Documente Academic
Documente Profesional
Documente Cultură
ANESHTESIOLOGY
2016
VSMU
Content
Standards for basic anesthetic monitoring.........................................................................................2
Noninvasive arterial blood pressure monitoring................................................................................3
Invasive pressure monitoring.............................................................................................................4
Electronic transducer.....................................................................................................................5
Factors affecting errors in invasive monitoring..............................................................................6
Arterial pressure. Arterial pressure waveform...............................................................................6
Arterial line placement...................................................................................................................8
Electrocardiography...........................................................................................................................9
Central venous catheterization........................................................................................................11
Techniques & Complications........................................................................................................11
Capnography....................................................................................................................................13
Indications & Contraindications...................................................................................................13
Techniques & Complications........................................................................................................13
Temperature....................................................................................................................................14
Urinary output..................................................................................................................................15
Peripheral nerve stimulation............................................................................................................16
Clearly, one of the most important roles is serving as the ever-vigilant set of “eyes” to
ensure patient safety. Despite the introduction of new methods to monitor the anesthetized
patient, hemodynamic monitoring and the analysis of these well-known parameters remain
central to assessing patient wellbeing.
Electronic transducer
The cannula is connected to high-pressure tubing that is filled with saline. This acts as
a continuous column of fluid that transmits intraluminal pressure changes to the transducer
diaphragm that oscillates in response to the pressure waveform. The movement is converted
to an electrical signal by a transducer. The transducer accomplishes this by acting as part of a
capacitor, inductor, or, most commonly, a strain gauge. The strain gauge uses variable
resistors, the electrical resistance of which increases with increasing length. The diaphragm
of the transducer moves a small plate connected to four strain gauges. With any one
movement, two gauges are compressed and the other two stretched. All four strain gauges
form part of a Wheatstone bridge, increasing the sensitivity.
Calibration
Calibration is the process of validating a measurement technique or equipment. It
was an important consideration for older electronic transducers. The standard calibration is
a 50-microvolt change in potential per 10 mm Hg. Modern systems typically do not require
external calibration because they are manufactured in a standardized manner to tight
standards.
Zeroing
Zeroing removes the effect of atmospheric pressure on transmural pressure of the
tubing system by exposing the transducer system to ambient atmospheric pressure. Zeroing
is performed by opening the transducer to atmospheric pressure and electronically setting
this atmospheric pressure to zero. Occasionally, this zero baseline may drift and should be
checked frequently, because even a 5 mm Hg drift can represent a significant difference in
low-pressure systems such as CVP.
Leveling
Leveling refers to the placement of an already zeroed transducer at a particular
height where the pressures are sought to be measured (a reference point). The electrical
transducer needs to be aligned to the superior aspect of the right atrium to measure the
pressure at the level of the heart. This eliminates errors of measurement that might occur
from the hydrostatic pressure exerted by the column of blood above or below the point of
reference to be measured. A 10 cm change in height will increase or decrease the pressure
reading by 7.5 mm Hg.
2. Leveling error
Figure 2. T The systolic components following the R wave on ECG consist of (1) steep pressure upstroke, peak; (2) systolic
peak pressure; and (3) decline, which correspond to the period of left ventricular systole. The down-slope is interrupted by
the (4) dicrotic notch, which reflects aortic valve closure at end systole. The remaining decay of waveform, (5) diastolic
runoff, occurs during diastole following the ECG T wave and reaches its nadir at end diastole, (6) end-diastolic pressure.
(Modified from Mark JB.Atlas of Cardiovascular Monitoring.)
Figure 3. Systolic (S) and diastolic (D) pressures are shown with arrows. Mean arterial pressure (MAP) is represented by the
area beneath the arterial pressure curve divided by the beat period, and it incorporates the S and D portions of the cardiac
cycle. (Modified from Mark JB.Atlas of Cardiovascular Monitoring.)
Electrocardiography
All patients should have intraoperative monitoring of their electrocardiogram (ECG).
There are no contraindications.
Lead selection determines the diagnostic sensitivity of the ECG. ECG leads are
positioned on the chest and extremities to provide different perspectives of the electrical
potentials generated by the heart. At the end of diastole, the atria contract, which provides
the atrial contribution to CO, generating the “P” wave. Following atrial contraction, the
ventricle is loaded awaiting systole. The QRS complex begins the electrical activity of systole
following the 120–200 msec atrioventricular (AV) nodal delay. Depolarization of the ventricle
proceeds from the AV node through the interventricular system via the His–Purkinje fibers.
