Documente Academic
Documente Profesional
Documente Cultură
CONTENTS
A. Preface 5
Intended Use 7
Intended Audience 7
Chapters and Contents 7
Conventions 9
Ordering Information 9
Technical Assistance 9
1. Description 10
Introduction 11
Installing the Biolyte 2000 12
Features 18
Analytes 18
Main Components 19
Tubing 20
Flow Controller 20
Pump 21
Waste Valve and Reference Valve 22
Waste Valve 22
Reference Valve 22
Sampler 23
Electrode Carrier 24
Measuring Electrodes 25
Reference Electrode 25
Air Detector 25
Input/Output Devices 26
Keypad 26
Liquid Crystal Display (LCD) 28
Alarm 28
Printer 28
Acceptable Samples 29
Sample Handling Requirement 29
Serum 30
Plasma 30
Whole Blood 31
Urine 31
Cerebrospinal Fluid 31
Acceptable Anticoagulants 32
Interfering Substances 32
Drugs Showing No Lithium Interference 33
Reference Intervals 34
Display Range for Analytes 34
2. Set Up 35
Introduction 36
Set Up Options in Menu 37
Selecting Electrodes 37
Selecting Units of Measurement 38
Setting Resolution Values 39
Setting Two-point Calibration Interval 40
Setting Date and Time 41
Setting Results/Printer 43
Setting Reference Values 44
Setting Panic Values 45
Setting Offsets Values 46
Setting Operation Options 47
3.Operation 49
Introduction 50
Software Description 50
Main Screen 50
Start/Standby Function Key 51
Samp. Type Function Key 52
Calibrate Function Key 53
Analyze Function Key 55
Status Function Key 59
Flowpath Function Key 60
Menu Function Key 61
Set Up Options 61
Operation Options 61
Data Options 61
Data Recall 62
Data Transfer 63
Data Clear 63
Software Version 65
Diagnostic 65
Maintenance 65
QC function Key 65
Biolyte 2000 Operation 66
Shortcut Procedures 67
Power On 67
Selecting Electrodes 67
Set Sample Type 67
Selecting Analysis Modes 67
Calibration 68
Manual Calibration 68
Auto-Calibration 68
Analysis : Serum/Plasma 68
Analysis : Whole Blood 69
Analysis : Urine 69
Analysis : CSF 69
4. Maintenance 70
Introduction 71
Maintenance Function 71
Condition 72
Condition Na 72
Condition f_path 75
Clean Flow Path 78
Reagent Information 81
Prime 82
Probe Adjustment 83
Calibrate Pump Frequency 84
Daily Maintenance 85
Checking the Reagent Pack Level 85
Cleaning the Sample Inlet Port 85
Conditioning Na Electrode 85
Conditioning the Flow Path with Serum 85
Checking Slope & Flow Time 86
Checking the Electrode Slope 86
Checking the Flow Time 86
Periodic Maintenance 86
Flow Path Maintenance Cycle 86
Replacing the W-line 87
Replacing the R-line 88
Replacing the S-line 89
Replacing the A-line/B-line 90
Replacing the Probe 91
Replacing the Electrodes 92
Non-routine Maintenance 93
Cleaning the Sample Inlet Port 93
Cleaning the Probe 93
Standby Mode 93
Replacing the Reagent Pack 94
Installation of Thermal Paper 97
Shutdown 98
6. Troubleshooting 118
Introduction 120
Troubleshooting Theory 120
Troubleshooting Procedure 121
Solving Flow Problems 121
Indicator of Flow Related Problem 122
Flow Times 122
Air and Fluid Readings 123
Confirming a Flow Problem 124
Possible Sources of Flow Problem 125
Flow Path Check 126
W-line Check 126
S-line/Electrode Carrier Check 127
Probe Check 128
A-line & B-line Check 129
R-line Check 129
Septum Check 130
Diagnostic 132
Device Exercise 132
Valve Exercise 133
Pump Exercise 134
Probe Exercise 135
Device Test 136
Speaker Test 136
Printer Test 137
LCD Test 138
Keypad Test 138
Sampler Sensor Test 139
Serial Port Test 140
Air Detector Test 141
Error Log 142
View Error Log 142
Print Error Log 143
Clear Error Log 144
System Test 144
Probe Position 145
Pump Movement 146
Valves Set 146
Sensor mV Data 147
Error Code 148
Introduction to Error Codes 148
Display Error Code 148
Problem Solving Procedures 148
Interpreting Error Codes 150
Error Code Reference 151
7. Theory 198
Introduction 199
Origin of the Potential Measurement 200
Calculation of Na, K, Cl, and Li Concentration 204
Electrode Calibration 206
Two-Point Calibration 206
Two-Point Calibration Sequence 206
One-point Calibration 207
Direct vs. Indirect Methodologies 208
Plasma Volume Displacement Error 208
Bicarbonate 208
Protein Binding 208
Combined Effects 209
Halogen Effects on Chloride Determination 209
8. Appendix 211
Appendix I : Warranty of Products 212
Appendix II : Specifications of Biolyte 2000 215
Appendix III : Performance of Biolyte 2000 219
Appendix IV : Warning Messages 220
Appendix V : Labels on the Biolyte 2000 221
Appendix VI : Description of IEC Letter Symbols 221
Appendix VII : Location of Serial Ports 222
Appendix VIII : Error Codes 222
Appendix IX : Spare Parts List of Biolyte 2000 227
Appendix X : Consumable List of Biolyte 2000 228
A. Preface
A. Preface 5
Intended Use 7
Intended Audience 7
Chapters and Contents 7
Conventions 9
Ordering Information 9
Technical Assistance 9
5
6
V2 Reference Manual
Intended Use
The Biolyte 2000 is intended for in vitro diagnostic use by healthcare professionals in the
quantitative determination of sodium, potassium, chloride, or lithium (option) in whole
blood, serum, and plasma; sodium, potassium, and chloride in whole blood, serum, plasma,
urine and CSF.
All the data reported from the Biolyte 2000 are for clinical reference. If abnormal
clinical data occur, further studies and evaluation of these data are suggested to
ensure clinical safety.
Intended Audience
The Biolyte 2000 V2 Reference Manual is written for healthcare professionals who operate
the Biolyte 2000. Users do not need prior experience with BioCare products, but should
have been trained on the instrument by BioCare’s designated personnel.
This manual describes the Biolyte 2000 and explains how to test samples with them.
Chapter 1, Description, gives instructions on installation, lists the Biolyte 2000’s operating
features, describes its operating components, and specimen handling requirements.
Chapter 2, Set Up, describes the instrument’s Set Up options and explains how to modify
them.
Chapter 3, Operation, explains how to enter sample data, analyze samples, and calibrate
the instrument.
Chapter 5, Quality Control, explains how to monitor the performance of the instrument
7
with quality control standards.
Chapter 6, Troubleshooting, describes error codes and explains how to repair the Biolyte
2000.
Chapter 7, Theory, describes the principles of electrode theory and the process of electrode
calibration. This chapter also includes the equations that the instrument uses to calculate
the Biolyte 2000’s concentration for each electrode.
Chapter 8, Appendix, lists all of Biolyte 2000’s system specifications, warning messages,
labels on the Biolyte 2000, description of IEC letter symbols, consumable information, error
codes, reference list and parts list.
8
V2 Reference Manual
Conventions
This manual uses the following conventions:
• Bulleted lists contain itemized or summarized information.
• Numbered lists contain procedure steps
• Notes are written in Italics and provide important or helpful operating information.
• WARNINGS are written in italicized capital letters and provide information that is
important for user or instrument protection.
• Initial Capital Letters begin the name of software screens & menu options.
• Square marked letters are used for the names of keys on the Biolyte 2000’s keypad.
Bold typed captions in the left-hand margin of some pages restate important information
that appears in the text or contain additional information that complements the text.
Throughout the software operating procedures, quick reference steps in the left-hand
margin graphically explain how to access the software screens that are explained in the text.
Ordering Information
Users can order more copies of the Biolyte 2000 V2 Reference Manual (Part. No.: 9010157),
from BioCare Corporation Marketing Department at +886 -3-2616678 or e-mail to
sales@biocare.com.tw.
When encountering any question arising from different language versions, the English
version will be the principal reference.
Technical Assistance
For technical assistance with users’ Biolyte 2000, call BioCare Corporation’s Customer
Service Center at +886-3-2616678.
9
1. Description
Introduction 11
Installing the Biolyte 2000 12
Features 18
Analytes 18
Main Components 19
Tubing 20
Flow Controller 20
Pump 21
Waste Valve and Reference Valve 22
Waste Valve 22
Reference Valve 22
Sampler 23
Electrode Carrier 24
Measuring Electrodes 25
Reference Electrode 25
Air Detector 25
Input/Output Devices 26
Keypad 26
Liquid Crystal Display (LCD) 28
Alarm 28
Printer 28
Acceptable Samples 29
Sample Handling Requirement 29
Serum 30
Plasma 30
Whole Blood 31
Urine 31
Cerebrospinal Fluid 31
Acceptable Anticoagulants 32
Interfering Substances 32
Drugs Showing No Lithium Interference 33
Reference Intervals 34
Display Range for Analytes 34
10
V2 Reference Manual
Introduction
This chapter lists the Biolyte 2000’s testing parameters, specimen requirements, describes
its main operating components, and provides an overview of major software functions. The
Biolyte 2000 appears in Figure 1-1.
11
Installing the Biolyte 2000
Those unfamiliar with the Biolyte 2000 should read chapter 1 before proceeding. Users must
complete the following steps to install the Biolyte 2000. Some procedures may not apply to
all instruments.
The Biolyte 2000 weighs approximately 24 lbs. (11 kg) (without reagent
pack). BioCare recommends that at least 1 person help to remove the
instrument from its shipping carton and place it on the counter. Complete
the following steps to install and assemble the instrument :
12
V2 Reference Manual
• Confirming Voltage Range
6. The power input module permits the Biolyte 2000 to run at 0.6A/100-240VAC.
7. Confirm that the voltage of users’ laboratory is in the range of Biolyte 2000’s
specification. Turn the instrument around in order to view the backside. The power-
input module is located where the power cord connects to the Biolyte 2000.
V A Fuse Requirement
13
• Assembling and Installing Electrodes
9. Remove the electrodes and electrode carrier from the packing material.
10. Place the electrode carrier on the table with the locking lever pointing towards the
installer. Press the locking lever down to its “Open” position (see Figure 1-3).
Note: Make sure there is a sealing gasket between each of the electrodes in the carrier.
14
V2 Reference Manual
Note: Make sure there is a sealing gasket between each of the electrodes in the carrier.
12. Engage the locking lever at the bottom of the carrier to secure all the electrodes in
the carrier. The lever should be in its “Closed” position.
13. Align the connector pins on the electrode and the electrode carrier with the
receptacles on the instrument.
14. Gently push the electrode carrier into place on the Biolyte 2000, pressing from the
bottom and the top simultaneously.
Note: Align the guiding ground rod on the electrode train to the hole on the analyzer. Make sure that you
push in the guiding ground rod in first.
15
• Installing the Septum
19. Remove a septum from the reagent pack.
20. Lift the sampler to the manual position and install the septum into the bayonet lock
located on the under surface of the sampler. Turn the septum clockwise 1/4 turn to
lock it in place.
16
V2 Reference Manual
• Installing the Reagent pack
25. Unscrew the bottle caps of all reagents and the waste bottle.
26. Place a reagent pack half-way onto the fluid deck.
27.Insert the tubing harness straws into the appropriate bottles.
28.Push the reagent pack all the way in.
29.Check all tubing connections. Incorrect installation may result in backward flow and
contamination of reagent packs.
30. Plug the RMS into the RMS port (See Figure 1-8 for the location of the RMS port).
The users may now proceed to Chapter 2, Set-up Options, followed by the analysis of
specimens as described in Chapter 3, Operation.
17
Features
The Biolyte 2000 is a state-of-the-art electrolyte analyzer that test serum, plasma, CSF,
whole blood, and urine. The Biolyte 2000 uses a microcomputer-based operating system to
measure analytes, calculate results, and rapidly perform state analyses. The Biolyte 2000’s
quantitated analytes are listed according to sample type and analysis mode in Table 1-2. The
Biolyte 2000 is highly flexible laboratory instrument and includes many operating features
that can be modified to accommodate users’ changing work environment. Users can adapt
the Biolyte 2000 to fit users’ laboratory testing needs and define its operating parameters
for the technicians who will use it.
Analytes
Table 1-2. Biolyte 2000’s Available Analytes by Analysis Mode and Sample Type.
Manual Sodium Ⓞ Ⓞ Ⓞ
Potassium Ⓞ Ⓞ Ⓞ
Chloride Ⓞ Ⓞ Ⓞ Ⓞ
Lithium* Ⓞ Ⓞ
SEMI-AUTO Sodium Ⓞ Ⓞ Ⓞ
Potassium Ⓞ Ⓞ Ⓞ
Chloride Ⓞ Ⓞ Ⓞ Ⓞ
Lithium* Ⓞ Ⓞ
18
V2 Reference Manual
Main Components
The Biolyte 2000’s Main Components (Figure 1-8) include tubing and flow controller to
aspirate fluids, an electrode carrier to analyze samples and several communication devices
to receive and display information. The Biolyte 2000 also includes a replaceable reagent
pack that contains all of the reagents to calibrate the instrument and analyze samples.
19
Tubing
A series of plastic tubing transports fluids throughout the Biolyte 2000. These tubing
include all tubing that extends from the reagent pack to the analytical components on the
front of the Biolyte 2000.
Note: The flow path is the route that fluids travel in, throughout the Biolyte 2000.
Single tubing called the W-line carries fluids out of the analytical components and back into
the reagent pack. Any fluids that exit the Biolyte 2000 are considered waste and are drained
into a waste bottle inside the reagent pack.
Note: Suction in the W-line draws all fluids through the flow path.
Flow Controller
The flow controller works to direct fluids through the flow path. The flow controller includes
a pump, a waste valve, a reference valve, and a sampler, all of which are controlled by the
Biolyte 2000’s microcomputer. The microcomputer turns the pump on and off, opens and
closes the waste valve and reference valve, and directs the sampler so that the proper fluids
are aspirated in the appropriate timing sequence. The Biolyte 2000 needs accurately timed
fluid flow to operate properly, and the microcomputer coordinates the flow controller so the
appropriate fluids are delivered to the correct components when they are needed.
20
V2 Reference Manual
• Pump
The pump (Figure 1-9) is mounted to the front of the Biolyte 2000 and consists of a roller
cage that is connected to a stepping motor.
21
• Waste Valve and Reference Valve
The waste valve and reference valve help control the flow of fluids during the instrument’s
operating cycles. All instruments include a waste valve and a reference valve.
• Waste Valve
The waste valve is located on the front of the instrument and on the top of the reference
valve.
• Reference Valve
The reference valve is located on the front of the instrument and below the waste valve.
22
V2 Reference Manual
• Sampler
The sampler, which is located on the front of the Biolyte 2000, controls which fluids are
aspirated into the electrode carrier for analysis. The 2 main components of the sampler are
the septum (Figure 1-5) and the probe (Figure 1-6).
The septum is a plastic cylinder that is partitioned into four compartments. Each
compartment holds air or a reagent from the reagent pack. Rubber washers called “septa”
separate the compartments and prevent fluids from leaking and mixing inside of the
septum.
The probe is a blunt-ended metal tubing that aspirates fluids into the Biolyte 2000. The
probe moves to different positions inside of the septum to aspirate reagents and extends
out of the septum to aspirate sample or air.
The sampler accepts fluids in 2 different positions. In the Manual Mode (Figure 3-27B), the
sampler aspirates a specimen from a hand-held container. In the Semi-Auto Mode (Figure
3-27A), the sampler automatically aspirates specimen from the sample cup. The sampler is
designed to aspirate reagents from inside the septum during specimen analyses, calibration
and maintenance procedures.
23
• Electrode Carrier
The electrode carrier (Figure 1-11) houses the analytical components that measure analyte
concentrations in a sample. It includes the measuring electrodes, a reference electrode, a
spacer, and an air detector. Each measuring electrode quantitates a specific analyte.
24
V2 Reference Manual
• Measuring Electrodes
During an analysis or calibration, fluid fills the electrode carrier and each electrode generates
an electrical potential (or current) in proportion to the amount of its specific analyte in the
fluid. Then the microprocessor calculates the concentration of each analyte in the fluid,
based on the output of each electrode.
• Reference Electrode
The reference electrode supplies a continuous electrical potential in contrast to a measuring
electrode. The reference electrode's potential varies with each particular sample. Because
electrical potential can only be measured in comparison to a known voltage, the Biolyte
2000 uses constant voltage supplied by the reference electrode, to measure the potential
produced by each measuring electrode.
• Air Detector
The air detector is located below the potassium electrode and has 2 functions: it verifies the
presence of air or fluid in the electrode train. The microcomputer monitors the air detector
so that it knows when to stop the pump from aspirating a fluid once it fills the electrode
train. When the air detector takes a reading during an analysis, it knows whether it is
detecting air or fluid. The air detector is an optical sensor that can detect air and fluids.
25
Input/Output Devices
Use the following input and output devices to communicate with the Biolyte 2000 :
• Keypad
• Liquid Crystal Display (LCD)
• Alarms
• Printer
• Keypad
The keypad (Figure 1-12) is located on the left, front side of the Biolyte 2000 and contains
both numeric and function keys.
26
V2 Reference Manual
The function keys are described in Table 1-3.
Key Function
Menu Displays the Main Screen from the sensor Status Screen.
QC Accesses QC functions.
27
• Liquid Crystal Display (LCD)
The Biolyte 2000 uses 40 characters * 4 lines LCD screen. The Biolyte 2000’s main screen is
shown in Figure 1-13.
The main screen contains the following information: the currents date and time, sample
type, and status.
• Alarm
The Biolyte 2000 produces an alarm in response to input errors or problems that occur
during automatic operation. The Biolyte 2000 beeps to inform users of the following
conditions:
• An attempt to enter too many characters.
• Entry of an illegal value.
• A system, calibration, or analysis error has occurred.
The Biolyte 2000 sounds an alarm to inform users of the following conditions :
• The flow path maintenance has timed out.
• A parameter has exceeded a panic limit.
• An operator warning is present.
Most alarms repeat until users shut them off by pressing any key on the keypad or by
accessing another procedure.
• Printer
The thermal printer provides hard copies of test results, calibration information, maintenance
information, and error messages.
28
V2 Reference Manual
Acceptable Samples
The following chart lists acceptable sample types for the different parameters measured:
The NCCLS Approved Guideline on Procedures for the Handling and Processing of Blood
Specimens indicates that serum or plasma should be physically separated from contact with
cells as soon as possible to a maximum time limit of 2 hours from the time of collection. It
is also recommended that a contact time of less than 2 hours be established for serum or
plasma sample intended for potassium analysis.
29
Serum
Collect blood for serum samples in plain vacuum tubing or serum separator tubing.
