Documente Academic
Documente Profesional
Documente Cultură
Cardiovascular X-ray
Imaging: Physics,
Equipment,
and Techniques
Arnold R. Cowen
References – 195
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
149 11
11.1 Introduction in cardiology examinations is often referred to as
digital cine acquisition.
Cardiovascular diagnosis and interventional This chapter aims to provide the practicing
therapy have a fundamental need for dynamic interventionalist with an update on the physical
X-ray images of suitable quality. These images and technical aspects of prevailing cardiovascular
are typically used to visualize blood flow, in X-ray imaging equipment. Special attention is
order to diagnose diseased vessel segments, ste- paid to the new generation of (solid-state) image
noses, aneurysms, etc. It is an unhelpful (but detectors, their physical characteristics, and how
currently unavoidable) characteristic of standard- they can augment clinical service. The aim is to
projection imaging with X-rays that blood is provide the interventionalist with a cogent under-
indistinguishable from the soft-tissue structures standing of these influential technologies. Such
in which the vessel is embedded. Therefore, visu- a working knowledge can have many benefits. A
alization of a blood vessel demands the admin- better understanding of these physical and tech-
istration of an artificial contrast agent, normally nical principles should lead to fuller utilization of
an iodine-loaded solution. The contrast medium equipment resources and facilities. For example,
is injected into the vessel to mix with the blood the interventionalist should be empowered to
and provide the absorption of the X-ray beam optimize fluoroscopic and image acquisition strat-
needed to depict the vessel lumen and therefore egies. It should also add to a better appreciation of
any associated disease. Furthermore X-ray image how equipment utilization impacts on radiation
guidance and visualization are indispensable safety, potentially to the benefit of both patients
in current interventional practice. Therefore, and staff. This knowledge may also contribute to
dynamic X-ray imaging is required to image better-informed selection of digital X-ray imag-
devices such as guide wires, catheters, balloons, ing equipment during the purchase process. This
stents, coils, cutters, or any other items on a should aid in future-proofing the selection of
burgeoning list of interventional tools. In order equipment against continuing advances in digital
to minimize any consequent radiation risks, the imaging technologies and new developments in
imaging must be performed at dose levels con- interventional techniques, which will no doubt
sistent with the ALARA (as low as reasonably require even more sophisticated X-ray image
achievable) principle, at all times. guidance.
Real-time dynamic X-ray imaging at low
dose rates, typically for the purpose of guid-
ing a clinical procedure, is known as fluoros- 11.2 Reprise of X-Ray Physics
copy. Historically, in this role, fluoroscopy has
been accepted to be an impermanent imaging Before considering cardiovascular imaging equip-
medium, requiring no storage of images for later ment in detail, this section reprises the basic phys-
review. The permanent recording of images—for ics of X-rays and their production for medical
example, to document clinical findings—has nor- use. Building upon this, the sections that follow
mally relied upon the acquisition (and storage) of go on to explore how X-rays interact with matter
dynamic image sequences of superior quality to (tissue) and in turn how the basic X-ray image is
fluoroscopy. Image acquisition in this form there- formed and then recorded using a suitable detec-
fore has to be performed at significantly higher tor. For further background, the interested reader
radiation dose rates than in fluoroscopy to ensure is referred to the many excellent textbooks on the
that this is achieved. At various times in the past, physics of medical X-ray imaging (e.g., [1, 2]).
serial images were recorded on radiographic film,
small-format film, cine film, or even video tape (as
clinically appropriate). Today, all dynamic X-ray 11.2.1 What Are X-Rays?
images, both in fluoroscopy and serial acquisition,
are acquired in digital form using computerized To cut a long story short, X-rays are a form of
X-ray imaging equipment. Hence, terms such as electromagnetic radiation, like many other more
digital fluoroscopy and digital acquisition are now familiar physical phenomena such as visible
used to differentiate these dynamic X-ray imaging light, ultraviolet light, infrared radiation, micro-
modalities. The high-speed acquisition of images waves, and radio waves. X-rays are subject to the
150 A.R. Cowen
vagaries of quantum physics and therefore exhibit in cardiovascular X-ray imaging; in practice,
the properties of (electromagnetic) waves while tungsten is alloyed with a small quantity (~5%)
at the same time behaving like particles. To date, of rhenium (Re) to increase target longevity.
the potentially very effectual wave-like properties Free electrons are derived by heating a filament
of X-rays [3] are unrealized in clinical routine coil (again made of tungsten) by passing an elec-
and therefore are not discussed further here. The tric current through it; this process is known as
X-ray beam used in medical X-ray imaging is bet- thermionic emission. The number of electrons
ter described as a stream (deluge) of individual freed and therefore the intensity of the X-ray
packets of X-ray energy or X-ray photons. If the beam produced are controlled by adjusting the
temporal frequency or wavelength of the X-ray is filament current and thereby the filament temper-
known, the equivalent X-ray photon energy (εp) ature. The intensity of the X-ray beam is defined
can be calculated using the simple equation: by the X-ray energy per unit area and unit time
passing a defined plane in the beam. The applica-
ep = h × f = h × c / l tion of a suitably high voltage (HV) between the
(−) cathode filament and the (+) tungsten anode
where: target, typically in the range of a 50–150 kilovolt
f is the temporal frequency (in cycles s−1) (kV) peak (kVp), accelerates the electrons up to a
λ is the wavelength kinetic energy suitable for producing X-rays. The
h is Planck’s constant (~6.626 × 10−34 J s) peak value of the accelerating voltage (kVp) that
c is the speed of light (3 × 108 m s−1 in vacuo) is applied controls the energy content of the X-ray
X-rays share many physical properties with beam and therefore its propensity to penetrate the
the electromagnetic phenomena listed above, but thickness of the body. At the same time, however,
have one notable distinguishing feature: X-rays kVp also affects the intensity of the X-ray beam.
have much greater photon energy. An X-ray pho- The electrical current that flows from the cathode
ton is typically several tens of thousands of times to the anode as a result of applying the accelerat-
11 more energetic than a light photon. Therefore, ing voltage is the tube current in milliamps (mA).
unlike light, an X-ray photon has the potential to For a given kVp, the X-ray beam intensity is pro-
penetrate a comparatively large thickness of mat- portional to the mA. Electrons accelerated across
ter, such as the human body. On the other hand, the X-ray tube collide with tungsten atoms on the
the X-ray photon also has the capacity to interact anode surface. X-rays are produced via one of two
with an atom that makes up the tissue, via one of physical mechanisms: bremsstrahlung and char-
the mechanisms described in 7 Sect. 11.3. When
acteristic X-ray production. Production of X-rays
such an interaction occurs, the X-ray photon by these two mechanisms is signally inefficient,
delivers whole or part of its energy to the tissue. viz., <1% of the electrical energy used is converted
X-ray photons (like the similarly energetic γ-rays) to X-rays. The balance of electrical energy (>99%)
have sufficient energy to ionize an atom; there- is converted unproductively to heat. The man-
fore, not surprisingly, X-rays are said to be ioniz- agement and disposal of heat remains one of the
ing radiation. Absorption of X-ray energy in body main constraints of modern X-ray tube design,
tissue results in the ionization of biomolecules, and this is discussed in 7 Sect. 11.7.1.
which can result in cell damage or death. The Bremsstrahlung production occurs when a
resultant biological risk increases with the mag- high-speed electron collides with the strong (+)
nitude of the X-ray dose delivered (and hence the coulomb field surrounding the nucleus of a target
degree of ionization produced). Radiation dose atom. The energy of the resulting X-ray photon
and its measurement are discussed in 7 Chap. 15.
depends upon the proximity of the collision to the
nucleus itself (viz., the local strength of the elec-
tric field). A consequence of the bremsstrahlung
11.2.2 How Are X-Rays Produced? mechanism is that during an exposure, a wide
range of X-ray photon energies is produced.
In medical imaging, X-rays are produced by Characteristic X-ray production occurs when
bombarding a high-atomic-number (Z) metal an incoming electron collides with a K-shell
target with a stream of high-speed electrons. electron; the K-shell contains the electrons
Tungsten (W) (Z = 74) is the target metal used that are most tightly bound to the nucleus. The
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
151 11
resulting ejection of a K-shell electron produces
a highly unstable atom. This is promptly followed ×104
16
by the transition of an electron from an L- or
14 60 kVp
M-shell to fill the K-shell vacancy. As a result, the
The physical characteristics of an X-ray beam can high-energy X-ray photons, making the X-ray
be conveniently described as a spectrum, which beam more penetrative.
represents the intensity of the beam at each con-
tributing X-ray photon energy. The exact form
of the X-ray spectrum is principally dependent 11.2.4 Rotating Anode X-Ray Tube
upon the electrical characteristics of the expo-
sure, the tube current (mA) and the peak tube The production of X-rays takes place inside an
voltage (kVp) settings. A change in mA produces evacuated metal (previously glass) enclosure,
a proportionate change in the intensity (height) which houses the cathode filament (in its focusing
of the X-ray spectrum, but otherwise, no detailed block), the rotating anode, and insulating fixtures.
change in spectral shape occurs. On the other In metal enclosures, the insulating fixtures are
hand, an increase in kVp has a more powerful often made of ceramic material. The anode target
effect on the X-ray spectrum, materially chang- comprises a disc-shaped metal structure attached
ing both the intensity of the X-ray beam and its to a rotor assembly. Around the anode disc is the
energy content (hence, spectral shape). The maxi- circumferential track, which is bombarded with
mum photon energy of the X-ray spectrum is electrons during the X-ray exposure. The cath-
defined by the kVp applied across the X-ray tube. ode block is mounted off-axis to accurately align
Spectra produced by a tungsten anode X-ray the filament with this track. Anode disks have a
tube are polyenergetic, as shown in . Fig. 11.1;
multilayer (or composite) construction, with the
this figure includes three computed X-ray spec- actual tungsten (target) layer superimposed upon
tra for exposures at 60 kVp (red), 80 kVp (blue), a molybdenum (Mo) alloy substrate. The anode
and 100 kVp (green), respectively. The 60 kVp disk is often backed by a thick layer of graphite
spectrum describes a continuous bremsstrahlung to increase the heat storage capacity and aid heat
envelope up to a maximum photon energy of dissipation via thermal radiation. During an
60 keV. An exposure at 80 kVp produces a spec- X-ray exposure, the anode is typically rotating
trum describing a broader bremsstrahlung enve- at 3000 revolutions per minute (but this rate can
lope up to a maximum photon energy of 80 keV; be significantly greater). Some anodes contain a
superimposed upon this are the X-ray emission set of narrow radial slots to condition the anode
lines of tungsten (at their characteristic energies). against the mechanical stress of repeated heating
By comparison, the 100 kVp spectrum contains and cooling cycles. A beryllium (Be) window is
X-ray photons up to the higher maximum energy built into the metal envelope to provide a low-
of 100 keV. This increases the complement of attenuation exit port for the X-ray beam. The
152 A.R. Cowen
X-ray tube is housed inside an earthed metal power delivery will overheat the anode surface
shield and surrounded by oil. The oil not only acts and cause permanent damage. X-ray tubes typi-
as an electrical insulator but also aids heat trans- cally offer two or three X-ray focal spot selections,
fer from the tube to the shield via convection. each with a different size and power capability. In
Except for the region adjacent to the exit port of a bifocus X-ray tube, the smaller fine focus typi-
the X-ray tube, the shield is lined with lead (Pb) cally has a size in the range of 0.3–0.5 mm, while
foil to prevent leakage of X-rays. Heat sensors are the larger more powerful broad focus is typically
often used to monitor the temperature of the oil to 0.6–1 mm in size. Associated with each focal
ensure safe operation of the X-ray tube. spot size selection is a filament of matched scale,
During the X-ray exposure, electrons flow smaller-gauge filaments being used with the fine
across the tube and impact upon the (rotating) focus and more robust filaments with the broad
anode. The instantaneous impact of electrons focus. The fine focus is commonly reserved for use
occurs over a rectangular zone on the anode in fluoroscopy applications, which operate at lower
target known as the actual X-ray focus (or line levels of X-ray tube current. For adult patients, the
X-ray focus); this represents the nascent source of X-ray broad focus is normally reserved for use in
X-rays. With respect to the emerging beam, the serial image acquisition. However, the fine focus
projected line X-ray focus is foreshortened by the may be used in pediatric image acquisition. Some
angled surface of the anode target. As a result, the modern cardiovascular X-ray imaging systems
effective focus or focal spot, which contributes to automatically select the optimum X-ray focal spot
X-ray image formation, has (roughly) a square size, depending upon the patient size, exposure
format. The anode target is typically beveled at an requirements, power limits, etc.
angle of between 8° and 12°. A shallower anode
angle provides greater foreshortening of the line
focus, and hence a smaller effective focal spot is 11.3 Interaction of X-Rays
realized. At the same time, however, the shallow with Matter
11 anode angle limits the available X-ray field cover-
age. In cardiac imaging, which requires restricted 11.3.1 Physics of X-Ray Interactions
anatomical coverage, this is usually not a prob-
lem. On the other hand, vascular imaging often Individual X-ray photons that are incident upon
requires coverage of quite large body sections. the patient during an X-ray exposure can subse-
Therefore, in general, vascular applications favor quently undergo a variety of different interaction
the use of a larger anode target angle to increase histories. It should be noted that the mechanisms
field coverage. To a first approximation at a given defining these interactions are probabilistic in
distance from the X-ray focal spot, the field of nature. For example, a small but important com-
X-rays incident upon the patient is approximately plement of the incident X-ray photons can travel
uniform. In practice, there is a 20–30% reduction unperturbed through a thickness (of even many
in X-ray intensity across the beam in the cathode– centimeters) of tissue. These photons exit the
anode direction due to the anode heel effect. This patient and contribute to the primary X-ray image,
effect is ignored here, as modern image detectors which is recorded by the image detector. Much
can compensate for this. more commonly, an X-ray photon will interact
The size and shape of the effective focal spot with an atom in the tissue and be removed from
influences the spatial resolution of the X-ray the X-ray beam either completely or partially.
image that is formed. The finite size of the X-ray Over the range of energies used in cardiovascular
source potentially represents a source of geomet- imaging, X-ray photons can interact via one of
rical blur (unsharpness) in the recorded images. two mechanisms: photoelectric absorption (PA)
Ideally for sharp (high-resolution) X-ray images, and Compton scatter (CS), which have distinc-
the effective focal spot should be as small as possi- tive properties. The combined effect of these two
ble. Unfortunately, there is a limit to the degree to interaction mechanisms reduces the intensity of
which the focal spot can be reduced in area. This the X-ray beam as it penetrates increasing thick-
limit is defined by the maximum instantaneous nesses of tissue. This process is known as X-ray
power (defined by the product P = kVp × mA) beam attenuation. The attenuation produced
that can be tolerated by the anode target. Excessive by a material is described by the linear X-ray
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
153 11
attenuation coefficient (μ). The greater the value For most elements, the ratio Z/A is approximately
of μ, the more strongly an X-ray beam will be constant, having a value of ~0.5. Uniquely, how-
attenuated by a given thickness of material. ever, hydrogen has a Z/A value of unity; this
enhances the sensitivity of the CS mechanism to
hydrogen-loaded materials, notably water and
11.3.2 Photoelectric Absorption hence soft tissue. Indeed, CS is the dominant
X-ray interaction process in soft tissue, but is
An interaction via the PA mechanism occurs less influential in bone. X-ray attenuation due to
when an incoming X-ray photon is fully absorbed CS falls much less rapidly with increasing pho-
by a tightly bound orbital electron in an atom. ton energy than for PA, and therefore CS tends
This results in the ejection of a photoelectron, to dominate X-ray image formation at increas-
which travels through adjacent tissue, depositing ing photon energies. The dynamics of the CS
its kinetic energy via ionization. The attenuation process are governed by the laws of momentum
of an X-ray beam due to the PA interaction (μPA) and energy conservation, but these details need
strongly increases with the mass density (ρ) and not trouble the reader. It is important however to
the atomic number (Z) of the attenuating mate- note that the secondary X-ray photon is emitted
rial, but falls rapidly with increasing X-ray photon in an arbitrary (random) direction with respect
energy (εp), viz.: to the primary photon and is said to be scat-
tered. This X-ray photon can be scattered at any
m PA µ r × Z 3 / e p3 angle, from a slight deflection through to full
back scatter (viz., over 180°). Again, subsequent
The PA interaction mechanism contributes absorption of the scattered photon in tissue will
strongly to X-ray beam attenuation for tissue contribute to the patient dose. Should the scat-
material with a higher Z value, such as bone (typi- tered X-ray photon exit the patient altogether, it
cal mean Z = ~14), but particularly so for iodine may then go on to irradiate clinical staff who are
(Z = 53), which is used in most cardiovascular in close proximity.
