2119701–100
Revision 3
do not duplicate
WARNING
X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR
UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x–ray beam becomes a source of
danger in the hands of the unauthorized or unqualified operator. Excessive exposure to x–radiation causes damage to human tissue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified persons from operating this equipment or exposing
themselves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with the Recommendations of the Interna-
tional Commission on Radiological Protection, contained in Annals Number 26 of the ICRP, and with applicable national standards.
ATENCION
LOS APARATOS DE RAYOS X SON PELIGROSOS PARA EL PACIENTE Y EL MANIPULADOR
CUANDO LAS NORMAS DE PROTECCION NO ESTAN OBSERVADAS
Aunque este aparato está construido según las normas de seguridad más estrictas, la radiación X constituye un peligro al ser manipulado
por personas no autorizadas o incompetentes. Una exposición excesiva a la radiación X puede causar daños al organismo.
Por consiguiente, se deberán tomar todas las precauciones necesarias para evitar que las personas incompetentes o no autorizadas
utilicen este aparato, lo que sería un peligro para los demás y para sí mismas.
Antes de efectuar las manipulaciones, las personas habilitadas y competentes en el uso de este aparato, deberán informarse sobre las
normas de protección fijadas por la Comisión Internacional de la Protección Radiológica, Anales No 26: Recomendaciónes de la Comi-
sión Internacional sobre la Protección Radiológica y normas nacionales.
ACHTUNG
RÖNTGENAPPARATE SIND EINE GEFAHR FÜR PATIENTEN SOWIE BEDIENUNGSPERSONAL,
WENN DIE GELTENDEN SICHERHEITSVORKEHRUNGEN NICHT GENAU BEACHTET WERDEN
Dieser Apparat entspricht in seiner Bauweise strengsten elektrischen und mechanischen Sichereitsnormen, doch in den Händen unbe-
fugter oder unqualifizierter Personen wird er zu einer Gefahrenquelle. Übermäßige Röntgenbestrahlung ist für den menschlichen Orga-
nismus schädlich.
Deswegen sind hinreichende Vorsichtsmaßnahmen erforderlich, um zu verhindern, daßunbefugte oder unqualifizierte Personen solche
Geräte bedienen oder sich selbst und andere Personen deren Bestrahlung aussetzen können.
Vor Inbetriebnahme dieses Apparats sollte sich das qualifizierte und befugte Bedienungspersonal mit den geltenden Kriterien für den ge-
fahrlosen Strahleneinsatz durch sorgfältiges Studium des Hefts Nr. 26 der Internationalen Kommission für Strahlenschutz (ICRP) vertraut
machen: Empfehlungen der Internationalen Kommission für Strahlenschutz und anderer nationaler Normenbehörden.
WARNING
VOLTAGE PRESENT
TABLE OF CONTENTS
'%3*/.
)#2' '#241' #.& /--43#3*/. /#1&
1–1 Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1–2 Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1–3 Test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1–4 Indicator lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
'%3*/.
)#2' ! *5*&'1 /#1&
2–1 Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
'%3*/.
)#2' 4'.%) /#1&
3–1 Test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
'%3*/.
)#2' "
4–1 Indicator lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4–2 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4–3 Test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
'%3*/.
)#2' !
5–1 INDICATOR LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5–2 FUSES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
'%3*/.
6–1 INDICATOR LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6–2 FUSES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
'%3*/.
)#2' '(4,#3*/. /#1&
7–1 INDICATOR LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7–2 SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7–3 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
i
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
" ,%('
$+"
(',*(& (*!
8–1 INDICATOR LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8–2 SWITCH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8–3 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
" ,%('
$+"
" -*%,. (*!
9–1 SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9–2 JUMPERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9–3 INDICATOR LIGHTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9–4 Test points (TP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
" ,%('
$+"
',"*# " (*!
10–1 SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
10–2 ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
10–3 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
" ,%('
$+"
',"*# " (*!
'! *!
11–1 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
11–2 ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
11–3 LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
" ,%('
$+"
(*!
12–1 SWITCHES AND JUMPERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
12–2 LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
" ,%('
$+"
","* +-))&. (*!
