Documente Academic
Documente Profesional
Documente Cultură
In the present study, surface melting of a magnesium alloy, ZE41, was performed with an
Nd:YAG laser using different laser parameters. The microstructure of the laser-treated and
untreated specimens was analyzed by optical and scanning electron microscopy and X-ray
diffraction. Corrosion resistance of the different laser-treated specimens along with the
untreated alloy was characterized using electrochemical impedance spectroscopy and weight loss
measurements in 0.001 M sodium chloride solution. Although the laser processing parameters
influenced the microstructure and the melt depth of the laser-treated zone, these had little effect
on the corrosion resistance of the alloy.
DOI: 10.1007/s11661-012-1590-x
The Minerals, Metals & Materials Society and ASM International 2013
Mg Zn Ce La Nd Pr Zr Fe Cu Ni
Bal. 3.82 0.64 0.27 0.14 0.06 0.69 0.002 0.002 <0.001
Specimen Nominal Laser Scan Rate Beam Dimensions Nominal Power Nominal Line
Specification Power (W) (m/min) Beam Profile (mm) Density (W/mm2) Energy (J/mm)
Sample 1 1856 19.875 D=1 1182 5.6
along the grain boundaries. The average grain size of the laser-treated zone of both the specimens, forming a
untreated alloy was approximately 50 lm. Figures 2(b) refined and continuous network of precipitates. Figure 3
and (c) show the microstructures of the laser-treated shows the XRD scans of the laser-treated and untreated
samples 1 and 2, respectively. Since the scan rate used for ZE41 alloy. It is evident that both the laser-treated
sample 1 was faster than that for sample 2, while the laser specimens had similar phases (i.e., a–Mg and
power was similar, the cooling rate would be more rapid Mg7Zn3(RE)) as that of the untreated alloy. The phase
in the case of the faster scan rate specimen (sample 1) and composition of the untreated alloy is in agreement with
hence the degree of microstructural refinement achieved earlier studies.[7,8]
was greater in the case of sample 1. On the other hand, a The cross sections of the two laser-treated specimens
more homogeneous microstructure was obtained in the are shown in Figure 4. Large surface cracks were
case of the slow scan rate specimen (sample 2) because of observed in the case of the faster scan rate specimen
the increased laser dwell time, which is given by[30] (sample 1), while the slower scan rate specimen (sample
2) was free of optical microscopically discernable cracks
L
t¼ ½1 (Figure 4(b)). Since ZE41 alloy has a high coefficient of
V thermal expansion and shrinkage due to solidifica-
where L is the beam interaction length in the direction of tion,[40] the presence of cracks in the faster scan rate
laser scan and V is the scan rate. The laser dwell time of specimen (sample 1) may be attributed to the greater
the slower scan rate specimen (2 9 103 min) was about power density and faster scan rate (Table II) employed,
an order of magnitude greater than the faster scan rate which provided a high input energy and a quicker
specimen (1 9 104 min). This led to an increased time cooling rate, causing a high thermal stress and cracking
in the liquid phase for dissolution of second phases and in the laser-treated zone. The melt depths of the slower
diffusion, and as a result a superior melt homogeniza- scan rate specimen (sample 2: ~750 lm) were consider-
tion occurred in the case of the slower scan rate ably greater than that of the faster scan rate specimen
specimen.[30] (sample 1: ~250 lm), which can be attributed to the
The intermetallic particles reprecipitated along the greater line energy and laser dwell time experienced at
grain boundaries of the refined structure throughout the the slower scan rate.
500
α - Mg
Mg7Zn3(RE)
Intensity (CPU) 400
300
200
100
0
20 25 30 35 40 45 50 55 60
2θ (degree)
70 faster scan rate (sample 1) at different times of immersion were similar to the
slower scan rate (sample 2)
60 representative plot.
