Differential Pressure Transmitters Handheld Pressure Gauges Pressure Calibration Systems Absolute Pressure Gauges
Properties of pressure gauges Differential pressure transmitters made by halstrup-walcher GmbH have been designed for non-aggressive, gaseous media. These gauges work according to an inductive measurement principle whereby an inductive displacement transmitter measures the deflection of a beryllium bronze diaphragm without making contact. The diaphragm is situated between two measurement chambers, thereby making it possible to record both positive and negative differential pressures. The measuring cell has no frictional parts or parts subject to mechanical wear. Beryllium bronze is a highly elastic material that is stabile for long periods of time, behaves well under a variety of temperature conditions and is extremely resistant to hysteresis. As a result, this technology can be used to create high-quality pressure gauges capable of taking measurements at pressures as low as a few Pa. halstrup-walcher utilizes three different measuring systems: Due to the excellent linearity afforded by its design, the linear variable differential transformer (LVDT) is primarily used for pressure calibration devices. The dual inductive pickup system sends a differential signal that is linearized by an electronic analysis unit. This system has been slated for use in manufacturing highquality differential pressure transmitters and hand-held pressure gauges. For basis applications will be used a piezoresistive precision cell (PS 27). Advantages perfect for positive or negative differential pressures and for either symmetrical or non-symmetrical measuring ranges devices can be calibrated especially suitable for very small measuring ranges available with a variety of different display unit options calibration certificates available in German or English from either the factory or from the German Calibration Service (DKD)
Type
PUC 24 PUC 28 P 26 P 29 P 82 R
Description
| Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter |
Special features
| Process monitoring device for clean rooms | Process monitoring device | Scalable differential pressure transmitter | For flammable gases | Pressure transmitter with root-extracted | output for measuring volume flow | For standard applications | In two-wire technology | Excellent price/performance ratio | Basic sensor for standard applications | With two limit switches | in the control panel housing | Portable, digital pressure gauge | with min./max. value memory | Portable, digital gauge for absolute pressures | with min./max. value memory | Provides highly accurate measurements | Portable calibration device | Portable, with integrated pressure generation | Absolute pressure transmitter | Digital precision barometer
Page
| 45 | 67 | 89 | 10 11 | | 12 13 | 14 15 | 14 15 | 16 17 | 18 19 | | 20 21 | | 22 23 | | 24 25 | 26 27 | 28 29 | 30 31 | 32 33 | 34 35 | | | | | | | | | | | | | | | | | | | | |
| Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter | Differential pressure transmitter |
EMA 200
EMA 210
| Hand-held pressure gauge | Pressure calibration device | Pressure calibration device | Absolute pressure transmitter | Absolute pressure transmitter
| 36 | 37
| |
www.halstrup-walcher.com
03.2011
Operating elements
PUC24
Connection diagram
Profibus DP M12
Supply voltage
RH
humidity transducer
display C display %RH microcontroller display Pa relay outputs audible signal BUS interface (optional) input console
power adapter
Profibus DP
D-sub socket (female connector)
2 3 5
6 9
1 2 3 4 5 6 7 8 9
A/D converter
2 3 5
Differential pressure transmitter PUC 24 Process monitoring device for clean rooms
Special features particularly suitable for use in clean rooms inputs for humidity and temperature sensors stable measurements thanks to cyclical self-calibration of the zero point (differential pressure) alarm thresholds (switching contacts) graphic LC display Profibus DP interface (optional) solvents resistant stainless steel surface acustic alarm when alarm thresholds are reached, may be reset by push-button bilingual menu (English / German)
Technical data
measurement range measurement ranges margin of error deflection drift / temperature zero point drift / temperature 100 Pa or 250 Pa freely scalable within this range 0.5 % of end value 0.03 % / K (+10 C...+50 C) 0 % (cyclical zero-point correction) data interface none Profibus DP (optional) RS232 (optional) overload capacity medium max. system pressure 200 x air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor bus connection 9-pin Sub D panel plug
(not for wall thickness above 5 mm applycable)
A 0 1 B 0 DP 2 C D DK RK
100 Pa 250 Pa
Sub D plug with 150 mm cable round plug M12 with 150 mm cable
output signals 0 10 V (Ri > 2 k) 0/4 20 mA (RL < 500 ) adjustable 2 contact points, 6 A, 230 VAC, may be configurated as desired within this pressure range
