Sunteți pe pagina 1din 9

Biochemical Oxygen Demand and

Carbonaceous Biochemical Oxygen


Demand
Written by: Sheree Gossett-Johnson
www.WaterWorldCE.com
Biochemical Oxygen Demand
and Carbonaceous Biochemical
Oxygen Demand
Educational Objectives
Upon completion of this course, the operator should understand what BOD and CBOD analytical tests mea-
sure and how to perform those tests. The operator should also have a clear understanding of the appropriate
quality assurance checks used for this analytical test. In addition, a template depicting data arrangement is
provided for the operator to use for test result validity.

I. Abstract
This operator education course explains in a step by step method BOD Nutrient Buffer solution: A solution of nutrients,
how to perform the Biochemical Oxygen Demand (BOD) and phosphorus, trace metals and “buffer” chemicals to ensure a
Carbonaceous Biochemical Oxygen Demand (Carb. BOD) ana- pH suitable for bacterial growth.
lytical tests, including quality assurance and control measures.
This method can be used to report BOD and Carb. BOD results Seed Solution (for seeded BOD’s): Commercial seed, using
to Ohio EPA as required in wastewater treatment plant National a quantifiable number of a micro-organism population.
Pollutant Discharge Elimination System (NPDES) Permits. All
needed equipment, chemicals, and glassware is listed. Data and Dissolved Oxygen (DO): The amount of dissolved oxygen
sample calculations are presented to provide templates for the present within the wastewater sample.
operator to use when performing this test.
Quality Assurance / Quality Control: To assure there was
II. Keywords not any known bias introduced to the test procedure and to
Take a glance through this list and be sure to read carefully provide a control mechanism that diminishes any bias intro-
any terms or abbreviations with which you are not familiar. duced in the test procedure.

Biochemical Oxygen Demand (BOD): Determination of the III. Introduction


oxygen requirements wastewater, as related to the biochemi- The Biochemical Oxygen Demand (BOD) determination is a
cal degradation of organic material (carbonaceous demand) test used to determine the oxygen requirements of wastewa-
and the oxygen used to oxidize inorganic (sulfides and ferrous ter. The measurements of oxygen are performed after 5-days
iron) and reduced nitrogen. of incubation to determine the biochemical degradation of or-
ganic (carbonaceous demand) and the oxygen used to oxidize
Carbonaceous Biochemical Oxygen Demand (Carb. BOD): inorganic materials (sulfides and ferrous iron) as well as re-
The BOD test containing a nitrification inhibitor is termed a duced nitrogen. The analytical test procedure is computed by
carbonaceous BOD analytical test. determining the difference between initial and final Dissolved
Oxygen (D.O.) measurements of the wastewater sample over
Glucose / Glutamic Acid: A “standard” check solution to a 5-day period. The sample’s carbonaceous BOD will be less
determine if the distilled water is contaminated. than or equal to the BOD result.
Historically, it has been found that the nitrogenous demand
KHP High and Low Standard: A “standard” Phosphate interferes with determining the carbonaceous demand. The
check solution to determine if the Dilution water is contami- BOD measurement is therefore generally not useful in assess-
nated. ing the oxygen needed to satisfy the carbonaceous demand.