Allow the sample to clot for approximately one half-hour. Following clot retraction,
centrifuge at 1000 RCF for 10 to 15 minutes, uncap the tubing, and use a syringe or
bulb pipette to obtain a serum sample. Take the sample from the area close to the
cells if possible, if the sample cannot be analyzed immediately, then separated serum
should remain at 22oC (room temperature) for no longer than 8 hours. If assays will
not be completed within 8 hours, the serum sample should be stored refrigerated
at 2 to 8oC. If assays will not be completed within 48 hours or if the serum sample
is to be stored beyond 48 hours, the samples are to be stored frozen at -20oC.
Plasma
Collect whole blood for plasma samples with minimal stasis, without exercise of the arm,
in vacuum tubing containing either sodium or lithium heparin. Whole blood specimens
are not to be chilled unless there are documented recommendations for doing so. Obtain
plasma by centrifuging heparinized whole blood within 1-hour of collection. Following
centrifugation at 1000 RCF for 10 to 15 minutes, remove the cap and use a syringe or bulb
pipette to obtain a plasma sample. Take the sample from the area close to the cells. Plasma
samples more than 1-hour old should be centrifuged immediately prior to analysis to remove
any fibrin clots. If assays will not be completed within 8 hours, the plasma sample should
be stored refrigerated at 2 to 8oC. If assays will not be completed within 48 hours or if the
plasma sample is to be stored beyond 48 hours, the samples are to be stored frozen at -20oC.
30
V2 Reference Manual
Whole Blood
Blood samples should be collected with minimal stasis, without exercise of the arm. Collect
blood for analysis on the Biolyte 2000 in vacuum tubing containing either sodium or
lithium heparin. Whole blood specimens are not to be chilled unless there are documented
recommendations for doing so. It is particularly important to ensure that samples are well
mixed before introduction into the Biolyte 2000.
Urine
Careful attention to urine sample handling is critical to ensure that accurate results are
obtained. Urine specimens should be collected in a clean, sterile container with a non-spill,
anti-evaporation lid. BioCare Corporation does not make any recommendations as to the
duration of collection.
To ensure the accuracy of the urine measurement, it is essential that urine be analyzed
within 2 hours of collection or refrigerated at temperatures between 2 and 8oC for extended
periods to prevent the growth and metabolism of microorganisms. Urine allowed to sit
unperceived at room temperature for extended periods of time will begin to decompose.
Acidification to a pH less than 3 may be used to preserve urine samples for extended storage
times. For urine electrolytes, boric acid may be used. The amount of acid preservative to be
used is dependent upon the collection period and urine output.
Cerebrospinal Fluid
Lumbar puncture to collect cerebrospinal fluid (CSF) for analysis must be performed by a
physician. Spinal fluid should be collected into plain sterile tubing labeled with the time of
collection and with numbers indicating the sequence in which the tubing were obtained if
multiple tubing are used. A serum or plasma sample or both should be obtained at the time
of spinal fluid collection to aid in diagnosis. CSF samples should be analyzed as soon as
possible after the time of collection and should not be stored.
References :
NCCLS. Procedures for the Handling and Processing of Blood. Document H18-A Vol. 10 No. 12.
Jacobs., Kasten, DeMott, and Wolfson, ed. 1990. Laboratory Test Handbook. Lexi-Comp Inc.
Tietz, N.W., ed. 1986. Textbook of Clinical Chemistry. W.B. Saunders Co.
31
Acceptable Anticoagulants
Sodium and lithium heparin is the recommended anticoagulants for plasma specimens.
EDTA, citrate, oxalate, or sodium fluoride is not recommended for use as anticoagulants.
Depending on the amount of heparin used in the collection container and whether it is filled
to capacity with blood, heparin concentrations may vary significantly. Sodium heparin may
elevate sodium results, and lithium heparin will elevate lithium results. Liquid heparin when
present in excess may cause errors by dilution.
Our experience suggests that lyophilized sodium or lithium heparin giving a final
concentration in blood of not more than 30 I.U./mL is acceptable in the critical care
laboratory. Biolyte 2000 users should take careful note of these considerations when
establishing reference intervals and interpreting results.
Interfering Substances
The following substances at the concentrations indicated have been shown
to produce an approximate 2% increase in the chloride concentration:
• β-hydroxybutyrate (25 mmol/L)
• acetoacetate (25 mmol/L)
• formate (10 mmol/L)
• lactate (25 mmol/L)
• salicylate (50 mg/dl)
• citrate (10 mmol/L)
32
V2 Reference Manual
Drugs Showing No Lithium Interference
Adding them to sera containing 1.5 mmol/L lithium tested the following medications. The
drug concentrations selected were based on NCCLS publication EP7--P 1986 Vol. 6, No. 13
(Proposed Guideline for Interference Testing in Clinical Chemistry). Expected lithium levels
were recovered in each case.
Phenobarbital 20 Mallaril 10
Primidon 20 Prolixin 1500
Carbamazepine 10 Allopurinol 60
Ethosuximide 70 Compazine 3
Phenytoin 20 Lorazepam 1
The possibility exists that of the numerous prescriptions and nonprescription medications
commonly administered some combination of drugs or their metabolites or both may be
encountered which will affect the lithium sensor.
33
Reference Intervals
Each laboratory should establish and maintain its own
reference intervals based on its patient population.
1
Tietz. Norbert W., Ed. 1986 Textbook of Clinical Chemistry. W.B. Saunders Co.
2
Statland. Bernaid. 1987. Clinical Decision Levels for Lab Tests. Medical Economics Books.
Note : 1. The Serum/Plasma display ranges also apply to serum or plasma specimens that are
analyzed as microsamples.
2. References to chloride or lithium or both are not applicable to instruments without these
electrodes.
34
V2 Reference Manual
2. Set Up
Introduction 36
Set Up Options in Menu 37
Selecting Electrodes 37
Selecting Units of Measurement 38
Setting Resolution Values 39
Setting Two-point Calibration Interval 40
Setting Date and Time 41
Setting Results/Printer 43
Setting Reference Values 44
Setting Panic Values 45
Setting Offsets Values 46
Setting Operation Options 47
35
Introduction
The Biolyte 2000’s Set Up options let users define the instrument’s operating parameters
to accommodate users’ laboratory’s testing needs. To access the Set Up Screen (Figure 2-1),
from the Menu Screen (Figure 3-16), input 1 and press Enter key.
Note: References to chloride or lithium or both are only applicable to instruments with these electrodes.
To select a Set Up option, input the number of the option and press Enter key.
36
V2 Reference Manual
Set Up Options in Menu
Selecting Electrodes
Use this function to turn “ON” or “OFF” each electrode. Figure 2-2 shows
the Test Select Screen.
Procedure
<< TEST SELECT >> 11:20:51
1. Na+ : ON 2. K+ : ON
MENU Screen
3. Cl- : ON 4. Li+ : OFF
>>Press Enter key to change. (ON,OFF) ↓
1
↓
Figure 2-2. Test Select Screen.
SETUP
Screen
2.Use keys to move the cursor, and press Enter key to turn “ON”
↓
or “OFF” the test electrode.
1
↓
3.After setting the test electrode, press Esc key twice to return to the Menu
TEST SELECT
Screen.
Screen
37
Selecting Units of Measurement
The Units Setting Option lets users select the units of measurement for
the analysis report of each test. Figure 2-3 shows the Units Setting Screen,
which lists all of the system tests and the units of measurement that, users
can select for each one.
1. The flowchart on the left side of this page is the step-by-step procedure
for Selecting Units of Measurement. Procedure
MENU Screen
<< UNITS SETTING >> 08:30:32
1. Na+ : mmol/L 2. K+ : mmol/L ↓
3. Cl- : mmol/L 4. Li+ : mmol/L 1
>>Press Enter key to change. (mmol/L,mEq/L) ↓
SETUP
Figure 2-3. Units Setting Screen. Screen
↓
2. Use keys to move the cursor, and press Enter key to change the 2
units of measurement. ↓
UNITS SETTING
3. After setting the units of measurement, press Esc key twice to return to Screen
the Menu Screen.
38
V2 Reference Manual
Setting Resolution Values
The Resolution Set Option lets users specify the number of decimal places
that the Biolyte 2000 reports for sodium, potassium, chloride, and lithium
results. Figure 2-4 shows the Resolution Set Screen with the alternate
resolution values.
1.The flowchart on the left side of this page is the step-by-step procedure
for Setting Resolution Values.
Procedure
<< RESOLUTION SET >> 08:30:32
1. Na+ : 123 2. K+ : 12.3
MENU Screen
3. Cl- : 123 4. Li+ : 1.23
>>Press Enter key to change. (123,12.3) ↓
1
↓
Figure 2-4. Resolution Set Screen. SETUP Screen
↓
2.Use keys to move the cursor, and press Enter key to change the 3
number of decimal places. ↓
RESOLUTION SET
3.After setting the resolution values, press Esc key twice to return to the Screen
Menu Screen.
Na+ x x
K+ x x
Cl- x x
Li+ x
39
Setting Two-point Calibration Interval
The 2-Pt Calibration Setup Option lets users specify the two-point
calibration interval for Biolyte 2000. Figure 2-5 shows the 2-Pt Calibration
Setup Screen.
1. The flowchart on the left side of this page is the step-by-step procedure
for Setting two-point Calibration Interval.
Procedure
<< 2-PT CALIBRATION SETUP >> 11:20:51
* Calibration interval : 2 hour
MENU Screen
>>Range : 1 - 12 hour ↓
1
Figure 2-5. 2-Pt Calibration Setup Screen. ↓
SETUP Screen
↓
2.Input the number 1~12 , and press Enter key to set the two-point 4
calibration interval. ↓
2-Pt CALIBRATION
3.After setting the two-point calibration interval, press Esc key twice to SETUP Screen
return to the Menu Screen.
4.If the input number is larger than 4, Biolyte 2000 will alarm, and return
to the previous value.
40
V2 Reference Manual
Setting Date and Time
The Biolyte 2000 displays the date and time on each screen and on every
printout. Users must enter the date and time when the instrument is
first powered up; users can change it at any time. The internal clock has
a battery that will maintain the correct date and time during long power
outages. Figure 2-6 shows the Date/Time Set Screen.
2.At the cursor, input a digit. Once input, cursor automatically jumps to
the next digit.
Note: For the year, once the last 2 digits are input, the first 2 digits automatically
change accordingly (e.g. Input 00, system automatically displays 2000).
41
3. Use keys to move the position of cursor to the desired place, and
use numerical key to input date or time. Then, at the end of the input,
press Esc key twice to return to the Menu Screen.
42
V2 Reference Manual
Setting Results/Printer
The Results Set Option lets users define the reference values, panic values,
and offsets values. Figure 2-8 shows the Results Set Screen.
1. The flowchart on the left side of this page is the step-by-step procedure
for accessing the Results Set Screen.
43
Setting Reference Values
The Reference Set Option lets users define the reference values of Biolyte
2000. Figure 2-9 shows the Reference Set Screen.
1. The flowchart on the left side of this page is the step-by-step procedure
for Setting Reference Values.
6. If the input number is larger than 1000, Biolyte 2000 will alarm, and
return to the previous value.
7. If the analysis data is out of the range of the reference values, Biolyte
2000 will automatically mark “<Ref” or “>Ref”.
44
V2 Reference Manual
Setting Panic Values
The Panic Set Option lets users define the panic values of Biolyte 2000.
Figure 2-10 shows the Panic Set Screen.
1.The flowchart on the left side of this page is the step-by-step procedure
for Setting Panic Values.
6.If the analysis data is out of the range of panic values, Biolyte 2000 will
automatically mark “<Pan” or “>Pan”.
45
Setting Offsets Values
The Offsets option lets users enter slope and intercept values for each
analytical parameter so that users can correlate results from the Biolyte
2000 with other instruments in users’ lab. Figure 2-11 shows the Offsets
Screen.
1.The flowchart on the left side of this page is the step-by-step procedure
for Setting Offsets.
Procedure
6.After setting the slope and intercept, press Esc key 3 times to return to
the Menu Screen.
Slope Intercept
46
V2 Reference Manual
Setting Operation Options
The Operation Options allow users to set the analysis result and
alarm option. Figure 2-12 shows the Operation Screen.
1. The flowchart on the left side of this page is the step-by-step procedure
for Setting Operation Options.
3.After setting the operation options, press Esc key to return to the Menu
Screen.
47
Table 2-2. Description of the Operation Options.
Patient copy OFF Biolyte 2000 will print two analysis results if it were set “ON”.
ID required OFF Users must input ID before analysis
procedure if it were set “ON”.
Calib. print ON Biolyte 2000 will print the calibration result
after calibration if it were set “ON”.
MV Print OFF Biolyte 2000 will print results of analysis & calibration
including voltage of signals if it were set “ON”.
Mess. Print ON 1. Biolyte 2000 will print results including worry
message if “occurred” if it were set “ON”.
Alarm ON 2. Table 2-3 shows the worrying messages.
System will alarm when error occurred if it were set “ON”.
Message Description
< Ref Means assay concentration is less than reference low limit.
> Ref Means assay concentration is greater than reference high limit.
< Pan Means assay concentration is less than panic low limit.
> Pan Means assay concentration is greater than panic high limit.
Drift Means E_0 drift is greater than E_0 drift limit.
NC Means the electrode isn’t calibrated.
48
V2 Reference Manual
3.Operation
Introduction 50
Software Description 50
Main Screen 50
Start/Standby Function Key 51
Samp. Type Function Key 52
Calibrate Function Key 53
Analyze Function Key 55
Status Function Key 59
Flowpath Function Key 60
Menu Function Key 61
Set Up Options 61
Operation Options 61
Data Options 61
Data Recall 62
Data Transfer 63
Data Clear 63
Software Version 65
Diagnostic 65
Maintenance 65
QC function Key 65
Biolyte 2000 Operation 66
Shortcut Procedures 67
Power On 67
Selecting Electrodes 67
Set Sample Type 67
Selecting Analysis Modes 67
Calibration 68
Manual Calibration 68
Auto-Calibration 68
Analysis : Serum/Plasma 68
Analysis : Whole Blood 69
Analysis : Urine 69
Analysis : CSF 69
49
Introduction
This chapter includes two parts: software description (software screen)
and analysis operation procedures. Most display screens and operation
procedures are described in this chapter.
Software Description
Main Screen
Push the Power Switch to “ON” position to turn the Biolyte 2000 on. Power
Switch is located at the backside of the Biolyte 2000 (see Figure 1-2).
After system initiated, the screen will change to the Main Screen.
50
V2 Reference Manual
1999/05/04 <<Biolyte>> 11:20:51
--------------------------------------
* Sample Type : Serum/Plasma
* Status : Na, K, Cl Ready
Status Description
“Parameter Indicates that the electrode(s) is ready and
ready” analysis can proceed.
Not ready Must perform calibration before proceeding to
analysis.
Flow path Must perform Flow Path procedure before
proceeding to other options.
In the Main Screen, users can use the waiting function keys as following :
Start/Standby, Samp. Type, Calibrate, Analyze, Menu, QC, Status, Flowpath.
1.In the Standby Screen, users only can use Start/Standby function key.
2.Under the Standby Mode, during auto-calibration, no calibration is carried out,
instead system will rinse STD-A for 5 seconds and REF for 2 seconds.
3.After pressing the Start/Standby key, system will run flush procedure, then return
to the Main Screen.
Note: After a power failure, when the electricity return, reload back to the previous standby or
normal mode.
51
Samp. Type Function Key
Users can press Samp. Type key in the Main Screen. Figure 3-3 shows the Sample Type Set Screen.
• Input 1~4 and press Enter key to change sample type, or press Esc key to return to the
Main Screen.
• When sample type is changed, the calibration status of Biolyte 2000 will change to NC.
Recalibrate must be done before proceeding to analysis procedure.
• When sample type change to urine, system will display the “HELP FOR URINE SAMPLE”
screen for analyze urine sample analyses (Figure 3-4)
52
V2 Reference Manual
Calibrate Function Key
Users can press Calibrate key in the Main Screen. Figure 3-5 shows 4 screens of the Calibration
Screen.
1.System will start to calculate calibration time when Calibrate key has been pressed.
2.During calibration procedure, press Status key to switch between the four Calibration Screens.
3.At the end of calibration, system will set and calculate next auto-calibration time.
4.If calibration procedure is interrupted, system will rinse the flow path with STD-A and REF.
5.At the end of calibration, system will printout Figure 3-6 (To set the calibration printout form,
see Setting Operation Options in chapter 2, Set Up for details).
53
1999/04/28 13:23:12
================================
<< Calibration >>
--------------------------------
Item Slope Range Mess.
---- ------ ------------- -------
Na+ 50.0 50.7 - 66.8 NC
1
K+ 50.0 48.4 - 64.5
Cl- -50.0 -59.9 - -39.1
================================
================================
Item STD-A [mV] STD-B [mV]
---- ------------ ------------
Na+ -120.55 -75.26 2
K+ 56.28 98.58
Cl- 123.45 78.02
================================
54
V2 Reference Manual
Analyze Function Key
Users can press Analyze key in the Main Screen. Figure 3-7A and Figure 3-7B shows
the Sample Data Screen.
Figure 3-7A. Sample Data Screen (for Serum/Plasma, Whole Blood and CSF).
1.System will clear error message of last analysis and begin calculating analysis time
when users press Analyze key.
◎ To input alphabets under the “ID” and “Patient Name” options, press . and
key to select the desired alphabet.
55
4. If barcode reader has been installed, users can read bar code of ID or Acc. by
barcode reader, when the cursor is at position of ID or Acc. When read data is out
of range, system will alarm.
5. The barcode reader model is CCD600 Serial Port interfaces made by GIGA-TMS
INC in TAIWAN. There is more operation information in the programming manual
of CCD Barcode Reader.
7. If operation option ID input is set “ON” (see Setting Operation Options in chapter
2,Set Up for details), ID must be input. If ID is not input when Analyze key is
pressed, system will show Figure 3-9.
10.Once sample analysis is finished, the printer will automatically print the results.
During printing, an Analysis Concentration Screen (Figure 3-11) will appear. Once
printing is finished, the system remains in the Analysis Concentration Screen for 5
seconds, or otherwise press any key to automatically return to the Analysis
Screen.
11. If the analysis procedure is interrupted, system will rinse the flow path with
STD-A and REF.
12. At the end of analysis, system will print results as shown in Figure 3-12 (set
analysis printout format, see Setting Operation Options in chapter 2, Set Up for
details).
57
1999/04/28 13:23:12
================================
<< Analysis >>
SAMPLE ID : A122113994
Access # : 1221159940
Name : S-M WU
Sequence # : 1
Analyze Date: 1999/04/28
Analyze Time: 13:24:10
Sample Type : Serum/Plasma
--------------------------------
Item Concentration Message 1
---- -------------- -----------
Na+ NC 2
K+ 3.12 mmol/L
Cl- 100.2 mEq/L Drift 3
================================
================================
Item STD-A [mV] Sample [mV] 4
---- ------------ ------------
Na+ -120.55 -75.26
K+ 56.28 98.58
Cl- 123.45 78.02
================================
(1) “Message Print” must be set “ON” in the Operation Option, for system to print this
message (See Setting Operation Options in chapter 2, Set Up for details).
(2) NC: Means that the sensor is not calibrated.
(3) Drift: Means that the voltage of STD-A taken during analysis is too large
compared with the voltage of STD-A during calibration.
