MOBIX-1000

IMPORTANT! X-ray Protection

X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS


MEASURES OF PROTECTION ARE STRICTLY OBSERVED.

X-ray equipment if not properly used may cause injury. Accordingly, the instructions
herein should be thoroughly read and understood before attempting to place this
equipment in operation. We will be glad to assist and cooperate in placing this
equipment in use.
 
Although this apparatus is built to the highest safety standards and incorporates a high
degree of protection against X-radiation other than the useful beam, no practical design
of equipment can provide complete protection, Nor can any practical design compel the
operator to take adequate precautions to prevent the possibility of any persons
carelessly unwisely, or unknowingly exposing themselves or others to X-radiation.
 
It is important that everyone working with X-radiation is properly trained and takes
adequate steps to insure protection against injury. It is also imperative that these
persons are fully acquainted with the recommendations of the international Commission
on Radiation Protection contained in Annals Number 26 of the ICRP, with the
recommendations of the National Council on Radiation Protection and Measurements, as
published in NCRP Reports, and with other applicable standards.
 
All persons using X-radiation equipment must be cognizant of the danger of excessive
exposure to X-radiation. The equipment herein described is sold with the understanding
that the LISTEM Corporation, its agents, and representatives are not liable for injury or
damage which my result from exposure to X-radiation.
 
Various protective material and devices are available. It is recommended that such
materials and devices be used.

LISTEM Corporation - 2 -
MOBIX-1000

CONTENT

1. Introdution .................................................................... 4
2. Configuration ................................................................ 5
2.1 System Configuration
2.2 System Environment and Rating
3. Installation and Adjustment .......................................... 10
3.1 Installation Location
3.2 Power Supply
3.3 Installation
3.4 Adjustment
3.5 Check Point and LED
3.6 Interface Connector
3.7 kV,mA Measurement
4. Display Console ............................................................ 22
5. Calibration ..................................................................... 24
5.1 Calibration Symbol
5.2 Anatomic Program
5.3 kV,mA Calibration
5.4 Program for Filament preheating
6. Error and Troubleshooting ........................................... 29
6.1 Mobix-1000 Operation
6.2 Error Display
6.3 Troubleshooting
7. Other Data .................................................................... 38
7.1 Tube Technical Data
7.2 kV/mAs Table
7.3 Standard Exposure factor table
8. Electric Schematics ...................................................... 42

LISTEM Corporation - 3 -
MOBIX-1000

1. Introduction

This sustem is specially designed for precise radiographic diagnosis in the case where
patients can't move to the X-ray room due to serious injury or some reason and is
integrated with Mobile body, X-ray generator( X-ray tube inside ) and X-ray controller.
Therefore it provides easy installation in the narrow space, easy use for high
performance micro-controller and easy handing thanks to compact size.
Especially this system is operated with house wall power or Power Storage Unit even in
the bad-power environment.

Its main features are

▶ Small body provides availability in the narrow space


▶ Smooth phantographic arm movement
▶ Arm swiveling movement covers wide radipgraphic range
▶ Adaptation of high-performance micro-controller for precise control
▶ Precise mAs control by mAs integrator
▶ Constant exposure density with optimum exposing factor storage(APR) reduces
exposure
▶ Easy X-ray radiation with anatomic porgram
▶ Soft touch panel adoption
▶ Self diagnosis with error message display
▶ Power supply installation cost reduction with house wall power usage

LISTEM Corporation - 4 -
MOBIX-1000

2. Configuration
2.1 Configuration
This radiographic mobile X-ray system is composed of the following units.
Refer to Fig 1.

▶ X-ray controller : Mobix-1000C


▶ X-ray generator : Mobix-1000H
▶ Collimator : BLD-150RK
▶ Body
▶ Didplay Console
▶ Operating Guide

LISTEM Corporation - 5 -
MOBIX-1000

① X-ray Controller
② X-ray Generator
③ Collimator
④ Display Console

< Fig 1. System Configuration >

LISTEM Corporation - 6 -
MOBIX-1000

2.2 System Environment and Rating


2.2.1 System Environment

Temperature : +5℃~40℃
During use
Humidity : 20% ~ 80%RH

During storage or Temperature : +5℃~40℃


non-use Humidity : 20% ~ 80%RH

2.2.2 Rating
2.2.2.1 MAIN Power
Single 190~230VAC, 30A, 50/60㎐
2.2.2.2 Electric Rating

Output : Maximum 2.5kW ( 125kV , 20mA )


