Documente Academic
Documente Profesional
Documente Cultură
Model 325
OPERATOR’S MANUAL
Revision C
mocon®
7500 Mendelssohn Avenue North
Minneapolis, Minnesota 55428 USA
Telephone: 763-493-7250 Fax: 763-493-6358
Technical Service 763-493-7250
Customer Service 763-493-7265
E-mail Address: instrument_support_services@mocon.com
http://www.mocon.com
Patents Pending
U.S. Patent Number 5,212,993 Part Number 143-162
Mocon and PAC CHECK are registered trademarks of Mocon, Inc.
LIQUI-BLOK is a registered trademark of Mocon, Inc.
PAC CHECK Model 325 Operator’s Manual About This Manual
Read Chapter 2, “Preparing the PAC CHECK Model 325 for Use.”
1. Understand the controls and operation of the PAC CHECK Model 325:
Safety Information
Be sure to read and understand this section and applicable chapters of this Operator's Manual before
setting up, operating and maintaining this machine.
To avoid personal injury and equipment damage, observe the following precautions:
Installation Precautions
• The power cord for the battery charger must be plugged into an appropriate power source as
specified in Chapter 4 "Specifications". Any other type of power source could damage the
instrument.
Operating Precautions
Your PAC CHECK Model 325 will provide you with years of maintenance-free use if the following
precautions and recommendations are observed.
• Use a Liqui-Block® filter (MOCON P/N 031-371) when liquid contamination of this instrument
is possible. Filters prevent moisture and/or other contaminants in the gas stream from
entering into the analyzer. Moisture and/or contaminants may damage the sensor.
• Use the proper needle with your PAC CHECK Model 325 (MOCON P/N 035-137).
• Periodically check the calibration of your PAC CHECK Model 325 for optimum accuracy.
• Clean the PAC CHECK Model 325 with a moistened cloth; never use water or any other
liquid to clean it.
• If the optional sticky foot assembly has been purchased, always lock the sticky foot assembly
in the closed position while not in use.
Service Precautions
SERVICE NOTE:
Should you have any questions about your system, please call MOCON Technical Services in the USA at
(763) 493-7250. We want you to receive the best in product support services
Table of Contents
About This Manual .............................................................................................................................................................iii
Chapter 1: Introduction....................................................................................................................................................1-1
Chapter 2: Preparing the PAC CHECK Model 325 for Use ............................................................................................2-1
Setting Up the Instrument .........................................................................................................................................2-1
Charging the Batteries ..............................................................................................................................................2-1
Choosing a Sampling Configuration .........................................................................................................................2-1
Automatic Mode using the Coiled Tube Assembly.............................................................................................2-2
Automatic Mode using a Filter and Needle ........................................................................................................2-3
Manual Mode using the optional Injection Port ..................................................................................................2-4
Connecting the RPPF ........................................................................................................................................2-5
Operating the RPPF...........................................................................................................................................2-6
Chapter 3: Operating your PAC CHECK Model 325 for Use ..........................................................................................3-1
Power-Up..................................................................................................................................................................3-1
Battery Monitoring.....................................................................................................................................................3-1
Power Down .............................................................................................................................................................3-1
Display Screens........................................................................................................................................................3-2
Idle Screen .........................................................................................................................................................3-2
Results Screen...................................................................................................................................................3-2
Measuring a Gas Sample in Automatic Mode ..........................................................................................................3-3
Measuring a Gas Sample in Manual Mode ..............................................................................................................3-3
Main Menu ................................................................................................................................................................3-4
Password Screen .....................................................................................................................................................3-4
Main Menu ► Products Menu ..................................................................................................................................3-5
Main Menu ► Users Menu .......................................................................................................................................3-8
Main Menu ► Setup Menu .....................................................................................................................................3-10
Main Menu ► Calibration Menu .............................................................................................................................3-16
Chapter 5: Maintenance..................................................................................................................................................5-1
Cleaning the PAC CHECK 325 ................................................................................................................................5-1
Replacing the Liqui-Blok® Filter ................................................................................................................................5-1
Maintaining the Manual Injection Port ......................................................................................................................5-2
Replacing the Filter Chip and Rubber Inlet...............................................................................................................5-3
Maintaining the RPPF...............................................................................................................................................5-4
Servicing the Batteries..............................................................................................................................................5-5
Replacing the Oxygen Sensor ..................................................................................................................................5-6
Figures
Figure 1-1: PAC CHECK Model 325 ............................................................................................................ 1-1
Figure 2-1: Automatic Mode with Tube Assembly........................................................................................ 2-2
Figure 2-2: Automatic Mode......................................................................................................................... 2-3
Figure 2-3: Manual Mode ............................................................................................................................. 2-4
Figure 2-4: Setting up the RPPF .................................................................................................................. 2-5
Figure 3-1: Idle Screen ................................................................................................................................ 3-2
Figure 3-2: Run Screen................................................................................................................................ 3-2
Figure 3-3: Results Screen .......................................................................................................................... 3-2
Figure 3-4: Main Menu ................................................................................................................................. 3-4
Figure 3-5: Password Screen ...................................................................................................................... 3-4
Figure 3-6: Products Menu .......................................................................................................................... 3-5
Figure 3-7: Select Product Screen............................................................................................................... 3-5
Figure 3-8: Product Name Screen ............................................................................................................... 3-5
Figure 3-9: Rename Product Screen ........................................................................................................... 3-6
Figure 3-10: Rename Product Advisory Screen........................................................................................... 3-6
Figure 3-11: Test Method Screen ................................................................................................................ 3-7
Figure 3-12: Users Menu Screen ................................................................................................................. 3-8
Figure 3-13: Users Select Screen................................................................................................................ 3-8
Figure 3-14: User Name Screen .................................................................................................................. 3-8
Figure 3-15: Rename User Screen .............................................................................................................. 3-9
Figure 3-16: Rename User Advisory Screen ............................................................................................... 3-9
Figure 3-17: Setup Menu Screen ............................................................................................................... 3-10
Figure 3-18: Language Screen .................................................................................................................. 3-10
Figure 3-19: Password Screen .................................................................................................................. 3-10
Figure 3-20: Screen/Audio Menu Screen................................................................................................... 3-11
Figure 3-21: Screen Backlight Menu Screen ............................................................................................. 3-11
Figure 3-22: Brightness Menu Screen ....................................................................................................... 3-11
Figure 3-23: Backlight Shutoff Time Screen .............................................................................................. 3-12
Figure 3-24: Audio Menu Screen ............................................................................................................... 3-12
Figure 3-25: Time/Date Screen.................................................................................................................. 3-13
Figure 3-26: Set Time/Date Screen ........................................................................................................... 3-13
Figure 3-27: Date Format Screen .............................................................................................................. 3-13
Figure 3-28: Shutdown Time Screen ......................................................................................................... 3-14
Figure 3-29: Test Settings Screen ............................................................................................................. 3-14
Figure 3-30: Alarms Screen ....................................................................................................................... 3-14
Figure 3-31: Low Alarm Screen ................................................................................................................. 3-15
Figure 3-32: Pump Time Screen ................................................................................................................ 3-15
Figure 3-33: Auto Store Screen ................................................................................................................. 3-15
Figure 3-34: Auto Send Screen ................................................................................................................. 3-16
Figure 3-35: Calibration Menu Screen ....................................................................................................... 3-16
Figure 3-36: Units Menu Screen ................................................................................................................ 3-17
Figure 3-37: Pressure Screen.................................................................................................................... 3-17
Figure 3-38: Temperature Screen.............................................................................................................. 3-17
Figure 5-1: Injection Port Assembly ............................................................................................................. 5-2
Figure 5-2: RPPF Components.................................................................................................................... 5-4
Figure 5-3: Servicing the Battery ................................................................................................................. 5-5
Figure 5-4: Removing the Back Cover ......................................................................................................... 5-6
Figure 5-5: Battery Removal ........................................................................................................................ 5-7
Figure 5-6: Unplugging the Keypad ............................................................................................................. 5-8
Figure 5-7: Moving the LCD ......................................................................................................................... 5-9
Figure 5-8: Removing the Circuit Board..................................................................................................... 5-10
Figure 5-9: Removing the Oxygen Sensor................................................................................................. 5-10
vi Revision C MOCON, Inc.
