Sunteți pe pagina 1din 95

APPLICATION FOR LICENSE TO AUTHORIZE NEAR-SURFACE LAND DISPOSAL OF LOW-LEVEL RADIOACTIVE WASTE Appendix 2.3.

1-2: Meteorological System

APPENDIX 2.3.1-2 METEOROLOGICAL SYSTEM

March 16, 2007

2.3.1-2-1

Revision 12a

APPLICATION FOR LICENSE TO AUTHORIZE NEAR-SURFACE LAND DISPOSAL OF LOW-LEVEL RADIOACTIVE WASTE Appendix 2.3.1-2: Meteorological System

TABLE OF CONTENTS
Models in use as of January 2005 are listed below. Section Cover Introduction Met One Model 014A Wind Speed Sensor Met One Model 024A Wind Direction Sensor Met One Model 090D Barometric Pressure Sensor Met One Model 083C-1 Relative Humidity w/ Platinum Temperature Sensor Met One Model 370C Rain Gauge Met One Model 96-1 Solar Radiation Sensor Novalynx Model 255-100 Analog Output Evaporation Gauge Met One Model 455A Micromet Data Logger Page

March 16, 2007

2.3.1-2-2

Revision 12a

EEMet One Instruments


Met One Instruments, Inc. Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, Oregon 97526 Tel(541) 471-7111 Fax (541) 471-7116 Regional Sales & Service: 3206 Main Street, Suite 106, Rowlett, Texas 75088 Tel(972) 412-4747 Fax (972) 412-4716 http:l/www.rnetone.corn metoneQmetone.com

Introduction

Your Meteorological Monitoring System has been completely wired and tested to specifications at the f a c t o y

When the system is installed, there should not be a need for recalibratioii.
Please refer to the System lnterccnnect Diagram(s) at the end of this manual, if required, when installifig the system.

If a datalogger is included in this system, it has been set up at the factory for this configuration. No adjustments are necessary except for time and date information.

A copy of the test certificate is enclosed, along with sensor calibration certificates, in a separate e n v e l o p .

Met One
Instruments
1600 NW Washington Blvd. Grants Pass, Oregon 97526 Telephone: 541 -471-71 1 1 Facsimile: 541-471-71 16

Regior,al Sales & Service 3206 Main St. Suite 106 Rowlett, Texas 75088 Telephone: 972-412-4715 Facsimile: 972-412-4716

PEWATl b Document No. 014A-9800

1600 Washington Blvd. Grants Pass, Oregon 97526 Telephone 541-471-7111 Facsimile 541-477-7116

Met One instruments

Regional Sales & Service 3206 Main St., Suite 106 Rowlett, Texas 75088 Telephone 972-412-4715 Facsimile 972-412-4716

0144 WIND SPEED SENSOR 1 .0 1 .1

GENERAL INFORMATION
The Met One Ol4.A Wind Speed Sensor uses a dwable 3-cup anemometer assembly and simple magnet-reed switch assembly to produce a series of contact closures whose frequency is proportional to wind speed. This sensor is usually used in corijunction with the 191 Crossarrn Assembly and a translator module, but may also b e used directiy with a variety of dataiaggers. Sensor Cable has a quick-connect connector with vinyl-jacketed, shielded cable. Cable length is given in -XX feet on each cable part number.

1.2

--

Table 1 - 1 Model O14A Wind Speed Sensor Specifications Performance Characteristics Maximum Operating Range Starting Speed Calibrated Range Accuracy Temperature Range Distance Constant* Standard (181 2 Aluminum Cup Assembly) Optional (1 708 Lexan Cup Assembly)
0-60 meterdsec or 0-125 rnph .5 metersJsec or 1 mph 0-50 melerdsec or 0-100 +I .5% or 0.25 mph -50" C to 85' C

Less than 15 feet Less than 5 feet

"The distance traveled by the air after a sharp-edged gust has occurred for the anemometer rate to reach 63% of the new speed. Electrical Characteristics Output Signal Phvsical Characteristics Weight Finish Mounting Cabling 1.5 Ibs. Anodized Use with 191 Crossarm 2-Conductor Cable, XX is cable length in feet Contact closure at frequency V= (f x 1.7892) + 1 mph

,
\

Fixtures

Optional Accessories (a)1708 Lexan Cup Assembly, Fast Response Type

014/4-9800 REV. 9/96

2.0
2.1

INSTALlATION
01 4A Wind Spesd Sensor Installation

A.
B.

Check to see that the cup assembly rotates freely (threshcld, bearing check).