2 www.WaterWorldCE.com
To adjust the BOD to accurately determine the biochemical mainly used to balance the influent wastewater BOD load and
degradation of only the carbonaceous demand, an inhibitor the mass of micro-organisms, termed F/M ratio.
can be added to prevent oxidation of the nitrogen compounds. In this course, the type of wastewater treatment analyti-
The BOD analytical test containing the inhibitor is cal test examined uses a method that determines the oxygen
termed a carbonaceous BOD analytical test. Comparing the demand present after the micro-organisms metabolize and
BOD with the carbonaceous BOD is useful when developing stabilize the wastewater. This wastewater contains many
wastewater treatment trends with relation to time and seasons. pollutants in the form of suspended and dissolved organic
It should be understood that nitrogenous demand can not be compounds. The operator must be familiar with how his/her
determined by subtracting the BOD from the carbonaceous individual wastewater treatment plant is designed and oper-
BOD because the BOD calculation includes the degradation ates.
of inorganic materials (sulfides and ferrous iron). To deter- In general, wastewater is delivered from homes, public
mine the nitrogen content of the sample (ammonia) an am- entities, businesses and industries through a sewer collection
monia probe is used. system to the wastewater treatment plant (WWTP), termed
Due to the number of factors that may affect the accuracy the influent. The collection system can be a closed system,
and precision, “quality assurance and control” checks are used relying on pumps to push or lift the sewage to the WWTP
to assure the operator that the test results are not influenced by or a gravity system that requires greater than or equal to a
contaminants within the distilled water or the BOD dilution 2% slope or a combination of gravity and pumps. Once the
water. To assure that the human error factor does not cause a wastewater is delivered to the WWTP, the micro-organisms
bias, duplicate analysis is performed. To preserve the sample, in the treatment portion of the WWTP use the wastewater
the wastewater samples must be maintained at or below 4 compounds as a food source to reproduce.
degrees Celsius (refrigeration), no chemical preservative is The food source used first by the micro-organisms is the
used in addition, the test procedure for carbonaceous BOD carbon compounds, Car. BOD. The Carb. BOD in the waste-
and BOD must begin within 6 hours of collection for both water is metabolized by the micro-organisms lowering the
24-Hour composite and Grab samples. Carb. BOD in the effluent. The operator can determine the
The purpose of a wastewater treatment plant is to take percent of Carb. BOD removal the WWTP has achieved by
in polluted water and remove the pollutants. One type of comparing the influent Carb. BOD to the effluent Carb. BOD.
wastewater treatment uses micro-organisms. The micro- The lower the oxygen demand found in the effluent the more
organisms biologically convert and stabilize the pollutants, percent removal. In addition, low Carb. BOD results indicate
removing them from the water. The treated water may then that the quality of the effluent is good and will not deterio-
be discharged into a stream without deteriorating the receiv- rate the stream. After the Carb. BOD is lowered to less than
ing stream. To assure that the receiving stream is not being 10 mg/L; other organic compounds are metabolized by the
adversely affected, the operator tests the treated discharged micro-organisms.
water, effluent, for the biochemical oxygen demand (BOD). Once the operator has achieved a low Car. BOD result in
The BOD analytical test is performed on the wastewater the effluent, the challenge is to maintain it at this low con-
treatment plant influent and effluent. The tests are performed centration. To maintain the food with the number of micro-
in a laboratory and their values are expressed in mg/l units organisms, the operator must “balance” both variables. To
(ppm). The results of these tests indicate the affect the effluent determine the balance, the operator divides the influent BOD
will have on the receiving stream. If the BOD of the effluent is by the mass of micro-organisms, measured as suspended sol-
found to be high than that means the dissolved oxygen avail- ids. This ratio is called the F/M Ratio.
able for the stream biota is limited. Limited dissolved oxygen
causes impairment to the fish and macro-invertebrate habitat. IV. Procedure
A high BOD causes excessive oxygen demands on the stream
which can cause the death, elimination or deformity of the MATERIALS
fish community. In addition, the macro-invertebrates (fish 1. BOD Bottles, stoppers and caps: BOD Buffer Pillow
food) may be eliminated due to the lack of sufficient dissolved 2. Volumetric Pipettes: BOD Siphon
oxygen. The BOD results indicate the oxygen requirements to 3. Graduated Cylinders: DO Meter
metabolize and stabilize all of the organic compounds in the 4. KHP High Standard: DO Probe
wastewater. Due to the interference of different organic com- 5. KHP Low Standard: BOD Bench Sheets
pounds within this test method, the BOD test is modified to 6. Polyseed BOD Capsules: 0-Toulidine
reflect only the carbonaceous oxygen demand ( Carb. BOD). 7. Sodium Sulfite: Acetic Acid
Although both the BOD and Carb. BOD analytical test re- 8. Erlenmeyer flask: Starch
sults indicate the effluent quality; the Carb. BOD can be used 9. Stir Plate: Distilled Water
by the operator to determine treatment efficiency and trend 10. Calcium Chloride: Phosphate Buffer
analysis through out the year. The BOD analytical results are 11. Magnesium Chloride: Ferric Chloride

