Sunteți pe pagina 1din 20

FUMISENSE

[UNIPHOS-251(PM)F]

METHYL BROMIDE FUMIGATION MONITOR

INSTRUCTION MANUAL

United Phosphorus Limited Readymoney Terrace, 167, Dr. Annie Besant Road, Worli, Mumbai - 400 018. India Tel: (91-22) 2493 0681, Fax: 2493 8826

PDF created with pdfFactory trial version www.pdffactory.com

TABLE OF CONTENTS
PG.NO 1. INTRODUCTION--------------------------------------------------------------------------4 1.1.INSTRUMENT DESCRIPTION--------------------------------------------------------4 1.2. WORKING PRINCIPLE-----------------------------------------------------------------4 2. DESCRIPTION OF DIFFERENT KEYS, KNOBS ETC. ON FRONT & REAR PANEL---------------------------------------5 FIGURE-1 FRONT PANEL ------------------------------------------------------------5 FIGURE-2 REAR PANEL---------------------------------------------------------------6 2.1.FRONT PANEL----------------------------------------------------------------------------6 2.2.REAR PANEL------------------------------------------------------------------------------8 3. OPERATION-------------------------------------------------------------------------------9 3.1. POWER ON THE INSTRUMENT----------------------------------------------------9 3.2. INTIALISATION---------------------------------------------------------------------------9 3.3. LCD BACKLIGHT------------------------------------------------------------------------10 3.4. MEASUREMENT MODE--------------------------------------------------------------10 3.5. PROCEDURE FOR ZEROING THE INSTRUMENT --------------------------10 3.6. PROCEDURE FOR MEASUREMENT -------------------------------------------11 4. HOW TO SELECT THE MODES ---------------------------------------------------12 4.1.SELECTING PARAMETER SET MODE-------------------------------------------12 4.2.SELECTING CALIBRATION MODE------------------------------------------------12 5. HOW TO GO TO THE MEASUREMENT MODE WHEN THE INSTRUMENT IS IN SOME OTHER MODE---------------------12 6. PROCEDURE FOR PERFORMING THE OPERATIONS IN OTHER MODES---------------------------------------------------------------------12 6.1.PARAMETER SET MODE-------------------------------------------------------------12 6.1.1. SET DATE------------------------------------------------------------------------------12 6.1.2. SET TIME-------------------------------------------------------------------------------13 6.1.3.RESETTING DATA LOGGER MEMORY----------------------------------------13 6.1.4.ENABLING/DISABLING DATA LOGGER---------------------------------------13 6.1.4.1. PROCEDURE TO DISABLE DATA LOGGER------------------------------13 6.1.4.2. PROCEDURE TO ENABLE DATA LOGGER-------------------------------14 6.2.CALIBRATION MODE------------------------------------------------------------------14 6.2.1.ADJUSTING THE ZERO DRIFT --------------------------------------------------14 6.2.2.ADJUSTING THE GAIN -------------------------------------------------------------15 7. HOW TO DOWNLOAD DATA ON PC/PRINTER--------------------------------15 8. HOW TO PURGE THE INSTRUMENT---------------------------------------------16 9. HOW TO MONITOR THE BATTERY CHARGE LEVEL------------------------16 10. MAINTENANCE---------------------------------------------------------------------------16 11. INSTRUMENT SPECIFICATIONS---------------------------------------------------17 12. SENSOR SPECIFICATIONS----------------------------------------------------------18 LIMITATION OF WARRANTY AND LIABILITY ---------------------------------------19

PDF created with pdfFactory trial version www.pdffactory.com

Dear Customer,
Congratulations on your purchase of Uniphos-251 (PM) F Methyl Bromide (FumiSense) monitor. This monitor has been produced using the state of the art Technology in our Chemo-Electronic Laboratory at Vapi, to provide you an instrument with the best of features and yet at a moderate cost. It represents the high degree of craftsmanship and reliability that has made UNIPHOS a leader in the field of gas detection. Please read the instruction manual carefully. It gives a brief description of the instrument and the full details of operation for measurement, calibration, parameter setting etc. To ensure a long trouble free performance, please take due care and ensure proper maintenance of your monitor as recommended in this Instruction manual. Please do not attempt to use any other accessories and other components other than UNIPHOS. Also do not attempt to open the instrument at site, as it would lead to termination of warranty for your instrument. If you have any queries regarding the operation or maintenance of your UNIPHOS monitor, please do feel free to consult our factory at the address and we will be too happy to help you.