PAC CHECK Model 325 Operator’s Manual Table of Contents
Tables
Table 2-1: Automatic Mode with Tube Assembly......................................................................................... 2-2
Table 2-2: Automatic Mode .......................................................................................................................... 2-3
Table 2-3: Manual Mode .............................................................................................................................. 2-4
Table 5-1: Injection Port Assembly Components ......................................................................................... 5-2
Table 5-2: RPPF Components .................................................................................................................... 5-4
Chapter 1: Introduction
This chapter provides an overview of the PAC CHECK Model 325.
The analyzer measures the oxygen and carbon dioxide concentration of a gas mixture by drawing a sample
through an internal infrared and electro-chemical cell. The electro-chemical sensor generates an electrical
current that is converted to an oxygen percentage on an easy-to-read LCD display. The IR sensor
generates a voltage that is inversely proportional to the carbon dioxide concentration. The analyzer utilizes
a high-quality NiMH rechargeable battery pack for power.
The PAC CHECK Model 325 uses Cal-Smart™ technology to insure the best possible oxygen sensor
accuracy. The Cal-Smart™ feature allows the oxygen sensor to be accurately calibrated using only room
air.
The PAC CHECK Model 325 is also equipped with an internal memory that can store up to 240
measurements at a time. Data is kept in memory even when the machine is turned off, and can be
transferred to a PC using an optional software package available from MOCON Inc.
To take full advantage of your analyzer’s capabilities, we recommend that you read through this manual in
its entirety before operating the PAC CHECK Model 325 for the first time.
The instrument is factory calibrated before shipment; however, you may want to verify the accuracy of the
calibration before putting your instrument into operation. See Appendix A for information on checking
calibration on your PAC CHECK Model 325.
MOCON, Inc. Revision C 2-1
Preparing the PAC CHECK Model 325 for Use PAC CHECK Model 325 Operator’s Manual
2 Connection for the AC Adapter for charging the battery pack N/A
2 Connection for the AC Adapter for charging the battery pack N/A
2. Install the Manual Inlet Guide to the Sample Inlet Port as shown Figure 2-3.
2 Connection for the AC Adapter for charging the battery pack N/A
1. Mount the shaft of the piercing assembly onto the platform base. Make sure the alignment
pin located on the bottom of the shaft fits into the slot in the hole of the platform base.
2. Using a ¾ inch wrench, secure the piercing assembly to the platform by tightening the nut.
3. Install the Y-shaped package support on the lower section of the shaft.
4. Connect the wide orifice of the injection tube (MOCON part number 380-047) to the
discharge end of the two-way valve (MOCON part number 300-040). Refer to Figure 2-4
below.
5. Connect the luer adapter (MOCON part number 035-255) to the narrow orifice of the injection
tube. Refer to Figure 2-4 below.
6. Connect the narrow orifice of the LIQUI-BLOK filter to the inlet of the two-way valve (MOCON
part number 300-040) on the RPPF.
7. Connect the suction tube (MOCON part number 380-046) to the wide orifice of the LIQUI-
BLOK filter.
8. Press the other end of the suction tube into the hole in the flow insert.
Two Way
Valve
LIQUI-
BLOK
Filter
Flow Insert
Injection Tube Knurled
Thumb Nut
Luer Adapter
Suction Tube
2. Connect the luer adapter on the end of the Injection Tube Assembly to the inlet port on the
PAC CHECK.
3. If the sample is liquid-filled with limited headspace, use the tilt fixture. This will move the
trapped gas to a position where it can be extracted. You can move the tilt fixture to either the
front or back mounting holes.
4. Loosen the vertical adjustment knob and position the piercing assembly at the appropriate
height for the sample to be tested.
6. Pull down on the piercing control lever, until the black sealing rubber is compressed to about
half the original size. This forms a leak free seal as the needle pierces the sample package.
7. Use the tension control knob to secure the piercing needle assembly after the package has
been pierced. This eliminates the possibility of the needle withdrawing after the seal is
obtained.
8. Pull back on the syringe piston to draw a liquid-free gas sample from the test package and
into the suction tubing.
9. Push in the syringe piston to force the gas sample from the syringe into the analyzer.
• Run/Results Screen.
• Setup Parameters.
Power-Up
When the OFF/ON button is pressed the PAC CHECK Model 325 display will illuminate and the “Idle”
screen will be displayed as shown in Figure 3-1.
Battery Monitoring
The battery level is represented by a 5-segment bar graph located on the right hand of the Idle and Run
screens. The battery level indication is replaced with a charge icon (lightening bolt) when the external power
is plugged in and the battery is charging.
If the battery voltage drops too low, an audible alarm is sounded and the analyzer enters a power down
mode. When the main battery pack is fully discharged or disconnected, a separate backup battery maintains
the real time clock, user names, product names and recorded measurements so no information is lost.
Power Down
There are two ways for the analyzer to go into a power down mode:
1. Press and hold the ON/OFF button for 4 seconds when the Idle, Results or Main menu
screens are displayed.
2. If no key has been pressed for a period of time defined in the shutoff menu, the unit will
automatically enter a power-down mode. The factory default time for automatic shutoff is 100
seconds and the selectable range is 40-300 seconds.
Display Operation
This section describes the various screens the PAC CHECK Model 325 will display.
Idle Screen
Any time the analyzer is powered up, the Idle Screen will appear as shown in Figure 3-1. The screen will
display the product, user, barometric pressure, temperature, date and time. As the analyzer will store up to
240 sample readings, the Idle Screen provides an indicator of how full the memory is. The mode of
operation is also displayed. When a test is initiated by pressing the test button the Run screen will be
displayed as shown in Figure 3-2 and will remain until the test has completed.
If a Cal-Smart calibration has not been performed for more than 24 hours a “Cal-Smart” message will be
displayed on the Idle and Run screens as shown in Figure 3-1. When the Cal-Smart message is displayed a
calibration should be performed. A Cal-Smart calibration can be performed by pressing and holding the Cal-
Smart key for 4 seconds from the Idle or Results screen.
MODE:AUTO VENT
11:00:00 10/24/05
Figure 3-1: Idle Screen
MODE:AUTO VENT
TESTING . . . .
Figure 3-2: Run Screen
Results Screen
When the test is complete, the Results Screen will appear as shown in Figure 3-3 and the display will show
the O2% and CO2% reading for the test performed.