Tube Voltage : 40~125kW , 81kV~125kV
Tube Current : 30mA , 20mA
Exposure Time : Within 6.6 seconds
Overload Protection
40~80kV : Up to 150mAs
81~ 125kV : Up to 60mAs
MVP tolerance : Within 5kVp at the line voltage of 210VAC
Rising Time : Within 10ms to the 75% of output
Overshoot : 2kVp
mAs step : 42 steps
Stability : Within ± 1KV at the fluctuation of main input power

Monitoring
Tube Voltage : Test point TP2-GND

20kV = 1V (Within 5% )

Tube Current : Test Point TP9-GND

10mA = 3V (Within 5% )

LISTEM Corporation - 7 -
MOBIX-1000

2.2.2.3 Alarm Signal


This system geneerates ALARM signal (5A-5B in CN3) with HV OFF when
trouble occurs in the HV unit. LED2 on the Display PCB turns off

2.2.2.4 Protection
Tube Voltage : 135kV ± 5%
Tube Current : 40mA ± 5%

2.2.2.5 Withstanding
Insulation Resistance : More than 50MΩ under 500V between AC input and
GND & System
Withstanding Voltage : Has to withstand for 1 minute under AC1500V
between AC input and GND & System
Self-withstanding voltage : for 3 minutes under 125kV ×1.2

2.2.2.6 AC input overcurrent portection


AC input current is blocked by the circult braker in case of error in the HV
unit.

☞ Note
Circuit braker is recommended by the system manufacturer.

2.2.2.7 Filament Unit (Frequency about 50kHz)

Voltage : 3 ~ 7V
Current : 2 ~ 5A
Operating Time : Depends on the exposure time
Duty : Depends on the radiographic duty.

Filament is controlled by the average value of primary current of the Filament


Transformer.

LISTEM Corporation - 8 -
MOBIX-1000

2.2.3 Specification

NO ITEM Specification Decription

1 MODEL Mobix-1000

2 Maximum Rating 40-125kVp/200mAs

3 Tube Focus 1.2mm×1.2mm

4 Maximum Height 2055mm

Single 220VAC
5 Power Supply
50/60Hz, 3kVA

6 mAs 1~200mAs

7 Display Panel Membrane Touch panel

8 X-ray Controller Installed inside body

mAs Display : Digital Display


kVp Display : Digital Display
Power ON LED
Body size and Organ selection LED
PREP,EXP indicator
Overload Error LED
9 Console
Filament Error LED
mAs selector : UP/DOWN
kVp selector : UP/DOWN
Power ON/OFF SW
SET SW : Exposure Factor storage
PREP,EXP SW : Pushbutton

Phantographic Arm Movement


10 Arm Rotaion Phntographic X-ray generator Movement
ARM Swivelling

Collimator : Manual Driven, Rectangular


11 Others
Film Storage

LISTEM Corporation - 9 -
MOBIX-1000

3. Installation and Adjustment


3.1 Installation Location
Take the following cautions before installation
1) Floor Condition for the system weight
2) Power supply required for the system operation
3) Location power supply and the system Within 5m between power sypply and
the system
4) Cooling Environment
5) Apprepriate area for the installation
6) Location of auxiliary Device

3.2 Power Supply


Power supply must have the following tolerance to ensure the stable operation
1) AC Power : Single phase 220VAC
2) Power supply frequency :50/60Hz
3) Maximum input power : 3.0kVA
4) Power fluctuation without load : Within ±5% of nominal power supply
5) Frequincy Fluctuation : Within ±0.5% of normal frequincy
6) Power supply Impedance : Less than 0.2Ω
7) Power Line : Thickness must be more than 5㎟
8) Braker : Over 220VAC, 30A

LISTEM Corporation - 10 -
MOBIX-1000

3.3 Installation
Various cables and connecctors should be connected after system positioning and
basic assembly

Required Document and Tools for Installation


․ Instruction
․ System wiring Sch.
․ System Dimemsional schematic
․ A set of tools for installation
․ General-purpose Multimeter
․ Oscilloscope (if necessary)

Precaution on Installation
All work must be carried out with extal care not to be shocked by AC electricity or
electrostatic discharge.