Manufactured At United Phosphorus Ltd. Chemo-Electronic Laboratory At: Nahuli, Post: Umbergaon Dist: Valsad, Gujarat State Pin: 396195, INDIA

Marketing Office United Phosphorus Limited Readymoney Terrace, 167, Dr. Annie Besant Road, Worli, Mumbai - 400 018. India Tel: (91-22) 2493 0681 Fax: 2493 8826

PDF created with pdfFactory trial version www.pdffactory.com

1. INTRODUCTION:
Methyl bromide is a highly specific fumigant used to control a number of pests in a wide range of agricultural and horticultural crops and in wood products. Its primary application is for quarantine treatment, soil fumigation and post harvest protection. Fumigation is generally carried out by covering the commodities within a leak proof enclosure and applying methyl bromide at a concentration ranging from 50 mg/l to 150 mg/lit. As a process control measure it is necessary to monitor the gas concentration at various stages of fumigation. For this application FumiSense ( Uniphos-251(PM)F ) the fumigation gas monitor is the right choice. This manual gives the full description of FumiSense and its operation, calibration, maintenance and other relevant information. 1.1. INSTRUMENT DESCRIPTION: FumiSense is a microprocessor based instrument for the measuring the methyl bromide gas in the concentration range of 0 200 mg/l. The instrument is specially designed for monitoring the gas concentration in the fumigation silo or container. The air sample from the fumigation silo / container is brought to the sensor by means of the inbuilt air-sampling pump of the instrument. A dust filter is placed in the inlet of the gas sample line to prevent dust particles entering the instrument. Being a microprocessor based unit it has data logging facility and it can store data such as gas concentration along with the silo/ container serial number, godown number etc. with real time clock. There is also provision to download the stored data on computer or serial printer. The instrument works on inbuilt 14.8 V/ 4 Ah capacity Li-Ion rechargeable batteries. 1.2. WORKING PRINCIPLE: The FumiSense is equipped with Thermal conductivity sensor for the detection of Methyl Bromide gas. Thermal conductivity sensor: It consists of a pair of thermal conductivity detectors secured in the gas-sampling hood made up of brass. The two detectors form the two arms of a wheat stones bridge, which is balanced in clean air. An out of balance signal is generated when one of the detectors is

PDF created with pdfFactory trial version www.pdffactory.com

placed in sample air containing Methyl Bromide. This occurs due to thermal conductivity difference between clean air and air containing Methyl Bromide. The out of balance signal, which is proportional to the Methyl Bromide concentration in the sample air is digitized and displayed on the digital display of the monitor.

2. DESCRIPTION OF THE DIFFERENT KEYS, KNOBS etc ON FRONT AND REAR PANEL:
The instrument has got four membrane keys viz. UP, DN, SET & ENTER, one knob, two LEDs, LCD display, Power on/off switch etc. on the front panel and pair of gas inlet/outlet ports, RS232 socket, battery charging socket, a potentiometer for MBr coarse tuning etc. on the rear panel. Figures 1 & 2 give the layout of the front panel and rear panel. The various functional keys, knobs etc. are numbered and described in the following paragraphs.

FIGURE: 1 - FRONT PANEL


LCD-PANEL (3)

UNIPHOS-251(PM)FE
POWER-ON/OFF SWITCH (4)

ON
SAMPLING

ZERO-FINE KNOB (5)

FUMISENSE

ZERO-FINE

ENTER KEY (6) SET KEY (7) DOWN KEY (8) POWER (1) SAMPLING (2) UP KEY (9)

PDF created with pdfFactory trial version www.pdffactory.com

FIGURE: 2 - REAR PANEL

MBr-ZEROCOARSE

BATTERY CHARGERSOCKET (11)

RS-232 SERIAL PORT FOR PC/PRINTER (12)

MBr GAS-OUTLET (13)

MBr GAS-INLET ((14)

MBr ZEROCOARSE KNOB (10)

2.1. FRONT PANEL (FIGURE 1)


SET KEY (7): This key has three functions. 1) In the Measurement mode: Press this key to go to the Purging mode / parameter setting mode /calibration mode. 2) In other modes: To escape from the current page and move over to the next page. 3) During initialization if SET key is pressed the instrument skips all prompts, viz Model no, Range, Resolution etc and enters the sampling mode.

