WARRANTY
The product being described in this manual is warranted against defects in materials or
workmanship for one year from the date of shipment, with the following exceptions.
1. All consumable and disposable products are guaranteed to be free from defects upon shipment
only.
2. Calibrations are considered normal maintenance and are not included in the 1-year warranty.
During the warranty period any defective parts other than those listed above will be replaced at no charge
to the customer. This warranty is rendered void and our company cannot be held liable for conditions
resultant there from if:
1. Damage to the unit is incurred as a result of mishandling.
2. The customer fails to maintain the unit in a proper manner.
3. The customer uses any parts, accessories, or fittings not specified or sold by our company.
4. Sale or service is performed by the non-certified service/dealer agency.
This warranty is in lieu of all other warranties, expressed or implied, and our company shall in no
event be liable for incidental or consequential damages including loss of use, property damage, or
personal injury resulting from breach of warranty.
Please test each piece of equipment prior to initial use and at least annually thereafter. To comply
with this standard, we recommend that you participate in our training program before selling it. This
service can be performed by certified technicians through our Product Service Group and authorized
dealers.
SERVICE
For optimal performance, product service should be performed only by qualified service personnel.
Technical Services representatives can be reached for fixations and are dispatched for required
maintenance by calling 0086-574-87800008, 87800009. Customers outside China should contact their
local factory-authorized distributor for service.
COMPLEMENTARY NOTICE
This manual provides instructions for installation and maintenance of Infant Radiant Warmer Series
HKN-93A, HKN-93B, HKN-93. And HKN-93A is the most complex mode among warmer HKN-93series.
So the contect of this manual is based on HKN-93A, and the difference between these models is
explained in this manual.
Since Ningbo David, conducts a continuous product improvement program, circuit and component
improvements are sometimes incorporated into equipment before they can be incorporated into the
printed manuals. This manual contains proprietary information, repairs and authorized modifications
should be performed only by qualified service personnel to maintain your warranty and to avoid creating
safety hazards. We cannot assume responsibility for any conditions affecting the proper operation of this
equipment which may result from unauthorized repair modification. Therefore, some parts used in your
equipment may be different than those which appear in the parts list of this manual. This sometimes
occurs due to difficulty in parts procurement, but does not alter the function of the equipment
The device, accessories and the packaging have to be disposed of waste correctly at the end of the
usage. Please follow Local Ordinances or Regulations for disposal.
i
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
TECHNICAL DEFINITIONS
SKIN TEMPERATURE SENSOR: A sensing device including the link with the equipment intended to
measure the infant’s baby temperature.
CONTROL TEMPERATURE: The preventative control temperature to the temperature control.
PRE-WARM MODE: A warm mode which can keep the mattress at a properly temperature. The heater
output will work automatically according to the set programme in this mode.
MANUAL MODE: A operation mode of in which the heater output is either at a fixed level or a
proportion of its maximum output set by the operator.
BABY MODE: A mode of operation in which the power output varies automatically in response to the
temperature of the baby, to achieve a temperature close to a value set by the operator.
STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature, measured
at the center of the TEST DEVICE positioned on the mid point of the EQUIPMENT mattress, does not
vary by more than 1℃ over a period of 1 hour.
TEMPERATURE ALARM CHECKOUT STATE: The difference between the indicated temperature and
control temperature is within ±0.5℃ and such state lasts for over 10 minutes. When checkout the
temperature alarm function, operation should be enter this state.
APGAR TIMER: It offers the functions of three periods alarming indication: 1min, 5min, and 10min for
clinical treatment.
TOTAL IRRADIANCE FOR BILIRUBIN Ebi:Irradiance equal to the evaluated irradiance in the range
between 400 nm and 550nm.
UNIFORMITY G2 OF THE TOTAL IRRADIANCE FOR BILIRUBIN: Ratio of the lowest TOTAL
IRRADIANCE FOR BILIRUBIN Ebi min to the highest TOTAL IRRADIANCE FOR BILIRUBIN Ebi max on the
EFFECTIVE SURFACE AREA in section 5.
LIFETIME OF LIGHT SOURCE: Time after the average total irradiance for bilirubin Ebi is attenuated by
25%
CAUTION: A caution is inserted in text to call attention of a procedure which, It not followed exactly, can
lead to damage or destruction of the equipment.
WARNING: A warning is inserted in text to call attention to dangerous or hazardous conditions inherent
to the operation, cleaning, and maintenance of the equipment which may result in personal
Injury or death of the operator or patient.
ii
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SYMBOLS
NOTE: The label marked with * is usually sticked on the surface of phototherapy
iii
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
iv
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
TABLE OF CONTENTS
SECTION PAGE
1. GENERAL INTRODUCTION………………………………………………………………………………1-1
1.1 INTRODUCTION……………………………………………………………………………………………1-1
1.2 DESCRIPTION…………………………………………………………………………………...……… 1-1
1.3 COMPOSITION OF PRODUCTS………………………………………………………………...……… 1-1
2. INSTALLATION……………………………………………………………………………………………..2-1
2.1 UNPACKING…………………………………………………………………………………...………….2-1
2.2 INSTALLATION………………………………………………………………...………………….….…….2-1
3. TECHNICAL INFORMATION………………………………………………………………………….…3-1
3.1 SPECIFICATION……………………………………………………………………………………………3-1
3.2 WORKING PRINCIPLE………………………………………………………………………...………….3-3
4. PREVENTATIVE MAINTENANCE………………………………………………………………………...4-1
4.1 GENERAL………………………………………………………………………………………………....4-1
4.2 CLEANING………………………………………………………………………………………...………...4-1
4.3 STERILIZATION……………………………………………………………………………...………...4-3
4.4 MAINTENANCE…………………………………………………………………………………………….4-3
5. SERVICE……………………………………………………………………………………………………...5-1
5.1 GENERAL………………………………………………………………………………………………….5-3
5.2 REPLACEMENT OF CHARGING BATTERY/FUSE/QUARTZ HEATER/
ILLUMINATION LIGHT/ PHOTOTHERAPY UNIT………………………………………………………5-1
5.3 SYSTEM SETTING PROCEDURES……………………………………………………………………5-4
5.4 TROUBLESHOOTING PROCEDURES…………………………………………………………………5-7
7. DIAGRAMS……………………………………………………………………………………………………7-1
7.1 GENERAL……………………………………………………………………………………………7-1
v
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 1
GENERAL INTRODUCTION
1.1 INTRODUCTION
This manual provides instructions for installation and maintenance of Infant Radiant Warmer Series
93.