Especially during technical service after system power OFF, pay special attention to
the main charge capacitor which is charge to DC 310V under normal operation.
Wait at least more than 1 minute for the capacitor to discharge.

Make sure that all protective ground wires are connected porperly before starting
the system after installation. This connection applies to all metallic covers which
may generate unexpexted contact voltage in case of a fault, They must be
connceted ot the protective ground wire.

means " Terminal for Connection of Equipment protective ground wire".

After installation, the protective ground wire must be tested with the test.

LISTEM Corporation - 11 -
MOBIX-1000

3.3.1 Installation Area


1) Secire enough space for normal system operation and placement.
2) Check floor condition before installation for better parking state.

☞ NOTE : Maximum body size 2055mm×1645mm

Minimum Space for Installation

LISTEM Corporation - 12 -
MOBIX-1000

3.4 Adjustment
Since this equipment is all calibrated and adjusted before delivery, compensates tube
voltage and current automatically within ±10% of input power fluctuation, it is
recommended that additional ajuustment not be made at the user site without prior
notice. Also, safety inspection should be performed regularly during use. System must
be calibrated and adjusted every time it is replaced. Make sure that all connections
must be ,checked before system adjustment.
Since this type is specially designed with integration of HV transfomer and x-ray
tube, non-invasive measuring instruments are required to measure tube voltahe
non-invasively. Or Connecting oscillocope or multimeter to the check point is possible
ot read and measure kVp,mA(between the TP2(+),TP9(+) -GND(-) of HV control
Board).Another way to measure tube current is check frequency between TP6 and TP3
on the Console Display Board. Refer chapter 5 Error and Troubleshooting.

☞ NOTE: All ground (Green with Orange stripe) must be connected

3.4.1 Power ON
Ensure that all cables and ground wires are connceted porperly before applying
power to the system. Blank dashes are displayed for about 10 seconds when the
system in ON. All initalizing tasks are done during this time.

☞ NOTE : Last setting is maintained when the system starts.

LISTEM Corporation - 13 -
MOBIX-1000

3.4.2 Pre-adjustment
Take the following steps using the various variable resistors before going to
adjustment,
1. Put the oscilloscope probes to test points TP3 and GND.
Adjust the frequency to 200kHz turning VR2. This frequency is main IGBT
inverting frequency(Actual frequency is 50kHz. 100kHz is measured form TL594)
2. Put the voltimeter porbes to test points TP1 and GND.
Adjust the voltage to 6.75VDC turning VR1. This value is tube overvoltage
protection level.
3. Put the voltimeter probes to test points TP8 and GND.
Adjust the voltage to 12VDC turning VR4. This value is tube overcurrent
portection level.
4. Put the oscilloscope probes to test points TP4 and GND. also to TP9 and GND.
Make an exposure and measure the frequency between TP4 and GND, voltage
between TP9 and GND.
Adjust the frequency to 2kHz turning VR5 and calibration tube mA data. Details
explanation is given later in Chapter 5 Calibration.

3.4.3. Measuring Instrument Connection Diagram for the System Adjustment

LISTEM Corporation - 14 -
MOBIX-1000

mA Measurement

Connect an oscilloscope to the check points on the Display Console PCB

mA CHECK POINT
TP6(+) - TP3(-) : 30mA = 6kHz

Connect an oscilloscope to the check points on the X-ray control PCB

mA CHECK POINT
TP4(+) - GND(-) : 30mA = 6kHz

Connect an oscilloscope or voltimeter to the check points on the HV Control PCB

mA CHECK POINT
TP9(+) - GND(-) : 10mA = 3V

kVp measurement
Non-Invasive kVp Meter
: Place non-invasive kVp meter 60-100cm from the X-ray tube focus.

Connect an oscilloscpoe to the check points on the X-ray Control Board

kVp CHECK POINT


TP2(+) - GND(-) : 20kVp = 1V

LISTEM Corporation - 15 -
MOBIX-1000

3.4.4 Compensation for the Input Power Fluctuation


Since this system is designed mainly for input power of single 220VAC 50/60Hz,
check the power input before installation.
Serious damage may be caused to the system unless the following tolerance is
observed.

Input Power : Single 220VAC±10VAC, 50/60Hz, 3kVA


LINE IMPEDANCE : Less than 0.2Ω

LISTEM Corporation - 16 -
MOBIX-1000

3.5 Check Point and LEDs


Check Points on the Display Console Board

CHECK POINT Description

TP1 +5VDC

TP2 +15VDC

TP3 0V

ALARM signal
TP6
Generates when serious error occurs in HV unit.