PDF created with pdfFactory trial version www.pdffactory.com

UP KEY (9): This key has two different functions. 1) Measurement Mode: For Gas Sampling. 2) Other modes: To increment the value of the digits over which the cursor is placed. DOWN KEY (8): This Key is has got two functions. 1) Calibration Mode: In the calibration mode this key is used as decrement key. 2) Parameter Set Mode: In this mode this key is used to move the cursor horizontally over different digits on the same line. 3) In the main menu on key press it displays the battery percentage on the LCD. ENTER KEY (6): This Key has two functions. 1) When in Measurement Mode: To go directly to the Data Downloading page. 2) When in other Modes: To confirm the value of the parameter or to confirm the desired operation. POWER ON/OFF SWITCH (4): This switch when in the ON position turns ON the monitor. NOTE: Putting the Power ON/OFF switch to the OFF position turns off the internal circuit. POWER LED (1): It glows when the monitor is powered ON.

SAMPLING LED (2): It glows when the air sampling process is in progress.

LIQUID CRYSTAL DISPLAY (3): It is a 16X2 alphanumeric display, which displays Various prompts during initialization immediately after power on, Gas concentration in the measurement mode User settable parameters during parameter setting mode
7

PDF created with pdfFactory trial version www.pdffactory.com

Calibration data viz. Cal-zero, Cal-gain during calibration mode. User friendly prompts during sampling, calibration and data downloading and other functions. During initialization the backlight glows for 30 seconds and then switches off automatically to save the Battery life. It can be activated on any key press viz. UP, DN, SET & ENTER.

ZERO CONTROL KNOB (FINE CONTROL)(5): For fine adjustment of the base line (zero gas) reading before taking the sample for measurement each time.
NOTE: The markings provided on the dial of the knobs are of no significance to the user.

2.2. REAR PANEL (FIGURE 2)


BATTERY CHARGER SOCKET (11): Connect the battery charger into this socket for charging the internal battery when the unit shows low battery indication. Charge the battery for about 5 to 8 hours using the battery charger supplied along with this monitor.

RS232 SOCKET (12): The RS 232 Cable provided with the instrument is to be inserted in to this socket to download data on PC/Printer. The other end of the cable is to be connected to the COM port of PC.The settings of the PC are mentioned below. GAS INLET PORT METHYL BROMIDE (14): The tube through which the sample air is drawn from the fumigation chamber/silo is to be connected to the gas inlet port. This is used specifically for Methyl Bromide GAS OUTLET PORT METHYL BROMIDE (13): A tube is to be connected to the gas outlet port and other end of this tube should be put back to the fumigation chamber. This is used specifically for Methyl Bromide. *MBr ZERO CONTROL KNOB (Coarse CONTROL)(10):
8

PDF created with pdfFactory trial version www.pdffactory.com

For Coarse adjustment of the base line (zero gas) reading before taking the sample ( MBr ) for measurement each time.
* The functioning of both the knobs is later described in section 3.6

3. OPERATION
3.1 Power ON the instrument: Put the power ON switch located on the top right of the front panel to ON position. A green lamp (LED) named POWER on the front panel of the Instrument illuminates confirming the same. 3.2 Initialization: Once the instrument is powered ON the instrument enters through a series of prompts such as Model no, gas name, operating range & resolution, current date & time etc. After the initialization the instrument is ready for sampling. During the initialization mode following prompts one after the other comes on the LCD panel
UNIPHOS 251 (PM) F (Device name)

METHYL BROMIDE MONITOR (Gas name)

RANGE: MBr: 0-200 mg/l RESOLUTION: MBr: 1 mg/l

(Range for MBr measurement) (Resolution for MBr measurement ) (Current Date) (Current Time) (Key operation) (Key operation) (Key operation)

28: 10: 2005 16:51: 23

PRESS UP KEY FOR SAMPLING DOWN KEY FOR BATTERY STATUS SET KEY PURGE/ SET/CAL MODE

ENTER KEY FOR DOWN LOADING DATA TO PC/ PRINT (Key Operation) Note: The above-mentioned prompts can be skipped and the instrument can be made to enter the sampling mode by pressing the SET key as mentioned above.