+ 10 C + 50 C 10 C + 70 C
Order Key
power consumption weight pressure ports protection class testing approx. 7 VA
PUC 24
accessories plug with switchable bus terminator DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0104 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
terminal strip
PUC 28
230 0,1
250
Operating elements
PUC 28
pressure display temperature display humidity (RH) display
Connection diagram
Profibus DP M12
Supply voltage
RH
humidity transducer
display C display %RH microcontroller display Pa relay outputs audible signal BUS interface (optional) input console
power adapter
Profibus DP
D-sub socket (female connector)
2 3 5
6 9
1 2 3 4 5 6 7 8 9
A/D converter
2 3 5
Special features inputs for humidity and temperature sensors stable measurements thanks to cyclical self-calibration of the zero point (differential pressure) alarm thresholds (switching contacts) graphic LC display Profibus DP interface (optional) acustic alarm when alarm thresholds are reached, may be reset by push-button bilingual menu (English / German) External tubing connections for calibrating (PUC 28 K optional)
Technical data
typ / measurement range measurement ranges margin of error deflection drift / temperature zero point drift / temperature 100 Pa or 250 Pa freely scalable within this range 0.5% of end value 0.03 % / K (+10 C...+50 C) 0 % (cyclical zero-point correction) PUK 28 100 Pa PUK 28 250 Pa PUK 28 K 100 Pa PUK 28 K 250 Pa data interface none overload capacity medium max. system pressure 200 x air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor Profibus DP (optional) RS232 (optional) bus connection 9-pin Sub D panel plug
(not for wall thickness above 5 mm applycable)
A 0 1 2 3 B 0 DP 2 C D DK RK
Sub D plug with 150 mm cable sensor response time time constants 25 ms 25 ms 40 s (adjustable) round plug M12 with 150 mm cable supply voltage 24 VDC, 10 % smoothed input signal humidity/temperature module (galvanically separated) 0 10 V, Ri = 470 k 0 / 4 20 mA, Ri = 50 adjustable output signals 0 10 V (Ri > 2 k) 0/4 20 mA (RL < 500 ) adjustable 2 contact points, 6 A, 230 VAC, may be configurated as desired within this pressure range
+ 10 C + 50 C 10 C + 70 C
Order Key
A power consumption weight pressure ports protection class testing approx. 7 VA
PUC 28
accessories plug with switchable bus terminator DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0104 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
P 26 Dimension drawing
P 26 with LCD
no LCD
P 26 Configuration software
Connection diagram
D-sub socket (female connector) mating connector (pins)
OUTPUT SIGNAL 1
I
RELAY 1 4 5 6
RELAY 2 7 8 9
POWER SUPPLY 11 12 13
11-12(+) 24 VDC 12-13 115 VAC 11-13 230 VAC 11-13 24 VAC
3
U
GND
PCCONFIGURATION
P 26
Differential pressure transmitter P 26 Intelligent differential pressure transmitter with scalable range
Special features range and display scalable switching contacts with adjustable switching thresholds time constants and output characteristics can be configured (root-extraction / linear) automatic zero-point calibration prevents zero-point drift unit conversion (e. g. mmH2O, mmHg, etc.) integrated valve provides a high level of overpressure protection available with interface RS232 (optional) also for top-hat rail mounting multilingual menu (English, German, Italian, French) measuring ranges
Technical data
output* measurement ranges (others available upon request) 10 / 50 /100 / 250 / 500 Pa 1/ 2.5 / 5 / 10 / 20 / 50 / 100 kPa free scalable from 10..100% within a range 0.5% + 0.3 Pa of scaled range (40...100% of end value) 0.03 % / K (+10 C...+50 C) 0 % (cyclical zero-point correction) 0 ...10 V (RL 2 k) 0...20 mA (RL 500 ) 4...20 mA (RL 500 ) 5 V (RL 2 k)
* output singal selectable
A 1 0 4 5
230/115 VAC
margin of error (0.3 Pa margin of error for reference) deflection drift / temperature zero point drift / temperature
measurement range measurement range e. g., 0 10 Pa, mbar, mmHg, etc. LCD none LCD and buttons for configuration interface
D S 2
600 kPa for measurement ranges 2.5 kPa 200 x for measurement ranges < 2.5 kPa air, all non-aggressive gases 600 kPa for measurement ranges 2.5 kPa 200 x for measurement ranges < 2.5 kPa
E 0 LC
F 0 2
G 0 RS
25 ms 25 ms 60 s (adjustable)
Order key
power consumption weight cable glands pressure ports protection class testing approx. 6 VA approx. 0,75 kg 3 x M 16 for hose NW 6 mm, others available upon request IP 65 CE, CSA, GOST
P 26
accessories RS 232 cable DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0085 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
P 29 Dimension drawing
with LCD
P 29
Flush for an adequate amount of time prior to operation! Warning: Risk of explosion!
no LCD
Flush for an adequate amount of time prior to operation! Warning: Risk of explosion!