This manual provides the guidance for the qualified service personnel. As for the instruction, please
refer to the user's manual.
1.2 DESCRIPTION
Based on HKN-93, the Infant Radiant Warmer HKN-93B added with the Phototherapy; Infant Radiant
Warmer HKN-93A added the Breath Resuscitation Bag and Infant Head Fixing Unit based on HKN-93B.
The Infant Radiant Warmer Series 93 adopts 3 kinds of control mode: Pre-warm mode, Manual mode
and Baby mode. Once the device starts working, it can enter into the Pre-warm mode automatically
without pressing any key.
The setting temperature and baby temperature can be indicated separately.
Except for the listed parts in 1.3, the warmer also has the quartz halogen lamp, X-ray tray and 4pcs
casters. ( Two of them with brakes).
Bassinet
Stand
1-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
A kind of bearing part, which is used for hanging the infusion bottle.
I.V. Pole
Max. Load: 2Kg
The core part with 3 temperature control mode: pre-warm mode, manual
mode, baby mode, APGAR timer, timing function, and it is used for
Temperature Controller
automatic controlling of infrared radiation heat output, please refer to the
user's manual.
A kind of bearing part, which is used for putting some small objects.
Shelf
Max. Load: 2 Kg
It is used for place the infant. The bassinet can be tilted. Max. Load: 10
Kg. This bassinet is equipped with 4pcs of panels to avoid the patient
Bassinet
falling off.
Size of mattress: 760mm × 595mm
A part which can support the main body of warmer. We have two kinds of
stand: fixed stand and vertical height adjustment stand. Usually we equip
Stand the former one, and the second one is optional part. If equip the former
one the distance from the bassinet to the ground is 890mm; if equip the
vertical height adjustment one, the distance is 890mm~1030mm.
Breath Resuscitation Bag It is intended to rescuing the neonate when the suffocation happens or
(For HKN-93A) breath is blocked. Operate according to it’s manual.
Infant Head Fixing Unit It is intended for fix neonate’s head during treatment. Operate according
(For HKN-93A) to it’s manual.
NOTE: 1.Size of whole unit with fixed stand after installation: W730 × H1795 × D1100mm
2. Size of whole unit with vertical height adjustment stand: W730×H1795×D1100mm ~
W730×H1935×D1100mm
3. Weight of whole unit with fixed stand: 105Kg
4. Weight of whole unit with vertical height adjustment stand: 140Kg
1-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 2
INSTALLATION
2.1 UNPACKING
Generally, the WARMER is usually packed to one carton. When taking out the equipment from the
cartons, take care not to damage the spare parts of the WARMER.
FIGURE 2.1
2.2 INSTALLATION
At least two professionals are required to do the installation of the WARMER with spanners and
screwdriver.
Installation step:
A. INSTALL THE UPPER COLUMN GROUPWARE (UPPER COLUMN + WARMER MODULE)
a. The installation of the warmer with VHA.
① Unscrew the fixing bolt on the stand cover with screwdriver and remove the cover, and pull out
the power cord of warmer module.
② Unscrew the Hexagon bolt M8X20, Spring washerΦ8, Flat washerΦ8 and external teeth
serrated lock washerΦ8 in figure 2.1 with spanner.
2-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
③ Unscrew the pan head bolt M4X10, Spring washerΦ4, Flat washerΦ4 and external teeth
serrated lock washerΦ4 in figure 2.1 with screwdriver to disassemble the back cover of upper column.
④ Insert the upper column into the stand column, and tighten it with the previous unscrewed
Hexagon bolt M8X20, Spring washerΦ8, Flat washerΦ8 and external teeth serrated lock washerΦ8;
insert the power cord of warmer module into the socket in figure 4.2 long with the path inside the upper
column.
⑤ Assemble the stand cover into its original position with the unscrewed bolts and tighten the
cover with the unscrewed pan head bolt M4X10, Spring washerΦ4, Flat washerΦ4 and external teeth
serrated lock washerΦ4.
b. The installation of the warmer with fixed stand
① Unscrew the Hexagon bolt M8X20, Spring washerΦ8, and Flat washerΦ8 on the upper
column.
② Assemble the decoration slice into the column of stand as figure 2.2 indicates.
Cummerbund
Column
FIGURE 2.2
③ Insert the upper column groupware into the stand’s column as figure 2.3 indicates, and tighten it
with the previous unscrewed Hexagon bolt M8X20, Spring washerΦ8, and Flat washerΦ8.
Upper column
FIGURE 2.3
IMPORTANT: Please make sure that the upper column and the stand are vertical, or else, it will
affect the temperature uniformity on the mattress.
2-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Panel
Tightening Clamp
Shelf
2-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
FIGURE 2.7
F. INSERT THE POWER CORD
Insert the power cord into its relevant socket. Check the WARMER according to the instruction in
section 2.3.
2-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
stare, and then start the warmer to make the device enter into the pre-warm mode automatically for
30min, and please read the baby temperature to compare the indicated temperature after pre-warm and
the one in the cold state, it should not less than 4℃. If not, please replace the heater.
② please select the manual control mode, the system default of heat output proportion is 30%,
there is 3grades in the heat indicator, press Up or Down key to adjust the value within range of
0%-100%.
Please set the heat output proportion >30%. There is alarm every 15min, “CHECK” alarm light
flashes, and the set temperature indicator displays alarm code E0.9, press silence/reset key to silence
alarming. The device still works according to the current mode and heat output proportion. If do not press
silence/reset key when “CHECK” alarm actives, the device will decrease the heat output proportion to
③ In Baby mode, pull off the skin sensor, the audible and visual alarm will occur, the alarm light of
sensor and the set temperature indicator will show alarm code E0.1. Press silence/reset key to silence
alarming for 4min. connect again the skin sensor, the device will come back to the normal working
condition.
F. CHECK SETTING ALARM
In Baby mode, put the skin sensor into the water cup at temperature lower 3.5℃ than the set
temperature for 2min, the setting temperature alarm will active. The “Set” light should flash with
continuous audible sound, and the set temperature indicator will show alarm code E0.7. Press
silence/reset key to silence alarming for 4min.