1. The PAC CHECK 325 must come to a stable temperature for accurate results. If large
ambient temperature changes occur, allow 20 minutes for system temperature to stabilize.
2. Apply a sticky nickel or an adhesive rubber strip to the package to be tested. The sticky
nickel or adhesive rubber strip allows accurate sampling without air leaking from the package
or room air leaking into the package.
3. Insert the needle through the sticky nickel or adhesive rubber strip into the package
headspace.
4. Press the Test key. The internal sampling pump will turn on and draw sample gas from the
package before shutting off automatically.
As the sample is drawn, a “Testing …” line will be displayed. When the test is complete, the
O2% and CO2% readings will be displayed.
1. The PAC CHECK 325 must be at a stable temperature for accurate results.
2. Apply a septum (a sticky nickel or an adhesive rubber strip) to the package to be tested.
4. Insert the syringe needle through the septum into the package headspace.
5. Pull back on the plunger to extract a gas sample from the package.
6. Remove the syringe from the package and insert the needle into the Injection Port.
7. Press the Test key, push the plunger in, at a steady rate, to inject the sample.
8. When the test is complete, the O2% and CO2% readings will be displayed
Main Menu
When the “MENU” button is pressed from the Run or Results Screen, a Main Menu display will appear as
shown in Figure 3-4.
MAIN MENU From this menu, the following keys are recognized:
1 MEMORY
2 PRODUCTS (a) Numeric keys 1-5: Enter 1 to 5 to select an item from the list.
3 USERS
4 SETUP (b) On/Off: Enter power-down mode.
5 CALIBRATION
(c) Test: Return to the Run menu and begin test sequence.
HH:MM:SS M/DD/YY
(d) Menu: Return to the Idle screen, does not start a test. (Shows the
Idle screen.)
The current time and date are displayed in real time at the bottom of the main menu. All other instances of
time/date are fixed, such as the time/date captured at the start of a test run sequence.
Password Screen
Upon selecting Setup or Calibration, the screen shown in Figure 3-5 will appear and the user will be asked
to enter the setup/configuration password via the following menu. The password is entered in “alpha-
numeric entry” mode. The password is not shown as it is entered on the screen, but characters are
displayed as they are entered. The default password is "MOCON”. The setup menu can be used to change
the password. Once a valid password has been entered it will be in effect until the instrument is turned off.
PASSWORD The following keys are recognized while in the Password menu:
Upon selecting the Products menu, the screen shown in Figure 3-6 will appear. This menu allows the user
to select or rename the product, change the ”Test Method”, or the “Auto Store” feature.
PRODUCTS From this menu, the following keys are recognized:
1 SELECT
2 RENAME (a) Numeric keys 1-3: Enter 1 to 3 to select an item from the list.
3 TEST METHOD
(b) CANCEL/CLEAR or MENU: Returns to main menu.
PRODUCTS The following keys are recognized while in the Select menu:
>1 PRODUCT1
2 PRODUCT2 (a) Numeric 1-6: Select one of the products shown.
3 PRODUCT3
4 PRODUCT4 (b) Left/Right Arrows: Move the indicator up and down the list.
5 PRODUCT5
6 PRODUCT6 (c) MENU: Save the selected product, return to the previous menu.
The PAC CHECK 325 comes with one product name defined. Once a product name has been created, it
can only be renamed, not deleted. When a blank product name is selected, a screen will appear as shown
in Figure 3-8.
PRODUCT The following keys are recognized while in product name menu:
(b) Left/Right Arrows: Move the indicator left or right within the name.
(c) MENU: Save the name and return to the previous menu.
PRODUCT The following keys are recognized while in the Product Rename menu:
(b) Left/Right Arrows: Move the indicator left or right within the name.
(c) MENU: Save the name and return to the previous menu.
When the “Rename’ selection is made for a product that has been used to store test data, a User Advisory
screen will appear as shown in Figure 3-10.
DATA STORED (a) The 1 key confirms the request and displays the Product Rename
FOR THE OLD NAME screen.
PLEASE CONFIRM
OK? (1-YES, 2-NO) (b) The 2 key returns to the previous menu.
When a product with existing data is renamed the stored data will be associated with the new name. The
new name must be used in the “Memory” menu to select, view, send or delete the data.
TEST METHOD The following keys are recognized while in the Test Method menu:
>1 AUTO HOLD
2 AUTO VENT (a) Numeric 1-3: Select one of the methods shown.
3 MANUAL
(b) Left/Right Arrows: Toggle between test methods.
The default Test Method is Auto Hold. The Auto Hold method should be used when sampling with the
pump and any of the following conditions exists: The sample is a rigid or semi-rigid package. The sample
does not contain enough headspace to insure it remains at or above ambient pressure. The sample is
“vacuum packed”.
The Auto Vent mode should be used when the Auto Hold or Manual methods are not appropriate.
The Manual test method should be used when using the optional manual injection port or the optional rigid
package fixture.
When the Users menu selection is made, a screen will appear as shown in Figure 3-12. This menu allows a
user to be selected or renamed.
USERS The following keys are recognized while in the Users menu:
1 SELECT
2 RENAME (a) Numeric keys 1-2: Enter 1 or 2 to select an item from the list.
The PAC CHECK 325 comes with one user name defined. Once a user name has been created, it can only
be renamed, not deleted. When a blank name is selected, a screen will appear as shown in Figure 3-14.
(b) Left/Right Arrows: Move the indicator left or right within the name.
(c) MENU: Save the name and return to the previous menu.
USER The following keys are recognized while in the User Rename menu:
(b) Left/Right Arrows: Move the indicator left or right within the name.
(c) MENU: Save the name and return to the previous menu.
When the “Rename’ selection is made for a user that has been used to store test data, a User Advisory
screen will appear as shown in Figure 3-16.
DATA STORED (a) The 1 key confirms the request and displays the User Rename
FOR THE OLD NAME screen.
PLEASE CONFIRM
OK? (1-YES, 2-NO) (b) The 2 key returns to the previous menu.
When a user with existing data is renamed the stored data will be associated with the new name. The new
name must be used in the “Memory” menu to select, view, send or delete the data.
SETUP The following keys are recognized while in the Setup menu:
1 LANGUAGE
2 PASSWORD (a) Numeric keys 1-6: Enter 1 to 6 to select an item from the list.
3 SCREEN/AUDIO
4 TIME/DATE (b) CANCEL/CLEAR or MENU: Return to the Main menu.
5 SHUTDOWN TIME
6 TEST SETTINGS
LANGUAGE The following keys are recognized while in the Language menu:
>1 ENGLISH
2 SPANISH (a) Numeric 1-5: Selects the corresponding language.
3 FRENCH
4 ITALIAN (b) Right / Left Arrow: Moves the cursor up or down the list.
5 GERMAN
(c) MENU: Saves current selection and returns to the Main menu.
PLEASE CONFIRM (c) Menu: Save new password, return to the previous menu
OK? (1-YES 2-NO)
NOTE: A new password will not take effect until the instrument has
been powered off.
Figure 3-19: Password Screen
BACKLIGHT The following keys are recognized while in the Screen Backlight Menu:
1 BRIGHTNESS
2 SHUTOFF TIME (a) Numeric keys 1 and 2: Select the Brightness or Shutoff Time
menu.
The following keys are recognized while in the Shutoff Time menu:
SHUTOFF TIME
(0-300 SECONDS) (a) Numeric keys: Enter 0 to 300.
