Sunteți pe pagina 1din 7

Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576

The linear variable differential transformer (LVDT)


position sensor for gravitational wave interferometer
low-frequency controls
Hareem Tariqa,b,*, Akiteru Takamoria,c, Flavio Vetranod,e, Chenyang Wanga,
Alessandro Bertolinia,f, Giovanni Calamaid,g, Riccardo DeSalvoa,
Alberto Gennaih, Lee Hollowayi,h, Giovanni Losurdod, Szabolcs Ma! rkaa,
Massimo Mazzonid,j, Federico Paolettih, Diego Passuelloh, Virginio Sannibalea,
Ruggero Stangad,j
a
LIGO project, California Institute of Technology, 1200 E. California Blvd, Pasadena, CA 91125, USA
b
Florida Institute of Technology, 150 W. University Bl., Melbourne, FL 32901, USA
c
Department of Physics, University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan
d
INFN Sezione di Firenze, via G. Sansone 1, 50019 Sesto Fiorentino, Firenze, Italy
e
Istituto di Fisica, Universita di Urbino, via S. Chiara 27, 61019 Urbino, Italy
f
Dipartimento di Fisica, Universita! di Pisa, via F. Buonarroti, 4, 56100 Pisa, Italy
g
Osservatorio di Arcetri, Largo E. Fermi 4, 51125 Firenze, Italy
h
INFN Sezione di Pisa, via Livornese 1291, S. Piero a Grado, 56100 Pisa, Italy
i
Department of Physics, University of Illinois at Urbana-Champaign, 901 West Illinois Street, Urbana, IL 61801, USA
j
Dipartimento di Astronomia e Scienze dello Spazio, Universita! di Firenze, Largo E. Fermi 2, 51125 Firenze, Italy

Received 5 December 2001; received in revised form 10 February 2002; accepted 17 February 2002

Abstract

Low-power, ultra-high-vacuum compatible, non-contacting position sensors with nanometer resolution and
centimeter dynamic range have been developed, built and tested. They have been designed at Virgo as the sensors
for low-frequency modal damping of Seismic Attenuation System chains in Gravitational Wave interferometers and
sub-micron absolute mirror positioning. One type of these linear variable differential transformers (LVDTs) has been
designed to be also insensitive to transversal displacement thus allowing 3D movement of the sensor head while still
precisely reading its position along the sensitivity axis. A second LVDT geometry has been designed to measure the
displacement of the vertical seismic attenuation filters from their nominal position. Unlike the commercial LVDTs,
mostly based on magnetic cores, the LVDTs described here exert no force on the measured structure. r 2002 Elsevier
Science B.V. All rights reserved.

PACS: 04.80.Nn; 95.30.Sf; 95.55.Ym

Keywords: Gravitational wave interferometers; Seismic isolation; Viscous modal damping; Position sensors; LVDT

*Corresponding author. LIGO project, California Institute of Technology, 1200 E. California Blvd, Pasadena, CA 91125, USA.
E-mail address: htariq@ligo.caltech.edu (H. Tariq).

0168-9002/02/$ - see front matter r 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 1 6 8 - 9 0 0 2 ( 0 2 ) 0 0 8 0 2 - 1
H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576 571