Actual tube current measuring point


TP7
10mA = 2kHz

LEDs on the Display Console Board

LED Description

LED1 +5VDC

LED2 +15VDC

LED3 X-ray radiation indicator

LISTEM Corporation - 17 -
MOBIX-1000

Check Points on the X-ray Control Board

CHECK POINT Description

TP1 Tube Overvoltahe limit

kVp Feedback
TP2
20kVp = 1V

Main Inverting Frequency


TP3
Set 200kHz

TP4 Tube Current

TP5 IGBT Drive signal

TP6 IGBT Drive signal

TP7 kVp Set

TP8 Tube Overcurrent limit

mA Feedback
TP8
10mA = 3V

TP10 kVp Feedback

TP11 mA Set

GND GND

LEDs on the X-ray Control Board

LED COLOR Description

LED1 RED Reserved

LED2 RED Tube Overcurrent

LED3 RED Reserved

LED4 RED Tube Overcurrent

LISTEM Corporation - 18 -
MOBIX-1000

CHECK POINT on Filament Drive Board

CHECK POINT Description

TP1 Filament Drive Voltage

TP2 Filament Drive Signal 1

TP3 Filament Drive Signal 2

TP4 Drive Signal for Filament Voltage

TP5 DC0V(for DC310)

TP6 Drive Signal for Filament Voltage

LISTEM Corporation - 19 -
MOBIX-1000

3.6 Inter Connector

Connector Pin NO Function Level

+ -

1A 1B RAD ON Opto-coupler ON

2A 2B Filament ON Opto-coupler ON

3A 3B LARGE or SMALL Focus Opto-coupler ON=SMALL

4A 4B RAD or FLU Opto-coupler ON=FLU

5A 5B Tube current 10mA=2kHz

6A 6B ALARM Opto-coupler OFF=ALARM

7A 7B Reserved

8A 8B MSB

9A 9B

10A 10B
8bit Digital Signal
11A 11B
kVp setting Opto-coupler ON=Logic 1
12A 12B
1 STEP=1kVp
13A 13B

14A 14B

15A 15B LSB

16A 16B Reserved

17A 17B MSB

18A 18B

19A 19B
8bit Digital Signal
20A 20B
mA setting Opto-coupler ON=Logic 1
21A 21B
1 STEP=20mA
22A 22B

23A 23B

24A 24B LSB

25A 25B Reserved

LISTEM Corporation - 20 -
MOBIX-1000

3.7 kV,mA Measurement

Actual kVp Measurement


Connect an oscilloscpoe to the check point on the HV Control PCB.

kVp CHECK POINT


TP2(+) - GND(-) : 20kVp = 1V

Actual mA Measurement
Connect an oscilloscope between check points on the Display Console PCB
to measure the frequency

mA CHECK POINT
TP6(+) - TP3(-) : 30mA = 6kHz

Connect an oscilloscope between check points on the Display Console PCB


to measure the voltage

mA CHECK POINT
TP9(+)-GND(-) : 30mA = 9V

LISTEM Corporation - 21 -
MOBIX-1000

4. Display console

① Power On Switch and Indocator => Power On


② Power OFF Switch => Power OFF
③ BUCKY 1 selection button and indicator
④ BUCKY 2 selection button and indicator
⑤ Body size selection buttons => Selects body size for APR Program
⑥ Buttons and indicators for the exposure => Save exposure factor with a buzzer
⑦ Exposure selection buttons and indicators and indicators
⑧ Filament Error Indicator => Reserver Combined with fatal error
⑨ kVp UP/DOWN Switch and display window => 40-125kVp, 1STEP
⑩ mAs UP/DOWN Switch and display window => 1mAs to 150mAs
⑪ Exposure Factor storage switch => Stores exposure factors (kVp, mAs)

LISTEM Corporation - 22 -
MOBIX-1000

kV, mAs Interaction

kV range Available mAs


40-50kV 1-200mAs
51-60kV 1-180mAs
61-70kV 1-150mAs
71-80kV 1-120mAs
81-90kV 1-100mAs
91-100kV 1-80mAs
101-110kV 1-40mAs
111-120kV 1-26mAs

LISTEM Corporation - 23 -
MOBIX-1000

5. Calibration
Required Tools for Calibration
Multimeter - Tube Voltage and Current measurement
Oscilloscope - Tube Voltage, Current and its waveform measurement
Various tools for calibration
Instrucions
System Wiring Schematic
Electric Schematics

☞ NOTE : All calibration mode or Program procedure begins with Image initial
Position SW

Make the following calibration in case of system aging, replacement, maintenacen


system error or when adjustments are required. The system has the following
calibrations and porgrams.

