Sunteți pe pagina 1din 84

EDITION/REVISION A/0

R NINGBO DAVID

SERVICE MANUAL

HKN-93 SERIES
INFANT RADIANT WARMER
NINGBO DAVID MEDICAL DEVICE CO., LTD.
ADD: NO.158 DAQING ROAD, SHIPU, NINGBO, ZHEJIANG PROVINCE, CHINA
POST CODE: 315731 FAX: 0086-574-65962111
MARKETING CENTER: NO. 11, ALLEY 722 SANGTIAN ROAD NINGBO, ZHEJIANG

POST CODE: 315040

TEL: 0086-574-87800008, 87800009 FAX: 0086-574-87801111


E-MAIL: sales@chinadavid.cn david@mail.nbptt.zj.cn
WEBSITE: http://www.nbdavid.com
EUROPEAN REPRESENTATIVE: SHANGHAI INTERNATIONAL HOLDING CORP. GMBH (EUROPE)
ADD: EIFFESTRASSE 80, 20537 HAMBURG, GERMANY
TEL: 0049-40-2513175 FAX: 0049-40-255726

EDITION/REVISION A/0
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

WARRANTY
The product being described in this manual is warranted against defects in materials or
workmanship for one year from the date of shipment, with the following exceptions.
1. All consumable and disposable products are guaranteed to be free from defects upon shipment
only.
2. Calibrations are considered normal maintenance and are not included in the 1-year warranty.
During the warranty period any defective parts other than those listed above will be replaced at no charge
to the customer. This warranty is rendered void and our company cannot be held liable for conditions
resultant there from if:
1. Damage to the unit is incurred as a result of mishandling.
2. The customer fails to maintain the unit in a proper manner.
3. The customer uses any parts, accessories, or fittings not specified or sold by our company.
4. Sale or service is performed by the non-certified service/dealer agency.
This warranty is in lieu of all other warranties, expressed or implied, and our company shall in no
event be liable for incidental or consequential damages including loss of use, property damage, or
personal injury resulting from breach of warranty.
Please test each piece of equipment prior to initial use and at least annually thereafter. To comply
with this standard, we recommend that you participate in our training program before selling it. This
service can be performed by certified technicians through our Product Service Group and authorized
dealers.

SERVICE
For optimal performance, product service should be performed only by qualified service personnel.
Technical Services representatives can be reached for fixations and are dispatched for required
maintenance by calling 0086-574-87800008, 87800009. Customers outside China should contact their
local factory-authorized distributor for service.

COMPLEMENTARY NOTICE
This manual provides instructions for installation and maintenance of Infant Radiant Warmer Series
HKN-93A, HKN-93B, HKN-93. And HKN-93A is the most complex mode among warmer HKN-93series.
So the contect of this manual is based on HKN-93A, and the difference between these models is
explained in this manual.
Since Ningbo David, conducts a continuous product improvement program, circuit and component
improvements are sometimes incorporated into equipment before they can be incorporated into the
printed manuals. This manual contains proprietary information, repairs and authorized modifications
should be performed only by qualified service personnel to maintain your warranty and to avoid creating
safety hazards. We cannot assume responsibility for any conditions affecting the proper operation of this
equipment which may result from unauthorized repair modification. Therefore, some parts used in your
equipment may be different than those which appear in the parts list of this manual. This sometimes
occurs due to difficulty in parts procurement, but does not alter the function of the equipment
The device, accessories and the packaging have to be disposed of waste correctly at the end of the
usage. Please follow Local Ordinances or Regulations for disposal.
i
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE OF DEFINITIONS AND SYMBOLS

TECHNICAL DEFINITIONS
SKIN TEMPERATURE SENSOR: A sensing device including the link with the equipment intended to
measure the infant’s baby temperature.
CONTROL TEMPERATURE: The preventative control temperature to the temperature control.
PRE-WARM MODE: A warm mode which can keep the mattress at a properly temperature. The heater
output will work automatically according to the set programme in this mode.
MANUAL MODE: A operation mode of in which the heater output is either at a fixed level or a
proportion of its maximum output set by the operator.
BABY MODE: A mode of operation in which the power output varies automatically in response to the
temperature of the baby, to achieve a temperature close to a value set by the operator.
STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature, measured
at the center of the TEST DEVICE positioned on the mid point of the EQUIPMENT mattress, does not
vary by more than 1℃ over a period of 1 hour.
TEMPERATURE ALARM CHECKOUT STATE: The difference between the indicated temperature and
control temperature is within ±0.5℃ and such state lasts for over 10 minutes. When checkout the
temperature alarm function, operation should be enter this state.
APGAR TIMER: It offers the functions of three periods alarming indication: 1min, 5min, and 10min for
clinical treatment.
TOTAL IRRADIANCE FOR BILIRUBIN Ebi:Irradiance equal to the evaluated irradiance in the range
between 400 nm and 550nm.
UNIFORMITY G2 OF THE TOTAL IRRADIANCE FOR BILIRUBIN: Ratio of the lowest TOTAL
IRRADIANCE FOR BILIRUBIN Ebi min to the highest TOTAL IRRADIANCE FOR BILIRUBIN Ebi max on the
EFFECTIVE SURFACE AREA in section 5.
LIFETIME OF LIGHT SOURCE: Time after the average total irradiance for bilirubin Ebi is attenuated by
25%

NOTE, IMPORTANT, CAUTION AND WARNING


NOTE: A note is inserted in text to point out procedures or conditions, which may otherwise be
misinterpreted or overlooked. A note may also be used to clarify apparently contradictory or
confusing situations.

IMPORTANT: Similar to a Note but be used where greater emphasis is required.

CAUTION: A caution is inserted in text to call attention of a procedure which, It not followed exactly, can
lead to damage or destruction of the equipment.

WARNING: A warning is inserted in text to call attention to dangerous or hazardous conditions inherent
to the operation, cleaning, and maintenance of the equipment which may result in personal
Injury or death of the operator or patient.

ii
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SYMBOLS

Attention, Consult Accompanying Documents.

Input Power Type F Fuse 5AH/250V

Auxiliary Mains Output, MAX:0.3A Type F Fuse 6.3AH/250V

Type T Fuse 800mAL/250V Type T Fuse 315mAL/250V

Type T Fuse160mAL/250V Heater Power Indicator

Type T Fuse 1.25AL/250V Type T Fuse 400mAL/250V

Type T Fuse 1.6AL/250V Type T Fuse 3.15AL/250V

Height adjustment down key Height adjustment upper key

Data Communication Connector Load Symbol

Off (Power:Disconnection to the


On (Power:Connection to the mains)
mains)

On (only for a part of equipment) Off (only for a part of equipment)

Class I Equipment Type BF Applied Part

Protective Earth (ground) Earth (ground)

Danger! High voltage! Timer Key

Set Up Key Set Down Key

Keypad Lock Key Calibration Key

Silence/Reset Key Mode Select Key

Eye mask mark/ Switch for


LED mark *
phototherapy unit

LED mark * CE Marking

Date of Manufacture Serial Number

NOTE: The label marked with * is usually sticked on the surface of phototherapy

iii
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SEASONAL SAFETY CHECK


1. Please clean the plug of power cord at least once a year. Too much dust on plug may cause the fire.
2. The following safety checks should be performed at least every 12 months by a qualified person who
has adequate training, knowledge, and practical experience to perform these tests. The data should be
recorded in an equipment log. If the device is not functioning properly or fails any of the above tests, the
device has to be repaired.
1 . Inspect the equipment and accessories for mechanical and functional damage.

2 . Inspect the safety relevant labels for legibility.

3 . Inspect the fuse to verify compliance with rated current and breaking characteristics.

4 . Verify that the device functions properly as described in the instructions for use.

5 . Test the protection earth resistance according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit

0.1Ω.
6 . Test the earth leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: NC

500μA, SFC: 1000μA.
7 . Test the enclosure leakage current according to IEC 60601-1:1988 + A1:1991 + A2:1995: Limit:

NC 100μA, SFC: 500μA.
8 . Test the patient leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: for

a.c.: 100μA (BF), for d.c.: 10μA (BF).
9 . Test the patient leakage current under single fault condition with mains voltage on the applied

part according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: for a.c.: 500μA (BF), for d.c.: 50μA (BF).
10 . Test the patient auxiliary leakage current according IEC 60601-1:1988 + A1:1991+ A2:1995:

Limit: NC for a.c.: 100μA (BF), for d.c.: 10μA (BF).SFC 500μA (BF), for d.c.: 50μA (BF).

iv
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE OF CONTENTS
SECTION PAGE
1. GENERAL INTRODUCTION………………………………………………………………………………1-1
1.1 INTRODUCTION……………………………………………………………………………………………1-1
1.2 DESCRIPTION…………………………………………………………………………………...……… 1-1
1.3 COMPOSITION OF PRODUCTS………………………………………………………………...……… 1-1

2. INSTALLATION……………………………………………………………………………………………..2-1

2.1 UNPACKING…………………………………………………………………………………...………….2-1

2.2 INSTALLATION………………………………………………………………...………………….….…….2-1

2.3 OPERATION CHECKOUT PROCEDURE…………………………………...………………….….…….2-4

3. TECHNICAL INFORMATION………………………………………………………………………….…3-1
3.1 SPECIFICATION……………………………………………………………………………………………3-1
3.2 WORKING PRINCIPLE………………………………………………………………………...………….3-3

4. PREVENTATIVE MAINTENANCE………………………………………………………………………...4-1
4.1 GENERAL………………………………………………………………………………………………....4-1
4.2 CLEANING………………………………………………………………………………………...………...4-1
4.3 STERILIZATION……………………………………………………………………………...………...4-3
4.4 MAINTENANCE…………………………………………………………………………………………….4-3

5. SERVICE……………………………………………………………………………………………………...5-1
5.1 GENERAL………………………………………………………………………………………………….5-3
5.2 REPLACEMENT OF CHARGING BATTERY/FUSE/QUARTZ HEATER/
ILLUMINATION LIGHT/ PHOTOTHERAPY UNIT………………………………………………………5-1
5.3 SYSTEM SETTING PROCEDURES……………………………………………………………………5-4
5.4 TROUBLESHOOTING PROCEDURES…………………………………………………………………5-7

6. REPLACEMENT OF SPARE PARTS………………………………………………………………………6-1


6.1 GENERAL ……………………………………………………………………………………………………6-1

7. DIAGRAMS……………………………………………………………………………………………………7-1
7.1 GENERAL……………………………………………………………………………………………7-1

v
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 1
GENERAL INTRODUCTION

1.1 INTRODUCTION
This manual provides instructions for installation and maintenance of Infant Radiant Warmer Series
93.
This manual provides the guidance for the qualified service personnel. As for the instruction, please
refer to the user's manual.

1.2 DESCRIPTION
Based on HKN-93, the Infant Radiant Warmer HKN-93B added with the Phototherapy; Infant Radiant
Warmer HKN-93A added the Breath Resuscitation Bag and Infant Head Fixing Unit based on HKN-93B.
The Infant Radiant Warmer Series 93 adopts 3 kinds of control mode: Pre-warm mode, Manual mode
and Baby mode. Once the device starts working, it can enter into the Pre-warm mode automatically
without pressing any key.
The setting temperature and baby temperature can be indicated separately.
Except for the listed parts in 1.3, the warmer also has the quartz halogen lamp, X-ray tray and 4pcs
casters. ( Two of them with brakes).

1.3 COMPOSITION OF PRODUCTS


The following diagram shows the main parts of the infant radiant warmer (Based on the HKN—93A):
Warmer Module I.V. Pole

Phototherapy unit Phototherapy unit

Shelf Temperature controller

Breath Resuscitation Bag


Infant Head Fixing Unit

Bassinet

Stand

1-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

DESCRIPTION OF PART EXPLANATION

This main part is composed of heater, reflective cover and so on, it is


intended to provide the radiation for electromagnetic spectrum infrared
Heater Head
range with weight of 16Kg, and its angle can be adjusted horizontal in two
ways. The lifetime of heater is 2000hrs.

It is intended for treating the bilirubin of patient.


Phototherapy The light source of this device is LED, and its lifetime is 5000hrs.
(For HKN-93B. 93A) Please wear the eye mask for patient when this part works, as for the
detail, please refer to the user’s manual.