1. Introduction complement them with a set of high-precision


position sensors. The specifications of these posi-
One of the most simple and effective Seismic tion sensors are as follows:
noise Attenuation Systems for Gravitational Wave * linearity within 1% over 710 mm, to match the
Detection interferometers [1–3] is a chain of
typical IPs dynamic range;
pendula [4–7] suspended from a very low-fre- * position resolution B10 nm r.m.s. to allow the
quency stage called Inverted Pendulum (IP). The
reduction of mirror residual motion;
IP [8,9] has the triple function of: * insensitivity to movements in the directions
* pre-filtering low-frequency seismic noise to orthogonal to the sensing axis, to allow for free
diminish the excitation of the rigid body motion movements of the IP on the horizontal plane
resonance of the attenuation chain; and not to re-inject seismic signals from the
* providing a suitable platform to damp the orthogonal directions in the feed back loop;
above-mentioned resonance. * force-free and contact-free, to avoid short
* providing the mechanical compliance to allow circuiting of the seismic excitation to the IP table;
precision positioning of the mirror suspended * only passive components inside the vacuum
from the chain using small forces. system, for reliability reasons; and
* fully Ultra-High-Vacuum (UHV) compatible,
The attenuation chains provide seismic attenua-
to prevent the contamination of the vacuum
tion in the frequency region of interest for
conditions.
gravitational wave detection and are capable of
delivering the required seismic attenuation factors Special linear variable differential transformer
(B1010) above 5–10 Hz, depending on the (LVDT) position transducers have been designed
pendula lengths. for the GW interferometer Virgo, to satisfy all the
The IP is composed of three flex joints, each above requirements. Also, all the electronic
supporting a leg. At the top, the three legs are circuitry is placed outside the vacuum enclosure
connected to a rigid table by means of small to satisfy the passive instrument requirement. The
flexures. The IP has two translational modes and LVDT signals are driven and read-out at the end
one torsional mode. of 20 or more meters of twisted shielded cable
A triplet of accelerometers [10,11] is mounted at without loss of resolution. These transducers turned
1201, in pinwheel geometry, at the periphery of the out to be so versatile that a modified, smaller
IP top table. It is designed to detect the accelera- version (referred in this paper as the ‘‘second type’’)
tion in the IP’s three degrees of freedom. The has been designed, which is being used in several
accelerometer signal is fed to actuators [12] to other locations of the GW interferometer seismic
damp the rigid body resonant modes of the chain attenuation systems. For example, LVDTs are
[13–15], and hence reduce the mirror r.m.s. being used as a sensor for all the Virgo super-
residual motion of the suspended mirror to a few attenuator filter vertical working points, as a sensor
tens of nanometers, in the frequency band between in the TAMA MGASF top filter, and as generic
10 mHz and 5 Hz. precision position sensor in mechanics R&D [16].
The accelerometers are inertial sensors, totally
insensitive to velocity, and, therefore, cannot
provide positioning of the IP table (and of the 2. The LVDT working principle
mirror suspended from it) with respect to a local
frame. Additionally, at low frequency (typically The LVDT is composed of three coils [17] whose
below 100 mHz, depending on the mechanics cross-section is shown in Fig. 1. The central
precision and ground stability) inertial sensors emitter coil, driven with a sinusoidal signal at a
can be more sensitive to tilts than to real frequency between 10 and 20 kHz, is mounted
accelerations. In order to take full advantage of between two larger receiver coils; the two receiver
the accelerometer performance it is necessary to coils are identical, counter-wound and connected
572 H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576

are, therefore, completely UHV compatible.


Although the coils were accurately spooled, the
lumped coil of this design makes it quite insensitive
to the spooling precision.
Unlike conventional magnetic core LVDTs [18],
this arrangement has no ferromagnetic compo-
nents and does not generate unwanted forces when
exposed to external magnetic fields.
The 1% linearity and transversal motion in-
sensitivity requirements are met by using two
simple circular coils wired in series and placed in a
Maxwell Pair (MP) configuration.
The MP configuration is related to the better-
known Helmoltz Coils (HC) configuration. The
HC coil is optimized to deliver a magnetic field
intensity as uniform as possible over a volume as
large as possible at the center of the pair. The MP
is designed to deliver a maximally uniform field
Fig. 1. Schematic view, sectioned along the symmetry axis, of a gradient over the same volume. The MP config-
LVDT position transducer. The three coils appear as the six uration is slightly less compact than the HC one; in
darkly shaded rectangles. The coil spools are made of either the MP, the coil separation s is
glass or Peek, the coil supports are made of slotted and pffiffiffi
mutually isolated aluminum parts to avoid eddy current loops. s¼r 3
No effort is made to make thermally compensated supports
because they are intended for use in a thermally stabilized where r is the coil radius
vacuum envelope. The MP pairs also differ from the HC because
the current flows in opposite directions in the two
receiver coils to generate the linear position signal
either in series or in parallel. The emitter is of the LVDT. If the two MP coils had parallel and
mounted on the IP table while the two receivers same direction field, the emitter coil would induce
are attached on a reference structure such that the a signal insensitive to position. Opposite current
emitter is exactly in the mid-point between the flow in the windings induces null signal at the
twin receiver coils; therefore, no net signal is central position while a remarkably linear signal is
induced. When the table displacements move the obtained along the displacement when the emitter
emitter coil in a direction, a sinusoidal signal coil moves along the MP symmetry axis. The
appears on the receiver coils. This signal has instrument remains insensitive to movements
amplitude roughly proportional to the displace- along other two axes. The MP has a reasonably
ment from the LVDT geometrical center. If the large sweet spot (B20% of the volume). As long as
coil is moved in the opposite direction the sign of the emitter coil remains in the sweet spot, good
the induced sinusoid is changed. linearity and insensitivity to transversal motion is
By a careful choice of the coil geometry, it was achieved. An example of the calculated linearity is
possible to get better than 1% linearity over 25 mm shown in Fig. 2.
of movement range with less than a percent
sensitivity to transversal movements. Also the
sophisticated design of the lock in amplifier circuit 3. LVDT measured performances
allowed for better than 10 nm r.m.s. position
resolution over the aforementioned range. The LVDT is expected to be linear well below a
The coils are made of Kapton-coated copper percent over a range of 30 mm and it is quite
wire, wound around glass or Peek supports, and insensitive to movements in the other two
H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576 573

