Wireless:

WiFi &W R Quick Start


This sheet is not intended to replace a thorough reading of the manual.

To connect WiFiX:
1. Locate the Antenna Cover at the top of the ICU and unscrew it.
Warning: Ensure that the Antenna Cover is put back in place when you are not working wireless
2. Attach an Antenna to the screw, turn until you feel resistance.
3. With the cable attached to the USB Module, connect the USB Module to the CDU.
Note: Use the USB Extension Cable provided in the WiFiX Kit to position the USB Module at a distance from the
CDU when required.
4. Screw the bottom of the Antenna into the connector located at the top of the USB Module until you
begin to feel resistance.
5. Affix the screw on the Mounting Bracket to the opening at the back of the USB Module. Use the
butterfly fastener attached to the side of the Mounting Bracket to firmly secure the Mounting
Screw into place, while adjusting the bracket joint at the angle desired.
Note: The ideal angle for wireless transmission is 90especially when working in rough terrain. Therefore, the
Antenna is positioned vertically and the Mounting Bracket is positioned perpendicularly forming a 90 angle with
the Antenna.
6. Press pliers located at the other end of the Mounting Bracket to open them. Position the now
attached Antenna, USB Module, and Mounting Bracket affixing them on top of the opened
carrying case lid or anywhere else desired so that the Antenna is pointed upwards.
Warning: Opening the pliers takes some force. Beware that the Pliers do not snap shut on your fingers.
Note: The Mounting Bracket can be connected to other places as needed. The module can also be mounted onto
a standard (off-the-shelf) tripod.
Note: The higher the location of the USB Module and Antenna, the better wireless connection attained.
Note: For the highest quality reception, ensure that both Antennas on the CDU and ICU are parallel to each
other.
7. Press the Wireless button located at the top of the ICU after all the necessary connections have
been made.
8. Return to the CDU and login to the system.

9. From the Grab Mode toolbar, click .


Two things occur:
a) Wait until the status indicator at the bottom right corner of the Status Bar changes to
.
b) Ensure that the Imager is Online by the indicator in the right corner of the status bar

.
Note: When using the WiFiX option, it may take some time for the Imager to become online depending on your
working environment.
Note: When using the WiFiX Wireless connection the link status changes to 54 Mbps from 100 Mbps as with a
standard wire connection.
Note: The WiFiX’s range is up to a maximum of 200m
300m line of sight.

Code: 1QSWIR00100 V06/2011


V08/10
Wireless:
WiFi &W R Quick Start
This sheet is not intended to replace a thorough reading of the manual.

To initialize the WXR:


1. Log-in to the Xbit / XbitPro system.
2. From the Options menu, select Environment.
3. From the menu bar on the left side of the Environment window, select X-ray Source.
4. Under Wireless Properties and beside the words Pin Code, in the text field boxes enter the 12
digit Pin Code as it appears on the sticker on your WXR.
5. When you are finished, you can:
a) Click Apply to add the Pin Code and remain in the Environment window.
b) Click OK to add the Pin Code and close the Environment window.
You are now ready to connect your WXR.

To connect the WXR:


1. Ensure that the ICU/X-ray Connecting Cable from the ICU and X-ray Source is disconnected.
2. Insert one end of the 5-pin Adapter Cable labeled WXR TO GOLDEN X.R 5 PIN into the terminal on
the bottom of the WXR Wireless X-ray module.
The red points on the connector and WXR terminal must align to make the correct connection.
3. Connect the other end of the 5-pin Adapter Cable to the cable connector at the back of the X-ray Source.
The red point on the connector and the black point above the X-ray Source cable connector must
align to make the correct connection.
4. Fasten the Velcro on the back of the WXR to the Velcro area on the X-ray Source.
5. Return to the CDU and login to the system.
6. You can either:

a) From the Grab Mode toolbar, click .


b) From the Options menu, select Internal WXR Wireless X-ray.
Note: The Internal WXR Wireless X-ray option is only available in the Options menu when you are
connected to the ICU.
7. The X-ray Source is now ready to take a Wireless X-ray.
Note: The WXR receives its power from the X-ray Source’s internal batteries. When the X-ray Source is turned
on the LED on the WXR is green.
Caution: Make sure the X-ray Source battery is fully charges before you begin your inspection.

Code: 1QSWIR00100 V06/2011