P 29
P 29 Configuration software
Connection diagramm
Output Signal 1
I GND
Power Supply 3
U
PCConfiguration
P 29
10
Differential pressure transmitter P 29 Intelligent differential pressure transmitter for flammable gases
Special features
TV-tested, differential pressure transducer for flammables gases does not generate any ATEX zone; design changes and technical modifications keep ignition source and gas safely mixture separated. time constants and output characteristics can be configured (root-extraction / linear) automatic zero-point calibration prevents zero-point drift unit conversion (e.g. mmH2O, mmHg, etc.) integrated valve provides a high level of overpressure protection also for top-hat rail mounting multilingual menu (English, German, Italian, French)
Technical data
output measurement ranges (other available upon request) 250/500 Pa 1/2.5/5/10/20/50/100 kPa free scalable from 10...100% within a range 0.5 % + 0.3 Pa of scaled range (40...100 % of end value) 0,03 %/K (+ 10 C...+ 50 C) 0% (cyclical zero-point correction) measurement range overload capacity medium max. line pressure 100 kPa for measurement ranges 2.5 kPa 200x for measurement ranges < 2.5 kPa flammable gases, all non-aggressive gases 100 kPa measurement range e.g., 0 250 Pa, mbar, mmHg etc. C 0...10 V (RL 2 k) 0...20 mA (RL 500 ) 4...20 mA (RL 500 ) 5 V (RL 2 k) A 1 0 4 5
margin of error (0.3 Pa margin of error for reference) deflection drift / temperature zero point drift / temperature
power supply 24 V DC
B 24 DC
margin of error standard 0.2% (40...100 % of end value) but min. 0.3 Pa
D S 2
E 0 LC
Order key
A
P 29
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate www.halstrup-walcher.com 03.2011 9601.-0003 9601.-0004 9601.-0002
11
P 82 R Dimension drawing
Connection diagram
DC power supply AC power supply analogue outputs
or or
12
Differential pressure transmitter P 82 R Pressure transmitter with root-extracted output for measuring volume flow
Special features highly accurate and stable for long periods very little hysteresis; largely independent of temperature differential pressure p at the measuring orifice is expressed as either a linear (UL,IL) ) or root-extracted function URAD = 10 V x UL, IR = 20 mA x IL or IR = 4 mA + 16 mA x (IL 4 mA)
Technical data
output measurement ranges (others available upon request) margin of error deflection drift / temperature zero point drift / temperature zero point drift / time 10 0/ 250 / 500 Pa 1 / 2.5 / 5 /10 / 20 kPa 1 % of end value 0.04 % / K (+10 C...+50 C) 0.05 % / K (+10 C...+50 C) 0.5 % / year measurement range measurement range in Pa, kPa, mmHg, etc. (e.g., 0 100 Pa) time constants medium max. system pressure air, all non-aggressive gases none 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor 1s 2s 5s sensor response time leak flow suppression 20 ms adjustable from 0 10% of end value LCD none 3 1/2 digit operating temperature storage temperature +10 C ... +60 C 10 C ... +70 C 4 1/2 digit D 0 1 2 5 E 0 3 4 0...10 V (RL 5 k) 0...20 mA (RL 500 ) 4...20 mA (RL 500 ) A 1 0 4 power supply 24 VDC 24 VAC 115 VAC 230 VAC B 24D 24A 115 230 C
overload capacity
5x
power consumption weight cable glands pressure ports protection class testing
Order key
A
P 82 R
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
13
no LCD
Pa
PIZ
DC power supply 420 mA output
or or
14
Special features for positive and negative differential pressures highly accurate and stable for long periods little zero point drift or hysteresis; largely independent of temperature also available as a two-wire system (PIZ model)
Technical data
output signals measurement ranges (others available upon request) margin of error 50 /100 / 250 / 500 Pa 1 / 2.5 / 5 /10 / 20 / 50 /100 kPa 1% of end value, 0.5 % of end value for measurement ranges 250 Pa, 0.2 % of end value for measurement ranges 250 Pa 0.04 % / K (+10 C...+50 C) 0.04 % / K (+10 C...+50 C) 0.5 %/ year 0...10 V (RL 2 k) 0...20 mA (RL 500 ) 4...20 mA (RL 500 ) 4...20 mA two-wire (RL 50 [UB (V) 10 V] ) measurement range measurement range (e. g., 0 ...100 Pa, mbar, mmHg etc.) B margin of error 1 % of end value 0.5%*, 250 Pa only 0.2 %*, 250 Pa 50 kPa