A kind of bearing part, which is used for hanging the infusion bottle.
I.V. Pole
Max. Load: 2Kg

The core part with 3 temperature control mode: pre-warm mode, manual
mode, baby mode, APGAR timer, timing function, and it is used for
Temperature Controller
automatic controlling of infrared radiation heat output, please refer to the
user's manual.
A kind of bearing part, which is used for putting some small objects.
Shelf
Max. Load: 2 Kg

It is used for place the infant. The bassinet can be tilted. Max. Load: 10
Kg. This bassinet is equipped with 4pcs of panels to avoid the patient
Bassinet
falling off.
Size of mattress: 760mm × 595mm

A part which can support the main body of warmer. We have two kinds of
stand: fixed stand and vertical height adjustment stand. Usually we equip
Stand the former one, and the second one is optional part. If equip the former
one the distance from the bassinet to the ground is 890mm; if equip the
vertical height adjustment one, the distance is 890mm~1030mm.

Breath Resuscitation Bag It is intended to rescuing the neonate when the suffocation happens or
(For HKN-93A) breath is blocked. Operate according to it’s manual.

Infant Head Fixing Unit It is intended for fix neonate’s head during treatment. Operate according
(For HKN-93A) to it’s manual.

NOTE: 1.Size of whole unit with fixed stand after installation: W730 × H1795 × D1100mm
2. Size of whole unit with vertical height adjustment stand: W730×H1795×D1100mm ~
W730×H1935×D1100mm
3. Weight of whole unit with fixed stand: 105Kg
4. Weight of whole unit with vertical height adjustment stand: 140Kg

1-2

W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 2
INSTALLATION

2.1 UNPACKING
Generally, the WARMER is usually packed to one carton. When taking out the equipment from the
cartons, take care not to damage the spare parts of the WARMER.

FIGURE 2.1

2.2 INSTALLATION
At least two professionals are required to do the installation of the WARMER with spanners and
screwdriver.
Installation step:
A. INSTALL THE UPPER COLUMN GROUPWARE (UPPER COLUMN + WARMER MODULE)
a. The installation of the warmer with VHA.
① Unscrew the fixing bolt on the stand cover with screwdriver and remove the cover, and pull out
the power cord of warmer module.
② Unscrew the Hexagon bolt M8X20, Spring washerΦ8, Flat washerΦ8 and external teeth
serrated lock washerΦ8 in figure 2.1 with spanner.

2-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

③ Unscrew the pan head bolt M4X10, Spring washerΦ4, Flat washerΦ4 and external teeth
serrated lock washerΦ4 in figure 2.1 with screwdriver to disassemble the back cover of upper column.
④ Insert the upper column into the stand column, and tighten it with the previous unscrewed
Hexagon bolt M8X20, Spring washerΦ8, Flat washerΦ8 and external teeth serrated lock washerΦ8;
insert the power cord of warmer module into the socket in figure 4.2 long with the path inside the upper
column.
⑤ Assemble the stand cover into its original position with the unscrewed bolts and tighten the
cover with the unscrewed pan head bolt M4X10, Spring washerΦ4, Flat washerΦ4 and external teeth
serrated lock washerΦ4.
b. The installation of the warmer with fixed stand
① Unscrew the Hexagon bolt M8X20, Spring washerΦ8, and Flat washerΦ8 on the upper
column.
② Assemble the decoration slice into the column of stand as figure 2.2 indicates.

Cummerbund

Column

FIGURE 2.2
③ Insert the upper column groupware into the stand’s column as figure 2.3 indicates, and tighten it
with the previous unscrewed Hexagon bolt M8X20, Spring washerΦ8, and Flat washerΦ8.

Upper column

FIGURE 2.3
IMPORTANT: Please make sure that the upper column and the stand are vertical, or else, it will
affect the temperature uniformity on the mattress.

2-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

B. INSTALL THE PANEL


Insert the panel into the Fixed seat as the arrow indicates in figure 2.4.1, and turn it upward to make
it vertical and press it as the figure 2.4.2 indicates.

Panel

Fixed Patch For Panel

FIGURE 2.4.1 FIGURE 2.4.2


NOTE: 1.Total 4 pieces panels, the panel with groove is the back panel.
2. Make the bassinet tilt when installing the back panel.
C. INSTALL SHELF AND I.V. POLE
① Unscrew the pan head bolt M4X12, Spring washerΦ4, Flat washerΦ4 on upper column with
screwdriver.
② Put the shelf on the upper column as figure 2.5.1 indicates, and tighten it with the previous
unscrewed pan head bolt M4X10, Spring washerΦ4, Flat washerΦ4.
③ Install the I.V. Pole on the upper column as figure 2.5.2 indicates and fix it with the tightening
clamp.

Tightening Clamp

Shelf

FIGURE 2.5.1 FIGURE 2.5.2


D. INSTALL THE PHOTOTHERAPY UNIT (FOR HKN-93B, HKN-93A)
① In figure 2.6.1, connect the plug from phototherapy unit and the one from the Heater Head.
② In figure 2.6.2, make the 3pcs round holes of calabash shape aim at the 3pcs bolts on the top of
Heater Head, and tighten it with the screwdriver.

Round holes of calabash shape

FIGURE 2.6.1 FIGURE 2.6.2

2-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

E. INSTALL BREATH RESUSCITATION BAG (FOR HKN-93A)


① In figure 2.7, install the hook of Breath Resuscitation Bag onto the column, and fix it with the
tightening clamp.
② Put Breath Resuscitation Bag on the hook.

FIGURE 2.7
F. INSERT THE POWER CORD
Insert the power cord into its relevant socket. Check the WARMER according to the instruction in
section 2.3.

2.3 OPERATION CHECKOUT PROCEDURE


Please follow the below operation checkout procedure before use.
A. CHECK CASTERS
Check whether the caster can drop when lifting the radiant warmer 2cm above the ground. The drop
of the caster will cause danger during transporting.
IMPORTANT: At least two persons are required, one lifts the warmers, and the other has to check
the casters. It needs Min. 140Kg to light the whole unit. Please support the stand
with block to avoid the warmer inclining if the casters drop.
B. CHECK POWER FAILURE ALARM
When the general power supply and the switch of controller are turned off, pull off the power cord of
whole unit, the Power failure alarm should active, and the “Power” light will flash with continuous
audible sound.
This operation is used for checking if the power failure is normal or not. Insert the power cord again
or switch off the controller to silence alarming after finishing checking.
IMPORTANT: Please make sure that the internal charging battery is full before use it. If it is not
full, it may cause the power failure without any alarming. If the battery is full, and
there is no hint when disconnect the power cord, please refer to the qualified
service personnel.

2-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

C. CHECK THE FUNCTION OF THE STAND (if applicable)


Step on the “upper” and “down” button on the bottom of the Stand after the power is supplied and
the adjustment to the height of the Stand should be actualized. If not, please refer to the qualified service
personnel.
NOTE: The vertical height adjustment Stand should be restarted after 30s.
D. START-UP THE CONTROLLER
Turn on the switch of the main power supply and temperature controller, the temperature controller
performs a long period sound “ringing” .All indicators are on, at this time, the device can do self-test
automatically for 5s. After then, the setting temperature indicator will display “--.-”, at the same time, the
baby temperature indicator displays the baby temperature, and the timer indicates the current time, and
the temperature controller can enter into the pre-warm mode automatically, and the pre-warm light will
be on. If one of them does not work, or the indication is incomplete, please refer to the qualified service
personnel.
E. CHECK WORKING MODE
① Place the skin sensor on the central of mattress, and record the baby temperature in the cold

stare, and then start the warmer to make the device enter into the pre-warm mode automatically for

30min, and please read the baby temperature to compare the indicated temperature after pre-warm and

the one in the cold state, it should not less than 4℃. If not, please replace the heater.

② please select the manual control mode, the system default of heat output proportion is 30%,

there is 3grades in the heat indicator, press Up or Down key to adjust the value within range of

0%-100%.

Please set the heat output proportion >30%. There is alarm every 15min, “CHECK” alarm light

flashes, and the set temperature indicator displays alarm code E0.9, press silence/reset key to silence

alarming. The device still works according to the current mode and heat output proportion. If do not press

silence/reset key when “CHECK” alarm actives, the device will decrease the heat output proportion to

30% in 1min. If not, please refer to the qualified service personnel.

③ In Baby mode, pull off the skin sensor, the audible and visual alarm will occur, the alarm light of

sensor and the set temperature indicator will show alarm code E0.1. Press silence/reset key to silence

alarming for 4min. connect again the skin sensor, the device will come back to the normal working

condition.
F. CHECK SETTING ALARM
In Baby mode, put the skin sensor into the water cup at temperature lower 3.5℃ than the set
temperature for 2min, the setting temperature alarm will active. The “Set” light should flash with
continuous audible sound, and the set temperature indicator will show alarm code E0.7. Press
silence/reset key to silence alarming for 4min.
2-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Set the temperature at 36℃, after the device enters into the temperature alarming checkout state,

put the skin sensor into the water cup at temperature 35.3℃±0.1℃ for 3min, the setting temperature

alarm will active. The “Set” light should flash with continuous audible sound, and the set temperature

indicator will show alarm code E0.8. Press silence/reset key to silence alarming for 4min.

G. CHECK DEVIATION ALARM

In baby mode, set the temperature at 35.0 ºC. Enter temperature alarm checkout state, put the

skin sensor into the water cup at 37 ºC. When the indicated temperature reaches 36.0 ºC, the upper

deviation alarm will active. The “H/L” light should flash with continuous audible sound, and the set

temperature indicator should show alarm code E0.5; Set the temperature at 35.0℃, after the device

enters into the temperature alarming checkout state, put the skin sensor into the water cup at

temperature 33℃, when the indicated temperature decreases to 34.0 ºC, the lower deviation alarm

should active. The “H/L” light should flash with continuous audible sound, and the set temperature

indicator should show alarm code E0.6. Press silence/reset key to silence alarming for 4min.

NOTE: If the system can not enter into the temperature alarm checkout state or the baby

temperature does vary ±1℃ than the setting temperature, the deviation alarm can

not occur.

H. CHECK OVER-TEMPERATURE ALARM

In baby mode, put the skin sensor into the water cups at 39.5℃±0.5℃, the over-temperature alarm

should active. The “Over” light should flash with continuous audible sound, and the set temperature indicator
should show alarm code E0.4. Press silence/reset key to silence alarming for 4min. reduce the baby
temperature and press silence/reset key to silence alarming.
I. CHECK APGAR TIMER

When the temperature controller is under the normal working state, press the APGAR key, the timer

start to indicate from 0s, when the timer indicates 50”~1’, 4’50”~5’, 9’50”~10’, it sounds “du…”, at the

same time, the APGAR timer is flashing.

J. CHECK THE TIMER

Under the normal working condition, press to indicate the current running time, If the indicated time

is not accurate, please reset, and the specific setting method is as follows:

Press the timer key, at the same time, turn on the controller for 3min, loosen the key when the

indicated setting interface appears.


2-6
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

At this time, Set temperature indicator shows the setting code, baby temperature indicator will
indicate the relevant value (as for setting code, setting item and setting range, please refer to the
following table ), the timer indicator will indicates P002(it means it has entered into the setting interface).
Operator can choose the setting items by pressing the timer key or APGAR key, and press the up and
down key to choose the data setting. After then, press the set key to store the setting value, or you can
press the silent/reset key to store and exit after finishing all setting procedure. If turn off the power supply
without pressing the silent/reset key, and this setting is invalid, and system will store the previous setting
value.
TIME SET CODE

SETTING CODE INDICATED ITEM SET RANGE

001 The former two figure of year 19 ~20

002 The latter two figure of year 00 ~99

003 month 01 ~12

004 date 01 ~31

005 week 01~07(01means Sunday)

006 hour 00~23(24hrs)

007 minute 00~59

008 second 00~59

K. CHECK THE PHOTOTHERAPY AND THE TIMER


In normal working condition, press the timer mode key and choose the accumulative time mode of
phototherapy, the light is on. Press the switch of Phototherapy to active the unit, and the timer indicator
will indicates its timing from zero. Pressing the power key of the phototherapy again to turn it off, and the
timer indicator remains the time recorded and stops timing, which will keep on timing on the basis of the
previous time recorded when restarting the phototherapy. If not, please refer to the qualified personnel to
repair the phototherapy unit.
L. CHECK BASSINET TILT MACHANISM
Pull the pull handle as the arrow indicates to adjust the bassinet tilt.