50 It is evident from the representative comparisons in
40 Figure 10 that the simulated impedance plots (Bode
30 plots), that were generated using the EEC in Figure 8,
20
were in good agreement with the experimental plots
obtained at different times of immersion. Although the
10
impedance data for all the specimens were recorded in
0
0.1 1 10 100 1000 10000 100000 1000000 the frequency range 1 MHz to 10 mHz, the impedance
Frequency (Hz) data were generally simulated in the frequency range of
100 kHz to 100 mHz in order to avoid any artifact in the
(c) high frequency range and/or noise in the low frequency
range,[44] and this practice is consistent with several
Fig. 5—Impedance plots of the laser-treated and untreated speci- other researchers.[41,45–47] The interfacial capacitances
mens at 3 hours of immersion in 0.001 M NaCl: (a) Nyquist, (b)
Bode modulus, and (c) Bode phase angle plots. and resistances associated with the EEC (Figure 8) were
calculated in order to understand the electrochemical
kinetics at the different interfaces (the corrosion product
specimen produced with a faster scan rate had the lowest film interface and the electrical double layer).
corrosion resistance. This may be attributed to the Figure 11 shows the evolution of Cf and Rf of all the
presence of cracks in the case of this specimen, which specimens at different times of immersion. It is evident
was generated by a faster scan rate and higher power that Cf and Rf of all specimens were similar and
density (laser power/area of the beam). remained reasonably unaltered with time, suggesting
EIS is a powerful tool to investigate the nature of no significant change in the hydroxide layer/solution
surface films and associated corrosion mechanisms of interface at different durations of immersion. The
metals and alloys. In order to attain a mechanistic evolution of Qdl and Rc is shown in Figure 12. Qdl of
understanding of the influence of laser parameters on all the specimens increased with time. The exponent (n)
the time-dependent corrosion process, a thorough anal- value of Qdl was close to 1, so it can be approximated as
ysis of the impedance data was carried out. Hence, in the a capacitance.[43] Since capacitance is directly propor-
present study, the impedance of substrate (both un- tional to the area of the capacitor, the trend of increase
treated and laser-treated)/corrosion product film/NaCl of Qdl over time for both the specimens indicates an
solution interfaces was analyzed using electrical equiv- increasing area fraction of exposure of the metal/
alent circuits (EEC). In corrosion processes, each of the hydroxide interface to the electrolyte with time. Qdl of
two interfaces behaves as a capacitance combined in all the specimens was similar until 7 hours of immersion;
5 5
0 0
0 5 10 15 20 0 5 10 15 20
Z' (kΩ cm2) Z' (kΩ cm2)
10
0
0 5 10 15 20
Z' (kΩ cm2)
Fig. 6—Nyquist plots of the laser-treated and the untreated ZE41 at different times of immersion in 0.001 M NaCl.
however, the slower scan rate specimen (sample 2) had of the faster scan rate specimen was a little higher than
lower Qdl than the faster scan rate (sample 1) and the the other two specimens. The slightly higher corrosion
untreated specimens, suggesting a smaller area fraction rate observed in the case of the faster scan rate specimen
of exposure of the metal/hydroxide interface than the was attributed to the cracks on its surface. This trend in
other two specimens. Rc of the faster scan rate specimen the corrosion resistance of the laser-treated and un-
(sample 1) had a decreasing trend, whereas Rc of the treated specimens is consistent with that obtained using
slower scan rate specimen (sample 2) and the untreated the electrochemical method (EIS).
specimen remained constant with the increase in immer- Figure 15 shows the typical evolution of the electrolyte
sion time. pH with increasing immersion time. The change in
The polarization resistance (Rp) is represented as the electrolyte pH was similar for all the specimens. After
algebraic sum of the corrosion film resistance and the 1 day (24 hours) of immersion, the initial pH (pH 5) of the
charge transfer resistance.[41] Rp of all the specimens electrolyte increased to pH 9 and was constant for the rest
(Figure 13) was similar till 7 h; however, after 24 hours of the test duration. This suggests that during the initial
of immersion, the Rp of the faster scan specimen hours of immersion (1 day), there was local alkalization
(sample 1) was the lowest, whereas the Rp values of due to active metal dissolution and the formation of
the untreated ZE41 and the slower scan rate specimen hydroxide film on the metal surface. This is also supported
(sample 2) were similar. The lower Rp of the faster scan by the Rc value (Figure 12(b)), which suggests that the
rate specimen (sample 1) is attributed to the presence of charge transfer resistance decreased (due to active metal
cracks on its surface. Although a refined, homogeneous, dissolution) until 24 hours of immersion (1 day), but
and crack-free microstructure was obtained in the case remained constant after that (due to the formation of the
of the slower scan rate specimen (sample 2), the corrosion product film). Furthermore, the constancy of
polarization resistance of this specimen was similar to the electrolyte pH agrees well with the constancy in the
that of the untreated ZE41. values of the parameters (Cf and Rf) related to the
corrosion product film/electrolyte interface.