2-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Set the temperature at 36℃, after the device enters into the temperature alarming checkout state,
put the skin sensor into the water cup at temperature 35.3℃±0.1℃ for 3min, the setting temperature
alarm will active. The “Set” light should flash with continuous audible sound, and the set temperature
indicator will show alarm code E0.8. Press silence/reset key to silence alarming for 4min.
In baby mode, set the temperature at 35.0 ºC. Enter temperature alarm checkout state, put the
skin sensor into the water cup at 37 ºC. When the indicated temperature reaches 36.0 ºC, the upper
deviation alarm will active. The “H/L” light should flash with continuous audible sound, and the set
temperature indicator should show alarm code E0.5; Set the temperature at 35.0℃, after the device
enters into the temperature alarming checkout state, put the skin sensor into the water cup at
temperature 33℃, when the indicated temperature decreases to 34.0 ºC, the lower deviation alarm
should active. The “H/L” light should flash with continuous audible sound, and the set temperature
indicator should show alarm code E0.6. Press silence/reset key to silence alarming for 4min.
NOTE: If the system can not enter into the temperature alarm checkout state or the baby
temperature does vary ±1℃ than the setting temperature, the deviation alarm can
not occur.
In baby mode, put the skin sensor into the water cups at 39.5℃±0.5℃, the over-temperature alarm
should active. The “Over” light should flash with continuous audible sound, and the set temperature indicator
should show alarm code E0.4. Press silence/reset key to silence alarming for 4min. reduce the baby
temperature and press silence/reset key to silence alarming.
I. CHECK APGAR TIMER
When the temperature controller is under the normal working state, press the APGAR key, the timer
start to indicate from 0s, when the timer indicates 50”~1’, 4’50”~5’, 9’50”~10’, it sounds “du…”, at the
Under the normal working condition, press to indicate the current running time, If the indicated time
is not accurate, please reset, and the specific setting method is as follows:
Press the timer key, at the same time, turn on the controller for 3min, loosen the key when the
At this time, Set temperature indicator shows the setting code, baby temperature indicator will
indicate the relevant value (as for setting code, setting item and setting range, please refer to the
following table ), the timer indicator will indicates P002(it means it has entered into the setting interface).
Operator can choose the setting items by pressing the timer key or APGAR key, and press the up and
down key to choose the data setting. After then, press the set key to store the setting value, or you can
press the silent/reset key to store and exit after finishing all setting procedure. If turn off the power supply
without pressing the silent/reset key, and this setting is invalid, and system will store the previous setting
value.
TIME SET CODE
2-7
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
R. END
2-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 3
TECHNICAL INFORMATION
3.1 SPECIFICATION
Specifications for the Radiant Warmer are provided in table 3.1. All specifications are subject to
change without notice. Changing of the air flow indoor or the tilt of the bassinet, or using heat-producing
devices near the Radiant Warmer are all possibly effect on the correlation among the temperature
uniformity of the mattress, the indicated temperature or the baby temperature.
TABLE 3.1 SPECIFICATIONS
Power Requirements………………………………………………………………………AC220V-230V/50Hz
Maximum Power Output…………………………………………………………………750W(for fixed stand)
Maximum Power Output………………………………………800W(for vertical height adjustment stand)
Maximum Heater Power Output………………………………………………………580W/240V
Auxiliary Mains Power Output……………………………AC220V-230V/50Hz, MAX. CURRENT 0.3A
Heater power display…………………………………………… 0 to 100%, adjustable in 10% increments
Protective earthed resistance…………………………………………………………………………<0.1Ω
Earth leakage current …………………………………………………………………………N.C.: <0.5mA
S.F.C.: <1mA
Patient leakage current …………………………………………………………………………N.C.: <0.1mA
S.F.C.: <0.5mA
Patient Leakage Current (MAINS VOLTAGE ON THE APPLIED PART) …………………S.F.C: <5mA
This equipment belongs to Class I, Type BF applied part, normal device(IP20)continuously operated.
W3 3-1
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
ENVIRONMENT TEMPERATURE
Operating Range………………………………………………….……..……………………+18℃~+30℃
Storage Range……..………………………………………………….……………………… -40℃~+55℃
ENVIRONMENT HUMIDITY
Operating Range…………………………………………………………………30%RH~75%RH
Storage Range………………………………………………………………………………………≤93%RH
ATMOSPHERE PRESSURE
Shipment and store atmospheric pressure range…………………………………500hPa~1060hPa
Working atmospheric pressure range…………………………………………………700hPa~1060hPa
OTHER SPECIFICATION
Working noise…………………………………………Ambient noise≤40dB(A), working noise≤50dB(A)
APGAR Timer…………………………………………………………………………………Tones at 50” ~1’,
4’50” ~5’, 9’50” ~10’
Maximum bilirubin radiation on the Mattress effective range…………………………≥800uW/cm2
Bilirubin uniformity of radiating on the mattress effective range…………………………………>0.4
Average value general bilirubin radiation of on the mattress effective range…………600 W/cm2
ALARM NOISE…………………………………………………………………………… ≤80dB(A) on mattress
SILENCE/RESET KEY
Press to silence the alarming, press twice to reset alarming.
Silence………………………………………………………………………………silence alarming for 4min,
alarming will silence if there is any other alarming actives
Reset……………………………………………………silence alarming to come back to the setting state
The lifetime ……………………………………………………………………………………………8 years
(Lifetime means the period from sell-by date to the date of discarding as useless.)
3-2 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
In the range of 320nm to 550nm, Schematic diagram of Ebi full spectrum radiation
1.2
1
0.8
s (λ ) 0.6
0.4
0.2
0
390 430 470 510 550
λ (nm)
Testing device integralE bicalibration trend
NOTE: During the process of getting the radiation and heat, the losing heat in convection,
evaporation, radiation, conduction will affect the heat balance. Therefore, to make the
baby decrease the dissipation, it is necessary to use the device in the environment of no
rapid air flowing ,and cover waterproof velum (polyethylene velum) the naked skin or to
use the waterproof net to increase the humidity around the skin to decrease the water
evaporation.