2-7
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

M. CHECK THE ADJUSTMENT FUNCTION OF ANGLE OF WARMER MODULE


Pull the handle for lock downward to rotate the heater module at horizon angle.

Handle for lock

N. CHECK ILLUMINATION LIGHT


Turn on the power switch of illumination light, and the light will be on.

O.CHECK THE PANEL


Push and pull the panel and check if the panels are firm or not and whether they are broken.

P. CHECK THE X-RAY TRAY


Push and pull the X-Ray tray and check if it is agile and handiness, or whether it is broken.

Q. CHECK I.V. POLE


Check if the I.V. Pole stand is tight or not.

R. END

2-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 3
TECHNICAL INFORMATION
3.1 SPECIFICATION
Specifications for the Radiant Warmer are provided in table 3.1. All specifications are subject to
change without notice. Changing of the air flow indoor or the tilt of the bassinet, or using heat-producing
devices near the Radiant Warmer are all possibly effect on the correlation among the temperature
uniformity of the mattress, the indicated temperature or the baby temperature.
TABLE 3.1 SPECIFICATIONS

Power Requirements………………………………………………………………………AC220V-230V/50Hz
Maximum Power Output…………………………………………………………………750W(for fixed stand)
Maximum Power Output………………………………………800W(for vertical height adjustment stand)
Maximum Heater Power Output………………………………………………………580W/240V
Auxiliary Mains Power Output……………………………AC220V-230V/50Hz, MAX. CURRENT 0.3A
Heater power display…………………………………………… 0 to 100%, adjustable in 10% increments
Protective earthed resistance…………………………………………………………………………<0.1Ω
Earth leakage current …………………………………………………………………………N.C.: <0.5mA
S.F.C.: <1mA
Patient leakage current …………………………………………………………………………N.C.: <0.1mA
S.F.C.: <0.5mA
Patient Leakage Current (MAINS VOLTAGE ON THE APPLIED PART) …………………S.F.C: <5mA

This equipment belongs to Class I, Type BF applied part, normal device(IP20)continuously operated.

TEMPERATURE CONTROL RANGE AND ITS SPECIFCIATION


Temperature control range……………………………………………………………………34.5℃~37.5℃
Baby Temperature display range………………………………………………………………5℃~65℃
Deviation between baby temperature and control temperature…………………………………≤0.5℃
Accuracy of skin temperature sensor………………………………………………………………≤0.3℃
Temperature uniformity of mattress……………………………………………………………………≤2℃

ALARM (See table 3.2)


Power alarm or Probe failure or disconnection of skin sensor………………………Alarm code E0.1
The probe for over-temp failure……………………………………………………………Alarm code E0.2
Deviation of two probes failure………………………………………………………………Alarm code E0.3
Over-temp alarm for skin sensor……………………………………………………………Alarm code E0.4
Upper deviation alarm…………………………………………………………………………Alarm code E0.5
Lower deviation alarm…………………………………………………………………………Alarm code E0.6
Wrong temperature set alarm in baby mode………………………………………………Alarm code E0.7

W3 3-1
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE 1.1 SPECIFICATIONS (continued)

The sensor position alarm in baby mode………………………………………………Alarm code E0.8


The tip every min in manual mode……………………………………………………Alarm code E0.9
Internal system failure…………………………………………………….……..Alarm code H0.1~H1.1

ENVIRONMENT TEMPERATURE
Operating Range………………………………………………….……..……………………+18℃~+30℃
Storage Range……..………………………………………………….……………………… -40℃~+55℃

ENVIRONMENT HUMIDITY
Operating Range…………………………………………………………………30%RH~75%RH
Storage Range………………………………………………………………………………………≤93%RH

ATMOSPHERE PRESSURE
Shipment and store atmospheric pressure range…………………………………500hPa~1060hPa
Working atmospheric pressure range…………………………………………………700hPa~1060hPa

AIR FLOW RATE


AMBIENT AIR MOVEMENT RATE………………………………………………………………… <0.3M/S

OTHER SPECIFICATION
Working noise…………………………………………Ambient noise≤40dB(A), working noise≤50dB(A)
APGAR Timer…………………………………………………………………………………Tones at 50” ~1’,
4’50” ~5’, 9’50” ~10’
Maximum bilirubin radiation on the Mattress effective range…………………………≥800uW/cm2
Bilirubin uniformity of radiating on the mattress effective range…………………………………>0.4
Average value general bilirubin radiation of on the mattress effective range…………600 W/cm2
ALARM NOISE…………………………………………………………………………… ≤80dB(A) on mattress

SILENCE/RESET KEY
Press to silence the alarming, press twice to reset alarming.
Silence………………………………………………………………………………silence alarming for 4min,
alarming will silence if there is any other alarming actives
Reset……………………………………………………silence alarming to come back to the setting state
The lifetime ……………………………………………………………………………………………8 years

(Lifetime means the period from sell-by date to the date of discarding as useless.)

FACTORY DEFAULT SETTING (ORIGINAL STATE)


Heater mode…………………………………………………………………………………Pre-warm mode
Heater output proportion under Manual mode ……………………………………………………… 30%
Set temperature under baby mode……………………………………………………………………36℃
Time Mode…………………………………………………………………………………………Clock mode

3-2 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE 1.1 SPECIFICATIONS (continued)

Spectral energy distribution Spectral distribution

In the range of 320nm to 550nm, Schematic diagram of Ebi full spectrum radiation
1.2
1
0.8
s (λ ) 0.6
0.4
0.2
0
390 430 470 510 550
λ (nm)
Testing device integralE bicalibration trend

3.2 WORKING PRINCIPLE


3.2.1 GENERAL
This section covers the function and the working principle, as for the figure for control system,
please refer to figure 3.2 and 3.3.
Surface temperature on mattress can be reached during the radiant system in figure 3.1, the radiant
heater outputs in the way of fixed proportion to make the body temperature of patient resume; in Baby
mode, the device will adjust the heat output proportion automatically by comparing the baby temperature
and the set temperature to keep the heat balance.

NOTE: During the process of getting the radiation and heat, the losing heat in convection,
evaporation, radiation, conduction will affect the heat balance. Therefore, to make the
baby decrease the dissipation, it is necessary to use the device in the environment of no
rapid air flowing ,and cover waterproof velum (polyethylene velum) the naked skin or to
use the waterproof net to increase the humidity around the skin to decrease the water
evaporation.

FIGURE 3.1 RADIATION PRINCIPLES

W3 3-3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Keypad control

Temperature indication

Function alarm

Output port
Heat control output
A
MIAN
MCU
A/D Temperature control probe of skin sensor

Communication connector

A/D Over-temp checking probe of skin sensor

Auxiliary
Output Port
MCU Heat control output B

FIGURE 3.2 CONTROL BLOCK DIAGRAM

3-4 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Start

Initialization
Timer 2 interrupt

System self-test
Timing

Data display
Return

YES
Keypad checking

Keypad handling
NO

Data handling

INTO interrupt

YES ADC data


Trouble checking

Return
NO
Alarm
Heating control identification
Series port
interruption

Charging control
Data transfer

Data display
Return

FIGURE 3.3 PROGRAMMED BLOCK DIAGRAM

W3 3-5
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

3.2.2 Temperature Control

Incubator temperature is adjusted through the compurgation between the baby temperature and the

set temperature. The Infant Radiant Warmer adopts 3 kinds of control mode: Pre-warm mode, Manual

mode and Baby mode, and the different mode had different way of heating. The user can choose the

wanted control mode by pressing the keys on the controller.

Pre-warm mode: system will output the heat according to the programme, after a while, the system

will output the heat at 30% to maintain the temperature on the mattress until the mode changes.

Manual mode: system will output the fixed heat according to the heat output proportion to make the

temperature on the mattress reach the wanted value so as to make the body temperature of patient

resume. The system default of heat output proportion is 30%.

Baby mode: System will adjust through the heat output proportion according to the baby

temperature and the set temperature to keep the heat balance of patient.

3.2.3 ALARM
Alarm is used for monitoring the control system state. The system will check the state by the alarm
checkout procedure. When the checked data is abnormal, it will alarm. Table 3.2 listed alarm condition.

TABLE 3.1 ALARM

ALARM MESSAGE DESCRIPTION

Power Failure Alarm There is no power supply.

Short-circuit or open-circuit occurs on the temperature control probe


code inside of skin temperature sensor, or connection with warmer is not
E0.1 well, the heater will stop working, please press the Silence/reset key to
silence alarming for 4min, and then it can reset if there is no failure.
Short-circuit or open-circuit occurs on the over-temp probe inside of
Sensor Alarm code skin temperature sensor, the heater will stop working, please press the
indicator is on E0.2 Silence/reset key to silence alarming for 4min, and then it can reset if
there is no failure.
Deviation of baby temperature of two probes inside of skin temperature
code sensor is more than 0.8℃, the heater will stop working, please press
E0.3 the Silence/reset key to silence alarming for 4min, and then it can reset
if there is no failure.
In the Baby mode, when the temperature measured by skin
Over-Temp
code temperature sensor is >38.5℃, the heater will stop working , please
Alarm indicator is
E0.4 press the Silence/reset key to silence alarming for 4min, and then it can
on
reset if there is no failure.

When the indicated baby temperature is more than 1℃(set value), the
Deviation Alarm code
heater will stop working , please press the Silence/reset key to silence
indicator is on E0.5
alarming for 4min, and then it can reset if there is no failure.

3-6 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE 3.1 ALARM (CONTINUE)

ALARM MESSAGE DESCRIPTION

When the indicated baby temperature is lower than 1℃ (set value), the
Deviation Alarm code
heater will stop working , please press the Silence/reset key to silence
indicator is on E0.6
alarming for 4min, and then it can reset if there is no failure.
When the baby temperature is lower 3.5℃ than the setting temperature,
code
the heater will stop working , please press the Silence/reset key to silence
E0.7
alarming for 4min, and then it can reset if there is no failure.
Setting Alarm When the device has entered into the steady temperature condition
indicator is on because of the accidental or unreasonable out of the steady temperature
code condition, and it will not enter into this condition in 3min without the
E0.8 deviation alarm, the heater will stop working , please press the
Silence/reset key to silence alarming for 4min, and then it can reset if there
is no failure.
In the Manual mode, the “CHECK” alarm with audible and visual will active
Checking Alarm code every 15min, the heater will keep working , please press the Silence/reset
indicator is on E0.9 key to silence alarming for 4min; the heat output proportion will be limited
to 30% without pressing the Silence/reset key.

code RAM inside of main MCU fails, the heater will stop working during
H0.1 alarming, it is invalid to press Silence/reset key.
code ROM inside of main MCU fails, the heater will stop working during
H0.2 alarming, it is invalid to press Silence/reset key.
code Communication of main and auxiliary MCU fails, the heater will stop
H0.3 working during alarming, it is invalid to press Silence/reset key.
code E2ROM inside main y MCU fails, the heater will stop working during
H0.4 alarming, it is invalid to press Silence/reset key.
code RAM inside of auxiliary MCU fails, the heater will stop working during
H0.5 alarming, it is invalid to press Silence/reset key.
code ROM inside of auxiliary MCU fails, the heater will stop working during
H0.6 alarming, it is invalid to press Silence/reset key.
The alarming
ADC1 failure(e.g., main system TLV2544, including wrong sample, chip
indicator is on code
damage, no response and so on), the heater will stop working during
H0.7
alarming, it is invalid to press Silence/reset key.
ADC2 failure(e.g., main system TLV2544, including wrong sample, chip
code
damage, no response and so on), the heater will stop working during
H0.8
alarming, it is invalid to press Silence/reset key.

code SRAM failure(including system cord failure), the heater will stop working
H0.9 during alarming, it is invalid to press Silence/reset key.

code The timer RTC inside the system fails or is not accurate, the heater will
H1.0 stop working during alarming, it is invalid to press Silence/reset key.

Heat circuit failure(e.g., heater, solid relay, mechanical relay and all control
code
circuit fails), the heater will stop working during alarming, it is invalid to
H1.1
press Silence/reset key.

W3 3-7
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 4
PREVENTIVE MAINTENANCE
4.1 GENERAL
The section provides cleaning and maintenance instructions.