Then the display shows PRESS UP KEY FOR SAMPLING. This is the measurement mode. Follow the procedure mentioned below for making measurements.
9

PDF created with pdfFactory trial version www.pdffactory.com

3.3 LCD back light: LCD panel has the backlight provision. After initialization this back light gets switched off automatically. It can be turned on any time by pressing any one of the membrane keys., Once turned on it remains ON for about 30 seconds & then again gets turned off automatically in order to save the battery life. Whenever required user can press any of the 4 membrane keys to activate the backlight. Please note that when backlight is OFF then for any operation it is necessary to turn on the backlight by pressing any one of the four membrane keys. 3.4 MEASUREMENT MODE 1. Once you are in the measurement mode the instrument is ready for making measurement. The display shows PRESS UP KEY FOR SAMPLING. 2. When UP key is pressed, the display will show READY FOR TAKING SAMPLE ADJUST ZERO KNOB TO READ ZERO, followed by the zero gas concentration and the prompt PRESS UP KEY. 3. Before making measurement you have to zero the instrument. For this, follow the procedure given in section 3.5. 3.5. PROCEDURE FOR ZEROING THE INSTRUMENT 1. The instrument is meant for measurement of Methyl Bromide in silo. There are two sets of tubes provided, one as gas inlet tube and other as gas outlet tube. The gas outlet port of monitor should be connected to gas outlet tube(Figure-3).

FIGURE-3 SAMPLING TUBE CONNECTION-ZERO ADJUSTMENT


10

PDF created with pdfFactory trial version www.pdffactory.com

2. If the sensor is exposed to target gas free clean air, the LCD should display zero reading. If the reading is not zero adjust it to zero by following the procedure mentioned below. 1.Turn COARSE ZERO knob provided on the rear panel of the instrument and adjust the reading displayed on the LCD to almost zero. 2.Then by using the FINE ZERO knob provided on the front panel of the instrument adjust the reading exactly to zero. NOTE: Usually the control knobs will be in the locked position. If they are locked, unlock them by shifting the notch provided on the control knobs to the opposite position. 3.6. PROCEDURE FOR MEASUREMENT 1. After zeroing the instrument connect the gas inlet tube to the Gas Inlet Port of the monitor ensuring no leaks (Figure-4). 2. Press the UP Key. The instrument will now start sampling. Sampling time provided is 3 minutes. At the bottom right corner of the LCD, time left for sampling will be displayed. 3. The pump will be turned on during sampling period. The pointer on the flow meter should indicate a flow rate in the range of 0.5 to 1 LPM. If not adjust the flow rate to required level using the flow control pot provided on the left side of the instrument. 4. After 3 minutes the pump & Sampling On LED will be automatically turned off. The display will now show PROCESSING ON PLEASE WAIT. After 10 seconds the display will show the gas concentration and SAMPLING OVER prompt. Note down the reading, which corresponds to the gas concentration. (This reading will be saved in the memory of the instrument provided the data logging option is ON. This data can be downloaded to the PC whenever required. Refer to Section 7 for details on data down loading.

FIGURE-4 SAMPLING TUBE CONNECTION DURING SAMPLING


11

PDF created with pdfFactory trial version www.pdffactory.com

Press UP Key to go to the measurement mode. Now the monitor is ready for another measurement. NOTE: You dont have to adjust zero after every measurement, even if the reading does not fall down to zero. This instrument has got two other modes for performing the following operations: 1. PARAMETER SETTING 2. CALIBRATION OF THE INSTRUMENT In your monitor the parameters are set already. But if you want to change any of them you have go to the PARAMETER SET MODE of the instrument.