100 (b) CANCEL/CLEAR: Clear the current setting; if zero, return to the
Screen Backlight Menu without saving changes
(c) MENU: Save the setting and return to the Screen Backlight Menu.
(d) MENU: Save the setting and return to the Screen/Audio Menu.
Figure 3-24: Audio Menu Screen
(d) MENU: Saves the settings and returns to the Time/Date menu.
Figure 3-26: Set Time/Date Screen
The following keys are recognized while in the Date Format Menu:
FORMAT
1 DD-MM-YY (a) Numeric keys 1 and 2: Select the date format.
>2 MM/DD/YY
(b) CANCE/CLEAR: Returns to the Time/Date menu without saving.
(c) MENU: Saves the setting and returns to the Time/Date Menu.
SHUTDOWN TIME The following keys are recognized while in the Shutdown Time menu:
(40-300 SECONDS)
(a) Numeric keys: See numeric entry mode (all keys apply).
100
(b) MENU: Save the setting and return to the previous menu.
TEST SETTINGS The following keys are recognized while in the Test Settings menu:
1 ALARMS
2 PUMP TIME (a) Numeric keys 1-4: Enter 1 to 4 to select an item from the list.
3 AUTO STORE
4 AUTO SEND (b) CANCEL/CLEAR or MENU: Return to the Setup menu.
ALARMS The following keys are recognized while in the Alarms menu:
1 LOW ALARM O2
2 HIGH ALARM O2 (a) Numeric keys 1- 4: Select the desired alarm.
3 LOW ALARM CO2
4 HIGH ALARM CO2 (b) CANCEL/CLEAR or MENU: Return to the Test Settings menu.
Main Menu ► Setup Menu ► Test Settings Menu ► Low Alarm Menu
When a “Low Alarm” or “High Alarm” selection is made, a screen will appear similar to that shown in Figure
3-31. The first line indicates the alarm level (High or Low) and the type of gas (O2% or CO2%).
LOW ALARM O2 The following keys are recognized while in the Low and High Alarm
(0.00-100.00 %) menus:
Main Menu ► Setup Menu ► Test Settings Menu ► Pump Time Menu
When the “Pump Time” selection is made, a screen will appear as shown in Figure 3-32. This menu allows
the pump time to be changed. The Pump Time is defined as the amount of elapsed time (in seconds) the
sample pump will run when a test is performed.
PUMP TIME The following keys are recognized while in the Pump Time menu:
(1-90 SECONDS)
(a) Numeric keys: Enter 1 to 90 seconds.
4
(b) MENU: Save the setting and return to the previous menu.
AUTO STORE The following keys are recognized while in the Auto Store menu:
1 NO
>2 YES (a) Numeric 1-2: Select No or Yes.
The following keys are recognized while in the Auto Send menu:
AUTO SEND
1 NO (a) Numeric 1-2: Select No or Yes.
>2 YES
(b) Left/Right Arrows: Toggle between 1 & 2.
CALIBRATION
1 GASES
2 UNITS The following keys are recognized while in the Calibration menu:
(d) Numeric keys 1-2: Enter 1 or 2 to select an item from the list.
PACCHECK MODEL 325 (e) MENU or CANCEL/CLEAR: Return to the Main menu.
NOTE: Your PAC CHECK Model 325 comes calibrated from the factory. See Appendix A for detailed
information about calibrating your PAC CHECK Model 325.
UNITS
1 PRESSURE The following keys are recognized while in the Calibration menu:
2 TEMPERATURE
(a) Numeric keys 1-2: Enter 1 or 2 to select an item from the list.
The following keys are recognized while in the Units Pressure menu:
PRESSURE
(a) Left/Right arrows: toggle between MMHG and MBAR.
>1 MMHG
(b) Numeric 1 or 2: Select MMHG or MBAR.
2 MBAR
(c) CANCEL/CLEAR: Return to the Units menu without saving the
changes.
(d) MENU: Save the setting and return to the Units Menu.
Figure 3-37: Pressure Screen
Chapter 4: Specifications
The table below describes the specifications of the Pac Check Model 325:
O2 Resolution 0.01% O2
Less than a 1% change in a 24 hour period at a
O2 Temperature Stability
concentration of 100% O2
CO2 Accuracy 2% absolute @ 0-20%; 3% absolute @ 20-100%
Chapter 5: Maintenance
This chapter will assist you in keeping your PAC CHECK 325 clean and working properly.
To avoid damage to your instrument, please follow the precautions listed below:
1. Unplug the instrument from the power source before you begin cleaning it.
2. Do not remove the instrument cover; there are no user serviceable parts inside.
4. Never use acetone or other solvents to clean the instrument. These chemicals could damage
the instrument as well as the housing.
To replace the LIQUI-BLOK® filter on the PAC CHECK 325, follow these steps:
2. Remove the LIQUI-BLOK® filter from the sample probe. Discard the filter.
How often you will need to replace the filter chip depends on the frequency and type of testing you do.
MOCON recommends checking and changing the filter chip at once least every week with normal use, more
often with frequent use. Filter chips will not operate properly if they are dirty, punctured or discolored.
Replace the rubber inlet if it is torn or if its orifice has stretched so that it no longer seals against the needle.
4. Pull out the filter holder from the end of the filter cartridge.
5. Tap the filter cartridge gently to remove the old filter chip.
6. Remove and check the rubber inlet to see if it needs to be replaced. Replace the rubber inlet
if is torn or if it no longer seals around the needle.
7. Use compressed air to clean out any dust and debris from the filter cartridge.
9. Put a new filter chip in the filter cartridge and reinstall the filter holder.
10. Install the filter cartridge in the inlet guide, making sure the rubber inlet end is inserted first.
11. Screw the Inlet Guide onto the Inlet Guide Adapter.
12. Install the Injection Port assembly on the PAC CHECK 325.
A LIQUI-BLOK filter should always be used to minimize potential damage to the sensor from liquid or dust in
the test sample. You should replace the filter if it appears discolored or ruptured.
How often you will need to replace the serviceable components depends on the frequency and type of
testing you do.
2
5
10
2
1 (Inside knurled Nut)
3
9
6
4
7 8
To order replacement parts, call the MOCON Spare Parts Order Desk at (763) 493-7265.
6. Plug the connector, from the battery, into the mating connector that connects to the circuit
board.
7. Install the battery in the instrument case, position it so that the wires exit the battery towards
the left-center side of the case.
8. Push the battery connector and wires into the small open space at the top of the battery. The
connectors and wires must be flush with or below the surface of the battery.
9. Replace the battery cover and secure it with the tow screws.
Battery Connector
Battery Cover
Please read this entire procedure before proceeding. You may choose to have MOCON Technical Services
replace the sensor. If you have any questions, please call MOCON Technical Services in the USA at (763)
493-7250. We want you to receive the best in product support services.
To replace the Oxygen Sensor on the PAC CHECK 325, follow these steps:
1. Turn off the instrument by pressing and holding the ON/OFF button for 4 seconds.
4. Remove the four screws, shown in Figure 5-4, securing the instrument cover and the battery
cover.
Remove Screws
Battery Connector
Remove Screws
Battery Cover
8. Turn the instrument over and place it face up on the work surface.
9. Carefully separate the top and bottom parts of the instrument case.
10. Carefully rotate the top of the case until it is approximately at a right angle to the bottom part.
This will provide access to the keypad connector, shown in Figure 5-6 below.