WARNING: Please cut off all switch connected with the power supply before cleaning.

4.2 CLEANING

This device must be cleaned and sterilized for the first time for initial use, or after used it for
one week.

4.2.1 DISASSEMBLY BEFORE CLEANING

A. Please cut off the main power connection and turn off all switches before cleaning.

B. Take out the skin temperature sensor from the temperature controller.

C. Take out the mattress from the bassinet, and take off the bedspread.

D. Please disassemble the panel from the bassinet according to the contrary sequence as

the step B of 2.2 in section 2.

E. Pull out the X-ray tray from the bassinet.

F. Please disassemble the phototherapy unit according to 2.2. in section 2.

4.2.2 CLEANING PROCEDURE

Please use the detergent registered by nation, and please follow the instruction for detergent

usage. Please proceed the following procedure after cleaning all dirt on the disassembled parts.

Please dry them after cleaning.

CAUTION: Some chemical detergent will conduct electricity or leave some remains so as

to the accumulation of dust that is easy conduct electricity. Please keep the

detergent away from all electric parts. Please do not leave the detergent on

these parts .

A. Clean the skin temperature sensor.

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

4-1
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

B. Clean the bedspread

Scrub the bedspread with the disinfectant-detergent. After washing in the clean water, then

make it dry.

C. Clean the panel

NOTE: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol, acetone,

or any organic solvents for cleaning. Do not expose the panel assembly to direct

ultraviolet radiation.

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

D. Clean the X-ray tray

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

E. Clean the bassinet, tray and the I.V pole

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

F. Clean the phototherapy unit

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

NOTE: The light resource protective panel is made of acrylic glass, in order to avoid the

crack ,please do not clean it with alcohol, acetone or other organic solutions, do

not place it under the direct radiation ultraviolet.

G. Clean the surface of the device.

Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or

paper towel.

NOTE: You should wait for the heater head became cooling completely, then the surface of

it can be cleaned.

CAUTION: 1. Please avoid the liquor flowed into the device during cleaning.

2.The reflect cover is an important parts which used to reflect the infrared

radiation heater to the mattress. It is easy to damage and you must be very

careful to clean it to prevent damaging it, or change the shape of it.

H. Clean the resuscitate bag and infant head fixing unit

Please clean according to the instruction in User’s manual.

4-2
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

4.2.3 ASSEMBLE AFTER CLEANING


NOTE: Before install the parts onto the radiant warmer, please check each parts carefully
and to see whether there is any broken. If there is any broken, it should be replaced
immediately.
A. Inset into the X-ray tray into the bassinet.
B. Please assemble the panel into the bassinet according to the step B of 2.2 in section 2.
C. Put the mattress into the bedspread, and then put the mattress on the bassinet.
D. Put the skin temperature sensor into the sensor socket.
E. Please assemble the phototherapy unit according to the step D of 2.2. in section 2.

4.3 STERILIZATION
NOTE: Please do not steam autoclave.
Please use the low temperature sterilization or gas sterilization.
Please clean the device completely before gas sterilization. All replacements should be
moved from the warmer. The new replacement should be installed after the sterilization.
NOTE: Sterilization must not exceed 54.5℃.
After gas sterilization, please keep ventilation for 16-24hrs. During the ventilation, the warmer
should work in the day environment of 32℃-35℃. If you do not use the warmer after ventilation,
please cover the warmer with the dustproof mantle.

4.4 MAINTENANCE
4.4.1 RECHARGEABLE BATTERY MAINTENANCE.
Please check the condition of the build-in rechargeable battery before the first use of device
or in the alternation of device using.
A. Operate the unit for a period of 12 to 24 hours.
B. Trigger a power failure alarm by disconnecting the AC power cord.
C. The power failure alarm should activate and continue to alarm for at least 10 minutes.
D. Reconnect the unit to the AC line and recharge the battery.
If the power failure alarm cannot last more than 10 minutes, please replace the rechargeable
battery. For this battery, it should be replaced by qualified service personnel.

4.4.2 HEATER’S REPLACE


In order to ensure the effect of the infrared radiant, when the lifetime of heater ends, it must
be replaced even it can still work. The reason is:
The electromagnetism spectrum infrared radiance of the heater will be reduced with the long
working time. This will make the device not achieve the standard as table 3.1 in this manual.
Therefore, the previous effect will not be good when the doctor uses it to keep the patient
warm.
For the heater’s replacing, please refer to section 5.

4-3
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

4.4.3 ACCURACY CALIBRATUON OF BABY TEMPERATURE SENSOR


Skin temperature sensor must be calibrated every half a year. The reason is:

With aging of skin temperature sensor, and the accuracy of sensor will changes a little to
make the baby temperature and the real temperature different so that the output heat can not
neet the clinical need heat.
Calibration method: in working state, put the skin sensor into the water cup at 36℃±0.1℃ or
inset the 36℃ calibration device on the socket of skin sensor, press the calibration key until the
Baby temperature indicator displays “--,--” for 10s, after the flashing stops, the calibration will
end when it indicates 36℃, and the you can continue to use the skin sensor. When it indicates
“EL.L”, it means the calibration error, please replace the skin temperature sensor.

NOTE: The temperature of the water cup must keep at 35.3℃±0.1℃ during calibration.

4.4 MAINTENANCE OF PHOTOTHERAPY UNIT


When the lifetime of phototherapy unit ends, please replace it for the phototherapy effect
even though it can still work. The reason is:
The radiation of LED will decline with the long working time, so the device will not meet the
requirements of Max. General bilirubin radiation in table 3.1, so that the preventative effect of
bilirubin radiation will lose.
The following table lists some common trouble within the lifetime.

Failure Reason Troubleshooting


4 pcs LED in series is one Connect the power supply;
Group LED lamp is not on group, usually one of them is please check the LED failure
damaged. with multi-voltage meter.
1. fuse on LED power board 1. Replace the fuse;
is damaged; 2. check the connection
2. bad connection between between the LED power
the LED power board and board and LED board;
All LED lamp is not on LED board; 3. Check the power cord of
3. Plug of phototherapy unit phototherapy unit and the
is not inserted or the power output socket of
connection is not good. warmer module.
1. bad connection between 1. check the connection
the LED power board and between the LED power
LED board; board and LED board;
A lot of LED lamp is not on 2. the power spruce on the 2. Check whether the power
LED power board can not cord of the damaged LED
work. is open circuit or short
circuit.
Please refer to section 5 regarding the replacement of phototherapy unit.

4-4
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 5
SERVICE
5.1 GENERAL
This section provides system setting procedure and the troubleshooting procedure of these

parts: replacement of illumination light, Quartz heater, Fuse, Internal charging Battery and

Controller.

5.2 REPLACEMENT OF CHARGING BATTERY, FUSE, QUARTZ HEATER,


ILLUMINATION LIGHT AND PHOTOTHERAPY UNIT.
5.2.1 REPLACEMENT OF ILLUMINATION LIGHT

See figure 5.1, open the protective cover of warmer module, disconnect the cable for

illumination light, and then open its fixing seat with pincher in order to take out the lamp and

replace it.

Fixing seat

Cable for illumination light

Illumination light

FIGURE 5.1

5.2.2 EPLACEMENT OF QUARTZ HEATER

See figure 5.2, open the protective cover of warmer module, unscrew the cable of two sides

for quartz heater and the mounting nut for bracket, take out the spring, disassemble the bracket

to take out the quartz heater.

Bracket Quartz heater

The mounting nut for bracket Cable of two sides for quartz heater

FIGURE 5.2
NOTE: Replacing of the quartz heating pipe must after it is cooled down.

5-1
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

5.2.3 REPLACEMENT OF CHARGING BATTERY

Open Controller and replace internal battery. See figure 5.3, disonnect the battery fastener

and the internal charing battery, and take out the plastic warsher to replace the battery.

Internal charing battery Battery fastener

Plastic warsher

Fuse’s seat Fuse’s seat

FIGURE 5.3

NOTICE: 1.During replacing the internal battery, the plastic washer should be placed the

original position, and keep correct connection between the anode/ cathode of

battery and its fastener;the faulty connection will cause the damage to the

battery and no hont for power failure alarm.

2.Please replace theb internal battery after using it for three years.

3. Battery will be automatically charged when the temperature controller works,

charging the battery outside of the incubator is forbidden.

4.The charging battery inside the wartmer is GP17R8H Ni-MH8.4V/170mAh, to

ensure safety, do not disassemble battery. Connect the battery correctly, do not

use other type battery.

5.It will not influence battery charging and the battery capacity when the battery

is full. (Switch on controller)

6. Please do not throw the old battery into fire or water to avoid explosion or

leakage.

7.Please dispose the Old battery accoring to local law, and please do not throw

them away randomly.

5-2

W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

5.2.4 REPLACEMENT OF FUSE

A.Open the fuse’s covers in figure 5.4.1 and 5.4.2 to replace 2 fuses F6.3AH/250V on VHA

and 2 fuses F5AH/250V on warmer module.(for infant radiant warmer with VHA)

Fuse’s Seat Fuse’s Seat

FIGURE 5.4.1 FIGURE 5.4.2

FIGURE 5.4

B.Open the fuse’s covers in figure 5.5 to replace 2 fuses F5AH/250V on warmer module.(for

infant radiant warmer with fixed stand)

Fuse’s Seat

FIGURE 5.5

C.See figure 5.3, open the protective cover on warmer module, and open the the fuse’s

cover of T400mAL/250V and T800mAL/250V on power board to replace 2 fuses. Open the the

fuse’s cover of T160mAL/250V to replace the fuse.

IMPORTANT: Please cut off the power supply, and pull off the power cord to replace the

fuse according to the specification.

5.2.5 REPLACEMENT OF PHOTOTHERAPY UNIT (For HKN-93A, 93B)

Disassemble the phototherapy unit and unscrew the limit bolt on the protective cover and

LED board. Take out the LED board as figure 5.6 indicates to replace the LED power board and

LED board. Open the fuse’s cover to replace fuse T1.6AL/250V.

5-3
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

LED board Fuse T1.6AL/250V LED power board

Pull out the LED board as the arrow indicates LED board

FIGURE 5.6

5.3 SYSTEM SETTING PROCEDURES


5.3.1 GENERAL

This paragraph provides controller system setting procedure. This procedure is only available

to the qualified service personnel.

NOTICE: Please follow by the procedures of item 5.3.3. if it is unnecessary, please do not

change data of system setting to avoid hazards.

5.3.2 BRIEF INTRODUCTION

Controller system setting procedures listed below:

1. Brightness

2. Spare

3. Baby temperature compensation

4. Machine Number

5. High baby temperature deviation

6. Low baby temperature deviation

7. Upper limit of baby temperature set

8. Lower limit of baby temperature set

9. Over temperature alarm value

5-4

W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

CODE 00.1 BRIGHTNESS

Stipulate the brightness of the indicators and the set range is 00.0~01.5. The higher the

value is, the brighter the data is.

CODE 00.3 BABY TEMPERATURE COMPENSATION

Stipulate the compensationof the baby temperature and the set range is -5.0~05.0. The first

figure for set means increasing or decreasing. Example:00.2℃ means that the baby temperature

will be added 0.2℃ as compensation. -1.3℃ means that the baby temperature will be reduced

1.3℃ ascompensation.

CODE 00.4 MACHINE NUMBER

Set range is 00.0~04.0. Example 00.1 means No.1 machine, 01.5 means No.15 machine.

CODE 00.5 HIGH BABY TEMPERATURE DEVIATION

Set high deviation under baby mode, set range is 00.0~03.0. Example: the data 01.0

means high deviation 1.0 ℃.

CODE 00.6 LOW BABY TEMPERATURE DEVIATION

Set low deviation under baby mode, set range is 00.0~03.0. Example: the data 01.0 means

low deviation 1.0 ℃.

CODE 00.7 UPPER LIMIT OF BABY TEMPERATURE SET

Set Max. set value of baby temperature, set range is 35.0~38.0. Example: the data 37.5

means high deviation 1.0 ℃.

CODE 00.8 LOWER LIMIT OF BABY TEMPERATURE SET

Set Min. set value of baby temperature, set range is 30.0~35.0. Example: the data 34.5

means.

CODE 00.9 OVER TEMPERATURE ALARM SET

Set range is 37.0~41.0. Example 38.5 means that alarm value of over temperature is

38.5.℃.