X-ray contrast media. By comparison, the atomic
number of soft tissue is much lower (typical mean
Z = ~7). At low X-ray photon energies, PA is the 11.3.4 Beer–Lambert’s Law
dominant interaction mechanism, but its influ- (Exponential Attenuation)
ence falls rapidly with increasing energy. The
absorption of X-ray photon energy in tissue via The physical description of the attenuation of an
the PA mechanism is a significant contributor of X-ray beam by tissue can be more easily appreci-
radiation dose to the patient. ated by first considering a simple phantom model,
rather than the much more complex problem
of a patient. This phantom model assumes that
11.3.3 Compton Scatter parallel ray geometry is applicable (viz., diver-
gence of the X-ray beam does not occur), that
An interaction via the CS mechanism occurs the X-ray beam is monoenergetic, and that X-ray
when the incoming photon collides with a loosely scatter can be ignored. Making these simplify-
bound (peripheral orbital) electron in an atom. ing assumptions unburdens the analysis but still
This causes the electron to be ejected from the provides some useful insights. Such a phantom
atom, accompanied by the emission of a second- model (see . Fig. 11.2) typically comprises a
ary X-ray photon of lower energy. Attenuation given thickness Τ of tissue of linear X-ray attenu-
of an X-ray beam via the CS mechanism (μCS) ation μT (where μT results from the sum of the PA
increases with the tissue mass density and the and CS interactions, therefore μT = μPA + μCS). If
ratio of the atomic number (Z) to the atomic the intensity of the X-ray beam incident on the
weight (A) of the tissue, but falls (slowly) with phantom is I, the resulting exit beam intensity IΤ
X-ray photon energy, viz.: is predicted by Beer–Lambert’s law, viz.:
mCS µ r × ( Z / A ) / e pZ I T = I exp - ( mT × T )
154 A.R. Cowen
of ~0.22 cm−1; however, the value does vary result in a featureless (uniform) X-ray image. The
with X-ray beam energy. Exponential attenua- analysis of X-ray image formation becomes much
tion describes a nonlinear function, where equal more informative if a signal detail is included in
increments in thickness of a material produce the phantom. In other words, a specific signal with
disproportionate reductions in the intensity of the different X-ray attenuation characteristics to the
transmitted X-ray beam. background is considered.
Rearranging this equation shows that if the val-
ues of I and μT are known, then the effective thickness
of the body section can be estimated by measuring 11.4.1 Iodine Signal
the transmitted X-ray intensity (IΤ) where:
In cardiovascular imaging, a signal is conve-
T = ( log e I / I T ) / mT niently represented by a blood vessel opacified
with iodinated contrast medium. The presence
This result can be used to estimate the equivalent of the contrast medium causes stronger attenu-
water thickness of a body section. Computation of ation of the X-ray beam than the background
this parameter is often used by fluoroscopy auto- tissue alone. The X-ray attenuation properties of
matic dose rate control systems to select X-ray iodine, soft tissue, blood, and bone are compared
exposure factors. across the relevant X-ray photon energy range
Another useful parameter that can be derived in . Fig. 11.3. Note that here the attenuation is
from Beer–Lambert’s law is the half-value thick- expressed in terms of the mass X-ray attenua-
ness (HVT) of a material. The HVT of tissue (the tion coefficient (μ/ρ) (which is equal to the linear
thickness required to attenuate the X-ray beam by X-ray attenuation (μ) divided by the mass den-
a factor of 2) can be estimated using the equation: sity (ρ)). Referring to this figure, we see that the
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
155 11
I T = I exp - ( mT × T )
Mass X-ray Attenuation Coefficient [m/p]
100
and in the shadow of the iodinated vessel:
33.2 keV
Iodine I C = I exp - ( mT × [T - d ] + mC × d )
10
Soft = I exp - ( mT × T + [ mC - mT ] × d )
Tissue
0.1 CX = ( I C - I T ) / I T
50 100
X-ray Photon Energy [keV]
Substituting for IC and IT from above gives:
.. Fig. 11.4 Phantom
model including a signal Entrance X-ray Intensity I
feature (blood vessel) Blood Vessel Containing
Iodinated Contrast Medium
d
T
11 IT
X-ray Intensity Profile
IT
equation that takes account of the properties of f is the fractional contrast medium content—
the contrast medium: this will vary with the patient’s condition, clinical
technique, injection site, and target vessel
mC = ( m / r )I × X I × f Substituting for μC and μΤ in the equation for
subject X-ray contrast yields:
where:
(μ/ρ)I describes the mass X-ray attenuation CX ~ - éë( m / r )I •X I • f - mT ùû•d
of iodine—this factor varies with X-ray beam
energy, but typically has a value of ~15 cm2 gm−1 If we assume that f = 1/5, then this equation pre-
ΧI defines the iodine concentration of the con- dicts that a contrast medium–enhanced vessel
trast medium—this varies with design and manu- 1.5 mm in diameter will yield a subject X-ray con-
facture, but 300 mg iodine ml−1 is representative trast of ~0.1 (viz., ~10%).
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
157 11
11.4.3 -Ray Scatter and Contrast
X of rejecting scatter includes the introduction of an
Degradation air gap between the patient and the image detec-
tor. This allows (some of) the X-ray scatter to
The description of subject X-ray contrast devel- bypass the image detector. The air gap technique
oped here has focused upon the primary X-ray is commonly used in pediatric imaging.
beam and ignored the influence of scattered X-rays.
X-ray scatter, which is an inevitable consequence
of Compton interactions particularly in soft tissue, 11.4.4 Antiscatter Grid
produces a fog of X-rays across the image field.
These scattered X-rays are geometrically uncor- The scatter rejection technique most commonly
related with the body structures and contribute used in adult cardiovascular imaging is the X-ray
no useful information to the image. In practice, antiscatter grid. The grid is positioned after the
the primary subject X-ray contrast is reduced by patient and mounted adjacent to the image detec-
the scattered X-rays, and hence the quality of the tor. A grid comprises an array of thin, parallel slats
recorded image is degraded. The contribution of of strongly absorptive material, typically lead,
X-ray scatter can be quantified in terms of the interleaved with thin spacers made of a radiolu-
scatter-to-primary ratio (SPR), viz., the ratio of the cent material, such as fiber or aluminum (Al). The
intensities of the scatter and primary content of grid is covered with a low-attenuation material
the X-ray beam. SPR varies with a number of char- (e.g., carbon fiber). The grid preferentially allows
acteristics of the X-ray image formation process. primary X-ray photons to pass between the lead
Important factors include the area of the X-ray slats and therefore to contribute to the recorded
beam entering the patient and the thickness of the image. Scattered X-rays, on the other hand, which
body section, essentially the volume of tissue irra- are deflected at an arbitrary angle, have a much
diated. SPR also increases with increasing X-ray greater probability of striking a lead slat and
beam energy (principally the kVp). The reduction hence being removed from the beam.
in subject X-ray contrast produced by scatter is The degree of scatter rejection by a grid is
described by the simple equation: described by the grid selectivity:
CS = CX • [1 + SPR ]
-1 SG = TG {P} / TG {S } > unity
accepted practice in pediatric imaging, but may the source. Associated with the divergence of the
also have a role in adult procedures [4]. Larger X-ray beam is a reduction in intensity, which is
savings in dose would accrue if the grid could be described by the familiar inverse square law, viz.:
removed during examinations of larger patients.
Unfortunately, this benefit comes at a cost of a Intensity µ 1 / D 2
marked deterioration in subject X-ray contrast
(and hence image quality). This problem can be Doubling the distance of the X-ray source to
ameliorated somewhat if an air gap is used to par- the imaging plane by a factor of 2 decreases the
tially offset the loss of the grid selectivity. incident X-ray beam intensity by a factor of 4
(viz., 22). The (attenuating) object intercepts the
X-ray beam and produces a projection image of
11.4.5 Imaging Geometry the object in the plane of the detector.
From the standpoint of X-ray physics, the
The geometry used in standard-projection X-ray body section essentially represents a three-
imaging is depicted schematically in . Fig. 11.5.
dimensional (3D) framework of attenuating
The X-ray beam emanates from a quasi-point structures. A diverging cone beam of radiation
source of radiation, and the resulting field of X-ray is used to address these structures and form the
photons diverges with increasing distance from basic X-ray image. The image is typically recorded
.. Fig. 11.5 Schematic
diagram of the geometry X-ray Source
used in standard-projec-
tion X-ray imaging
11 X-ray
Beam
SOD
SID
Object
X-ray
Penumbra
(Unsharpness)
Projected Object
Image
Detector
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
159 11
as a two-dimensional (2D) projection of the X-ray image detectors also have a limiting spatial
patient volume on an area detector. As a result, resolution. Factors affecting the spatial resolution
structures lying along a given ray are superim- of a modern dynamic X-ray image detector are
posed, and information as to their exact depth discussed in 7 Sect. 11.9.4. For contact geom-
within the body is essentially lost. This superpo- etry (minimum geometrical magnification), the
sition of body structures can lead to ambiguities unsharpness of the X-ray image detector nor-
when interpreting the final image. This can only mally dominates spatial resolution.
be avoided if a more sophisticated X-ray imaging The recording of time-resolved information
geometry is adopted, as we will see 7 Sect. 11.6.3.
is essential in cardiovascular imaging. The imag-
The divergent nature of the X-ray beam means ing of dynamic structures and devices however
that an object (in the body) projected on the may be subject to a further source of blur due
image detector is subject to geometrical magnifi- to motion unsharpness. The degree of motion
cation, defined by the ratio: unsharpness depends upon the speed of the
moving structure and the temporal resolution of
M = SID : SOD the image acquisition process. This issue can be
particularly important when imaging the heart
where SOD is the X-ray source-to-object distance and coronary vessels and, of course, in pediatric
and SID is the X-ray source-to-image-detector patients. Today, modern dynamic X-ray image
distance. detectors exhibit negligible temporal unsharp-
Objects in the body situated at different dis- ness (lag); therefore, high-fidelity recording of
tances from the X-ray source are subject to differing dynamic structures is essentially limited by the
degrees of geometrical magnification. For example, duration of the X-ray exposure pulses. In turn,
iodinated vessels of the same diameter at the front this encourages the use of powerful X-ray sources
and back of the body section will be projected that can deliver sequences of very intense X-ray
onto the image detector with differing widths. pulses of suitably short duration. Such arguments
This geometrical process can distort the pattern of apply equally well to (low-dose) pulsed fluoros-
body structures that are recorded. Obviously, pro- copy and (high-dose) serial image acquisition.
vided that absolute estimates of size–scale are not
required, as is the case in many imaging tasks, this
need not be a problem. Where quantitative esti- 11.5 Signal-to-Noise Ratio
mates of size are important, however, calibration of Description of X-Ray Images
scale against a device of known size is essential.
When geometrical magnification is used, the Quantum physics reveals that an X-ray beam
finite size of the X-ray focus produces an X-ray comprises a stream of individual packets of
penumbra around projected objects, as illustrated energy or photons, not a continuum of energy.
in . Fig. 11.5. This can lead to blurring of signals
Consequently, the interaction mechanisms that
recorded in the image, particularly those that are underpin the formation of the X-ray image are
very small in size and/or fine in texture. However, statistical in nature. Inevitably, the fabric of the
larger structures with sharp discontinuous edges recorded X-ray image is perturbed by random
will also be affected. This blurring process, fluctuations in the photon count contributing to
known as geometrical unsharpness, can result in each of the picture elements forming the image.
a visible loss of spatial resolution in the images. These fluctuations are often referred to as X-ray
Geometrical unsharpness can be kept to a mini- quantum noise. In fluoroscopy, the noise fluctua-
mum by correct choice of X-ray focus, combined tions are a function of both space and time, sub-
with minimizing the degree of magnification. The jectively giving the appearance of a scene viewed
latter can be achieved by keeping the body sec- during a heavy snowstorm. These dynamic noise
tion in close contact with the image detector and fluctuations overlay the useful signal information
by keeping the detector at some distance from the and are most obvious in regions of uniform tis-
X-ray source. Contact geometry is also favorable sue background. The presence of noise fluctua-
from the standpoint of minimizing the patient tions sets a fundamental limit to the quality of the
dose. Geometrical unsharpness is only one of sev- image that can be formed using X-rays, as illus-
eral potential sources of blur in an X-ray image. trated in . Fig. 11.6.
160 A.R. Cowen
Ratio
tion of low-contrast details. On the other hand, can be developed further to take account of the
in serial image acquisition, the dose per frame is antiscatter grid. As a result, FOM analysis can also
usually increased by a factor of 10, 100, or even be used to evaluate the viability of X-ray scatter
1000 (depending upon the modality involved). As rejection strategies (such considerations, how-
a result, the ΔSNR will ramp up, reducing noise ever, go beyond the needs of this chapter).
and progressively improving the visibility of low-
contrast details.
It is also useful to appreciate that the ΔSNR 11.6 Digital X-Ray Imaging
is strongly linked to a physical interpretation of
information. Increasing the count of primary 11.6.1 The Digital Image
X-ray photons will increase the quantity of infor-
mation available in the image. Obviously, such a All modern cardiovascular X-ray image detectors
gain in information, however, must be balanced record images in digital form. The use of the term
against the increase in radiation dose incurred by digital implies a number of attributes, which are
the patient. Clearly, there is no justification for briefly reviewed in this section. A digital image is
increasing the radiation dose beyond the mini- an electronic image represented in a data format
mum level required to derive the information compatible with computer hardware and software
needed to perform the clinical task (in accordance protocols. As such, the image can be captured,
with the ALARA principle). On the other hand, processed, stored, and communicated rapidly and
this model also illustrates the risk associated with efficiently. It should be understood that dynamic
overenthusiastic reductions in the dose and hence X-ray images comprise a sequence of individual
in the X-ray photon count. Reducing the photon frames over time, as illustrated schematically in
count to too low a level will seriously degrade the . Fig. 11.7. Hence, dynamic processes are dis-
ΔSNR and, as a result, the image quality. This will cretely sampled in time at a frame rate that can
reduce the information content of the images, hopefully depict relevant moving structures and
potentially beyond the point where it is possible devices within the body. The acquired image
to successfully perform the clinical task. sequences are often viewed as a movie presenta-
The trade-off between the image quality and tion. A modern imaging system incorporates a
the patient dose can be explored systematically digital video frame buffer, making it possible to
using a suitable figure of merit (FOM) [5]. The display movie images at a rate independent of the
162 A.R. Cowen
.. Fig. 11.7 Schematic
diagram of dynamic X-ray
imaging: sequence of indi-
vidual digital image frames
acquired over time
Discrete
Grey-Scale
Value
Sequence of
Image Frames
2D Array
of Pixels
Time
11 acquisition rate. Image frames in the sequence may The equivalent pixel resolution of the detector is
also be stepped through individually and viewed expressed in terms of the linear density of pixel
as a static display. Salient image frames may then pairs and is normally expressed in line pairs (lp)
be selected to document clinical findings. per millimeter. This spatial frequency is com-
Each frame in the image sequence is a spa- monly referred to as the Nyquist frequency fN
tially sampled (discretized) representation of the where:
input X-ray image, not a continuously varying (or
analog) version. The image frame is represented f N = 1 / 2Dp = N / 2W
as a regular 2D array of minute, square picture
elements or pixels. For a square-format image, For example, if N = 1024 pixels and W = 20 cm,
the matrix contains [N × N] pixels (for example then fN = 1024/400 = 2.56 lp mm−1
[1024 × 1024] pixels). A rectangular format image Dynamic detectors have been developed with
contains [N × M] pixels where N ≠ M (for exam- different fields of view, form factors, and pixel
ple [2560 × 2048] pixels). As we see in 7 Sect.
resolutions to match the differing needs of the
11.8.2, these pixels are built into the electronic various types of cardiac and vascular imaging
superstructure of modern solid-state digital X-ray application. The total number of pixels forming
image detectors. The size of the pixels has a strong an image frame can be calculated from the ratio
influence upon the fidelity of the images. The of the active area of the detector to the area of
smaller the pixel size, the greater the spatial reso- an individual pixel. In a modern dynamic digital
lution of the recorded images, but the greater the detector, an image frame can contain anything
quantity of data that have to be managed, stored, between 1 million and over 5 million pixels.
processed, displayed, etc. Associated with each pixel in the image is a
The pixel-sampling interval of a detector Δp grayscale value that represents the intensity of the
can be calculated from the number of pixels along image at that pixel location. This grayscale value is
a dimension N and the equivalent field of view W selected from a set of discrete levels, not a contin-
such that: uum of intensity values. By convention, digitiza-
tion of the overall grayscale range is expressed as a
Dp = W / N power of 2 and quantified in terms of the number
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
163 11
of bits (binary units) of information per pixel. (including pixel array size and bit depth) but also
Increasing the number of bits per pixel improves contains patient demography plus other technical
the grayscale resolution of the image data. The information about the image acquisition device,
grayscale resolution is defined by the specifica- the radiographic examination, and display char-
tion of the analog-to-digital converter (ADC), acteristics. In the future, the header will explicitly
which digitizes the analog input signal (e.g., to 8, include patient radiation dose information. The
10, 12, 14, or greater bit resolution). The number DICOM structures used in cardiac and vascular
of grayscale levels G allocated per pixel for digiti- digital X-ray imaging share common features—
zation with g bit resolution is given by the simple for example, both organize image files into runs
equation: and studies. DICOM also supports both lossless
and lossy image compression—for example, via
G = 2g the JPEG compression standard. Compliance with
the DICOM standard facilitates interfacing image
Modern (solid-state) image detectors typically acquisition system(s) with a picture archiving and
digitize the image to 14 bits of grayscale resolu- communication system (PACS) environment and
tion. Substituting in the above equation, we see should ensure portability of image data across dif-
that a 14-bit image enjoys a generous choice from ferent hardware platforms.
some 214 = 16,384 grayscale levels at each pixel
site. Such a specification is adequate for stan-
dard dynamic X-ray imaging applications. More 11.6.2 Digital X-Ray Imaging System
sophisticated image acquisition schemes such as
3D reconstructive imaging may require improved A modern cardiovascular digital X-ray imag-
grayscale resolution (e.g., 16 or even 18 bits per ing system is an assemblage of several key high-
pixel). technology components. The overall layout of a
The acquisition of a sequence of digital images typical imaging system is illustrated schematically
generates copious quantities of data. The total in . Fig. 11.8. The components identified in this
quantity number of bits of data (Ν) acquired in block diagram each make a conspicuous contribu-
a single run of τ seconds duration and at frame tion to the dynamic imaging process: from X-ray
rate f is given by the product: production to image acquisition to computer pro-
cessing and storage, through to final display.