Sofeware Calibrations
EEPROM RESET
Anatomic Program
kVp, mA calibration
Preparation Time Program

LISTEM Corporation - 24 -
MOBIX-1000

5.1 Calibration Symbol

BUCKY1 SW: kV,mA Calibration, Filament Preheating Program

HEAD LAT SW: kV,mA Calibration

HEAD AP/PA SW: Filament Preheating Program

5.2 Anatomic Program

Follow the below instructions to set the exposure parameters (kV, mAs )
of anotomic program

1) Select NO BUCKY、BUCKY1, BUCKY2, BUCKY1,2 , buttons are TOGGKE type.


2) Select body size
3) Select body part
4) Select details(6) from body part section
5) Select kV, mAs
6) Press SET switch, SET LED and Buzzer shall be operated and then kV,mAs
shall be saved.

LISTEM Corporation - 25 -
MOBIX-1000

5.3 kV, mA Calibration

Make this calibration when kV or mA values are out of range or untolerable.


Select 40kV and 2mAs as start point. This is recommended not mandatory order
Put the oscilloscope probes between TP2 and GND , YP9 and GND.

☞ NOTE : TP2 and GND -> kVp measurement


TP9 and GEN -> kVp measurement

1) Press BUCKY1 button and then press HEAD LAT switch

The following message will be displayed with a buzzer.

P1
kV Data : 123
30mA Data : 123
P2 20mA Data : 123
Heating offset : -2

40-80kV : 30mA,20mA select


81-125kV : 20mA select

This message is the calibration value for 40kV chosen.


The 40kV should have a setting value
for 30mA, 20mA, so the "kV Data : 123" on LCD is the kV value for the setting
value 40kV and "30mA Data : 123" is 30mA value for 40kV.
As mentioned above, the value of 40-80kV should be calibratied at 20mA,
so " 20mA Data" is 20mA-calibrated value at 40kV.

LISTEM Corporation - 26 -
MOBIX-1000

First of all, the calibration at 20mA should be done.


2) Increase av by pressing CHEST SW and decrease kV by pressing SMALL SW
3) The mark "▶"is indicating 30mA Data, and now 20mA calibration at 40~80kV is
being done. ( if ▶ indicates 30mA Data, press P2 ). Increase mA by pressing
ABDOMEN SW and decrease mA by pressing MEDIUM SW.
4) Take X-ray
5) Check the wave and change Heating offset to make mA wave plane
6) Obtain 40kV, 30mA by measuring kV,mA

kVp measurement

kVp CHECK POINT


on X-RAY Control PCB
TP2(+) - GND(-) : 20kVp = 1V

mA measurement

mA CHECK POINT
on X-RAY Control PCB
TP9(+) - GND(-) : 30mA = 9V

7) In case the output is within a permitted limit, press SET. It will be saved
with buzzer.
8) Select 40,50,60,70,80,90,100,110 and 120 kV in turn and repeat STEP 2)-7).
Calibration Data is stored automatically the moment SET SW is pressed.

Calibrate against 30mA


9) Select 40kV and press P1 and then check if ▶ indicates 30mA Data.
10) Repeat 2)-7). Increase and decrease mA data in mA measurement
terminal to be 30mA=9V
11) Select 40,50,60,70,80 kV in turn and repeat STEP 2)-7)
12) Exit form the calibration mode by pressing BUCKY 1 SW.

LISTEM Corporation - 27 -
MOBIX-1000

5.4 Program for Filament Preheating

Programs Filament Preheating luminosity

1) Press BUCKY1 button and HEAD AP/PA at the same time.

The following message will be displayed on LCD Window with buzzer.

2) Increase a value for Filament Preheating by pressing CHEST SW and decrease


it by pressing SMALL SW.

3) Press SET SW after setting up the value for Filament Preheating


4) Exit from program mode by pressing BUCKY 1 SW.