5-5
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

5.3.3 PROCEDURE

At first, connected power supply, press and hold for 2s as soon as switch on until Baby

temprature indicator displays “ --.-“, and no display on Set temprature indicator. And then,

press Or until Baby temprature indicator displays “78.6”, after then, press to enter

into the system set screen, or press to exit

.After entering into the set screen, Set temprature indicator displays set code, Baby

temprature indicator displays the specific set data.

Turn on the power switch

Press and hold the for 3s

Enter password

Press key, EXIT Press or key to enter password 78.6,


Then press key to confirm

Error Password check

YES

System setting screen

Press key or Key (For HKN-93,press or


key ) set temperature indicator displays the set code

Brightness set of indicator

Press or Key can change


current set value

END

Press key Press and together

SAVE and EXIT Return to original set

FLOW CHART 5.1 SYSTEM SET PROCEDURE FLOW CHART

5-6

W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

5.4 TROUBLESHOOTING PROCEDURES


5.4.1 GENERAL

Troubleshooting guides for the Controller are provided in Paragraph 5.4.3 and Paragraph 5.4.4.

Paragraph 5.4.3 provides the alarm code when the controller alarms. Paragraph 5.4.4 provides

troubleshooting in the form of flowcharts.

z It is confirmed that the system settings are correct before troubleshooting.

5.4.2 TEST EQUIPMENT REQUIRED

The test devices listed below are required for troubleshooting the Controller. Equivalent test

device may be substituted.

z Digital VOM, FLUKE 8060A

z 31.0℃ Baby Temperature Simulation

Simulator probe for sensor

Resistor R1, 2=23166Ω±182Ω/0.25W/0.5%

5.4.3 TROUBLESHOOTING

All alarms have the relevant alarm code except for the power ٛ ailure alarm , please refer to Table

5.1.

TABLE 5.1 ERROR CODES

ALARM EXPLANATION AND


TYPE
CODE CORRECTIVE ACTION

Main RAM failure alarm H0.1 Replace main MCU

Main ROM failure alarm H0.2 Replace main MCU

1. Check the connection of circuit between


Double MCU communication failure alarm H0.3 the main MUC and auxiliary MCU
2. Replace main MUC and auxiliary MCU

E2PROM main MCU failure alarm H0.4 Replace main MCU

5-7
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

TABLE 5.1 ERROR CODES(CONTINUE)

RAM failure alarm inside auxiliary MCU H0.5 Replace auxiliary MCU

ROM failure alarm inside auxiliary MCU H0.6 Replace auxiliary MCU

Main ADC failure alarm H0.7 Refer to Flow chart 5.1

Auxiliary ADC failure alarm H0.8 Refer to Flow chart 5.2

SRAM failure alarm H0.9 Replace U7 on mian board

Real timer RTC failure alarm H1.0 Replace U21 on mian board

Heat circuit failure alarm H1.1 Refer to Flow chart 5.3

E0.1 Refer to Flow chart 5.4

Skin temperature sensor failure alarm


E0.2 Refoer to Flow chart 5.5

Deviation failure alarm Skin


E0.3 Refer to Flow chart 5.6
temperature sensor

Over-temp alarm E0.4 Refer to Flow chart 5.7

Upper deviation E0.5 1. Check the position of sensor


Deviation alarm 2. Check the ambient temperature
3. Make the device far away the heat resource
Lower deviation E0.6

Please set the temperature or change the cotrol


E0.7
mode correctly
Set alarm
Please check the condition of probe and the
E0.8
patient’s skin

Power failure alarm NO Refer to Flow chart 5.8

5-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

5.4.4 TROUBLESHOOTING FLOW CHART


Before troubleshooting, forst, please switch off, and disconnect the power cable, and then open

the top cover of warmer module and the back cover of column. And then connect the power cable and

turn onthe switch, at the same time, make sure that power +5V, +12V of controller, element, and all

cables are in good condition.In working state, press Key and Key , and the Baby temperature

indicator displays the temperature measured by the Over-temp probe inside skin sensor, and Set

temperature indicator displays the voltage of battery, so it is easy to find the reason of failure.

Main ADC failure alarm

Primary power
NO
On main board between R8 and R9 Check R8 and R9.

is +2.5V ?

YES

On main board Pin of U5-9 is +2.5V ? NO Check the associated circuits

Between U5-9 and R8, R9

YES

Connect U5 to NO
On main board Check the associated circuits.
U16 is normal?

YES

Replace U5

FLOW CHART 5.1 MAIN ADC FAILURE ALARM

5-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Auxiliary ADC failure alarm

Primary power NO

On main board between R14 and R15 Check R14 and R15.

is +2.5V ?

YES

NO
On main board Pin of U6-7 is +2.5V ? Check the associated circuits

Between U6-7and R14, R15

YES

On main board Connect U6 to NO


Check the associated
U19 is normal? circuits.

YES

Replace U6

FLOW CHART 5.2 AUXILIARY ADC FAILURE ALARM

5-10
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Heat circuit failure alarm

Inside warmer module open circuit of YES Replace heater


heater?

Checking point: Point A

Controller is in the heating stae of 100%

Solid relay is Pin 1of solid relay YES


Checking point: Point B
inside the and J1-2 is AC power?
power frame,
and J1 on the
power board NO

Control terminal YES


On power board of Solid relay K1 is Check solid relay abd replay K1
+12V?
NO

YES
On power board Triode Q2-B, Q3-B is Replace Q2, Q3
>0.7V?

NO

On main board U10-9, U10-7 is high level? Check connection between

the main board and power

NO board

Replace main board

FLOW CHART 5.3 HEAT CIRCUIT FAILURE ALARM (Sheet 1 of 4)


5-11
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Checking point: Point A

Controller is in the state of non-heat

Solid relay is inside Pin 1 of NO


the power frame sole relay and J1-2 is Checking point: Point C
J1 is on powre board AC power?

YES

The control terminal NO


On Power board Check solid relay and
of solid replay K1 is +12V?
relay K1

YES

YES Replace Q2, Q3


On power board Triode Q2-B, Q3-B is >0.7V?

YES

U10-9, NO

On main board U10-7 is high level? Check connection between

the main board and power

board

YES

Replace main board

FLOW CHART 5.3 HEAT CIRCUIT FAILURE ALARM (Sheet 2 of 4)

5-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Checking point: Point B

NO

On power board Voltage of U1-2 is >5mV? Check inductor T1

YES

NO

On power board Voltage of U1-1 is >2V? Check U1, R1, C3

YES

On power board Triode Q1-B is NO Check R2, R4, C4

>0.7V?

NO

On main board U11-2 is high level? Check the connection between

main board and the power

board

YES

Replace main board

FLOW CHART 5.3 HEAT CIRCUIT FAILURE ALARM (Sheet 3 of 4)

5-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Checking point: Point C

YES

On power board voltage of U1-2 is >5mV? Check inductor T1

NO

YES

On power board voltage of U1-1 is >2V? Check U1, R1, C3

NO

YES
On power board Triode Q1-B is >0.7V? Check R2, R4, C4

NO

YES

On main board U11-2 is high level? Check the connection

between main board and

power board

NO

Replace the main board

FLOW CHART 5.3 HEAT CIRCUIT FAILURE ALARM (Sheet 4 of 4)

5-14
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Skin Temperature Sensor Failure Alarm (E0.1)


``

Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J4 on main board

The 31.0℃±0.2℃Baby Temperature Simulation (decade


, equal 23166Ω±182Ω) will instead skin sensor, turn on
power supply

Baby temperature indiccator NO


Display 31.0℃±0.2℃ alarm code E0.1? Replace the skin temperature sensor

YES

NO
On the main board U5-7 is 2.3V? Check relay R4 and the connecting
cable between resistor R4and U5-7

YES

Check U5

FLOW CHART 5.4 Skin Temperature Sensor Failure Alarm (Sheet 1 of 1)

5-15
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Skin Temperature Sensor Failure Alarm (E0.2)


``

Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J3 on main board

The 31.0℃±0.2℃Baby Temperature Simulation (decade


, equal 23166Ω±182Ω) will instead skin sensor, turn on
power supply

Probe temperature of skin NO


sensor 31.0℃±0.2℃ alarm code E0.2? Replace the skin temperature sensor
on page 5-8

YES

NO
On the main board U6-6 is 2.3V? Check relay R5 and the connecting
cable between resistor R5 and U6-6

YES

Check U6

FLOW CHART 5.5 Skin Temperature Sensor Failure Alarm

5-16
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Skin Temperature Sensor deviation Failure Alarm (E0.3)


``

Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J3 on main board

The 31.0℃±0.2℃ Baby Temperature Simulation (decade


, equal 23166Ω±182Ω) will instead skin sensor, turn on
power supply

2 Probes temperature of skin NO


sensor at 31.0℃±0.2℃ alarm? Replace the skin temperature sensor
on page 5-8

YES

NO
On the main board U5-7 and U6-6 Check resistor R3 and R5
is 2.3 V?

YES

Check U5 or U6

FLOW CHART 5.6 Skin Temperature Sensor devoation Failure Alarm

5-17
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

OVER TEMPERATURE ALARM

YES
Heating source nearby Far away from
heating source

NO
Over-temp probe NO
On page 5-8 Temperature of Over-temp probe Troubleshooting
Is normal? according to flow
chart 5.5
YES

Solid relay is inside NO


the power frame, and Pin1 of sold relay and J1-2 is Checking point:
J1 on the power board AC power? Point C ( page 5-13)

YES

NO
On powr board Control terminal of solid relay K1 Check solid relay and
is +12V? relay K1

YES

Triode Q2-B and NO


On power board Q3-B is >0.7V? Replace Q2, Q3

YES

FLOW CHART 5.7 Over Temperature Failure Alarm


5-18
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

From sheet 1

U10-9, U10-7 NO Check connection btween


On main board
Is high level? Main board and power
board

YES

FLOW CHART 5.7 Over Temperature Failure Alarm (Sheet 2 of 2)

Power failure alarm

NO
Controller has AC power input? Check main power input

YES

Fuse on the back cover of YES Replace fuse


controller is damaged?

NO

NO
Second output of power transformer Check transformer
is normal? and initial fuse

YES

FLOW CHART 5.8 Power Failure Alarm (Sheet 1 of 2)


5-19
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

From sheet 1

NO
On power board Output voltage of U3-3 is +5V? Check F1, B1, U3,
C5, C9, C13,
C6, D4
YES

NO
On power board J8-8 is high level? Check connetion between

U5, U2, R17-R22 and


power board and
main board
YES
NO
On main board J2-8 is high level? Check the data bus
between the main
YES board and power
board

Replace U11

FLOW CHART 5.8 Power Failure Alarm (Sheet 2 of 2)

5-20
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 6
REPLACEMENT PARTS

6.1 GENERAL
This section provides parts lists for infant radiant warmer. Part numbers of accessories are

provided below:

Figure 6.1 Parts Location Diagram, whole unit

6-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.1 Replacement Parts List, whole unit

ITEM No. DESCRIPTION PART NUMBER

1 Warmer module K05 000 00

2 column K03 000 00

3 bassinet K04 000 00

4 VHA stand P01 000 00

5 controller K02 000 00

6 Phototherapy unit K06 000 00

Figure 6.2 Parts Location Diagram, Warmer Module

6-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.2 Replacement Parts List, Warmer Module

ITEM No. DESCRIPTION PART NUMBER

1 Shell for warmer module K05 001 00

2 Switch panel K05 002 00

3 Reflective mantle K05 003 00

4 Reflection bar K05 004 00

5 Protective rod K05 005 00

6 Frame K05 006 00

7 Spring II) K05 007 00

8 Snap spring K05 008 00

9 Cold lamp K05 009 00

10 R4.0 transformer K05 010 00

11 Top cover K05 011 00

12 Cross groove pan head bolt M6×14 See No. of standard Parts

13 Cross groove pan head bolt M6×12 See No. of standard Parts

14 Cross groove pan head bolt M4×12 See No. of standard Parts

15 Flat washer 4 See No. of standard Parts

16 Spring washer 4 See No. of standard Parts

17 external teeth serrated lock washers 4 See No. of standard Parts

18 Wiring slice 4 See No. of standard Parts

19 Hexagon nut M4 See No. of standard Parts

20 Cross groove pan head bolt M3×12 See No. of standard Parts

6-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.2 Replacement Parts and Accessories List

ITEM
DESCRIPTION PART NUMBER
No.