11. Carefully unplug the keypad cable from the connector on the circuit board.
LCD Display
Keypad Connector
Remove Screws
12. Remove the two screws below the keypad connector, shown in Figure 5-6, securing the
circuit board to the case.
13. Carefully lift the LCD display off the locating pins and lay it face down on the circuit board as
shown in Figure 5-7 below. Be careful to avoid damaging the display cable, connector or
scratching the face of the LCD.
14. Remove the four remaining screws, shown in Figure 5-7, securing the circuit board to the
case.
16. Turn the instrument over and carefully place it face down on the work surface while holding
the display in place.
17. Carefully lift the case back off the circuit board and position next to the case as shown in
Figure 5-8 on the next page.
Remove Screw
Remove Screws
Locating Pin
LCD Display
Remove Screw
18. Remove the O-Ring from the new oxygen sensor; take care not to damage the threads.
19. Unplug the signal cable from the connector at the back of the oxygen sensor.
20. Remove the oxygen sensor from the housing by turning it counterclockwise.
21. Screw the new sensor into the housing, by turning it clockwise until it is finger tight.
22. Tighten the sensor an additional ¼ turn clockwise to seat the O-Ring in the housing.
23. Plug the signal cable into the connector at the back of the oxygen sensor.
Oxygen Sensor
Signal Cable
24. Place the circuit in the bottom part of the case and position the LCD as shown in Figure 5-7.
25. Replace the 4 screws, shown in Figure 5-7, that were removed in step 14.
26. Replace the LCD on the locating pins as shown in Figure 5-6. Be careful to avoid kinking or
creasing the LCD cable.
27. Replace the 2 screws, shown in Figure 5-5, that were removed in step 12.
28. Plug the keypad cable into the keypad connector as shown in Figure 5-6. Be careful to avoid
kinking or creasing the cable.
29. Replace the instrument front cover. Inspect the fit carefully to insure that everything is
correctly aligned.
30. Turn the instrument over, while carefully holding the cover in place and place it face down on
the work surface.
31. Replace the 2 screws, shown in Figure 5-5, that were removed in step 7.
33. Replace the 2 screws securing the back, (shown in Figure 5-5) that were removed in step 4.
34. Install the battery cover and replace the final 2 screws to secure it in place.
35. Turn the instrument over and turn it on by pressing and holding the ON/OFF button for 4
seconds. The instrument should power up and display the “Idle” screen.
36. A quick functional test of the new oxygen sensor can be performed by doing a Cal-Smart®
calibration.
37. To insure accuracy across the entire instrument range the oxygen sensor must be calibrated
using the “Calibrate O2” function in the calibration menu. Refer to Appendix A page A-10.
• Checking the PAC CHECK Model 325 oxygen and carbon dioxide calibration.
• Calibrating the PAC CHECK Model 325 the oxygen and carbon dioxide sensors.
This appendix is set up so you can easily access information on checking calibration and calibrating the
PAC CHECK Model 325.
Overview
In order to perform your own calibration checks and calibrations on the PAC CHECK Model 325, you should
have available reference gases of known O2 and CO2 concentrations in a balance of an inert gas, typically
N2 or a known combination of these gases. When choosing reference gases, it is best to pick those with
concentrations close to the range of O2 and CO2 you will be testing.
NOTE: Before checking the calibration, ensure there has been sufficient time for the instrument to come to
equilibrium with the ambient temperature. To assure optimum results, calibrate at or near the normal
working temperature.
2. Verify the instrument is in Auto Hold or Auto Vent mode; if necessary, change the Test Mode.
If the values are outside these tolerances the instrument should be calibrated. Refer to
“Calibrating the PAC CHECK Model 325” on page A-4.
3. If the instrument is in Auto Hold or Auto Vent mode, change the Test Mode to manual.
6. The reference gas must be allowed to flow through the system until it reaches equilibrium. A
minimum time period of 1 to 2 minutes is recommended.
The O2 value should be within the stated tolerance for the reference gas used.
If the value is outside this tolerance, the instrument should be calibrated. Refer to
“Calibrating the PAC CHECK Model 325” on page A-4.
To check the CO2 reference gas calibration, use the following procedure.
3. If the instrument is in Auto Hold or Auto Vent mode, change the Test Mode to manual.
4. Connect a Carbon Dioxide Reference gas to the instrument as illustrated in Figure A-1.
6. The reference gas must be allowed to flow through the system until it reaches equilibrium. A
minimum time period of 1 to 2 minutes is recommended.
9. The CO2 value should be within the stated tolerance for the reference gas used.
If the value is outside this tolerance, the instrument should be calibrated. Refer to
“Calibrating the PAC CHECK Model 325” on page A-4.
The figure below shows how to connect a reference gas to the instrument using the certified gas calibration
equipment available from MOCON. When using gas and equipment not supplied by MOCON, a similar
equipment configuration should be used. A regulator capable of controlling output pressure and flow is
required. The reference gas flow must be maintained at a stable value between 50 and 100 cc/min.
Calibration Supplies
Calibration requires a regulated supply of reference gases with known oxygen and carbon dioxide
concentrations. MOCON can provide these optional accessories to allow for quick and easy calibration:
Please call your sales representative or the MOCON Spares Desk in the USA at (763) 493-7265 for more information.
When the oxygen concentration of the sample headspace is normally above 20% (ambient air) the three
point calibration method should be used. The three point calibration method requires a “High O2” reference
gas for calibration. Commercial grade 100% oxygen is a suitable reference gas for a three point calibration.
The carbon dioxide sensor is calibrated using two reference points. Ambient air is used as one of the
reference points. For the second or “High Reference” point of the CO2 sensor, a gas with a certified
composition at or near 100% is often used.
Using the Cal-Smart feature to calibrate the oxygen sensor regularly insures the best possible sensor
accuracy. It is recommended that a Cal-Smart calibration be performed daily.
1. Select “Gases” from the Calibration menu and then select the Cal-Smart menu item.
A three point calibration should be performed under the following conditions: when the oxygen sensor has
been replaced, when a Calibration Check, using a high concentration oxygen reference gas, indicates
calibration is needed and a Cal-Smart calibration has not corrected the problem.
1. Select “Gases” from the Calibration menu and then select the “Calibrate O2” menu item.
4. Wait until the screen instructing the user to connect the reference gas appears.
Note: The remainder of the procedure can be skipped, when performing a calibration after the
sensor has been replaced, if the instrument will not be used for testing above 20% (ambient
air). To skip the “High O2” calibration sequence press the “Cancel” key.
5. Connect the O2 reference gas to the instrument. See Calibration Figure A-1: Reference Gas
Connection for information on connecting a reference gas.
7. Input the actual O2% reading that was sampled into the field.
The Carbon Dioxide sensor should only be calibrated when the result of a Calibration Check indicates the
sensor is outside of the allowable range.
3. Connect the High CO2 reference gas at a flow rate of 50 - 100 cc/min. See Calibration Figure
A-1: Reference Gas Connection for information on connecting a reference gas.
4. Allow the reference gas to flow through the system for a minimum period of 1 to 2 minutes.
6. Input the actual CO2% reading that was sampled into the field.
10. After the test has completed the “Calibration Gases” menu will be displayed.
CALIBRATION
1 GASES The following keys are recognized while in the Calibration menu:
2 UNITS
(a) Numeric keys 1-2: Enter 1 or 2 to select an item from the list.