*of end value
A U I0 I4 IZ C 1 05 02
deflection drift / temperature zero point drift / temperature zero point drift / time
10 x for measurement ranges 20 kPa 2 x for measurement ranges > 20 kPa air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor
supply voltage 24 VDC, +20 % / 15% 24 VAC, +6 % / 15% (50 / 60 Hz) 115 VAC, +6 % / 15% (50 / 60 Hz) 230 VAC, +6 % / 15% (50 / 60 Hz) +10...+32 VDC (two-wire system)
D 24D 24A 115 230 PIZ LCD none 3 1/2 digit 4 1/2 LCD (PU / PI only) F 0 3 4
20 ms time constants E 0 1 2 5
none 1s 2s 5s
power consumption weight cable glands pressure ports protection class testing
approx. 3 VA 0.8 kg PU/PI: 2xPG 7 others available upon request , PIZ: 1 x PG 7 others available upon request , for hose 6 mm IP 65 CE, CSA accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0003 9601.-0004 9601.-0002
Order key
A
P
www.halstrup-walcher.com
03.2011
15
PS 10 Dimension drawing
with LCD
no LCD
Pa
Connection diagram
DC power supply AC power supply analogue outputs
or or
16
Special features for positive and negative differential pressures highly accurate and stable for long periods little zero point drift or hysteresis; largely independent of temperature excellent price/performance ratio
Technical data
output signals measurement ranges (others available upon request) margin of error deflection drift / temperature zero point drift / temperature zero point drift / time 250 / 500 Pa 1 / 2.5 / 5 /10 / 20 / 50 /100 kPa 2 % of end value 0.1% / K (+10 C...+50 C) 0.1% / K (+10 C...+50 C) 0.5 %/year 0...10 V (RL 2 k) 0 ... 20 mA (RL 500 ) 4 ... 20 mA (RL 500 ) measurement range measurement range in Pa, kPa, mmHg, etc. (e. g., 0 ... 250 Pa) power supply overload capacity medium max. system pressure 10 x for measurement ranges 20 kPa 2 x for measurement ranges > 20 kPa air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor 24 VDC, +20% / 15% 24 VAC, +6% / 15% (50 / 60 Hz) 115 VAC, +6% / 15% (50 / 60 Hz) 230 VAC, +6% / 15% (50 / 60 Hz) time constants none sensor response time 20 ms 1s 2s operating temperature storage temperature +10 C ... + 60 C 10 C ... +70 C 5s D 0 1 2 5 LCD none 3 1/2 digit 4 1/2 digit C 24D 24A 115 230 E 0 3 4 A 1 0 4 B
power consumption weight cable glands pressure ports protection class testing
approx. 3 VA approx. 0.8 kg 2 x PG7 others available upon request , for hose 6 mm IP 65 CE, CSA
Order key
A
PS 10
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
17
PS 27 Dimension drawing
no LCD
with LCD
PS 27
halstrup wa cher
ST4
power supply 24 VAC/DC
ST2
ST4
power supply 15...32 VDC
18
Special features for positive and negative differential pressures little zero point drift or hysteresis; largely independent of temperature also for top-hat rail mounting Zero-point calibration can be run via an external signal switching contact with adjustable thresholds (optional) output signals selectable via jumper four measurement ranges selectable via jumper (optional) with LCD (optional)
Technical datas
output signals measurement ranges
(others available upon request)
A 1 2 0 4 5 B AC/DC ZWL C
100 / 200 / 500 Pa 1/2,5/5/10/20/50/10 0 kPa 2 % of end value 0,1% / K 0,1% / K 12x for measurement ranges 20 kPa 4x for measurement ranges 20 kPa air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges 10 kPa max. nominal pressure of sensor
0 ...10 V (RL 50 k) 2 ... 10 V (RL 50 k) 0 ... 20 mA (RL 500 ) 4 ... 20 mA (RL 500 ) 0 ... 5 V (RL 50 k)
output signals adjustable via jumper
margin of error deflection drift / temperature zero point drift / temperature overload capacity
power supply 24 V AC/DC (without galvanic separation) 15 ... 32 VDC (two wire system) measurement range standard (e. g. 0 -100 Pa) switchable: 100, 250, 500, 1000 Pa switchable: 250, 500, 1000, 2500 Pa switchable: 1, 2,5, 5, 10 kPa switchable: 10, 25, 50, 100 kPa
1 2 3 4 D S 1
20 ms 20 ms ... 4 s adjustable (factory provided) contact point (not for two-wire system) none
1 switch relay max. 230 VAC, 6 A (min. required switching capacity 300mW) LCD
E 0 1
power consumption weight cable glands pressure ports protection class testing
none 4 digit
Order key
