Calibration certificate

Model: TeraOhm 10 kV - Ml 3200 Date: 24.11.2009


Serial No.: 09340098 Performed By:Franci Ogrin Signature:
Date Placed In Service: Due Date:
Metrel Recommended Cal Interval: 12 months
* The due date may be established (by the customer) by adding the "Recommended Cal Interval" to
the "Date Placed In Service."
No. Function / Ubat=7,4V Reference Low limit Reading Uncertainty High limit
1 Outlook, battery indication, keys and serial communication PASS
2.1 Voltage 10V 6V 10 1V 14V
AC Voltage (50 Hz) 500V 481V 502 4V 519V
2.2 Insulation resistance
Test voltage measured with V-meter Ri = 100

UN = 500V 535V 517V 540 3V 553V

UN = 1000V 1058 V 1025V 1067 5V 1091 V


UN = 2500 V 2620 V 2542V 2647 13V 2698 V

UN = 5000 V 5220 V 5067V 5255 25V 5373 V

UN = 10000 V 10,4kV 10,1 3 kV 10,5 0,1 kV 10,7kV

2.3 Short current 5,0 mA 4,68 mA 4,97 0,02 mA 5,22 mA

2.4 Insulation resistance UN=500V 100GQ 94,7 G 99,8 2GQ 108GQ

UN= I OOOV Okfl OkQ 0 1 kQ 3kQ


100k Q. 92 kQ 100 1 kQ 108 kQ
1 MQ 0,947 MQ 0,997 0,005 M Q 1.080MQ
100 MQ 94,7 MQ 99,8 1 MQ 108 MQ
3GQ 2,82 GQ 2,84 0,04 GQ 3.18GQ
50 G Q. 47,2 G Q 49,9 0,6 GQ 52,8 GQ
500 G ii 4/2 Gii ~f$F GGi" 528 G Q
UN=10000V 3,13 TQ 2,63 TQ 3,13 0,2 TQ 3,63 TQ
open circuit >10TQ PASS
2.5 Generator capability at Ubat = 7,4 V, UN = 10 kV, R = 10 M Q.
Test voltage >10 kV 10 2 kV PASS 0,1 kV

2.6 Capacitor measurement


U N =1000V, 4,8//F 4,55 /jF 4,95 0,05/;F 5,05 //F
3 Withstanding voltage test AC 10,7kV PASS

All results in accordance with technical specification.


Reference instruments:
No. Instrument Type Certificate No. Due
1 Calibration System 9100, Wavetek 09C00876 31.7.2010

20 751 061

METREL d.d. Tel.: +386 1 7558 200


METREL Ljubljanska c. 77 Fax.: +386 1 7549 095
Metrel Certificate of
Calibration Practices
The Metrel Quality System is certified by BVQI
according to Quality standard ISO 9001 : 2000,
Certificate No. 137295. The Metrel hereby certifies that
the above instrument was calibrated in accordance with
applicable Metrel calibration procedures during the
manufacturing process. These processes are ISO 9001
controlled and are designed to assure that the
instrument will meet its published specification.

The Metrel further certifies that the measurement


standards and instruments used during the calibration of
this instrument are traceable to the (inter)national
standards.
The policies and procedures at this facility comply with
EN ISO/IEC 17025.
Environmental conditions: Temperature: 23 °C ± 5 °C