(4) “MV Print” must be set “ON” in the Operation Option, for system to print this
message (See Setting Operation Options in chapter 2, Set Up for details).
58
V2 Reference Manual
Status Function Key
The Status Screen can call up in any screen by pressing the Status key.
Figure 3-13A, B, C shows the Status Screens.
1.Press Status key to change the Status Screens or press the Enter key to
print all status information.
2.The Status Screen 1 (Figure 3-13A) shows the last-conducted calibration slopes
of each sensor. The analysis test number will auto-count from 1 to 10,000.
3.The Status Screen 2 (Figure 3-13B) shows voltage values of each sensor
that has been selected. All signals change per second.
4.The Status Screen 3 (Figure 3-13C) shows the flow rate of calibration and analysis
procedure. All signals change per second.
59
Flowpath Function Key
Press Flowpath key in the Main Screen. System will prime air into
the analyzer until S-line, electrode train, and W-line are empty. Once
empty, screen will display to the Flowpath Screen 1 (Figure 3-14).
Either press Flowpath key again while in the Flowpath Screen 1 to prime all
fluids into the tubing and return to the Main Screen (Figure 3-1), or, system will
automatically prime all fluids after one hour of idling. Before priming all fluids into
the tubing, system will beep 10 times (once per second) and display the Flowpath
Screen 2 (Figure 3-15). To stay at flowpath mode (or empty tubing), press ESC key
during these 10 seconds.
60
V2 Reference Manual
Menu Function Key
Press Menu key in the Main Screen to enter Menu Screen. Figure 3-16 shows the
Menu Screen.
Input 1~6 and press Enter key to select function, or press Esc to return to the Main
Screen.
• Set Up Options
When users press Menu key in the Main Screen, system will enter the Menu Screen.
Input 1 and press Enter key to run set up procedures (see Set Up Options in chapter
2, Set Up for details).
• Operation Options
When users press Menu key in the Main Screen, system will enter the Menu Screen.
Input 2 and press Enter key to run operation procedures (see Setting Operation
Options in chapter 2, Set Up for details).
• Data Options
When users press Menu key in the Main Screen, system will enter the Menu Screen.
Input 3 and press Enter key to call up data options. Figure 3-17 shows the Data
Screen.
61
• Data Recall
Input 1 and press Enter key in the Data Screen to enter the Data Recall Screen.
Figure 3-18 shows the Data Recall Screen.
1.For seeks, users can input ID and press Enter key or read ID by barcode reader.
If users don’t input ID, press Enter key to seek all data (Figure 3-17). (NONE: means
for all ID)
62
V2 Reference Manual
• Data Transfer
Input 2 and press Enter key in the Data Screen to enter the Transfer Data Screen.
Figure 3-20 shows the Transfer Data Screen.
1.To run this function, users must connect analyzer to PC by serial port and use
ASTM protocol (see Service Manual for details).
2.When Biolyte 2000 is transferring data, screen shows Figure 3-18. The line 2 of
screen will display the data being transferred at the time.
• Data Clear
Input 4 and press Enter key in the Data Screen to enter Clear Data Screen. Figure 3-21
shows the Clear Data Screen.
1.Press Enter key to clear the saved data in memory, or press Esc key to return to the
Data Screen.
63
3.Figure 3-23 shows the Data Clear OK Screen. Press any key to return to the Data
Screen.
4.If system shows Figure 3-24, press any key to return to the Data Screen and
contact BioCare Customer Service Center.
64
V2 Reference Manual
• Software Version
When Menu key was pressed in the Main Screen, system will enter the Menu Screen.
Input 4 and press Enter key in the Menu Screen to enter the Software Version Screen.
Figure 3-25 shows the Software Version Screen.
• Diagnostic
When Menu key was pressed in the Main Screen, system will enter the Menu Screen.
Input 5 and press Enter key to run diagnostic procedures (for details, see Diagnostic
in chapter 6, Troubleshooting).
• Maintenance
When Menu key was press in the Main Screen, system will enter the Menu Screen.
Input 6 and press Enter key to run maintenance procedures (for details, see chapter
4, Maintenance).
QC function Key
Press QC key in the Main Screen to run QC analysis (for details, see chapter 5, QC).
65
Biolyte 2000 Operation
This prologue contains shortcut instructions for running samples.
The Biolyte 2000 is a sophisticated instrument with many operating functions and
convenient user features. This chapter explains in detail how to operate the Biolyte
2000. However, the Biolyte 2000 is also designed to accommodate users’ need to
run analyses quickly and simply without using all of the offered functions.
The following operation shortcuts allow users to analyze samples quickly without
altering the instrument’s set-up parameters.
Note: 1. Reference to chloride or lithium or both are only applicable to instruments with these
electrodes.
2. In order to test lithium, both sodium and potassium should also be tested at the same time.
66
V2 Reference Manual
Shortcut Procedures
The shortcut procedures are as shown on Figure 3-26.
→ Calibration → Analysis
Note: The Shortcut Procedures are for reference use only. For more detailed information, please
refer to the respective sections in this Reference Manual.
Power On
See Powering on the Instrument (in Chapter 1, Description) for details.
Selecting Electrodes
See Selecting Electrodes (in Chapter 2, Set Up) for details.
67
Calibration
Press Calibrate key in the Main Screen to run calibration procedures. There are two
modes of calibration build in Biolyte 2000, Manual Calibration and Auto-calibration.
Manual Calibration
To initiate a manual calibration on the Biolyte 2000, press Calibrate key in the Main
Screen.
Auto-Calibration
The Biolyte 2000 automatically initiates a calibration sequence 30 minutes after
POWER ON and approximately every 2 hours thereafter, or 2 hours after the last
manual calibration. To set two-point calibration interval, see Setting Two-point
Calibration Interval in chapter 2, Set Up for details.
Analysis : Serum/Plasma
1.All installation and calibration procedures must be done before running samples.
2.Table 3-3 shows the procedures for running samples on the two analysis modes.
Note: If users don’t press Analyze key when the probe is at the outside position, Biolyte 2000
will automatically alarm after 10 seconds and the probe will return to the inside position after
20 seconds.
3.Biolyte 2000 will print results and save data into memory automatically at the end
of each analysis, and show result on the screen.
4.After printing result, screen will return to the Analysis Screen (Figure 3-7).
68
V2 Reference Manual
Analysis : Whole Blood
1.All procedures for running whole blood are the same as running serum/plasma.
Sample type must be selected as “Whole Blood” (for details, see Set Sample Type
in chapter 3, Operation).
2.Biolyte 2000 will print result, save data into memory automatically at the end of
each analysis, and show result on the screen.
3.After printing result, screen will return to the Analysis Screen (Figure 3-7).
Analysis : Urine
1.Before analyzing urine samples, urine samples must be diluted with Urine Diluent
for Biolyte 2000 (Part. No.: 1030072). Dilute one part urine sample with two parts
diluent and mix well. After diluting urine sample, users can analyze urine sample.
All procedures of running urine are the same as running serum/plasma. Sample
type must be selected as “Urine” (for details, see Set Sample Type in chapter 3,
Operation).
2.Biolyte 2000 will print result, save data into memory automatically at the end of
each analysis, and show result on the screen.
3.After printing result, screen will return to the Analysis Screen (Figure 3-7).
Note: If the printed result of urine sample shows “Too low”, press Enter key to select “Diluted urine”
item as “No” in the Analysis Screen (Figure 3-7B), then analyze undiluted sample directly.
Analysis : CSF
Note: Only the concentration of chloride can be measured in CSF.
1.All procedures for running CSF are the same as running serum/plasma. Sample
type must be selected as “CSF” (for details, see Set Sample Type in chapter 3,
Operation).
2.Biolyte 2000 will print result, save data into memory automatically at the end of
each analysis, and show result on the screen.
3.After printing result, screen will return to the Analysis Screen (Figure 3-7).
69
4. Maintenance
Introduction 71
Maintenance Function 71
Condition 72
Condition Na 72
Condition f_path 75
Clean Flow Path 78
Reagent Information 81
Prime 82
Probe Adjustment 83
Calibrate Pump Frequency 84
Daily Maintenance 85
Checking the Reagent Pack Level 85
Cleaning the Sample Inlet Port 85
Conditioning Na Electrode 85
Conditioning the Flow Path with Serum 85
Checking Slope & Flow Time 86
Checking the Electrode Slope 86
Checking the Flow Time 86
Periodic Maintenance 86
Flow Path Maintenance Cycle 86
Replacing the W-line 87
Replacing the R-line 88
Replacing the S-line 89
Replacing the A-line/B-line 90
Replacing the Probe 91
Replacing the Electrodes 92
Non-routine Maintenance 93
Cleaning the Sample Inlet Port 93
Cleaning the Probe 93
Standby Mode 93
Replacing the Reagent Pack 94
Installation of Thermal Paper 97
Shutdown 98
70
V2 Reference Manual
Introduction
This chapter explains how to maintain Biolyte 2000. Maintenance includes both
daily and periodic maintenance procedures, as well as component replacement,
non-routine maintenance and instrument shutdown/start up procedures.
Note: References to chloride or lithium or both are only applicable to analyzer with these electrodes.
Maintenance Function
Biolyte 2000 maintenance functions shall be operated as described below.
Input 6 and press Enter key in the Menu Screen to enter the Maintenance Screen.
Figure 4-1 shows the Maintenance Screen.
Input 1~6 and press Enter key to select function or press Esc key to return to the
Menu Screen.
Table 4-1 shows the description of Maintenance Options.
Option Description
Condition Running procedures for conditioning
sodium electrode and flow path.
Clean flow path Running procedures for cleaning flow path.
Reagent info Showing the information of the reagent pack.
Prime Running procedures for priming the selected solution or air.
Probe adjustment Adjusting the probe position.
Cal. pump frequency Calibrating pump frequency.
71
Condition
Input 1 and press Enter key in the Maintenance Screen to enter the
Condition Screen. Figure 4-2 shows the Condition Screen.
• Condition Na
Note: When users perform this procedure, be sure to condition the sodium electrode first, then
condition the flow path with serum.
Instruction
• The sampler must be pulled out (push sample arm up) for sampling
Conditioning Na Solution.
• Fill the sample cup with 1ml CONDITIONING Na SOLUTION
(Part. No.: 1030019).
• From the Menu Screen, press 6 and Enter key to access the Maintenance
Screen.
• Select the Condition option (select 1 to access this option).
• Select the Condition Na option (select 1 to access this option).
• Follow the directions on the screen.
The cycle takes about 1 minute. When the cycle is finished, all
electrodes will be uncalibrated, and the status will display “NC”. Users
must run calibration procedures before running analysis.
72
V2 Reference Manual
Condition Na Screen 1
Input 1 and press Enter key in the Condition Screen to enter the Condition
Na Screen. Figure 4-3A, B, C, D, E, F shows the Condition Na Screens.
• Press Enter key to move the probe out, or press Esc key to return to the
Maintenance Screen.
Condition Na Screen 2
Press Enter key in the Condition Na Screen 1 to switch
to the Condition Na Screen 2 (Figure 4-3B).
73
Condition Na Screen 3
Press Analyze key in the Condition Na Screen 2 to switch
to the Condition Na Screen 3 (Figure 4-3C).
• Press Esc key once for Figure 4-3F, or press Esc key twice to return to the
Maintenance Screen.
Condition Na Screen 4
After priming Conditioning Na Solution, screen switches
to the Condition Na Screen 4 (Figure 4-3D).
Condition Na Screen 5
After condition Na electrode, screen will change to
the Condition Na Screen 5 (Figure 4-3E).
74
V2 Reference Manual
• Condition f_path
Note: Always condition the flow path with serum if sodium electrode has been conditioned.
Instruction
• The sampler must be pulled out (push sample arm up) for sampling serum.
• Fill the sample cup with serum.
• From the Menu Screen, press 6 and Enter key to access the Maintenance
Screen.
• Select the Condition option (select 1 to access this option).
• Select the Condition f_path option (select 2 to access this option).
• Follow the directions on the screen.
The cycle takes about 1 minute. When the cycle is finished, all
electrodes will be uncalibrated, and the status will display “NC”. Users
must run calibration procedures before running analysis.
75
Condition f_path Screen 1
Input 2 and press Enter key in the Condition Screen to enter the Condition f_path
Screen. Figure 4-4A, B, C, D, E, F show the Condition f_path Screens.
Press Enter key to move the probe out, or press Esc key to return to the Maintenance
Screen.
76
V2 Reference Manual
Condition f_path Screen 3
Press Analyze key in the Condition f_path Screen 2 to switch to the Condition f_path
Screen 3 (Figure 4-4C).
• Press Esc key once for Figure 4-4F, or press Esc key twice to return to the
Maintenance Screen.
77
Clean Flow Path
Instruction
• The sampler must be pulled out (push sample arm up) for sampling Cleaning
Solution.
• Fill the sample cup with CLEANING SOLUTION (Part. No.: 1030012).
• From the Menu Screen, press 6 and Enter key to access the Maintenance
Screen.
• Select the Clean Flow Path option (select 2 to access this option).
• Follow the directions on the screen.
• After cleaning flow path, run Condition f_path procedures and calibration
again before analyzing sample.
The cycle takes about 1 minute. When the cycle is finished, all electrodes will be
uncalibrated, and the status will display “NC”. Users must run calibration procedures
before running analysis.
78
V2 Reference Manual
Clean Flow Path Screen 1
Input 2 and press Enter key in the Maintenance Screen to enter the Clean Flow Path
Screen. Figure 4-5A, B, C, D, E, F shows the Clean Flow Path Screens.
Press Enter key to move the probe out, or press Esc key to return to the Maintenance
Screen.
79
Clean Flow Path Screen 3
Press Analyze key in the Clean Flow Path Screen 2 to switch to the Clean Flow Path
Screen 3 (Figure 4-5C).
• Press Esc key once for Figure 4-5F, or press Esc key twice to return to the
Maintenance Screen.
80
V2 Reference Manual
Reagent Information
Input 3 and press Enter key in the Maintenance Screen to enter the Reagent
Information Screen. Figure 4-6 shows the Reagent Information Screen.
81
Prime
Input 4 and press Enter key in the Maintenance Screen to enter the Prime Screen.
Figure 4-7 shows the Prime Screen.
Figure 4-8 shows the Priming Screen. Press Esc key to interrupt prime procedure,
and screen will show Figure 4-9.
Press Esc key to return to the Prime Screen when screen shows Figure 4-9.
Note: When users use the full prime function to empty the tubing, it is normal that the system
alarms three times.
82
V2 Reference Manual
Probe Adjustment
Input 5 and press Enter key in the Maintenance Screen to enter the Prime Screen.
Figure 4-10A shows the Probe Adjustment Screen 1.
• Input 1 or 2 and press Enter key to enter the Probe Adjustment Screen 2
(Figure 4-10B) or 3 (Figure 4-10C).
• Press Esc key to return to the Maintenance Screen.
• Option “2. SAMPLE”: Means the probe position of the Semi-Auto Mode (when
the sampler is inside).
83
Calibrate Pump Frequency
Input 6 and press Enter key in the Maintenance Screen to enter the Calibrate Pump
Frequency Screen 1 (Figure 4-11A).
After Calibrating pump frequency, screen will change to the Calibrate Pump
Frequency Screen 2 (Figure 4-11B).
84
V2 Reference Manual
Daily Maintenance
The following maintenance procedures must be performed
before users use Biolyte 2000 for the first time each day:
☉ Check the reagent pack level.
☉ Clean the sample inlet port.
☉ Condition the flow path with serum.
☉ Check the slope and flow time.
Note: 1. BioCare recommends conditioning flow path before daily operation. Users
must calibrate the instrument if electrode conditioning has just been performed.
2. All fluids in tubing are infectious, thus, must be handled with care! Wear gloves if necessary!
Perform the following steps to complete the daily maintenance procedures.
Note: The amount of the reagent shown in the Status Screen 2 is approximate. In some
cases a zero indicator can appear with some reagent remaining, or in other cases the
reagent will be entirely consumed and the indicator will show some reagent remaining.
Note: If whole blood is analyzed frequently, clean the sample inlet port more often.
Conditioning Na Electrode
See Condition Na in the earlier section of this chapter for details.
85
Checking Slope & Flow Time
Calibrate the instrument if electrode conditioning has just been performed.
Calibration is initiated by pressing Calibrate key.
If any electrode slope is not within the acceptable range, perform the appropriate
corrective procedures in Chapter 6, Troubleshooting.
Press Status key repeatedly until the Status Screen 2 (Figure 3-13B)
appears. This screen displays the flow times during the last calibration
and analysis cycles, as well as their acceptable ranges.
If either flow time is not within the acceptable range, perform the
appropriate corrective procedures in Chapter 6, Troubleshooting.
Periodic Maintenance
Periodic maintenance procedures must be performed according to
users’ scheduled timetables to assure optimal instrument performance.
Unless otherwise indicated, the flow path maintenance cycle must
be initiated before the start of any maintenance procedures.
The flow path maintenance cycle purges the flow path of fluid and
holds the instrument in an idle state while it is being serviced.
Note: Unless otherwise indicated, the Flow Path Maintenance Cycle must be initiated before any
component of the flow path is disconnected for maintenance or troubleshooting procedures.
86
V2 Reference Manual
Replacing the W-line
Figure 4-12 shows the location of the W-line. Replace the W-line according to the
following procedures:
87
Replacing the R-line
Figure 4-13 shows the location of the R-line. Replace the R-line according to the
following procedures:
88
V2 Reference Manual
Replacing the S-line
Figure 4-14 shows the location of the S-line. Replace the S-line according to the
following procedures:
89
Replacing the A-line/B-line
Figure 4-15 shows the location of the A-line and B-line. Replace the A-line and B-line
according to the following procedures:
90
V2 Reference Manual
Replacing the Probe
Figure 4-16 shows the location of Probe. Replace the Probe according to the
following procedures:
91
Replacing the Electrodes
Figure 4-17 shows the Electrode Carrier. Replace the electrodes according to the
following procedures:
92
V2 Reference Manual
Non-routine Maintenance
Non-routine maintenance procedures are performed in response to trends in
instrument operation that may generate error codes if left uncorrected, or as
corrective actions in response to inappropriate or incorrect procedures.
Standby Mode
The Standby mode is designed to keep the instrument in a functional state
indefinitely while conserving fluids during long periods of inactivity. In this mode
recalibration is suspended; however, an idle sequence is performed periodically
to advance a small segment of STD-A and REF through the flow path.
Note: In order to conserve fluids, the Biolyte 2000 does not perform recalibration while it’s in
Standby Mode.
93
Replacing the Reagent Pack
Replace the reagent pack and septum when the reagent pack has expired or is empty.
Note: 1. All fluids in tubing are infectious, thus, must be handled with care! Wear gloves if necessary!
2. Septum must be replaced together with the reagent pack.
2.After emptying all standard solutions, system will show Figure 4-18B.
94
V2 Reference Manual
6.After emptying all tubing, system will show Figure 4-18D.
95
16.Press Calibrate key to start full priming and system will show Figure 4-18E.
96
V2 Reference Manual
Installation of Thermal Paper
This section does not only explain installation of thermal paper but also explain the
replacement of thermal paper. Users can install/replace thermal paper according to
the following procedures.