A
PS 27
IP 65 CE accessories
DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate www.halstrup-walcher.com 03.2011
19
Panel housing
begin and end settings select values or parameters call up parameter setting and display function
Connection diagram
relay output
DC power supply AC power supply
analogue
analogue output analogue output
power adapter
transistor output
DC power supply AC power supply
optocoupler
analogue output
20
analogue output
REG21/relay output
REG21/transistor output
Special features for positive and negative differential pressures can be used as a two- or three-position controller highly accurate and stable for long periods very little hysteresis; largely independent of temperature automatic zero-point calibration switching contacts available as relay or transistor outputs with adjustable switching thresholds panel housing
Technical data
output signals measurement ranges (others available upon request) margin of error deflection drift / temperature zero point drift / temperature 50/100/250 / 500 Pa 1 / 2.5 / 5 /10 / 20 / 50 /100 kPa 1 % of end value, 0.5 % of end value for measurement ranges 250 Pa 0.04 % / K (+10 C...+50 C) 0 % (cyclical zero-point correction) measurement range measurement range (e. g., 0100 Pa, mbar, mmHg, etc.) B margin of error 1% of end value 0.5 % of end value 250 Pa only C 1 05 0...10 V (RL 2 k) 5 V (RL 2 k) 0 ... 20 mA (RL 500 ) 4 ... 20 mA (RL 500 ) A 1 5 0 4
200 x for measurement ranges < 2.5 kPa 600 kPa for measurement ranges 2.5 kPa air, all non-aggressive gases 10 kPa for measurement ranges 10 kPa for measurement ranges > 10 kPa max. nominal pressure of sensor
supply voltage 24 VDC, +20 % / 15% 24 VAC, +6 % / 15% (50/60 Hz) 115 VAC, +6 % / 15% (50/60 Hz) 230 VAC, +6 % / 15% (50/60 Hz)
20 ms adjustable up to 10 s switching contacts 2 relays with floating changeover contacts 230 VAC (50 / 60 Hz), 6 A
Order key
A B C D E
REG 21
for hose 6 mm zero-point calibration for sensor control method (two- or three-position controller) switching point and hysteresis switching signal inversion response delay of relay outputs and analogue output CE 03.2011
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0003 9601.-0004 9601.-0002
testing www.halstrup-walcher.com
21
Connection diagram
zero point calibration on/ off select values or parameters select decimal places
pneumatic connections
22
Hand-held pressure gauge EMA 200 Portable, digital pressure gauge with min./ max. value memory
Special features flow-rate measurements taken in conjunction with a pitot tube displays pressure and flow rate stores min. and max. value scalable analogue output of 0 2 V can convert between Pa, kPa, mmHg, mmH2O, inH2O temperature measurement measuring ranges
Technical data
measurement range accuracy overload capacity 0.5 % of end value 10 x for measurement ranges 10 kPa 2 x for measurement ranges > 10 kPa 1.2 x in the 200 kPa measurement range v = 1.291*p air-speed given in m/s and p= differential pressure at pitot tube in Pa electronically by pressing zero point key air, all non-aggressive gases 02 V (RL 2 k) only for positive range 012 V (RL 2 k) for negative and positive range 3 1/2 digit LCD, character height = 10 mm 1 - 10 s 200 Pa 2 kPa 20 kPa 200 kPa (2 mbar) (20 mbar) (200 mbar) (2000 mbar) 0 (1.5) 18 m/s 0 (5) 58 m/s A 0 1 10 100
air-speed calculation
power supply
9 V battery (lifetime 100h) (display reads low bat when power falls below a certain minimum level) power automatically switches off after approx. 20 min. approx. 0.4 kg for hose 6 mm CE
Order key
A
EMA 200
accessories carrying bag DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9074.-0001 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
23
Connection diagram
zero point calibration on/ off select values or parameters select decimal places
pneumatic connection
24
Hand-held pressure gauge EMA 210 Portable instrument for measuring absolute pressures with min./max. value memory
Special features for measuring absolute pressures stores min. and max. value scalable analogue output of 0 2 V can convert between kPa, mbar, hPa, inH2O temperature measurement
Technical data