www.WaterWorldCE.com 3
12. Manganous Sulfate: Alkali Iodide Azide 2. Put in the bottle to be read, wait for it to stabilize,
record the reading in the correct space correspond-
PREPARATION OF MATERIALS ing to the numbered bottle.
A. Glucose / Glutamic Acid - Dry the Glucose and Glutamic 3. Repeat the above being sure to rinse probe with
Acid at 104 degrees Celsius for a minimum of one (1) distilled water between bottles.
hour. Place in a desiccator for a minimum of one/half (½)
hour. Weigh 0.15 grams of each and add to a one (1) liter B. Line up BOD Bottles in rows 1-5 for a quality control
volumetric. Using distilled water, bring the water up to check of the test.
level appropriately (ensuring that the bottom of meniscus 1. Row 1 - (1) distilled water, (1) Glucose/Glutamic
is level with the line). Acid
B. KHP High and Low Standard - Dry KHP for two (2) 2. Row 2 - (1) Nutrient Buffer Solution
hours at 104 degrees Celsius for two (2) hours. Place in 3. Row 3 - (3) Seed Standards
desiccator for one (1) hour. Weigh 0.8503 grams and add 4. Row 4 - Low Standard Control 68 mg/L KHP Stan-
to a one (1) liter volumetric. Using distilled water, bring dard
the water up to level mark appropriately (ensuring that 5. Row 5 - High standard Control 136 mg/L KHP Stan-
the bottom of meniscus is level with the line). dard.

1. High Standard = 136 mg/I. Put 20 mI of KHP solu- a. Row 1 - Perform once for each distilled jug. Put
tion in a one-hundred milliliter (100 ml) volumetric. the BOD bottle full of distilled water first in
Using distilled water, bring the water up to the level the row. To the next bottle add 6 ml Glucose/
mark appropriately (ensuring that the bottom of me- Glutamic Acid, 2 ml seed solution and fill with
niscus is level with the line). nutrient water.
2. Low Standard = 68 mg/l. Put 10 ml of KHP solution
in a one-hundred milliliter (100 ml) volumetric. Us- Read and record IDO.
ing distilled water, bring the water up to the level Stopper bottle, add distilled water on top of the stop-
mark appropriately (ensuring that the bottom of per and cap.
meniscus is level with the line).
b. Row 2 - Fill one (1) B.O.D bottle full of nutrient
C. BOD Nutrient Buffer solution – Using a five (5) gallon water.
jug of distilled water, remove one (1) three-hundred mil-
liliter (300 ml) bottle full of distilled water, then add a Read and record IDO.
BOD Buffer pillow (poly seed) or the nutrient solutions Stopper bottle, add distilled water on top of the stop-
from the storage area and mix the contents thoroughly. per and cap.
See below.
c. Set up three (3) BOD bottles, while using a
Nutrients: Volumetric and Class A pipette add 20 ml of
• 1 ml - Calcium Chloride (Ca Cl) Dissolve 27.5 g seed solution to the first bottle, then fill with
CaCl2 in distilled water and dilute to 1 L. nutrient water, in the second bottle add 15 ml
• 1 ml - Magnesium Sulfate (Mg SO) Dissolve 22.5 g of seed solution, then fill with nutrient water, in
MgSO4*7H2O in distilled water and dilute to 1 L. the third bottle add 10 ml of seed solution, then
• 1 ml - Ferric Chloride (Fe Cl) Dissolve 27.5 g CaCl in fill with nutrient water.
distilled water and dilute to 1 L.
• 1 ml - Phosphate Buffer, Dissolve 8.5 KHP2PO4, Read and record IDO.
21.75 g K2HPO4, 33.4 g Na2HPO4* 7 H20 and 1.7 g
NH4CL in 500 Ml distilled water and dilute to 1 L. Stopper bottle, add distilled water on top of the stop-
per and cap.
D. Seed Solution - Follow the directions on the manufac-
turer’s label; Example: Poly Seed BOD capsules. d. Row 4 - Set up three (3) BOD bottles, while
using a Volumetric and Class A pipette add 2
SETTING UP SEEDED BODS ml seed solution, 25 ml of low KHP standard to
A. To properly read Initial Dissolved Oxygen (IDO) and the first bottle, then fill with nutrient solution,
Final Dissolved Oxygen (FDO): in the second bottle add 2 ml seed solution, 15
1. Turn the D.O. probe off and remove it from the cali- ml of low KHP standard, then fill with nutrient
bration bottle. solution, in the third bottle add 2 ml seed solu-