W3 3-3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Keypad control
Temperature indication
Function alarm
Output port
Heat control output
A
MIAN
MCU
A/D Temperature control probe of skin sensor
Communication connector
Auxiliary
Output Port
MCU Heat control output B
3-4 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Start
Initialization
Timer 2 interrupt
System self-test
Timing
Data display
Return
YES
Keypad checking
Keypad handling
NO
Data handling
INTO interrupt
Return
NO
Alarm
Heating control identification
Series port
interruption
Charging control
Data transfer
Data display
Return
W3 3-5
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Incubator temperature is adjusted through the compurgation between the baby temperature and the
set temperature. The Infant Radiant Warmer adopts 3 kinds of control mode: Pre-warm mode, Manual
mode and Baby mode, and the different mode had different way of heating. The user can choose the
Pre-warm mode: system will output the heat according to the programme, after a while, the system
will output the heat at 30% to maintain the temperature on the mattress until the mode changes.
Manual mode: system will output the fixed heat according to the heat output proportion to make the
temperature on the mattress reach the wanted value so as to make the body temperature of patient
Baby mode: System will adjust through the heat output proportion according to the baby
temperature and the set temperature to keep the heat balance of patient.
3.2.3 ALARM
Alarm is used for monitoring the control system state. The system will check the state by the alarm
checkout procedure. When the checked data is abnormal, it will alarm. Table 3.2 listed alarm condition.
When the indicated baby temperature is more than 1℃(set value), the
Deviation Alarm code
heater will stop working , please press the Silence/reset key to silence
indicator is on E0.5
alarming for 4min, and then it can reset if there is no failure.
3-6 W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
When the indicated baby temperature is lower than 1℃ (set value), the
Deviation Alarm code
heater will stop working , please press the Silence/reset key to silence
indicator is on E0.6
alarming for 4min, and then it can reset if there is no failure.
When the baby temperature is lower 3.5℃ than the setting temperature,
code
the heater will stop working , please press the Silence/reset key to silence
E0.7
alarming for 4min, and then it can reset if there is no failure.
Setting Alarm When the device has entered into the steady temperature condition
indicator is on because of the accidental or unreasonable out of the steady temperature
code condition, and it will not enter into this condition in 3min without the
E0.8 deviation alarm, the heater will stop working , please press the
Silence/reset key to silence alarming for 4min, and then it can reset if there
is no failure.
In the Manual mode, the “CHECK” alarm with audible and visual will active
Checking Alarm code every 15min, the heater will keep working , please press the Silence/reset
indicator is on E0.9 key to silence alarming for 4min; the heat output proportion will be limited
to 30% without pressing the Silence/reset key.
code RAM inside of main MCU fails, the heater will stop working during
H0.1 alarming, it is invalid to press Silence/reset key.
code ROM inside of main MCU fails, the heater will stop working during
H0.2 alarming, it is invalid to press Silence/reset key.
code Communication of main and auxiliary MCU fails, the heater will stop
H0.3 working during alarming, it is invalid to press Silence/reset key.
code E2ROM inside main y MCU fails, the heater will stop working during
H0.4 alarming, it is invalid to press Silence/reset key.
code RAM inside of auxiliary MCU fails, the heater will stop working during
H0.5 alarming, it is invalid to press Silence/reset key.
code ROM inside of auxiliary MCU fails, the heater will stop working during
H0.6 alarming, it is invalid to press Silence/reset key.
The alarming
ADC1 failure(e.g., main system TLV2544, including wrong sample, chip
indicator is on code
damage, no response and so on), the heater will stop working during
H0.7
alarming, it is invalid to press Silence/reset key.
ADC2 failure(e.g., main system TLV2544, including wrong sample, chip
code
damage, no response and so on), the heater will stop working during
H0.8
alarming, it is invalid to press Silence/reset key.
code SRAM failure(including system cord failure), the heater will stop working
H0.9 during alarming, it is invalid to press Silence/reset key.
code The timer RTC inside the system fails or is not accurate, the heater will
H1.0 stop working during alarming, it is invalid to press Silence/reset key.
Heat circuit failure(e.g., heater, solid relay, mechanical relay and all control
code
circuit fails), the heater will stop working during alarming, it is invalid to
H1.1
press Silence/reset key.
W3 3-7
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 4
PREVENTIVE MAINTENANCE
4.1 GENERAL
The section provides cleaning and maintenance instructions.
WARNING: Please cut off all switch connected with the power supply before cleaning.
4.2 CLEANING
This device must be cleaned and sterilized for the first time for initial use, or after used it for
one week.
A. Please cut off the main power connection and turn off all switches before cleaning.
B. Take out the skin temperature sensor from the temperature controller.
C. Take out the mattress from the bassinet, and take off the bedspread.
D. Please disassemble the panel from the bassinet according to the contrary sequence as
Please use the detergent registered by nation, and please follow the instruction for detergent
usage. Please proceed the following procedure after cleaning all dirt on the disassembled parts.
CAUTION: Some chemical detergent will conduct electricity or leave some remains so as
to the accumulation of dust that is easy conduct electricity. Please keep the
detergent away from all electric parts. Please do not leave the detergent on
these parts .
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
4-1
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Scrub the bedspread with the disinfectant-detergent. After washing in the clean water, then
make it dry.
NOTE: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol, acetone,
or any organic solvents for cleaning. Do not expose the panel assembly to direct
ultraviolet radiation.
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
NOTE: The light resource protective panel is made of acrylic glass, in order to avoid the
crack ,please do not clean it with alcohol, acetone or other organic solutions, do
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
NOTE: You should wait for the heater head became cooling completely, then the surface of
it can be cleaned.
CAUTION: 1. Please avoid the liquor flowed into the device during cleaning.
2.The reflect cover is an important parts which used to reflect the infrared
radiation heater to the mattress. It is easy to damage and you must be very
4-2
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
4.3 STERILIZATION
NOTE: Please do not steam autoclave.
Please use the low temperature sterilization or gas sterilization.
Please clean the device completely before gas sterilization. All replacements should be
moved from the warmer. The new replacement should be installed after the sterilization.
NOTE: Sterilization must not exceed 54.5℃.
After gas sterilization, please keep ventilation for 16-24hrs. During the ventilation, the warmer
should work in the day environment of 32℃-35℃. If you do not use the warmer after ventilation,
please cover the warmer with the dustproof mantle.
4.4 MAINTENANCE
4.4.1 RECHARGEABLE BATTERY MAINTENANCE.
Please check the condition of the build-in rechargeable battery before the first use of device
or in the alternation of device using.
A. Operate the unit for a period of 12 to 24 hours.
B. Trigger a power failure alarm by disconnecting the AC power cord.
C. The power failure alarm should activate and continue to alarm for at least 10 minutes.
D. Reconnect the unit to the AC line and recharge the battery.
If the power failure alarm cannot last more than 10 minutes, please replace the rechargeable
battery. For this battery, it should be replaced by qualified service personnel.