The normal QRS lasts approximately 120 msec, which can be prolonged in patients with
cardiomyopathies and heart failure. The T wave represents repolarization as the heart
prepares to contract again. Prolongation of the QT interval secondary to electrolyte
imbalances or drug effects can potentially lead to life-threatening arrhythmias (les torsade
de pointes). The electrical axis of lead II is approximately 60° from the right arm to the left
leg, which is parallel to the electrical axis of the atria, resulting in the largest P-wave voltages
of any surface lead. This orientation enhances the diagnosis of arrhythmias and the
detection of inferior wall ischemia. Lead V5 lies over the fifth intercostal space at the
anterior axillary line; this position is a good compromise for detecting anterior and lateral
wall ischemia. A true V5 lead is possible only on operating room ECGs with at least five lead
wires, but a modified V5 can be monitored by rearranging the standard three-limb lead
placement. Ideally, because each lead provides unique information, leads II and V5 should be
monitored simultaneously. If only a single-channel machine is available, the preferred lead
for monitoring depends on the location of any prior infarction or ischemia. Esophageal leads
are even better than lead II for arrhythmia diagnosis, but have not yet gained general
acceptance in the operating room. Electrodes are placed on the patient’s body to monitor
the ECG. Conductive gel lowers the skin’s electrical resistance, which can be further
decreased by cleansing the site with alcohol. Needle electrodes are used only if the disks are
unsuitable (eg, with an extensively burned patient).
Figure 5. Rearranged three-limb lead placement. Anterior and lateral ischemia can be detected by placing the left arm lead
(LA) at the V5 position. When lead I is selected on the monitor, a modified V5 lead (CS5) is displayed. Lead II allows
detection of arrhythmias and inferior wall ischemia. RA, right arm; LL, left leg
The ECG is a recording of the electrical potentials generated by myocardial cells. Its
routine use allows arrhythmias, myocardial ischemia, conduction abnormalities, pacemaker
malfunction, and electrolyte disturbances to be detected. Because of the small voltage
potentials being measured, artifacts remain a major problem. Patient or lead-wire
movement, use of electrocautery, 60-cycle interference from nearby alternating current
devices, and faulty electrodes can simulate arrhythmias. Monitoring filters incorporated into
the amplifier to reduce “motion” artifacts will lead to distortion of the ST segment and may
impede the diagnosis of ischemia. Digital readouts of the heart rate (HR) may be misleading
because of monitor misinterpretation of artifacts or large T waves—often seen in pediatric
patients—as QRS complexes. Depending on equipment availability, a preinduction rhythm
strip can be printed or frozen on the monitor’s screen to compare with intraoperative
tracings. To interpret ST-segment changes properly, the ECG must be standardized so that a
1-mV signal results in a deflection of 10 mm on a standard strip monitor. Newer units
continuously analyze ST segments for early detection of myocardial ischemia. Automated ST-
segment analysis increases the sensitivity of ischemia detection, does not require additional
physician skill or vigilance, and may help diagnose intraoperative myocardial ischemia.
Commonly accepted criteria for diagnosing myocardial ischemia require that the ECG be
recorded in “diagnostic mode” and include a flat or down sloping ST-segment depression
exceeding 1 mm, 80 msec after the J point (the end of the QRS complex), particularly in
conjunction with T-wave inversion. ST-segment elevation with peaked T waves can also
represent ischemia. Wolff–Parkinson–White syndrome, bundle-branch blocks, extrinsic
pacemaker capture, and digoxin therapy may preclude the use of ST-segment information.
The audible beep associated with each QRS complex should be loud enough to detect rate
and rhythm changes when the anesthesiologist’s visual attention is directed elsewhere.
Some ECGs are capable of storing aberrant QRS complexes for further analysis, and some
can even interpret and diagnose arrhythmias. The interference caused by electrocautery
units, however, has limited the usefulness of automated arrhythmia analysis in the operating
room.
Capnography
Indications & Contraindications.
Determination of end-tidal CO2 (EtCO2) concentration to confirm adequate
ventilation is mandatory during all anesthetic procedures, but particularly so for general
anesthesia. A rapid fall of EtCO2 is a sensitive indicator of air embolism, a major complication
of sitting craniotomies. There are no contraindications.
