Phased Array

Collection Hardware Set-up


Sonatest VEO

Collection Hardware set-up

 This presentation provides an explanation of the setup


procedure for a Phased Array inspection with the Sonatest
Veo

 This scenario starts with a pre-configured example


configuration file, which minimizes the number of
parameters to modify, it is possible to start from any type
of setup and reconfigure parameters

09 Phased Array Collection Hardware Set-up VEO


Rev0 1
Collection Hardware set-up
- Press the power button to boot the unit

- Press the key to select the file list, use the click wheel
to navigate

- Select the file which matches the probe and wedge you are using

- Press to open the selected file

- With the selected pre-configured configuration file, parameters of the


‘Probe’ and ‘Wedge’ menus do not need to be modified

Collection Hardware set-up


- Go to the ‘Part’ menu and adjust the parameters according to the sample:
- Material
- Thickness
- Sound velocities
- Weld geometry

- While adjusting the parameters, confirm with the 3D view that they match
the weld preparation.

09 Phased Array Collection Hardware Set-up VEO


Rev0 2
Collection Hardware set-up
- If a parameter turns red, it means the weld geometry is not valid and a
parameter needs to be adjusted

Invalid
Geometry

Top Right
Width is
too wide

Collection Hardware set-up


- Go to the ‘Scan’ menu and adjust the parameters accordingly:
- Focalisation
- Focal distance
- Angle Resolution
- Start and Stop angle
- Start and Range path
- Number of Elements (First, Last, Tx, Rx)

- In the ‘Geometry’ menu set the Index offset distance, which positions the
front of the wedge in reference to weld center.

09 Phased Array Collection Hardware Set-up VEO


Rev0 3
Collection Hardware set-up
- Go to the ‘View’ menu and set the
number of skips to see if the scan
covers the whole weld volume
1st skip

2nd skip

- If you use an encoder, go to the


‘Encoding’ menu and select ‘Scan
Axis Only’ to activate the encoder

Collection Hardware set-up


- Once your Set-up is ready, press , to start ‘acquisition mode’.

- Press and select the prefered lay-out

- Go to the ‘Calibration menu’ and perform the calibrations. If you use a


multi-scan set-up, each scan must be calibrated independently.

09 Phased Array Collection Hardware Set-up VEO


Rev0 4
Collection Hardware set-up
- After finishing each calibration wizard the appropiate indicator in the S-
scan image will turn to green.

- Save the set-up, including calibrations, press


Select ‘Setup’ in the Save Options window and save the file.

Collection Hardware set-up


- Check the movement direction of the encoder:
Move the encoder in the scanning direction
and confirm scan position (bottom left
corner) is going up with positive number.
If not, invert the scanning direction in the
‘Encoding’ menu

- Place the probe on the right start position


(Scan start position and offset position)

- When ready to go, you can directly press as the encoder position will
be automatically set to the ‘Scan Start Pos.’ value.

- Once the area has been scanned, press and save the recorded file.
It will be saved with a .utdata file extension.

10

09 Phased Array Collection Hardware Set-up VEO


Rev0 5
Collection Hardware set-up
- To review the data, press to recall your data file.

- Press to see the data

- To create, during analysis, a report or screen capture, press and


select ‘Report’ or ‘Screen Capture’ and enter the file name.