Documente Academic
Documente Profesional
Documente Cultură
6-02-07
Procedures Page 1 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
CONTENTS
PURPOSE
5. Treatment of the calibration results
WORK INSTRUCTION and calibration period
5.1 Treatment of Calibration Results.
1. Introduction
5.2 Calibration period
2. Technical Requirements
2.1 Interior Crystal Oscillator APPENDICES
2.2 Time Interval Measuring
Appendix 1
3. Calibration Conditions
3.1 Environmental Conditions Contents and format of the calibration
3.2 Standard equipment for calibration certificate
4. Subject and method of calibration
Appendix 2
4.1 Examination of the exterior and
during normal working Calibration of the crystal oscillator
4.2 Calibration of the interior crystal standard if the chronometer does not have
oscillation an output of the crystal oscillator
4.3 Calibration of the measuring error frequency
Appendix 3
The calibration of the crystal-oscillator
stopwatch
Source:
Verification regulation of time interval measuring instrument with digital indication (trial usage)
[Issued on May 31, 1995 and put into effect since May 1, 1996 by National Technical Bureau - JJG 238—95, National
Measuring Verification Regulation of People’s Republic of China]
Prepared by Approved
Date Date
ITTC – Recommended 7.6-02-07
Procedures Page 2 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
Signal B Filter B
Fig. 1 Fundamental diagram of the time interval measuring instrument with digital indication
These chronometers are used for the fol- opening of two mechanical contacts
lowing measurements: duration of closing or opening for a single
mechanical contact
time interval between two electronic pulses time interval between the actions of a sin-
(rising or falling edge) gle electronic pulse and a single mechanical
time of one single electronic pulse contact
time interval between closing and (or)
ITTC – Recommended 7.6-02-07
Procedures Page 3 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
4.1.1 The calibrated chronometer should 4.2 Calibration of the interior crystal oscilla-
not have any mechanical damages, which tion
could affect the normal use of the instru-
ment. All the control switches should work The connection of the instruments is shown in
easily and reliably. Fig. 2.
4.1.2 Switch the current on. A standard time
Calibrated chro-
nometer comparator Reference frequency standard
(a)
Calibrated chro-
counter Reference frequency standard
nometer
(b)
Fig. 2 Connection diagram of the instruments for calibrating the interior crystal oscillation
where: f − f0
fi measured value of each sample by A= (4)
use of the electronic counter. f0
f0 nominal value of the output fre- The crystal oscillation frequency must be
quency for the calibrated crystal oscillation. aligned to the same order as that of the
frequency fluctuation. However, the accu-
Or: S = ymax (τ ) - ymin (τ ) (2)
racy ought to be noted on the the calibra-
tion certificate.
1 3
y (τ ) = ∑ yi (τ )
3 i =1
If the calibrated chronometer does not
have an output of the crystal oscillation,
where: the calibration of the crystal oscillation
yi(τ) average relative frequency devia- norm can be carried out on the basis of the
tion directly measured using a comparator method given in the Appendix 2.
of the frequency standard.
4.2.2 Alignment of the frequency accuracy 4.3 Calibration of the measuring error
To ensure the reliability of the calibration
result during the process the accuracy of 4.3.1 Selection of the calibration values and
the crystal oscillation must be restricted to the error calculation.
a value less than one-tenth of the fre- The measurement begins at the minimum
quency fluctuation value. value, followed by a series of values ten
The frequency accuracy may be calculated times greater each time and then ends with
by formula (3) using the data of the fre- the maximum value.
quency fluctuation calibration. If the measuring range of the chronometer
is divided into sub-ranges, the calibration
f − f0 value at the lowest sub-range can be se-
max
A= (3)
f0 lected according to the above mentioned
If the result of the calculation is better than method. For the calibration only the
the required accuracy, the crystal oscillat- maximum values of the other sub-ranges
ing frequency need not be adjusted. Oth- need to be taken.
erwise it must be aligned. The measurement must be done three times
When being adjusted the sampling time for each calibration value. The arithmetic
can be taken as 10s and only one meas- mean value will be taken as the measuring
urement is sufficient. The frequency accu- value.
racy may be calculated by formula (4): The measuring error can be calculated by
formula (5):
ITTC – Recommended 7.6-02-07
Procedures Page 6 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
A calibrated
Time interval generator
B chronometer
Reference of frequency
Standard
Reference of frequency
Standard
4.3.2.2 Pulse Width Measurement of the calibrated chronometer. The stop signal is
Negative Pulse put on B channel. The trigger slopes of
The time interval generator gives a single both channels are set to be negative.
negative pulse signal. The trigger slopes of 4.3.3 Action Time Measurement of the me-
A and B input channels of the calibrated chanical contacts
chronometer are set to be negative and The connection of the instruments is shown
positive respectively. in Fig. 4.
