TDLS Gas Analyzer

Process Oxygen Analysis

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time>
Advanced Process and Combustion Analysis

O2 CO

H2O

TruePeak Tunable Diode Laser Spectrometer

– O2 concentration in vapor recovery and process applications


– Combustion Gas Analysis of O2, CO, CH4, gas temperature, H2O
– Trace H2O in Corrosive Processes

<Document Number>
Copyright © Yokogawa Electric Corporation Proprietary info goes here…
<date/time> 2
Process Oxygen Analysis
REACTOR CONTROL
VAPOR RECOVERY SYSTEMS
FLARE AND VENT HEADERS

O2
O2
O2

TruePeak technology provides interference free measurement


Rapid 5 second analysis
In process measurement
No sensor contact = lower ownership costs
No sample system, no analyzer shelter, no sample lines = lower installed cost

ACCURATE MEASUREMENT for CONTROL IMPROVEMENT


FAST MEASUREMENT for SAFETY IMPROVEMENT
LOW INSTALLED AND OPERATING COST
<Document Number>
Copyright © Yokogawa Electric Corporation Proprietary info goes here…
<date/time> 3
Traditional

Traditional
Traditional
With traditional Process Oxygen With TRUEpeak you no longer have to
measurements the analyzer requires compromise:
you to:
Accuracy
Extract and transport the sample TRUEpeak method is interference free
Adding Cost and Slowing Response TRUEpeak measures the process without
changing the composition
Condition the sample to protect the analyzer
Rapid 5 second measurement in the process
Adding Cost and Compromising the Accuracy
Cost and Reliability
Protect the analyzer from the elements Installation without sample extraction,
Adding Cost transport and conditioning
No analyzer shelter required
Often the measurement also suffers from No sensor contact
Interferences – Reducing Accuracy
Sensor Degradation – Reducing Accuracy
and Adding Cost

<Document Number>
Copyright © Yokogawa Electric Corporation Proprietary info goes here…
<date/time> 4
TruePeak Features

Low Long Term Reduced Cost


Measurement Quality Reliability
Cost of Ownership To Install

- No moving parts - No analyzer shelter - No cross interference - Continuous diagnostics


- No sensor contact required - Fast measurement speed (every measurement)
- No sensor replacement - Typically no sample (5 seconds typical) - Diagnostics stored on
- High MTTF for all system required - Wet measurement (no the analyzer, history
components volume errors) data available
- Minimal - Positive indication of
calibration/validation operation
requirement - Automatic validation
- Low failure rate
- RUGGED

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 5
Tunable Diode Laser Analysis

At 760 nm Oxygen also absorbs Infrared light

With precise wavelength tuning isolated absorption peaks


can be selected

Only TruePeak area integration gives accurate analysis


with varying background gases, temperature and
pressures

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 6
Installed Cost Comparison vs Paramagnetic

Analyzer Cost Estimate


Analyzer
%O2 on Vent Header Paramagnetic TDL
Description

Analyzer hardware $ 10,000 $ 25,000

Low Cost Of Ownership Sample system $ 25,000 $ 10,000

Free standing Enclosure $ 15,000


- No moving parts
Analyzer Hardware Total $ 50,000 $ 35,000
- No sensor contact and No sensor replacement
Packaging
- High MTTF for all components
Sample transport lines $ 4,000
- Minimal calibration/validation requirements
Calibration gases & regulators $ 1,000 $ 1,000
Spare parts (est. at 10% of HW and SS
costs) $ 3,500 $ 3,500
Reduced Installation Costs Auxiliary Equipment Total $ 8,500 $ 4,500
- No analyzer shelter
Construction/On-site Installation $ 10,000 $ 5,000
- Typically, no sample system required
Other-Specify in remarks

Field Installation Total $ 10,000 $ 5,000

Contingency $ 12,400 $ 8,600

System Total $ 80,900 $ 53,100

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 7
Accuracy vs. Paramagnetic

Measurement Quality
- No cross interference
- Fast measurement speed - (5 seconds typical) Dry vs. Wet Measurement
- Wet measurement – (no volume errors)
A rough calculation of the dry
measurement errors as follows:

Paramagnetic Interference - Moisture varies from 5-15%, O2 level of 10%


- With 15% H2O, paramagnetic will read 11.76%
Table below shows a sample hydrocarbon stream [O2 dry = 10% (1/(1-.15)) ]
and he amount of paramagnetic interference for - With 5% H2O, paramagnetic will read 10.52%
the two most common process paramagnetic [O2 dry = 10% (1/(1-.05) ]
analyzers. This is not a worst case example.
Servomex data (most extensive) lists 57 (out of - Using a mid level offset during calibration the
138) gases with a higher interference than those error will be +/- 0.62% oxygen
shown below.

Vol. Interference Interference


Gas % (Servomex) (Siemens)
propane 10 -0.10 -0.09
methane 40 -0.08 -0.07 Siemens Oxymat 6 manual 02/04.
ethane 10 -0.06 -0.05
butane 10 -0.15 -0.13 Servomex Bulletin:
ethylene 10 -0.03 -0.02
“The Effects of Variations in Sample Compositions on Servomex
CO2 10 -0.03 -0.03
Magneto-Dynamic Oxygen Analyzers” using 110°C cell
Total Interference -0.44 -0.39 temperature

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 8
ACCURACY VS. 1st GENERATION(2f) TDL

Measurement Quality
- No cross interference

- TruePeak is the ONLY process TDL that integrates


the peak area, eliminating background gas effects

- Other TDL analyzers either measure peak height


only (bottom table), or “fit” the peak

- Peak fitting has to deal with background gas +


PRESSURE + TEMPERATURE (top chart)

Test of 2f TDL - Calibrated with O2 in N2


(using peak height only) - Background gas changed to 3 different hydrocarbons

PRESSURE
TEST GASES 2f TDL Reading REL ERROR
(PSIG)
0 10.00% 0.00%
10% O2 in N2
5.64 9.60% -3.96%
0 7.82% -21.78%
10% O2 in CH4
5.47 7.03% -29.70%
0 3.37% -32.67%
5% O2 in C2H4
5.66 2.87% -42.57%
0 5.94% -42.32%
10.3% O2 in C3H8
5.54 4.85% -52.90%

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 9
Reliability / Diagnostics

Paramagnetic
Reliability sensors contact the - Pressure spikes can damage the cell
gas. Aggressive - Corrosive/aggressive gases can
samples can destroy damage the cell
or compromise the
- TDL has no sample contact with sensor
measurement: - Liquids can damage the cell
- Windows can be purged to eliminate
analyzer-gas contact

- Spectra (IF stored on analyzer) can


allow historical measurement validation

- Continuous diagnostics

- TruePeak diagnostics stored on the


analyzer

- TruePeak allows Automatic validation - Diagnostics are STORED


with each measurement
- Spectra is stored during
warning/fault conditions,
TruePeak including high/low
Diagnostics concentration
- Warning/fault, calibration
history, configuration
history all recorded and
stored

<Document Number>
Copyright © Yokogawa Electric Corporation
<date/time> 10
Designed for the Most Demanding Applications

O2 CO

H2O

<Document Number>
Copyright © Yokogawa Electric Corporation Proprietary info goes here…
<date/time> 11

S-ar putea să vă placă și