CALIBRATION GASES The following keys are recognized while in the Gases menu:
1. CAL-SMART
2. CALIBRATE CO2 (a) Numeric keys 1-3: Enter 1 to 3 to select an item from the list.
3. CALIBRATE O2
(b) CANCEL/CLEAR or MENU: Return to the Main menu
SAMPLE AT AIR The following keys are recognized while in the Sample at Air screen:
PRESS [TEST] WHEN (b) TEST: Begins the “Cal-Smart” calibration sequence. The screen
READY TO CALIBRATE will appear as shown in Figure A-5 during the testing sequence.
To begin the calibration sequence, place the needle in room air and press the test key. The screen will
display a “Testing …” message and the analyzer will automatically zero the oxygen sensor.
AUTO ZERO O2 The following keys are recognized while the “Auto Zero O2” screen is
displayed:
COLLECTING DATA
(a) CANCEL/CLEAR: Returns to the calibration menu without
completing the test sequence.
TESTING………
When the Automatic Zero test sequence has completed the screen will appear as shown in Figure A-6 and
the analyzer will make three consecutive ambient air measurements.
SAMPLE AT AIR The following keys are recognized while the Sample at Air data entry
screen is displayed:
COLLECTING DATA
(a) CANCEL/CLEAR: Returns to the calibration menu without
completing the test sequence.
TESTING………
When the Cal-Smart calibration sequence has completed the Calibration Gases menu will be displayed.
HIGH CO2 The following keys are recognized while in the “High CO2” screen:
(20.00-100.00 %)
(a) CANCEL/CLEAR: Return to the previous menu without performing
FLOW REFERENCE GAS a High Reference calibration.
INTO INPUT
(b) TEST: Begins the “High CO2” calibration sequence. The screen
PRESS [TEST] WHEN will appear as show in Figure A-8 during the testing sequence.
READY TO CALIBRATE
Follow these steps to prepare for a High CO2 Reference gas calibration:
2. Connect the High CO2 reference gas as illustrated in Figure A-1 in this appendix.
3. Establish a stable flow of gas between 50 and 100 cc/min. When using an adjustable
MOCON calibration gas regulator set the secondary pressure to between 7 and 8 psi.
4. Allow the reference gas to flow through the system for a minimum period of 1 to 2 minutes
prior to beginning the Test.
To begin the reference calibration sequence, press the test key. The screen will display a “Testing …”
message and the analyzer will automatically take three consecutive samples.
HIGH CO2 The following keys are recognized while the “High CO2 Testing” screen
(20.00-100.00 %) is displayed:
TESTING………
After the sampling sequence is complete, a screen will appear as shown in Figure A-9. This screen is used
to enter the appropriate percentages for the High CO2 reference gas in the field.
The following keys are recognized while in the High CO2 screen:
HIGH CO2
(20.00-100.00 %) (a) Numeric keys: Enter desired concentration for the CO2 in the field.
After data entry is complete and the Menu key is pressed, a screen will appear as shown in Figure A-10.
This screen is used to confirm that the calibration information should be saved.
HIGH CO2 The following keys are recognized while the High CO2 confirmation
(20.00-100.00 %) screen is displayed:
32.80 (a) Numeric 1- 2: Confirm or cancel the calibration and return to the
previous menu.
PLEASE CONFIRM (b) CANCEL/CLEAR: Cancel the calibration and return to the previous
1-YES 2-NO menu.
SAMPLE CO2 AT AIR The following keys are recognized while in the Sample at Air screen:
PRESS [TEST] WHEN (b) TEST: Begins the “Sample CO2 At Air” sequence. The screen will
READY TO CALIBRATE appear as shown in Figure A-12 during the testing sequence.
To the complete the CO2 calibration sequence, place the needle in room air and press the test key. The
screen will display a “Testing …” message and the analyzer will automatically take three consecutive
ambient air samples. When the sequence has completed the Calibration Gases menu will be displayed.
SAMPLE CO2 AT AIR The following keys are recognized while the Sample CO2 at Air Testing
screen is displayed:
COLLECTING DATA
(a) CANCEL/CLEAR: Returns to the calibration menu without
completing the test sequence.
TESTING………
A three point calibration should be performed when a Calibration Check, using a high concentration oxygen
reference gas, indicates that calibration is needed and a Cal-Smart calibration has already been performed.
A three point calibration must be performed whenever the oxygen sensor has been replaced.
To begin the 3 point calibration sequence, place the needle in room air and press the Test key.
SAMPLE AT AIR The following keys are recognized while in the Sample at Air screen:
PRESS [TEST] WHEN (b) TEST: Begins the “Sample at Air” calibration sequence. The
READY TO CALIBRATE screen will appear as shown in Figure A-14 during the testing
sequence.
SAMPLE AT AIR The following keys are recognized while the Sample at Air Testing
screen is displayed:
COLLECTING DATA
(a) CANCEL/CLEAR: Returns to the calibration menu without
completing the test sequence.
TESTING………
When the Sample at Air test sequence has completed the screen will appear as shown in Figure A-15. The
analyzer will display a “Testing …” message and automatically zero the oxygen sensor.
AUTO ZERO O2 The following keys are recognized while the Testing Sample at Air
screen is displayed:
COLLECTING DATA
(a) CANCEL/CLEAR: Returns to the calibration menu without
completing the test sequence.
TESTING………
When the Auto Zero O2 sequence has completed, a screen will appear as shown in Figure A-16. The “High
O2” screen is used to perform a High Reference calibration of the oxygen sensor.
HIGH O2 The following keys are recognized while in the “High O2” screen:
(80.00-100.00 %)
(a) CANCEL/CLEAR: Return to the previous menu without performing
FLOW REFERENCE GAS a High Reference calibration.
INTO INPUT
(b) TEST: Begins the “High O2” calibration sequence. The screen will
PRESS [TEST] WHEN appear as shown in Figure A-17 during the testing sequence.
READY TO CALIBRATE
2. Connect the High O2 reference gas as illustrated in Figure A-1 in this appendix.
3. Establish a stable flow of gas between 50 and 100 cc/min. When using an adjustable
MOCON calibration gas regulator set the secondary pressure to between 7 and 8 psi.
4. Allow the reference gas to flow through the system for a minimum period of 1 to 2 minutes
prior to beginning the Test.
To begin the reference calibration sequence, press the test key. The screen will display a “Testing …”
message and the analyzer will automatically take three consecutive samples.
HIGH O2 The following keys are recognized while the Testing High O2 screen is
(80.00-100.00 %) displayed:
TESTING………
After the sampling sequence is complete, a screen will appear as shown in Figure A-18. This screen is used
to enter the appropriate percentages for the High O2 reference gas in the field.
After data entry is complete and the Menu key is pressed, a screen will appear as shown in Figure A-19.
This screen is used to confirm that the calibration information should be saved.
HIGH O2 The following keys are recognized while the High O2 confirmation
(80.00-100.00 %) screen is displayed:
87.80 (a) Numeric 1- 2: Confirm or cancel the calibration and return to the
previous menu.
PLEASE CONFIRM (b) CANCEL/CLEAR: Cancel the calibration and return to the previous
1-YES 2-NO menu.