measurement range accuracy overload capacity zero point calibration medium analogue output display response time 0.5 % of end value 2 x for measurement ranges > 20 kPa electronically by pressing zero point key air, all non-aggressive gases 02 V (RL 2 k) 012 V (RL 2 k) 3 1/2 digit LCD, character height = 10 mm 0.5 s 0 - 2000 hPa A 1
power supply
9 V battery (lifetime 120h) (display reads low bat when power falls below a certain minimum level) power automatically switches off after approx. 20 min. approx. 0.4 kg for hose 6 mm CE
Order key
A
EMA 210
accessories carrying bag DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9074.-0001 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
25
analogue output pressure port for neg. pressure vent pressure port for pos. pressure vent
26
Special features highly accurate and stable for long periods extremely durable little zero point drift or hysteresis; largely independent of temperature analogue output of 0 1 V (optional) easy to operate
Technical data
measurement ranges margin of error 1 % of end value 0.5 % of end value for measurement ranges 1 kPa (optional) 0.2 % of end value for measurement ranges 1 kPa - 50 kPa (optional) 10 x for measurement ranges 10 kPa 2 x for measurement ranges > 10 kPa via potentiometer on front face air, all non-aggressive gases 0 ... 1 V (RL 2 k) BNC connector 3 1/2 digit LCD, character height = 13 mm toggles between 0.02 s; 0.2 s; 1 s analogue output operating temperature storage temperature +10 C ... +60 C 10 C ... +70 C none 0...1 V (optional) C 0 1 0 ...100 Pa 0 ...1 kPa 0 ...10 kPa 0 ...100 kPa accuracy 1% of end value zero point calibration medium analogue output display time constants 0.5% of end value (only for measurement ranges 1 kPa) 0.2% of end value (only for measurement ranges 1 kPa - 50 kPa) (0 ...1 mbar) (0 ...10 mbar) (0 ...100 mbar) (0 ...1000 mbar) A 0 1 10 100 B 1 5 2
overload capacity
Order key
A B C
accessories carrying bag shoulder bag DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9063.-0001 9064.-0001 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
27
depth
on/off button
display light
analogue output
vent for neg. vent for pos. pressure port pressure port
28
Special features highly accurate, reproducible results internal pressure generation extremely durable; excellent for service applications unit conversion, e.g. mmHg/kPa, mbar/kPa rechargeable battery allows for portable operation
Technical data
measurement ranges margin of error 0.5 % of end value 1 digit 0.2 % of end value 1 digit for measurement ranges 1 kPa - 50 kPa All measurement ranges have a 99 % overrange. Linearity data pertains to a measurement range of 0 100 %. 0.1% not applicable; panel button available for resetting zero point 0.04 %/ K (+10 C ... +50 C) +22 C air, all non-aggressive gases 0...100 Pa 0...1 kPa 0...10 kPa 0...100 kPa (0...1 mbar) (0...10 mbar) (0...100 mbar) (0...1000 mbar) A 0 1 10 100 300 750
hysteresis temperature effect (zero point) temperature effect (span) calibration temperature medium
margin of error 0.5 % of end value 0.2% of end value (measurement ranges 1 kPa - 50 kPa) (optional)
B 1 2
pressure transmitter, approx. 100 cm3 (1, 10, 100 kPa) approx. 200 cm3 (100 Pa) 0 ... 1 V (RL 2 k) two connectors 4 mm 4 1/2 digit LCD, character height = 10 mm toggles between 0.1 s; 1 s
C 230 115
Order key
A
KAL 84
accessories carrying bag hand pump DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9062.-0001 9601.-0036 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
29
menu on/off
menu selection
User software
30
Pressure calibration device KAL 100 / 200 Pressure calibration device with integrated pressure generation
Special features high mobility by battery-operation (optional) run-time with accumulator acc. 8h automatic zero point calibration provides high zero-point stability quickly provides positive or negative differential pressures up to 100 kPa USB Interface optional (standard for KAL 200) unit conversion (e.g., mmHg, mmH2O, psi etc.) multilingual menu (English, German, Italian, French, Spanish) With power supply and measuring input for the external test object.
