Sunteți pe pagina 1din 8

Chemistry

MASS SPECTROMETER - MASS SPECTROSCOPY

Introduction to the Mass Spectrometer

The mass spectrometer is an instrument by which you can separate


ionised/charged (+) particles of different mass and determine the amounts
of each particle in a mixture.

The technique is called mass spectroscopy or mass spectrometry ('mass-spec'


and 'MS' in shorthand!).

Mass spectrometry gives accurate


information on the relative masses of
isotopes and their relative abundance
(proportions).

Mass spectrometry is an important


method of analysis in chemistry and can
be used to identify elements by their
characteristic mass spectrum pattern -
the technique is used in planetary space probes e.g. mass spectrometer
instrumentation is incorporated in the Mars explorer vehicles.

The substance to be analysed is introduced/injected into a high vacuum


(extremely low pressure) tube system (at K left diagram) where the particles are
ionised by colliding with beam of high speed electrons (at Q in left diagram).

Note: If the sample is not already a gas, then substance must be vapourised, i.e.
the material must be in the gaseous state to be analysed in a mass
spectrometer.

The resulting (+) ions are accelerated down a tube (from + to - plates, P in left
diagram) and then through a powerful magnetic field.

The charged or ionised particles are deflected by this powerful magnetic field (R in
left diagram).

How much they are deflected depends on the particle mass and the speed of
the particle and the strength of a magnetic field i.e. lighter particles of lower mass
(and momentum) are deflected more than heavier particles of bigger mass (see
right diagram below) for a given set of conditions.

By varying the strength of the magnetic field, it is possible to bring into focus
onto an ion detector (N in left diagram) at the end of the tube (effectively an
electrical event is detected), every possible mass in turn and a measure the
strength of the ion current, which is a measure of how much of that ion has
been formed from the sample under analysis.

A simplified diagram of a mass spectrometer tube system is shown below (left)


with further explanation as to what is going on and an extra diagram to show the
relative paths of light to heavy ions for a given strength of magnetic field.

KEY TO DIAGRAM and more detail of each component's function

K = sample injection point, it must be a gas, so a liquid/solid must be vaporised at


the injection point.

IONISATION

Q = high voltage (high +/- p.d.) electron gun which fires a beam of high
speed/energy electrons from a heated 'metal element' into the vaporised sample
under analysis and causes ionization of the atoms (or molecules) forming
positive ions (mainly monopositive in charge).

The collision of high KE electrons with atoms or molecules causes another electron
to be knocked off the particle leaving a negative deficit i.e. a positively charged
particle is formed e.g.

M(g) + e- ==> M+(g) + 2e-, usually written as just

M(g) ==> M+(g) + e- (M might represent e.g. a metal atom or a molecule)

The ions formed should be written as [M]+, a notation that is handy if you are
dealing with ionised molecule fragments with an overall single positive charge e.g.
[CH3]+ is seen in the mass spectrum of methane gas, CH4.

The low pressure (~vacuum) is needed to prevent the ions from colliding with air
particles which would stop them reaching the ion detector system.

ACCELERATION

P = are negative plates which accelerate the positive ions down the tube (there
are positive plates at the start of the tube). A moving beam of charged particles
creates a magnetic field around itself, and this 'ion beam' magnetic field interacts
with the magnetic field at R.

DEFLECTION-SEPARATION

R = the magnetic field that causes deflection of ions, this is can be varied to
change the extent of deflection for a given mass and to focus a beam of ions of
particular mass down onto the detector. Hence, by programming the mass
spectrometer to 'sweep' through all likely particle masses, in terms of the right hand
diagram, you can increase the strength of the magnetic field to bring into focus
onto the ion detector monopositive ions of increasing mass.

DETECTION

N = an ion detection system which essentially generates a tiny electrical current


when the ions hit it. The strengths of the 'electronic' signals from the various ion
peaks are sent to a computer for analysis, computation and display. They tell you
the particle masses present and their relative abundance (see the mass spectrum
diagram for the element strontium below).