21 Spring washer 3 See No. of standard Parts

22 Flat washer 3 See No. of standard Parts

23 Cross groove pan head bolt M4×40 See No. of standard Parts

24 Hexagon nut M3 See No. of standard Parts

26 Cross groove pan head bolt M3×8 See No. of standard Parts

26 External teeth serrated lock washer 3 See No. of standard Parts

27 Cross groove pan head bolt M4×30 See No. of standard Parts

28 Powe switch K05 012 00

29 Cover nut M3 See No. of standard Parts

30 Cold pressure wiring terminal 4 K05 013 00

31 Quartz heater K05 014 00

32 Compressed spring P01 013 00

33 Scaleboard K05 015 00

6-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.3 Parts Location Diagram, Column

Table 6.3 Replacement Parts List, Column

ITEM No. DESCRIPTION PART NUMBER

1 Tightening clamp K03 001 00

2 shelf K03 002 00

3 Seat for shelf K03 003 00

4 Cross arm of I.V. Pole K03 004 00

5 Infusion rod K03 005 00

6 Ferrite K03 006 00

7 Nylon strap K03 007 00

8 Sping washer 5 See No. of standard Parts

9 Cross groove pan head bolt M5×16 See No. of standard Parts

10 Cross groove pan head bolt M4×10 See No. of standard Parts

11 Flat washer 4 See No. of standard Parts

12 Spring washer 4 See No. of standard Parts

6-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.4 Parts Location Diagram, Bassinet

6-6
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.4 Replacement Parts List, Bassinet

ITEM No. DESCRIPTION PART NUMBER

1 Back panel K04 001 00

2 Front panel K04 002 00

3 Left panel K04 003 00

4 Right panel K04 004 00

5 Fixing seat K04 005 00

6 Left socket K04 006 00

7 X-ray tray K04 007 00

8 Right socket K04 008 00

9 Hexagon nut M6 See No. of standard Parts

10 Spring washer 6 See No. of standard Parts

11 Flat washer 6 See No. of standard Parts

12 Pull handle K04 009 00

13 Spring washer 4 See No. of standard Parts

14 Cross groove pan head bolt M4×12 See No. of standard Parts

15 Flat washer 4 See No. of standard Parts

16 Waving handle K04 010 00

17 Slotted tightening clamp M4×10 See No. of standard Parts

18 Regulate and compress packing K04 011 00

19 External teeth serrated lock wahser 4 See No. of standard Parts

20 Bearing groupware K04 012 00

21 Sliding block K04 013 00

22 Cross connector K04 014 00

23 transmission shaft K04 015 00

24 Limit knot K04 016 00

25 Hexagon slotted nut M12 See No. of standard Parts

26 Slot pin 3.2×18 See No. of standard Parts


6-7
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.5 Parts Location Diagram, fixed Stand

Table 6.5 Replacement Parts List, fixed Stand

ITEM No. DESCRIPTION PART NUMBER

1 Pin head K01 001 00

2 Pin column K01 002 00

3 spring K01 003 00

4 Bakelite handle K01 004 00

5 Flat washer 8 See No. of standard Parts

6 Seal for guard rail K01 005 00

7 shelf K01 006 00

8 Caster with brake K01 001 00

9 Caster without brake K01 002 00

6-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.6 Parts Location Diagram, VHA stad

Table 6.6 Replacement Parts List, VHA stad

ITEM No. DESCRIPTION PART NUMBER

1 Cross groove pan head bolt M4×10 See No. of standard Parts

2 Spring washer 4 See No. of standard Parts

3 Flat washer 4 See No. of standard Parts

4 Switch groupware P01 009 00

5 Tread switch P01 013 00

6 Hexgon round headbolt M6×25 See No. of standard Parts

7 Spring washer 6 See No. of standard Parts

8 Flat washer 6 See No. of standard Parts

9 Height adjustment transformer P01 014 00

10 Height adjustment column P01 007 00

11 Caster with brake K01 001 00

12 Caster without brake K01 002 00

6-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.7 Parts Location Diagram, Controller

6-10
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.7 Replacement Parts List, Controller

ITEM No. DESCRIPTION PART NUMBER

1 Fuse’s seat Y02 025 00

2 Switch 129-432 K02 001 00

3 Power socket(include power filter FN9222-10-06) 02 043 01

4 Front push K02 002 00

5 RS-232 communication connector groupware H02 019 00

6 Cross groove pan head bolt M3×12 See No. of standard Parts

7 Flat washer 3 See No. of standard Parts

8 Spring washer 3 See No. of standard Parts

9 Hexagon nut M3 See No. of standard Parts

10 Cross groove pan head bolt M3×30 See No. of standard Parts

11 Buzzer 02 040 00

12 Transformer groupware 02 032 00

13 Cross groove countersunk head bolt M4×12 See No. of standard Parts

14 Spring washer 4 See No. of standard Parts

15 Hexagon nut M4 See No. of standard Parts

16 Solid relay K02 003 00

17 Cross groove countersunk head bolt M3×6 See No. of standard Parts

18 Power board K02 004 00

19 Battery box 02 028 00

20 Cross groove countersunk head bolt M3×12 See No. of standard Parts

21 Wiring slice 3 See No. of standard Parts

22 Earthed label 02 017 00

6-11
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.7 Replacement Parts List, Controller

ITEM
DESCRIPTION PART NUMBER
No.

23 Hexagon nut M5 See No. of standard Parts

24 Spring washer 5 See No. of standard Parts

25 Flat washer 5 See No. of standard Parts

26 External teeth serrated lock washer 5 See No. of standard Parts

27 Retagualr socket CB2-16S K02 004 00

28 Support rod 3 K02 005 00

29 Paper washer Φ3 See No. of standard Parts

30 Adjustable wiring pin K02 006 00

31 Power output socket SS-8B K02 007 00

32 AC power cord 02 030 00

6-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Figure 6.8 Parts Location Diagram, Controller

Table 6.8 Replacement Parts List, Controller

ITEM No. DESCRIPTION PART NUMBER

1 Operatin panel for controller K02 008 00

2 Cross groove pan head bolt M3×6 See No. of standard Parts

3 Spring washer 3 See No. of standard Parts

4 Paper washer Φ3 See No. of standard Parts

5 Main board K02 009 00

6 Display frame K02 010 00

7 Skin temperature sensor H02 030 00

6-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.9 Satandard parts list

ITEM
DESCRIPTION AND SPECIFICATION PART NUMBER
No.

1 Cross groove pan head bolt M6×14 BZ 084

2 Cross groove pan head bolt M6×12 BZ 015

3 Cross groove pan head bolt M4×12 BZ 009

4 Flat wahser 4 BZ 037

5 Spring washer 4 BZ 044

6 External teeth serrated lock washer 4 BZ 088

7 Wiring slice 4 BZ 077

8 Hexagon nut M4 BZ 054

9 Cross groove pan head bolt M3×12 BZ 004

10 Spring washer 3 BZ 043

11 Flat washer 3 BZ 036

12 Cross groove pan head bolt M4×40 BZ 100

13 Hexagon nut M3 BZ 053

14 Cross groove pan head bolt M3×8 BZ 002

15 External teeth serrated lock washer 3 BZ 102

16 Cross groove pan head bolt M4×30 BZ 083

17 Cover nut M3 BZ 159

18 Spring washer 5 BZ 045

19 Cross groove pan head bolt M5×16 BZ 116

20 Cross groove pan head bolt M4×10 BZ 008

21 Hexagon nut M6 BZ 056

22 Srping washer 6 BZ 046

23 Flat washer 6 BZ 039

6-14
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Table 6.9 Satandard parts list

ITEM No. DESCRIPTION AND SPECIFICATION PART NUMBER

24 Slotted tightening clamp M4×10 BZ 169

25 Hexagon slotted nut M12 BZ 170

26 Slot pin 3.2×18 BZ 171

27 Flat washer 8 BZ 040

28 Hexagon round head bolt M6×25 BZ 147

29 Cross groove pan head bolt M3×30 BZ 082

30 Cross groove countersunk head bolt M4×12 BZ 023

31 Cross groove countersunk head bolt M3×6 BZ 017

32 Cross groove countersunk head bolt M3×12 BZ 020

33 Wiring slice 3 BZ 076

34 Hexagon nut M5 BZ 055

35 Flat washer 5 BZ 038

36 External teeth serrated lock washer 5 BZ 051

37 Paper washer Φ3 BZ 050

38 Cross groove pan head bolt M3×6 BZ 001

6-15
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SECTION 7
DIAGRAMS
7.1 GENERAL
The section provides wiring diagrams for Infant radiant warmer, parts location diagrams, and
schematic diagrams of controller.

CZ1: Retagular faucet CB2-16 CZ2,CZ3: Connector 25471-4-B B: R40 transformer


RL2: Illumination light D: Earthed slice
RL1: Heater 580W/240V
SD-14.5V/50W OT1.5-4
K2: Switch of illumination light
K1: Controller ‘s switch RK1-01C/2×2
RK1-01C/2×2
F1: T type fuse 1.25AL/250V F2: T type fuse 3.15AL/250V

WIRING DIAGRAMS OF WARMER MODULE

7-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

F3:T type fuse


F1, F2:F type fuse 5AH/250V F4: T type fuse 315mAL/250V
160mAL/250V
X52: Retagular faucet
X51: RS232 connector B: Buzzer EFM-236LB/12VDC
CB2-16
S: General power switch
T: Transformer R.20 D: Wiring slice OT1.5-4
129-432
G: Charging battery GP17R8H Ni-MH8.4V/170mAh K: Solid relay GN84137010-25A
X53: Auxiliary power output faucet SS-8B X50: Power filter FN9222-10-06

WIRING DIAGRAMS OF CONTROLLER (Fixed stand)

7-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

F3: T type fuse


F1, F2:F type fuse 5AH/250V F4:T type fuse 315mAL/250V
160mAL/250V
X52: Retagualr faucet
X51:RS232 connector B: Buzzer EFM-236LB/12VDC
CB2-16
T:Transformer R.20 D:Wiring slice OT1.5-4 K: Solid relay GN84137010-25A
G:Charging battery GP17R8H Ni-MH8.4V/170mAh X50: Power filter FN9222-10-06
X53:Auxiliary power output faucet

WIRING DIAGRAMS OF CONTROLLER( VHA stand)

7-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

PARTS LOCATION DIAGRAM FOR POWER BOARD

7-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Parts List of power board

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

1 R1,R2,R11,R23,R24,R25 Resistor 0805-10KΩ 6

2 R3 Resistor RJ-0.25-10KΩ 1

3 R4~R7,R16,R28,R31 Resistor 0805-1KΩ 7

4 R8,R12,R14,R15,R17~R20 Resistor 0805-2KΩ 8

5 R9,R10,R13,R29 Resistor 0805-200Ω 4

6 R21 Resistor 0805-7.5KΩ 1

7 R22 Resistor 0805-3KΩ 1

8 R26 Resistor 0805-100Ω 1

9 R27 Resistor RJ-0.25-30KΩ 1

10 R30 Resistor 0805-510Ω 1

11 C1,C2,C5~C8,C15~C17,C25 Capacitor 0805-0.1μF 10

12 C3 Capacitor 0805-2200pF 1

13 C4 Capacitor 1210-10μF 1

14 C9~C12 Capacitor CD11-25V-470μF 4

15 C13,C14 Capacitor CD11-16V-2200μF 2

16 C18 Capacitor 0805-0.01μF 1

17 C19,C20 Capacitor CA45D-50V--4.7μF 2

18 C21~C24 Capacitor CA45D-25V--1μF 4

19 U1,U2 Integrated circuit LM358D 2

20 U3,U4 Integrated circuit MC7805CT 2

21 U5 Integrated circuit SN74AHCT00 1

22 U6 Integrated circuit 7812CT 1

23 U7,U11 Integrated circuit TLP521-1 2

24 T1 Inductor TVA1421-01 1

7-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Parts List of power board

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

25 U8 Integrated circuit TLP521-2 1

26 U9 Integrated circuit MAX232CSE 1

27 U10 Integrated circuit MAX489CSD 1

28 U12 Integrated circuit SN74LVC1G3157 1

29 K1,K2 Relay JW1Afsn-DC12V 2

30 Q1~Q5,Q7,Q8 Triode S8050 7

31 Q6 Triode S8550 1

32 B1,B2 Rectifier Bridge GBU10005 2

33 D1,D2,D3,D10 Diode 1N4148 4

34 D4,D5 Diode 1N5339B 2

35 D6,D7 Diode 1N4002 2

36 D8 Diode 1N5350B 1

37 D9 Diode HZ11B1 1

38 DC1 DC/Dc power module F0505S-1W 1

39 J1 Socket ME040-50802 1

40 J2,J6 Connector Receptacle XY2.5-3 2

41 J3 Connector Receptacle XY2.5-4 1

42 J4 Connector Receptacle XY2.5-2 1

43 J5 Connector Receptacle XY4-4 1

44 J7 Connector Receptacle XY2.5-10 1

45 J8 DC-2-16(vertical) 1

46 J9 Connector Receptacle ME040-50803 1

47 F1 Fuse T 800mAL/250V 1

48 F2 Fuse T 400mAL/250V 1

7-6
W3
EDITION/REVISION A/0

J4
C1 +12 J3
4
VCC 3 BATTERY
bat+
104 2
U1A 1
3 C4 R5
J1 1 R4 10uF 1K 4PIN
1 2 1k
2 R2 heater_state U3
T1 D1 D2 LM358 10K B1 Va MC7805CT VCC
2PIN 1N4148 1N4148 Q1 3 1 1 3
R1 AC + Vin +5V
8050
F1 C5 C9 C13 C6 D4