13–1 LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
13–2 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
ii
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
!,%('
$+! -/%&%*0 (&,#! (*
14–1 Indicator lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
!,%('
$+! *% %+ ',!*"! (*
15–1 ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
15–2 LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
15–3 TEST POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
!,%('
!,%('
$+!
!,%('
$+!
iii
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
iv
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
MECHANICAL OR OTHER HAZARDS.
v
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
vi
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
REVISION HISTORY
CENTRAL LISTINGS
LIST OF EFFECTIVE PAGES
Title page 1
Safety Instructions 1
i thru viii 1
1 thru 64 1
vii
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
viii
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
1
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
X11
X5
X9
2
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
calibration
3
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
Measurement scale
DS11 DS7 DS5 mA range
on XJ6 (BNC)
0 0 1 10 to 80 mA 20 mA for 1 V
1 0 0 0 to 20 mA (fluoro) 2 mA for 1 V
0 0 0 Tube 1
1 0 0 Tube 2
1 1 0 Tube 3
1 1 1 Tube 4
1 = LED lit
4
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
5
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
P1 kV measure + D3
Adjust voltage ratio and
P2 overshoot adjustment
kV measure – D2 (1 V = 8 kV)
6
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
7
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
8
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
0V
V-10
10 V
TP4 Cathode voltage B4 Analog F2
9
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
10
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
11
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
F1
F2
F3
F5
F4 N1
F6
N2
F7
F8
F9
N3
F10
TP1 TP5
TP2 F11
TP4
TP3
N4
N5 DE1
N5 DE2 DE3 DE4
12
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
DE3 Green 3-B7 Line Supply present A7
N2 Amber 1-K11 Ø2 C5
3Ćphase
N3 Amber 1-K11 Ø1 voltages B4
present
N4 Amber 1-L11 Ø3 B4
F3 Not used
F9 Not used
13
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
14
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
15
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
N2
CENTRAL LISTINGS
N2 Amber DC POWER K2
16
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
17
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
E
CENTRAL LISTINGS
61
D
52
43
C
34
30
25
B
21
17
11
A
6
SLOT:
1 2 3 4
18
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
N 15 V Green 1-E3 - 15 V ON (from LV Supply) E4
P5V Green 1-E5 + 5 V ON (from LV Supply) E4
Neon NE1 Amber 5-E6 200 V AC OK (for auxiliary voltage board) E3
19
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
20
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
21
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
C
D
B
E
10
CENTRAL LISTINGS
9
8
7
6
5
4
3
2
1
22
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
KV75 Green 4-B7 kV MEAS > 75 % kV REF A5
KVTH Green 4-B7 kV MEAS > kV THRESHOLD A5
CAL Yellow 4-B7 CALIBRATION MODE A6
REG Yellow 4-B7 Non Linear regulation A6
SYNC Green 4-B7 Exposure started A6
TST1 Yellow 4-B7 Synchro simulation (diagnostic) A6
OPNL Yellow 4-B7 Open Loop Control A6
23
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
24
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
25
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
26
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
RX Yellow 1-G6 X-Ray Exposure Control A2
27
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
28
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
29
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
C
D
B
E
10
9
CENTRAL LISTINGS
8
7
6
ABI2
5
ABI1
4
3
2
1
30
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
Switch on position 1: Position 1 is used for
+ 300 V is always OFF troubleshooting
31
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
32
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
I1MX Current max in 3-B6 DC 5V
inverter 1
TP15 + 15 V ON 6-F3 DC + 15 V
TN15 - 15 V ON 6-F1 DC - 15 V
33
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
34
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
35