MASS SPECTRUM

The resulting record of the ion peaks is called the mass spectrum or mass
spectra. The highest peak is called the base peak and is often given the relative
and arbitrary value of 100, particularly in the mass spectra of organic
compounds.

MASS SPECTRA

For elements you get a series of signals or ion peaks for each isotope present and
the ratio of peak heights gives you the relative proportion of each isotope in the
element so that you can calculate the relative atomic mass of an element. This
'simple' spectra of mononuclear ions like Sr+ is only true for non-molecular
elements like metals (see mass spectrum of strontium diagram below) or
noble gases, but for molecular elements like nitrogen or the halogens things are not
so simple (see chlorine example below).

For larger e.g. organic molecules, things can be very complex indeed, as molecules
fragment and many different ions can be formed BUT you can get the relative
molecular mass of a molecule by identifying what is called the molecular ion
peak, that is, when one electron is knocked of the molecule but the molecule
retains its full molecular structure.
e.g. benzoic acid (Mr = 122) gives a molecular ion peak of m/e = 122, due
to [C6H5COOH]+

but you also get fragments such as [C6H5]+ with an m/e of 77 as the molecule
breaks up on further electron impact.

CHLORINE EXAMPLE

Chlorine is a good example of a molecular element whose mass spectra can


be a bit tricky when first encountered ...
37
Chlorine consists of two principal isotopes, chlorine-37 (25% is Cl) and
chlorine-35 (75% is 35Cl).

BUT, chlorine consists of Cl2 diatomic molecules, which, on ionisation, can split into
chlorine atoms.

The result is a series of 5 different mass peaks from the various isotopic atomic or
molecular ion possibilities.

1. [37Cl37Cl]+ or [37Cl2]+ m/z = 74

2. [37Cl35Cl]+ m/z = 72 (note that you must show the two isotopes separately)

3. [35Cl35Cl]+ or [35Cl2]+ m/z=70

4. [37Cl]+ m/z=37

5. [35Cl]+ m/z=35

m/z means the relative mass of the ion over its charge, which for our purposes the
charge is taken +1 (little z) and the mass (little m) is the relative atomic/formula
mass of the particle. m/z sometimes quoted as m/e. You should write the structure
of the ion in square brackets and put the charge on the outside of them in the top
right - this is an important and universally accepted notation in mass spectrometry.

The mass spectra of organic compounds can be very complex as the molecules
fragment under electron bombardment, but the resulting mass spectra can used
to identify compounds from their 'finger-print' pattern of ion peaks of different
mass and particular proportions for a given set of experimental conditions.

The largest m/z value gives the molecular mass of a molecule, i.e. the ion of
largest mass, prior to fragmentation, is formed when the original whole and neutral
molecule, loses one electron e.g. for ethane it would be due to the formation of
[C2H6]+, m/z = 30 and is called the molecular ion peak.

STRONTIUM EXAMPLE

A 'simple' element spectrum to interpret AND a subsequent relative


atomic mass calculation based on the mass spectrum of the element
strontium

The relative atomic mass of an element, Ar, is the weighted average mass
of the isotopes present, compared to 1/12th of the relative mass of the carbon-
12 isotope. [ 12C is given the relative mass value of 12.0000 ]

Quite often the highest m/e peak is arbitrarily given the relative value of 100, as in
this case and referred to as the base peak, but the peak lines might well indicate
% abundance of isotopes. The diagram of abundances is sometimes called a stick
diagram.

Relative peak height = relative abundance as measured from the ion current
detector signal.

The mass spectrum shows strontium consists of four isotopes, 84Sr (peak height =
0.68), 86Sr (peak height = 12.0),87Sr (peak height = 8.47) and 88Sr (peak height =
100.0)

The sum of the heights = 0.68 + 12.0 + 8.47 + 100.0 = 121.15


So we can now calculate the weighted average mass of ALL the isotopes.

