Sunteți pe pagina 1din 4

An Alternative to Silt Density Index (SDI)…..

Continuous Particle Counting


Author: Robert L. Bryant, Chemtrac Systems, Inc.

O ne of the biggest pro-


blems experienced
with reverse osmosis
(RO) systems is the need for
frequent chemical cleaning and/or
membrane replacement caused by
Background
Colloidal (particulate) fouling of r e-
verse osmosis membranes is often
traced back to the pretreatment
system. These problems are
sometimes the result of inadequate
design, or improper operation.
the rapid plugging caused by
smaller particles.
A typical warranty requirement for
RO systems is an SDI of 5.0 for
spiral wound membranes. Al-
though the test is published in the
American Society for Testing and
colloidal fouling. These colloids Inadequate design may be due to Materials (ASTM) manual several
also affect daily operations as seen an inaccurate analysis of the water “versions” have evolved to meet
by decreasing product flow over source, a change in the source, or individual pretreatment system
time, decreasing salt rejection, and increased product water needs conditions. Other names have
increasing differential in membrane without added pretreatment been associated with the test, such
feed versus reject pressure. They equipment. Oftentimes, fouling as “plugging index”, “silting index”,
may cause membrane failure be- may occur simply due to the and “plugging factor ”. There is
cause of uneven flow distribution inadequate monitoring and control sufficient data to verify that the test
and concentration phenomena of upstream processes such as provides a good correlation with
within the cartridge (1). clarifiers, filters, etc. Membrane changes in RO performance, at
Silt Density Index (SDI) has been throughput is directly affected by least for treatment of a particular
the standard test to predict mem- deposits on the feedwater side. water source. This correlation does
brane fouling for many years. The Production can be maintained, to a not always hold true when com-
method (ASTM D-4189) has been point, by increasing feed pressure, paring different waters and different
accepted worldwide. It has been a but operating costs increase, and sites (2).
useful test, and often shows rea- membrane damage can occur. The standard procedure uses a
sonable correlation with variations Before the maximum operating 47-millimeter diameter mixed cell-
in RO performance attributable to pressure is reached, membranes ulose esters filter with 0.45-µm pore
colloidal fouling. However, there must be cleaned, either in place, or size. Smaller diameter filters can
are some limitations to the test, and sent to another facility. If the plant be used as long as the collected
"variances" in plant operations can does not have “excess” capacity, volumes are adjusted in proportion
affect reliability, repeatability, and/ mobile units must be brought on to the filter area. Smaller pore
or operational usefulness. site to maintain required pro- sizes, such as 0.1 µm, are used if
duction. This is a very expensive the SDI of a RO effluent is desired.
1. SDI is not continuous, on-line process. Some operators have experienced
measurement. variability in the test results on
2. Repeatability can be affected SDI identical samples. The “cross flow”
by differences in operator Silt Density Index is a measure- fluid dynamics in an operating RO
technique. ment, or “relative index” of the are different from the “dead end”
3. T e s t r e s u l t s c a n b e v e r y amount of particulates in water. filtration on a SDI filter paper. Also,
“scattered depending on up- The test consists of flowing a water concentrations of ionic, organic,
stream equipment operations sample through a 0.45 micron (µm) and biological materials on an RO
at time/place SDI test is done. membrane filter at a constant 30 membrane surface are very diff-
Based on these limitations and pounds per square inch gauge erent than on the SDI filter surface.
variances, an alternative to SDI, or (psig) pressure. As the filter plugs, The interactions that can contribute
a “supplemental” method would be the unit measures the decrease in to fouling are complex, and not well
desirable. Preferably, a method flow versus time. There is usually a defined. Some procedures advise
that was on-line, continuous, had close relationship between the SDI that two or more tests be ran to
“multipoint” sampling capability, number and the rate at which confirm results, and if there is a
required no operator interven- colloidal and particulate fouling large difference, to run a third.
tion……and, of course, was cost occurs. It is assumed that the test These multiple tests can be dis-
effective. On-line Particle Counters, is mostly a measurement of couraging to an operator because
and Particle Monitors have demon- colloidal concentration because of the time required and deciding
strated the capability to meet these large particulates affect the SDI the which result to believe. Some
needs. least when they are compared to users have found a significant diff-
-erence in SDI results due to var- ducts are limited by federal reg- er than a turbidimeter. In fact, for a
iances in filter paper properties. ulations from exceeding specific 5.0-µm particle, the PM can detect
Thus, values obtained with filters numbers of certain size particles concentrations of less than 10p/mL
from different manufacturers may Large volume parentrals (LVP) may (3).
not be comparable. A particle’s contain more than 5 p/mL larger There are several factors that en-
surface charge can have a signifi- than 25 µm, and not more 50 larger able the PM to be a very effective
cant effect on whether it passes than 10 µm. tool for operating the plant.
through the pore, or adheres and There is a size range of particles in
plugs the filter. Temperature aff- water, which is difficult to remove in 1. The useable measurement is not
ects water viscosity, which changes a multimedia filter. These particles affected by particle size, shape, color,
the flowrate through the filter. All of follow stream lines of the water flow. or composition. It gives an “index” of
these “variables” affect the SDI Because their inertia is low, they do the number of particles per unit
value. not settle with gravity as their mass volume. The sensor is extremely
is small, and they are too large to be sensitive to any particle that passes
through the sample cell greater than 1
Particle Detection Instruments affected by molecular (Brownian)
µm.
Turbidimeters: “Light scattering” forces. This size range is between
2. There is no calibration requirement.
turbidimeters (nephelometers) have 2.0 to 10.0 µm, which is above the 3. The cost of a PM is lower than a
been the traditional on-line instru- size range where turbidimeters are standard particle counter.
ments used for monitoring/control- most effective. 4. Maintenance is accomplished by
ling the pretreatment systems of Particle counters measure the size simply replacing the sample tubing.
clarifiers, media filters, and car bon and concentration of particles using The sample does not contact any optic
filters. However, as the demand for “light blocking” methods with a laser surfaces, which eliminates signal drift