GND
10k
Heater K1B C3 T800mA 4 2 104 470uF 2200uF 104 1N5339B
AC -
JW1aFSN 222 GBU10005
VCC +12 +5 J5

2
K1A Q3 4
3
R16 ctr_relay R17 R21 R22 2 U6
+12 2K 7.5K 3K
1K 1 MC7812CT

SCHEMATIC DIAGRAM FOR POWER BOARD


B2 +12
JW1aFSN 8050 4PIN 3 1 1 3

GND
AC + IN OUT
D3 TRANS-OUT
SERVICE MANUAL FOR INFANT RADIANT WARMER

R18 R19 R20 C15 C11 C14 C7 D8


2K 2K 2K F2 104 470uF 2200uF 104 1N5350B
4 2
1N4148 Q2 AC -
J2 +5

2
T400mA
3 GBU10005
+12 R6
2 ctr_ssr
1 3 U2A C25 U4
1k
1 D6 MC7805CT +5
+5
3PIN 8050 2 1 3
Va Vin +5V
SSR ctr LM358 U5A 104
2 1N4002 C10 C12 C8 D5

GND
K2A 3 pwr_state D7 470uF 470uF 104 1N5339B
Q8 1
R31 bat+
+12
ctr_led U2B
1K 5 1N4002

2
7 SN74AHCT00
JW1aFSN 8050 6
D10 LM358 VCC DC1 +5A

7-7
1 7
1N4148 VIN Vout+
J9
3 C19 C20
2 4.7uF 2 5 4.7uF Q7
GND Vout- 8050
1 +5A
+12 H0505S-1W C16
3PIN K2B
BLUE LED U7 C22 C23
Q6 1uF U9 104 1uF
JW1aFSN R8 2 6
R29 R11 C21 VS+ VS- C24
2K VCC +5A
R25 8550 bat+ 10K 1uF 1 4 1uF
TLP521-1 200 C1+ C2+
10k 3 5
C1- C2-
U11 +5A 11 14
R26 C2 R27 R9 T1IN T1OUT
10 7
104 30k T2IN T2OUT
100 200 12 13
R1OUT R1IN
Q4 9 8 J7
R2OUT R2IN
ctr_bat R23 8050 bat_voltage 1
D9 TLP521-1 R12 MAX232 2
10k
R24 HZ11B1 R3 2K U10 +5A C17 3
10k 10k C18 U8B 1 14 4
103 R10 NC VCC
2 13 5
RO NC
200 3 12 A 104 6
/RE A
4 11 B 7
DE B

ctr_bat
5 10 Z 8
TLP521/2 R14 DI Z
VCC +5 6 9 Y 9
2K GND Y
7 8 10
GND NC

16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
J6 +5 U8A MAX489CSD RS232
3 R13 U12
2 200 3 5
R28 B1VCC +5A
1 1
B2
1K Q5 J8 4
TLP521/2 R15 A
3PIN 16PIN 6 2
R7 2K S GND
Bell
1K SN74LVC1G3157
8050 R30
510

W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

PARTS LOCATION DIAGRAM OF MAIN BOARD 1

7-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

PARTS LOCATION DIAGRAM OF MAIN BOARD 2

7-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Parts List of main board

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

1 R1,R2 Resistor 0805-1KΩ 2

2 R3,R4,R5 Resistor RJ-0.25-20KΩ 3

3 R6,R7,R10~R13,R16,R17 Resistor 0805-10KΩ 8

4 R8,R9,R14,R15 Resistor RJ-0.25-10KΩ 4

5 RG1 Resistor Group RAA09103G-1/8-10KΩ 1

6 C1~C17,C21,C22 Capacitor 0805-0.1μF 19

7 C18~20,C25~C34,C41,C42 Capacitor 0805-0.01μF 15

8 C23,C24 Capacitor CA45B-16V-10μF 2

9 C35,C36 Capacitor CD11-16V-220μF 2

10 C37~C40 Capacitor 0805-22pF 4

11 U1 Integrated circuit SN74AHCT139D 1

12 U2 Integrated circuit SN74AHCT32 1

13 U3,U4 Integrated circuit SN74LVC1G3157DBVR 2

14 U5,U6 Integrated circuit TLV2544CD 2

15 U7 Integrated circuit DS1230Y-100 1

16 U8 Integrated circuit SN74AHC273DW 1

17 U9 Integrated circuit SN74AHCT04 1

18 U10~U12 Integrated circuit SN74LS244 3

19 U13 Integrated circuit SN74AHCT74A 1

20 U14,U15 Integrated circuit TPS3824-50 2

21 U16 Integrated circuit AT89C51ED2-SLSIM 1

22 U17 Integrated circuit SN74AHCT373DW 1

23 U18 Integrated circuit DS12C887 1

24 U19 Integrated circuit AT89C4051-24PU 1

25 JZ1,JZ2 crystal oscillator CCS11.0592MHz 2

26 L1~L6 Inductor NL3225-220/1210 6

27 J1 Group leads PJ-2×40(10 contacts) 1

28 J2 DC-2-16(vertical) 1

29 J4 Connector Receptacle XY2.5-6 1

30 J5 single row , male DPCZ-60T(9contacts) 1

7-10
W3
W3
J1
10PIN

J2
1
2

1
2
3
4
5
6
7
8
9
U2A U2B

10
U1A U1B CS2 1 CS4 4 3
VCC
A14 2 4 A12 14 12 CS4 3 G1 6 G2 TO-RXD 4
A Y0 A Y0
A15 3 5 A13 13 11 CS2 /RD 2 /RD 5 +5 5
B Y1 B Y1

DI N
CLK
6 CS1 10 CS3 U10A 6
Y2 Y2 L1

led_alm
1 7 15 9 CS5 SN74AHCT32 SN74AHCT32 1 heater_state 7

LEDCS

led_pwr
G Y3 G Y3 G L2
18 power_state 8
Y1 L3
SN74LS139A SN74LS139A C21 U2C RENB 2 16 9
A1 Y2 L4
CS3 9 WENB 4 14 10
+5 A2 Y3 L5
C17 8 CLK1 TXD2 6 12 RXD2 bat_voltage 11
A3 Y4 L6
104 /WR 10 TO-RXD 8 ctr_bell 12
+5 A4
U3 VCC U4 VCC C1 ctr_bat 13
104 TXD1 3 5 RXD1 3 5 C5 SN74AHCT32 C6 SN74LS244 14
B1VCC B1VCC +5 +5
TXD2 1 RXD2 1 U10B ctr_led 15
B2 B2 C4 VCC +5
TXD 4 C3 RXD 4 104 19 16
A A 104 U9B G VCC
S0 6 2 104 S0 6 2 U7 104 U8 104 9 +5 VCC
S GND S GND Y1
A0 10 11 D0 D0 3 2 ctr_ssr ctr_ssr 11 7 16PIN
A0 D0 D1 Q1 A1 Y2
SN74LVC1G3157 SN74LVC1G3157 A1 9 12 D1 D1 4 5 bell 3 4 ctr_relay 13 5
A1 D1 D2 Q2 A2 Y3 C35 C36
A2 8 13 D2 D2 7 6 led_pwr 15 3
A2 D2 D3 Q3 A3 Y4 220uF 220uF
C2 A3 7 15 D3 D3 8 9 17
A3 D3 D4 Q4 A4
A4 6 16 D4 D4 13 12 WENB SN74AHC04
VCC A4 D4 D5 Q5 U9A
VCC A5 5 17 D5 D5 14 15 RENB SN74LS244 VCC
A5 D5 D6 Q6
U5 104 A6 4 18 D6 D6 17 16
C25 A6 D6 D7 Q7
DOUT 1 16 ADCS A7 3 19 D7 D7 18 19 led 1 2 led_alm C22
SDO /CS C23 103 A7 D7 D8 Q8 C8
DIN 2 15 A8 25 28
SDI REFP 10uF A8 VCC +5 104
CLK 3 14 A9 24 CLK1 11
C18 C19 C20 SCLK REFM A9 CLK
J3 VCC ADC_INT 4 13 A10 21 VCC /rst 1 104
ctr_relay

EOC/(-INT) FS A10 CLR SN74AHC04


1 103 103 103 5 12 A11 23
VCC VCC /PWDN VCC A11
4
9
8

2 envi 6 11 A12 2 27 /WR SN74AHC273DW R1


A0 GND A12 WE R17 1K
7 10 CSTART A13 26 22 /RD
A1 /CSTAR A13 OE 10K
2PIN bat_voltage 8 9 A14 1 20 A15 2 1 clear U13B
Q
Q

A2 A3 A14 CE GND RST


J4 VCC C7 U11B SN74LS74A
WDI

6 TLV2544 R8 DS1230Y-100 19 C13 C15 VCC


skin1 +5 G VCC
5 C41 10K 9 ctr_led 1045 3
R9 Y1 VCC VDD RST
4 C11 U11A 104 11 7 R16
PRE
CLK
D
CLR

VCC A1 Y2
3 1 G2 13 5 U15 10K 104

7-11
10K VCC G A2 Y3
2 103 D0 18 15 3 TPS3824-50 U19
VCC Y1 A3 Y4
1 skin2 U17 104 D1 16 2 heater_state 17 1 20
10
11
12
13

Y2 A1 A4 RST VCC
D0 3 2 A0 D2 14 4 power_state C38 TXD1 2 19
D0 Q0 Y3 A2 RXD P1.7 VCC
6PIN D1 4 5 A1 D3 12 6 SN74LS244 22pF RXD1 3 18
R3 R4 R5 D1 Q1 Y4 A3 JZ1 TXD P1.6
C10 D2 7 6 A2 8 4 17
20K 20K 20K D2 Q2 A4 11.0592M XTAL2 P1.5
D3 8 9 A3 5 16 cstart2
clear

+5 D3 Q3 XTAL1 P1.4
D4 13 12 A4 SN74LS244 C39 6 15
D4 Q4 /INT0 P1.3
U16 104 D5 14 15 A5 22pF 7 14
D5 Q5 /INT1 P1.2
D0 43 2 S0 D6 17 16 A6 U12A 8 13
R2 +5 P0.0/AD0 P1.0 D6 Q6 T0 P1.1
wdi D1 42 3 wdi D7 18 19 A7 1 G1 9 12
R6 P0.1/AD1 P1.1 D7 Q7 G T1 P1.0
1K D2 41 4 D0 18 J5 10 11
P0.2/AD2 P1.2 Y1 GND P3.7

4
10K D3 40 5 ADCS 1 D1 16 2 key1 1 VCC
P0.3/AD3 P1.3 OE Y2 A1 R10
D4 39 6 LEDCS ALE 11 D2 14 4 key2 2 AT89C4051
/rst P0.4/AD4 P1.4 LE Y3 A2 key3
2 1 D5 38 7 CLK D3 12 6 3
GND RST P0.5/AD5 P1.5 Y4 A3 R11 10K
D6 37
P0.6/AD6 P1.6
8 DIN SN74AHC373
A4
8 key4 4