If the printer fails to print after a new roll of paper is installed, investigate the
following causes:
☉ The printer paper is jammed. Check that the part number of the paper
(Part. No.: 5020097) is correct for the Biolyte 2000's printer.
☉ After installing the thermal paper, check the power light indicator. If the
power light indicator flashes, press the printer cover into the case.
If the above-mentioned are checked and none of them is the cause of the printer
failure, power off the Biolyte 2000 and try power on again.
97
Shutdown
Use the following shutdown procedures to prevent salt buildup in the flow path
when the Biolyte 2000 is turned off or unplugged.
Note: All fluids in tubing are infectious. Thus, they must be handled with care! Wear gloves if
necessary!
98
V2 Reference Manual
5. Quality Control (QC)
Introduction 100
Analyze Control 101
QC Setup 105
Print Westgard Rules Violation 108
Statistics 110
Browse QC Data 111
Print QC Data 112
Print L-J Chart 113
Browse Statistics 115
Print Statistics 116
Clear QC Data 117
99
Introduction
Monitor the performance of Biolyte 2000 by analyzing quality control samples.
Biolyte 2000 can track up to 7 QC levels and store the results in instrument’s
memory. These results will be used to calculate any statistics data that users regret.
Reports of historical data, including Levey-Jennings charts, can be printed when
necessary.
This package can check whether the QC test Results obey the 5 Westgard Rules. If
any rule is violated and error is reported, the Biolyte 2000 will alarm. Results that
fall outside the QC assay ranges will require verification by the users before they
become part of the historical data.
Access any QC functions from the Main Screen by pressing the QC key on
instrument.
Input 1~5 and press Enter key to select function or press Esc key to return to the
Main Screen.
100
V2 Reference Manual
Analyze Control
When this option is selected from the QC Screen, the screen displays the following:
Input the level number of QC samples and select which test items to
analyze. Operate as normal analysis procedures. Finally, the results will
be printed automatically and a message will be displayed on screen
asking users to confirm whether to accept the record or not.
Instruction
•From the QC Screen, press 1 , Analyze control, to access this option.
•Press key to move cursor.
•Use numerical key to input level numbers of QC samples, and use Enter key to turn
on or off the test items.
•Press Analyze key to start analysis.
•Check analysis results (as shown in Figure 5-3, on the following page).
101
1999/04/28 13:23:12
================================
<< Control Analysis >>
Level : 1
Lot No. : 12345678
Expiry Date : 1
Analyze Date: 1999/04/28
Analyze Time: 13:24:10
--------------------------------
Item Concentration Message 1
---- -------------- -----------
Na+ NC 2
K+ 3.12 mmol/L
Cl- 100.2 mEq/L Drift 3
================================
================================
Item STD-A [mV] Sample [mV] 4
---- ------------ ------------
Na+ -120.55 75.26
K+ 56.28 98.58
Cl- 123.45 78.02
================================
(1)Message Print must be set “ON” in the Operation Option, for system to
print this message (See Setting Operation Options in chapter 2, Set Up
for details).
(2)NC: Means that the sensor is not calibrated.
(3)Drift: Means that the voltage of STD-A taken during analysis is too large
compared with the voltage of STD-A during calibration.
(4)Voltage Print must be set “ON” in the Operation Option, for system to print
this message (See Setting Operation Options in chapter 2, Set Up for
details).
102
V2 Reference Manual
After printing the analysis results, screen will change to the Accept Screen (Figure
5-4).
If users input 1, screen will change to the Save QC Data Screen 1 (Figure 5-5). If users
input 2, screen will return to the Analyze Control Screen (Figure 5-2).
103
If saving QC data is OK, screen will change to the Save QC Data Screen 2 (Figure 5-6).
• Press any key to return to the Analyze Control Screen (Figure 5-2).
If there is not enough memory, screen will change to the Save QC Data Screen 3
(Figure 5-7).
• Press any key to return to the Analyze Control Screen (Figure 5-2).
• Use the Clear QC data function to clear all the old data before saving this data (See
later section in this chapter for details).
104
V2 Reference Manual
QC Setup
Biolyte 2000 can check the results of QC test by comparing the QC setup. The
QC setup includes QC samples levels, expiry date, lot number, and theoretical
concentration ranges.
Instruction
• From the QC Screen, press 2 , QC setup, to access this option.
• Press key to move cursor.
• Use numerical key to input level number, lot number and expiry date.
• Press Enter key to set range, mean & SD, and Westgard rule setting.
• Press Esc to return to the QC Screen.
105
QC Range Screen
Instruction
• Press key to move cursor.
• Use numerical key and Enter key to input the QC range.
Note: Analytes for which no ranges are entered or are missing will not be tracked. The resolution
depends on the Resolution Set (see Setting Electrolyte Resolution Values in chapter 2, Set Up for
details).
106
V2 Reference Manual
QC Mean & SD Screen
Instruction
• Press key to move cursor.
• “AUTO” means that the values of Mean and SD are determined by the
pervious QC data.
• Press QC key, then use numerical key and Enter key to input the QC range.
• Press Esc to return to the QC Setup Screen.
Instruction
• Press key to move cursor.
• Press Enter key to switch ON or OFF.
• Press Esc to return to the QC Setup Screen.
107
Print Westgard Rules Violation
This option provides users to print Westgard Rules violation chart. There are 5 rules
which can be applied to the QC sample.
(1) 1-2S: One sample with result greater than 2 standard deviations (SD) from
the Mean.
(2) 1-3S: One sample with result greater than 3 SD’s from the Mean.
(3) 2-2S: Two consecutive samples with results greater than 2SD’s from the
Mean.
(4) 4-1S: Four consecutive samples with result greater than 1 SD from the
Mean.
(5) 10X: Ten consecutive samples with result that fall on the same side of the
Mean.
108
V2 Reference Manual
Figure 5-14 shows the Westguard Rules Violation Chart
Format as printed by the Biolyte 2000.
1999/04/28 13:23:12
=================================
Westgard Rules Violation
--------- Level: 1234567---------
-
Cl 1-2S 1 3 *
1-3S * * *
2-2S * 3 *
4-1S * * *
10X * *
--------------------------------
+
Na 1-2S 1 3 *
1-3S * * *
2-2S * 3 *
4-1S * * *
10X * *
--------------------------------
+
K 1-2S 1 3 *
1-3S * * *
2-2S * 3 *
4-1S * * *
10X * *
--------------------------------
+
Li 1-2S
1-3S
2-2S
4-1S
10X
=================================
(1)Number (1-7): means the level in which violation for at least one data has
occurred.
(2) * : Means no violation.
(3)Blank: means that not enough data has been checked.
109
Statistics
This function group provides users to process, calculate, and print QC results. Once
selected, the screen displays the following. Users may input the desired function
number to access the function.
Input 1~5 and press Enter key to select function or press Esc key to return to the QC
Screen.
110
V2 Reference Manual
Browse QC Data
By this function, user can browse every historical data.
Instruction
• From the Statistics Screen, press 1 , Browse QC Data, to access the option.
• Press key to move cursor.
• Use numerical key to set date interval.
• Use key to move cursor to proper fields and use Enter key to
turn on or off the test items.
• Press QC key to switch screen to the QC Data Browse Screen 2 (Figure 5-17).
111
Print QC Data
Use this function to print the historical QC test data.
Instruction
• From the Statistics Screen, press 2 , Print QC data, to access the option.
• Use numerical key to set date interval.
• Use key to move cursor to specific fields and use Enter key to
turn on or off the test items.
• Press QC key to print data table (Figure 5-19 on the following page).
• Press Esc to return to the Statistics Screen.
Figure 5-19 shows the QC Data Format printed by Biolyte 2000.
1999/04/29 12:32:45
================================
<< QC DATA FOR Cl- LEVEL 1 >>
From 1999/04/01 to 1999/04/24
--------------------------------
Date Concentration
1999/04/02 132.5 mmol/L
1999/04/02 132.4 mmol/L
: : :
1999/04/24 132.0 mmol/L
================================
112
V2 Reference Manual
Print L-J Chart
When the Levey-Jenings charts option is selected, the screen displays the following:
Instruction
• From the Statistics Screen, press 3, Print L-J chart, to access this option.
• Use numerical key to set date interval.
• Use key to move cursor to specific fields and use Enter key to
turn on or off the test items.
• Press QC key to print L-J chart (Figure 5-21 on the following page).
• Press Esc to return to the Statistics Screen.
113
Figure 5-21 shows the L-J Chart Print Format printed by Biolyte 2000.
1999/04/29 12:32:45
================================
<< L-J CHARTS FOR Cl- LEVEL 1 >>
From 1999/04/01 to 1999/04/24
Date ---------------------------
04/02 0
04/02 1 |
04/03 | 4
04/04 |
04/10 |1
04/10 1 |
04/15 3 |
04/16 3 |
04/17 2 |
04/18 | 1
04/20 | 1
04/24 0
04/24 | 2
================================
114
V2 Reference Manual
Browse Statistics
Biolyte 2000 helps users to calculate results such as mean value, SD and %CV.
Instruction
•From the Statistics Screen, press 4 , Browse Statistics data, to access the
option.
• Use numerical key to set date interval.
•Use key to move cursor to proper fields and use Enter key to
turn on or off the test items.
• Press QC key to browse statistics data (Figure 5-23).
115
Print Statistics
Instruction
• From the Statistics Screen, Press 5, Print Statistics data, to access the
option.
• Use numerical key to set date interval.
• Use key to move cursor to specific fields and use Enter key to
turn on or off the test items.
• Press QC key to print statistics data (Figure 5-25).
• Press Esc to return to the Statistics Screen.
Figure 5-25 shows the Statistics Data Printout Format printed by the Biolyte 2000.
1999/04/29 12:32:45
================================
<< STATISTICS DATA FOR Cl- >>
From 1999/04/01 to 1999/04/24
Unit : mmol/L
--------------------------------
Control n Mean SD CV%
--------------------------------
Level 1 13 132.4 1.52 1.23
Level 3 13 102.4 0.95 0.89
Level 5 13 80.2 -0.62 0.51
================================
116
V2 Reference Manual
Clear QC Data
When the data stored in memory are full, users have to clear the
QC data in order to save more test results in the future.
Instruction
• From the QC Screen, press 5 , Clear QC data, to access the option.
• Ensure the decision again.
Note: Once cleared, data will be lost forever.
• Use key to move cursor to specific fields and use Enter key to select
level.
• Press QC key to clear data (Figure 5-27) or press Esc key to quit.
• Press Esc to return to the QC Screen.
117
6. Troubleshooting
Introduction 120
Troubleshooting Theory 120
Troubleshooting Procedure 121
Solving Flow Problems 121
Indicator of Flow Related Problem 122
Flow Times 122
Air and Fluid Readings 123
Confirming a Flow Problem 124
Possible Sources of Flow Problem 125
Flow Path Check 126
W-line Check 126
S-line/Electrode Carrier Check 127
Probe Check 128
A-line & B-line Check 129
R-line Check 129
Septum Check 130
Diagnostic 132
Device Exercise 132
Valve Exercise 133
Pump Exercise 134
Probe Exercise 135
Device Test 136
Speaker Test 136
Printer Test 137
LCD Test 138
Keypad Test 138
Sampler Sensor Test 139
Serial Port Test 140
Air Detector Test 141
Error Log 142
View Error Log 142
Print Error Log 143
Clear Error Log 144
System Test 144
Probe Position 145
Pump Movement 146
118
V2 Reference Manual
Valves Set 146
Sensor mV Data 147
Error Code 148
Introduction to Error Codes 148
Display Error Code 148
Problem Solving Procedures 148
Interpreting Error Codes 150
Error Code Reference 151
119
Introduction
This chapter describes troubleshooting procedure for the Biolyte 2000 and provides
instructions on how to use error codes and the Diagnostics Screen.
Troubleshooting Theory
The most efficient troubleshooting procedures focus on the simplest and fastest
way to resolve error codes. Since most error codes are due to flow problem in the
Biolyte 2000’s flow path, it is logical to investigate flow problems first. It is helpful
while troubleshooting to view the system as a whole and to consider how the flow
path components interact to transport fluids throughout the Biolyte 2000. When
multiple error codes occur, resolve the flow problems before troubleshooting the
remaining error codes. In most cases, the accompanying errors are resolved along
with the flow problems. Error codes that remain unresolved after flow problem has
been corrected must be investigated individually.
Do not replace any flow path components until you resolved the flow problems,
unless instructed to do so during the flow path check.
The information in this chapter and an understanding of the flow path components
and their functions are all you need to resolve most of the problems you will
encounter with your Biolyte 2000.
120
V2 Reference Manual
Troubleshooting Procedure
The recommended troubleshooting procedure consists of a series
of steps that are designed to resolve a malfunction in the most
direct and logical way. The following approach is used:
1. Eliminate obvious causes.
2. Troubleshooting individual error codes and groups of error codes by first
eliminating flow problems. Then resolve any remaining error codes.
121
• Indicator of Flow Related Problem
The Biolyte 2000 monitors flow during every calibration and analysis
cycle and provides the following useful flow indicators :
☉ Flow times
☉ Air and fluid readings
These flow indicators can help you to determine if a problem is flow related.
• Flow Times
Press Status key repeatedly to view the last calibration and analysis flow rate.
The calibration flow rate is the time (in seconds) it takes for Standard B to pass
through the air detector. When Standard B first reaches the air detector, the Biolyte
2000 begins measuring the time it takes for the air detector to see the end of the
Standard B stream.
Analysis flow rate is the time (in seconds) it takes for a sample to pass through the
air detector. When sample first reaches the air detector, the Biolyte 2000 begins
measuring the time it takes for the air detector to see the end of the sample stream.
The analysis flow rate of aqueous standards should be close to the calibration flow
rate. More viscous samples (such as whole blood) will have slightly longer flow rate.
It is possible for the instrument to have a good analysis flow time while it has a bad
calibration flow time, since each measures the flow rate of a different fluid. Because
a calibration flow time measures the flow rate of standard B, a problem such as
insufficient standard B in the reagent pack or a problem in the B-line can cause a
poor calibration flow time without affecting the analysis flow time.
The Biolyte 2000 performs most efficiently when the flow times fall within their
optimal ranges. Flow times that fall just outside of these ranges indicate a subtle
flow problem that will worsen over time (such as minor protein buildup in the flow
path, slightly worn tubing, or a slightly worn septum). You should investigate a flow
problem as soon as the flow times deviate from their optimal ranges.
Flow times that fall outside of their acceptable limits generate error codes 0404 (Flow
Time Too Short) or 0405 and 1405 (Flow Time Too Long). Error codes 0404 and 0405
refer to the calibration flow times, and error code 1405 refers to the analysis flow
times. Each of these errors indicates a more significant flow problem and precludes
accurate analytical results.
122
V2 Reference Manual
A calibration or analysis flow time of zero indicates that no calibration or analysis
has been performed since the instrument was powered down; therefore, data is
unavailable.
The air reading is the signal reading from the air detector when air should be
passing through it. The fluid reading is the signal reading from the air detector
when fluid should be passing through it.
The calibration or analysis cycle is a series of timed events. The air detected takes
the air reading when it expects air to be passing through it. Likewise, the air
detector takes a fluid reading when if expects fluid to be passing through it.
Flow readings like these usually generate error code 0400, Air Detector
Uncalibration, for calibration sequence, or 1400, Can’t Tell Air from Fluid, for analysis
sequence. These errors indicate a flow problem such as an obstruction or an air
leak.
If the air detector does become defective, if will gradually degenerate over time. As
the air detector accumulates protein buildup, it becomes less able to distinguish
between air and fluid readings; fluid readings increase while air readings decrease
so that the 2 readings become closer.
123
• Confirming a Flow Problem
If the flow indicators suggest a flow problem, confirm its existence by
performing the flow test with deionized water. The procedure requires a
small beaker or similar container of deionized water and a new S-line. Test
the entire flow path first. Then disconnect components where indicated
and retest in order to isolate the portion of the flow path in which the flow
problem is located. Complete the following steps to perform the Flow Test:
If you have confirmed but were unable to diagnose a specific flow problem,
perform the following steps before proceeding to the flow path check :
1.Flush the flow path with deionized water. Refer to Priming the Flow Path in
Chapter 4, Maintenance, for detailed instructions.
124
V2 Reference Manual
2.If flow is now unrestricted, close the door and perform a calibration to
confirm that the flow problem has been resolved.
3.If the flow problem persists, proceed to the flow path check.
1.W-line: The W-line is the only components that come in direct contact
with the pump. Damaged, blocked, or improperly installed pump tubing
can hinder suction.
2.Probe/S-Line: Clots, salt, or protein buildup in the probe and S-line can
hinder suction.
3.Electrode Carrier: Clots, salt, or protein buildup in the electrodes can
block the electrodes and cause instabilities. Missing or damaged gaskets
between electrodes can cause loss of suction.
4.Septum: The constant movement of the probe can eventually wear out
the septum and cause it to leak.
5.A-line, B-line and R-line: Damaged or blocked tubing in the harness can
cause loss of suction.
Detailed instructions for identifying and correcting these problems
are given in the following flow path check procedure.
125
• Flow Path Check
For troubleshooting purposes, the flow path has been
divided into the following 6 sections :
1.W-line Check
2.S-line/Electrode Carrier Check
3.Probe Check
4.A-line & B-line Check
5.R-line Check
6.Septum Check
If you have determined the exact location of a flow problem, troubleshoot only that
section of the flow path in which the problem originates.
If you have been unable to locate the source of the problem, troubleshoot the flow
path in the recommended order to avoid duplicating steps. Investigate and correct
each section completely before proceeding to the next one. Discontinue the flow
path check as soon as the flow problem has been eliminated.
• W-line Check
1.Disconnect the W-line from the top of the electrode carrier.
2.From the Menu Screen, input 6 to access the Maintenance function.
Select the “Prime” option.
Select the “Full” option.
3.Immerse the free end of the W-line into a beaker of deionized water and try
to aspirate water into the W-line. Watch for leaks and obstructions.
4.If there is no leak or blockage in the W-line, proceed to S-line/Electrode
Carrier Check. If the flow is slow or completely blocked, there is a leak or
obstruction in the W-line.
a.Detach the fitting on the W-line of the reagent harness tubing.
b.Remove the W-line from the pump.
c.Remove the tubing from the blunt syringe needle. Fill the syringe
with deionized water and connect the W-line to needle.
d.Flush the W-line to clear any obstruction.
e.Reinstall the W-line into the pump.
f.Reconnect the W-line to the top of the electrode carrier.
g.Repeat step 2 and step 3.
h.If the W-line is blocked or damaged and cannot be cleared, discard
the W-line and install a new one, then perform a calibration to see
if the flow problem is resolved.
126
V2 Reference Manual
i.If there is no leak or blockage, connect the W-line to the top of
the electrode carrier and perform a calibration to see if the flow
problem is resolved.
j.If the problem still persists, proceed to S-line/Electrode Carrier
Check.
• Probe Check
Warning: Be careful not to place your hands on or near the septum when extending the probe.
1.Pull the sampler up to the manual position.
2.Put a cup of deionized water underneath the probe.