4. HOW TO SELECT THE MODES:


4.1 SELECTING PARAMETER SET MODE: When you are in the measurement mode if you press the SET Key twice, the instrument will go to the Parameter set mode. 4.2 SELECTING CALIBRATION MODE: Press SET Key thrice when you are in the measurement mode. The instrument goes to the calibration mode.

5. HOW TO GO TO THE MEASUREMENT MODE WHEN THE INSTRUMENT IS IN ANOTHER MODE:


Press SET Key until the Measurement Page appears.

6. PROCEDURE FOR PERFORMING OPERATIONS IN OTHER MODES:


6.1. PARAMETER SET MODE As mentioned in Section 4.1 press SET Key twice to go to PARAMETER SET MODE. The display will now show SET PARAMETERS? CONFIRM? Press ENTER Key. Now follow the steps mentioned below to set the various parameters. Press SET Key to exit SET PARAMETER Mode and go to the next page.
6.1.1. SET DATE:

1. Once you go to the parameter set mode the display will show the current date. 2. The cursor will also be blinking on the first digit. Use DOWN Key to move the cursor horizontally on different digits. 3. If any digit is to be changed move the cursor on the digit and press UP Key. Using UP Key the digits can be changed from 0-9.
12

PDF created with pdfFactory trial version www.pdffactory.com

4.Press ENTER Key to confirm the changed digit. Incase the displayed date is not to be changed you can press SET Key to escape from the date page to next page i.e. set time page.
6.1.2. SET TIME:

1. When you are in the measurement mode press the SET Key until the display shows the current time. The cursor will now be blinking on the first digit. 2. As explained in previous section use DOWN Key to move to different digits horizontally and Up Key to change the digits on which cursor is blinking. Using UP Key the digits can be changed from 0-9. 3.Every change of digit is to be confirmed by pressing ENTER Key. If time is not to be changed press SET Key to escape from this page and go to next page.

6.1.3. RESETTING DATA LOGGER MEMORY:

1. Press the SET Key until the display shows RESET DATA LOGGING MEMORY Page. The display will now show RESET DATALOGGER 2. Press ENTER key to reset Data Logger. CAUTION: Resetting Data Logger Memory refers to deletion of all the records present in the memory of the instrument and not available to the user during later periods. If the records in the data logging memory are not to be deleted press SET Key to escape from this page and go to next page.
6.1.4. ENABLING/DISABLING DATA LOGGER:

Press the SET Key until the display shows ENABLE/DISABLE DATALOGGER Page. When this page is selected, if the data logger is enabled the display will show DISABLE DATA LOGGER. Follow the steps mentioned in section 6.1.4.1. to disable data logger. If the data logger is disabled the display will show ENABLE DATA LOGGER. Follow the procedure mentioned in section 6.1.4.2. to enable data logger.
6.1.4.1. PROCEDURE TO DISABLE DATA LOGGER:

1. Press ENTER key. The display will now show DISABLE DATA LOGGER CONFIRM? . 2. Press ENTER key again to DISABLE DATA LOGGER.

13

PDF created with pdfFactory trial version www.pdffactory.com

6.1.4.2. PROCEDURE TO ENABLE DATA LOGGER:

1.Press ENTER key. The display will now show ENABLE DATA LOGGER CONFIRM? . 2. Press ENTER key again to ENABLE DATA LOGGER. NOTE: ENABLING Data logger enables the records to be logged on to the instruments memory, DISABLING Data logger disables the records to be logged on to the instruments memory. Please ensure that the Data Logger is enabled/Disabled as per your convenience. Press SET Key to escape from this page and go to measurement mode. 6.2.CALIBRATION MODE Calibration of the instrument involves two steps: 1) Zeroing the instrument in clean air 2) Applying gas of known concentration and adjusting the gain so that the reading matches with the calibration gas concentration As mentioned in Section 4.2 press SET Key thrice to go to CALIBRATION MODE. The display will show CALIBRATE?. Press ENTER Key. Now the display will show CALIBRATE? CONFIRM?. Press ENTER Key again. The unit will now go to the zero drift adjustment page. Press SET Key to escape from this mode and go to measurement mode
6.2.1. ADJUSTING THE ZERO DRIFT

1. In the ZERO DRIFT ADJUSTMENT Page, If the reading is already zero, press SET Key. When SET Key is pressed the unit will go to CAL GAIN Adjustment Page. 2. If the reading is negative, using UP Key increment the CAL OFF zero value until the reading becomes zero. If the reading is positive, using DOWN Key decrement the CAL OFF zero value until the reading becomes zero.
NOTE: The

range of CAL OFF zero value is from 125 to +125. If the CAL OFF zero value has reached the extreme value and the reading is still not zero, then zero adjustment is not possible with the Keys.

