MT - 96

10298, 14298 ■ Analysis Flow

Check the error log in M-Utility.

■ Error Occurrence Conditions


N
Errors 10230-10236 logged?
10298 Before-reading scanner retry error
Y
<Service Procedure>
Check the error code table
This error occurs when the version of the M-Utility is checked or error log backup is executed again and troubleshoot.
during RU bootup. “2. Error Code Table”

<Occurrence Condition>
1
During routine processing, because the scanner-related error is detected, the IP is returned to the
cassette and the scanner is initialized. The routine processing is retried again, but the scanner-
Other software exited before RU N
related error recurs. startup? Exit the other software.

Y
14298 Before-reading scanner retry error “Maintenance Utility Volume:
2.1.3 VER: Procedures for Checking the Software Version of the RU
During routine processing, because the scanner-related error is detected, the IP is returned to the (Implemented with RU Software Version of A07 or Later, or PCMUTL 1.2
cassette and the scanner is initialized. or Later)”

TP2 voltage on the SCN12A N Replace the power supply.


board normal?
“Check, Replacement,
and Adjustment Volume:
Y
“9.1 Checking the Voltage 11.8 Power Supply”
on the Board Test Pins
(Connectors)”
Fuses H11, H12, H13, and H14 N Replace the fuse.
on the SCN12A board normal?
“Check, Replacement,
and Adjustment Volume:
Y
12.2 SCN12A Board Fuses”
“10.4 Scanner-Related Fuses”

N
Error recurs? 1
Y

CPU12A and SCN12A boards Reseat the error-causing board


tested in M-Utility normal? N and check if the error occurs N
again.
1
Y
“8. Board Tests Y
in M-Utility”

Replace the error-causing board.


“Check, Replacement, and
Adjustment Volume”
“11.2 CPU12A Board”
N
Error recurs? 1 “11.3 SCN12A Board”

Reseat the boards in the order named,


and if the error is detected, replace the board.
1. SCN12A board
2. CPU12A board
3. MTH12A board
4. Scanning optics unit

“Check, Replacement, and Adjustment Volume”


“11.3 SCN12A Board”
“11.2 CPU12A Board”
“11.7 MTH12A Board”
“8.1 Scanning Optics Unit” FR6H2471.EPS

010-051-06 CR-IR346RU Service Manual [5.] MT - 96