3.From the Menu Screen, input 6 to access the Maintenance function.
Select the “Prime” option.
Select the “Sample” option.
4.Try to aspirate water into the probe. Watch for leaks and obstructions.
5. If there is no leak or blockage in the probe, proceed to Septum Check.
6. If the flow is slow or completely blocked, there is a leak or obstruction in
the probe.
a.Detach the fitting on the S-line of the probe.
b.Unscrew the probe cap and take off the probe.
c.Remove the tubing from the blunt syringe needle. Fill the
syringe with deionized water and connect the probe to needle.
d.Flush the probe to clear any obstruction.
e.Reinstall the probe into the probe cap.
128
V2 Reference Manual
f.Screw the probe cap tightly.
g.Reconnect the S-line to the top of the probe.
h.Repeat step 2 to step 4.
i. If the probe is blocked or damaged and cannot be cleared,
perform the following procedure :
(1) Discard the probe and install a new one.
(2) Perform a calibration to see if the flow problem is resolved.
j. If there is no leak or blockage, perform a calibration to see if the
flow problem is resolved.
k. If the problem still persists, proceed to Septum Check.
• R-line Check
1.Disconnect the R-line from the reference electrode.
2.From the Menu Screen, input 6 to access the Maintenance function.
Select the “Prime” option.
Select the “Reference” option.
129
3.Immerse the free end of the R-line into a beaker of deionized water and
try to aspirate water into the R-line. Watch for leaks and obstructions.
4.If there is no leak or blockage in the R-line, proceed to Septum Check.
5.If the flow is slow or completely blocked, there is a leak or obstruction in
the R-line.
a. Detach the fitting on the R-line of the reagent harness tubing.
b. Remove the tubing from the blunt syringe. Fill the syringe with
deionized water and connect the R-line to needle.
c. Flush the R-line to clear any obstruction.
a.Reconnect the R-line to the left-hand side of the electrode
carrier..
e. Repeat step 1 and step 2.
f. If the R-line is blocked or damaged and cannot be cleared,
discard the R-line and install a new one, then perform a
calibration to see if the flow problem is resolved.
(1) If there is no leak or blockage, perform a calibration to see if the
flow problem is resolved.
(2) If the problem still persists, proceed to Septum Check.
• Septum Check
1.Pull the sampler up to the manual position (remove the sampler cover if
necessary).
2.From the Menu Screen, input 6 to access the Maintenance function.
Select the “Prime” option.
Select the “Full” option.
3.Make sure that the probe is centered in the correct chamber of the
septum during each prime cycle.
Note: It is easier to see a reagent aspirating through the tubing harness if you create an air bubble in
the tubing. While pumping, withdraw and reinsert the straw into the reagent pack to create the air
bubble.
4.Observe each line during a prime cycle to see if reagent is being
aspirated.
5.If all of the reagents prime well, the septum is eliminated as a cause of the
flow problem. The flow path check is complete. If the preceding steps
failed to resolve the flow problem, contact BioCare Corporation’s
Customer Service Center.
6.If one or more of the reagents are not priming, there may be a problem in
the septum or reagent harness. Proceed as follows :
a.Verify the fitness of the septum :
(1)Check the fit of the septum on the sampler and tighten it if
130
V2 Reference Manual
needed.
(2)Reprime the reagents as before to see if the flow problem is resolved.
b.If the flow problem persists, install a new septum (see Chapter 4,
Maintenance) and prime the reagents as before. During the
prime observe that the probe moves to the correct positions in
the septum chambers.
7.If all the reagents are priming and the probe is properly positioned, the
septum and the tubing harness are now eliminated as a cause of the flow
problem.
8.If the reagents are not priming with the new septum and the probe is
improperly positioned, contact BioCare Corporation’s Customer Service
Center.
9.If reagents are not priming with the new septum, but the probe is
properly positioned, the flow problem lies in the tubing harness.
10.The septum is now eliminated as a cause of the flow problem.
131
Diagnostic
Press Menu key in the Main Screen. System will enter the Menu Screen. Input 5
and press Enter key to run diagnostic procedures. Figure 6-1 shows the Diagnostic
Screen.
Input 1~6 and press Enter key to select function, or press Esc key to return to the
Menu Screen.
Device Exercise
Input 1 and press Enter key in the Diagnostic Screen to enter the Device
Exercise Screen. Figure 6-2 shows the Device Exercise Screen.
Input 1~3 and press Enter key to select function, or press Esc key to return to the
Diagnostic Screen.
132
V2 Reference Manual
• Valve Exercise
Input 1 and press Enter key in the Device Exercise Screen to enter the Valve Exercise
Screen. Figure 6-3 shows the Valve Exercise Screen where exercise times and valve
status are displayed.
2.Press Esc key to interrupt this procedure and return to the Device Exercise Screen.
3.Figure 6-4 shows the Valve Exercise Finish Screen. Press Esc key to return to the
Device Exercise Screen.
133
• Pump Exercise
Input 2 and press Enter key in the Device Exercise Screen to enter the Pump Exercise
Screen. Figure 6-5 shows the Pump Exercise Screen where exercise times and pump
speed are displayed.
2. Press Esc key to interrupt this procedure and return to the Device Exercise Screen.
3. Figure 6-6 shows the Pump Exercise Finish Screen. Press Esc key to return to the
Device Exercise Screen.
134
V2 Reference Manual
• Probe Exercise
Input 3 and press Enter key in the Device Exercise Screen to enter the Probe Exercise
Screen. Figure 6-7 shows the Probe Exercise Screen where exercise times and probe
position are displayed.
2. Press Esc key to interrupt this procedure and return to the Device Exercise Screen.
3. Probe returns to initial position and the Probe Exercise Finish Screen (Figure 6-8)
appears. Press Esc key to return to the Device Exercise Screen.
135
Device Test
Input 2 and press Enter key in the Diagnostic Screen to enter the Device Test Screen.
Figure 6-9 shows the Device Test Screen.
Input 1~6 and press Enter key to select function, or press Esc key to return to the
Diagnostic Screen.
• Speaker Test
Input 1 and press Enter key in the Device Test Screen to enter the Speaker Test
Screen. Figure 6-10 shows the Speaker Test Screen.
• Press key to move cursor, then press Enter key to beep or alarm.
• Press Esc key to return to the Device Test Screen.
136
V2 Reference Manual
• Printer Test
Input 2 and press Enter key in the Device Test Screen to enter the Printer Test Screen.
Figure 6-11 shows the Printer Test Screen.
• After printing all character, system will automatically return to the Device
Test Screen.
• Press Esc key to interrupt printing and return to the Device Test Screen.
• Figure 6-12 shows the “finished” printout format.
• If Esc key is pressed to interrupt the printer test, on the printout, “<<PRINT
INTERRUPT>>” will be printed instead of “<<PRINT FINISH>>”.
137
• LCD Test
Input 3 and press Enter key in the Device Test Screen to enter the LCD Test Screen.
Figure 6-13 and Figure 6-14 shows the LCD Test Screen.
▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊
▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊
▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊
▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊
▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊▊
• Press Esc key to change screen, and to return to the Device Test Screen.
• Keypad Test
Input 4 and press Enter key in the Device Test Screen to enter the Keypad Test
Screen. Figure 6-15 shows the Keypad Test Screen.
• Press any key except Esc key. When a key is being pressed, system will display
the key pressed on line 2 of LCD screen.
• Press Esc key to return to the Device Test Screen.
138
V2 Reference Manual
• Sampler Sensor Test
Input 5 and press Enter key in the Device Test Screen to enter the Sampler Sensor
Test Screen. Figure 6-16 shows the Sampler Sensor Test Screen.
139
• Serial Port Test
Input 6 and press Enter key in the Device Test Screen to enter the Serial Port Test
Screen. Figure 6-17 shows the Serial Port Test Screen.
140
V2 Reference Manual
Air Detector Test
Press 3 and press Enter key in the Diagnostic Screen to access the Air Detector Test
function. Figure 6-18 shows the Air Sensor Test Screen and Figure 6-19 shows the
Test Result Screen.
141
Error Log
Input 4 and press Enter key in the Diagnostic Screen to enter the Error Log Screen.
Figure 6-20 shows the Error Log Screen.
Input 1~3 and press Enter key to select function, or press Esc key to return to the
Diagnostic Screen.
Press key to view more error log or press Esc key to return to the Error Log
Screen.
142
V2 Reference Manual
• Print Error Log
Input 2 and press Enter key in the Error Log Screen to enter the Printer Error Log
Screen. Figure 6-22 shows the Printer Error Log Screen.
1.When system is printing error log, users can press Esc key to stop printing.
2.Figure 6-23 shows printout format of Print Error Log. System will print the log
results from the last error to the first error (to a maximum of 50 errors in memory).
================================
Error Log
--------------------------------
(1) 1999/04/24 09:45:35
0105 Na+ Slope Too Low
(2) 1999/04/24 09:45:35
0205 K+ Slope Too Low
(3) 1999/04/22 10:35:12
1400 Can t Tell Air from Fluid
(4) 1999/04/20 16:54:03
2000 Illegal Probe Position
================================
3.After printing, system will display Figure 6-24. Press any key to return to the Error
Log Screen.
143
• Clear Error Log
Input 3 and press Enter key in the Error Log Screen to enter the Clear Error Log
Screen. Figure 6-25 shows the Clear Error Log Screen.
After clearing, users can press any key to return to the Error Log Screen.
System Test
Input 5 and press Enter key in the Diagnostic Screen to enter the System Test Screen.
Figure 6-26 shows the System Test Screen.
Input 1~3 and press Enter key to select function, or press Esc key to return to the
Diagnostic Screen.
144
V2 Reference Manual
• Probe Position
Input 1 and press Enter key in the System Test Screen to enter the Probe Position
Screen. Figure 6-27 shows the Probe Position Screen.
1. Input 1~6 and press Enter key to move the probe to selected position.
☉ Option “5. SAMPLE (IN)”: Means the probe position of the Semi-Auto
Mode (when the sampler is inside). If the sampler is at the “OUT” position
when users select this option, system will alarm.
☉ Option “6. SAMPLE (OUT)”: Means the probe position of the Manual Mode
(when the sampler is outside). If the sampler is at the “IN” position when
users select this option, system will alarm.
145
• Pump Movement
Input 2 and press Enter key in the System Test Screen to enter the Pump Movement
Screen. Figure 6-28 shows the Pump Movement Screen.
• Valves Set
Input 3 and press Enter key in the System Test Screen to enter the Valves Set Screen.
Figure 6-29 shows the Valves Set Screen.
• Use key to move cursor between Waste Valve and Reference Valve, then
press Enter key to close or open.
• Press Esc key to return to the System Test Screen.
146
V2 Reference Manual
Sensor mV Data
Input 6 and press Enter key in the Diagnostic Screen to enter the Sensor mV Data
Screen. Figure 6-30 shows the Sensor mV Data Screen.
Input 1 or 2 and press Enter key to select function, or press Esc key to return to the
Diagnostic Screen.
Figure 6-31 shows the Calibration mV Data Screen and Figure 6-32 shows the
Analysis mV Data Screen. They show the voltage values in the last calibration or
analysis procedure.
147
Error Code
This section describes error codes and explains how to display, interpret, and use
them while troubleshooting.
Error codes do not state what part of the Biolyte 2000 is malfunction; rather they
describe how the Biolyte 2000 is affected by a malfunction.
First troubleshoot the error codes that are flow related. Many flow problems
generate numerous slope and instability errors, which you should resolve before
you address any individual channel errors.
Do not try to change any electrodes or troubleshoot slope errors until you have
resolved the instrument’s flow problems.
Before you troubleshoot error codes, it is best to
☉ Check the flow time
☉ Perform the flow path check as described earlier in this chapter
148
V2 Reference Manual
This section identifies possible causes of error codes and includes instructions on
how to remedy them. When referring to the error code sheet outlining corrective
action, we recommend performing corrective actions in Solution Block A before
going to Solution Block B. If the recommendations giving here do not resolve an
instrument problem, contact BioCare Corporation’s Customer Service Center for
troubleshooting assistance.
Note: It is helpful to record the Biolyte 2000’s error codes, flow times, and slope performance
numbers before contacting BioCare Corporation’s Customer Service Center.
149
Interpreting Error Codes
☉ Error code consists of 4 characters.
☉ Character means :
X X XX
Class Location Particular
0- Na+ Channel
1- K+ Channel
0- Calibrate error Two digits between 01 and 19
2- Cl- Channel
3- Li+ Channel
4- Others
0- Na+ Channel
1- K+ Channel
1- Analyze error 2- Cl- Channel Two digits between 01 and 19
3- Li+ Channel
4- Others
0- Operation
1- Device
2- System error 2- Software Two digits between 01 and 19
3- QC &
Communication
4- Others
150
V2 Reference Manual
Error Code Reference
Code Message Page
0001 Na+ Instability STD-A 161
0002 Na+ Instability STD-B 161
0003 Na+ Overload STD-A 163
0004 Na+ Overload STD-B 163
0005 Na+ Slope Too Low 165
0006 Na+ Slope Too High 165
0101 K+ Instability STD-A 167
0102 K+ Instability STD-B 167
0103 K+ Overload STD-A 169
0104 K+ Overload STD-B 169
0105 K+ Slope Too Low 171
0106 K+ Slope Too High 171
0201 Cl- Instability STD-A 155
0202 Cl- Instability STD-B 155
0203 Cl- Overload STD-A 157
0204 Cl- Overload STD-B 157
0205 Cl- Slope Too Low 159
0206 Cl- Slope Too High 159
0305 Li+ Slope Too Low 173
0306 Li+ Slope Too High 173
0307 Li+ Slope Can’t Calculate 175
0400 Air Detector Uncalibrated 176
0401 No STD-A 177
0402 No STD-B 178
0403 No Sample 179
151
Code Message Page
0404 Flow Rate Too Short 180
0405 Flow Rate Too Long 181
0406 ADC Not Ready 197
1001 Na+ Instability STD-A 161
1002 Na+ Instability STD-B 161
1003 Na+ Overload STD-A 163
1004 Na+ Overload STD-B 163
1005 Na+ Conc. Too Low 184
1006 Na+ Conc. Too High 184
1007 Na+ E-0 Drift 185
1008 Na+ Instability Sample 161
1009 Na+ Overload Sample 163
1101 K+ Instability STD-A 167
1102 K+ Instability STD-B 167
1103 K+ Overload STD-A 169
1104 K+ Overload STD-B 169
1105 K+ Conc. Too Low 186
1106 K+ Conc. Too High 186
1107 K+ E-0 Drift 187
1108 K+ Instability Sample 167
1109 K+ Overload Sample 169
152
V2 Reference Manual
Code Message Page
1201 Cl- Instability STD-A 155
1202 Cl- Instability STD-B 155
1203 Cl- Overload STD-A 157
1204 Cl- Overload STD-B 157
1205 Cl- Conc. Too Low 182
1206 Cl- Conc. Too High 182
1207 Cl- E-0 Drift 183
1208 Cl- Instability Sample 155
1209 Cl- Overload Sample 157
1305 Li+ Conc. Too Low 188
1306 Li+ Conc. Too High 188
1307 Li+ E-0 Drift 189
1310 Li+ Cannot Calculate 175
1400 Can’t Tell Air from Fluid 190
1401 No STD-A 177
1402 No STD-B 178
1403 No Sample 179
1404 Short Sample 179
1405 Flow Rate Too Long 192
1406 ADC Not Ready 197
1410 Unreachable Pump Speed 191
2000 Illegal Probe Position 197
2001 Illegal Probe Speed 197
2002 Illegal Pump Speed 197
153
Code Message Page
2002 Illegal Pump Speed 197
2005 Illegal Error Code 197
2006 Can't Tell Air from Fluid 190
2007 No STD-A 177
2008 No STD-B 178
2009 No Sample 179
2010 LCD Over Flow 197
2102 Printer Out of Paper 193
2103 Printer Error 197
2104 Printer Off-Line 197
2106 ADC Error 197
2205 RMS I/O Error 194
2206 RMS Write Error 194
2210 RMS Analyzer Type Error 194
2211 RMS Pack Type Error 194
2212 Pack Lot Expired 195
2213 Pack Uselife Expired 195
2214 Standard A Empty 195
2215 Standard B Empty 195
2216 Reference Empty 195
2300 Westgard Rule 1-2s 193
2301 Westgard Rule 1-3s 193
2302 Westgard Rule 2-2s 193
2303 Westgard Rule 4-1s 193
2304 Westgard Rule 1x 193
2305 QC Control Outside Assay Range 196
2306 Transmit to PC Time-out 195
2307 Too Many NAKs from PC 195
154
V2 Reference Manual
Error Code
0201 Cl- Instability STD-A
0202 Cl- Instability STD-B
1201 Cl- Instability STD-A
1202 Cl- Instability STD-B
1208 Cl- Instability Sample
Solution Block A
1. Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, there may be bubble hang-up in the electrode.
Perform a flow path conditioning cycle with serum. Perform a calibration.
3. If the error persists, enter flow path maintenance.
☉ Clean electrode pins and connectors on sensor board with alcohol to eliminate
salt bridges.
4. If the error persists, enter flow path maintenance.
a. Remove the chloride and reference electrodes and flush them with
deionized water.
b. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
c. Reinstall the chloride and reference electrodes, exit flow path maintenance,
and perform a calibration.
5. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
6. If the error persists, there may be an obstruction. Perform a flow path check
(Chapter 6, Troubleshooting). Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the chloride electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice.
2. If the error persists, replace the reference electrode (Chapter 4,
155
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. Electrode slope error codes will also appear when an electrode overload condition exists.
Once the electrode overload error is resolved, the electrode slope error codes should disappear.
2. If an electrode overload exists on more than 1 electrode, the problem may stem from
an improperly attached electrode carrier.
3. If the sodium, potassium, or chloride overload errors occur, investigate the reference electrode.
4. Math errors will also appear if overload errors occur. Once the electrode overload error
is resolved, the math error should disappear.
156
V2 Reference Manual
Error Code
0203 Cl- Overload STD-A
0204 Cl- Overload STD-B
1203 Cl- Overload STD-A
1204 Cl- Overload STD-B
1209 Cl- Overload Sample
Description: A “Chloride Overload” error occurs when the millivolt readings for
standard A, standard B, or the sample exceed the acceptable operational limits of
the chloride electrode.
Solution Block A
1. Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, troubleshoot this error first.
3. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
4. If the error persists, enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
b. Remove the chloride and reference electrode and flush them with
deionized water.
c. Reinstall the electrodes
d. Exit flow path maintenance.
5. Check the tubing harness for air leaks near the straw connections.
6. Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1. Clean electrode pins and connectors on sensor board with alcohol. Salt bridges
on pins can cause overloads.
2. Replace the chloride electrode (Chapter 4, Maintenance).
3. Perform a flow path conditioning cycle with serum. Perform calibration twice.
4. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
157
Notes:
1. If the slope error is accompanied, resolve these error codes first.
2. Normal but negative slopes on all channels indicate Standard A and B are mixed up.
3. High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
4. If all channels have Slope Low or High errors, it is usually due to one or more of the following :
a. Flow problems
b. Incorrectly installed reagent pack
c. Contaminated reagent pack
d. Reference electrode problem
158
V2 Reference Manual
Error Code
0205 Cl- Slope Too Low
0206 Cl- Slope Too High
Solution Block A
1. Repeat the calibration to see if the error recurs. Note the slope before and after
conducting the calibration.