0.2 15
percentage deviation from constant gradient

0 10

-0.2 5

readout (mm)
-0.4 0

-0.6 -5

-0.8 -10

-1 -15
-15 -10 -5 0 5 10 15 (a) -15 -10 -5 0 5 10 15
displacement (mm)
0.3
Fig. 2. Percentage deviation of the magnetic field gradient
along the sensitive axis of the LVDT.
residuals (mm)

0.2

0.1

-0.1
orthogonal directions. The expected linearity is
confirmed by the calibration of the actual instru- -0.2
-15 -10 -5 0 5 10 15
ments. The emitter coil is moved along the LVDT (b) displacement (mm)
axis with the help of a micrometric positioning
stage. A calibration constant of about 500 mV/mm Fig. 3. Calibration measurement of an IP LVDT (top) and its
residuals from a linear fit (bottom). The error bars (voltmeter
was obtained as shown in Fig. 3 (top), with
resolution) are within the marker size.
o71% deviation from the linearity illustrated in
Fig. 3 (bottom). The very symmetric and almost
sinusoidal residuals point either to geometrical
effects present due to the non-negligible thickness sensitivity of the LVDT to transversal motion is
of the three coils or to the imperfect spacing of the very small, up to a radial distance of 12 mm.
two receiver coils. In principle, the LVDT position resolution is
Alignment errors of the emitting coil with the only limited by the electronics noise. The electro-
receiving coils’ symmetry axis affect the LVDT nics noise is kept low by using a simple phase
sensitivity only to the second order and may locked signal detection technique. The signal sent
introduce some unwanted sensitivity to transversal to the emitter coil is also used to synchronize a low
motions. The insensitivity of the LVDT to the noise, double input linear gate. The amplified
transversal motion is shown in Fig. 4. Only the signal of the receiver coils is sent with opposite
trend of the measured distribution is important. polarity to two alternate gate inputs. For a more
The equations of magnetic field in vacuum require detailed discussion of the electronics see Ref. [19].
smoothness of the LVDT response, which is The actual circuit can be found in Ref. [20].
incompatible with the measured scatter. The In the IP configuration of the LVDTs, described
dispersion of the measurement points is due to a above, a resolution of 20 nm r.m.s. was achieved,
small wobble of the micrometric sled in the as shown in Fig. 5. Further electronics noise
magnetic field gradient direction. The flat trend improvements have led to LVDT resolution twice
of the measurement proves that the actual as good as what is shown in Fig. 5.
574 H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576

0.015 shown in Fig. 6 (the actual mechanical design can


be found for example in Ref. [21]), has been used
0.01 to generate diagnostics of passive filters of the
attenuation chains in the Virgo superattenuators
readout variation (mm)