4-3
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
With aging of skin temperature sensor, and the accuracy of sensor will changes a little to
make the baby temperature and the real temperature different so that the output heat can not
neet the clinical need heat.
Calibration method: in working state, put the skin sensor into the water cup at 36℃±0.1℃ or
inset the 36℃ calibration device on the socket of skin sensor, press the calibration key until the
Baby temperature indicator displays “--,--” for 10s, after the flashing stops, the calibration will
end when it indicates 36℃, and the you can continue to use the skin sensor. When it indicates
“EL.L”, it means the calibration error, please replace the skin temperature sensor.
NOTE: The temperature of the water cup must keep at 35.3℃±0.1℃ during calibration.
4-4
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 5
SERVICE
5.1 GENERAL
This section provides system setting procedure and the troubleshooting procedure of these
parts: replacement of illumination light, Quartz heater, Fuse, Internal charging Battery and
Controller.
See figure 5.1, open the protective cover of warmer module, disconnect the cable for
illumination light, and then open its fixing seat with pincher in order to take out the lamp and
replace it.
Fixing seat
Illumination light
FIGURE 5.1
See figure 5.2, open the protective cover of warmer module, unscrew the cable of two sides
for quartz heater and the mounting nut for bracket, take out the spring, disassemble the bracket
The mounting nut for bracket Cable of two sides for quartz heater
FIGURE 5.2
NOTE: Replacing of the quartz heating pipe must after it is cooled down.
5-1
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Open Controller and replace internal battery. See figure 5.3, disonnect the battery fastener
and the internal charing battery, and take out the plastic warsher to replace the battery.
Plastic warsher
FIGURE 5.3
NOTICE: 1.During replacing the internal battery, the plastic washer should be placed the
original position, and keep correct connection between the anode/ cathode of
battery and its fastener;the faulty connection will cause the damage to the
2.Please replace theb internal battery after using it for three years.
ensure safety, do not disassemble battery. Connect the battery correctly, do not
5.It will not influence battery charging and the battery capacity when the battery
6. Please do not throw the old battery into fire or water to avoid explosion or
leakage.
7.Please dispose the Old battery accoring to local law, and please do not throw
5-2
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
A.Open the fuse’s covers in figure 5.4.1 and 5.4.2 to replace 2 fuses F6.3AH/250V on VHA
and 2 fuses F5AH/250V on warmer module.(for infant radiant warmer with VHA)
FIGURE 5.4
B.Open the fuse’s covers in figure 5.5 to replace 2 fuses F5AH/250V on warmer module.(for
Fuse’s Seat
FIGURE 5.5
C.See figure 5.3, open the protective cover on warmer module, and open the the fuse’s
cover of T400mAL/250V and T800mAL/250V on power board to replace 2 fuses. Open the the
IMPORTANT: Please cut off the power supply, and pull off the power cord to replace the
Disassemble the phototherapy unit and unscrew the limit bolt on the protective cover and
LED board. Take out the LED board as figure 5.6 indicates to replace the LED power board and
5-3
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Pull out the LED board as the arrow indicates LED board
FIGURE 5.6
This paragraph provides controller system setting procedure. This procedure is only available
NOTICE: Please follow by the procedures of item 5.3.3. if it is unnecessary, please do not
1. Brightness
2. Spare
4. Machine Number
5-4
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Stipulate the brightness of the indicators and the set range is 00.0~01.5. The higher the
Stipulate the compensationof the baby temperature and the set range is -5.0~05.0. The first
figure for set means increasing or decreasing. Example:00.2℃ means that the baby temperature
will be added 0.2℃ as compensation. -1.3℃ means that the baby temperature will be reduced
1.3℃ ascompensation.
Set range is 00.0~04.0. Example 00.1 means No.1 machine, 01.5 means No.15 machine.
Set high deviation under baby mode, set range is 00.0~03.0. Example: the data 01.0
Set low deviation under baby mode, set range is 00.0~03.0. Example: the data 01.0 means
Set Max. set value of baby temperature, set range is 35.0~38.0. Example: the data 37.5
Set Min. set value of baby temperature, set range is 30.0~35.0. Example: the data 34.5
means.
Set range is 37.0~41.0. Example 38.5 means that alarm value of over temperature is
38.5.℃.
5-5
W3
OPERATOR’S MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
5.3.3 PROCEDURE
At first, connected power supply, press and hold for 2s as soon as switch on until Baby
temprature indicator displays “ --.-“, and no display on Set temprature indicator. And then,
press Or until Baby temprature indicator displays “78.6”, after then, press to enter
.After entering into the set screen, Set temprature indicator displays set code, Baby
Enter password
YES
END
5-6
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Troubleshooting guides for the Controller are provided in Paragraph 5.4.3 and Paragraph 5.4.4.
Paragraph 5.4.3 provides the alarm code when the controller alarms. Paragraph 5.4.4 provides
The test devices listed below are required for troubleshooting the Controller. Equivalent test
5.4.3 TROUBLESHOOTING
All alarms have the relevant alarm code except for the power ٛ ailure alarm , please refer to Table
5.1.
5-7
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
RAM failure alarm inside auxiliary MCU H0.5 Replace auxiliary MCU
ROM failure alarm inside auxiliary MCU H0.6 Replace auxiliary MCU
Real timer RTC failure alarm H1.0 Replace U21 on mian board
5-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
the top cover of warmer module and the back cover of column. And then connect the power cable and
turn onthe switch, at the same time, make sure that power +5V, +12V of controller, element, and all
cables are in good condition.In working state, press Key and Key , and the Baby temperature
indicator displays the temperature measured by the Over-temp probe inside skin sensor, and Set
temperature indicator displays the voltage of battery, so it is easy to find the reason of failure.
Primary power
NO
On main board between R8 and R9 Check R8 and R9.
is +2.5V ?
YES
YES
Connect U5 to NO
On main board Check the associated circuits.
U16 is normal?
YES
Replace U5
5-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Primary power NO
On main board between R14 and R15 Check R14 and R15.
is +2.5V ?
YES
NO
On main board Pin of U6-7 is +2.5V ? Check the associated circuits
YES
YES
Replace U6
5-10
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
YES
On power board Triode Q2-B, Q3-B is Replace Q2, Q3
>0.7V?