4 www.WaterWorldCE.com
tion, 10 ml of low KHP standard, then fill with Stopper bottle, add distilled water on top of the stop-
nutrient solution. per and cap.

Read and record IDO. c. Row 8 - Set up three (3) BOD bottles, while
Stopper bottle, add distilled water on top of the top- using a graduated cylinder add the following
per and cap. amount of influent to the bottles,
1) If very clean samples 12, 7.5, and 3 ml.
e. Row 5 - Set up three (3) BOD bottles, while us- 2) If cloudy samples 7.5, 3, and 1 ml.
ing a Volumetric and Class A pipette add 2 ml
seed solution, 15 ml of high KHP standard to while using a wide mouth pipette add two (2) ml of
the first bottle, then fill with nutrient solution, seed solution, then fill with nutrient water.
in the second bottle add 2 ml seed solution, 7.5
ml of high KHP standard, then fill with nutrient Read and record IDO
solution, in the third bottle add 2 ml seed solu- Stopper bottle, add distilled water on top of the stop-
tion, 3 ml of high KHP standard, then fill with per and cap.
nutrient water.
d. Row 9 - Set up three (3) BOD bottles, add 0.16
Read and record IDO. grams of nitrification inhibitor then while using
Stopper bottle, add distilled water on top of the stop- a graduated cylinder add the following amount
per and cap. of influent to the bottles,
1) If very clean samples 12, 7.5, and 3 ml.
C. Line up BOD Bottles in rows 6-9 for influent, effluent 2) If cloudy samples 7.5, 3, and 1 ml.
B.O.D and Carb. BOD analysis the test.
while using a wide mouth pipette add two (2) ml of
1. Row 6 – (3) Effluent samples for BOD analysis seed solution , then fill with nutrient water.
2. Row 7 - (3) Effluent samples for Carb. BOD analysis
3. Row 8 - (3) Influent samples for BOD analysis Read and record IDO
4. Row 9 - (3) Influent samples for Carb. BOD analysis Stopper bottle, add distilled water on top of the
stopper and cap.
a. Row 6 - Set up three (3) BOD bottles, while
using a graduated cylinder add the following READING FINAL D.O.
amount of effluent to the bottles, 1. Turn the D.O. meter on and allow it warm up for at least
15 minutes.
1) If very clean samples 150, 60, and 25 ml. 2. Remove the 5 Day BODs/Carb. BODs out of the incuba-
2) If cloudy samples 100, 50, and 20 ml. tor on the appropriate day. The BOD bottles are situated
in the incubator from the left to the right. Remove them
while using a wide mouth pipette add two (2) ml of in the order that they are placed in the incubator.
seed solution fill with nutrient water.
3. Obtain the BOD work sheets with all of the necessary
Read and record IDO information, and then see the two Example Tables shown
Stopper bottle, add distilled water on top of the stop- here (Example Tables : Rows 1 – 3; Rows 4 – 9).
per and cap. 4. Perform a Winkler Titration to calibrate the DO meter or
if applicable use Air Calibration.
b. Row 7 - Set up three (3) BOD bottles, add 0.16 5. Uncap all of the BOD bottles and place the plastic caps in
grams of nitrification inhibitor then while using the incubator.
a graduated cylinder add the following amount 6. Remove the stopper from the last bottle in the first row in
of effluent to the bottles, measure the Final D.O.
1) If very clean samples 150, 60, and 25 ml. 7. Allow the probe to stabilize in the sample, after stabiliza-
2) If cloudy samples 100, 50, and 20 ml. tion has been achieved, record the value on the Bench
work sheet in the appropriate space and column.
while using a wide mouth pipette add two (2) ml of 8. Remove the probe from the last bottle of the first row and
seed solution, then fill with nutrient water. place it in the middle bottle of the first row.
9. Allow the sample to be read until stabilization has been
Read and record IDO. achieved, and record the result. Proceed with the mea-