N = g ´ N ´ M ´ f ´t Successful interventions rely upon (among
many other things) the matched and synergetic
Total data content is often expressed as bytes of operation of the key components of the imaging
data, where 8 bits equals 1 byte. A single digital equipment. Powerful but also flexible produc-
cardiac image acquisition run can generate hun- tion of X-rays is essential to ensure dose-efficient
dreds of megabytes of data, where 1 megabyte recording of high-quality dynamic images.
equals 1 million bytes of data. Obviously, acquisi- Important technological developments in this
tion of multiple runs increases the data load pro- area include X-ray tubes with a much-improved
portionately, unless data compression is used. anode heat capacity and management, and also
As described so far in this section, a digital grid-controlled X-ray tubes. These developments
X-ray image is represented as a simple bitmap. have had a significant impact on fluoroscopy-
Such an image format is too primitive to be reli- guided cardiovascular interventions in particu-
ably used to communicate and display image lar. To facilitate dose-efficient system operation,
files in a clinical department. The image file must imaging systems utilize increasingly sophisticated
also contain relevant administrative and techni- automatic dose rate control (ADRC). ADRC sys-
cal information. This typically takes the form of tems are designed to automatically set optimized
metadata in a header to the raw image data. Such X-ray exposure factors for each individual patient
a file format has been developed specifically for and type of examination [7].
medical imaging applications, known as the Image detectors play a pivotal role in the
DICOM (Digital Imaging and Communications capture of high-quality X-ray image sequences
in Medicine) standard. The DICOM header and their conversion into a stream of digital
defines not only the basic image specification data. Recent years have seen a paradigm shift in
164 A.R. Cowen
PACS
System
Computer
Patient ADRC
GEO
C-arm
Operator Interface
Spectral Filter
Beam
Shaping
Power
Generator
X-ray Source
image detector technology from electron-optical The X-ray source and image detector are
television-based devices to the new generation affixed to a C-arm (or a similar design) attached
of solid-state image detectors [8, 9]. All acquired to a floor- or ceiling-mounted stand. The C-arm
images must undergo digital image processing can be rotated to different positions around the
to ensure optimum image quality. In modern patient [10] to optimize the projection geometry
detectors, this includes digital image calibration at each stage of the procedure. The geometry of
to correct for cosmetic deficiencies in the image the C-arm is calibrated (GEO), enabling the X-ray
detector and to restore the physical integrity of the projection angle(s) to be set up automatically by
acquired data. In addition, digital image enhance- the system if required. Continuous motorized
ment is used to ensure that the visual quality of rotation of the C-arm (plus the source and detec-
the images presented to the clinician is also opti- tor) about the isocenter is also used to acquire
mized. Today, images are viewed on high-quality a set of projections over a wide angular sweep
flat-screen (FS) monitors. To meet the immedi- around the patient. The resulting set of images
ate and short-term image storage needs of the can be viewed on the display as a rotating pan-
interventionalist, the system must provide access orama. Alternatively, the acquired projection data
to high-speed data storage (of medium capac- can be used to compute a 3D image. The system
ity), typically a magnetic hard disc. To meet the described is a single-plane C-arm unit. Biplane
long-time image storage needs, the system must configurations, which incorporate two C-arms
provide access to a secure, large-capacity archive; and two orthogonal X-ray imaging channels, are
this is typically achieved via an institutional (or also widely used in cardiovascular imaging, par-
possibly a dedicated) PACS network. Some envi- ticular for pediatric examinations. The economic
ronments also use portable archive media, such as cost and clinical role of single- and biplane con-
recordable optical disks, to store results of indi- figurations may differ, but the physics and tech-
vidual procedures. nologies remain the same.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
165 11
Salient enabling technologies that underpin the transit of the contrast medium. Clinical expe-
modern cardiovascular imaging, from the X-ray rience has shown that optimum DSA imaging
equipment to the final display, are reviewed requires logarithmic (rather than linear) subtrac-
in 7 Sects. 11.7, 11.8, 11.9, 11.10 and 11.11.
tion of image frames; this algorithm is described
Preparatory to this, however, the basic physics of by the equation:
three widely used cardiovascular imaging tech-
niques—digital subtraction angiography, rota- DS = log I Contrast - log I Mask or alternatively
tional angiography (RA), and 3D reconstruction log {I Contrast / I Mask }
imaging—are briefly discussed.
This result indicates that DSA actually represents
the (log) division of images rather than (lin-
11.6.3 Digital Subtraction ear) subtraction. Contrast enhancement is then
Angiography (normally) applied to the log-subtracted image
to optimize detail visibility in the final display.
A digital image is represented by an array of num- The overall DSA algorithm is summarized in
bers that encode pixel locations and associated . Fig. 11.9.
intensity values. Outputting the arrays of numbers Logarithmic subtraction eliminates the
to a suitable display device such as a solid-state FS imprint of the background anatomy, depicting the
monitor makes it possible to view dynamic image opacified vessels (in either black or white polar-
sequences and/or selected image frames. Prior to ity) laid out upon a uniform (gray) background.
viewing the display image, the data are subject In practice, this is not quite correct as DSA pro-
to digital enhancement (numerical manipula- duces on the gray background a heterogeneous
tion of the image data) to optimize the presented field of random noise; this artifact is known as
subjective quality. This is widely used in dynamic the DSA noise print. The noise print reflects the
grayscale imaging in order to improve contrast varying contributions of quantum and electronic
and spatial resolution and to reduce noise fluc- noise over regions of different X-ray attenuation
tuations. Other imaging algorithms can be used (defined by the background anatomy). Random
to process the image data to achieve more fun- noise in the mask and contrast medium–enhanced
damental improvements in the presentation of image frames is uncorrelated and therefore adds
clinical information. A typical example of such (rather than subtracts). Therefore, DSA eliminates
an algorithm is digital subtraction angiography the anatomical structure noise but boosts the
(DSA) [11–13]. overlay of random noise. Assuming that the vari-
The aim of DSA is to generate a sequence of ance of the noise in the contrast-enhanced image
images that clearly depicts the flow of iodinated (σContrast2) and the mask images (σMask2) are of sim-
contrast medium through the arteries of interest, ilar magnitude, the DSA algorithm will amplify
unencumbered by overlying anatomical struc- the noise variance (power) by a factor of 2, viz.:
tures. Here the contrast medium–enhanced vessel
is the signal of interest. This is achieved by first sDSA2 = éësContrast 2 + sMask 2 ùû ~ 2•s 2
acquiring a mask image of the background anat-
omy. The contrast medium is then injected into Correspondingly, DSA reduces the ΔSNR (of the
the vessel, and a sequence of images is acquired, opacified vessel) by a factor of √2, viz.:
which records the inflow and washout phases. In
a standard grayscale presentation, X-ray beam DSNR DSA = DS / s Ö 2 = DSNR / Ö 2
attenuation due to the competing background
anatomy represents a powerful source of noise. Despite this, provided that the X-ray dose per
Fortunately, this anatomical structured noise can, frame is sufficiently high, the conspicuity of the
in principle, be eliminated by subtracting the con- opacified vessel will be significantly improved by
trast medium–enhanced image frames (IContrast) the subtraction of the background anatomy, com-
and the mask image (IMask) [14]. Following digi- pared with an individual grayscale image (which
tal subtraction, the static (common mode) image retains this structured noise). The dose per frame
content is eliminated, leaving an impression of used in DSA can be a factor of 10 (or so) greater
the changing opacification of the vessels due to than that used in grayscale image acquisition.
166 A.R. Cowen
Log
Mask Image
Contrast
Subtract
Enhancement
DSA Image
Log
Contrast Image
.. Fig. 11.9 Schematic diagram of the digital subtraction angiography (DSA) imaging algorithm
11 DSA is a temporal subtraction algorithm, anatomy still may not be complete. As a result,
as the image frames are subtracted across time. hybrid subtraction has also been investigated as it
Therefore, in principle, DSA relies upon the potentially combines the benefits of temporal and
patient remaining immobile during image acqui- energy subtraction [15]. To date, however, neither
sition to prevent (minimize) spatial misregistra- the energy subtraction nor the hybrid subtraction
tion. In practice, this is not always achievable; algorithms have proved to be a suitable replace-
therefore, remedial postprocessing may be needed ment for DSA in clinical routine.
to ameliorate motion-related artifacts; techniques
include remasking, pixel shifting, and automatic
reregistration. To date, DSA has proved to be 11.6.4 Rotational Angiography
successful in all clinical areas apart from coro-
nary artery imaging. This is due to the rapid and The 2D projection images routinely produced
sometimes irregular motion of the heart plus lung in fluoroscopy and digital image acquisition
motion due to respiration, which can overlap the (including DSA) provide an incomplete geo-
heart. Recent advances in computer processing, metrical representation of cardiovascular struc-
however, may resolve this issue. Nevertheless, in tures. Individual projection images of opacified
all other clinical areas, DSA has been the imaging vessels are susceptible to overlapping and fore-
yardstick in vascular diagnosis and intervention shortening, which limit the precision of the clini-
for well over a quarter of a century. cal information that is derived. Recording two
Alternative algorithms to DSA include energy orthogonal views (in biplane imaging) offers an
subtraction, whereby images acquired at two improved, but still incomplete, perspective on
different X-ray beam energies are digitally sub- target vascular structures. Even better-quality
tracted. Energy subtraction has certain attrac- clinical information is derived if multiple angular
tions in cardiovascular imaging as it is potentially views are acquired (. Fig. 11.10). For example,
less susceptible to movement misregistration, as a set of digital images can be acquired at equi-
no precontrast mask image is required. However, angular intervals over a semicircular sweep of
in practice, the subtraction of the background the C-arm around the body section. In its most
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
167 11
.. Fig. 11.10 Schematic
diagram of the geometry of
rotational angiography (RA)
X-ray image acquisition
Image Detector
r
to
Im
c
ag
te
De
e
De
e
ag
te
c
Im
to
r
Image Detector
Image Detector
Isocentre
basic form, this imaging technique is known as a rotating panorama of the vasculature. RA relies
rotational angiography (RA). In RA, the C-arm upon motion parallax to engender an impression
typically rotates at a constant angular speed of depth, which otherwise is missing in 2D pro-
of 30–60 degrees s−1 and describes a sweep of jection images. Either a grayscale or a subtractive
180–240° around the body. A single bolus of approach can be adopted in RA imaging. The lat-
contrast medium is injected for the duration of ter requires a double sweep of the C-arm with
the acquisition phase. Reasonably rapid data projection images acquired at equivalent angular
acquisition is essential to ensure maximum ves- positions. A set of mask images is acquired dur-
sel filling and minimum motion blur. Projection ing the first sweep, and this is followed by a sec-
images are typically acquired at a frame rate ond (reverse) acquisition sweep while the bolus
of 10–30 frames s−1. A higher frame rate than of contrast medium is injected. RA has proved
this, such as 60 frames s−1, may occasionally be to be particularly useful in neurovascular imag-
required, but this usually requires a compensa- ing—for example, in improving the depiction
tory reduction in pixel resolution (to balance the of the shape, location, and orientation of intra-
quantity of data acquired). Projection images are cranial aneurysms [16, 17]. Over the years, RA
often acquired at dose levels suitable for grayscale has found a role in other clinical areas, including
acquisition rather than for DSA; the dose per abdominal interventions [18] and renal artery
frame in RA can be 4–10 times lower than that assessment, and also in improved visualization
used in DSA exposures. of coronary arteries. For example, three studies
The C-arm is rotated around a fixed reference have investigated the safety and clinical utility
point, known as the isocenter. The clinical region of RA in coronary artery procedures, reporting
of interest must be accurately positioned at the significant reductions in the contrast medium
isocenter before the image acquisition begins. load, radiation dose, and procedure time, com-
The acquired image sequence can then be viewed pared with standard 2D digital cardiac imaging
on the display screen as a movie loop, depicting [19–21].
168 A.R. Cowen
detector and cone-beam geometry is inevitably is now also possible to reconstruct 3D images of
more susceptible to X-ray scatter than multislice low-contrast (soft-tissue) structures.
CT (MSCT); the latter benefits from tight col-
limation of the X-ray beam both at the source
and at the detector array. Reportedly, the X-ray 11.7 X-Ray Equipment
scatter-to-primary ratio found in C-arm-based
cone-beam CT can exceed 3:1, while the SPR of 11.7.1 irect Heat Removal
D
an MSCT scanner typically measures 0.2:1 [27] from the Anode
or less. The presence of X-ray scatter reduces the
ΔSNR and degrades the visibility of low-contrast During extended periods of fluoroscopy and/or
sensitivity in 3D reconstructed images. In addi- acquisition of multiple runs of serial images, the
anode disk of the X-ray tube accumulates a con-
11 tion, scatter induces artifacts in the reconstructed
siderable quantity of heat. Build-up of heat in the
images, such as cupping (in regions of homoge-
neous attenuation) and streaks between high- anode limits the duration and frequency of the
attenuation structures. The presence of scatter fluoroscopy and associated exposure runs that can
also perturbs the calibration of reconstruction be implemented. Early designs of the X-ray tube
in terms of Hounsfield units (HUs). A variety of relied upon thermal (infrared) radiation from the
strategies is employed to reduce X-ray scatter in anode to dissipate the accumulated heat. More
C-arm-based cone-beam CT. These include use of efficient heat management is achieved by direct
an antiscatter grid and/or air gap and a bow tie or removal of heat from the anode disk, and this
wedge-shaped beam-compensating filter. In addi- strategy is now exploited by leading equipment
tion, software corrections are often applied to cor- manufacturers. One notable example of such a
rect for quantitative estimates of the X-ray scatter. design is the Philips Maximus Rotalix Ceramic
Together, these techniques improve image quality (MRC) X-ray tube [28], shown in . Fig. 11.11.
and reduce the artifacts induced by X-ray scatter This design of X-ray tube is well matched to the
in C-arm-based cone-beam CT (but cannot cor- demanding exposure conditions prevailing in
rect for them completely). cardiovascular interventions, being able to sus-
Cone-beam CT images are reconstructed tain high X-ray tube output for extended periods
offline using a dedicated workstation, typically as of time.
a 3D array containing 5123 or 2563 voxels (volume The MRC X-ray tube utilizes a liquid metal–
cells). Implementation of surface-shaded display lubricated rotating sleeve. The anode disc and
(or volume rendering) is then used to achieve a sleeve are mounted upon a stationary hollow
photorealistic representation of the vessels in metal shaft. Spiral grooves are machined on
3D. Such 3D visualization techniques can help the the outside surface of the shaft, as illustrated in
clinician interpret complex webs of overlapping . Fig. 11.12. The rotating sleeve glides over the
vessels [22, 23]. For example, vessel segments can thin film of liquid metal lubricant that separates
be rotated on the workstation display and viewed the sleeve and shaft. The lubricant comprises a
from different perspectives. Access to 3D vascular eutectic gallium–indium–tin (Ga–In–Sn) metal
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
169 11
alloy, which is liquid at room temperature. The
spiral groove bearing and liquid metal lubrica-
tion produce a friction-free interface between the
two metal surfaces. This mechanism is analogous
to a car aquaplaning on a wet road, whereby tire
traction is lost when a driver (unwisely) drives at
high speed. The rotating sleeve spins around the
support shaft at 4200 revolutions per minute. The
anode need only be accelerated up to full running
speed once a day, via a motor contained within the
tube itself. Having reached the running speed, the
anode continues to rotate throughout the work-
ing day. The bearing supports a large-diameter
(viz., 200 mm) anode disk, affording both high
.. Fig. 11.11 Modern high-performance X-ray tube
X-ray tube output and a large anode heat storage
featuring a spiral groove bearing and liquid metal lubrica-
tion. (See «Acknowledgments») capacity.
