customer’s premises, especially of opened samples.
LEGAL NOTICE
Neither IRMM, its contractors nor any person acting on their behalf:
(a) make any warranty or representation, express or implied, that the use of any information, material, apparatus,
method or process disclosed in this document does not infringe any privately owned intellectual property rights;
or
(b) assume any liability with respect to, or for damages resulting from, the use of any information, material,
apparatus, method or process disclosed in this document, save for loss or damage arising solely and directly
from the negligence of IRMM.
NOTE
A detailed technical report is available on www.irmm.jrc.be. A paper copy can be obtained from the Joint
Research Centre, Institute for Reference Materials and Measurements on request.
European Commission – Joint Research Centre
Institute for Reference Materials and Measurements (IRMM)
Retieseweg 111, B - 2440 Geel (Belgium)
Telephone: +32-(0)14-571.722 - Fax: +32-(0)14-590.406
Page 3 of 3