When the calibration has been confirmed or canceled the Calibration Gases menu will be displayed.
MEMORY 4% FULL The following keys are recognized while in the Memory menu:
1 SELECT DATA
2 VIEW (a) Numeric keys 1-4: Enter 1 to 4 to select an item from the list.
3 SEND
4 DELETE (b) CANCEL/CLEAR or MENU: Return to the Main menu.
7 OF 7 SELECTED
The instrument can store the results of up to 240 tests. The percentage of memory available for data
storage that has been used is displayed on the top line of the screen. When the memory is full the display
will show “Memory 100% Full”. When the memory is full the oldest data will automatically be discarded as
new data is generated. No warning is generated, except for the 100% indication, when data is discarded.
The results from the last test will be stored automatically if the “Auto-Store” setting for the currently selected
product is set to “Yes”. If the Auto-Store setting is set to “No” the data will not be saved unless the “Store”
key is pressed. More information on setting the Auto-Store function is contained in Chapter 3 Operating the
PAC CHECK model 325.
SELECT DATA The following keys are recognized while the Select Data screen is
1 PRODUCT POTATOCHIP displayed:
2 USER TIMOTHY
3 FROM 10/25/03 (a) Numeric 1-4: Select the desired item for editing of the selection
4 THRU 10/27/03 criteria.
Main Menu ► Memory Menu ► Select Data ► From and Thru Time/Date
When the “From” or “Thru” items are selected in the Select Data menu, a screen as shown in Figure B-5 will
appear for the respective item. These screens are used to set the Date criteria that will be used when
selecting data for viewing, sending or deleting. The Time/Date selected on these screens will be displayed
on the appropriate line (from or thru) of the Select Data screen.
The following keys are recognized while these screens are displayed
FROM TIME/DATE
(a) Numeric 0-9: Change the value of the highlighted field
11:00:00 10/15/04
HH:MM:SS MM/DD/YY (b) Right / Left Arrow: Moves to the next field. The validity of the data
in the field is checked before moving to the next field.
Default values will appear in this screen if they have not been selected since the last time the instrument
was turned on. The default product and user will be the currently selected Product and User. The selected
product and user are retained when the instrument is turned off. The default “From” date will be the date of
the oldest record in memory. The default “Thru” date will be today’s date.
The “All” data’ may be made selected by setting the product and user filters to “All”, and letting the date
range default to oldest date through the current date. The default values may be restored, by pressing the
CANCEL/CLEAR key.
NOTE: The FROM/THRU range is primarily effective for memory send and memory clear. When this
feature is used for memory view, all the stored data will be viewable, but viewing will start
with the first record found at the “FROM” date. Viewing will not be stopped while scrolling
ahead in time if data is available. Viewing will not be prevented while scrolling back in time if
data is available at a date older than the starting point of “from” date.
Data will be sorted by time/date for viewing and sending, with the oldest record in the given range appearing
first. All users and/or products will be viewed or sent unless filtered by selection of the product and/or user
as shown on the Select Data screen. When all products or all users are selected, different products and
users may be mixed within the data set, the data will be sorted by time/date only.
If the data set contains more than seven records, it will be displayed seven records at a time. The pointer in
the first column indicates the currently selected record. The top line displays the product and user name of
the currently selected record. The currently selected record can be changed using number keys 1-7. The
detailed test data can be displayed by selecting the desired record and then pressing the Menu key.
The following keys are recognized while the Memory View (or data
review) screen is displayed:
(a) Numeric 1-7: Set that respective record as the current record.
(b) Left Arrow: Scroll to previous group of records in the data set. If the
PRODUCT 01 TEST USER first group of records is currently displayed, an invalid key beep will
>1 11:00:00 10/27/03 indicate the top of the list is displayed.
2 10:58:00 10/27/03
3 10:56:00 10/27/03 (c) Right Arrow: Scroll to next group of records in the data set. If the
4 10:54:00 10/27/03 last group of records is currently displayed, an invalid key beep will
5 10:52:00 10/27/03 indicate the end of the list is displayed.
6 10:49:00 10/27/03
7 10:48:00 10/27/03 (d) CANCEL/CLEAR: Returns to the Memory menu.
(e) MENU: Displays the currently selected data record’s detailed test
results using the Single Data Point screen see (Figure B-7).
Figure B-6: View Screen
740 MMHG 23.5 ◦C While viewing a single data point from previous test, the following keys
are recognized:
0.02%O2
(a) Left/Right arrows: Display the previous/next record in the data set.
30.00%CO2
(b) Clear/Cancel: Delete the displayed record from memory a
confirmation screen will be displayed.
SEND The following keys are recognized while the Send screen is displayed:
PRODUCT PRODUCT 01
USER TEST USER (a) CANCEL/CLEAR: Abort the data transmission and return to the
FROM 10/25/03 Memory menu.
THRU 10/27/03
(b) MENU: Abort the data transmission and return to the Memory
SENDING 14 OF 22 menu.
The count will increment as records are sent. Data transmission may be aborted by pressing the Menu or
Test key. Data transmission will continue until complete or aborted. When data transmission ha completed
the display automatically returns to the previous menu.
If the RS-232 port is not connected the record count will not increment. Clear/Cancel will return to the
Memory menu.
DELETE The following keys are recognized while the Send screen is displayed:
PRODUCT PRODUCT01
USER TEST USER (a) Numeric 1-2: Confirm or cancel the deletion and return to the
15:55:36 10/21/03 memory menu.
THRU
13:34:56 10/27/03 (b) CANCEL/CLEAR: Returns to the Memory menu without deleting
any data.
OK? (1-YES 2-NO)
DELETE
PRODUCT –ALL-
USER –ALL-
15:55:36 10/21/03
THRU
13:34:56 10/27/03
Data Log
A maximum of 400 data points may be stored in memory. Each data point stored in the data log will be of
the following format:
1. Time HH:MM:SS
2. Date MM-DD-YY (applies to storage format, not display format)
3. Test Mode
4. User Index
5. Product Index
6. O2 reading (scaled integer, 0.01% resolution)
7. CO2 reading (scaled integer, 0.01% resolution)
8. Temperature reading at time of test.
9. Barometric pressure reading at time of test.
NOTE: The format of Date data stored in the internal memory will always be HH:MM:SS MM-DD-YY
format; however, display of that data will be according to the format selected by the user in the
Setup Time/Date menu.
This data may be viewed using the View menu or exported through the RS-232 port using the Send feature.
New data will always be added to memory as it is obtained (and the Store key pressed or in auto store
mode). New data will be inserted in the first available unused location in memory. In the event there are no
empty records, a search will be made for the oldest record and it will be discarded. This may cause a
noticeable slow-down in operation; therefore, it is advisable to clear memory from time to time.
9600 Baud, 8 Data bits, No Parity, 1 Stop Bit, No Hardware flow control
Minimum Unit
Part Number Description Order of
Quantity Measure
mocon
®
EFFECTIVE DATE: January 4, 2011
WARRANTY POLICY
STATEMENT OF LIMITED WARRANTY
MOCON warrants that any part of any MOCON instrument or accessory which proves to be defective in material or workmanship
during the warranty period will be repaired by MOCON “certified” service personnel only, or at MOCON’s option replaced, free of
charge, F.O.B. MOCON’s factory. Please consult the MOCON technical services department to determine which warranty statement
applies to your instrument. This warranty applies to the original purchaser only, and is subject to the following terms and conditions:
1. Units or systems with MOCON offered field training, either purchased as an option or included in the purchase price.
• One year from date of shipment from MOCON’s factory if field training is performed by MOCON personnel or
“certified” representatives.
