Documente Academic
Documente Profesional
Documente Cultură
K A R L F IS H E R A P P A R A T U S A N D IT S
P E R F O R M A N C E V E R IF IC A T IO N
14.1 I N T R O D U C T I O N
The K arl Fisher titration is one of the most common and most sensitive methods
used in the analytical laboratory. The titrimetric determination of water is based
on the quantitative reaction of water with an anhydrous solution of sulfur diox ide
and iodine in the presence of a buffer that reacts with hydrogen ions. This titration
is a two-stage process:
A naly tical M ethod V alidation and Ins tr u ment P er for mance V er ifi cation, E dited by Chung Chow
Chan, Herman Lam, Y. C. Lee, and Xue-Ming Zhang
IS B N 0 -47 1-25 9 5 3-5 Copyright 20 0 4 John W iley & S ons, Inc.
221
222 KARL FISHER APPARATUS AND ITS PERFORMANCE VERIFICATION
Pyridine was used in the beginning of the development of the method. The
reaction was slow and the endpoint unstable because of weak basicity of pyridine.
The pyridine system buffers at about pH 4. A stronger base, imidazole, has been
used to replace pyridine since it gives a faster response and has the advantages of
lower toxicity and decreased odor. The optimal pH range for the SO2 imidazole
buffer is at pH 6. It is important that the pH of the Karl Fisher reaction be
maintained within the range 5 to 7. Outside this recommended pH range, the
endpoint may not be reached.
There are two types of Karl Fisher titrations: volumetric and coulometric.
Volumetric titration is used to determine relatively large amounts of water (1
to 100 µg) and can be performed using the single- or two-component system.
Most commercially available titrators make use of the one-component titrant,
which can be purchased in two strengths; 2 mg of water per milliliter of titrant
and the 5 mg of water per milliliter of titrant. The choice of concentration is
dependent on the amount of water in the sample and any sample size limita-
tions. In both cases, the sample is typically dissolved in a methanol solution. The
iodine/SO2 /pyridine (imidazole) required for the reaction is titrated into the sam-
ple solution either manually or automatically. The reaction endpoint is generally
detected bivoltametrically.
Coulometric titration is used to determine relatively low concentrations of
water (10 µg to 10 mg) and requires two reagents: a catholyte and an anolyte
(the generating solution). The iodine required for the reaction is generated in situ
by the anodic oxidation of iodide.
2I− → I2 + 2e− (14.3)
The iodine then reacts with the water that is present. The amount of water titrated
is proportional to the total current (according to Faraday’s law) used in generating
the iodine necessary to react with the water. One mole of iodine reacts quantita-
tively with 1 mol of water. As a result, 1 mg of water is equivalent to 10.71 C.
Based on this principle, the water content of the sample can be determined by
the quantity of current that flows during the electrolysis. For this reason, the
coulometric method is considered an absolute technique, and no standardization
of the reagents is required.
14.3 INSTRUMENTATION
Karl Fisher apparatus has to be designed to exclude moisture, deliver titrant, and
to detect the endpoint. The air in the system is kept dry with a suitable desiccant,
INSTRUMENTATION 223
and the titration vessel may be purged by means of a stream of dry nitrogen or
air. For endpoint detection, most commercially available units use a bivoltametric
method to indicate that the endpoint has been reached. For this method a constant
current of about 20 µA is applied across a pair of platinum electrodes that are
about 2.5 mm apart. As the titration proceeds, water reacts with iodine and is
consumed. When the endpoint is reached and all the water is consumed, there
is a buildup of free iodine in solution. The free iodine causes ionic conduction
in the solution. As a result, the voltage must be reduced to keep the polarization
current constant [1,2]. When the voltage drops below a defined value, the titration
is stopped. Figure 14.1 shows a schematic diagram of a typical KF titrator.
In some cases (see below) a KF drying oven is required to get the water from
a sample into the titration vessel. For these special cases, a solid sample (usually)
is placed into a specially designed KF oven where the sample is heated, and the
water goes into the vapor phase. A stream of dry carrier gas (usually, N2 or
air) sweeps the liberated moisture into the reaction vessel, where it is titrated
by either the coulometric or the volumetric method. It is critical that the carrier
gas is dry and that there are no leaks along the pathway to the reaction vessel.
Passing the carrier gas over activated molecular sieve prior to the sample will
ensure that the gas is dry.
The KF reaction depends on free water available for titration. Instruments may
be designed to homogenize the sample to release water prior to the titration. In
I = 20 mA
V = v ariable
T o pumping system
Methanol solution
Double Pt electrode
Magnetic Stirrer
this case, a high-speed blender may be used directly in the titration vessel. This
is a particularly useful accessory for determination of water in tablets or certain
food products that are difficult to prepare for analysis.
glass syringe, a known volume of purified water can be added in the titration
vessel, and the KF value for the titrant can be determined. This method is highly
operator dependent and is not recommended for measurements where the accu-
racy of the determination is critical. The preferred method would be to add a
known quantity of water by weight.
In the coulometric method, standardization is not necessary, since the current
consumed can be measured absolutely. However, a standard with known water
content should be checked periodically to ensure that the system is functioning
properly. In this case, a certified water standard is generally used, and the amount
of water is determined and compared with the amount that is certified to be
present. Some coulometric titrators are equipped with an oven for liberating the
moisture from samples that are either insoluble in methanol or that react with I2 ,
methanol, or one of the other reagents. Solid standards (e.g., potassium citrate
monohydrate) are available for checking the oven, and this check is performed
after the coulometer function has been verified.
REFERENCES