NO
NO board
YES
YES
YES
U10-9, NO
board
YES
5-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
NO
YES
NO
YES
>0.7V?
NO
board
YES
5-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
YES
NO
YES
NO
YES
On power board Triode Q1-B is >0.7V? Check R2, R4, C4
NO
YES
power board
NO
5-14
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J4 on main board
YES
NO
On the main board U5-7 is 2.3V? Check relay R4 and the connecting
cable between resistor R4and U5-7
YES
Check U5
5-15
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J3 on main board
YES
NO
On the main board U6-6 is 2.3V? Check relay R5 and the connecting
cable between resistor R5 and U6-6
YES
Check U6
5-16
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
Check connection
between air sensor and NO
its seat, the connection Good conection Inset the sensor or J3
seat and J3 on main board
YES
NO
On the main board U5-7 and U6-6 Check resistor R3 and R5
is 2.3 V?
YES
Check U5 or U6
5-17
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
YES
Heating source nearby Far away from
heating source
NO
Over-temp probe NO
On page 5-8 Temperature of Over-temp probe Troubleshooting
Is normal? according to flow
chart 5.5
YES
YES
NO
On powr board Control terminal of solid relay K1 Check solid relay and
is +12V? relay K1
YES
YES
From sheet 1
YES
NO
Controller has AC power input? Check main power input
YES
NO
NO
Second output of power transformer Check transformer
is normal? and initial fuse
YES
From sheet 1
NO
On power board Output voltage of U3-3 is +5V? Check F1, B1, U3,
C5, C9, C13,
C6, D4
YES
NO
On power board J8-8 is high level? Check connetion between
Replace U11
5-20
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 6
REPLACEMENT PARTS
6.1 GENERAL
This section provides parts lists for infant radiant warmer. Part numbers of accessories are
provided below:
6-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
6-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
12 Cross groove pan head bolt M6×14 See No. of standard Parts
13 Cross groove pan head bolt M6×12 See No. of standard Parts
14 Cross groove pan head bolt M4×12 See No. of standard Parts
20 Cross groove pan head bolt M3×12 See No. of standard Parts
6-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
ITEM
DESCRIPTION PART NUMBER
No.
23 Cross groove pan head bolt M4×40 See No. of standard Parts
26 Cross groove pan head bolt M3×8 See No. of standard Parts
27 Cross groove pan head bolt M4×30 See No. of standard Parts
6-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
9 Cross groove pan head bolt M5×16 See No. of standard Parts
10 Cross groove pan head bolt M4×10 See No. of standard Parts
6-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
6-6
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
14 Cross groove pan head bolt M4×12 See No. of standard Parts
6-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
1 Cross groove pan head bolt M4×10 See No. of standard Parts
6-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
6-10
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
6 Cross groove pan head bolt M3×12 See No. of standard Parts
10 Cross groove pan head bolt M3×30 See No. of standard Parts
11 Buzzer 02 040 00
13 Cross groove countersunk head bolt M4×12 See No. of standard Parts
17 Cross groove countersunk head bolt M3×6 See No. of standard Parts
20 Cross groove countersunk head bolt M3×12 See No. of standard Parts
6-11
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
ITEM
DESCRIPTION PART NUMBER
No.
6-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
2 Cross groove pan head bolt M3×6 See No. of standard Parts
6-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
ITEM
DESCRIPTION AND SPECIFICATION PART NUMBER
No.
6-14
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
6-15
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
SECTION 7
DIAGRAMS
7.1 GENERAL
The section provides wiring diagrams for Infant radiant warmer, parts location diagrams, and
schematic diagrams of controller.
7-1
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-2
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-3
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-4
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
2 R3 Resistor RJ-0.25-10KΩ 1
12 C3 Capacitor 0805-2200pF 1
13 C4 Capacitor 1210-10μF 1
24 T1 Inductor TVA1421-01 1
7-5
