Sunteți pe pagina 1din 2

DIST D-2892/5236 CC

Computer Controlled-Combined Distillation System according to


ASTM D-2892 (TBP)
ASTM D-5236 (Potstill)

The i-Fischer® Dist D-2892 / 5236 CC is a vers and the determination of the fraction
fully computer controlled unit of turn-key volume as well as for the direct distillation
design, and ready for use after installation rate control.
and commissioning. The system is fully hou- The automatic fraction collector (in the
sed and equipped with doors in the front system section Potstill: with 12 receivers)
and rear to satisfy safety requirements and includes a built-in internal balance and vo-
to facilitate service aspects. lume follower system, which is used for the
The automatic fraction collector (in the sy- simultaneous determination of the fraction
stem section TBP: with 20 receivers) inclu- weight, volume and for the direct distillation
des a built-in internal balance for the deter- rate control.
mination of the fraction weight, while the The vacuum equipment and the control
separate volume follower system is used for system are used for alternative operation
discharging the fractions into the final recei- of both distillation processes.

4
expect essentials

Technical Data automatic closing of the final receivers sophisticated vacuum control with
Operation Temperature: IBP ... 575°C (TBP section only) dynamic vacuum reduction procedure
AET (depending on product) automatic fraction/receiver changes sophisticated safety and product
Operation Pressure: 0.1 ... 760 mmHg according to pre-selected cut tempera- cracking checks
(Torr) tures and/or receiver volume media supply check (nitrogen, instru-
Fraction collector: 20 receivers (TBP) determination of fraction volumes mentation air, cooling water)
12 receivers, heated by IR-radiators and weights by built-in volume follo- all alarms & safety checks with atten-
(Potstill) wer system and internal balance ded/unattended mode
"supervisory”-mode for manual over- various options available: e.g. "water
Design Essentials ruling of automatic control mode – removal/dehydration”, "density mea-
operation through PC and flow chart with indication of every manual inter- surement”, "safety devices”, "calibra-
(process parameters are indicated digi- vention in the flow chart tion devices”, "high vacuum extension
tally and are presented in coloured => manual/semi-automatic operation of (final cut range up to 620°…650°C
curves as a diagram "parameter vs. the unit (in emergency cases or for AET)”, "software extension 2x 20 re-
time") – easy unit management also maintenance etc.) ceivers for collecting narrow cuts”
by less experienced operators => manual receiver change
user-software application operated automatic debutanisation procedure
under WINDOWS™XP automatic continuation of the TBP-
"log-book” with permanent data process, e.g. with the next vacuum run
report and data storage to be effected
reported history of distillation automatic cooling down procedure
processes after a distillation run
storage of distillation processes with programmable washing/cleaning runs
parameters and data, limited by OS for different solvents
("operator system”) only evaluation of final data and TBP-curve
pre-selectable heating modes, e.g. electrical lifts for heating jackets
indirect/direct distillation rate control integrated UPS-system for PC, control
fully automatic fraction discharge into system for data storage up to 10
final receivers without any disturban- minutes during a power failure
ce of the column equilibrium

S-ar putea să vă placă și