cleaner water increases, the need diode as the light source. This is caused by fouling.
grows for more sensitive instru- important information because it can
ments. Turbidimeters do not pro- be used to determine the effect- A Case History
vide the required sensitivity. In iveness of different type filters. For An electric generating station in-
fact, water with a turbidity of 0.5 example, if 5.0 µm particles are stalled an RO system in an effort to
Nephelometric Turbidity Units passing through a cartridge filter reduce demineralizer regenerations,
(NTU) can contain hundreds of with a pore size rating of 1.0 µm, and reduce wastewater disposal
particles per milliliter (p/mL). there must be an internal problem, problems. The pretreatment system
Turbidity is defined as “an expres- or possibly the filter has been mis- is shown in Figure 1. The water
sion of the optical property that labeled. source is a small river that has
causes light to be scattered and rapidly changing turbidity. An inter-
absorbed, rather than transmitted in Particle monitors: The particle mediate settling/storage basin mini-
straight lines through the sample”. monitor (PM) uses a unique “light mizes large turbidity swings in the
Simply stated, turbidity is the inter- fluctuation” technique that com- clarifier inlet. The primary coagulant
action between light and suspend- bines features and benefits of the was originally alum, but was chan-
ed particles in liquid. It is not a truly turbidimeter and particle counter. It ged to a polymer/inorganic blend.
quantitative value, but more of a is an extremely sensitive, but simple Turbidimeters were used to monitor
relative value, or an “index” of device that gives a “relative” index of clarifier and filter effluents. The RO
water quality. An important pro- water cleanliness. The PM measur- system consists of two trains with
perty of NTU measurement is that es fluctuations in intensity of a nar- each having a primary and second-
for a given mass concentration of row light beam (infrared LED) trans- ary set of membranes. Start-up flow
particles, the turbidity is less as mitted through a flowing sample. was 250 gallons per minute (gpm)
particle size increases. This means These fluctuations are caused by per train with a differential pressure
that larger size particles could be in random variations in the number of of 36 psig. SDI tests were initially
a sample, but not contribute ap- particles in the light beam. Since performed on a regular basis, but
preciably to the turbidity reading. only the AC component measures were eventually discontinued be-
particle motion, it can be easily iso- cause of no useable correlation
Particle Counters: On-line, liquid lated and amplified. Turbidimeters between SDI values and cleaning
particle counters have been used produce a DC electrical signal that is frequency.
for many years in the “final product” proportional to the measured para- Almost immediately after start-up
stage of high-purity water systems. meter. This makes the device sus- (1994), the RO trains had to be
Some instruments detect part icles ceptible to “drift” caused by DC cleaned often, usually twice a week,
in the submicron range. For ex- noise, light source variances, and because of high differential pres-
ample, in the microelectronics optical surface fouling. Sensitivity sure. The cost of membrane clean-
industry, particle size measurement tests have shown that for particles ing chemicals, manpower, and waste
down to 0.05 µm is often required. larger than 2.0 µm, the PM has a disposal was excessive.
Injectable, pharmaceutical pro- detection capability 100 times great- Differential pressure also caused
Figure 1. Pretreatment System cartridge filter to monitor inlet and
outlet particulate levels. The particle
Polymer counter tracked the number and size
Coagulant Particle
Counter of particles leaving the 5-µm and 3-
µm filters. A small amount of cat-
Clear Multimedia 5 1
ionic polymer was fed to the multi-
Clarifier
Well Filters Micron Micron media influent to improve removal of
sub-micron particles. Although the
counters/monitors do not detect par-
Particle
ticles below 1 µm, it is well establish-
Particle
SCM Monitor
Counter
ed that colloids are adsorbed on filt-
(3 ea.)
er media that is properly conditioned
with the correct amount of the cor-
rect polymer. A streaming current
Reverse monitor (SCM) was used to optimize
Boiler Feedwater Demineralizers
Osmosis the polymer feed to ensure that the
minimum amount was being used.
Condensate A polymer can also contribute to RO
fouling if it is overfed (4).
Results are shown in Figure 2.
the cartridge filters to be changed ent of each pressure filter. Two filt- Before polymer addition, particle
out often. Each of the 4 vessels ers were shown to be passing a high counts in the filter effluent were
contains 52 elements with each number of particles and were re- more than 120 per mL. After the
element costing approximately moved from service. This change polymer addition, counts dropped to
$100.00. A complete change out forced all of the flow to go through less than 20 per mL and almost (0)
could cost up to $21,000. Over the three filters, which matched the in the cartridge filter effluent. The
next several years, attempts were design capacity for multimedia. polymer dose was approximately 0.4
made to solve the problems: Particle removal improved 50%. parts per million (ppm) and an im-
A program was then established to mediate improvement was seen in
1. A particle “charge-modifying use particle counters and particle the RO operation. Cleaning fre-
device” that (reportedly) prevents monitors to continuously measure quency went from once per week to
particulates from building up on the performance of every system com- once per month. Feedwater pres-
membranes was attempted. There ponent. Particle monitors were in- sure buildup was less than 2.0 psig
was no improvement in membrane stalled on the effluents of the clarifier after 30 days.
performance. clearwell, first and second stage The particle counters and monitors
2. Different membrane manufact- multimedia filters. These strategic also provided a couple of unexpect-
urers were evaluated. A new mem- installations allowed observation of ed benefits. Operators found that
brane was discovered that gave more how backwashes and flow changes changing filter backwash procedures
flow with less fouling and lower
affected filter performance. The reduced particulates in the effluent.
differential pressure.
particle counter was installed at the This finding was not detectable us-
3. Anthracite was replaced with multi-
media (anthracite, sand, garnet,
quartz) to get better removal of small Figure 2. Treatment Results
particulates. There was no “measur-
able” improvement in water quality
going to the cartridge filters or the
performance of the RO.
4. Different brands of cartridge filters
were evaluated. Seven (7) brands
gave varying results in performance
and cost.
5. A chemical technology company
was contacted to assess the entire
coagulation/filtration process using on-
line monitoring and control instru-
ments….