4.3.2.3 Time interval measurement of two 4.3.3.1 Duration Measurement of the clos-
positive electronic pulses ing for one contact
The time interval generator gives a single A single start signal (short circuit) given by
positive pulse signal in two traces. The the time interval generator, connected to
start-up signal is put on A channel of the the relevant measurement input terminal of
calibrated chronometer. The stop signal is the chronometer.
put on B channel. The trigger slopes of 4.3.3.2 Duration of the opening for a single
both channels are set to be positive . contact
4.3.2.4 Time interval measurement of two The single turn-off signal (open circuit)
negative electronic pulses given by the time interval generator is put
The time interval generator gives a single onto the relevant measurement input ter-
negative pulse signal in two lines. The minal of the chronometer.
start-up signal is put on A channel of the
4.3.3.3 Measurement of the time interval 5. Treatment of the calibration results and
for closing between two contacts calibration period
The time interval generator gives single
start signals as two outputs, one is the 5.1 Treatment of Calibration Results.
start-up signal and the other is the stop
signal. Both of them are connected to the For chronometers which meet the re-
relevant measurement input terminals of quirement of this working instruction, a cali-
the chronometer respectively. bration certificate will be supplied: For the
4.3.3.4 Measurement of the time interval ones which do not meet the requirement, an
for the turn-off of two contacts advice note of the calibration result should be
Single turn-off signals given by the time supplied with the items which do not comply
interval generator one after another are being pointed out.
connected to the relevant measurement
input terminals of the chronometer re- 5.2 Calibration period
spectively.
The calibration period of the chronometer
is one year
ITTC – Recommended 7.6-02-07
Procedures Page 9 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
APPENDICES
Appendix 2
The connection of the instruments is shown in
Calibration of the crystal oscillator stan- the following Fig.
dard if the chronometer does not have an
output of the crystal oscillator frequency
The functions of the chronometer are taken per hour, in total N times (N = 4, 8 or
measured by use of the electronic pulse. The 24). The frequency fluctuation value can be
time interval generator gives the pulse width of calculated by use of formulae (1):
the single pulse or the interval between two S = (yi)max –(yi)min (1)
single pulses.
Ti − T0
yi =
The given time interval may be supposed to T0
be T0. The time unit corresponding to the
where:
minimum display position of the chronometer
is τ0. Then the value of T0 must meet the fol- Ti measured value every time.
lowing formula:
2. Alignment of the frequency accuracy
τ0 1
≤ v If the result of the calculation using the data
T0 10 from the frequency fluctuation measurements
satisfies the following formula, the alignment
where:
of the crystal oscillator frequency is not
v frequency fluctuation of the internal needed.
crystal oscillator of the chronometer.
10 S ≥ yi max
≥S (2)
1. Calibration of the frequency fluctuation else: open the cover of the chronometer, ad-
After the chronometer has warmed up ac- just the frequency of the crystal oscillation till
cording to the working instruction the meas- it meets the requirement of formula (2). This
urement can be started. One measurement is time the value of |yi|max need only be calcu-
ITTC – Recommended 7.6-02-07
Procedures Page 11 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication
lated for one measured value after the align- |yi|max must be stated in the calibration report
ment. as the frequency accuracy.
Or
working instruction.
1. Calibration of the measuring error
The calibration values are selected as 1s, 2. Calibration of the daily difference
10s, 1min and 1h.
The daily difference means the measuring
Three measurements should be carried out error when the time interval is taken as one day.
for the first three measuring values. The arith- For the normal stopwatch no one will measure
metic mean of their measurement values may such a long time interval. But the crys-
be taken. For the last case, two measurements tal-oscillator stopwatch also posses the display
can be performed. Take the arithmetic mean function of hour, minute and second. So the
value of its measurement value. requirement of the index of the travel-time –
The measuring error can be calculated by “daily difference” similar as watch and crystal
use of the formula (5) in the main body of this clock is required.
ITTC – Recommended 7.6-02-07
Procedures Page 13 of 13
Sample Work Instructions
Effective Date Revision
Calibration of Chronometers with 2002 00
Digital Indication