2. Check the level of Standard A and Standard B in the reagent pack. If levels are
low, replace the reagent pack (Chapter 4, Maintenance). If contamination is
suspected. Replace the reagent pack. Perform a calibration.
3. If the error persists, perform a flow path conditioning cycle with serum to
eliminate any bubble hang-up. Perform a calibration.
4. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
5. If the error persists, enter flow path maintenance and remove the reference and
chloride electrodes. Flush both with deionized water. Reinstall the electrodes.
Verify that sealing gaskets are present between all electrodes. Exit flow path
maintenance and perform a calibration.
6.If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the chloride electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice.
2.If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. If the slope error is accompanied, resolve these error codes first.
2. Normal but negative slopes on all channels indicate Standard A and B are mixed up.
3. High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
159
4.If all channels have Slope Low or High errors, it is usually due to one or more of the following :
a. Flow problems
b. Incorrectly installed reagent pack
c. Contaminated reagent pack
d.Reference electrode problem
160
V2 Reference Manual
Error Code
0001 Na+ Instability STD-A
0002 Na+ Instability STD-B
1001 Na+ Instability STD-A
1002 Na+ Instability STD-B
1008 Na+ Instability Sample
Solution Block A
1. Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, there may be bubble hang-up in the electrode. Perform a
flow path
conditioning cycle with serum. Perform a calibration.
3. If the error persists, enter flow path maintenance.
☉ Clean electrode pins and connectors on sensor board with alcohol to eliminate
salt bridges.
4. If the error persists, enter flow path maintenance.
a. Remove the sodium and reference electrodes and flush them with
deionized water.
b. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
c. Reinstall the sodium and reference electrodes, exit flow path maintenance,
and perform a calibration.
5. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting).
Change the R-line if necessary (Chapter 4, Maintenance). Perform a calibration.
6. If the error persists, there may be an obstruction. Perform a flow path check
(Chapter 6, Troubleshooting). Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1.Replace the sodium electrode (Chapter 4, Maintenance). Perform a sodium
conditioning cycle with Conditioning Na Solution, then a flow path conditioning
cycle with serum. Perform calibration twice.
2.If the error persists, replace the reference electrode (Chapter 4, Maintenance).
161
Perform calibrationtwice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. If the stability error code is accompanied by flow error codes or slope error codes, resolve these
error codes first, and the instability error codes should disappear.
2. Instabilities on patient samples may occur randomly as a result of interfering substances in the
sample.
3. If all channels are giving instabilities, either the electrode carrier is not attached correctly, or there
is a flow problem, or the reagent pack is empty.
4. If more than 1 electrode channel is giving instabilities, check the reference electrode.
162
V2 Reference Manual
Error Code
0003 Na+ Overload STD-A
0004 Na+ Overload STD-B
1003 Na+ Overload STD-A
1004 Na+ Overload STD-B
1009 Na+ Overload Sample
Description : A “Sodium Overload” error occurs when the millivolt readings for
standard A, standard B, or the sample exceed the acceptable operational limits of
the sodium electrode.
Solution Block A
1. Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, troubleshoot this error first.
3. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
4. If the error persists, enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
b. Remove the sodium and reference electrode and flush them with
deionized water.
c. Reinstall the electrodes
d. Exit flow path maintenance.
5. Check the tubing harness for air leaks near the straw connections.
6. Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1. Clean electrode pins and connectors on sensor board with alcohol. Salt bridges
on pins can cause overloads.
2. Replace the sodium electrode (Chapter 4, Maintenance).
3. Perform a sodium conditioning cycle with Conditioning Na Solution and then
perform a flow path conditioning cycle with serum. Perform calibration twice.
4. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
163
Notes:
1. Electrode slope error codes will also appear when an electrode overload condition exists. Once
the electrode overload error is resolved, the electrode slope error codes should disappear.
2. If an electrode overload exists on more than 1 electrode, the problem may stem from an
improperly attached electrode carrier.
3. Electrode overload errors typically stem from either a faulty electrode or an electrical problem.
4. If overload errors occur on more than 1 electrode channel, investigate the reference electrode.
5. Math errors will also appear if overload errors occur. Once the electrode overload error is
resolved, the math error should disappear.
164
V2 Reference Manual
Error Code
0005 Na+ Slope Too Low
0006 Na+ Slope Too High
Solution Block A
1. Repeat the calibration to see if the error recurs. Note the slope before and after
calibrating.
2. Check the level of Standard A and Standard B in the reagent pack. If levels are
low,replace the reagent pack (Chapter 4, Maintenance). If contamination is
suspected. Replace the reagent pack. Perform calibration twice.
3. If the error persists, perform a sodium conditioning cycle with Conditioning Na
Solution. Perform a flow path conditioning cycle with serum. Perform calibration
twice.
4. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
5. If the error persists, enter flow path maintenance and remove the reference and
sodium electrodes. Flush both with deionized water. Reinstall the electrodes.
Verify that sealing gaskets are present between all electrodes. Exit flow path
maintenance and perform a calibration.
6. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the sodium electrode (Chapter 4, Maintenance). Perform a sodium
conditioning cycle with Conditioning Na Solution.
Perform a flow path conditioning cycle with serum. Perform calibration twice.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
165
Notes:
1.If the slope error is accompanied, resolve these error codes first.
2.Normal but negative slopes on all channels indicate Standard A and B are mixed up.
3.High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
4.If all channels have Slope Low or High errors, it is usually due to one or more of the following:
a.Flow problems
b.Incorrectly installed reagent pack
c.Contaminated reagent pack
d.Reference electrode problem
166
V2 Reference Manual
Error Code
0101 K+ Instability STD-A
0102 K+ Instability STD-B
1101 K+ Instability STD-A
1102 K+ Instability STD-B
1108 K+ Instability Sample
Solution Block A
1 Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, there may be bubble hang-up in the electrode. Perform a flow
path conditioning cycle with serum. Perform a calibration.
3. If the error persists, enter flow path maintenance.
☉ Clean electrode pins and connectors on sensor board with alcohol to eliminate
salt bridges.
4. If the error persists, enter flow path maintenance.
a. Remove the potassium and reference electrodes and flush them with
deionized water.
b. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
c. Reinstall the potassium and reference electrodes, exit flow path
maintenance, and perform a calibration.
5. If the error persists, check the fluid delivery rate of the R-line (Chapter 6
, Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
6. If the error persists, there may be an obstruction. Perform a flow path check
(Chapter 6, Troubleshooting). Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the potassium electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice.
2. If the error persists, replace the reference electrode (Chapter 4,
167
Maintenance). Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. If the stability error code is accompanied by flow error codes or slope error codes, resolve these
error codes first, and the instability error codes should disappear.
2. Instabilities on patient samples may occur randomly as a result of interfering substances in the
sample.
3. If all channels are giving instabilities, either the electrode carrier is not attached correctly, or there
is a flow problem, or the reagent pack is empty.
4. If more than 1 electrode channel is giving instabilities, check the reference electrode.
168
V2 Reference Manual
Error Code
0103 K+ Overload STD-A
0104 K+ Overload STD-B
1103 K+ Overload STD-A
1104 K+ Overload STD-B
1109 K+ Overload Sample
Description : A “Potassium Overload” error occurs when the millivolt readings for
standard A, standard B, or the sample exceed the acceptable operational limits of
the potassium electrode.
Solution Block A
1.Repeat the analysis or calibration to see if the error recurs.
☉ Verify that the attachment of the electrode carrier to the Biolyte 2000 is secure.
2. If the error persists, troubleshoot this error first.
3. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
4. If the error persists, enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are formed.
b. Remove the potassium and reference electrode and flush them with
deionized water.
c. Reinstall the electrodes
d. Exit flow path maintenance.
5. Check the tubing harness for air leaks near the straw connections.
6. Perform a calibration.
7. If the error persists, continue to Solution Block B.
Solution Block B
1. Clean electrode pins and connectors on sensor board with alcohol. Salt bridges
on pins can cause overloads.
2. Replace the potassium electrode (Chapter 4, Maintenance).
3. Perform a flow path conditioning cycle with serum. Perform calibration twice.
4. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
169
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. Electrode slope error codes will also appear when an electrode overload condition exists.
Once the electrode overload error is resolved, the electrode slope error codes should disappear.
2. If an electrode overload exists on more than 1 electrode, the problem may stem from an
improperly attached electrode carrier.
3. Electrode overload errors typically stem from either a faulty electrode or an electrical problem.
4. If overload errors occur on more than 1 electrode channel, investigate the reference electrode.
5. Math errors will also appear if overload errors occur. Once the electrode overload error is
resolved, the math error should disappear.
170
V2 Reference Manual
Error Code
0105 K+ Slope Too Low
0106 K+ Slope Too High
Solution Block A
1. Repeat the calibration to see if the error recurs. Note the slope before and after
calibrating.
2. Check the level of Standard A and Standard B in the reagent pack. If levels are
low, replace the reagent pack (Chapter 4, Maintenance). If contamination is
suspected. Replace the reagent pack. Perform a calibration.
3. If the error persists, perform a flow path conditioning cycle with serum to
eliminate any bubble hang-up. Perform a calibration.
4. If the error persists, check the fluid delivery rate of the R-line (Chapter 6,
Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
5. If the error persists, enter flow path maintenance and remove the reference and
potassium electrodes. Flush both with deionized water. Reinstall the electrodes.
Verify that sealing gaskets are present between all electrodes. Exit flow path
maintenance and perform a calibration.
6. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the potassium electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
171
Notes:
1. If the slope error is accompanied, resolve these error codes first.
2. Normal but negative slopes on all channels indicate Standard A and B are mixed up.
3. High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
4. If all channels have Slope Low or High errors, it is usually due to one or more of the following:
a. Flow problems
b. Incorrectly installed reagent pack
c. Contaminated reagent pack
d. Reference electrode problem
172
V2 Reference Manual
Error Code
0305 Li+ Slope Too Low
0306 Li+ Slope Too High
Solution Block A
1. Repeat the calibration to see if the error recurs. Note the slope before and after
calibrating.
2. Check the level of Standard A and Standard B in the reagent pack. If levels are low
, replace the reagent pack (Chapter 4, Maintenance). If contamination is
suspected. Replace the reagent pack. Perform a calibration.
3. If the error persists, perform a flow path conditioning cycle with serum to
eliminate any bubble hang-up. Perform a calibration.
4.If the error persists, check the fluid delivery rate of the R-line (Chapter 6
, Troubleshooting). Change the R-line if necessary (Chapter 4, Maintenance).
Perform a calibration.
5.If the error persists, enter flow path maintenance and remove the reference and
lithium electrodes. Flush both with deionized water. Reinstall the electrodes.
Verify that sealing gaskets are present between all electrodes. Exit flow path
maintenance and perform a calibration.
6.If the error persists, continue to Solution Block B.
Solution Block B
1.Replace the lithium electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform 2 calibration.
2.If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1.The sodium and potassium channels must be calibrated for lithium analysis to work.
2. Normal but negative slopes on all channels indicate Standard A and B are mixed up.
3. If the slope error is accompanied, resolve these error codes first.
173
4. High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
5. If all channels have Slope Low or High errors, it is usually due to one or more of the following :
a. Flow problems
b. Incorrectly installed reagent pack
c. Contaminated reagent pack
d. Reference electrode problem
174
V2 Reference Manual
Error Code
0307 Li+ Slope Can’t Calculate
1310 Li+ Can’t Calculate
Description : A “Lithium Slope Can’t Calculate” error and “Lithium Can’t Calculate”
error occurs during a calibration cycle or analysis cycle when sodium and potassium
electrodes are unable to obtain a proper slope or reading.
Solution Block A
1. Check and resolve any error related to sodium and potassium channels. This error
will be solved automatically.
Notes:
1. The sodium and potassium channels must be calibrated for lithium analysis to work.
2. Normal but negative slopes on all channels indicate that Standard A and B are switched.
3. If the slope errors occur, resolve these error codes first.
4. High negative slopes indicate that the reagent pack may have been exposed to freezing
temperatures.
175
Error Code
0400 Air Detector Uncalibrated
Description : The “Air Detector Uncalibrated” error occurs during a calibration cycle
when the air detector cannot distinguish air from fluid by their signals.
Solution Block A
1. From the Menu Screen, input 5 to access the Diagnostic function.
Select the “Air detector test” option.
2. If the “Test result” shows “PASS”, perform a calibration.
3. If the “Test result” shows “FAIL”, press Flowpath key to remove the bubble in the
air detector. Perform a calibration.
4. Repeat step 3 several times.
5. If the error persists, replace the air detector (Chapter 4, Maintenance). Perform a
flow path conditioning cycle with serum. Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
176
V2 Reference Manual
Error Code
0401, 1401, 2007 No STD-A
Description : A “No Standard A” error occurs during an analysis cycle when the air
detector does not see Standard A at the appropriate time.
Solution Block A
1. Troubleshoot the reagent pack :
a. Check the level of Standard A in the bottle. If there is sufficient Standard
A, perform a Prime A cycle to confirm that Standard A is being pumped.
Perform a calibration.
b. If there is sufficient Standard A in the bottle, replace the reagent pack
(Chapter 4, Maintenance). Perform a calibration.
2. If the error persists, perform a flow path check (Chapter 6, Troubleshooting).
a. Check the reagent straw and the tubing harness for correct installation.
b. Verify that the probe is positioned in the center of the septum chambers
when performing the flow path check.
☉ If the probe position is incorrect, contact BioCare Corporation’s Customer
Service Center.
☉ If the probe position is correct, perform a calibration.
c. If an obstruction is found, clean the appropriate components with
deionized water. Perform a calibration.
3. If any damaged tubing is found during the flow path check, replace the reagent
harness or the pump tubing or both as needed (Chapter 4, Maintenance). Perform
a calibration.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. From the Menu Screen, input 5 to access the Diagnostic function.
Select the “Air detector test” option.
2. If the “Test result” shows “PASS”, perform a calibration.
3.If the “Test result” shows “FAIL”, press Flowpath key to remove the bubble in the air
4.Repeat step 3 several times.
5.If the error persists, replace the air detector (Chapter 4, Maintenance). Perform a
flow path conditioning cycle with serum. Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
177
Error Code
0402, 1402, 2008 No STD-B
Description : A “No Standard B” error occurs during an analysis cycle when the air
detector does not see Standard B at the appropriate time.
Solution Block A
1. Troubleshoot the reagent pack :
a. Check the level of Standard B in the bottle. If there is sufficient Standard
B, perform a Prime B cycle to confirm that Standard B is being pumped
Perform a calibration.
b. If there is sufficient Standard B in the bottle, replace the reagent pack
(Chapter 4, Maintenance). Perform a calibration.
2. If the error persists, perform a flow path check (Chapter 6, Troubleshooting).
a. Check the reagent straw and the tubing harness for correct installation.
b. Make sure the probe is positioned in the center of the septum chambers
when performing the flow path check.
☉ If the probe position is incorrect, contact BioCare Corporation’s Customer
Service Center.
☉ If the probe position is correct, perform a calibration.
c. If an obstruction is found, clean the appropriate components with
deionized water. Perform a calibration.
3. If any damaged tubing is found during the flow path check, replace the reagent
harness or the pump tubing or both as needed (Chapter 4, Maintenance). Perform
a calibration.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. From the Menu Screen, input 5 to access the Diagnostic function.
Select the “Air detector test” option.
2. If the “Test result” shows “PASS”, perform a calibration.
3. If the “Test result” shows “FAIL”, press Flowpath key to remove the bubble in the
air detector. Perform a calibration.
4. Repeat step 3 several times.
5. If the error persists, replace the air detector (Chapter 4, Maintenance). Perform a
flow path conditioning cycle with serum. Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
178
V2 Reference Manual
Notes :
If the flow error #0400 (Air Detector Uncalibrated) appear along with the error No Standard B,
change the septum.
Error Code
0403, 1403, 2009 No Sample
1404 Short Sample
Description :
A “No Sample” error code occurs during an analysis cycle when the air detector does
not see the sample within the allotted time.
A “Short Sample” error occurs during an analysis when an insufficient amount of
sample is seen by the air detector.
Solution Block A
1. Check the amount of sample. If it is sufficient, continue to Solution Block B.
Solution Block B
1. From the Menu Screen, input 5 to access the Diagnostic function.
Select the “Air detector test” option.
2. If the “Test result” shows “PASS”, perform a calibration.
3. If the “Test result” shows “FAIL”, press Flowpath key to remove the bubble in the
air detector. Perform a calibration.
4. Repeat step 3 several times.
5. If the error persists, replace the air detector (Chapter 4, Maintenance). Perform a
flow path conditioning cycle with serum. Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
179
Error Code
0404 Flow Time Too Short (Calibration Only)
Description :
A “Flow Time Too Short” error occurs during a calibration when the time required
for Standard A to travel from the probe tip (in the septum) to the air detector is less
than the allotted time.
Solution Block A
1. If the pump tubing was just replaced, release each tubing from the bracket, let it
relax, stretch it several times, and reposition it around the pump. Replace the
tubing in the bracket without twisting it.
2. If the error persists, perform a flow path check (Chapter 6, Troubleshooting).
☉ Check that the probe is positioned in the center of the septum chambers when
performing septum of the flow path check.
3. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace a S-line. Make sure correct part number is installed (Chapter 4,
Maintenance).
2. If the error persists and the pump tubing was not recently replaced, replace the
pump tubing (Chapter 4, Maintenance).
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
A flow error code has the highest priority in troubleshooting. If other electrode error codes appear
with the flow error, resolve the flow problem first, and the other errors should disappear.
180
V2 Reference Manual
Error Code
0405 Flow Time Too Long (Calibration Only)
Description :
A “Flow Time Too Long” error occurs during a calibration when the time required for
Standard A to travel from the probe tip (in the septum) to the air detector is greater
than the allotted time.
Solution Block A
1. If the pump tubing was just replaced, release each tubing from the bracket, let it
relax, stretch it several times, and reposition it around the pump. Replace the
tubing in the bracket without twisting it. Perform a calibration.
2. If the error persists, perform a flow path check (Chapter 6, Troubleshooting).
☉ Check that all sealing gaskets are in place.
☉ Verify that the probe is positioned in the center of the septum chambers when
performing septum check of the flow path check.
3. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the S-line. Make sure correct part number is installed (Chapter 4,
Maintenance).
2. If the error persists and the pump tubing was not recently replaced, replace the
pump tubing (Chapter 4, Maintenance).
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
A flow error code has the highest priority in troubleshooting. If other electrode error codes appear
with the flow error, resolve the flow problem first, and the other errors should disappear.
181
Error Code
1205 Cl- Conc. Too Low
1206 Cl- Conc. Too High
Description :
A “Chloride Concentration Too Low or Too High” error occurs when an analysis is
performed on a sample and the chloride is calculated to be above or below the
acceptable linearity range of the chloride electrode.
Solution Block A
1. If a flow error appears, resolve the flow problem first. Perform a flow path check
(Chapter 6, Troubleshooting).