Rotating Anode
Rotating
Sleeve
Evacuated Metal
Envelope Stationary Shaft with
Spiral Grooves Coolant Circulating
Through Hollow Shaft
Electron
Flow Liquid Metal Lubricant Motor
X-ray Beam
.. Fig. 11.12 Schematic diagram of an X-ray tube featuring a spiral groove bearing and liquid metal lubrication
170 A.R. Cowen
Heat accumulating in the anode is removed not 30 frames s−1. For certain applications, particu-
only by thermal radiation, but importantly also via larly pediatric cardiac imaging, a frame rate of
a second direct heat transfer pathway. This pathway 60 frames s−1 may be used. In examinations where
allows heat to be directly transferred through the motion is less prevalent and/or dose reduction is
rotating anode assembly to the coolant (oil), which paramount, fluoroscopy may be carried out with a
is circulated through the hollow shaft. The hot oil lower pulse rate such as 10, 7.5, 6, or 3.75 frames s−1
is cooled via an external heat exchanger, before (or even as low as 0.5 frames s−1). The low pulse
recirculation through the shaft. The resulting rates are used in cardiac electrophysiology and
improvement in heat removal facilitates interven- in many areas of vascular fluoroscopy. Pulsed
tional imaging, eliminating unwelcome time delays fluoroscopy at low frame rates requires digital
between and/or during clinical procedures, which gap filling to suppress the disturbing flicker that
were frequently experienced in the past [29]. otherwise would appear in the displayed images.
Gap filling signifies the repeated presentation of
successive image frame(s) to achieve an apparent
11.7.2 Grid-Controlled X-Ray Source increase in the displayed frame rate.
long lengths of HV cables. Consequently, the expo- in modern cardiovascular X-ray imaging sys-
sure pulses had substandard characteristics, with a tems typically have a maximum power rating of
11 slow kV rise at the front edge and a slow decay at 80–100 kW. The generator supplies the X-ray tube
the end of the pulse. This prolonged the duration with electrical power, as a rapid sequence of pre-
of the X-ray pulses and delivered spurious low- cise pulses. The resulting X-ray exposures must
energy radiation. The former degraded the spatio- be tailored to the requirements of (high-dose)
temporal resolution of moving vessels, while the serial image acquisition and (low-dose) pulsed
latter increased the patient skin dose unnecessarily. fluoroscopy as appropriate. Modern power gen-
Modern X-ray tubes used in cardiovascular erators utilize high-frequency (HF) power inverter
fluoroscopy now incorporate a third electrode, technology [32]. HF generators are powerful and
known as a control grid [30]. This control grid flexible, but compact in design and economical to
comprises a fine wire mesh mounted in the gap manufacture. The HF generator functions by tak-
between the cathode and anode. Exposure switch- ing mains voltage and up-converting the frequency
ing is implemented directly at the X-ray source by and delivering the HF power to the primary coils
applying a suitable bias voltage to the control grid of the HV transformer. In practice, this is achieved
either allowing tube current to flow or extinguish- in two steps: firstly, mains frequency is stepped
ing the flow. The response of the grid-controlled down from 50 Hz (in the UK and Europe) to direct
X-ray tube to the control signal is instantaneous, current (DC), which is then up-converted to the
essentially producing X-ray pulses with a perfectly operating frequency, typically in a range from tens
square kV profile. This enables rapid sequences to hundreds of kilohertz. At these frequencies,
of precise X-ray pulses to be reliably generated. the HV transformer works very efficiently, step-
Pulsed fluoroscopy is the standard fluoroscopy ping up the voltage to the required peak anode
technique in modern cardiovascular X-ray imag- voltage with minimum power loss. The amplified
ing. Pulsed fluoroscopy improves the spatiotem- HF voltage is then rectified and smoothed before
poral resolution of rapidly moving blood vessels application across the X-ray tube. HF generators
and saves the radiation dose [31]. use fast-response closed-loop regulation of expo-
In adult coronary vessel imaging, fluo- sure factors (kV and mA). The kV pulses applied
roscopy is commonly performed at a pulse across the tube approximate a constant potential
rate of 15 frames s−1, but very occasionally at waveform, apart from a few percent of HF ripple.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
171 11
The values of the tube voltage used in cardio- pulses must be kept as short as possible, con-
vascular examinations typically fall in the range sistent with power loading restrictions. In prin-
of 50–125 kVp, in either pulsed fluoroscopy or ciple, minimum switching times of 1 ms can be
serial digital image acquisition. HF generators achieved using quartz-controlled switching of
use fast-response, closed-loop feedback control to the electrical power. In cardiac image acquisi-
automatically regulate the X-ray tube voltage. A tion, the runs are often captured at a similar
potential divider is used to continually sense the frame rate to that used in pulsed fluoroscopy—
kV generated and measure the divergence from for example, 15 frames s−1. In cardiac image
the reference kV. The reference kV in this case will acquisition, the dose per frame will typically be
be defined by the ADRC system. A device known 10 (or so) times greater than that used in pulsed
as a voltage-controlled oscillator (VCO) is used to fluoroscopy. In vascular applications, image
generate the kV correction signal. The correction acquisition is often performed at a compara-
signal consists of a series of trigger pulses, which tively low frame rate, say, from a maximum of
have a frequency proportional to the divergence 6 frames s−1 down to 1 or 0.5 frames s−1. Again,
of the actual kV from the reference kV. These trig- the optimum frame rate varies with the clini-
ger pulses are fed back to the power converter to cal application. At the same time, the dose per
re-establish the reference tube voltage. As a result, frame, certainly in DSA, has to be increased by a
the closed-loop regulation system enables the factor of 100 (or so) compared with that used in
peak tube voltage (kVp) to be rapidly and accu- cardiac acquisition. This is necessary to ensure
rately established and maintained throughout the that the subtracted images have a suitably high
imaging sequence. In DSA, the kVp must remain ΔSNR and therefore low-contrast resolution.
constant (within 1% or so) throughout the image X-ray pulse durations in vascular image acqui-
acquisition sequence to prevent artifacts follow- sition are significantly longer than those used in
ing image subtraction. cardiac image acquisition.
During pulsed fluoroscopy, the X-ray expo-
sure is delivered as a rapid series of pulses of short
duration. The tube current value typically used in 11.7.4 Spectral Filter
pulsed fluoroscopy is ~150 mA. An even higher
tube current (from hundreds up to 1000 mA or X-ray imaging systems incorporate hardware
so) is commonly required in digital image acqui- in the X-ray tube head to shape and condition
sition. A separate fast-response closed-loop regu- the X-ray beam (see . Fig. 11.8) [33, 34]. This is
lation circuit is used to control and stabilize the required to optimize the radiation field and X-ray
filament current. A sensing voltage is measured spectrum for each type of clinical examination.
across a reference resistor and is used to moni- The X-ray field is shaped using rectilinear lead
tor the actual X-ray tube current. The measured collimators and semitransparent wedge filters.
current is compared with the reference current The former are automatically set to delimit the
(mA) selected by the automatic exposure regime. X-ray field of view (FOV) against changes in the
Again, any divergence from the reference cur- zoom field and the source-to-image detector dis-
rent is c ompensated for via a VCO feedback loop. tance. The operator is also able to manually col-
The VCO supplies charge pulses to update the limate the X-ray beam to delineate a smaller FOV
filament current control circuit and stabilize the matched to the clinical region of interest. This
tube current. Closed-loop regulation of mA auto- is good practice as it both avoids unnecessary
matically compensates for filament aging and the patient (and staff) dosing and improves image
space-charge effect of the X-ray tube, ensuring quality by reducing X-ray scatter. Wedge filters
accurate and reliable dose calibration. are used to equalize the radiation field incident on
Pulsed fluoroscopy and image acquisition the detector across the body section. This reduces
both rely upon the production of a sequence of flare and contrast loss at interfaces between ana-
precise X-ray pulses, sometimes delivered over tomical regions of comparatively low and high
an extended time period. In the case of fluo- X-ray transmission (e.g., the heart/lung border).
roscopy, the run can be many minutes long. In Similar techniques are also used in DSA to mini-
image acquisition, the run is typically 10–20 s mize overexposure of the detector in regions of
long. The duration of the individual X-ray very high X-ray transmission.
172 A.R. Cowen
a spectral filter (shaping filter). Spectral filters are etration and as a result will reduce subject X-ray
used to reduce the risk of deterministic skin inju- contrast. For example, addition of a 0.1 mm–thick
ries, and their use has become standard radiation copper spectral filter reduces vessel subject X-ray
protection practice in interventional fluoroscopy. contrast by up to 5%. Provided there is sufficient
Copper spectral filters of a thickness between 0.1 X-ray tube power and heat loading capacity in
and 0.9 mm are well established in cardiovascu- reserve, this loss of contrast can be ameliorated
11 lar fluoroscopy. Copper spectral filters are also by retuning the X-ray exposure factors. Typically,
increasingly being used in cardiac and vascular this would be achieved by reducing the kVp to
image acquisition. Successful implementation of soften the X-ray beam (somewhat) and boosting
spectral filters demands the availability of very the mA to compensate for the fall in X-ray tube
powerful X-ray tubes with excellent heat manage- output.
ment properties, as discussed in 7 Sect. 11.7.3. Rare earth series elements (such as gado-
linium (Gd) and erbium (Er)) have been con-
sidered as possible alternatives to copper as
spectral filter material. For example, recently, a
4 ×10
4
No added Cu
gadolinium spectral filter was investigated for
3.5 0.1 mm Cu use in pediatric cardiac angiography [35]. One
X-ray Beam Intensity [AU]
0.2 mm Cu
3 0.4 mm Cu commercial cardiovascular imaging system cur-
rently employs a tantalum (Ta) (Z = 73) spec-
2.5
tral filter. Tantalum has a K-absorption edge at
2 67.4 keV and as a result reduces not only the con-
1.5 tribution of low-energy X-ray photons but also
1 the high-energy X-ray photons that exceed the
0.5 K-absorption edge. Reportedly, tantalum filters
preferentially transmit photons in the midenergy
0
0 20 40 60 80 100 range of the X-ray spectrum (at energies below
Photon Energy [keV] 67.4 keV), providing good dose savings while
better preserving subject X-ray contrast [36].
Currently, however, copper remains the spectral
.. Fig. 11.13 X-ray spectra for 80 kVp exposure: effects
of 0.1, 0.2, and 0.4 mm copper spectral filters. kVp peak filter material of choice in the majority of cardio-
kilovoltage. (See «Acknowledgments») vascular imaging installations.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
173 11
11.7.5 Automatic Dose Rate Control manufacturer. Nevertheless there are some shared
principles, which are worth reviewing here.
Cardiac and vascular imaging systems incorpo- Obviously the ADRC system must operate within
rate specialized hardware and software to auto- the appropriate maximum heat load restrictions
matically select and stabilize the X-ray exposure and power limits of the X-ray tube and generator.
factors (and hence operating dose rates) during In addition, there are legal limits on the maxi-
clinical operation. Until recently, this type of con- mum patient dose rates in fluoroscopy, which
trol facility was referred to as automatic brightness must never be exceeded. For example, in the
control (ABC) [37]. Today, the term automatic dose USA, the Food and Drug Administration (FDA)
rate control (ADRC) is more commonly used (see prescribes a maximum entrance surface dose rate
. Fig. 11.8) [38]. To a certain extent, this change
in fluoroscopy of 8.8 cGy min−1; however, such
in name also marks the increased sophistication limits vary in different parts of the world. Within
of modern automatic dose rate control facilities. these restrictions, however, the ADRC can be
ADRC applies to both pulsed fluoroscopy and designed to function in a variety of different ways
digital image acquisition. An ADRC system is by exploiting different exposure parameter trajec-
designed to maintain a constant (target) dose rate tories. These parameter trajectories are designed
at the image detector entrance independently of to achieve particular performance outcomes in
the patient size (age), the thickness of the body clinical use. Four potential fluoroscopic ADRC
section, the projection angle, the degree of X-ray trajectories are depicted in . Fig. 11.14. These tra-
beam collimation and filtration, and the details of jectories essentially describe the kVp versus mA
the clinical procedure (e.g., the use of a contrast response of the ADRC to patients of increasing
medium or otherwise). During fluoroscopy, the thickness (as indicated by the directions of the
ADRC typically measures the mean value of the arrows in the graph). The periphery of the zone
detector signal over a defined (dominant) area at delineating the ADRC response, given by the
the center of the image field. Feedback control is hatched area, is defined by the maximum electri-
then used to automatically adjust the X-ray expo- cal limits of the X-ray source and generator.
sure factors to stabilize the detector entrance dose Curve A (in black) represents the so-called
rate. Factors controlled by the ADRC include the isowatt ADRC algorithm, which functions by
peak tube voltage (kVp), the tube current (mA), holding the [kVp × mA] product at the constant
and the exposure pulse duration. Certain designs maximum value. As patient thickness increases,
of ADRC also automatically select the spectral fil- the kVp increases monotonically, while the mA
ter. ADRC ensures that the displayed fluoroscopic falls, following the maximum isowatt limit curve.
images have stable brightness, with no need for This ADRC algorithm makes maximum use of the
the operator to adjust exposure factors manually. power capabilities of the X-ray source.
In addition, stabilizing the detector entrance dose Curve B (in green) represents a low-dose
rate ensures that the level of X-ray quantum noise ADRC algorithm where the kVp (and the pene-
in the image is constant to help maintain a con- tration of the X-ray beam) increases rapidly as the
sistent level of displayed image quality. In fluo- patient thickness increases, keeping the entrance
roscopy, the operator can normally select a target surface dose rate at a minimum. It is only after
dose rate, typically from two ADRC settings (say, the maximum kVp limit is reached that the mA
providing low dose rate versus standard dose rate increases significantly, and this can only continue
options) or three settings (say, low dose rate ver- until the maximum isowatt rating curve is inter-
sus standard dose rate versus high image quality sected. Associated with the rapid increase in kVp
options). These selections typically step the dose is a reduction in subject X-ray contrast however.
rate in increments of a factor of 2 (or so). The This ADRC algorithm might be used where dose
operator is free to change the ADRC target dose reduction is the overriding consideration and
rate to meet the evolving image quality and dose image quality is secondary.
constraints of each patient examination. Curve C (in red) represents a high-contrast
The technical details of an ADRC system vary (iodine-sensitive) ADRC algorithm, which is often
with the design philosophy adopted by individual used in vascular imaging. Here the kVp is held at a
manufacturers. Often this information is pro- value in the range of 63–66 kV to maximize sub-
prietary and must be solicited from the relevant ject X-ray contrast of iodinated contrast media.
174 A.R. Cowen
.. Fig. 11.14 Typical
automatic dose rate control
(ADRC) parameter trajecto-
ries (peak kilovoltage (kVp)
versus tube current in milli-
amps (mA)): curve A isowatt
algorithm (black), curve B
kVpmax
low-dose algorithm (green),
curve C high-contrast
B A
D
C
This kVp is used to exploit the K-absorption edge between dose and image quality, across the range
11 of iodine at 33.2 keV. An increase in patient thick- of patient sizes. Such designs of ADRC algo-
ness inevitably demands an increase in mA up to rithm may measure/estimate the image quality
the point where the maximum isowatt curve is and dose, and derive an FOM to use in param-
intersected. Beyond this, the kVp must increase eter selection. Such efforts are aimed at further
and the mA must fall, again following the maxi- improving the congruence of ADRC (and hence
mum isowatt curve. This ADRC algorithm maxi- imaging system) operation with the ALARA
mizes subject X-ray contrast but potentially at a principle.
cost of an increased patient dose. In serial image acquisition, the selection of
Curve D (blue) is a so-called anti-isowatt exposure factors (and potentially the spectral fil-
ADRC algorithm where the kVp and mA increase ter) by the ADRC is typically founded upon the
together, up to the point where the maximum parameters set up in the preceding fluoroscopy
isowatt curve is intersected. Beyond this, the run. The fluoroscopic exposure factors are typi-
kVp increases, and the mA falls and follows the cally first used to estimate an equivalent water
maximum isowatt curve. This curve represents a thickness for the patient. It should be noted that
compromise between algorithms B and C, with a identical factors are not used, but a suitable trans-
balanced trade-off between the patient dose and formation is required to cross over from optimum
image quality. fluoroscopy to optimum acquisition parameters.