LISTEM Corporation - 28 -
MOBIX-1000

6. Error and Troubleshooting


6.1 Mobix-1000 Operation

1) When the power is on, "Initializing..." will be displayed on two LCD.


EEPROM and SYSTEM will be automatically checked.1)
After 5 seconds, if there is no error, kV,mAs will be on LCD1 and
"Anatomic Mode" will be on LCD 2.
In case an error shows up, ERROR message appears on kV indicator.
Press any keys if you want to continue.

2) Select kv/mAs with kV and mAs UP/DOWN SW. If it is necessary, select body
size and exposing parts and press SET SW to store exposure factors.

3) Make preparation and exposure with the set values. It some error occurs,
corresponding error message is displayed on the kV window.

NOTE
When operating the phantographic Arm, it shall unfasten the lock system always.
After the operation, it shall fasten and fix it
Arm should not operate while fastening it, otherwise it should unfasten to operate
it. When operating the lock while fastening it, it can bring about the abrasion and
prevent it from fixing.

LISTEM Corporation - 29 -
MOBIX-1000

6.2 Error Display

Error Code Description

Fatal Error
Probable cause
E00
. Over Tube Voltage
. Over Tube Current

E01 Serial EEPROM ACK Error

E02 Serial EEPROM READ Error

E04 RAD mA less than allowable current

E05 RAD mA more than 40mA

E06 EEPROM Error

E09 System Overload

E10 No X-ray Raiated

LISTEM Corporation - 30 -
MOBIX-1000

6.3 Troubleshooting

E00 Fatal error occurred in the HV unit or abnormal working.

This is usually caused by over/under voltage, over/under current or


HV spark.
a. Check the line to primary of HV transformer and filament
connection.
b. Check the feedback line(CN5 on control board PC1)
CN5-6 should be connected rightly.
c. Check the 4 error-indicating LEDs on the Control PCB(PC1).
Depending on the LED, follow the trouble shooting Flow Chart
d. Check the filament main heating.

E01 Serial EEPROM I2C Protocol ACK Error


This occurs when CPU cannot accept ACK signal from the serial
EEPROM. This is usually caused by the serial EEPROM malfunction.

a. Turn off the system and then start again in a few minutes.
b. If E01 keeps occuring, replace with a new one. Do the
re-calibration after replacement. Refer to Chapter Calibration.

E02 Serial EEPROM I2C Protocol READ Error


This occurs when CPU cannot accept READ signal from the serial
EEPROM. This is usually caused by the serial EEPROM malfunction.

a. Turn off the system and then start again in a few minutes.
b. If E01 keeps occuring, replace with a new one. Take the
re-calibration after replacement. Refer to Chapter Calibration.

OTE : E01, E02 always occurs after EEPROM is replaced to new


one. Take re-calibration after replacement.

LISTEM Corporation - 31 -
MOBIX-1000

E04 Exposure with mA less than allowable current in the RAD mode.

kVp Expected mA E04 range

40~80 30mA,20mA less thnn l10mA

81~125 20mA less than 5mA

Connect a voltmeter or oscilloscope on the check point TP2(+), GND(-)


on X-ray control PCB and then check the voltage.

E05 Exposure with mA more than allowable current in the RAD mode.

E06 EEPROM Error

E09 System Overload


Refer to error code E00

E10 No X-ray radiated


Check the interface connector between console and control PCB.
Check the digital signal of interface connector and check the value
of kV,mA set analog.
Check the IGBT drive signals(TP5, TP6 on control board)
Check the input signal of HV transformer. normal wave is close to

LISTEM Corporation - 32 -
MOBIX-1000

NO Error Porbable caues Service

▶ Check Fuse F2
▶ No maon power supply ▶ Check 15VAC from
▶ Transformer T1 error Transformer T1
1 NO Power on
▶ Relay RLY1,2 error ▶ Check Relay RLY1,2
▶ Console panel error ▶ Check Connector P3
▶ Replace console panel

▶ Control PCB error


▶ Bad connection between ▶ Check pulse output of
Connector P7 of TL494
2 NO X-ray
Console PCB and CN9 ▶ Check IGBT drive signal
of Control PCB ▶ Replace monotank
▶ Filament Drive error

