Documente Academic
Documente Profesional
Documente Cultură
IV. METHODOLOGY
The design’s block diagram is shown in Figure 1.
Signal uController
Sensing PC interface
Conditioning Interface
Figure 4. Integrator Circuit with Drift regulator
Figure 1. Apnea Detector Block Diagram Any change that the capacitive sensor experienced has
little effect on the variation of the amplitude of the
The sensing and the signal conditioning blocks comprise integrated output. Therefore a need to amplify this
the Instrumentation stage of the project. This stage will change without amplifying the whole output is
be the bulk of the hardware of the project and is divided introduced in the circuit. A difference amplifier was the
into three main stages (Figure 2). first choice for this task, but the many advantages that an
instrumentation amplifier offers made us decide to use
the latter. The inputs that will be fed to the
Integrator Instrumentation Peak instrumentation amplifier will come from two
Circuits Amplifier Detector integrating op-amp circuits. The first integrator circuit
bears the capacitive sensor as the feedback while the
Figure 2. Hardware Block Diagram other integrator circuit has a fixed valued capacitor. The
second integrator will serve as the reference integrator
The basic configuration of an op-amp integrator circuit circuit which will be subtracted from the first integrator.
is shown in Figure 3. The expected output of the instrumentation amplifier
will then contain, at the very least, the small change or
variation that the sensor will introduce if any change in
humidity is sensed. The instrumentation amplifier should
ECE 117 3
also have a gain control in order to change the The operation of the peak detector is based on the idea of
magnitude of the small change that is captured. holding off the charge that a capacitor holds. The diode
will conduct positive half-cycles, charging the capacitor
to the waveform’s peak. When the input waveform falls
below the DC peak stored on the capacitor, the diode is
reversed biased, blocking current flow from capacitor
back to the source. Thus, the capacitor retains the peak
value even as the waveform drops to zero.
Where, R8=R7, R10=R9, R5=R6, Rpot = VR1. VR1 is a Figure 7. Schematic Full
10kΩ potentiometer place in between the difference op-
amps to serve as the gain of the overall instrumentation
amplifier. (Figure 5)
VII. CONCLUSION
The project was successful in detecting apnea and
monitoring breathing patterns using only a humidity
sensor. Apnea may not be detected successfully through
this project but can be prevented at a greater probability.
Hardware construction was not very tedious once the
required stages were planned accordingly. This project
also aided in proving the instrumentation amplifier from
the lecture class because of the accurate output of the
circuit. Software programming helped in giving a more
reliable and real time measurement. The actual sensor
also assisted the project because it easily fitted the
breathing mask generally available in the public (does
not need a specific breathing mask).
VIII. RECOMMENDATIONS
The project could have been more marketable if it was
contained in a mobile package wherein it only requires a
single voltage supply. To do this, the operation
amplifiers that were used could be replaced with single
supplied operational amplifiers like the LM358. An
alarm system can be interfaced in the project when a
warning needs to be noticed using a simple buzzer. A
more in depth data recording could also improve this
project.
IX. REFERENCES
[1] C-5Rice Drying Principles.pdf.
[2] HS1101 Capacitive Humidity Sensor Datasheet
[3] LM741 Datasheet
Dry Rice 1.1V 25%RH
Wet Salt 1.4V 75%RH
Human Breath 1.5V-1.6V 95%RH