Technical data
typ margin of error KAL 100 0.5 % of end value 1 digit (100 Pa measurement range) 0.2 % of end value 1 digit (only for 1, 10 and 100 kPa measurement ranges) 0.3% of end value 1 digit (100 Pa measurement range) 0.1% of end value 1 digit (only for 1, 10 and 100 kPa measurement ranges) 0,1 % 600 kPa for 10 kPa and 100 kPa measurement ranges 200 kPa for 100 Pa and 1 kPa measurement ranges 0 % (cyclical zero-point correction) KAL 100: 0,04 %/ K (+10 C ... +50 C) KAL 200: 0,03 %/ K (+10 C ... +50 C) +22 C air, all non-aggressive gases data interface none measuring input / power supply (external test object) display 0-10 V, 0/4-20 mA Measuring accuracy of end value: 0,2% 24 VDC / 100 mA alphanumerical display with 2 x 20 characters USB + measuring input (standard for KAL 200) D 0 1 KAL 100 KAL 200 measurement ranges 0...100 Pa 0...1 kPa 0...10 kPa 0...100 kPa supply voltage 115 VAC, +6 % / 15 % (50 /60 Hz) 230 VAC, +6 % / 15 % (50 /60 Hz) 115 VAC, +6 % / 15 % (50 /60 Hz) and lithium-ionen accumulator 230 VAC, +6 % / 15 % (50 /60 Hz) and lithium-ionen accumulator (0...1 mbar) (0...10 mbar) (0...100 mbar) (0...1000 mbar) A 100 200 B 0 1 10 100 C 1 2 1A 2A
Order key
A B C D
KAL
operating temperature storage temperature power consumption weight pressure ports testing www.halstrup-walcher.com
+10 C ... +40 C 10 C ... +70 C 10 VA approx. 4.5 kg for hose 6 mm CE 03.2011
accessories carrying case DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9220.-0001 9601.-0003 9601.-0004 9601.-0002
31
with LCD
no LCD
kPa
Connection diagram
DC power supply AC power supply analogue outputs
or or
32
Special features highly accurate and stable for long periods little zero point drift or hysteresis; largely independent of temperature display can be adjusted (reduced) to correspond to the height of installation site (see DIN ISO 2533) AD 1000 with pressure port for measuring absolute pressure BA 1000 for measuring barometric pressure
Technical data
measurement ranges margin of error temperature effect calibration temperature operating temperature storage temperature long-term drift reduction 1 %, reference 0.5 hPa with respect to sea level 0.04 %/K (+10 C ... +50 C) +22 C +10 C +60 C 10 C ... +70 C 0.3 hPa / year 0 850 m above sea level (please indicate when placing your order) approx. 3 VA 2 x PG 7 (for a 80 x 120 housing) 2 x PG11 (for a 120 x 122 housing) IP 65 approx. 0.6 kg for hose 6 mm CE LCD none 3 1/2 digit 4 1/2 digit reduction none (please indicate in meters) 80...120 kPa 85...115 kPa 90...110 kPa 95...115 kPa 0...50 kPa
2) 1)
1)
1)
1)
0...100 kPa
2)
100A
2)
80A 90A 0A
2) AD 1000 (w. pressure port)
2)
output signals 0...10 V (RL 2 k) 0... 20 mA (RL 500 ) 4 ... 20 mA (RL 500 )
B 1 0 4
Order key
A
AD-BA 1000
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate www.halstrup-walcher.com 03.2011 9601.-0003 9601.-0004 9601.-0002
33
BA 90 Dimension drawing
Connection diagram
AC power supply DC power supply analogue outputs
34
U/I output
Special features a potentiometer can be used to adjust (reduce) display to correspond to the height of installation site (see DIN ISO 2533) highly accurate and stable for long periods little zero point drift or hysteresis; largely independent of temperature 3 supply voltages in one unit
Technical data
output signals measurement range margin of error resolution 913.3 1113.3 hPa 0.4 hPa 1 digit, reference 0.5 hPa with respect to sea level 0.1 hPa 2 + 2 V (RL 5 k) 0...20 mA (RL 250 ) 4...20 mA (RL 250 ) A 1 0 4
0.2 hPa / C, for temperatures ranging between +20 C+50 C + 22 C 0 C +50 C (temperature compensation between + 20 C...+50 C) 10 C... +70 C 0.3 hPa / year
supply voltage
230 VAC +6 / 15 % or 115 VAC +6 / 15 % or 12 28 VDC (universal voltage adapter) 0...850 m above sea level, via potentiometer approx. 5 VA approx. 0.8 kg CE
Order key
A
BA 90
accessories DAkkS-DKD calibration certificate, German DAkkS-DKD calibration certificate, English factory calibration certificate 9601.-0003 9601.-0004 9601.-0002
www.halstrup-walcher.com
03.2011
35
Sample applications Volume flow measurement If measured in conjunction with a differential pressure transmitter (measuring orifice, Wilson flow grid, pitot tube, etc.), the recorded differential pressure will be directly proportional to the volume flow as a result of the rootextracted function of the P 82 R / P 26 model pressure transmitter. The resulting value is then displayed on the built-in LCD. The output signal is also proportional to the volume flow, thereby eliminating the need for time-consuming calculations in the master control module. P 26
flow grid
MEN
Measuring differential pressure Monitoring static room pressure, as is done, for instance, in clean rooms or operating rooms, requires a pressure transmitter that can operate within very small measurement ranges, often only a few Pascals. The P 26 pressure transmitter is perfectly suited for this type of task, as it is designed to operate in measurement ranges as small as 0 ...10 Pa.