ADRC systems designed for cardiovascular Finally, adaptive tuning of exposure factors
imaging may also feature automatic selection of occurs during the early frames of the acquisition
the thickness of the copper spectral filter against run to compensate for the actual patient attenu-
patient size [38]. This facility uses a motorized ation characteristics (including composition of
wheel containing a set of copper plates of differ- the body section). In practice, dose stabilization
ent thickness, mounted in the collimator hous- is achieved within the first few frames of the
ing. Automatic copper spectral filter selection start of the run and is therefore not normally a
is integrated with the more familiar autoregula- distraction to the viewer. Significantly, the dose
tion of kV and mA (discussed above). Arguably, overhead is therefore also kept to a minimum.
this arrangement can further refine the trade-off This ADRC approach is adequate for purpose in
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
175 11
cardiac X-ray image acquisition, where constancy camera such as a Plumbicon, Chalnicon, Saticon,
of the exposure factors throughout the run is not or Newvicon. More modern X-ray IITV systems
essential. replaced the electron-tube camera with a semi-
In DSA, however, the image signal, and hence conductor sensor known as a charge-coupled
exposure factors, must be held constant within device (CCD). CCD sensors are solid-state TV
tight limits, throughout both the mask (precon- cameras with superior, more stable technical
trast) and the contrast medium–enhanced phases characteristics than the earlier electron-tube
of the exposure run. Even modest variations in camera [43]. The resulting video images can be
exposure can result in unacceptable artifacts in the presented live on a display monitor and/or digi-
viewed image sequence following digital subtrac- tized prior to image processing and viewing. A
tion. Consequently, ADRC algorithms designed number of review articles have described the
for DSA often use initial low-dose test shots to design characteristics of X-ray IITV (including
estimate the exposure factors to be used in the CCD)–based fluoroscopy systems.
acquisition run proper. The selected exposure fac- The emergence of DSA imaging equipment
tors are then held constant (locked) throughout in the early 1980s established the clinical ben-
the image acquisition sequence. Locked exposure efits of integrating computerized video proces-
factors are also required during RA image acqui- sors and magnetic disk storage with X-ray IITV
sition. systems. The success of DSA established digital
Fluoroscopy systems normally incorporate serial image acquisition in vascular diagnosis
a second feedback control, which automatically and treatment. This subsequently led to further
stabilizes electronic gain, not the dose rate. This important innovations, such as dedicated digi-
facility is known as automatic gain control (AGC). tal grayscale X-ray imaging devices for cardiac
The AGC is particularly helpful under conditions applications. Advances in digital X-ray IITV
where the ADRC is no longer able to achieve the detectors led to the demise of film as the prin-
target detector dose rate; this condition is some- cipal recording medium in cardiac and vascular
times referred to as «running out of steam» and imaging. Digital X-ray IITV imaging made it
can occur, for example, when imaging the bariat- possible to acquire, access, and replay sequences
ric patient and where the maximum patient dose of images online and view the images in a digi-
rate limit has been reached. The AGC automati- tally enhanced form. At the same time, image-
cally adjusts the electronic gain to compensate for processing algorithms to facilitate patient dose
the shortfall in signal amplitude. This maintains savings—such as digital recursive filtering, last
the brightness of the displayed image, so the inter- image hold, and 2D road mapping—boosted the
ventionalist can continue the examination, albeit clinical utility and acceptance of digital fluoros-
that the image quality is degraded by increased copy [42, 44, 45]. By the turn of the millennium,
noise fluctuations. digital X-ray image- intensified CCD (IICCD)
systems had proved to be an effective and flex-
ible platform for cardiovascular imaging [46].
11.8 ynamic X-Ray Image
D Around this time, however, a new generation of
Detectors dynamic X-ray image detectors emerged, based
upon solid-state electronics. In the intervening
11.8.1 -Ray Image Intensifier
X decade, these detectors have gone on to replace
Television System digital X-ray IITV technologies in many cardio-
vascular facilities across the world.
Up to the turn of the new millennium, the
dynamic image detector used in cardiovascular
imaging was the X-ray image intensifier television 11.8.2 Solid-State X-Ray Image
(IITV) system [39, 40]. This system comprised a Detector
chain of electron-optical components, includ-
ing an X-ray image intensifier tube plus a high- Solid-state detectors were originally designed
specification closed-circuit TV channel, along for use in standard-projection radiography [47].
with coupling lenses [41, 42]. Historically, TV These digital radiography (DR) detectors pro-
images were recorded via a vacuum electron-tube vided online access to the electronic data, so the
176 A.R. Cowen
a light-sensitive cell (a photodiode) plus a switch- Pixium 4800 (Trixell SAS, Moirans, France). This
ing component in the form of a thin film transis- FD is manufactured specifically for use in cardiac
tor (TFT) switch. The quantity of charge carriers imaging applications and has a square FOV of a
released in each photodiode is proportional to the maximum diagonal size of ~25 cm (248 mm to be
intensity of the X-ray beam at that location. Each precise), with two zoom field modes of ~202 mm
TFT switch in the array is addressed via a matrix and 156 mm diagonal size, respectively. The pixel-
of (horizontal) gate control lines and (vertical) sampling interval is 184 μm, and the bit resolution
data transfer lines. During the X-ray exposure, all of the ADC is 14 bits. This detector can support
of the TFT switches in the array are held «open» an imaging rate of up to 30 frames s−1 at full pixel
so that packets of electronic charge accumulate resolution and 60 frames s−1 with pixel binning
in local storage capacitors. At the end of the X-ray (binning is explained in 7 Sect. 11.9.4).
exposure, the resulting 2D pattern of the electri- A biplane dynamic FD–based vascular
cal charge is read out by gating each row of pixels interventional lab is shown in . Fig. 11.16; this
in the AM array in turn. The readout charge is is a mock-up but clearly shows salient equip-
amplified prior to digitization and transfer to the ment component. The digital imaging channels
system computer. The dynamic FD is exposed to that are featured incorporate the Pixium 4700
repeated pulses of X-rays, and the resulting pat- dynamic FD (Trixell SAS), which also exploits
terns of charge are read out, producing a sequence the indirect-conversion principle. This FD has
of frames. Depending upon the dose levels used, a nominal maximum FOV of [38 cm × 30 cm]
the sequence of frames will constitute fluoroscopy ([381.9 cm × 294.1 cm] to be precise). The FD also
or serial image acquisition. provides multiple zoom field options; details of the
The internal construction of a commonly available zoom field modes vary with the equip-
used dynamic FD is shown in . Fig. 11.15. The
ment manufacturer. For the system presented,
components of the detector identified in this seven zoom selections are provided, covering
figure include the CsI:Tl X-ray absorption layer, from [30 cm × 30 cm] through [19 cm × 19 cm]
the surface light reflector, the amorphous silicon down to [11 cm × 11 cm] in the maximum zoom
.. Fig. 11.15 Construc-
tion of a dynamic flat-panel
detector (FD) designed
for cardiac imaging. (See
«Acknowledgments»)
Refresh Light
a-Si Photodiode
Array
Csl:TI
Scintillator
178 A.R. Cowen
.. Fig. 11.16 Biplane
neurovascular interven-
tional lab (mock-up) fea-
turing large field-of-view
(FOV) dynamic flat-panel
detectors (FDs). (See
«Acknowledgments»)
field selection. The pixel-sampling interval is IITV channels and therefore less susceptible to
154 μm, and the bit resolution of the ADC is 14 changes in performance over time—for example,
bits. The zoom field selections present images due to thermal drift. They are also insensitive to
with a square format. Square-format images are the magnetic and environmental effects that can
11 also used in fluoroscopy. High-frame-rate pulsed adversely affect IITV systems. For cardiovascular
fluoroscopy is achieved by using pixel binning. imaging applications, overall, the dynamic FD
The Pixium 4700 detector is incorporated into represents a superior imaging platform to the
several vascular (and some cardiac) imaging sys- digital X-ray IITV system.
tems manufactured by Philips Healthcare (shown The physical image quality of medical X-ray
in . Fig. 11.16) and Siemens Medical Solutions in
image detectors including dynamic FDs can be
Europe. Other major manufacturers of (indirect- evaluated using a measurement tool kit includ-
conversion) dynamic FD–based imaging sys- ing the dynamic range, geometrical distortion,
tems include GE Healthcare and Varian Medical vignetting and veiling glare, spatial resolution,
Systems in the USA. Certain other manufacturers lag (memory effect), and detective quantum effi-
use direct-conversion solid-state detector systems ciency (DQE). These measures are transportable
in their cardiovascular imaging products. across different designs of X-ray image detec-
tor, and the results can be used to compare per-
formance on an objective basis. 7 Sects. 11.9.1,
11.9 I mage Quality of Dynamic 11.9.2, 11.9.3, 11.9.4, 11.9.5, and 11.9.6 focus on
Flat-Panel Detectors the physical imaging performance of the new
dynamic FDs, where appropriate results are com-
All of the diagnostic and interventional cardio- pared with those produced by a digital X-ray IITV
vascular imaging roles traditionally supported system.
by the X-ray IITV chain are now reproduced—
indeed improved upon—by dynamic FDs. These
detectors are much more compact and lighter in 11.9.1 Dynamic Range
weight than the bulky IITV imaging chain. The
compact nature of the FD streamlines C-arm In basic terms, the dynamic range of an FD
design, increasing maneuverability around and describes the range of the detector entrance dose
accessibility to the patient. FDs are electrically over which information can be usefully recorded.
and opto-mechanically much more stable than The dynamic range is related to the ratio of the
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
179 11
maximum to the minimum entrance dose levels imaging, an application that places even heavier
that the detector can accommodate. More spe- demands on detector dynamic range. For exam-
cifically, the former is defined by the maximum ple, the dynamic range of an MSCT scanner
signal capability of the detector and the latter by approaches 106:1 with 20-bit equivalent digitiza-
the noise base. A dynamic FD normally operates tion. To compete with this, future designs of FD
as a multifunctional image receptor and, as such, are likely to implement 16- or 18-bit digitiza-
must be capable of operating over a wide range tion plus signal-sensitive gain selection [70, 71].
of dose levels [68]. For example, low-dose pulsed FDs employing such an extended dynamic range
fluoroscopy demands effective image record- design will improve the depth of information in
ing down to a mean detector entrance dose of the projection images and hence the quality of the
only 10 nGy per frame (or so). In high-quality 3D reconstructions.
dynamic image acquisition, the mean detector
entrance dose is typically increased to ~100 nGy
or 1 μGy per frame, depending upon the imag- 11.9.2 Geometry and Vignetting
ing mode. The former (lower) value is typical of
the dose per frame used in cardiac digital image Images acquired with an X-ray IITV system
acquisition; these images are commonly acquired are susceptible to certain field-dependent arti-
at a frame rate of 15 frames s−1. The higher dose facts, which can degrade their cosmetic quality.
value is typical of that used in vascular grayscale Geometrical distortion in an X-ray IITV image
image acquisition. In DSA, the mean detector arises from the electron-optical design of the
entrance dose can reach radiographic dose levels, X-ray image intensifier tube. The mapping of
viz., up to 4 or 5 μGy per frame or so. FDs have a electrons from the concave photocathode onto a
maximum dose capability somewhat greater than planar output screen results in a form of geomet-
this (by a factor of ~10); above this dose, overload rical distortion known as pincushion distortion.
of the FD will occur, producing signal saturation. Pincushion distortion describes the progressive
Tube camera–based X-ray IITV systems were increase in geometrical magnification toward the
particularly susceptible to highlight overloads and periphery of the image field. At the same time, the
blooming (the spurious spread of the saturated luminance of the image field falls off toward the
signal); these effects would, for example, occur periphery, producing a nonuniform distribution
between the lower limbs during peripheral angi- in brightness (so-called vignetting). Image inten-
ography or over the lungs in pulmonary artery sifiers are also subject to a characteristic form of
studies. Semitransparent wedge filters were devel- geometrical distortion due to extraneous magnetic
oped to ameliorate these effects; this facility has fields, which is known as S-distortion. The impact
been retained with FD-based imaging equipment. of S-distortion on image quality is compounded
For dynamic FDs, signals are normally by the fact that it varies with the changing angula-
digitized with 14 bits of grayscale resolution tion of the C-arm; this effect is felt most strongly
(corresponding to 16,384 levels). FDs used in in RA and related reconstructive imaging applica-
fluoroscopy reportedly offer a dynamic range tions. FDs are subject to none of these forms of
10–100 times wider than an X-ray IITV system geometrical distortion and therefore can record
[8, 69]. In order to accommodate this range of images with excellent field homogeneity. FDs are
dose levels, dynamic FDs incorporate more than insensitive to magnetic fields and therefore can
one preamplifier gain setting. The appropriate be successfully implemented in hybrid X-ray/MR
gain setting is selected automatically to scale the imaging laboratories [72] or in conjunction with
signal range achieved in each imaging mode. This magnetic catheter navigation equipment [68].
ensures that the signal digitization yields opti-
mum grayscale resolution and the impact of the
noise injected by the readout electronics is kept 11.9.3 eiling Glare (Low-Frequency
V
to minimum. Dynamic FDs have an accurate Drop)
linear signal response across the dynamic range
and can support accurate logarithmic subtraction Images produced by an X-ray IITV system are
of image data during DSA [61]. Dynamic FDs subject to the scatter of X-rays, electrons, and light
are increasingly being used in 3D reconstructive over a large area, which arise at various stages of
180 A.R. Cowen
image conversion. The term veiling glare is used to electronics all contribute to this. In the case of
describe the overall effect of these scatter mecha- an indirect-conversion dynamic FD, the scatter
nisms. The presence of veiling glare degrades and diffusion of fluorescent light in the X-ray
the contrast of recorded images. Veiling glare is absorption layer also contribute to image blur,
often quantified in terms of the low-frequency albeit to a moderate degree due to the channeled
drop (LFD); this concept is derived from the structure of the CsI:Tl scintillator. Typical spa-
modulation transfer function (MTF). MTF is tial resolution properties of indirect-conversion
used to describe the detailed spatial resolution dynamic FDs designed for cardiac and vascular
properties of X-ray image detectors [47]. The applications are summarized in . Table 11.1.
LFD measures the deterioration in MTF that is Equivalent values for a digital X-ray IITV system
attributable to the large-area scatter mechanisms. are included for reference. Comparing results in
A high value of LFD implies poor reproduction the largest FOV selection, the spatial resolution of
of large-area contrast by the image detector, and a large-FOV dynamic FD typically exceeds that of
vice versa. Dynamic FDs naturally exhibit a much a digital X-ray IITV system by over a factor of 2
lower-value LFD and therefore less contrast loss or so. The excellent spatial resolution of the FD is
than an X-ray IITV system; contrast is typically maintained across the whole image field. For the
improved by a factor of 5–10 with the FD [55]. digital X-ray IITV system, the spatial resolution is
Consequently, dynamic FDs produce 2D projec- only maintained within a quality area at the cen-
tion and 3D reconstruction images with superior ter of the image field and deteriorates toward the
low-contrast sensitivity (and a wider dynamic periphery.
range) than X-ray IITV systems. lp line pairs
For a large-FOV dynamic FD, typical of that
used in vascular imaging, the high frame rates
11.9.4 Spatial Resolution (say, 15 or 30 frames s−1) sometimes required
in fluoroscopy can usually only be achieved by
11 The spatial resolution of an X-ray image detector binning (averaging) data over blocks of pixels.
measures its capacity to resolve small and fine- The process of binning pixel data over [2 × 2]
detail structures in an image. The spatial resolu- blocks is illustrated schematically in the top two
tion of a dynamic FD is influenced by a number diagrams in . Fig. 11.17. It must be noted that
of physical and technical factors. For example, the binning inevitably reduces the spatial resolution
sizes of the FD pixel-sampling interval and effec- (by a factor of 2 in each linear dimension) (see
tive aperture plus the bandwidth of the readout . Table 11.1). At the same time, however, binning
.. Table 11.1 Spatial resolution of dynamic flat-panel detectors (FDs) designed for cardiac and vascular
imaging: results compared with a large field-of-view (FOV) digital X-ray image intensifier television (IITV) system
FOV Square Rectangular Square field Circular Square field Circular field
field field field
[2 × 2] Binning
Full Resolution Half Resolution
reduces the total number of pixels to be read out accompanying this is a reduction in detector
in each fluoroscopic frame, by a factor of 4. As FOV and therefore anatomical coverage.) For a
a result, the presentation of sequences of large- digital X-ray IITV system, the spatial resolution
FOV images at a 4× higher frame rate becomes is materially improved when an electronic zoom
feasible (for the same readout bandwidth). In field is selected; this is due to the rescaling of the
other words, if the maximum frame rate for image projected on the TV sensor. Selecting the
full-resolution imaging is 7.5 frames s−1, then if maximum IITV zoom field can more than double
pixel binning is used, the maximum frame rate the spatial resolution compared with that for the
increases to 30 frames s−1, as illustrated in the largest FOV selection. Dynamic FDs also provide
bottom two diagrams in . Fig. 11.17. Fortunately,
multiple zoom field selections. In contrast to the
in fluoroscopy the associated reduction in spatial digital X-ray IITV system, the intrinsic pixel reso-
resolution is usually not intrusive, as this is par- lution of an FD remains constant, independently
tially offset by the accompanying smoothing of of the zoom field that is selected. This obviously
noise fluctuations. is not the case if pixel binning is applied; in this
During fluoroscopy the interventionalist often case, the spatial resolution is typically reduced by
needs to select a detector zoom field to magnify a factor of 2.
the presented image and therefore improve the In IITV-based fluoroscopy, the selection of an
visibility of fine-detail structures. (Obviously electronic zoom field is invariably accompanied
182 A.R. Cowen
by an increase in the X-ray dose per frame deliv- 11.9.5 Lag and Memory Effect
ered at the image intensifier, II, entrance. In turn,
this leads to an increase in the patient dose. In Understandably, dynamic FDs must be capable
going from the largest FOV to the maximum of recording fluoroscopic images with sufficient
zoom field selection, the II (and potentially temporal resolution to meet the needs of the
patient) entrance dose may be increased by a fac- target clinical examination(s). For example, in
tor of 4 or more. This is required (at least in part) cardiac angiography, the mean velocity of a coro-
to compensate for the fall in the II brightness gain nary vessel [73] typically measures ~50 mm s−1,
with an increasing zoom factor. In a digital X-ray while the peak velocity can exceed 100 mm s−1;
IITV system, the zoom-dependent increments in clearly, high temporal resolution and frame rates
dose are fixed. In theory, for a dynamic FD, such are essential in coronary angiography. Imaging
zoom-dependent dose increments are not nec- other types of vascular structures can often be
essary; in practice, however, they are still often performed satisfactorily using a lower temporal
used to maintain the visual quality of the images. resolution and frame rate. The temporal response
The dose increments used with an FD are often of a dynamic FD is determined by two physical
less severe than those required by an X-ray IITV mechanisms [74]: the memory effect (or ghosting)
system [69]. With an FD, these dose increments and lag. These mechanisms occur concurrently;
must be carefully matched to the image quality under a given set of circumstances, a detailed
needs and dose constraints of the individual clini- experimental analysis is required to distinguish
cal examination. Some examinations may favor and quantify their contributions.
an FD entrance dose increment of, say, 20–40% Memory effect refers to the production of a
(or so) between zoom field selections, while other spurious frozen pattern (a so-called ghost image),
examinations may not need an increment in dose which mirrors the image content present in the
with zoom field selection. The spatial resolu- preceding X-ray exposure. This phenomenon can
tion of a large-FOV dynamic FD can be gauged persist for some time (even for several minutes),
11 from the abdominal DSA image presented in particularly after an intense X-ray exposure of a
. Fig. 11.18. Small vessels in the vascular bed are
high-contrast structure is made [75]. Memory
uniformly depicted with excellent spatial resolu- effect reflects a variation in the detector response
tion throughout the whole image field. depending upon the exposure history. Both indi-
rect- and direct-conversion FDs are susceptible
to memory effect; however, they result from
quite different physical mechanisms. In general,
the latter design of FD is more susceptible to
ghosting artifacts [32]. With regard to indirect-
conversion FDs, memory effect represents an
increase in CsI:Tl light emission and a-Si:H
photodiode gain, following the X-ray exposure.