[4] and position feedback signals in the top filter


0.005
both in the superattenuators and in the TAMA-
SAS [22]. They have been used also as position
0 sensor in R&D creep measurements [16,23].
These smaller LVDTs have a simpler, mono-
-0.005 lithic geometry and rely on a different geometry to
achieve good linearity. They are made with long,
-0.01
thin, uniformly wound cylindrical coils rather than
with short, lumped, loop-like coils. Good linearity
is achieved by choosing the right coil aspect ratios
-0.015
0 2 4 6 8 10 12 14 (Fig. 6). While mechanically much simpler and
transversal displacement (mm) self-supporting, a feature quite useful for small
Fig. 4. Sensitivity of a LVTD to transversal motion. The error
geometries, these coils are more sensitive to
bar shown is the resolution of the voltmeter used in this transversal motions and to the smoothness of the
measurement. coiling of the wire around the spool. Even small
imperfections of coiling are easily detectable as
local perturbations on the linearity of the re-
sponse.

Fig. 5. Spectral r.m.s. position resolution of the LVDT. The


apparent drop off of the residual noise above 50 Hz is generated
by the low-pass filters after the demodulation and it is not an
indicator of improved sensitivity.
Fig. 6. Geometry of a small LVDT. The emitter coil is 9 mm in
diameter and 18 mm long. The receiver coils are continuously
wound on a spool 24 mm in diameter, 28 mm long; the first
4. Other applications of LVDTs 14 mm are wound in one direction and the remaining 14 mm are
counter-wound; in the figure a gap is shown in the center to
show where the winding direction changes. The receiver coils
A second, smaller version of the LVDT (the are wound in a single layer without any space between each wire
actual mechanical design of this LVDT can be turn. In the emitter more than one layer is often wound to
found in Ref. [22]), with a different geometry, improve the sensitivity.
H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576 575

20 position resolution below nanometer level. This is


obtained at the price of a dynamic range reduced
to a few mm only. Position resolution at the
15
1012 m level has been obtained with LVDTs at
the price of proportionally, narrower dynamic
readout (mm)

ranges [24].
10

5. Problems due to electronics


5
The LVDT drivers are sensitive to variations of
amplitude of the excitation signal. Special care is
taken in designing the LVDT driver card in order
0
0 5 10 15 20
to generate a suitably stable excitation signal. The
Displacement (mm) internal frequency generators for the LVDT
(a)
drivers are perfectly adequate when absolute
0.06 positioning resolution is not an issue, like for
0.04 example in mode damping feedback circuits, where
residuals (mm)

0.02 drifts of 5mHz are irrelevant.


0 Thermal stabilization of the frequency generator
-0.02
and of the LVDT driver card is needed when
-0.04
absolute precision is necessary. For example,
thermal drifts have been found to be the dominat-
-0.06
0 5 10 15 20 ing noise source in long-term precision position
(b) Displacement (mm) measurements [16].
Fig. 7. Linearity performance of the small LVDT of Fig. 6. The
For very high stability use, the LVDT driver
error bars (voltmeter resolution) are within the marker size. The card is provided with an input for an external,
mechanical movement used has an error of 10 m. amplitude-stabilized oscillator.

The LVDTs are obviously very sensitive to the


coil spool stability. In a creep measurement, where 6. Conclusion
nylon spools were used, the creep signal was
actually dominated by the spool shrinking while High-resolution and high-linearity non-contact-
the spool itself progressively lost humidity in the ing position sensors have been developed to satisfy
dry thermally stabilized measurement chamber the requirement of GW detector passive Seismic
[23]. Ceramics or Peek coil spools have been Attenuation Systems. They are fully UHV compa-
successfully used as a replacement to avoid this tible. Only passive LVDT coils are mounted inside
problem and to achieve the desired signal stability the vacuum enclosure, thus guaranteeing a very
[16]. high level of reliability. Position resolutions below
While the linear dynamic range of these smaller the nanometer, within useful ranges of a few
LVDTs is similar to that of the bigger LVDT millimeter and percent linearity have been
versions (Fig. 7), they do not allow much lateral achieved. The low sensitivity of these LVDTs to
movement and are used only where the mechanical movements in the orthogonal degrees of freedom
configuration allows only linear motion. These from the symmetry axis is particularly well suited
small LVDTs are routinely used with higher gain for use in feedback loops. This precision position
(in part obtained with more coil layers in the detector, designed for position feedback in GW
emitter and in part electronically) to drive their Seismic Attenuation Systems, has proved useful
576 H. Tariq et al. / Nuclear Instruments and Methods in Physics Research A 489 (2002) 570–576

