With this kind of Flexibility in an Analyzer, You Can Design Your Own IH/Safety Analyzer PID Analhyzers, LLC http://www.hnu.com
PID Analyzers, LLC copyright 2003-2011
History of Photoionization
In
May 2011, Dr. Jack Driscoll of HNU and now with PID Analyzers received the Outstanding Achievement Award from NESACS for the development of the first handheld photoionization detector The Model 102+ represents the fifth generation of that device
PID Analyzers, LLC copyright 2003-2011
Schematic of PID
Photoionization Process
Where: R= a molecule with an IP <10.5 for the 10.6 or 11.9 for the 11.7 lamp h = a photon with an energy >= IP of the molecule R+ is the ion formed by the process that is collected in the ion chamberconcentration is proportional to the current produced
PID Analyzers, LLC copyright 2003-2011
R + h = R+ + e
Snap on Heads
Interchange PID lamp heads (9.5, 10.6, 11.7 eV) without the need for recalibration Library for 9.5, 10.6, 11.7 >300 response factors The best option is to have a 10.6 eV lamp with an 11.7 head to use when looking for chlorinated HC
PID Analyzers, LLC copyright 2003-2011
11.7 Lamps
Customer comment the 11.7 lamp that I got from your competitor lasted just 5 days We have been making 11.7 eV lamps since 1979 and they are still better than any others out there
Our
Electronic Schematic
ADC
Keyboard Display Diagnostics
AMP
1-4 DAC to 0-1V Sensors RS232
CPU
PID Assembly
Analyzers
102+ PID
Optional
ppb mode (1-60,000 ppb) 1-3 additional sensors can be added for cost of sensor at time of purchase Sensors: CG or TCD, EC (select from > 20) Optional 9.5 & 11.7 Snap on Heads
PID Analyzers, LLC copyright 2003-2011
102+
In
IP (eV)
Response
Response
Response
Detector
Methanol Acetonitrile Toluene 10.85 11.72 8.83
10.6
0 0 9.3
11.7
2.3 1.3 11.4
FID
0.6 1.6 7
THF
Acetic Anhydride Dichloromethane Pyridine Benzene
9.45
10.35 11.35 9.32 9.25
6.2
0.9 0.2 6.1 10.0
6.3
4 6.8 7 10.0
4.5
2.5 0.6 5 6
Snap-On Modules
It is possible to have multiple heads for the 102+ or other models CG, TCD & IR use the same circuit but are setup differently (current)-can have separate heads for each Unbiased EC sensors can be interchanged by customer, recalibrated then run Readout unit will recognize type of PID or type of WB/sensor
Safety/IH Monitoring Area Monitoring VOCs+ LEL +Toxics Leak Detection for VOCs in chemical/petrochemical plants Transportation- Loading/Unloading Chemicals Monitoring Pilot Plants Laboratory Monitoring Short term; medium term monitoringusing the sleep mode