3. Press the Enter Key to confirm the new CAL OFF zero value. Now the unit will go to the CAL GAIN Adjustment Page. Press SET Key to escape from this page and go to the next page.

14

PDF created with pdfFactory trial version www.pdffactory.com

6.2.2. ADJUSTING THE GAIN


NOTE: The

range of CAL GAIN value is from 125 to +125. If the CAL GAIN value has reached the extreme value and the reading has still not reached the expected reading, then gain adjustment is not possible with the Keys and the unit should be sent to the factory for calibration.

1. 2.

Now press the Enter Key to confirm the new CAL GAIN value. Now the unit comes to the default measurement mode.

Press SET Key to escape from this page and go to the default measurement mode.

7. HOW TO DOWN LOAD DATA ON PC/PRINTER


1.Connect the cable given for data downloading to the RS232 Slot provided at the rear panel of the instrument. Connect the other end of the cable (9-Pin Connector) to the COM1 Port of PC or Port of Printer. 2.To download data on PC, select *HyperTerminal. This is an application generally provided along with the operating system ranging from Windows 98 and upwards. 3.Press the ENTER Key on the front panel of the instrument. The instrument will go to the DATA DOWNLOAD Page. Now the LCD will display DOWNLOAD? 4.Press the ENTER Key. The LCD will display DOWNLOAD? CONFIRM?. 5.Press the ENTER Key once again to download the data. The display will now show the number of records being downloaded on to the PC/Printer. After downloading the records the unit will go to the measurement mode. Press SET Key to escape from this page and go to the measurement mode. *The COM1 Port Settings of the Hyper terminal settings to be carried out in the PC are as follows: 1.Baud Rate: 9600 bps (bits per second) 2.Data Bits: 8 3.Stop Bits: 1 4.Parity Bits: None 5.Flow Control: None
15

PDF created with pdfFactory trial version www.pdffactory.com

8.HOW TO PURGE THE INSTRUMENT


1. Now press SET Key in the measurement mode. The display will show PURGE?. 2. Press ENTER Key. The display will show PURGE? CONFIRM? 3. Press ENTER Key again. The display will show the prompt REMOVE TUBES. Remove the tubes from the gas inlet port(14) and gas outlet port(13). The instrument will start purging after a few seconds. The purging period given is 3 minutes which will be indicated on the bottom line of LCD. After purging the instrument will go to the measurement mode.

9. HOW TO MONITOR THE BATTERY CHARGE LEVEL.


You can know the battery charge level by pressing the DOWN ARROWKEY before entering into the sampling mode i.e. when the LCD display is showing the following prompt PRESS UP ARROW -KEY FOR SAMPLING

LCD will indicate the battery charge level in terms of the % of the full capacity of the battery. The battery status will be displayed in four different levels viz. >100%, >75%, >50% and <25% Beyond 25% the instrument flashes the prompt LOW BATTERY RECHARGE

10. MAINTENANCE
NOTE: 1. NEVER allow the unit to be used in direct sunlight. You may please provide the canopy/sun shade while using on the site. 2. ZERO it in a target gas free clean air. 3. NEVER expose the sensor to gas concentration higher than its prescribed range. 4. Occasionally the casing of the instrument can be wiped with a wet cloth. But care should be taken to see to it that no water enters the instrument. 5. Ensure that the battery is sufficiently charged every time you use the unit.
16

PDF created with pdfFactory trial version www.pdffactory.com

11. INSTRUMENT SPECIFICATIONS:


1. Detection Principle : 2. Gas Sampling : Thermal Conductivity Detector (TCD) Batch Sampling.Sample Draw Type (with inbuilt pump). Red LED - Indication for pump sampling ON. 3 minutes Gas selection by using down key and selector switch 16X2 Alphanumeric LCD Display with backlight (auto-switch off provision) Microprocessor Based electronic instrument with inbuilt data logger Data Logger; Day, time and measurement results recorded for each measurement More than 1000 data points