Therefore Ar = {(0.68 x 84) + (12.0 x 86) + (8.47 x 87) + (100.0 x 88)}/121.15 =


87.7

The book value is 87.62, BUT this calculation does NOT take into account the very
accurate relative atomic masses based on the carbon-12 scale, it merely uses the
mass numbers, which are always integer.

ISOTOPIC MASSES - definition and uses

Very accurate isotopic masses are usually a tiny fraction different from a whole
number but provide invaluable information.

Modern mass spectrometers are exceedingly accurate and very sophisticated


instruments and can measure mass to at least 4 decimal places. They can readily
distinguish between N2, CO and C2H4 molecules, all with an integer Mr of 28.

The very accurate molecular ion masses are [N2]+ = 28.0061, [CO]+ = 27.9949 and
[C2H4]+ = 28.0313

A very accurate mass spectrometer (for high resolution mass spectroscopy) can
even differentiate between organic molecules of the same integer molecular mass.

e.g. for the molecular mass 103, some possible, however unlikely, molecular
formulae could theoretically be

C5HN3 = 103.0170, C3H5NO3 = 103.0269, C2H5N3O2 = 103.0382, C7H5N =


103.0427, CH5N5O = 103.0494

The Time of Flight Mass Spectrometer (for advanced level students only!)

Ion mass separation using a time-of-flight mass spectrometer - a more


modern instrument
• In a time-of-flight mass spectrometer the ions are formed in a similar manner
by electron bombardment, and the resulting ions accelerated between
electrically charged plates.

• Again, the sample must be a gas or vapourised and is bombarded with an


electron beam or laser beam to knock off electrons to produce positive ions.

• However, the method of separation due to different m/e


(mass/charge) values is then dependent on how long it takes the ion
to travel in the drift region' i.e. the region NOT under the influence of an
accelerating electric field.

• The ions are accelerated in the same way between positive to


negative plates in an electric field of fixed strength i.e. constant
potential difference.

• The smaller the mass of the ionised particle (ionized atom, fragment
or whole molecule) the shorter the time of flight in the drift region
where no electric field operates.

• This is because for a given accelerating potential difference, a lighter


particle is accelerated more to a higher speed than a heavier ion, so
the 'time of flight' down the tube is shorter.

• Therefore the ions are distinguished by different flight times NOT by


different masses being brought into focus with a magnetic field as described
in section 4a BUT the separation by time of flight is still determined by the
m/e (m/z) value of the ion.

• The general principles of the separation are required knowledge but the
mathematics is NOT needed by A level students, but if you are
interested, a simplified summary is given below

o t = Kinst√(m/q)

o t = time of flight, m = mass of ion, q = charge on ion,

o Kinst = a proportionality constant based on the instrument settings and


characteristics e.g. the electric field strength, length of analysing tube
etc.

o Therefore t is proportional to the square root of the mass of the


ion for particles carrying the same charge - the bigger the mass the
longer the 'flight time'.

o The first equation is derived partly from the extra mathematics


outlined below.
o KE = qV, the kinetic energy imparted to the ion is given by its charge
x the potential difference of the accelerating electric field.

o The acceleration, for a fixed electric field, results in an ion having the
same kinetic energy (KE) as any other ion of the same charge q but
the velocity v of the ion depends on the m/e (m/z) value.

o v = d/t (or t = d/v), where v = velocity of accelerated particle in the


drift region, d = length of tube in the drift region. (or t = d/v)

o KE = 1/2mv2, so the bigger m, the smaller is v in the drift region


and hence the basis of detecting ions of different mass by different
'flight times'.

o The diagram makes the method look simple, but far from it, the
instrument works in a pulsed manner i.e. pulsed electric field, and
some pretty sophisticated electronics are used to analyse the signals
from the detector and the software calculates the mass of the ion
based on the drift flight time.

S-ar putea să vă placă și