In October 2003, a supplier* was


contacted to evaluate the coagula-
tion/filtration process. A particle
counter was used to check the efflu-
ing turbidimeters. It was also dis-
covered through the particle counter
that the cartridge filter supplier had
mistakenly shipped 3.0-µm filters
instead of 1.0-µm filters. Currentl y,
the plant is controlling multimedia
filter operations with particle moni-
tors and monitoring membrane feed-
water with the particle counter.

Conclusion
The return on investment (ROI) of
using sensitive and reliable on-line
particle counters and particle moni-
tors in high-purity water pretreat-
ment systems can be substantial.
Operators no longer have to rely on
SDI “spot checks” to operate the
plant. Savings in chemicals, filter
replacements, and membrane clean-
ing/replacement can be quantified.
Economic losses can be avoided in
microelectronics and pharmaceutical
production because of water assoc-
iated problems. Boiler feedwater
quality and boiler reliability can be
ensured.

References

1. Paul, D.H. “Pretreatment Workshop”,


ULTRAPURE WATER Expo 1996,
Chicago, Ill. (May 1-3, 1996).

2. Kaakinen, J.W. “SDI Instrmentation To


Estimate the Fouling Potential of RO
Feedwater”, ULTRAPURE WATER.
11(5), pp. 42-54 (July/August 1994)

3. Gregory, J. “A Simple Particle Monitor


for Low Turbidity Waters”, AWWA
Water Quality Technology Confer-
ence, St. Louis, Mo. (1988).

4. May, L.M.; Carlson W.M. “Effects of


Coagulant Overdose in Clarifiers on
Downstream Ion Exchange Process”,
International Water Conference,
Pittsburgh, Pa. (Nov. 4-7, 1985)

Endnote
*Chemtrac Systems, Inc. of Norcross, GA.,
is the supplier mentioned in the text.

Author Robert L. Bryant is the president of


Chemtrac Systems, Inc. He began his
career with Nalco Chemical in 1969 and
introduced streaming current technology to
the water treatment industry in 1982. Mr.
Bryant founded Chemtrac Systems Inc. in
1985. He has written more than 20 articles
for technical publications and Chemtrac
holds several patents on water instrument-
ation devices.

S-ar putea să vă placă și