Service Note: Replacement of high-pass filter board and its power supply
modification
CyFlow® Counter II (CY-S-3022)

All procedures described in this service note should only be performed by authorized and trained technicians.

Scope: This service note is restricted to all CyFlow® Counter devices equipped with old version of
Photomultiplier Tube (PMT) connectors (see Fig. 1).

Fig. 1

1. Turn off the CyFlow® Counter.

2. Disconnect all PMT cables at the upper part of the device.

3. Disconnect high-pass filter board from controller board (see Fig. 2).

Fig. 2

Partec GmbH  Am Flugplatz 13  02828 Görlitz  Germany


Fon +49 3581 87460  Fax: +49 3581 874670  Mail: mail@partec.com  Web: www.partec.com 1/3
Rev001_2011-07-11

4. Remove all old PMT connectors carefully (see Fig. 3).

Fig. 3

5. Insert all new sockets in predefined order (see Fig 4).


(From left to right: SSC/white, FL2/orange, FL3/red)

SSC FL2 FL3


(white) (orange) (red)

Fig. 4

6. Connect all high-pass filter cables in correct colour coded order (see Fig. 5).

Fig. 5

Partec GmbH  Am Flugplatz 13  02828 Görlitz  Germany


Fon +49 3581 87460  Fax: +49 3581 874670  Mail: mail@partec.com  Web: www.partec.com 2/3
Rev001_2011-07-11

Note: Do not connect the power supply cable without the supplied adapter (see Fig. 6).

Fig. 6

7. Connect the supplied adapter to the high-pass filter power supply cable (see Fig. 7+8) and the
high-pass filter board (see Fig.9).

Fig. 7 Fig. 8 Fig. 9

8. Connect all PMT cables in appropriate order at the upper part of the device.

9. Turn on the CyFlow® Counter, perform an offset calculation as described in the “CyFlow® Counter
Instrument Service Manual” and check the device.

Partec GmbH  Am Flugplatz 13  02828 Görlitz  Germany


Fon +49 3581 87460  Fax: +49 3581 874670  Mail: mail@partec.com  Web: www.partec.com 3/3