2. If the sample’s concentration is too low or too high, the sample cannot be assayed
for chloride.
3. If there is no flow problem, perform a flow path conditioning cycle with serum.
Run a control with adequate chloride concentration to check the electrode
performance. Repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. The analytical range for chloride is 60-200 mM for serum, plasma, whole blood, and CFS. The
urine range is 40-400 mM.
2. No chloride results are available on samples that generate these 2 error codes.
182
V2 Reference Manual
Error Code
1207 Cl- E-0 Drift
Description :
A “Chloride E-0 Drift” error occurs during an analysis when the Standard A millivolt
drift between the last 2-point calibration and the 1-point analysis calibration
exceeds the acceptable operational limits of the chloride electrode.
Solution Block A
1. If flow errors also appear, resolve the flow problems first.
2. Enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are found.
b. Remove the sodium and reference electrodes and flush them with
deionized water.
c. Reinstall the electrodes.
d. Exit flow path maintenance.
3. Check the fluid delivery rate of the R-line (Chapter 6, Troubleshooting). Change
the R-line if necessary (Chapter 4, Maintenance). Perform a calibration and repeat
the analysis.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the chloride electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice and repeat the analysis.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice and repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
If all channels are giving E-0 Drift errors, the reagent pack may be empty. Troubleshoot reagent pack
before flow errors.
183
Error Code
1005 Na+ Conc. Too Low
1006 Na+ Conc. Too High
Description :
A “Sodium Concentration Too Low or Too High” error occurs when an analysis is
performed on a sample and the sodium is calculated to be above or below the
acceptable linearity range of the sodium electrode.
Solution Block A
1. If a flow error appears, resolve the flow problem first. Perform a flow path check
(Chapter 6, Troubleshooting).
2. If the sample’s concentration is too low or too high, the sample cannot be assayed
for sodium.
3. If there is no flow problem, perform a flow path conditioning cycle with
serum. Run a control with adequate sodium concentration to check the electrode
performance. Repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. The analytical range for sodium is 60-200 mM for serum, plasma, and whole blood. The urine
range is 40-400 mM.
2. No sodium results are available on samples that generate these 2 error codes.
184
V2 Reference Manual
Error Code
1007 Na+ E-0 Drift
Description :
A “Sodium E-0 Drift” error occurs during an analysis when the Standard A millivolt
drift between the last 2-point calibration and the 1-point analysis calibration
exceeds the acceptable operational limits of the sodium electrode.
Solution Block A
1. If flow errors also appear, resolve the flow problems first.
2. Enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are found.
b. Remove the sodium and reference electrodes and flush them with
deionized water.
c. Reinstall the electrodes.
d. Exit flow path maintenance.
3. Check the fluid delivery rate of the R-line (Chapter 6, Troubleshooting). Replace
the R-line if necessary (Chapter 4, Maintenance). Perform a calibration and repeat
the analysis.
4. If the error persists, perform a sodium conditioning cycle with Conditioning Na
Solution. Perform a flow path conditioning cycle with serum. Perform calibration
twice and repeat the analysis.
5. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the sodium electrode (Chapter 4, Maintenance). Perform a sodium
conditioning cycle with Conditioning Na Solution, then a flow path conditioning
cycle with serum. Perform calibration twice and repeat the analysis.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice and repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
If all channels are giving E-0 Drift errors, the reagent pack may be empty. Troubleshoot reagent pack
before flow errors.
185
Error Code 1105 K+ Conc. Too Low
1106 K+ Conc. Too High
Description :
A “Potassium Concentration Too Low or Too High” error occurs when an analysis is
performed on a sample and the potassium is calculated to be above or below the
acceptable linearity range of the potassium electrode.
Solution Block A
1. If a flow error appears, resolve the flow problem first. Perform a flow path check
(Chapter 6, Troubleshooting).
2. If the sample’s concentration is too low or too high, the sample cannot be assayed
for potassium.
3. If there is no flow problem, perform a flow path conditioning cycle with serum.
Run a control with adequate potassium concentration to check the electrode
performance. Repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. The analytical range for potassium is 1.0-10.0 mM for serum, plasma, and whole blood. The urine
range is 3-300 mM.
2. No potassium results are available on samples that generate these 2 error codes.
186
V2 Reference Manual
Error Code
1107 K+ E-0 Drift
Description :
A “Potassium E-0 Drift” error occurs during an analysis when the Standard A millivolt
drift between the last 2-point calibration and the 1-point analysis calibration
exceeds the acceptable operational limits of the potassium electrode.
Solution Block A
1. If flow errors also appear, resolve the flow problems first.
2. Enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are found.
b. Remove the sodium and reference electrodes and flush them with
deionized water.
c. Reinstall the electrodes.
d. Exit flow path maintenance.
3. Check the fluid delivery rate of the R-line (Chapter 6, Troubleshooting). Replace
the R-line if necessary (Chapter 4, Maintenance). Perform a calibration and repeat
the analysis.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the potassium electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice and repeat the analysis.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice and repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
If all channels are giving E-0 Drift errors, the reagent pack may be empty. Troubleshoot reagent pack
before flow errors.
187
Error Code
1305 Li+ Conc. Too Low
1306 Li+ Conc. Too High
Description :
A “Lithium Concentration Too Low or Too High” error occurs when an analysis is
performed on a sample and the lithium is calculated to be above or below the
acceptable linearity range of the lithium electrode.
Solution Block A
1. If a flow error appears, resolve the flow problem first. Perform a flow path check
(Chapter 6, Troubleshooting).
2. If the sample’s concentration is too low or too high, the sample cannot be assayed
for lithium.
3. If there is no flow problem, perform a flow path conditioning cycle with
serum. Run a control with adequate lithium concentration to check the electrode
performance. Repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. The analytical range for lithium is 0.1-5.0 mM for serum, plasma, and whole blood.
2. No lithium results are available on samples that generate these 2 error codes.
188
V2 Reference Manual
Error Code
1307 Li+ E-0 Drift
Description :
A “Lithium E-0 Drift” error occurs during an analysis when the Standard A millivolt
drift between the last 2-point calibration and the 1-point analysis calibration
exceeds the acceptable operational limits of the lithium electrode.
Solution Block A
1. If flow errors also appear, resolve the flow problems first.
2. Enter flow path maintenance.
a. Verify that sealing gaskets are present between all electrodes and that
tight seals are found.
b. Remove the sodium and reference electrodes and flush them with
deionized water.
c. Reinstall the electrodes.
d. Exit flow path maintenance.
3. Check the fluid delivery rate of the R-line (Chapter 6, Troubleshooting). Change
the R-line if necessary (Chapter 4, Maintenance). Perform a calibration and repeat
the analysis.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the lithium electrode (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice and repeat the analysis.
2. If the error persists, replace the reference electrode (Chapter 4, Maintenance).
Perform calibration twice and repeat the analysis.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
1. If all channels are giving E-0 Drift errors, the reagent pack may be empty. Troubleshoot reagent
pack before flow errors.
2. The sodium and potassium channels must be calibrated for lithium to work.
189
Error Code
1400, 2006 Can’t Tell Air from Fluid
Description :
The “Can’t Tell Air from Fluid” error occurs during an analysis cycle if the measured
signal difference between fluid and air did not fall within acceptable limits.
Solution Block A
1. Perform a calibration. If during the calibration the error #0400 (Air Detector
Uncalibrated) occurs, troubleshoot that error first (Chapter 6, Troubleshooting).
☉ When the Air Detector Uncalibrated error is resolved, the “Can’t Tell Air from
Fluid” error will disappear.
2. Enter flow path maintenance and remove the air detector. Flush it with Cleaning
Solution, allowing the solution to soak inside the air detector for 5 minutes. Flush
with deionized water. Reinstall the air detector, exit flow path maintenance, and
perform a calibration. Repeat the analysis.
3. If the error persists, continue to Solution Block B.
Solution Block B
1. Replace the air detector (Chapter 4, Maintenance). Perform a flow path
conditioning cycle with serum. Perform calibration twice.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
190
V2 Reference Manual
Error Code
1410 Unreachable Pump Speed
Description :
The instrument can auto-calibrate the speed of the peristaltic pump to optimize
the flow rate in the tubing set. If the instrument makes 3 attempts to calibrate the
peristaltic pump and still cannot adjust the flow rate to an acceptable range. Then
this warning message will be shown.
Solution Block A
1. Check that the tubing set, especially the W-line, is installed properly.
2. Run the “Calibrate Pump Frequency” again to properly set the pump speed.
Solution Block B
1. If the tubing set was not replaced for a period of time, replace the tubing set
(Chapter 4, Maintenance).
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
This error message is only a warning signal. It means the instrument is not ready to be used.
191
Error Code
1405 Flow Time Too Long (Analysis Only)
Description :
A “Flow Time Too Long” error occurs during an analysis (after the sample has already
reached the air detector) if the time required for the air detector to see the end of
the sample is greater than the allotted time.
Solution Block A
1. Check the sample for clots. Repeat the analysis.
2. If the error persists and the pump tubing was just replaced, release each tubing
from the bracket, let it relax, stretch it several times, and reposition it around the
pump. Replace the tubing in the bracket without twisting it. Perform
a calibration. Repeat the analysis.
3. If the error persists, perform a flow path check (Chapter 6, Troubleshooting).
☉ Check that all sealing gaskets are in place.
☉ Verify that the probe is positioned in the center of the septum chambers when
performing septum check of the flow path check.
4. If the error persists, continue to Solution Block B.
Solution Block B
1. If the pump tubing was not recently replaced, replace the pump tubing (Chapter 4,
Maintenance).
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
A flow error code has the highest priority in troubleshooting. If other electrode error codes appear
with the flow error, resolve the flow problem first, and the other errors should disappear.
192
V2 Reference Manual
Error Code
2102 Printer Out of Paper
Description :
A “Printer Out of Paper” error code occurs when the printer runs out of paper during
an analysis or calibration or when the printer cover is open.
Solution Block A
1. If there is sufficient paper or if the paper has just been changed, check the printer
cover. If it is open (lever forward), close it by pushing the lever back.
2. Press QC key, then input 3 (Print Westgard Rule Violation function) to reactivate
the printer.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Error Code
2300 Westgard Rule 1-2s
2301 Westgard Rule 1-3s
2302 Westgard Rule 2-2s
2303 Westgard Rule 4-1s
2304 Westgard Rule 1x
Description :
These error codes indicate that a Westgard Rule has been violated on a QC standard.
An explanation of the Westgard Rules is found in Chapter 5, Quality Control. These
errors do not necessarily indicate that there is anything wrong with the instrument.
Solution Block A
1. Make sure the correct QC standard is being analyzed.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
If other error codes appear along with the QC error, resolve those errors first.
193
Error Code
2205 RMS I/O Error
2206 RMS Write Error
Description :
The error means that the “Reagent Management System (RMS)” cannot access the
RMS or the RMS is missing.
Solution Block A
1. Check whether the RMS is installed.
2. Check the cable and the connection around the RMS plug.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Error Code
2210 RMS Analyzer Type Error
2211 RMS Pack Type Error
Description :
These error codes indicate that the reagent pack and the instrument are not
supposed to be used with each other.
Solution Block A
1. Check the type of the reagent pack and confirm its type with the supplier.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
194
V2 Reference Manual
Error Code
2212 Pack Lot Expired
2213 Pack Uselife Expired
2214 Standard A Empty
2215 Standard B Empty
2216 Reference Empty
Description :
These error codes indicate that the reagent pack is expired, out of date, or has been
used up.
Solution Block A
1. Please change to a new reagent pack.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Error Code
2306 Transmit to PC Time-out
2307 Too Many NAKs from PC
Description :
The errors indicate that a problem occurs during the communication between a PC
and the instrument. Error Code 2306 means that the instrument does not receive
any response from the PC and the initial handshake was not established correctly.
Error Code 2307 means that the information received by the PC does not correspond
to the information sent by the instrument.
Solution Block A
1. Check the cable’s connections and check that the type of cable being used is
correct.
2. Check that the settings of the PC are correct according to ASTM 1394 and ASTM
1381 protocol.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
195
Error Code
2305 QC Control Outside Assay Range
Description :
These error codes indicate that an analyte in a QC standard was outside the
specified range. An explanation of the QC package is found in Chapter 5, Quality
Control. This error does not necessarily indicate that there is anything wrong with
the instrument.
Solution Block A
1. Make sure the correct QC standard is being analyzed.
If these solutions do not resolve the problem, call BioCare Corporation’s Customer
Service Center.
Notes:
If other error codes appear along with the QC error, resolve those errors first.
196
V2 Reference Manual
Error Code
0406, 1406 ADC Not Ready
2000 Illegal Probe Position
2001 Illegal Probe Speed
2002 Illegal Pump Speed
2005 Illegal Error Code
2010 LCD Over Flow
2103 Printer Error
2104 Printer Off-Line
2106 ADC Error
Description :
These error codes usually stem from a hardware problem.
Solution Block A
1. Contact BioCare Corporation’s Customer Service Center.
197
7. Theory
Introduction 199
Origin of the Potential Measurement 200
Calculation of Na, K, Cl, and Li Concentration 204
Electrode Calibration 206
Two-Point Calibration 206
Two-Point Calibration Sequence 206
One-point Calibration 207
Direct vs. Indirect Methodologies 208
Plasma Volume Displacement Error 208
Bicarbonate 208
Protein Binding 208
Combined Effects 209
Halogen Effects on Chloride Determination 209
198
V2 Reference Manual
Introduction
The evolution of the ion selective electrode (ISE) as a measuring device began in
1875 with the recognition that glass is an electrolytic conductor.1 Initially, studies
were conducted using whatever kinds of glass were readily available, such as
laboratory glass tubing. As knowledge in the field increased, these glasses were
analyzed and their compositions altered to measure the effects of such change on
their diffusion rates and selectivity. Complex multi-component lithe glass and non-
glass materials that can function as ion selective membranes as they were identified
replaced the simple trioxide glasses originally used. In particular, ion exchange
resins with high selectivity were developed. When these resins were embedded
in structural matrices, their selectivity was further increased and their utility for
potentiometric measurement was enhanced. Ion selective electrodes like the pH
electrode are now commonly found in every clinical chemistry laboratory.
Regardless of the specific type of sensor used, all the ISE devices operate in the
same way as do the pH and blood gas electrodes developed in the 1950s by the
laboratories of MacInnes and Dole,2 and Severinghaus and Bradley.3
199
Origin of the Potential Measurement
To help understand the ion exchange phenomena that occur at the membrane
of the ion selective electrode, consider first the ion exchange taking place at
the surface of a simple permeable membrane, such as a dialysis membrane, in a
concentration cell between 2 salt solutions, A and B (Figure 7-1). Because solution B
has a smaller concentration of the ions in both solutions, a concentration gradient
is set up at the membrane. The concentration gradient creates a driving force to
send ions in solution from the cell where they are most concentrated across the
membrane to the cell where they are less concentrated. This force decreases, as
the concentration becomes more equal on both sides of the membrane. This
equilibration process results in equal salt concentrations on both sides of the
membrane and equal rates of migration both ways across the membrane.
Where subscripts o and i denote ions on either side of the membrane. In Figure
7-2, the dialysis membrane is replaced with an ion-selective membrane that can
differentiate among ions. For purposes of discussion, a K+ selective membrane is
used. This means the membrane can differentiate potassium ions from all other
ions in solution. As the ions in solutions A and B move about, the K+ ions enter into
the ion-selective sites on the membrane surface and leave behind the negatively
charged Cl- ions.
200
V2 Reference Manual
Figure 7-2 Establishment of membrane potential.
The chloride ions line up adjacent to the membrane and set up a potential that
prevents further movement across the membrane surface. At equilibrium, a
potential difference, or membrane potential, is established on the membrane
surface and the area adjacent to the membrane can be compared to a miniature
battery with the voltage equal to the potential difference of the membrane. The
membrane potential is dependent on the activities of the potassium ions on either
side of the membrane and was described by Nernst :
Where Em is the membrane potential, a is the activity of the potassium ions, R is the
universal gas constant, T is the temperature in degrees Kelvin, F is Faraday’s constant
and n is the charge of the measured ion:
+1 for sodium, potassium, and ammonium, -1 for chloride.
When the glass tubing in Figure 7-2 is filled with a salt solution of constant
composition, called the internal filling solution, (IFS) the electric potential of the
membrane depends only on the solution outside the membrane as follows :
Where Eo is a constant that includes a term for a0, the activity of the ion in the IFS.
We can convert from the natural log (ln) in Equation 3 to a base 10 log and device an
equivalent equation:
201
In order to measure the potential on the ion selective membrane; a complete
electrical circuit is needed. A typical ion selective electrode measuring circuit is
shown in Figure 7-3. There are 2 electrodes here: a reference electrode and the
ion selective electrode. The potential of the reference electrode is maintained
constant; that of the ISE varies, depending on the activity of the ion if interest in the
test solution. A voltmeter is used to measure the potential difference between the
ion selective electrode and the reference electrode.
The potential difference of the entire electrode measuring circuit, Ecell, is equal to
the algebraic sum of the potenial from the ion selective electrode, Em, the reference
electrode, Er, and the junction potential, Ej.
202
V2 Reference Manual
Equation 5 can be rewritten :
When the unknown activity of the ion of interest is measured in a test solution, the
potential of the test solution, Ex is compared with that of a known. Most of the terms
cancel out as follows :
203
Calculation of Na, K, Cl, and Li
Concentration
The activity of an ion is a measure of its “effective” concentration in solution. It
is equal to the product of C, its concentration in solution, and f, its ionic activity
coefficient. a=(f)(C). The activity coefficient relates concentration to activity, and
is a function of ionic strength. Thus, Equation 10 can be rewritten in terms of
concentration as follows :
The ionic strength of whole blood, plasma and serum tends to remain relatively
constant over the physiological range. 4 As a result, the activity coefficient of
sodium and potassium can be assumed to be constant. The internal standards are
formulated to reflect the same ionic strength as that of serum. Therefore, a given
ion’s activity coefficient can be assumed to be equal in the standard and sample.
Equation 11 can be rewritten to terms of concentration only as follows :
Where,
Thus, when Ecell, the potential of the test solution, is measured against that of a
standard solution in which the concentration of the ion of interest is known, the
potential difference between the 2 solutions is dependent only on the difference
between the concentrations of the ion of interest in the test and standard solutions :
Letting
204
V2 Reference Manual
By holding CStd in Equation 16 constant, E is dependent on only one variable, Cx, the
concentration of the ion of interest in the sample. Equation 16 can be rearranged to
isolate this variable :
205
Electrode Calibration
There are 2 major methods of electrode calibration: a 2-point calibration, and a
1-point calibration. A 2-point calibration, also called a full calibration, is used to
establish the electrode slopes for each sensor, and to verify electrode performance.
A 1-point calibration is used to monitor and correct the instrument for electrode
drift.
Two-Point Calibration
During a 2-point calibration, solutions of known ion concentrations are flushed
through the flow path. Each electrode of the sensor block measures the chemical
activity, in millivolt, of a particular ion in the solutions in order to obtain an electrode
slope. When samples with unknown ion concentrations are aspirated into the flow
path, each electrode takes a millivolt reading of chemical activity. To obtain the ion
concentration of the solution, these readings are compared to the electrode slopes
established during calibration.