These effects manifest themselves as a spurious
increase in detector conversion efficiency, pro-
ducing a so-called bright-burn artifact in images
that are acquired subsequently. Memory effect
can prove to be a particular problem in mixed-
mode imaging applications, specifically where
low-dose fluoroscopy might follow immediately
after the acquisition of images at a high dose
level—for example, as routinely occurs in DSA
[68]. The X-ray dose per frame during fluoros-
copy can be as low as one thousandth of that
.. Fig. 11.18 Sample abdominal digital subtrac-
used during DSA image acquisition. Therefore,
tion angiography (DSA) image acquired with a large even a modest degree of memory effect can in
field-of-view (FOV) dynamic flat-panel detector (FD). principle impinge upon the fluoroscopic images
(See «Acknowledgments») that follow.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
183 11
Lag is the property that quantifies the ability reconstructed images. It should be noted that in
of an image detector to accurately record time- these applications, afterglow in the CsI:Tl scintil-
varying changes in image content. Lag can cause lator also becomes a significant factor. For these
smearing of rapidly moving and time-varying reasons, the maximum frame rate used in RA
image structures. The larger the detector lag, acquisition is often limited to 15 or 30 frames s−1.
the poorer (slower) the temporal response, and
vice versa. Lag results from the carry-over of a
proportion of the recorded signal content into 11.9.6 Detective Quantum
succeeding frames in the sequence. In indirect- Efficiency
conversion FDs, lag largely results from the rela-
tively slow temporal response of the a-Si:H layer. Detective quantum efficiency (DQE) is currently
More specifically, lag arises from the trapping and the most influential physical parameter used
slow release (detrapping) of charge carriers in the to quantify (and compare) the performance of
photodiode array [76]. X-ray image detectors [78]. DQE describes the
Dynamic FDs must incorporate measures to efficiency with which an X-ray image detector
minimize lag and memory effect. Many modern can record information delivered in the incident
dynamic FDs achieve this via a so-called refresh (input) X-ray beam. As such, the DQE of an X-ray
light (or reset light), which conditions the detec- image detector is defined by the ratio [47, 78]:
tor with a bright flash of light prior to each new
image acquisition cycle [76, 77]. The refresh DQE detector = SNR recorded 2 / SNR input 2
light takes the form of an array of light-emitting
diodes, which instantaneously flood the FD with where:
light photons, thereby saturating the charge traps SNRinput2 is the square of the signal-to-noise
in the a-Si:H prior to each new X-ray exposure. ratio at the input of the image detector. This is
As a result, lag (and memory effect) is reduced to defined by the fluence (number or count per unit
an acceptably low level, ensuring an acceptable area) of X-ray photons contributing to an indi-
detector response for most 2D and 3D imag- vidual frame in the fluoroscopic image sequence.
ing applications. Signal retention due to lag in The value of SNRinput can be calculated knowing
a modern indirect-conversion FD is reportedly the detector entrance dose per frame utilized and
as low as 0.3% at a time 1 s after termination of the energy spectrum of the incident X-ray beam.
the X-ray exposure; after 10 s, the lag reduces by SNRrecorded2 is the square of the signal-to-noise
a further order of magnitude [68]. This ensures ratio recorded by the image detector. The value
that the temporal resolution is adequate even of SNRrecorded2 can be computed from the out-
for high- speed imaging applications such as put image data. In terms of counting statistics,
pediatric cardiac fluoroscopy. Equivalent values SNRrecorded2 is an estimate of the fluence of infor-
of lag for direct-conversion FDs are reportedly mation carriers (X-ray photons) that the recorded
much higher [68]. In some vascular applica- image is effectively worth. (Note: to simplify the
tions, a moderate degree of lag can be tolerated discussion here, it is assumed that the fluoro-
during fluoroscopy. Indeed, lag can sometimes scopic image detector exhibits zero lag, or that
help to improve fluoroscopic image quality by any lag that does exist is corrected.)
time-averaging (smoothing) noise fluctuations. The information recorded by an image detec-
In modern cardiovascular imaging systems, lag is tor can never exceed that delivered by the input
synthesized via a computer processing algorithm X-ray beam; therefore, DQE always has a value in
(such as recursive filtering). This makes it possible the range:
to automatically adjust the degree of lag to main-
tain image quality, as circumstances demand. 0 £ DQE detector £ 1
The presence of lag and memory effects in a
dynamic FD sets a limit on the maximum rate at A DQE value of unity implies that the detector
which projection data can be usefully acquired for records the incoming X-ray information perfectly.
3D reconstructive imaging. The spurious carry- At the other extreme, a DQE of zero implies that
over of signal content into subsequent projec- no information at all is recorded. Real-world
tions gives rise to artifacts and unsharpness in the X-ray image detectors, and obviously the DQE,
184 A.R. Cowen
have a value falling somewhere between these two Indeed, the variation of DQE with the incident
extremes. The deterioration in recorded informa- detector dose is an important characteristic of
tion occurs for two principal reasons. Firstly, no dynamic detector performance. This can be ana-
detector can absorb all of the incident X-ray photons lyzed by measuring how DQE varies as a function
with 100% efficiency, and therefore some photons of the detector entrance dose per frame. In early
are lost to the image detection process. Inevitably, designs of dynamic FD, the image quality in fluo-
some X-ray photons pass straight through the roscopy fell below that of the digital X-ray IITV
X-ray absorber, while others that are absorbed may system; this was due to the comparatively pow-
then be re-emitted and escape the detector. This erful contribution of electronic noise in the FD
loss in primary X-ray information is compounded designs at that time [48, 51, 63, 80]. Today, this is
by various other noise sources in the detector itself; no longer an issue. Typical DQE performance for
an important example is the electronic noise that a modern indirect-conversion FD is summarized
arises in the a-Si:H matrix array and the signal in . Fig. 11.19. Equivalent results for a modern
readout circuitry. The DQE of a modern indirect- digital X-ray IITV system are included for ref-
conversion dynamic FD typically lies in the range erence. The ranges of the detector doses used in
of 0.70–0.80 for X-ray beam energies used in car- standard fluoroscopy, digital cardiac acquisition,
diovascular imaging [57, 58, 79]. Direct-conversion digital grayscale angiography, and DSA (where
dynamic FDs typically have a somewhat lower the dose per frame may reach or even exceed
DQE, with a value in the range of 0.50–0.60 [65]. that used in screen film radiography) are demar-
The influence of electronic noise on detector cated for reader orientation. The high DQE of the
DQE performance varies strongly with the signal indirect-conversion FD is maintained across the
level (and therefore the detector entrance dose). majority of the dynamic range. At best, FD DQE
11 1.0
Fluoroscopy Digital
DSA
Cardiac
Acquisition
Digital
Angiography
DQE(X)
(Grey-scale)
FD Grey-scale FD DSA
IITV Grey-scale IITV DSA
0.1
1 10 100 1,000 10,000
Dose per frame [nGy]
.. Fig. 11.19 Imaging performance of an indirect- variation of detective quantum efficiency (DQE) as a func-
conversion dynamic flat-panel detector (FD) compared tion of the detector entrance dose per frame in various
with a digital image intensifier television (IITV) system: imaging modalities
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
185 11
performance is typically 10–20% greater than In grayscale imaging, digital X-ray IITV sys-
that of a digital IITV–based digital fluoroscopy tems do not benefit from the online correction of
system [81, 82]. This finding would suggest that fixed-pattern noise, and hence, DQE is degraded
a modest (but not pronounced) improvement in at a high dose per frame. On the other hand, in
image quality and/or saving in patient/staff dose DSA, fixed-pattern noise is eliminated (along with
should be expected when moving from digital the patient background anatomy) by logarithmic
IITV–based to dynamic FD–based fluoroscopy. subtraction. As such, for DSA applications, the
This proposition has been supported by studies in DQE of the digital X-ray IITV at a high dose
cardio-angiography [83] and cardiac electrophys- per frame is improved compared with that of the
iology imaging [84]. However, other studies have grayscale (nonsubtracted) images. However, fol-
reported increases in procedural dose with the FD lowing subtraction, the random noise fluctuations
detector [85–87], albeit that better image quality are more powerful than in an individual image
(spatial resolution) was also usually indicated. It frame, for reasons explained in 7 Sect. 11.6.3.
is unclear from these studies to what extent the Consequently, the DQE values produced during
reported findings were dominated by differences DSA for both dynamic FD and digital X-ray IITV
in overall system design, calibration, and opera- are typically reduced by a factor of 2. However, at
tor technique selection rather than the FD itself the dose levels used in DSA, the FD still exhib-
[88]. For example, it has been reported in the USA its superior DQE compared to the digital X-ray
that when using FD-based cardiac systems, 90% IITV system (albeit less markedly so than for
reductions in dose are available for adult patients the grayscale comparison). Over the dose range
and 70–90% for pediatric patients by judicious used in cardiovascular imaging, the FD exhibits
selection of technique factors [89]. a constant DQE, confirming that the indirect-
For modern indirect-conversion FDs in fluo- conversion FD is a quantum-limited (and hence
roscopy, good DQE performance is maintained dose-efficient) image detector.
down to dose levels of less than 10 nGy per frame The set of graphs presented in . Fig. 11.19
[57, 58, 60, 79]. At exceptionally low dose lev- implies that modern indirect-conversion FDs can
els (say, well below 5 nGy per frame), a digital comfortably support the full spectrum of clinical
X-ray IITV system reportedly might offer better applications previously underwritten by digital
imaging performance; however, such a dose level X-ray IITV systems. Indeed, over the practical
would rarely be used in clinical practice. At higher working dose range used in cardiovascular imag-
dose levels (say, 1 μGy per frame and above), ing, the FD is a more dose-efficient detector than
X-ray IITV systems exhibit a marked deteriora- the digital X-ray IITV. This becomes apparent
tion in DQE, and the superiority of the dynamic when a dynamic FD is used to acquire X-ray pro-
FD becomes clearly apparent. This deterioration jection data for 3D reconstruction imaging.
in digital X-ray IITV system DQE results from the DQE is a powerful and versatile concept
growing impact of the granular structure of the II but can be augmented to take account of other
input and output phosphor screens. This random image quality characteristics, such as unsharp-
structure acts as a fixed-pattern noise source, ness in space (blur) and temporal unsharpness
whose contribution becomes more pronounced as (lag). This produces an even more comprehensive
the X-ray quantum noise diminishes with increas- analysis of dynamic FD performance. However,
ing dose. The raw image data produced by an such an analysis goes well beyond the scope of
FD are subject to a number of (well-recognized) this chapter.
cosmetic imperfections. Therefore, FD systems
incorporate automatic calibration and correction
routines designed to remove these imperfections 11.10 Digital Image Processing
and therefore optimize detector response. These
routines eliminate fixed-pattern noise in the FD Digital image processing has proved to be a pow-
plus other artifacts such as dark current offset, erful vector in the evolution of modern cardio-
defective (and dead) lines and pixels in the AM vascular X-ray imaging equipment. Undoubtedly,
array, and the memory effect (in some designs of continuing advances in digital image processing
system). (The overall image restoration process is will be required to support the even more com-
identified as «RES» in . Fig. 11.8.)
plex and exacting interventions that will emerge
186 A.R. Cowen
in the future. Salient digital image-processing [10, 91]. In this case, the original image is blurred
algorithms currently in use are reviewed in not with a large-area kernel as used in image har-
7 Sects. 11.10.1, 11.10.2, and 11.10.3.
monization but with a small area kernel aimed at
producing localized blurring. Subtraction of this
unsharp mask image from the original image pro-
11.10.1 Generic Image duces a contour map containing the fine-detail
Enhancement information. A weighted contribution of the con-
tour image is then added back on to the original
Generic digital image processing is needed to opti- image, producing the contour-enhanced version
mize the visual quality of the images presented to of the original image. The form and degree of
the clinician. Firstly, automatic digital restoration contour enhancement are controlled by the area
(identified as «RES» in . Fig. 11.8) is required to
of the unsharp mask and the weighting factor that
optimize the technical quality of the raw image is applied. The degrees of digital contrast and/or
data that are initially acquired. Such routines contour enhancement that can be applied to an
are embedded in dynamic FD–based imaging image are ultimately limited by the noise fluctua-
systems, as discussed in 7 Sect. 11.9.6. Secondly,
tions; these limitations are clearly apparent when
digital enhancement algorithms (identified as processing low-dose fluoroscopy images.