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
D_GND
CENTRAL LISTINGS
MPPU MPG
MPPU ONLY
36
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
37
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
38
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
39
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
ILLUSTRATION 1–1
MP PHASE 41 A5 A6 INT/2 BOARD LAYOUT
CENTRAL LISTINGS
40
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
TP24 _KV>75%_KVREF 3-F6 B/CĆ6/7
TP25 _GRID_CTRL 3-F6 B/CĆ6/7
TP26 _F_MAX_EN 3-F7 B/CĆ6/7
TP28 _BIAS_EN 3-F7 B/CĆ6/7
TP29 _GRID_EN 3-F7 B/CĆ6/7
TP30 _GRID_PULSE_CTRL 3-E7 B/CĆ6/7
TP31 KV_ACQ_EN 3-F7 B/CĆ6/7
TP32 XRAY_EXP_CONTROL 3-E7 B/CĆ6/7
TP33 _EXP_75 3-E7 B/CĆ6/7
TP34 _PLSDSYS 3-D8 B/CĆ6/7
TP42 _MB_EN 3-C9 B/CĆ6/7
TP51 _MB_REF 3-B8 B/CĆ6/7
TP52 For factory use sheet 3 B/CĆ6/7
TP53 _MB_CDE 3-A8 B/CĆ6/7
TP66 QUENCH CTRL 3-B5 B/CĆ6/7
TP67 GRID_PULSE_OK 3-B5 B/CĆ6/7
TP68 mA Demand 5-B4 DC 10 V A6
TP69 Bias Feedback 8-D6 DC 10 V C2
TP70 Bias Demand 8-B7 DC 10 V B2
TP71 For factory use 9-D3 A4
TP72 ADC input 9-B5 10 V A4
TP73 For factory use 2-E7 C3
P5V + 5 V Supply 11-C4 DC +5 V A7
GND 0 / 15 V 10-F6 DC 0V A7
P15V + 15 V Supply 11-E3 DC +15 V A5
MAF mA Feedback 5-C7 D1
N15V -15 V 11-F3 DC -15 V B2
41
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
42
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
43
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
44
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
SW3 Red PB Schematic RESET
1 of 6 Hardware and firmware
D5 Red
D4 to D11
D6 Red Lit successively during
application (Normal mode)
D7 Red
Schematic
D8 Red 2 of 6
Different meanings in
D9 Red Diagnostic mode
(See direction 19 121)
D10 Red
D11 normally flashes in
D11 Red application.
45
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
46
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
47
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
48
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
MA15 Green 3-E2 -15 V ON A7
CENTRAL LISTINGS
PRES Yellow 1-A3 Ich >1 A rms A8
MOV 0 / ± 15 V 3-F2 DC 0V A6
FREQ Frequency running 1-C4 Pulse 15 kHz < Freq < 60 kHz B8
49
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
50
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
51
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
P12V
P24V
P200
52
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
53
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
54
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
55
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
DS1-4
CENTRAL LISTINGS
VR1
DS5-8
56
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
57
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
58
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
Position the voltage selector jumper (10) according to the line input (110 V/220 V) to select
220 V at the LV supply input. See Illustration 15.
Adjust the Output Voltage Adjust potentiometer (12) to obtain 5.1 V "1% (dc) at TP5V
(reference 0 VL) on the Interface 2 Board MP Phase 41 A5 A6.
" ! " ! ! ! ! ! # ! !
CENTRAL LISTINGS
Remember to move JUMPER 10 to select the correct
input voltage (200–240 V) for the LV Supply.
11
12
1
1 2
2 2 M4 x 8 3
Mounting Hole
3 4
4 5
93
5 6
6 7 10
7 7 8
9 9 9
10
205
223
13
2 M4 x 8
Mounting Hole
* Dimensions in mm
* Cover is optional
1 +5 V Output terminal 8 NC
2 +5 V GND terminal 9 AC (N) Input terminal
3 +12 V (+15 V) Output terminal 10 100/200 V Input voltage selector
4 ±12 V (±15 V) GND terminal (Short 100-120 V/Open 200-240 V)
11 Output indicator
5 –12 V (–15 V) Output terminal
12 Output voltage adjust
6 FG Signal Ground
13 Protection Ground
7 AC (L) Input terminal
59
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
60
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
Plug from stabilizer transformer XJ501 of MP Phase 41 A6 to connector XJ3 of Power Unit
CENTRAL LISTINGS
Electronics MP Phase 41 A5:
Cable W132 marked “50Hz” for 50Hz line frequency.
Cable W133 marked “60Hz” for 60Hz line frequency.
61
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
62
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
Each strap to be installed in the appropriate one of its seven possible positions, according to
the nominal line voltage (360, 380, 400, 420, 440, 460, or 480V).
CENTRAL LISTINGS
63
MP Phase 4 + Vascular Option Generator
GE Medical Systems Central Listings
REV 3 sm 2119701–100
CENTRAL LISTINGS
64