C. Weight Loss Measurements
A comparison of the corrosion rates of the laser- D. Corrosion Morphology
treated and untreated specimens is shown in Figure 14. Figure 16 shows the corrosion morphology of the
The corrosion rates of the slower scan rate specimen and untreated and the laser-treated specimens at 55 hours
the untreated specimen were similar after 7 days of immersion in 0.001 M NaCl. Thick corrosion prod-
(168 hours) of immersion, whereas the corrosion rate ucts with scattered pits covered the entire surface of
0 0
0.1 10 1000 100000 10000000 0.1 10 1000 100000 10000000
20
50
15 40
10 30
20
5
10
0
0.1 10 1000 100000 10000000 0
0.1 10 1000 100000 10000000
Frequency (Hz)
Frequency (Hz)
30 faster scan rate (Sample 1) 80 faster scan rate (Sample 1)
At 55 h At 55 h
slower scan rate (Sample 2) 70 slower scan rate (Sample 2)
25 Untreated Untreated
Phase angle (degree)
60
20
|Z| (kΩ cm2)
50
15 40
30
10
20
5
10
0 0
0.1 10 1000 100000 10000000 0.1 10 1000 100000 10000000
Frequency (Hz) Frequency (Hz)
Fig. 7—Bode impedance plots of the laser-treated and untreated specimens at different times of immersion in 0.001 M NaCl.
Fig. 10—Comparisons of the Bode modulus and phase angle plots at different durations of immersion: experimentally determined and simulated
data for a typical laser-treated specimen (sample 1 in this case).
parameters. Even though the present study employed a confirms the earlier proposition that in the absence of
wide variation in laser parameters, the resulting micro- any beneficial alloying elements (such as Al) in the ZE41
structure (Figure 2) still produced no observable alloy, the influence of the change in microstructure
improvement in corrosion resistance. This further due to laser processing parameters on the corrosion
Cf (nF/cm2)
Rf (kΩ cm2)
0.3 15
0.2 10
0.1 5
0 0
0 10 20 30 40 50 60 0 10 20 30 40 50 60
Time (h) Time (h)
(a) (b)
Fig. 11—Evolution of (a) Cf and (b) Rf of the laser-treated and untreated specimens in 0.001 M NaCl solution.
Rc (kΩ cm2)
150
10
100
5
50
0 0
0 10 20 30 40 50 60 0 10 20 30 40 50 60
Fig. 12—Evolution of (a) Qdl and (b) Rc of the laser-treated and untreated specimens in 0.001 M NaCl solution.
20 Untreated ZE41
1.2
1
15
Rp (kΩ cm2)
0.8
10 0.6
0.4
5 0.2
0
0 Faster scan Slower scan Untreated
0 10 20 30 40 50 60 rate rate ZE41
Time (h)
Fig. 14—Weight loss measurements of the laser-treated (the faster
Fig. 13—Evolution of Rp of the laser-treated and untreated speci- scan rate and the slower scan rate) and untreated ZE41.
mens in 0.001 M NaCl solution.
resistance of this alloy is insignificant.[29] The thorough chemical parameters (Figures 11 and 12) among the
time-dependent EIS characterization presented here laser-treated and untreated alloy, thus providing the
confirms the consistency of the interfacial (corrosion mechanistic basis for the insignificant improvement in
product film/electrolyte and metal/electrolyte) electro- corrosion resistance due to laser surface melting.
(a)
Completely
corroded
grains
Pits
100 µm
(b)
Pits
100 µm
(c)
Pits
100 µm
Fig. 16—Corrosion morphology of the (a) untreated, (b) faster scan rate (sample 1), and (c) slower scan rate (sample 2) specimens at 55 h of
immersion in 0.001 M NaCl.