«ENH» in . Fig. 11.8) must then be applied to
The noise fluctuations visible during fluoros-
optimize the visual quality of the images finally copy can be reduced by temporal averaging of
presented on the display(s). consecutive image frames. Each presented image
Standard image enhancement techniques frame may, for example, comprise the sum of the
include transformation of the image grayscale current frame with, say, the preceding {N − 1}
via a lookup table (LUT) in order to optimize the frames in the fluoroscopic sequence. If the signal-
rendition of signals across the dynamic range. to-noise ratio of an individual frame is given by
One familiar grayscale transformation is known SNR{1}, then the improvement in the signal-
11 as white compression, which is used to suppress to-noise ratio resulting from time-averaging N
signal highlights but amplify (stretch) signal con- frames, assuming 100% weighting for each frame
trast at indicative (normally lower) levels of mean (SNR{N}) is described by:
signal [90]. This technique is particularly useful
in cardiac grayscale imaging, where the heart is SNR { N } = SNR {1} • Ö N
adjacent to the lungs, which exhibit regions of
very high X-ray transmission. Application of In practice, temporal noise averaging is normally
a suitable LUT improves visibility of coronary applied via a technique known as recursive filter-
structures while suppressing such highlights. ing. This algorithm applies a weighted summation
Digital image harmonization is another prac- of the preceding frames; however, the recursive
tical method of equalizing the display of signal weighting factor progressively diminishes in mag-
contrast across the dynamic range. Here a heavily nitude for frames acquired earlier in the sequence.
blurred (hence, unsharp) mask is created by con- Recursive filtering acts like a computer-generated
volving the input image with a large-area kernel. lag (temporal unsharpness), whose magnitude
This operation averages the image data over large- can be programmed. The recursion factor can be
area domains, producing a heavily blurred version configured by the manufacturer’s service engi-
of the original image. Subtraction of a fraction of neer for each clinical application to achieve the
this harmonization mask from the original image desired degree of noise reduction. Inevitably, any
data removes large-area trends in the image that target signal in the image that is moving—such as
contain little (if any) relevant clinical information. an opacified vessel, a guide wire, a catheter, or a
This also equalizes the intensity of the image field, stent—may however be blurred by generic recur-
allowing a greater degree of contrast amplification sive filtering. To be a clinically viable method, the
to be applied to the images prior to display [91]. recursive filter must be locally sensitive to the
Digital contour enhancement is widely used presence of target signal motion. Therefore, the
to sharpen fine details, small structures, and edge algorithm must be primed to automatically reset
discontinuities in the image. This has traditionally the recursion factor to prevent signal blur, as cir-
been achieved using the unsharp mask algorithm cumstances demand.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
187 11
Multiresolution (M-R) processing of dynamic 11.10.2 Targeted Image
X-ray images is increasingly being used to opti- Enhancement
mize imaging performance in cardiovascular
imaging [92, 93]. M-R processing is applied Digital image processing can also be used to
equally successfully in pulsed fluoroscopy and enhance the visualization of a specific target
serial image acquisition. The M-R processing structure. Such algorithms are typically targeted
algorithm typically decomposes each image frame at interventional devices such as stents, emboliza-
(in the sequence) into a set of sub-bands of dif- tion coils, or atrial septal defect closure devices.
fering spatial scale. These sub-bands encompass Algorithms can also be targeted at the stenotic
the range of scales from the finest details, through lesion itself. A commonly used implementation of
succeeding sizes, up to the coarsest structures. targeted enhancement is designed to enhance the
Each sub-band is processed independently using visibility of stents during coronary artery inter-
an intelligent filter design to separate (likely) ventions [94, 95]. This algorithm, often referred
vascular signals from random background noise to as stent boost, is briefly discussed here, as it
at that scale. The algorithm adaptively processes represents a working example of targeted image
each sub-band to enhance the identified signals processing.
but diminish the competing noise fluctuations. The optimum positioning and full expansion
The processed sub-bands are then recomposed to of a stent are important factors in ensuring the
produce the final enhanced image. . Figure 11.20
viability of a revascularization procedure. For
illustrates the improvement in image quality example, mal-apposition and underexpansion of
that is achievable with an M-R enhancement a stent are often associated with restenosis and
algorithm. The coronary angiogram on the left late thrombosis. Correct deployment of the stent
is subject to a grayscale optimization but is oth- relies upon adequate visualization at each stage
erwise unprocessed. The image on the right is in the intervention. Coronary stents are often
subject to spatial M-R enhancement; this image depicted with a low ΔSNR due to the small diam-
illustrates the improvement in the visualization eter and low X-ray attenuation of the stent struts
of coronary vessels that is achievable. The M-R (typically only 100–50 μm in size). In addition, the
algorithm can also be adapted to take account of visibility of the stent may be compromised by the
temporal (frame-to-frame) changes in image con- rapid movement of the host coronary vessel. This
tent. Temporal cues can then be used to improve problem is compounded by the fact that cardi-
the signal enhancement and noise suppression. In ologists normally have to rely upon 2D projection
interventional fluoroscopy, the marked improve- images of what are often irregular 3D structures.
ment in image quality attainable with M-R Targeted digital image enhancement algorithms
processing is often traded against a significant have therefore been developed to improve (boost)
reduction in the patient dose. the visualization of the stent, independently of
.. Fig. 11.20 Example of
generic image enhance-
ment: an unprocessed
coronary artery acquisition
image versus an image pro-
cessed with the multireso-
lution (M-R) enhancement
algorithm. (See «Acknowl-
edgments»)
ography, which clearly shows stent struts despite FDs represent a much more uniform, stable, and
the presence of significant calcification [96]. distortion-free platform for implementing 3D
Results with the stent boost algorithm X-ray imaging than the digital X-ray IITV system.
described above can be augmented by includ- As a result, the new FDs yield 3D reconstruc-
ing a reference overlay of the vessel lumen in the tions of high-contrast angiographic structures
image. This reference overlay can be derived from of greater spatial resolution [102, 103] and with
a contrast medium–filled image of the vessel, fewer artifacts [104, 105]. Corrections may still
which is used as a subtraction mask [97]. All of however be needed for any mechanical flexing of
the images needed for the process, including the the C-arm under gravity as it is rotated around the
contrast medium–filled vessels, can be acquired patient. 3D reconstruction imaging is now used
in the same acquisition run. The contrast medium to improve the visualization of complex networks
is typically injected for the final couple of seconds of blood vessels, particularly in neurovascular
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
189 11
procedures [106]. Reportedly, this can ease some using CBCT. The improvement in technical per-
of the difficulties experienced in navigating a formance afforded by the dynamic FD has made
device through a web of tortuous vessels prior to it possible to reconstruct high-quality 3D and
the intervention. Use of 3D reconstructed images (CT-like) 2D slice images of not only contrast
can reduce the duration of vascular procedures medium–enhanced vasculature but also the sur-
and contrast medium load [103, 107]. Patient rounding brain [117]. Soft-tissue (low-contrast)
radiation dose can be reduced significantly by CBCT reconstruction demands the acquisition
replacing multiple DSA runs by a single rotational of a set of projections with a finer angular inter-
acquisition; this has been confirmed in a number val than is required for high-contrast vascular
of neurovascular dose surveys [108–110]. imaging. This is essential to minimize artifacts in
3D angiographic images reconstructed from the reconstructed images due to angular under-
data acquired using a dynamic FD mounted on sampling. Typically, the number of projections
a C-arm are now widely used in interventional acquired has to be increased by a factor of 3–6
cardiovascular imaging. In principle, C-arm (or so) to achieve a 3D soft-tissue reconstruction
cone- beam CT (CBCT) can exhibit superior of acceptable quality. As a result, soft-tissue scan
spatial resolution to multislice CT. For an MSCT times increase, typically to 20–25 s or so duration;
scanner, the spatial resolution (in high-resolution this compares with the more usual 4–5 s used
mode) can measure ~1.5 lp mm−1. At full pixel in high-contrast (angiographic) reconstructions
resolution, C-arm CBCT typically reaches a spa- [118]. In addition, the dose per projection may
tial resolution limit of 4 lp mm−1 (given a realistic also be increased by a factor of ~2–4× to boost the
degree of geometrical magnification). In practice, SNR. The [4 × 4] pixel binning mode is also often
binned mode operation is normally used in 3D used in soft-tissue reconstructions to further
reconstructive imaging, either with [2 × 2] or reduce image noise. Interventionalists are finding
[4 × 4] pixel binning. The corresponding spatial that the immediacy, flexibility, and convenience
resolution limits in these modes typically reduce of C-arm-based 3D reconstructive imaging in its
to 2 and 1 lp mm−1, respectively. The dynamic various forms make it a valuable adjunct to stan-
range of an MSCT scanner can measure up to dard 2D imaging. It must be noted however that
106:1. This value can be compared with 104:1 for soft-tissue CBCT comes with an overhead in the
an FD-based (direct-conversion type) CBCT sys- patient dose compared with high-contrast (angio-
tem [107]. The large area of the FD ensures that in graphic) CBCT, potentially exceeding that used in
CBCT, a substantial volume of anatomy is covered MSCT [119].
in a single rotation of the C-arm. On the other To illustrate the high quality of C-arm-based
hand, the wide-angle, open X-ray beam geometry CBCT images achievable with an FD, we will
that is used with CBCT makes the projection data consider one specific 3D digital image-processing
much more susceptible to scattered X-rays than technique known as dynamic 3D road mapping
with MSCT scanner geometry. This limits the [120, 121]. Standard-projection (2D) road map-
low-contrast performance of CBCT, as discussed ping has been around for over 20 years and has
in 7 Sect. 11.6.4. The minimum contrast sensitiv-
proved to be a useful digital image-processing tool,
ity of a modern MSCT scanner, expressed as the particularly in neurovascular interventions [44,
minimum detectable difference in Hounsfield 122]. Here a grayscale (or possibly subtractive)
units (HUs), typically measures ~2–3 HUs. Under image of the vascular region of interest is acquired
laboratory conditions, the contrast sensitivity and is then used as a road map for the live fluo-
of FD-based CBCT reportedly measures 5 HUs roscopy image. The road map is used to aid navi-
[111]. In clinical practice, the contrast sensitivity gation of guide wires, catheters, or other devices
will normally be poorer than this value (10 HUs through irregular and tortuous sections of vessels.
or greater) [112–114]. Reportedly, the contrast A major drawback of 2D road mapping however is
sensitivity of CBCT is sufficient to visualize intra- that each time the interventionalist repositions the
cranial hemorrhage (albeit not with the precision C-arm, as the procedure progresses, a new refer-
achieved with an MSCT scanner) [115, 116]. ence road map has to be acquired. The additional
Digital X-ray IITV systems lacked the road maps take time to acquire and result in an
dynamic range and ΔSNR required to reliably increase in the contrast medium load and radia-
reconstruct low-contrast soft-tissue structures tion dose to the patient (and staff).
190 A.R. Cowen
AMA suitably modulates the light transmission of than relying upon multiple displays, a (very)
the LC layer. As a result, the device generates a large-screen FS display (56″ in size) was recently
pattern of light output, which forms the displayed introduced to provide the interventionalist with
image. Presentation of a time-varying sequence of a more flexible viewing environment. This large-
image frames generates a dynamic display suit- screen color display features a [2160 × 3840] or
able for viewing fluoroscopy and other types of 8.3-megapixel array. This so-called FlexVision
dynamic X-ray image. display [126] enables clinical staff to compose,
Modern FS displays offer excellent spatial reso- inspect, and evaluate images and data from mul-
lution, minimum contrast loss (veiling glare), and tiple signal sources. . Figure 11.24 shows a typical
a wide dynamic range. The recommended maxi- FlexVision display composition seen in a cardiac
mum and minimum specification values of out- interventional room. This display system also fea-
put luminance for diagnostic FS display devices tures a so-called superzoom facility, incorporating
are typically 500 and >170 cd m−2, respectively an advanced contour enhancement algorithm,
[125]. FS displays are electrically and thermally which enables selected images to be enlarged by
stable and, under normal circumstances, require over a factor of 4. This facility is used to improve
minimal operator or service engineer adjustment. the visualization and interpretation of particularly
They are also much less bulky and more energy complex, fine-detail cardiovascular structures.
efficient than the CRT. The geometry and unifor-
mity of FS displays are excellent and are immune
to distortions and interference due to extraneous 11.12 Technique Considerations
magnetic fields. Early designs of FS display could
have a restricted angle of view; however, this is Knowledge and understanding of X-ray coronary
no longer seen as a major issue. Reportedly, some angiography and fluoroscopy equipment provide
FS displays can exhibit a degree of lag (temporal a base for the development of imaging concepts
unsharpness); in principle, however, this might applicable to the clinical setting. Relevant techni-
benefit the display of fluoroscopic images, par- cal concepts have been reviewed in 7 Sects. 11.2,
ticularly those produced at low dose rates. 11.3, 11.4, 11.5, 11.6, 11.7, 11.8, 11.9, 11.10, and
FS displays come with a variety of (diagonal) 11.11. High image quality is always critical for
sizes, typically ranging from 18″ to 22″ for indi- adequate depiction of coronary morphology and
vidual image displays. The equivalent pixel reso- consequently for accurate clinical interpretation.
lutions of these FS displays typically range from Inevitably, there are potential radiation hazards
[1024 × 1280] to [1200 × 1600], corresponding associated with achieving this aim. The operator-
to 1.3- to 2-megapixel arrays, respectively. Rather dependent technique factors that impact upon
192 A.R. Cowen
image quality and radiation dose are now con- source-to-image-detector distance (SID) to the
sidered. Experience has been drawn from X-ray source-to-object distance (SOD), indicates the
imaging in interventional coronary angiography operator’s options to adjust the geometrical
(ICA); however, the principles carry over to many magnification (see 7 Sect. 11.4.5). In practice,
cular iodine contrast media, all with the basic titled «Technique Considerations,» was written by
structure of a benzene ring with symmetrically Peter Lanzer, MD, Health Care Center, Bitterfeld-
coupled iodine atoms. Vascular iodine contrast Wolfen, Germany.
media differ in important properties, including: The author gratefully acknowledges many
Water solubility, which determines the maxi- helpful discussions during the preparation of this
mum concentration of iodine that can be achieved chapter with his colleagues Amber Gislason and
and, therefore, the maximum X-ray absorption, Andrew Davies (at the University of Leeds) and
which generates coronary artery subject contrast Professor Mohan Sivananthan (of the Yorkshire
Osmolality, i.e., the osmotic pressure of the Heart Centre, LGI Leeds).
solution, determining the magnitude of water Amber Gislason kindly provided the X-ray
shifts between vascular and extravascular spaces spectra shown in . Figs. 11.1 and 11.13. Mr. Davies
Viscosity, which determines the force required and Professor Sivananthan kindly provided the
for intracoronary injections (warming the con- images used in . Figs. 11.20 and 11.21. The images
trast media to 37 °C before administration sharply used in . Figs. 11.11, 11.15, 11.16, 11.18, 11.22,
categorical course in physics. Oak Brook: RSNA; 1995. HJ. Technical and clinical results of an experimental
p. 49–58. flat dynamic (digital) X-ray image detector (FDXD)
35. Baldazzi G, Corazza I, Rossi PL, Testoni G, Bernardi T, systems with real-time correction. SPIE Proc Phys Med
Zannoli R. In vivo effectiveness of gadolinium filter for Imaging. 1998;3336:33–44.
paediatric cardiac angiography in terms of image qual- 54. Bury RF, Cowen AR, Davies AG, Baker EL, Hawkridge P,
ity and radiation exposure. Phys Med. 2002;28:109–13. Bruijns AJC, Reitsma H. Technical report: initial experi-
36. Rossi PL, Mariselli M, Corazza I, Bianchini D, Biffi M, ences with an experimental solid-state universal digi-
Martignani C, Zannoli R, Boriani G. Decrease in patient tal X-ray detector. Clin Radiol. 1998;53:923–8.
radiation exposure by a tantalum filter during electro- 55. Bruijns AJC, Bury R, Busse F, Davies AG, Cowen AR, Rut-
physiological procedures. Pacing Clin Electrophysiol. ten W, Reitsma H. Technical and clinical assessments of
2009;32(Suppl 1):S109–12. an experimental flat dynamic X-ray image detector sys-
37. Geise RA. Fluoroscopy: recording of fl
uoroscopic tem. SPIE Proc Phys Med Imaging. 1999;3659:324–35.
images and automatic exposure control. Radiograph- 56. Jung N, Alving PL, Busse F, Conrads N, Meulenbrugge
ics. 2001;21:227–36. HM, Rutten W, Schiebel UW, Weibrecht M, Wieczorek
38. Lin PP. The operation logic of automatic dose control HK. Dynamic X-ray imaging based on an amorphous
of fluoroscopy system in conjunction with spectral silicon thin-film array. SPIE Proc Phys Med Imaging.
filters. Med Phys. 2007;34:3169–72. 1998;3336:974–85.
39. Krohmer JS. Radiography and fluoroscopy, 1920 to the 57. Busse F, Rutten W, Sandkamp B, Alving PL, Bastiaens
present. Radiographics. 1989;9:1129–53. RJM, Ducourant T. Design and performance of a high
40. Schueler BA. General overview of fluoroscopic imag- quality cardiac flat panel detector. SPIE Proc Phys Med
ing. Radiographics. 2000;20:1115–26. Imaging. 2002;4682:819–27.
41. Wang J, Blackburn TJ. X-ray image intensifiers for fluo- 58. Granfors PR, Aufrichtig R, Netel H, Brunst G, Boudry
roscopy. Radiographics. 2000;20:1471–7. JM, Luo D, Albagli D, Tkaczyk JE. Performance of a
42. Van Lysel MS. Fluoroscopy: optical coupling and the flat cardiac detector. SPIE Proc Phys Med Imaging.
video system. Radiographics. 2000;20:1769–86. 2001;4320:77–86.
43. Snoeren RM, ten Caat RB, Dillen BGM, Gieles P, van der 59. Sivananthan UM, Moore J, Cowan JC, Pepper CB,
Veen JCT. Solid state image sensor in X-ray television. Hunter S, Cowen AR, Davies AG, Kengyelics SM. A flat-
Medicamundi. 1991;36:203–11. detector cardiac cath lab system in clinical practice.
44. Pooley RA, McKinney JM, Miller DA. Digital fluoros- Medicamundi. 2004;48:4–12.
11 copy. Radiographics. 2001;21:521–34.
45. Neitzel U. Recent technological developments and
60. Granfors PR, Aufrichtig R, Possin GE, Giambattista BW,
Huang ZS, Liu J, Ma B. Performance of a 41 × 41 cm2
their influence. Radiat Prot Dosim. 2000;90(1–2):15–20. amorphous silicon flat panel X-ray detector designed
46. Powell A, Katzen B. First experiences with a CCD sys- for angiographic and R&F imaging applications. Med
tem in interventional radiology: the Integris V5000. Phys. 2003;30:2715–26.