Install the sensor into the fitting on the end of the 191 mounting arm. (THE E N D WITHOUT THE ALIGNMENT BUSHING) Instail just deep enough to allow cable connection. (Reference the mounting detaii in Figure 2-1). f Apply a small amount o silicone grease to t h e sst screws to prevent 'freezitlg up' in corrosive environments. Tighten the lockin!: set screws--do not over tighten. Connect the Cable Assembly to the connector receptscle on base of sensor. Secure the cable to the mounting arm using cable ties or tape.

C.
D.

Wiring

The Cable Assembly contains three wires.

Blsck

Signal
Common

Rd
WhiteIBrn

Cable Shield

01 4A-9800 REV. 9/96

0 1 4 A Wind Speed

Rev. 4/11/89

TABLE

3-1

MODEL 014A WlND SPEED SENSOR CALIBRATION

WIND VELOCITY
Speed V mph 10 20 30 40 50 60 70 80 90 100 110 120

VS

OUTPUT FREQUE3CY

S P E E D IN METERSISEC

RFS
1.284 2.846 4.409 5.972 7.535 9.098 10.661 12.224 13.787 15.350 16.913 18.476 20.039 21.602 23.165 24.728 26.291 27.854 29.41 7 30.980 32.543 34.106
35.669

STANDARD CALIBRATOR TEST POINTS

V mph

=RPFvl
16.767

+I

V m p s = RPbl
37.5067

+ ,44704

Based on equation f= ,5589 (V-I) where f is the output frequency.

V is wind speed miles per hour. RPS = cup revolution per second.
1 MPH = 0.44707 meterslsec

37.232

014A-9800 REV. 9/96

3.0
3.1

OPERATIONAL CHECK-OUT AND CALlBRallON


014A Wind Speed Sensor Check-Out

A.

Spinning the anemometer cup assembly will produce a series of pulses. To verify sensor output, monitor this signal with either a plug-in Translator Module, Datalogger or an Ohmmeter. Refer to Frequency vs. Wind Speed Table 3-1. Spin slowly and monitor output signal. A windspeed calibrator may be used to check operation at different RPM points. Inspect the cup assernbly for loose cup arms or other damage. The cup assembly cannot change calibration unlass a mechanical part has come loose or has been -broken.

E.

4.0

MAINTENANCEAND TROUBLESHOOTING
General Maintenance Schedule*

4.1

6-1 2 month intervals:


A. B, Inspect sensor for proper operation per Section 3.0. Replace Wind Speed Sensor bearings in extremely adverse environments per Section 4.6.

12-24 month intervals:

A.

Replacement of sensor bearings.

24-36 month intervals:

A.

Recommended complete factory overhaul of sensor.

*Schedule is based on average to adverse environments.

Table 4-1 TROUBLESHOOTING TABLE


Symptom

Probable Cause Faulty reed switch Faulty bearings Faulty cable

Remedy Replace reed switch

No sensor output
No sensor ouput No sensor OL?PU?

Replace bearings

CI-reck Connections

014A-9806 REV. 9/96

4.3

01 4A Wind Speed Sensor: 6-1 2 Month Periodic Service

A.

At the crossarm assembly, disconnect the Sensor Cable from the Sensor (leave the cable secured to the crossarm) and remove the Sensor from the fitting on the crossarm assembly.
Loosen the two set screws on the side of the hub and remove the anemometer cup assembly. Visually inspect the anemometer cups for cracks and breaks and make sure that each is securely attached to the cup assembly hub. Inspect the Sensor for any signs of corrosion and dust buildup. Rotate the Sensor hub assembly to make sure that it turns freely and that the Sensor bearings are not damaged. Make sure that the magnet assembly is not contacting the reed switch. (Ref. Fig. 4.1).
A moisture drain vent is located on the base of the Sensor. Make sure that this vent is clear.

8.

C .