206
V2 Reference Manual
• One-point Calibration
Electrode drift is a gradual change in electrode potential that can adversely affect
electrode performance. The Biolyte 2000 performs a 1-point calibration during
every analysis to determine if the drift for any electrode falls outside its acceptable
range. Detection of excessive drift generates the appropriate error codes,
depending on the analysis mode.
207
Direct vs. Indirect Methodologies
This section compares direct and indirect methodologies for the measurement
of analytes and explains how the Biolyte 2000 electrode offset values were
determined.
Bicarbonate
Sodium and potassium ions tend to bind to bicarbonate ions in solution to a small
but measurable degree. Because the Biolyte 2000 does not sense bound ions, this
bicarbonate binding introduces a normal sample. Citrate and acetate binding also
occur but normally with very little effect.
Protein Binding
Proteins are large molecules whose quaternary structures are associated with
varying charge densities. This property causes proteins to be zwitterions; that is,
they can exhibit anionic and cationic behavior as a function of pH. At physiological
pH, albumin and the globulin exhibit anionic behavior and therefore tend to bind
sodium and potassium ions to a small but measurable degree. This protein binding
introduces of sodium and potassium on normal samples. Other serum protein
molecules exhibit cationic behavior at physiological pH and tend to bind chloride
208
V2 Reference Manual
and bicarbonate. This tendency lowers direct ISE measurements of chloride and
bicarbonate by approximately 1.5% on normal samples.
Combined Effects
Serum electrolyte results from flame photometers and diluting instruments exhibit
a negative bias of 6-7% on normal samples, due to plasma volume error. Biolyte
2000 directs potentiometric instruments exhibit a negative bias of 3-4% due to ion
binding. For this reason, Biolyte 2000 electrolyte results tend to run 2-2% higher
than flame photometry or dilution ISE results for sodium and potassium.
Plasma Volume
+6% +6%
(Na+, K+, Cl-)
Protein Binding -1.5% —
209
1. Eisenmann, G., ed. Glass Electrodes for Hydrogen and Other Cations: Principles
and Practice. New York: Marcel Dekker, Inc. 1967.
2. D.A. MacInnes and M.Dole. The Behavior of Glass Electrodes of Different
Compositions. J Am Chem Soc Vol51. 1930:pp.29-36.
3. J.W. Severinghaus and A.F. Bradley. Electrodes for Blood pO2 and pCO2
Determination. J Appl Physiol Vol12. 1958:pp.515-520.
4. MS. Mohan and R.G. Bates Blood pH, Gases and Electrolytes. NBS Special
Publication 450,U.S Government Printing Office. 1977:p.293.
210
V2 Reference Manual
8. Appendix
Appendix I : Warranty of Products 212
Appendix II : Specifications of Biolyte 2000 215
Appendix III : Performance of Biolyte 2000 219
Appendix IV : Warning Messages 220
Appendix V : Labels on the Biolyte 2000 221
Appendix VI : Description of IEC Letter Symbols 221
Appendix VII : Location of Serial Ports 222
Appendix VIII : Error Codes 222
Appendix IX : Spare Parts List of Biolyte 2000 227
Appendix X : Consumable List of Biolyte 2000 228
211
Appendix I : Warranty of Products
The Products manufactured by BioCare Corporation (hereinafter called "BioCare")
are warranted against defects in materials and workmanship prior their expiration
date. The Instrument of Products is covered by a Warranty Period of 12 months
warranty from the date of delivery. BioCare's liability under this warranty is
limited, at the option of BioCare, for repair or replacement of the product. Item
expendable through normal use is not covered by this warranty. All warranty repair
or replacement of parts shall be limited to instrument malfunction, which, in the
sole opinion of BioCare, is due or traceable to defects in the original materials or
workmanship. All obligations of BioCare under this warranty shall cease in the event
of abuse, accident, alteration, misuse, or neglect of the Products. In warranty repair
or replacement, all parts are warranted for only the remaining unexpired portion
of the original Warranty Period, applicable to the repaired or replaced parts. After
expiration of the applicable Warranty Period, the customer will be charged at the
current prices for parts, labor and transportation.
Reasonable care must be used to avoid hazards to the Products. BioCare expressly
disclaims responsibility for loss or damage caused by the use of its ways other than
in accordance with proper operating procedures.
All claims under this warranty must be made promptly after the occurrence of
circumstances giving rise thereto, and must be received by BioCare or its authorized
representative within fifteen (15) days after the instrument malfunction. Such
claims should include the Products serial number, the date of purchase, and a full
description of the circumstances giving rise to the claim. Before any Products are
returned for repair and/or adjustment, written authorization from BioCare or its
authorized representative for the return and instruction as to how and where these
Products should be returned, must be obtained. Any Products returned to BioCare
for examination shall be sent freight prepaid via the means of transportation
indicated as acceptable to BioCare.
BioCare reserves the right to reject any warranty claim not promptly reported and
212
V2 Reference Manual
any warranty claim on any item that has been altered or has been returned by non-
acceptable means of transportation. When any product is returned for examination
and/or inspection, or for any other reason, the customer shall be responsible for
all damage resulting from improper packing or handling, and for loss in transit,
notwithstanding any defector nonconformity in the product. In all cases, BioCare
has the sole responsibility for determining the cause and nature of the failure;
BioCare's determination with regard thereto shall be final.
If it has been found that the Products has been returned without cause and is still
serviceable, the customer will be notified and the Products returned at its expense;
in addition, a charge for testing and examination may be made to the Products so
returned.
BioCare reserves the right to change, alter, modify or improve any of its Products
without any obligation to make corresponding changes to any Products previously
sold or shipped.
ALL THE DATA REPORTED FROM BIOLYTE ARE FOR CLINICAL REFERENCE, IF
ABNORMAL CLINICAL DATA OCCUR, THE USERS ARE SUGGESTED TO TAKE
FURTHER STUDIES AND EVALUATION OF THESE DATA TO ENSURE CLINICAL
SAFETY.
213
PLEASE FILL OUT YOUR REGISTRATION CARD (MANUFACTURER COPY),
CERTIFIED BY AUHORIZED SIGNATURE, THEN SENT IT BACK TO BioCare.
YOUR WARRANTY WITH BioCare CORPORATION FOR THE PRODUCTS WILL BE
ACTIVATED IMMEDIATELY AFTER THIS REGISTRATION.
214
V2 Reference Manual
Appendix II :
Specifications of Biolyte 2000
1. Measured Parameters :
☉ Sodium, Potassium, Chloride or Lithium.
2. Sample Type :
☉ Primary : Serum, Plasma, Whole Blood (Vacutainer, Syringe)
☉ Secondary : Urine (Manual Pre-dilution Required)
3. Sample Volume :
☉ 150 μL blood for measurement of sodium, potassium, and chloride or
lithium.
☉ 150 μL urine for measurement of sodium, potassium, and chloride.
4. Carryover :
☉ Less than 1% sample to sample
5. Analysis Time / Throughput
☉ 49 seconds / 73 samples per hour (maximum)
6. Calibration :
☉ Automatic 2-point calibration : User selectable intervals of 1-4 hours,
☉ 1-point calibration with each sample,
☉ Idle only in STANDBY MODE (i.e. no calibration)
7. Operation Modes :
☉ STANDARD MODE : Daily Use Operation (automatic calibration at user
selectable intervals, always ready).
☉ STANDBY MODE : Nights / Weekends Operation (idle only, flush on
startup).
8. LCD Screen :
☉ Character Format : 40 chars. x 4 lines
☉ Character Font 5 x 8 Dot + Cursor
☉ Character Color : Yellow Green
☉ Viewing Area : 148.0 mm x 30.3 mm
o
☉ Viewing Angle 30 C 150 o
☉ ype : Tranreflective with LED Back Light
9. Build-in Thermal Printer :
☉ Character Format : 32 characters per lines (32 Max)
☉ Character Font : 12 x 24 Dot
☉ Printing Width : 58 mm
☉ Printing Speed : 80 mm/s
215
10. User Interface :
☉ Data Transmission Port (RS232) Serial Interface : using ASTM protocol for
connection capabilities to personal computer
☉ Barcode Reader Port (PS/2 Port)
11. On-Board QC Software Package :
☉ QC data storage and reporting.
12. On-Board Patient Data Storage :
☉ Last 450 patient samples.
13. Data Power Protection :
☉ Time, date, last 2-point calibration parameters, last patient ID, setup
parameters and QC data are stored in nonvolatile memory chips. These
data are unaffected by power outages.
14. Power Requirement :
☉ 100-240 V~50-60 Hz 0.6A
15. Main supply voltage fluctuations not exceeding +/- 10% of the nominal voltage.
16. Equipment Pollution Degree :
☉ Pollution Degree 2
17. Equipment Installation Category :
☉ Installation Category II
18. Operation Condition :
o
☉ Relative humidity of 95% for temperatures up to 15~40 C at atmospheric
pressure 10.2~15.4 Psi.
19. Storage Condition :
o
☉ Relative humidity of 0-95% for temperatures up to -10~60 C at
atmospheric pressure 7.3~15.4 Psi.
20. Shipping :
☉ up to 2000m
21. Physical & Environmental :
☉ Height : 410 mm (16.1 inches)
☉ Width : 270 mm (10.6 inches)
☉ Depth : 245 mm (9.6 inches)
☉ Weight : 11 kg (24 lb)
☉ No hazardous waste, all disposable are land fill acceptable.
216
V2 Reference Manual
22. Power Supply Cords :
Countries Length (cm) Cord Color Wire Color Plug Power Rating Connector Power Rating Cable Type
Black
USA, etc 183 Black White WS-001;10A/125V WS-002;10A/125V STV 18AWG
Green
Brown H05VV-F 3G
UK, etc 183 Black Blue WS-012A;13A/250V WS-002;10A/250V 0.75MM
Green/Yellow SQ N.M
Blue H05VV-F 3G
French, etc 183 Black Brown WS-010;16A/250V WS-002;10A/250V 0.75MM
Green/Yellow SQ
Blue
Australia, 183 Pearl White Brown WS-015;10A/250V WS-002;10A/250V LTSA 3G
etc Green/Yellow 0.75MMSQ
Blue H05VV-F 3G
Italy, etc 183 Black Brown WS-013;10A/250V WS-002;10A/250V 0.75MM
Green/Yellow SQ
Blue H05VV-F 3G
Swiss, etc 183 Black Brown WS-014;10A/250V WS-002;10A/250V 0.75MM
Green/Yellow SQ
Blue H05VV-F 3G
South 183 Black Brown WS-016;10A/250V WS-002;10A/250V 0.75MM
Africa, etc Green/Yellow SQ
Blue H03VV-F 3G
Israel, etc 183 Black Brown WS-020;10A/250V WS-002;10A/250V 0.75MM
Green/Yellow SQ
23. Fuse :
☉ Time-Delay
☉ Low Breaking Capacity
☉ Physical Size: 0.197” x 0.788” (5mm x 20mm)
☉ Construction: glass tube, end caps: nickel plated brass
☉ Voltage Rating: 250V AC or less
☉Interrupting Rating: 35A at 250V (500mA – 3.15A), 10x In at 250V (4A – 6.3A)
☉ Current Rating Available: 2A
24. Regulatory Approvals :
☉ CE Mark ( EN/IEC61010-1, EN/IEC 61010-2-101 for safety & EN 61326-1, EN
61326-2 for EMC)
☉ FDA
217
25. Consumables and Accessories :
26.Warranty:
☉ The Products manufactured by BioCare Corporation are warranted
against defects in materials and workmanship.
218
V2 Reference Manual
Appendix III :
Performance of Biolyte 2000
1. Recommend operating range and general imprecision for Serum, Plasma, Whole
Blood
Range [mmol/L] Range [mmol/L] C.V.s Within Run,n=20 C.V.s Day-To-Day,n=20 Resolution [mmol/L]
219
Appendix IV: Warning Messages
1. The Biolyte 2000 is for indoor use only.
2. Do not use the Biolyte 2000 in a manner not specified by BioCare Corporation.
9. Do not use the service program, which is for service engineers only.
10. When the Biolyte 2000 alarms or there is any error message shown, stop
operating and check.
11. The electromagnetic environment should be evaluated prior to operation of the
device.
12. Use of this instrument in a dry environment and don’t use this instrument
in close proximity to sources of strong electromagnetic radiation, as these may
interfere with the proper operation.
13. User’s should ensure that a compatible electromagnetic environment for the
equipment can be maintained in order that the device will perform as intended.
14. This equipment complies with the emission and immunity requirements
described in this part of the EN 61326 series.
220
V2 Reference Manual
Appendix V :
Labels on the Biolyte 2000
Label Classification Label Name
Grounding Label
Use By Label
Manufacturer Label
Authorized Representative in
the European Community Label
Warning Label
Appendix VI :
Description of IEC Letter Symbols
Symbol Description Symbol Description
V Voltage | Power On
A Current
221
Appendix VII : Location of Serial Ports
222
V2 Reference Manual
2. Error Codes Table :
223
Code Message Description
1001 Na+ Instability STD-A See 0001
1002 Na+ Instability STD-B See 0001
1003 Na+ Overload STD-A See 0003
1004 Na+ Overload STD-B See 0003
1005 Na+ Conc. Too Low The concentration of a sample is below
the acceptable linearity range.
1006 Na+ Conc. Too High The concentration of sample is above the acceptable linearity range.
The standard A millivolt drifts between the last
1007 Na+ E-0 Drift 2-point calibration and the 1-point analysis calibration
exceeds the acceptable operational limits.
See 0001
1008 Na+ Instability Sample See 0003
1009 Na+ Overload Sample See 0001
1101 K+ Instability STD-A See 0001
1102 K+ Instability STD-B See 0003
1103 K+ Overload STD-A See 0003
1104 K+ Overload STD-B See 1005
1105 K+ Conc. Too Low See 1006
1106 K+ Conc. Too High See 1007
1107 K+ E-0 Drift See 0001
1108 K+ Instability Sample See 0003
1109 K+ Overload Sample See 0001
1201 Cl- Instability STD-A See 0001
1202 Cl- Instability STD-B See 0003
1203 Cl- Overload STD-A See 0003
1204 Cl- Overload STD-B See 1005
1205 Cl- Conc. Too Low See 1006
1206 Cl- Conc. Too High See 1007
1207 Cl- E-0 Drift See 0001
1208 Cl- Instability Sample See 0003
1209 Cl- Overload Sample See 1005
1305 Li+ Conc. Too Low See 1006
1306 Li+ Conc. Too High See 1007
1307 Li+ E-0 Drift The reading of the electrode cannot be interpolated to a
1310 Li+ Cannot Calculate meaningful value due to the improper reading of other electrodes.
224
V2 Reference Manual
Code Message Description
1400 Can’t Tell Air from Fluid See 0400
1401 No STD-A See 0401
1402 No STD-B See 0402
1403 No Sample See 0403
1404 Short Sample The air detector detected an insufficient amount of sample.
1405 Flow Time Too Long The time required for sample to travel from the probe tip
to the air detector is greater than the allotted time.
1406 ADC Not Ready See 0406
1410 Unreachable Pump Speed The peristaltic pump is not calibrating within
an acceptable range of speed.
2000 Illegal Probe Position The probe position value is out of its range.
2001 Illegal Probe Speed The probe speed value is not an acceptable value.
2002 Illegal Pump Speed The pump speed value is not an acceptable value.
2005 Illegal Error Code The error code is not acceptable.
2006 Can't Tell Air from Fluid See 0400
2007 No STD-A See 0401
2008 No STD-B See 0402
2009 No Sample See 0403
2010 LCD Over Flow The display on the LCD screen is over flow.
2102 Printer Out of Paper The printer has run out of paper.
2103 Printer Error Printer error exclude 2101 &2102
2104 Printer Off-Line Printer is on off line status new.
2105 - -
2106 ADC Error See 0406
2108 - -
2204 Value Over Flow Print data out of format
2205 RMS I/O Error Cannot initialize the communication and handshake with RMS.
2206 RMS Write Error Unable to write data into RMS.
2207 - -
2208 - -
2209 - -
2210 RMS Analyzer Type Error The reagent pack is used in the wrong type of the analyzer.
2211 RMS Pack Type Error The analyzer with the current settings does
not use this kind of reagent pack.
225
2212 Pack Lot Expired The reagent pack has passed it shelf life.
2213 Pack Uselife Expired The reagent pack has reached its open useful lifespan.
2214 Standard A Empty Standard A is used up.
2215 Standard B Empty Standard B is used up.
2216 Reference Empty Reference solution is used up.
2300 Westgard Rule 1-2s A Westgard Rule has been violated on a QC standard.
2301 Westgard Rule 1-3s See 2300
2302 Westgard Rule 2-2s See 2300
2303 Westgard Rule 4-1s See 2300
2304 Westgard Rule 1x See 2300
2305 QC Control Outside The concentration of control sample is out of acceptable limit.
Assay Range
2306 Transmit to PC Timed-out No response from the PC.
2307 Too Many NAKs from PC The checksum calculated by the PC does not
match the checksum sent by the analyzer.
226
V2 Reference Manual
Appendix IX :
Spare Parts List of Biolyte 2000
Part. No. Description
1010159 Main PCB
1010162 Keypad PCB
1010167 Peripheral PCB
1010161 Driver PCB
1040549 AMP Module
1040008 Electrode Carrier
2160015 Pinch Valve (Waste Valve or Reference Valve)
3010189 Keypad
3020070 Adhesive Backed Rubber Foot
2080191 Keypad Flat Cable
2080053 LCD Flat Cable
1990030 LCD Module (English & European)
2080079 Power Cord (Australia)
2080078 Power Cord (French)
2080083 Power Cord (Israel)
2080080 Power Cord (Italy)
2080082 Power Cord (South Africa)
2080081 Power Cord (Swiss)
2080077 Power Cord (UK)
2040011 Power Supply
2040038 Fuse
2080195 Printer Cable
1040508 Probe
3010192 Probe Screw
1040552 Pump M type
1040546 Sampler Module
1990001 Sampler Position Sensor
2990079 Thermal Printer Module
9010157 Biolyte 2000 V2 Reference Manual
227
Appendix X :
Consumable List of Biolyte 2000
Part. No. Description
1030033 Biolyte Na Electrode
1030031 Biolyte K Electrode
1030030 Biolyte Cl Electrode
1030032 Biolyte Li Electrode
1030034 Biolyte Ref Electrode
1030035 Biolyte Assay Electrode Spacer
1040010 Air Detector
1030108 Reagent Pack with septum
1030019 Conditioning Na Solution
1030012 Cleaning Solution
1030013 Chem Set A Level 1
1030014 Chem Set A Level 2
1030015 Chem Set A Level 3
1030016 Chem Set A Level 4
1030018 Chem Set A Level 6
3010193 Sample Cup
1040530 Septum
3030040 Septum Mainfold
5020097 Thermal Printer Paper
1030072 Urine Diluent
1990040 W-LINE
1040509 Tubing Set
228
V2 Reference Manual
2015 Biolyte 2000 V2 Reference Manual Version 1.0 9010157