• Ninety days from date of shipment from MOCON’s factory if field training is not purchased.
• One year from date of shipment from MOCON’s factory if an extended warranty is purchased as an option or
included in the purchase price (not available on all instruments).
3. Spare parts, repairs and accessories when purchased separately and not a part of a new instrument order.
4. This warranty covers normal use only. It does not cover damage that results from alteration, accident, misuse, abuse,
neglect, or failure to follow assembly, installation, operational, or other MOCON instruction.
5. All warranty repair items are to be shipped at purchaser’s expense, to and from MOCON.
6. THE FOREGOING SHALL CONSTITUTE THE SOLE AND EXCLUSIVE REMEDY FOR ANY PURCHASE OF MOCON
PRODUCTS, AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED, IMPLIED OR STATUTORY, WHICH ARE
HEREBY DISCLAIMED AND EXCLUDED BY MOCON, INCLUDING WITH LIMITATION, THE IMPLIED WARRANTIES
OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND USE AND ALL OBLIGATIONS AND
LIABILITIES ON THE PART OF MOCON FOR DAMAGES ARISING OUT OF OR IN CONNECTION WITH THE USE OF
REPAIR OR PERFORMANCE OF THE PRODUCT. IN NO EVENT SHALL MOCON BE LIABLE FOR ANY INDIRECT
INCIDENTAL, SPECIAL OR CONSEQUENTIAL DAMAGES OR FOR ANY DELAY IN THE PERFORMANCE BY MOCON
OF ANY OF ITS OBLIGATIONS UNDER THE FIELD SERVICE AGREEMENT OR THIS WARRANTY DUE TO CAUSES
BEYOND ITS CONTROL.
7. In no event shall MOCON be liable for any damages in excess of the price paid for software including, but not limited to,
direct, consequential (including, without limitation, lost profits), special, exemplary, incidental and indirect damages, arising
out of or in connection with the use, the results or the inability to use the software, and imposed under any cause of action
whatsoever, including contract, warranty, strict liability, or negligence, even if MOCON has been notified of the possibility of
such damages.
MOCON software is provided “as is” and MOCON makes no warranty as to the software. MOCON DISCLAIMS ALL WARRANTIES,
EXPRESS OR IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, RELATED TO
THE SOFTWARE, ITS USE OR ANY INABILITY TO USE IT, AND THE RESULTS OF ITS USE.
P/N 032-846 C
mocon
®
EFFECTIVE: January 4, 2011
1. PART REPLACEMENT
If a defective part can be located, a free replacement part can be shipped collect to the customer for installation by in-
house repair personnel. This is the fastest and least expensive way to get the equipment operating again.
5. PART REPLACEMENT
If a defective part can be located, a replacement can be shipped to the Customer for installation by in-house repair
personnel. Charges will include the cost of the parts plus shipping. This is the fastest and least expensive way to get the
equipment operating again.
P/N 032-847 B
mocon
7500 Mendelssohn Avenue North
EFFECTIVE DATE: January 4, 2011
MOCON offers a complete range of service options to purchasers of MOCON instrumentation and systems.
4. PART REPLACEMENT
If the customer can identify a defective part, a replacement can be shipped F.O.B. MOCON factory to the customer for
installation by in-house repair personnel. Charges will include the cost of the parts plus shipping. This is the fastest and the
least expensive way to get the equipment operating again.
5. RETURN UNIT TO MOCON
If replacing the part appears to be beyond the capability of the Customer’s in-house personnel, the instrument may be
shipped back to the MOCON factory in Minneapolis, Minnesota for repairs. The Customer is responsible for the cost of
round trip freight and insurance, and will be billed for parts and labor required accomplishing the repair.
6. SERVICE REPRESENTATIVE
If malfunction is considered by MOCON to be an emergency, a MOCON service representative will be dispatched from
MOCON to visit the Customer’s facility to correct the problem. The charges for this service include out-of-pocket travel
expenses including portal-to-portal travel time at prevailing labor rates, parts used, and labor at the prevailing labor rates to
repair the equipment.
Please refer to the latest version of MOCON’s Warranty Policy sheet for warranty information.
P/N 032-848 C
Glossary
Accuracy
The degree of agreement of a measurement with an accepted reference level or value.
Automatic Mode
The method of taking a sample using the automatic internal sampling pump in the PAC CHECK MODEL
325.
Auto Hold
The Auto Hold test method is one of two Automatic Mode methods that use the automatic internal
sampling pump in the PAC CHECK Model 302. The Auto Hold mode should be used when sampling
with the pump and any of the following conditions exists: The sample is a rigid or semi-rigid package.
The sample does not contain enough headspace to insure it remains at or above ambient pressure. The
sample is “vacuum packed”.
When making measurements under these conditions a vacuum can be created within the instrument.
The Auto Hold mode maintains or “Holds” the sample within the instrument, after the pump stops, until
the measurement can be completed and compensated for the actual sample pressure. After the sample
has been measured and compensated for the pressure difference it is vented to atmosphere.
Auto Vent
The Auto Vent test method is one of two Automatic Mode methods that use the automatic internal
sampling pump in the PAC CHECK Model 302. The Auto Vent mode should be used when the Auto
Hold or Manual modes are not appropriate.
The Auto Vent mode “Vents” the sample to atmosphere after the pump stops and then the sample is
measured and compensated to the ambient pressure.
Calibration
A procedure which systematically adjusts a quantitative measuring instrument to ensure that the
measurements it produces are accurate.
CO
The chemical abbreviation for the carbon monoxide molecule.
CO2
The chemical abbreviation for the carbon dioxide molecule.
Communications Software
Any software package that allows one computer to communicate with another. The microprocessor in
the PAC CHECK MODEL 325 communicates with a personal computer through a communications
software application.
Concentration
The amount of a particular substance in a unit amount of another substance. Typically, room air has an
oxygen concentration of 20.6%.
Headspace
The atmosphere surrounding a product in a sealed package.
Inlet Guide
The port into which the sample is injected. Also contains the injection filter assembly, which sends the
sample through to the oxygen sensors inside the instrument.
LIQUI-BLOK Filter®
A filter is used to minimize potential damage to the sensor from liquid in a test sample.
Manual Mode
A sampling method, were a headspace sample is taken from a package using a syringe and manually
injected into the PAC CHECK 325. This sampling method requires the optional Injection Port. This
method should be used when the headspace volume in the package does not meet the minimum
sample size specification.
O2
The chemical abbreviation for the oxygen molecule. Two oxygen atoms make an oxygen molecule.
Resolution
The minimum discernible difference in the change of output signal of an instrument.
Response Time
The time it takes for the oxygen sensor to give an accurate reading after a sample has been injected.
Sample Gas
The gas extracted from a package for which an oxygen concentration reading is desired.
Sensitivity
The degree to which an instrument responds to a defined incoming quantity of the phenomenon being
detected.
Sticky Nickel
A round, adhesive-backed septum which is placed on the package before the needle is inserted to
remove the gas sample from the package. It creates a leak-free seal between the needle and the
package being tested.