www.WaterWorldCE.com 5
surement and the recording of the final D.O. in the first 7. The fourth and fifth row of bottles contains the high and
bottle of the first row. low KHP standards.
10. Rinse the probe off with distilled water and place the a. I.D.O. - F.D.O. = D.O.
probe in the third bottle of the next row. b. D.O. - seed control = D.O. less the seed
11. Allow the stabilization of the D.O. to be achieved and c. D.O. (less the seed) X dilution factor = mg/L RE-
record the reading on the worksheet in the appropriate COV. of Standard
space and column. d. Average the recovery /Standard X 100 = % of Re-
12. Repeat steps described in items 7 - 11 until all of the final covery
D.O.s have been measured and recorded on the BOD
work sheets. In the Example Table the three bottles “recovery” is found
under the column “BOD”. Average the three values 70 + 70 +
Note: It is not necessary to rinse from one bottle to the 67 / 3 = 69 and divide by the standard used to determine %
next, because samples are read from an area of less concentra- Recovery.
tion to an area of higher concentration. However the probe
must be rinsed off before beginning measurements in a new 69/68 X 100 = 101 %
row.
NOTE* Acceptable Recovery range is 80 - 120 %. Repeat the
V. Calculations same steps with the high KHP standards.
1. Aliquot (Alq) = ml of sample.
2. Dilution (Dil.) = 300(BOD bottle volume)/Aliquot of 8. The sixth through ninth rows, calculation = D.O. Less
sample. Seed X Dil. = BOD or Carb. BOD.
3. X = Means to multiply in all listed calculations below.
4. Rows one and two are the distilled water blank and the VI. Quality Assurance and Quality Control
nutrient blank rows. 1. For each batch of nutrient water, one (1) blank distilled is
analyzed.
NOTE * The D.O. on these two rows should not decrease 2. For each wastewater sample three (3) samples are deter-
by more than 0.2 mg/L, if more than 0.2 mg/L decrease mined then averaged.
the test is not valid. 3. For each sample date three (3) nutrient blanks are ana-
lyzed.
5. Glucose/Glutamic Acid = F.D.O. X Dil = mg/L 4. For each sample date three (3) glucose/glutamic bottles
In the Example Table, 4.45 X 50 = 222.5 rounded to 223. are analyzed.
5. One set of KHP Standards are analyzed once (1) per
NOTE * Acceptable range is 200 +/- 37 week.

6. Seed Control = (I.D.O. - F.D.O.) X (dilution factor seed VII. Summary


control). Determining the Carb. BOD of the influent and effluent al-
lows the operator the ability to determine the percent of Carb.
After all three bottles have been calculated average the BOD removal the WWTP has achieved. The lower the oxygen
three for the total seed control to be subtracted from all demand found in the effluent the more percent removal, trans-
seeded samples. versely, the higher the oxygen demand found in the effluent
In the Example Table, (7.10 X 0.1) + (5.25 X 0.133) + (3.60 the less percent removal. To maintain treatment toward the
X 0.2)/3 = 0.710. more efficient low effluent Carb. BOD, the operator needs to
balance the influent BOD with the number of micro-organism,
termed the F/M ratio. A trend analysis of the effluent Carb.
BOD can be established through out the year to assist the
operator with adjusting the range of the F/M ratio.
The F/M Ratio represents the pounds of BOD applied per
day per pound of micro-organisms, measured as suspended
solids in the treatment portion of the wastewater treatment
plant. The operator has control of the micro-organism popula-
tion, but generally not the influent BOD. To control of the level
of treatment the operator increases or decreases the removal
of micro-organisms from the treatment system.