Figure 7. A nondiverting sensor placed in-line analyzes CO2 concentration at the sampling site.
Temperature
Indications
The temperature of patients undergoing anesthesia must be monitored.
Postoperative temperature is increasingly used as a quality anesthesia indicator.
Hypothermia is associated with delayed drug metabolism, increased blood glucose,
vasoconstriction, impaired coagulation, and impaired resistance to surgical infections.
Hyperthermia can likewise have deleterious effects perioperatively, leading to tachycardia,
vasodilation, and neurological injury. Consequently, temperature must be measured and
recorded perioperatively.
Contraindications
There are no contraindications, although a particular monitoring site may be
unsuitable in certain patients.
Techniques & Complications
Intraoperatively, temperature is usually measured using a thermistor or
thermocouple. Thermistors are semiconductors whose resistance decreases predictably with
warming. A thermocouple is a circuit of two dissimilar metals joined so that a potential
difference is generated when the metals are at different temperatures. Disposable
thermocouple and thermistor probes are available for monitoring the temperature of the
tympanic membrane, nasopharynx, esophagus, bladder, rectum, and skin. Infrared sensors
estimate temperature from the infrared energy that is produced. Tympanic membrane
temperatures reflect core body temperature; however, the devices used may not reliably
measure the temperature at the tympanic membrane. Complications of temperature
monitoring are usually related to trauma caused by the probe (eg, rectal or tympanic
membrane perforation). Each monitoring site has advantages and disadvantages. The
tympanic membrane theoretically reflects brain temperature because the auditory canal’s
blood supply is the external carotid artery. Trauma during insertion and cerumen insulation
detract from the routine use of tympanic probes. Rectal temperatures have a slow response
to changes in core temperature. Nasopharyngeal probes are prone to cause epistaxis, but
accurately measure core temperature if placed adjacent to the nasopharyngeal mucosa. The
thermistor on a pulmonary artery catheter also measures core temperature. There is a
variable correlation between axillary temperature and core temperature, depending on skin
perfusion. Liquid crystal adhesive strips placed on the skin are inadequate indicators of core
body temperature during surgery. Esophageal temperature sensors, often incorporated into
esophageal stethoscopes, provide the best combination of economy, performance, and
safety. To avoid measuring the temperature of tracheal gases, the temperature sensor
should be positioned behind the heart in the lower third of the esophagus. Conveniently,
heart sounds are most prominent at this location.
Urinary output
Indications
Urinary bladder catheterization is the only reliable method of monitoring urinary
output. Insertion of a urinary catheter is indicated in patients with congestive heart failure,
renal failure, advanced hepatic disease, or shock. Catheterization is routine in some surgical
procedures such as cardiac surgery, aortic or renal vascular surgery, craniotomy, major
abdominal surgery, or procedures in which large fluid shifts are expected. Lengthy surgeries
and intraoperative diuretic administration are other possible indications. Occasionally,
postoperative bladder catheterization is indicated in patients having difficulty voiding in the
recovery room after general or regional anesthesia.
Contraindications
Bladder catheterization should be done with utmost care in patients at high risk for
infection.
Techniques & Complications
Bladder catheterization is usually performed by surgical or nursing personnel. To
avoid unnecessary trauma, a urologist should catheterize patients suspected of having
abnormal urethral anatomy. A soft rubber Foley catheter is inserted into the bladder
transurethrally and connected to a disposable calibrated collection chamber. To avoid urine
reflux and minimize the risk of infection, the chamber should remain at a level below the
bladder. Complications of catheterization include urethral trauma and urinary tract
infections. Rapid decompression of a distended bladder can cause hypotension. Suprapubic
catheterization of the bladder with tubing inserted through a large-bore needle is an
uncommon alternative.
Clinical Considerations
An additional advantage of placing a Foley catheter is the ability to include a
thermistor in the catheter tip so that bladder temperature can be monitored. As long as
urinary output is high, bladder temperature accurately reflects core temperature. An added
value with more widespread use of urometers is the ability to electronically monitor and
record urinary output and temperature. Urinary output is a reflection of kidney perfusion
and function and an indicator of renal, cardiovascular, and fluid volume status. Inadequate
urinary output (oliguria) is often arbitrarily defined as urinary output of less than 0.5
mL/kg/hr, but actually is a function of the patient’s concentrating ability and osmotic load.
Urine electrolyte composition, osmolality, and specific gravity aid in the differential diagnosis
of oliguria.