Conversion table for the most common pressure units Pa hPa/mbar kPa bar psi mmH2O inH2O mmHg inHg 36
| bar | psi 0.00001 | 0.0001 | | 0.001 | 0.015 0.010 | | 0.145 |1 | 14.514 0.069 | |1 | 0.000098 | 0.001 | 0.00249 | 0.036 | 0.00133 | 0.019 | 0.03386 | 0.491
DakkS-DKD pressure calibration laboratory Germanys national metrology institute (Deutschen Akkreditierungsstelle GmbH) has certified Walcher Metechnik GmbH a member of the halstrup-walcher group of companies to perform pressure calibrations in accordance with DIN EN ISO / IEC 17025. This allows Walcher Metechnik GmbH to issue DAkkSDKD calibration certificates for differential pressure transmitters, calibration devices, absolute pressure transmitters and portable pressure gauges. Absolute pressures between 0.25 and 20 bar can be measured here, as can negative and positive differential pressures of 0 to 20 bar between gases. Measuring and calibration devices are calibrated independently of the manufacturer. Services also include recalibration of all of the products listed above as directed by the ISO 9000 quality management system for measuring equipment.
www.halstrup-walcher.com
37
Our expertise in pressure measuring technology For decades halstrup-walcher has always stood for quality and reliability concerning pressure measuring technology for differential pressures between 0-10 Pa and 0-100 kPa. Our inductive sensor element with its copper beryllium membrane guarantees a high degree of independence from varying temperature as well as long-term stability. For measuring volume flow and mass flow, we offer models with digital displays that have been precalibrated for these parameters. All pressure transmitters are available with a display and calibration protocol, alternatively in German or English. In addition to pressure transmitters, we also produce extremely cost-effective pressure calibration devices, e.g. to be used for medical engineering. Our expertise in mechatronics The most remarkable feature of our mechatronic positioning systems is the integration of engine, gear, performance electronics, measuring system, controls, and interfaces on a very confined space.
Our expertise in drive technology Our focal point in manufacturing spur Gears always lies on customer specific solutions offering a maximum of cost-effectiveness for every application. Our scope of delivery includes complete solutions including motor control gear with position feedback signal and/or adjustable limit switches. You can also have your gearbox produced according to your specific requirements.
38
halstrup-walcher GmbH precision for your success halstrup-walcher GmbH was founded as Erwin Halstrup Company in 1946. It was renamed Erwin Halstrup Multur GmbH in 1981 and assumed the name halstrup-walcher GmbH in the year 2000. It became a subsidiary of the Walcher Metechnik GmbH in 1990. Halstrup-walchers technical solutions stand out for their extraordinary quality, precision and innovative nature. Our product range covers the following devices differential pressure transmitter for low pressures handheld pressure gauges pressure calibration systems absolute pressure measuring systems barometers spur gearboxes actuators linear drives positioning systems
Distribution In-house salesmen and commercial agents take care of national sales; appointed retailers carry out international sales of our precision measuring instruments. Manufacturing Modern machines for fitting and soldering circuit boards are used for the manufacture of the electronic modules. Climatic chambers are available for burn-ins as well as air-controlled labs for quality control and/or calibration of the end products. The mechanical manufacturing process involves punches, milling cutters (CNC), lathes (CNC) as well as electronically controlled presses to mount the gear wheels. Production cells are responsible for portions of the assembly process and perform the final inspection of mechanical parts.
www.halstrup-walcher.com
39
Your competent partner in pressure measuring technology between 10 Pa and 100 kPa
halstrup-walcher GmbH Stegener Strae 10 D-79199 Kirchzarten Germany Tel. +49-76 61-39 63-0 Fax +49-76 61-39 63-99 www.halstrup-walcher.com info@halstrup-walcher.com
7150.000324