6 www.WaterWorldCE.com
The best F/M value to be used for process control must
be determined for each wastewater treatment plant. There
is no perfect number, but, most wastewater treatment plants
operate best in the 0.2 – 0.5, lb influent BOD/lbs solid inven-
tory per day. The operator must determine the optimum F/M
ratio or the individual wastewater treatment plant to achieve
the lowest Carb. BOD analytical value within the discharged
effluent.
The operator’s goals are to operate the wastewater treat-
ment plant in such a manner to efficiently lower the Carb.
BOD, maintain that efficiency and therefore assure that the
receiving stream is not being adversely affected. Generally,
poorly treated wastewater effluent causes impairment to the
fish and macro-invertebrate habitat through placing a high
demand on the streams dissolved oxygen. In addition, the
macro-invertebrates (fish food) may be eliminated due to the
lack of sufficient dissolved oxygen. Both analytical tests re-
sults, from BOD and Carb. BOD, indicate the effluent quality;
the Carb. BOD can be used by the operator to determine treat-
ment efficiency and trend analysis through out the year. The
BOD analytical results are mainly used to balance the influ-
ent wastewater BOD load and the mass of micro-organisms,
termed F/M ratio.

www.WaterWorldCE.com 7
Online Completion
Use this page to review the questions and choose your answers. Return to www.waterworldce.com and sign in. If you have not previously purchased
the program select it from the “Online Courses” listing and complete the online purchase. Once purchased the exam will be added to your Archives
page where a Take Exam link will be provided. Click on the “Take Exam” link, complete all the program questions and submit your answers. An
immediate grade report will be provided and upon receiving a passing grade (70%) your “Verification Form” will be provided immediately for viewing
and/or printing. Verification Forms can be viewed and/or printed anytime in the future by returning to the site, sign in and return to your Archives Page.

Questions
1. To be valid, a BOD sample dilution
must have a residual dissolved oxygen
(D.O.) of at least 1 mg/L after five
days incubation. True or false?
A. False
B. True

2. All National Pollutant Discharge Elimina-


tion System testing must be performed
within 6 hours of collection for only 24-
hour composite samples. True or false?
A. True
B. False

3. All samples tested for BOD must be


seeded with a commercial seed product
such as Polyseed. True or false?
A. False
B. True

4. In Standard Methods, the BOD


method recommends that the analyst
test a 2% glutamic acid with each batch
of samples. True or false?
A. True
B. False

5. The recommended maximum


storage time for the BOD test
is: Choose the best answer.
A. 24 hours
B. analyze immediately
C. 6 months
D. 6 hours
E. 7 days

6. What is the proper preservative used


for the Biochemical Oxygen Demand
test? Choose the best answer.
A. refrigeration
B. none, analyze immediately
C. add nitric acid to a pH
D. add sulfuric acid to a pH<2
E. add sodium hydroxide

7. A particular sample’s carbonaceous


BOD will be ___________ the BOD
result? Choose the best answer.
A. less than or equal to
B. greater than
C. less than
D. equal to

8 www.WaterWorldCE.com
Biochemical Oxygen Demand and
Carbonaceous Biochemical Oxygen Demand

PROGRAM COMPLETION INFORMATION


If you wish to purchase and complete this activity traditionally (mail or fax) rather than Online, you must provide the information requested below. Please be sure to
select your answers carefully and complete the evaluation information. To receive credit, you must receive a score of 70% or better.
Complete online at: www.WaterWorldCE.com

Name: Title: Specialty:

Address: E-mail:

City: State: ZIP: Country:

Telephone: Home ( ) Office ( ) Lic. Renewal Date:

Requirements for successful completion of the course and to obtain 1 professional development hour (PDH):
• Read the entire course.
• Take the test online at: www.WaterWorldCE.com
• A score of 70% on this test will earn you 1 PDH.
• Optional: Complete course evaluation below and return to:
PennWell, James Laughlin, 1421 S. Sheridan Rd., Tulsa, OK 74112 or jamesl@pennwell.com.
• Payment of $ will be required to take the test. Charges on your statement will show up as: PennWell

Course Evaluation
Please evaluate this course by responding to the following statements, using a scale of Excellent = 5 to Poor = 0.
1. Please rate the course’s effectiveness. 5 4 3 2 1 0
2. Was the overall administration of the course effective? 5 4 3 2 1 0
3. Do you feel that the references were adequate? Yes No
4. Would you participate in a similar program on a different topic? Yes No
5. If any of the test questions were unclear or ambiguous, please list them. ________________________________________________________
6. Was there any subject matter you found confusing? Please describe. ________________________________________________________
7. What additional power education topics would you like to see? ________________________________________________________

PAYMENT & CREDIT INFORMATION Answer Form


Examination Fee: $ Credit Hours: 1 Please check the correct box for each question below.

Should you have additional questions, please contact 1. ❑ A ❑ B ❑ C ❑ D 11. ❑ A ❑B ❑C ❑D


2. ❑ A ❑ B ❑ C ❑ D 12. ❑ A ❑B ❑C ❑D
James Laughlin, (918) 832-6320 (Mon-Fri 9:00 am-5:00 pm CST).
3. ❑ A ❑ B ❑ C ❑ D 13. ❑ A ❑B ❑C ❑D
❑ I have enclosed a check or money order.  ❑ I am using a credit card.
4. ❑ A ❑ B ❑ C ❑ D 14. ❑ A ❑B ❑C ❑D
My Credit Card information is provided below. 5. ❑ A ❑ B ❑ C ❑ D 15. ❑ A ❑B ❑C ❑D
❑ American Express  ❑ Visa  ❑ MC  ❑ Discover 6. ❑ A ❑ B ❑ C ❑ D 16. ❑ A ❑B ❑C ❑D
Please provide the following (please print clearly): 7. ❑ A ❑ B ❑ C ❑ D 17. ❑ A ❑B ❑C ❑D
Exact Name on Credit Card 8. ❑ A ❑ B ❑ C ❑ D 18. ❑ A ❑B ❑C ❑D
9. ❑ A ❑ B ❑ C ❑ D 19. ❑ A ❑B ❑C ❑D
Credit Card # Expiration Date
10. ❑ A ❑ B ❑ C ❑ D 20. ❑ A ❑B ❑C ❑D
Signature

SPONSOR/PROVIDER COURSE EVALUATION and PARTICIPANT FEEDBACK


All content has been derived from references listed, and or the opinions of WaterWorldCE faculty. Please direct all questions pertaining to We encourage participant feedback pertaining to all courses. Please be sure to complete the survey included with the course and mail to:
PennWell or the administration of this course to James Laughlin, 1421 S. Sheridan Rd., Tulsa, OK 74112 or jamesl@pennwell.com. jamesl@pennwell.com.
EDUCATIONAL DISCLAIMER RECORD KEEPING
The opinions of efficacy or perceived value of any products or companies mentioned in this course and expressed herein are those of the PennWell maintains records of your successful completion of any exam. Please contact our offices for a copy of your professional
author(s) of the course and do not necessarily reflect those of PennWell. Completing a single professional development course does not provide development hours report. This report, which will list all credits earned to date, will be generated and mailed to you within five business days
enough information to give the participant the feeling that s/he is an expert in the field related to the course topic. It is a combination of many of receipt.  © 2010 by the PennWell Corporation
educational courses and on-the-job experience that allows the participant to develop skills and expertise.

www.WaterWorldCE.com 9

S-ar putea să vă placă și