-1-
. . . ADDITIONAL INFORMATION (FAST ENTRY)
for AS800, 2000T, TH02 and 600T SAFETY TX HANDLING
Device: TH02
The TH02 device is an EEPROM Control device and two possible status can be set: BURN to EEPROM and NOT
BURN to EEPROM; BURN to EEPROM means that when a parameter (or a set) is changed, the data are copyed
from RAM to EEPROM by executing the Hart command 39 making changes permanent; NO BURN to EEPROM
means that parameters are changed only in RAM so when the device is reset, data are lost. The operator is
requested to set the BURN or NOT BURN mode in all the menus where data can be changed (CONFIG, TRIM).
Menù Structure:
PV:
Primary Variable - Temp. + Analog Output
Sensor Temperature (F1)
Sensor 1 (F2)
Sensor 2 (F3)
REVIEW:
TX Info (F1)
Tx type, Manufacturer, Tag, Descriptor, Message, Date, Device Number,H/W Write Protection,
Units, Range (LRV, URV), Damping, Limits Units, Minimum Span, Limits (LRL, URL), Sensor type,
Select Input, Channels, Main Supply, Wires, Serial Sensor, Break Monitoring, Short Monitoring (1
and 2) Cold Junction Compensation Temperature, Max temperature for Sensor, Limits (High, Low)
for saturation output, Threshold (1 and 2), ST1+Rel+led, ST2+Rel+led, Sensor error, Device Error,
Text 1, Text 2, Final Assembly number, H/W, S/W, UCD, TSD.
CONFIG:
Chng IO (F1)
Sensor type, Units, Range (LRV, URV), Damping limits (High, Low), Threshold (1 and 2)
Xmtr Info (F2)
Tag, Date, Descriptor, Message
Gen Conf (F3)
Polling Address
Dbase Mem (F4)
Tag List (F1)
Save Conf (F2)
Send Conf (F3)
Delete Conf (F4)
Data Stored: same as for Generic Hart device: see APPENDIX "D"
SERIAL LINK:
Dbase Mode (F1)
Modem Mode (F2)
Load Prog (F3)
TRIM:
4-20 mA Trim (F2)
Other Scale (F1)
4 to 20 mA (F2)
Factory Preset (F3)
Reset factory trim (F4)
Loop Test (F3)
4 mA (F1)
20 mA (F2)
Other (F3)
-3-
. . . ADDITIONAL INFORMATION (FAST ENTRY)
for AS800, 2000T, TH02 and 600T SAFETY TX HANDLING
-4-