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
31 Q6 Triode S8550 1
36 D8 Diode 1N5350B 1
37 D9 Diode HZ11B1 1
39 J1 Socket ME040-50802 1
45 J8 DC-2-16(vertical) 1
47 F1 Fuse T 800mAL/250V 1
48 F2 Fuse T 400mAL/250V 1
7-6
W3
EDITION/REVISION A/0
J4
C1 +12 J3
4
VCC 3 BATTERY
bat+
104 2
U1A 1
3 C4 R5
J1 1 R4 10uF 1K 4PIN
1 2 1k
2 R2 heater_state U3
T1 D1 D2 LM358 10K B1 Va MC7805CT VCC
2PIN 1N4148 1N4148 Q1 3 1 1 3
R1 AC + Vin +5V
8050
F1 C5 C9 C13 C6 D4
GND
10k
Heater K1B C3 T800mA 4 2 104 470uF 2200uF 104 1N5339B
AC -
JW1aFSN 222 GBU10005
VCC +12 +5 J5
2
K1A Q3 4
3
R16 ctr_relay R17 R21 R22 2 U6
+12 2K 7.5K 3K
1K 1 MC7812CT
GND
AC + IN OUT
D3 TRANS-OUT
SERVICE MANUAL FOR INFANT RADIANT WARMER
2
T400mA
3 GBU10005
+12 R6
2 ctr_ssr
1 3 U2A C25 U4
1k
1 D6 MC7805CT +5
+5
3PIN 8050 2 1 3
Va Vin +5V
SSR ctr LM358 U5A 104
2 1N4002 C10 C12 C8 D5
GND
K2A 3 pwr_state D7 470uF 470uF 104 1N5339B
Q8 1
R31 bat+
+12
ctr_led U2B
1K 5 1N4002
2
7 SN74AHCT00
JW1aFSN 8050 6
D10 LM358 VCC DC1 +5A
7-7
1 7
1N4148 VIN Vout+
J9
3 C19 C20
2 4.7uF 2 5 4.7uF Q7
GND Vout- 8050
1 +5A
+12 H0505S-1W C16
3PIN K2B
BLUE LED U7 C22 C23
Q6 1uF U9 104 1uF
JW1aFSN R8 2 6
R29 R11 C21 VS+ VS- C24
2K VCC +5A
R25 8550 bat+ 10K 1uF 1 4 1uF
TLP521-1 200 C1+ C2+
10k 3 5
C1- C2-
U11 +5A 11 14
R26 C2 R27 R9 T1IN T1OUT
10 7
104 30k T2IN T2OUT
100 200 12 13
R1OUT R1IN
Q4 9 8 J7
R2OUT R2IN
ctr_bat R23 8050 bat_voltage 1
D9 TLP521-1 R12 MAX232 2
10k
R24 HZ11B1 R3 2K U10 +5A C17 3
10k 10k C18 U8B 1 14 4
103 R10 NC VCC
2 13 5
RO NC
200 3 12 A 104 6
/RE A
4 11 B 7
DE B
ctr_bat
5 10 Z 8
TLP521/2 R14 DI Z
VCC +5 6 9 Y 9
2K GND Y
7 8 10
GND NC
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
J6 +5 U8A MAX489CSD RS232
3 R13 U12
2 200 3 5
R28 B1VCC +5A
1 1
B2
1K Q5 J8 4
TLP521/2 R15 A
3PIN 16PIN 6 2
R7 2K S GND
Bell
1K SN74LVC1G3157
8050 R30
510
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-8
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-9
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
28 J2 DC-2-16(vertical) 1
7-10
W3
W3
J1
10PIN
J2
1
2
1
2
3
4
5
6
7
8
9
U2A U2B
10
U1A U1B CS2 1 CS4 4 3
VCC
A14 2 4 A12 14 12 CS4 3 G1 6 G2 TO-RXD 4
A Y0 A Y0
A15 3 5 A13 13 11 CS2 /RD 2 /RD 5 +5 5
B Y1 B Y1
DI N
CLK
6 CS1 10 CS3 U10A 6
Y2 Y2 L1
led_alm
1 7 15 9 CS5 SN74AHCT32 SN74AHCT32 1 heater_state 7
LEDCS
led_pwr
G Y3 G Y3 G L2
18 power_state 8
Y1 L3
SN74LS139A SN74LS139A C21 U2C RENB 2 16 9
A1 Y2 L4
CS3 9 WENB 4 14 10
+5 A2 Y3 L5
C17 8 CLK1 TXD2 6 12 RXD2 bat_voltage 11
A3 Y4 L6
104 /WR 10 TO-RXD 8 ctr_bell 12
+5 A4
U3 VCC U4 VCC C1 ctr_bat 13
104 TXD1 3 5 RXD1 3 5 C5 SN74AHCT32 C6 SN74LS244 14
B1VCC B1VCC +5 +5
TXD2 1 RXD2 1 U10B ctr_led 15
B2 B2 C4 VCC +5
TXD 4 C3 RXD 4 104 19 16
A A 104 U9B G VCC
S0 6 2 104 S0 6 2 U7 104 U8 104 9 +5 VCC
S GND S GND Y1
A0 10 11 D0 D0 3 2 ctr_ssr ctr_ssr 11 7 16PIN
A0 D0 D1 Q1 A1 Y2
SN74LVC1G3157 SN74LVC1G3157 A1 9 12 D1 D1 4 5 bell 3 4 ctr_relay 13 5
A1 D1 D2 Q2 A2 Y3 C35 C36
A2 8 13 D2 D2 7 6 led_pwr 15 3
A2 D2 D3 Q3 A3 Y4 220uF 220uF
C2 A3 7 15 D3 D3 8 9 17
A3 D3 D4 Q4 A4
A4 6 16 D4 D4 13 12 WENB SN74AHC04
VCC A4 D4 D5 Q5 U9A
VCC A5 5 17 D5 D5 14 15 RENB SN74LS244 VCC
A5 D5 D6 Q6
U5 104 A6 4 18 D6 D6 17 16
C25 A6 D6 D7 Q7
DOUT 1 16 ADCS A7 3 19 D7 D7 18 19 led 1 2 led_alm C22
SDO /CS C23 103 A7 D7 D8 Q8 C8
DIN 2 15 A8 25 28
SDI REFP 10uF A8 VCC +5 104
CLK 3 14 A9 24 CLK1 11
C18 C19 C20 SCLK REFM A9 CLK
J3 VCC ADC_INT 4 13 A10 21 VCC /rst 1 104
ctr_relay
VCC A1 Y2
3 1 G2 13 5 U15 10K 104
7-11
10K VCC G A2 Y3
2 103 D0 18 15 3 TPS3824-50 U19
VCC Y1 A3 Y4
1 skin2 U17 104 D1 16 2 heater_state 17 1 20
10
11
12
13
Y2 A1 A4 RST VCC
D0 3 2 A0 D2 14 4 power_state C38 TXD1 2 19
D0 Q0 Y3 A2 RXD P1.7 VCC
6PIN D1 4 5 A1 D3 12 6 SN74LS244 22pF RXD1 3 18
R3 R4 R5 D1 Q1 Y4 A3 JZ1 TXD P1.6
C10 D2 7 6 A2 8 4 17
20K 20K 20K D2 Q2 A4 11.0592M XTAL2 P1.5
D3 8 9 A3 5 16 cstart2
clear
+5 D3 Q3 XTAL1 P1.4
D4 13 12 A4 SN74LS244 C39 6 15
D4 Q4 /INT0 P1.3
U16 104 D5 14 15 A5 22pF 7 14
D5 Q5 /INT1 P1.2
D0 43 2 S0 D6 17 16 A6 U12A 8 13
R2 +5 P0.0/AD0 P1.0 D6 Q6 T0 P1.1
wdi D1 42 3 wdi D7 18 19 A7 1 G1 9 12
R6 P0.1/AD1 P1.1 D7 Q7 G T1 P1.0
1K D2 41 4 D0 18 J5 10 11
P0.2/AD2 P1.2 Y1 GND P3.7
4
10K D3 40 5 ADCS 1 D1 16 2 key1 1 VCC
P0.3/AD3 P1.3 OE Y2 A1 R10
D4 39 6 LEDCS ALE 11 D2 14 4 key2 2 AT89C4051
/rst P0.4/AD4 P1.4 LE Y3 A2 key3
2 1 D5 38 7 CLK D3 12 6 3
GND RST P0.5/AD5 P1.5 Y4 A3 R11 10K
D6 37
P0.6/AD6 P1.6
8 DIN SN74AHC373
A4
8 key4 4
WDI