for many other non-contacting position measure- [14] G. Losurdo, Inertial control of the Virgo superattenuators,
ments. Proceedings of 1999 Amaldi Conference, Pasadena, CA.
[15] A. Takamori, et al., Mirror suspension system for the
TAMA-SAS, Proceedings of 2001 Amaldi Meeting, P-
010033-00-D available at http://admdbsrv.ligo.caltech.edu/
References dcc/; Classical and Quantum Gravity, 2002.
[16] M. Beccaria, et al., Nucl. Instr. and Meth. A 404 (1998)
[1] G. Ballardin, et al., Rev. Sci. Instr. 72 (9) (2001) 3643. 455.
[2] A. Bertolini, et al., New seismic attenuation system (SAS) [17] R. DeSalvo, et al., LIGO-II seismic attenuation system
for the advanced LIGO configurations (LIGO2), Proceed- (SAS), test tower LVDT (linear variable differential
ings of 1999 Amaldi Conference, Pasadena, CA. transducer). mechanical drawings, LIGO drawings, D-
[3] A. Bertolini, et al., Nucl. Instr. and Meth. A 461 (2001) 010243-00-R available at http://admdbsrv.ligo.caltech.edu/
300. dcc/.
[4] M. Beccaria, et al., Nucl. Instr. and Meth. A 394 (1997) [18] W. Kester, Practical design techniques for sensor signal
397. conditioning, Analog Devices Inc., 1999.
[5] S. Marka, et al., Anatomy of the TAMA SAS Seismic [19] F. Paoletti, L.V.D.T.: chi era costui? Elettronica, Flash 152
Attenuation System, Proceedings of 2001 Amaldi Meeting, (July/August) (1996) 37–46.
LIGO Report P-010034-00-D available at http://ad- [20] R. DeSalvo, et al., LIGO-II seismic attenuation system
mdbsrv.ligo.caltech.edu/dcc/; Classical and Quantum (SAS) test tower LVDT (linear variable differential
Gravity, 2002. transducer), electronics circuits diagrams, LIGO drawings,
[6] G. Cella, et al., Nucl. Instr. and Meth. A 487 (2002) 652. D-010248-00-R available at http://admdbsrv.ligo.caltech.
[7] J. Winterflood, et al., Using Buckling Euler Springs for edu/dcc/.
Vibration Isolation, Amaldi, 2001. [21] R. DeSalvo, et al., TOTEM, blade’s creep measuring
[8] G. Losurdo, et al., Rev. Sci. Instr. 70 (1999) 2507. facility, Mechanical drawings LIGO drawings, D-010252-
[9] R. De Salvo, et al., Nucl. Instr. and Meth. A 420 (1999) 00-R available at http://admdbsrv.ligo.caltech.edu/dcc/.
316. [22] R. DeSalvo, et al., TAMA seismic attenuation system
[10] A. Bertolini, High sensitivity accelerometers for gravity (SAS) tower, Mechanical drawings, LIGO drawings, D-
experiments, Doctoral Thesis, Universit!a di Pisa, June 010249-00-R, in: R. DeSalvo, et al., LIGO-II seismic
2001, LIGO preprint, P-010009-00-Z available at http:// attenuation system (SAS) test tower Filter-0, Mechanical
admdbsrv.ligo.caltech.edu/publications/default.htf?page= drawings, LIGO drawings, D-010241-00-R available at
theses. http://admdbsrv.ligo.caltech.edu/dcc/.
[11] S. Braccini, et al., Rev. Sci. Instr. 66 (3) (1995) 2572. [23] A. Bertolini, et al., The GAS blade creep measurements,
[12] C. Wang, et al., Nucl. Instr. and Meth. A 489 (2002) problems and some solutions, LIGO Report, T-10112-00-
563. R available at http://admdbsrv.ligo.caltech.edu/dcc/.
[13] G. Losurdo, et al., Rev. Sci. Instr. 72 (9) (2001) 3653. [24] S. Braccini, et al., Rev. Sci. Instrum. 66 (1995) 2672.

S-ar putea să vă placă și