5 ITU UPGRADE
5.1 IP configuration
Note: If ITU IP address already configured, write down all the current ITU settings and
continue with the next steps.
2. The Intelligent Touchscreen Unit (ITU) boots up in service mode automatically, if it does
not have the GE SW installed. (RAFI - the Supplier of the ITU - delivers the ITUs without the
GE software).
3. In case the ITU does not start in Service mode, then press the [Service] button under
the ITU. The ITU local service main page is displayed.
Illustration 1:
Illustration 2:
5. Touch the [Fixed IP address] button ( , Illustration 3) to configure the ITU with a fixed IP
address.
Illustration 3:
On the displayed screen you will need to enter all data, including DNS server name,
otherwise the ITU will not accept them!
To change a parameter, touch the radio button of the item to modify and use the virtual
keyboard to enter the data
The [Back space] is the upper right button of the keyboard and it is use to delete the
default values.
NOTE:
• Enter the ITU IP address
• Enter the Default gateway IP address otherwise fill with zero (0.0.0.0).
To validate all your data, touch the [enter] button ( , Illustration 3). To cancel the data
NOTE:
and come back to the main page, touch the [HOME] button ( , Illustration 3).
6. In order to make the ITU IP address setting effective, you need to reset the ITU using the
Reset button ( , Illustration 1) under the box.
5. Click on Internet Explorer. Make sure you have the checkmark clicked ON .
Illustration 7:
3. If pinging is successful, you can start Internet Explorer on that computer. If not, please go
back and check the steps of setting the ITU IP address correctly.
Step 2
Step 1
Step 4
Step 3
Step 6:
Step 5
Illustration 10:
7. On the displayed window on DL (or on the computer where you do the installation from),
press the [Installation] button .
Illustration 11:
1. On the displayed window, press the [OS] button . A file selection is requested.
Illustration 12:
o CE Image
2. Insert the SW installation CDROM in the DL: CD-ROM APPLI ZENITH RTAC + DL + JEDI +
POSI + ITU
3. Click on Browse and select the file ITU\OS Image\Bootloader\NK.bin [Open] > [Upload].
The OS sw upload begins, and a sandglass will rotate on the screen. This will take a few
NOTE: minutes (about 7min). When installation is finished successfully, you can continue with the
2nd step of OS installation.
4. Click [HOME] and press the [Installation] button ( , Illustration 11).
6. Click on Browse and select the file ITU\OS Image\CEImage\NK.bin [Open] > [Upload].
The OS sw upload begins, see bargraph at bottom (about 2min), and a sandglass
NOTE: will rotate on the screen. This will take a few minutes (about 7min).
After the OS Installation is done, you need to set the ITU IP address again
8. Redo section 5.1, and then continue with the RAFI sw installation.
3. Insert the SW installation CDROM in the DL: CD-ROM APPLI ZENITH RTAC + DL + JEDI +
POSI + ITU
3. Click on Browse and select the following file itu_install_zen_xxx.zip [Open] > [Upload].
4. Once the upload is completed (about 2 minutes), reset the ITU using the Reset button ( ,
Illustration 1) under the box and remove the CD from the CD-ROM reader.
1. Verify Innova IP configuration is done. If a MacLab CardioLab station is used, verify its IP
configuration is done.
6. On the displayed window on DL (or the other computer where you do the installation
from), press the [Configuration] button .
Illustration 13:
2. Enter the data and press the [Accept] button to validate your configuration. Press the
[Home] button to come back to the previous screen.
Illustration 15:
NOTE: If you want to modify settings, press the [Accept] button to validate changes, or
[Reset] button to cancel, before pressing the [Home] button to come back to the
previous screen.
1. Depending on the customer choice, press the [Sound ON/OFF] button to turn the
sound ON or OFF.
Illustration 16:
2. Reset the ITU using the Reset button ( , Illustration 1) under the box.
In case of the MacLab or CardioLab option enabled, the screen layout of the MacLab
NOTE:
needs to be installed.
1. Press the Service button under the ITU.
Illustration 17:
6. On the window displayed, press the [Modality layout] button . A filename is requested.
Illustration 19:
7. Insert the MacLab CDRom into the MacLab CD reader. Select the zip file in the folder
ITU_CONFIG. This contains the MacLab ITU layout configuration filesMLabCommands.xml
and MLabTabs.xml (defined in MacLab OM).
8. Once the download is completed, reset the ITU using the Reset button ( , Illustration 1)
under the box.
5.5 Backup
6. On the window displayed, press the [Backup] button , this will start the Backup
procedure (a rotating sandglass will be shown).
Illustration 21:
7. When the Backup procedure is finished, you will see the link “Click to save backup file”.
Click on the link and save the file in the selected folder. (Note: Make a note where you
saved the file, it is suggested to copy this file to DL into the C:\Backup folder. To restore
you will need to find the same file location.)
9. Once the download is completed and backup done, Plug a GE Service key on the DL and
reset the Innova system.