▶ Relay RLY3 error


3 NO Collimating light ▶ F3 error
▶ Transformer T3 error

LISTEM Corporation - 33 -
MOBIX-1000

LISTEM Corporation - 34 -
MOBIX-1000

LISTEM Corporation - 35 -
MOBIX-1000

LISTEM Corporation - 36 -
MOBIX-1000

LISTEM Corporation - 37 -
MOBIX-1000

7. Other Data
7.1 Tube Technical Data
1) Type : Stationary

2) Maximum kVp : 125kVp


3) Focus : 2.0mm 2.0mm
4) Target Angle : 20°
5) Heat Capacity : 35kJ(50KHU)
6) Maximum Tube Current ( Full wave rectified) : 80mA

7) Weight : 500g

LISTEM Corporation - 38 -
MOBIX-1000

7.2 kV, mAs Table (50/60Hz)

NO mAs 40-80kVp 81-125kVp


1 0.3 30mA 20mA
2 1 30mA 20mA
3 2 30mA 20mA
4 3 30mA 20mA
5 4 30mA 20mA
6 5 30mA 20mA
7 6 30mA 20mA
8 7 30mA 20mA
9 8 30mA 20mA
10 9 30mA 20mA
11 10 30mA 20mA
12 12 30mA 20mA
13 14 30mA 20mA
14 16 30mA 20mA
15 18 30mA 20mA
16 20 30mA 20mA
17 22 30mA 20mA
18 24 30mA 20mA
19 26 30mA 20mA
20 28 30mA 20mA
21 30 30mA 20mA
22 32 30mA 20mA
23 34 30mA 20mA
24 36 30mA 20mA
25 38 30mA 20mA
26 40 30mA 20mA

LISTEM Corporation - 39 -
MOBIX-1000

NO mAs 40-80kVp 81-125kVp


27 45 30mA 20mA
28 50 30mA 20mA
29 55 30mA 20mA
30 60 30mA 20mA
31 65 30mA OVERLOAD
32 70 30mA OVERLOAD
33 75 30mA OVERLOAD
34 80 30mA OVERLOAD
35 85 30mA OVERLOAD
36 90 30mA OVERLOAD
37 95 30mA OVERLOAD
38 100 30mA OVERLOAD
39 110 30mA OVERLOAD
40 120 30mA OVERLOAD
41 150 30mA OVERLOAD

LISTEM Corporation - 40 -
MOBIX-1000

7.3 Standard exposure factor table

50/60Hz, TO-440

With Table Without table


Part kVp
mAs Distance Drid mAs Distance Drid

Chest PA 100 3.75 140Cm 3.75 140Cm

L-spine AP 80 32.5 100Cm 75 140Cm

8:1, 8:1,
L-spine Lat 80 62.0 100Cm 120 140Cm
103line 103line
/ /
Pelvis 80 32.5 100Cm 62 140Cm
inch inch

Abdomen 80 32.5 100Cm 62 140Cm

Skull AP 80 24 100Cm 45 140Cm

Hand AP 80 2 100Cm X 3.75 140Cm X

Forearm AP 80 3.75 100Cm X 7.25 140Cm X

☞ Exposure facotrs above can be changed without prior notice depending

on the system enviroment ( using film, screen, developing condition... )

LISTEM Corporation - 41 -
MOBIX-1000

8. Electric Schematics

1) CONTROL SCH 1/3 ( CBDE020100DR0 )

CONTROL SCH 2/3 ( CBDE020200DR0 )

CONTROL SCH 3/3 ( CBDE020300DR0 )

2) DISPLAY SCH 1/2 ( CBDE010100DRA )

DISPLAY SCH 2/2 ( CBDE010200DR0 )

3) Filament Drive ( CBDE040100DR0 )

4) RELAY BOARD(PSU) ( CBDE030101DRA )

5) HV-MONOTANK ( CBDE140100DR0 )

6) IGBT-DRIVE ( CBDE050100DR0 )

7) Snubber SCH ( CBDE070100DR0 )

8) POWER CHARGE SCH ( CBDE060100DR0 )


9) Membrance switch schematic ( CBDE130100DR0 )

10) LCD PANEL SCH ( CBDE080100DR0 )

11) BLOCK DIAGRAM 1(PSU) ( CBDE100101SR0 )


12) BLOCK DIAGRAM 2 ( CBDE100100WR0 )

13) POWER WIRING DIAGRAM ( CBDE060100SR0 )

LISTEM Corporation - 42 -