CALIBRATION CERTIFICATE

REF: IVD/CAL/2019/4 24.04.2019

NAME OF THE CUSTOMER MANUFACTURER


Lakshmi Medical Center,
BIOSYSTEMS,SPAIN
35/1, Kabeer Dasar Street,
Venkatese Colony, Pollachi - 2
PRODUCT NAME SERIAL No:

BTS 350 801759458

CALIBRATED DATE: 24.04.2019

This is to Certify that the above mentioned equipment has been inspected and Calibrated as per the
protocol on 24.04.2019. The Equipment is working satisfactorily. This Calibration is valid through
25.04.2020. The Next Calibration is due on 24.04.2020.

UPS POWER Supply:


1. E to N 1 VAC In Range
2. E To P 228 VAC In Range
3. P To N 229 VAC In Range

S.No Power Supply Measured


01 Power adapter 15 VDC 14.97 VDC
02 Inbuilt Battery 12 VDC 11.98 VDC

S.No CHECK PARAMETERS DAC STATUS


01 340nm (50 - 90) 63
02 405nm (75 – 140) 101
03 505nm(75 – 140) 110
04 535nm (75 – 140) 98
05 560nm (60 - 120) 102
06 600nm (75 – 140) 105
07 635nm (75 – 140) 107
08 Perilstatic Pump Working Good
09 Aspiration Volume Good
10 Flowcell Temperature (37°C) 36.9°C

ANY REMARKS: Instrument working good


For Bio ‘e’ Services

Engineer’s Signature,