C9 D7 36 9 DOUT C12 SN74LS244 5
P0.7/AD7 P1.7 10K
104 5
VDD RST
3 rst VCC
U12B
G
19 6 U6 VCC
RTC_INT 15 24 A8 D4 9 7 1 16
R7 P3.3/INT1 P2.0/A8 Y1 SDO /CS C26
+5 U14 ADC_INT 14 25 A9 U18 104 D5 7 11 key5 8 2 15
10K P3.2/INT0 P2.1/A9 Y2 A1 SDI REFP C24 103
TPS3824-50
P2.2/A10
26 A10 1
MOT VCC
24 D6 5
Y3 A2
13 key6 9 3
SCLK REFM
14
SERVICE MANUAL FOR INFANT RADIANT WARMER
+5
EDITION/REVISION A/0
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-12
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-13
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
9 H1 LED Φ5 red 1
15 V1 Nixietube NET-5631AG 1
16 V2 Nixietube NET-3631AG 1
17 V3 Nixietube NEQ-2841AG 1
7-14
W3
EDITION/REVISION A/0
DIG14
C1
H15 H16 H17
LED LED LED
104
VCC C3
SA1
SB1
SC1
R1 10uF +5
24K V1
U1 VCC SA1 11
A R5
18 19 SB1 7 12 DIG1
RST V+ B DIG.1 1K
LEDCS 12 14 SA1 SC1 4
7-15
G
SDP1 3 6
DP NC
NET_3631A
pre_warm manual baby set environment bat_full/charge set_alm check_alm sensor _alm over_alm deviation_alm
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
C2 DIG13 DIG12
H13
H10 H11 H12 H2 H3 H5 H6 H7 H8 H9
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
A K
104 LED LED
VCC C4
LED2
SDP2
SA2
SB2
SC2
SD2
SE2
SF2
4
3
4
3
4
3
4
3
4
3
4
3
4
3
4
3
SB2
SC2
SD2
SE2
SF2
R2 10uF
24K
U2 VCC
18 19 V3
RST V+
LEDCS 12 14 SA2 SA2 11
CS SA A DIG11
next 1 16 SB2 SB2 7 12 DIG7
DIN SB B DIG1
CLK 13 20 SC2 SC2 4 DIG12
10
12
14
16
18
20
CLK SC C
8
DIG13 5 23 SD2 SD2 2 9 DIG8
DIG7 SD D DIG2
DIG12 10 21 SE2 SE2 1
DIG6 SE E
DIG11 3 15 SF2 SF2 10 8 DIG9 H14
DIG5 SF F DIG3
DIG10 7 17 SG2 SG2 5 LED10
DIG4 SG G
DIG9 6 22 SDP2 SDP2 3 6 DIG10
DIG3 SDP DOP DIG4
DIG8 11 24
11
13
15
17
19
1
9
DIG2 DOUT
DIG7 2 8 NEQ-2841A
DIG1 DIG8
SDP2
SA2
SB2
SC2
SD2
SE2
SF2
SG2
SDP2
SA2
9 4
GND CON
MAX7221
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-16
W3
W3
+15A +15B
R1 R2 R3 R4 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 R15 R16 R17 R18 R19 R20
180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180
H1 H5 H9 H13 H17 H21 H25 H29 H33 H37 H41 H45 H49 H53 H57 H61 H65 H69 H73 H77
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
H2 H6 H10 H14 H18 H22 H26 H30 H34 H38 H42 H46 H50 H54 H58 H62 H66 H70 H74 H78
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
H3 H7 H11 H15 H19 H23 H27 H31 H35 H39 H43 H47 H51 H55 H59 H63 H67 H71 H75 H79
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
LED
H4 H8 H12 H16 H20 H24 H28 H32 H36 H40 H44 H48 H52 H56 H60 H64 H68 H72 H76 H80
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
+
+1
+15
+15C +15D +1
7-17
R21 R22 R23 R24 R25 R26 R27 R28 R29 R30 R31 R32 R33 R34 R35 R36 R37 R38 R39 R40
180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180 180
H81 H85 H89 H93 H97 H101 H105 H109 H113 H117 H121 H125 H129 H133 H137 H141 H145 H149 H153 H157
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
H82 H86 H90 H94 H98 H102 H106 H110 H114 H118 H122 H126 H130 H134 H138 H142 H146 H150 H154 H158
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
SERVICE MANUAL FOR INFANT RADIANT WARMER
H84 H88 H92 H96 H100 H104 H108 H112 H116 H120 H124 H128 H132 H136 H140 H144 H148 H152 H156 H160
LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED LED
EDITION/REVISION A/0
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
7-18
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
4 B1 Transistor GBU10005 1
7 F1 Fuse T1.6AL/250V 1
7-19
W3
SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0
U1
MC7815 +15A
1 3
Vin +15V
C1 C5 C9 C13
0.1uF 2200uF 0.1uF
470uF
GND
2
U2
J1 B1 MC7815 +15B
F1
2 3 1 1 3
AC + Vin +15V
1
FUSE1 C2 C6 C10 C14
2PIN 4 2 0.1uF 2200uF 0.1uF
AC - 470uF
GND
GBU1005
2
U3
MC7815 +15C
1 3
Vin +15V
C3 C7 C11 C15
0.1uF 2200uF 0.1uF
470uF
GND
2
U4
MC7815 +15D
1 3
Vin +15V
C4 C8 C12 C16
0.1uF 2200uF 0.1uF
470uF
GND
2
J2
+15A 1
+15B 2
+15C 3
+15D 4
5
6
6PIN
7-20
W3
W3
X3'
F2 F1 S1
XZ9 1 2
X1 XZ3 XZ4
XZ7 1 3 XZ8
D 2 1 2 4
XZ21
3
XZ22
XZ5
1
1
XZ6
2
2
XZ1
XZ2
XZ11
XZ10
XZ23
XZ24
XZ5
XZ6
XZ2
XZ1
XZ11
XZ25
XZ26
XZ3 4 5 6 XZ4
XZ7 1 2 3 XZ8
X2
F3 V
XZ12-2 XZ12-1
2
1
T
7-21
To controller XZ13 S1:Power switch
X1:Socket
6 5 4
XZ13 3 2 F1,F2:F fuse 6.3AH/250V
XZ16
XZ15
1 XZ12-2
X2' (T 800mAL/250V)
F3:Wiring fuse F103
X2:Connector ZJLJQ-6
XZ18
XZ16
XZ18
XZ19
XZ17
XZ15
XZ20
XZ17
XZ20
XZ19
S2、S3:Micro switch Z15G1306
3
1
2
1
S4、S5: Micro switch VSI0N021C2
X3
2
T:Power transformer R50
S3 S2 V:Rectifier bridge GBU10005
D1、D2:Diode 1N5401