WDI
C9 D7 36 9 DOUT C12 SN74LS244 5
P0.7/AD7 P1.7 10K
104 5
VDD RST
3 rst VCC
U12B
G
19 6 U6 VCC
RTC_INT 15 24 A8 D4 9 7 1 16
R7 P3.3/INT1 P2.0/A8 Y1 SDO /CS C26
+5 U14 ADC_INT 14 25 A9 U18 104 D5 7 11 key5 8 2 15
10K P3.2/INT0 P2.1/A9 Y2 A1 SDI REFP C24 103
TPS3824-50
P2.2/A10
26 A10 1
MOT VCC
24 D6 5
Y3 A2
13 key6 9 3
SCLK REFM
14
SERVICE MANUAL FOR INFANT RADIANT WARMER

CSTART 17 27 A11 23 D7 3 15 key7 C14 ADC2_int 4 13 10uF


P3.5/T1 P2.3/A11 SQW Y4 A3 key8 104 EOC/(-INT) FS VCC
16 28 A12 17 9PIN 5 12

SCHEMATIC DISGRAM OF MAIN BOARD


+5 P3.4/T0 P2.4/A12 A4 VCC VCC /PWDN
+5
P2.5/A13
29 A13 D0 4
AD0
skin2 6
A0 GND
11
C37 35 30 A14 D1 5 SN74LS244 7 10 cstart2
EA/VP P2.6/A14 AD1 VCC A1 /CSTAR
22pF 31 A15 D2 6 19 RTC_INT 8 9
P2.7/A15 AD2 IRQ A2 A3
9
8
7
6
5
4
3
2
1

JZ2 21 D3 7 18 /rst C16 R14


11.0592M20 XTAL1 AD3 RST VCC
D4 8 17 /RD C42 TLV2544
XTAL2 AD4 DS VCC R15 10K
C40 D5 9 RG1
AD5
22pF rst 10 11 RXD D6 10 15 /WR 104 10K
RST P3.0/RXD AD6 R/W R13
13 TXD D7 11 14 ALE 10K 103
P3.1/TXD AD7 AS 10K
/RD 19 33 ALE 12 13 CS1
P3.7/RD ALE/P GND CS key1 key2 key3 key4 key5 key6 key7 key8
/WR 18 32
P3.6/WR PSEN
DS12C887
AT89C51ED2-P44 R12 C27 C28 C29 C30 C31 C32 C33 C34
10K VCC 103 103 103 103 103 103 103 103

+5
EDITION/REVISION A/0
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SCHEMATIC DIAGRAM OF DISPLAY BOARD 1

7-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

SCHEMATIC DIAGRAM OF DISPLAY BOARD 2

7-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Parts list of diplay board

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

1 R1,R2 Resistor 0805-24KΩ 2

2 R3,R4 Resistor 0805-10KΩ 2

3 R5,R6 Resistor 0805-1 KΩ 2

4 C1,C2 Capacitor 0805-0.1μF 2

5 C3,C4 Capacitor CA45D-16V-10μF 2

6 Q1,Q2 Triode 8050 2

7 U1,U2 Integrated circuit MAX7221CGW 2

8 J1 Double row, male SPCJ-2×60(10contacts) 1

9 H1 LED Φ5 red 1

10 H2,H15,H16,H17 LED Φ5 green 4

11 H4~H9 LED ENA-L10ZAD 6

12 H10~H12 LED ENA-L10ZAG 3

13 H13 LED ((yellow and green)) 1

14 H14 LED ENA-B10ZAG 1

15 V1 Nixietube NET-5631AG 1

16 V2 Nixietube NET-3631AG 1

17 V3 Nixietube NEQ-2841AG 1

7-14
W3
EDITION/REVISION A/0

DIG14
C1
H15 H16 H17
LED LED LED
104
VCC C3

SA1

SB1

SC1
R1 10uF +5
24K V1
U1 VCC SA1 11
A R5
18 19 SB1 7 12 DIG1
RST V+ B DIG.1 1K
LEDCS 12 14 SA1 SC1 4

SCHEMATIC DIAGRAM OF DISPLAY BOARD


CS SA C
DIN 1 16 SB1 SD1 2 9 DIG2 Q1 H1
DIN SB D DIG.2 R3
J1 CLK 13 20 SC1 SE1 1 led_alm 8050
CLK SC E
SERVICE MANUAL FOR INFANT RADIANT WARMER

10 DIG14 5 23 SD1 SF1 10 8 DIG3 10k


DIG7 SD F DIG.3
9 DIG6 10 21 SE1 SG1 5
DIG6 SE G LED
8 DIG5 3 15 SF1 SDP1 3 6 +5
DIG5 SF DP NC
7 led_alm DIG4 7 17 SG1 alarm led
DIG4 SG
6 led_pwr DIG3 6 22 SDP1 NET-5631A
DIG3 SDP R6
5 DIG2 11 24 next V2
DIG2 DOUT 1K
4 DIG1 2 8 SA1 11
DIG1 DIG8 A
3 9 4 SB1 7 12 DIG4 Q2 H4
GND CON B DIG.1 R4
2 SC1 4 led_pwr 8050 1 3
+5 C K A
1 MAX7221 SD1 2 9 DIG5 10k 2 4
D DIG.2 K A
SE1 1
E
10PIN VCC SF1 10 8 DIG6
F DIG.3
SG1 5 power alarm led

7-15
G
SDP1 3 6
DP NC
NET_3631A
pre_warm manual baby set environment bat_full/charge set_alm check_alm sensor _alm over_alm deviation_alm

2
1

2
1

2
1

2
1

2
1

2
1

2
1

2
1
C2 DIG13 DIG12
H13
H10 H11 H12 H2 H3 H5 H6 H7 H8 H9

A K
A K

A K
A K

A K
A K

A K
A K

A K
A K

A K
A K

A K
A K

A K
A K
104 LED LED
VCC C4
LED2

SDP2

SA2

SB2

SC2

SD2

SE2

SF2
4
3

4
3

4
3

4
3

4
3

4
3

4
3

4
3
SB2

SC2

SD2

SE2

SF2
R2 10uF
24K
U2 VCC
18 19 V3
RST V+
LEDCS 12 14 SA2 SA2 11
CS SA A DIG11
next 1 16 SB2 SB2 7 12 DIG7
DIN SB B DIG1
CLK 13 20 SC2 SC2 4 DIG12

10

12

14

16

18

20
CLK SC C

8
DIG13 5 23 SD2 SD2 2 9 DIG8
DIG7 SD D DIG2
DIG12 10 21 SE2 SE2 1
DIG6 SE E
DIG11 3 15 SF2 SF2 10 8 DIG9 H14
DIG5 SF F DIG3
DIG10 7 17 SG2 SG2 5 LED10
DIG4 SG G
DIG9 6 22 SDP2 SDP2 3 6 DIG10
DIG3 SDP DOP DIG4
DIG8 11 24

11

13

15

17

19
1

9
DIG2 DOUT
DIG7 2 8 NEQ-2841A
DIG1 DIG8

SDP2

SA2

SB2

SC2

SD2

SE2

SF2

SG2

SDP2

SA2
9 4
GND CON
MAX7221

W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

PARTS LOCATION DIAGRAM of LED board

Parts List of LED board

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

1 R1~R40 Resistor 180Ω±1% 40

2 H1~H160 Blue light LED RL804DBC030-0 160

3 J1 XY4-6S Connector Receptacle 1

4 LED base 160

7-16
W3
W3
+15A +15B

R1 R2 R3 R4 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 R15 R16 R17 R18 R19 R20
180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180

H1 H5 H9 H13 H17 H21 H25 H29 H33 H37 H41 H45 H49 H53 H57 H61 H65 H69 H73 H77
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED

H2 H6 H10 H14 H18 H22 H26 H30 H34 H38 H42 H46 H50 H54 H58 H62 H66 H70 H74 H78
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED

H3 H7 H11 H15 H19 H23 H27 H31 H35 H39 H43 H47 H51 H55 H59 H63 H67 H71 H75 H79
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
LED

H4 H8 H12 H16 H20 H24 H28 H32 H36 H40 H44 H48 H52 H56 H60 H64 H68 H72 H76 H80
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED

+
+1
+15
+15C +15D +1

7-17
R21 R22 R23 R24 R25 R26 R27 R28 R29 R30 R31 R32 R33 R34 R35 R36 R37 R38 R39 R40
180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180

H81 H85 H89 H93 H97 H101 H105 H109 H113 H117 H121 H125 H129 H133 H137 H141 H145 H149 H153 H157
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED

H82 H86 H90 H94 H98 H102 H106 H110 H114 H118 H122 H126 H130 H134 H138 H142 H146 H150 H154 H158
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
SERVICE MANUAL FOR INFANT RADIANT WARMER

SCHEMATIC DIAGRAM OF LED BOARD


H83 H87 H91 H95 H99 H103 H107 H111 H115 H119 H123 H127 H131 H135 H139 H143 H147 H151 H155 H159
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED

H84 H88 H92 H96 H100 H104 H108 H112 H116 H120 H124 H128 H132 H136 H140 H144 H148 H152 H156 H160
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
EDITION/REVISION A/0
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

PARTS LOCATION DIAGRAM OF LED BOARD

7-18
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

Parts List of LED BOARD

ITEM NO. CODE DESCRIPTION AND SPECIFICATION QUANTITY

1 C1~4,C13~C16 Capacitor CT4-0.1μF-63V 8

2 C5~C8 Capacitor CD11-470μ-35V 4

3 C9~C12 Capacitor CD11-2200μ-25V 4

4 B1 Transistor GBU10005 1

5 U1~U4 Integrated circuit MC7815CT 4

6 Heat sink DSJ-55 4

7 F1 Fuse T1.6AL/250V 1

8 J1 Connector, receptacle XY4-2 1

9 J2 Connector, receptacle XY4-6 1

7-19
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

U1
MC7815 +15A
1 3
Vin +15V
C1 C5 C9 C13
0.1uF 2200uF 0.1uF
470uF

GND
2
U2
J1 B1 MC7815 +15B
F1
2 3 1 1 3
AC + Vin +15V
1
FUSE1 C2 C6 C10 C14
2PIN 4 2 0.1uF 2200uF 0.1uF
AC - 470uF

GND
GBU1005

2
U3
MC7815 +15C
1 3
Vin +15V
C3 C7 C11 C15
0.1uF 2200uF 0.1uF
470uF
GND
2

U4
MC7815 +15D
1 3
Vin +15V
C4 C8 C12 C16
0.1uF 2200uF 0.1uF
470uF
GND
2

J2
+15A 1
+15B 2
+15C 3
+15D 4
5
6

6PIN

SCHEMATIC DIAGRAM OF LED BOARD

7-20
W3
W3
X3'
F2 F1 S1
XZ9 1 2
X1 XZ3 XZ4
XZ7 1 3 XZ8
D 2 1 2 4

XZ21
3

XZ22
XZ5

1
1
XZ6

2
2
XZ1

XZ2

XZ11
XZ10
XZ23
XZ24

XZ5

XZ6
XZ2
XZ1
XZ11

XZ25
XZ26
XZ3 4 5 6 XZ4
XZ7 1 2 3 XZ8

X2
F3 V
XZ12-2 XZ12-1

2
1
T

7-21
To controller XZ13 S1:Power switch
X1:Socket
6 5 4
XZ13 3 2 F1,F2:F fuse 6.3AH/250V

XZ16
XZ15
1 XZ12-2

X2' (T 800mAL/250V)
F3:Wiring fuse F103
X2:Connector ZJLJQ-6

WIRING DIAGRAM OF VHA STAND


SERVICE MANUAL FOR INFANT RADIANT WARMER

M:JC35A-200 Electric handspike

XZ18
XZ16

XZ18
XZ19

XZ17
XZ15

XZ20

XZ17

XZ20
XZ19
S2、S3:Micro switch Z15G1306

3
1

2
1
S4、S5: Micro switch VSI0N021C2
X3
2
T:Power transformer R50
S3 S2 V:Rectifier bridge GBU10005
D1、D2:Diode 1N5401