Medicamundi. 1999;43(4):38–44. 61. Ducourant T, Couder D, Wirth T, Trochet JC, Bastiaens R,
47. Cowen AR, Kengyelics SM, Davies AG. Solid-state flat- Bruijns T, Luijendijk HA, Sandkamp B, Davies AG, Didier
panel digital radiography detectors and their physical D, Gonzalez A, Terraz S, Ruefenacht D, et al. Image
imaging characteristics. Clin Radiol. 2008;63:487–98. quality of digital subtraction angiography using flat
48. Schiebel U, Conrads N, Jung N, Weibrecht M, Wiec- detector technology. SPIE Proc Phys Med Imaging.
zorek H, Zaengel T. Fluoroscopic X-ray imaging with 2003;5030:203–14.
amorphous silicon thin-film arrays. SPIE Proc Phys Med 62. Bruijns AJC, Bastiaens R, Hoornaert B, von Reth E,
Imaging. 1994;2162:129–40. Busse F, Heer VK, Ducourant T, Cowen AR, Davies AG,
49. Antonuk LE, Yorkston J, Huang W, Siewerdsen JH,
Terrier F. Image quality of a large-area dynamic flat
Boudry JM, El-Mohri Y. A real-time, flat-panel amor- detector: comparison with a state-of-the-art IITV sys-
phous silicon digital X-ray imager. Radiographics. tem. SPIE Proc Phys Med Imaging. 2002;4682:332–43.
1995;15:993–1000. 63. Colbeth RE, Boyce S, Fong R, Gray K, Harris R, Job ID,
50. Chabbal J, Chaussat T, Ducourant T, Fritsch L, Michailos Mollov IP, Nepo B, Pakovich JM, Taie-Nobarie N, Seppi EJ,
J, Spinnler V, Vieux G, Arques M, Hahm G, Hoheisel M, Shapiro EG, Wright MD, Webb C, Yu JM. 40 × 30 cm2 flat
Horbaschek H, Schulz RF, Spahn MF. Amorphous sili- imager for angiography, R&F and cone-beam CT appli-
con X-ray image sensor. SPIE Proc Phys Med Imaging. cations. SPIE Proc Phys Med Imaging. 2001;4320:94–102.
1996;2708:499–510. 64. Choquette M, Demers Y, Shukri Z, Tousignant O, Aoki K,
51. Colbeth RE, Allen MJ, Day DJ, Gilblom DL, Klaus Meijer- Honda M, Takahashi A, Tsukamoto A. Real time perfor-
Brown ME, Pavkovich J, Seppi EJ, Shapiro EG. Charac- mance of a selenium based detector for fluoroscopy.
terisation of an amorphous silicon fluoroscopic imager. SPIE Proc Phys Med Imaging. 2001;4320:501–8.
SPIE Proc Phys Med Imaging. 1997;3032:42–51. 65. Tousignant O, Demers Y, Laperriere L, Nishiki M,
52. Colbeth RE, Allen MJ, Day DJ, Gilblom DL, Harris R, Job Nagai S, Tomisaki T, Takahashi A, Aoki K. Clinical per-
ID, Klausmeier-Brown ME, Pavkovich JM, Seppi EJ, Sha- formances of a 14″ × 14″ real time amorphous sele-
piro EG, Wright MD, Yu J. Flat panel imaging system for nium flat panel detector. SPIE Proc Phys Med Imaging.
fluoroscopy applications. SPIE Proc Phys Med Imaging. 2003;5030:71–6.
1998;3336:376–87. 66. Asahina H. Selenium-based flat panel X-ray detector
53. Bruijns TJ, Alving PL, Baker EL, Bury RF, Cowen AR, for digital fluoroscopy and radiography. Toshiba Med
Jung N, Luijendijk HA, Meulenbrugge HJ, Stouten Rev. 1999;69:1–7.
Cardiovascular X-ray Imaging: Physics, Equipment, and Techniques
197 11
67. Tousignant O, Demers Y, Lapierre L, Marcovici S. (2007). 82. Baker EL, Cowen AR, Kemner R, Bastiaens R. A physi-
A-Se flat panel detectors for medical applications. Sen- cal evaluation of a CCD-based X-ray II fluorography
sors applications symposium IEEE, San Diego, Califor- system for cardiac applications. SPIE Proc Phys Med
nia USA. Imaging. 1998;3336:430–41.
68. Spahn M. Flat detectors and their clinical applications. 83. Vano E, Geiger B, Schreiner A, Back C, Beissel
Eur Radiol. 2005;15:1934–47. J. Dynamic flat panel detector versus image intensi-
69. Nikoloff EL. Survey of modern fluoroscopy imaging: fier in cardiac: dose and image quality. Phys Med Biol.
flat-panel detectors versus image intensifiers and 2005;50:5731–42.
more. Radiographics. 2011;31:591–602. 84. Davies AG, Cowen AR, Kengyelics SM, Moore J, Pepper
70. Roos PG, Colbeth RE, Mollov I, Munro P, Pavkovich J, C, Cowen C, Sivananthan UM. X-ray dose reduction
Seppi EJ, Shapiro EG, Tognina CA, Virshup GF, Yu M, in fluoroscopically guided electrophysiology proce-
Zentai G, Kaissi W, Matsinos A, Richters J, Riehm H. Mul- dures. Pacing Clin Electrophysiol. 2006;29:262–71.
tiple-gain-ranging readout method to extend the 85. Prasan AM, Ison G, Rees DM. Radiation exposure dur-
dynamic range of amorphous silicon flat-panel imag- ing elective coronary angioplasty: the effect of flat-
ers. SPIE Proc Phys Med Imaging. 2004;5368:139–49. panel detection. Heart Lung Circ. 2008;17:215–9.
71. Boyce SJ, Chawla A, Samei E. Physical evaluation of a 86. Trianni A, Bernardi G, Padovani R. Are new technolo-
high frame rate, extended dynamic range flat panel gies always reducing patient doses in cardiac proce-
detector for real-time cone beam computed tomog- dures. Radiat Prot Dosim. 2005;117:97–101.
raphy applications. SPIE Proc Phys Med Imaging. 87. Tsapaki V, Kottou S, Kollaros N, Dafnomili P, Kyriakidis Z,
2005;5745:591–9. Neofotistou V. Dose performance evaluation of charge
72. Fahrig R, Wen Z, Ganguly A, DeCrescenzo G, Rowlands coupled device and a flat-panel digital fluoroscopy sys-
JA, Stevens GM, Saunders RF, Pelc NJ. Performance of tem recently installed in an interventional cardiology
a static-anode/flat-panel X-ray fluoroscopy system in a laboratory. Radiat Prot Dosim. 2004;111(3):297–304.
diagnostic strength magnetic field: truly hybrid X-ray/ 88. Davies AG, Cowen AR, Kengyelics SM, Moore J, Sivana-
MR imaging system. Med Phys. 2005;32:1775–84. nthan MU. Do flat detector cardiac X-ray systems con-
73. Achenbach S, Ropers D, Holle J, Muschol G, Daniel WG, vey advantages over image intensifier-based systems?
Moshage W. In-plane coronary arterial motion veloc- Study comparing X-ray dose and image quality. Eur
ity: measurement with electron-beam CT. Radiology. Radiol. 2007;17:1787–94.
2000;216:457–63. 89. Nikoloff EL, Lu ZF, Dutta A, So J, Balter S, Moses J. Influ-
74. Zhao W, DeCresenzo G, Rowlands JA. Investigation ence of flat-panel fluoroscopic equipment variables on
of lag and ghosting in amorphous selenium flat- cardiac radiation doses. Cardiovasc Intervent Radiol.
panel X-ray detectors. SPIE Proc Phys Med Imaging. 2007;30:169–76.
2003;4682:9–20. 90. Cowen AR. Image processing in digital radiography.
75. Siewerdsen JH, Jaffray DA. A ghost story: spatio-tem- Imaging. 1994;6:77–99.
poral response characteristics of an indirect-detection 91. Cowen AR, Hartley PJ, Workman A. The computer
flat-panel imager. Med Phys. 1999;26:1624–41. enhancement of digital grey-scale fluorography
76. Overdick M, Solf T, Wischmann H-A. Temporal artefacts images. Br J Radiol. 1988;61(726):492–500.
in flat dynamic X-ray detectors. SPIE Proc Phys Med 92. Aach T, Mayntz C, Rongen P, Schmitz G, Stegehuis
Imaging. 2001;4320:47–58. H. Spatiotemporal multiscale vessel enhancement for
77. Ducourant T, Michel M, Vieux G, Peppler T, Trochet JC, coronary angiograms. SPIE Proc Phys Med Imaging.
Schulz RF, Bastiaens RJM, Busse F. Optimization of key 2002;4684:1010–21.
building blocks for a large area radiographic and fluo- 93. Wu Z, Fang M, Qian J, Schramm H. A multi-scale
roscopic dynamic X-ray detector based on a-Si:H/CsI:Tl adaptive method for blood vessel enhancement
flat panel technology. SPIE Proc Phys Med Imaging. in X-ray angiography. SPIE Proc Phys Med Imaging.
2000;3977:14–25. 1997;3036:326–35.
78. Dainty JC, Shaw R. Image science. London: Academic 94. Koolen JJ, Van Het Veer M, Hanekamp CEE. Stentboost
Press; 1975. image enhancement: first clinical experience. Medica-
79. Tognina CA, Mollov I, Yu JM, Webb C, Roos PG, Batts M, mundi. 2005;49(2):4–8.
Trinh D, Fong R, Taie-Nobriae N, Nepo B, Job IS, Gray 95. Mishell JM, Vakharia KT, Ports TA, Yeghiazians Y,
K, Boyce S, Colbeth RE. Design and performance of a Michaels AD. Determination of adequate coronary
new a-Si flat panel imager for use in cardiovascular stent expansion using stentboost, a novel fluoroscopic
and mobile C-arm imaging systems. SPIE Proc Phys image processing technique. Catheter Cardiovasc
Med Imaging. 2004;5368:648–56. Interv. 2007;69:84–93.
80. Davies AG, Cowen AR, Kengyelics SM, Bury RF, Bruijns 96. Sivananthan UM, Blackburn M, Cowan JC, Mclenachan
TJ. Threshold contrast detail detectability measure- J, Pepper CB, Hunter S, Moore J, Cowen AR, Davies AG,
ment of the fluoroscopic image quality of a dynamic Kengyelics SM. Cardiac cath lab upgrade improves
solid-state digital X-ray image detector. Med Phys. efficiency and reduces dose. Medicamundi. 2006;
2001;28:11–5. 50(2):1–9.
81. Spekowius G, Boerner H, Eckenbach W, Quadflieg
97. Agostini P, Verheye S. Bifurcation stenting with dedi-
P, Laurenssen GJ. Simulation of the imaging perfor- cated biolimus-eluting stent: X-ray visual enhance-
mance of X-ray image intensifier TV camera chains. ment of the angiographic result with “StentBoost”.
SPIE Proc Phys Med Imaging. 1995;2432:12–23. Catheter Cardiovasc Interv. 2007;70:233–6.
198 A.R. Cowen
98. Eng MH, Klein AP, Wink O, Hansgen A, Carroll JD, TJ. Flat-panel CT: fundamental principles, technol-
Garcia JA. Enhanced stent visualization: a case series ogy & applications. Radiographics. 2008;28:2009–22.
demonstrating practical applications during PCI. Int 112. Smyth JM, Sutton DG, Houston JG. Evaluation of
J Cardiol. 2010;141(1):e8–e16. the quality of CT-like images obtained using a com-
99. Schoonenberg G, Florent R. Advanced visibility mercial flat panel detector system. Biomed Imaging
enhancement for stents and other devices: image Interv J. 2006;2(4):e48.
processing aspects. Cardiol Clin. 2009;27:477–90. 113. Wallace MJ, Kuo M, Glaiberman C, Binkert CA, Orh
100. Fahrig R, Fox AJ, Lownie S, Holdsworth DW. Use of RC, Soulez G. Three dimensional C-arm cone-beam
a C-arm system to generate true three-dimensional CT: applications in the interventional suite. J Vasc
computed rotational angiograms: preliminary Interv Radiol. 2008;19:799–813.
in vitro and in vivo results. AJNR Am J Neuroradiol. 114. Miracle AC, Mukherji SK. Conebeam CT of the head
1997;18:1507–14. and neck, part 1: physical principles. AJNR Am J Neu-
101. Fahrig R, Moreau M, Holdsworth DW. Three dimen- roradiol. 2009;30:1088–95.
sional computed tomographic reconstruction using 115. Orth RC, Wallace MJ, Kuo MD. C-arm cone-beam CT:
C-arm mounted XRII: correction of image intensifier general principles and technical considerations for
distortion. Med Phys. 1997;24:1097–106. use in interventional radiology. J Vasc Interv Radiol.
102. Baba R, Konno Y, Ueda K, Ikeda S. Comparison of 2008;19:814–21.
flat-panel detector and image intensifier detector 116. Struffert T, Eyopglu IY, Huttner HB, Engelhorn T,
for cone-beam CT. Comput Med Imaging Graph. Doelken M, Saake M, Ganslandt O, Doerfler A. Clini-
2002;6:153–8. cal evaluation of flat-panel detector compared with
103. Hirota S, Nakao N, Yamamoto S, Kobayashi K, Maeda multi-slice computed tomography in 65 patients
H, Ishikura R, Miura K, Sakamoto K, Ueda K, Baba with acute intracranial haemorrhage: initial results.
R. Cone-beam CT with flat-panel detector digital J Neurosurg. 2010;113:901–7.
angiography system: early experiences in abdomi- 117. Heran NS, Song JK, Mamba K, Smith W, Niimi Y,
nal interventional procedures. Cardiovasc Intervent Berenstein A. The utility of DynaCT in neurovascu-
Radiol. 2006;29:1034–8. lar procedures. AJNR Am J Neuroradiol. 2006;27:
104. Hirai T, Korogi Y, Ono K, Yamura M, Uemara S, Yamashita 330–2.
Y. Pseudostenosis phenomenon at volume-rendered 118. Kamran M, Nagaraja S, Byrne JV. C-arm flat detector
three-dimensional digital angiography of intracranial computed tomography: the technique and its appli-
11 arteries: frequency, location and effect on image
evaluation. Radiology. 2004;232:882–7.
cation in interventional neuro-radiology. Neuroradi-
ology. 2010;52:319–27.
105. Kakeda S, Korogi Y, Ohnari N, Hatakeyama Y, Moriya J, 119. Kyriakou Y, Richter DA, Kalendar WA. Neuroradiologic
Oda N, Nishino K, Miyamoto W. 3D digital subtraction applications with routine C-arm flat panel detector
angiography of intracranial aneurysms: comparison CT: evaluation of patient dose measurements. AJNR
of flat panel detector with conventional IITV system Am J Neuroradiol. 2008;29:1930–6.
using a vascular phantom. AJNR Am J Neuroradiol. 120. Soderman M, Babic D, Homan R, Andersson T. 3D
2007;28:839–43. roadmap in neuroangiography: technique and clini-
106. Sugahara T, Korogi Y, Nakashima K, Hamatake S,
cal interest. Neuroradiology. 2005;47:735–40.
Honda S, Takahashi M. Comparison of 2D and 3D 121. Wilhelm K, Babic D. 3D angiography in the interven-
digital subtraction angiography in evaluation of tional clinical routine. Medicamundi. 2006;50:24–31.
intracranial aneurysms. AJNR Am J Neuroradiol. 122. Turski PA, Stieghorst MF, Strother CM, Crummy
2002;23:1545–52. AB, Lieberman RP, Mistretta CA. Digital subtrac-
107. Hatakeyama Y, Kakeda S, Korogi Y, Ohnari N, Moriya tion angiography “road map”. Am J Roentgenol.
J, Oda N, Nishino K, Miyamoto W. Intracranial 2D 1982;139:1233–4.
and 3D DSA with flat panel detector of the direct 123. Racadio JM, Babic D, Homan R, Rampton JW, Patel
conversion type: initial experience. Eur Radiol. MN, Racadio JM, Johnson ND. Live 3D guidance in
2006;16:2594–602. the interventional radiology suite. Am J Roentgenol.
108. Schueler BA, Kallmes DF, Cloft HJ. 3D cerebral angi- 2007;189:357–64.
ography: radiation dose comparison with digital 124. Badano A. AAPM/RSNA tutorial on equipment selec-
subtraction angiography. AJNR Am J Neuroradiol. tion: PACS equipment overview. Display systems.
2005;26:1898–901. Radiographics. 2004;24(3):879–89.
109. Bridcut RR, Murphy E, Workman A, Flynn P, Winder 125. The Royal College of Radiologists. (2008). Picture
RJ. Patient dose from 3D rotational neurovascular archiving and communication systems (PACS) and
studies. Br J Radiol. 2007;80:362–6. guidelines on diagnostic display devices. This guid-
110. Tsapaki V, Vano E, Mavrikou I, Neofotistou V, Gallego ance is only available electronically from: www.rcr.
JJ, Fernandez JM, Santos E, Mendez J. Comparison of ac.uk (Accessed March 2012).
patient dose in two-dimensional carotid arteriogra- 126. Drost MM. Evaluation of a recently developed
phy and three-dimensional rotational angiography. 56″ monitor in CV interventions. Medicamundi.
Cardiovasc Intervent Radiol. 2008;31:477–82. 2009;53(3):24–8.
111. Gupta R, Cheung AC, Bartling SH, Lisauskas J,
127. Gurley JC. Flat detectors and new safety aspects of
Grasruck M, Leidecker C, Schmidt B, Flohr T, Brady radiation safety. Cardiol Clin. 2009;27:385–94.