3. Sampling Time 4. Gas Selection

: :

5. Display

6. Electronic Circuitry :

7. Output

8. Data Logging Function 9. Inbuilt Pump

Flow rate: 500 ml/min - 1000 ml/min Power: 4.5 V DC, 35 mAh 14.8 VDC, 4.0 Amp-hr, inbuilt Li-Ion rechargeable battery operated Height - 150 mm Width - 230 mm Depth - 210 mm 3.3 Kg ABS

10. Power Supply

11. Dimensions

12. Weight 13. Construction Material 14. Standard Accessories

: :

Cable for data logger, Dust filter , Batt charger

17

PDF created with pdfFactory trial version www.pdffactory.com

12.SENSOR SPECIFICATIONS:
1. Gas detected : Methyl Bromide(MBr)

2. Concentration range : 0-200 mg/L 3. Resolution 4. Warm up time : 1 mg/L : Less than 3 min

5. Response time (T95): Typically less than 10 sec after establishment of the gas concentration in the sensor hood 6. Accuracy :Inherently 2% (but also depends on the accuracy of the calibrating gas & sampling system) :0 to + 50 oC

7. Operating temperature 8. Operating life of sensor

:Expected life over 5 years Warranted for two years.

18

PDF created with pdfFactory trial version www.pdffactory.com

LIMITATION OF WARRANTY AND LIABILITY


Seller warrants that this product, under normal use and service as described in the operators manual shall be free from defects in workmanship and material for a period of twelve (12) months, or the length of time specified in the operators manual, from the date of shipment to the customer. This limited warranty is subject to the following exclusion: a. Batteries and certain other components when indicated in specifications are warranted for a period of 90 days from the date of shipment to the customer. b. With respect to any repair services rendered, Seller warrants that the parts repaired or replaced will be free from defects in workmanship and material, under normal use, for a period of 90 days from the date of shipment to the customer c. Seller does not provide any warranty on finished goods manufactured by others. Only the original manufacturers warranty applies. d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, any goods, which are incorporated into other products or equipment by the Buyer. All goods returned under warranty shall be at the Buyers risk of loss; Sellers factory prepaid, and will be returned at Sellers risk of loss, Buyers Factory prepaid. The foregoing is IN LIEU OF all other warranties and is subject to the conditions and LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. THE EXCLUSIVE REMEDY OF THE USER OR PURCHASER, AND THE LIMIT OF LIABILITY OF SELLER FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES IN CONNECTION WITH THIS PRODUCT (INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE, STRICT LIABILITY, OTHER TORT, OR OTHERWISE) SHALL BE THE RETURN OF THE PRODUCT TO THE FACTORY OR DESIGNATED LOCATION AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF THE SELLER, THE REPAIR OR REPLACEMENT OF THE PRODUCT. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING, REASSEMBLY OR REINSTALLATION COSTS OR CHARGES. NO ACTION, REGARDLESS OF FORM, MAY BE BROUGHT AGAINST THE SELLER MORE THAN ONE YEAR AFTER THE CAUSE OF ACTION HAS ACCRUED. The purchaser and all users are deemed to have accepted the terms of this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended or modified nor may any of its terms be waived except by a writing signed by an authorized representative of the Seller.

19

PDF created with pdfFactory trial version www.pdffactory.com

(Due to continuing development we reserve the right to Change specifications without prior notice)

Manufactured At United Phosphorus Limited Chemo Electronic Laboratory P. O. Nahuli, Tal. Umbergaon, Dist. Valsad - 369 108, India Tel.: 91-260-2730156/7/8/9 Fax: 91-260-2730160/2431823

Marketing Office United Phosphorus Limited Readymoney Terrace, 167, Dr. Annie Besant Road, Worli, Mumbai - 400 018. India Tel: (91-22) 2493 0681, Fax: 2493 8826

http://www.chemo-electronic.com/

20

PDF created with pdfFactory trial version www.pdffactory.com

S-ar putea să vă placă și