D.
E.

F.
G 4.4

Re-install Sensor as per installation procedure (Section 2.0) and verify proper operation using procedures in Section 3.0.

0144 Wind Speed Sensor General Assembly (refer to 014A Assembly Drawing)

The following steps cover basic disassembly: A. At the crossarm assembly, disconnect the Sensor Cable from the Sensor (leave the cable secured to the crossarm) and remove the Sensor from the crossarm assembly. Loosen the two set screws and remove the anemometer cup assembly. Remove the three (3) flathead screws at the tog of the Sensor and lift out the bearing mount assembly, taking care not to break the wires.

B.

C.
4,5

Reed Switch Replacement Procedure Use the foilowing procedure to repiace Sensor Reed Switch: A. Remove bearing mount assembly as per Section 4.4. Unsolder the leads of the Reed Switch (10) and remove the switch from the mounting terminals ( 1 3). Solder the new switch onto the sides on the switch mount terminals, taking care not to stress the point where the leads enter the glass reed switch body. (Solder quickly to reduce excess heat to reed switch.) Measure the distance between the bottom of the rotating magnet and the top of the switch envelope, as shown in Figure 4 -1. The spacing should measure between .010 and ,020 inch.

B.
C.

01 48-9800 REV. 9/96

D.

Monitor the output of the translator module and spin shaft for an upscale indication, li switch seems to falter, adjust switch slightly closer to magnet. If possible, connect the shaft to an 1800 RPM motor, using a flexible coupling and verify an output of 108 mph with a 50% duty cycle. Reassemble Sensor by reversing procedure. FIGURE 4-1 : REED SWITCH INSTALLATION

E.
F.

4.6

Bearing Replacement Procedures The bearings used in 014A Sensor are special stainless steel ball bearings with a protective shield. Bearings are lubricated and sealed. Do not lubricate bearings as the lubrication will attract dust and wili form an oilldust glue. Use the following proczdure for bearing replacement:

A.

Remove bearing mount assembly as per Section 4.4. Loosen set screws(21) in magnet assembly (4), lift shaft (7) and collar (3) up and out of bearing mount (2). Be sure to retain lower spacer (19) Insert a right-angle type of toot such as an alien wrench into bearing, cock it slightly to one side and remove bearing. Remove both hearings. Clean bearing seats. Install new bearings. Be careful not to introduce dirt particles into bearings. CLEAN WANDS ONLY! DO NOT ADD LUBRICATION OF ANY KIND. Reassemble the Sensor in reverse order. Be sure to include spacers (19) over the bearings when repbcing the shaft in the bearing mount. After t h e magnet assembly (4) has been tightened, a barely perceptible amount of endplay sho~ild be fe!t when the shaft is moved up and down (approximately ,004 inch).

B.
C.

D,

E.

014.A-9800 REV. 9/96

4.7

O14A Wind Speed Sensor Repair and Recalibration Service

This service provided by the factory enables fast, economical service for the user. This repair and calibration service includes disassembly and detailed electronic components. Service includes replacement of bearings regardless of apparent condition. Service also includes replacement of the following items.

A.

Shaft
All set screws.

B.

Service also includes functional test of Sensor. Other compcnenk will be replaced as required. Additional charges for additional materials only will be added to the . basic service charge. Table 4 2 REPMCEABLE PARTS LIST PART #
101685-2 101685-4 101685-7 101715 101812 101898 860001 101812-1 2844 51 0 0 2 0 970062 860250 601250 601 2 3 0

DESCRIPTION WS SUPPORT BEARING MOUNT C O W I\hrlGNErASSY CUP ASSEMBLY (ALUM) BEAillNG SHAFT CUP ASSEMBLY, LOCllN SWITCH REED CAP FOR CONNECTOR TERMINAL HH SMlTil SPACER SET SCREW 4-40>: 118 FLAT HD.4-40X 114 SCREW

* Parts included in 2402 Rebuild Kit

01 4/4-9800 REV. 9/96

1805 SENSOR CABLE

TERMINATES IN SPADE LUGS

DASH--NO. = LENGTH IN FEET

El==-

BLX--------------

Si'GNbL

BHT OR RZD ----------C C O W O N

DASB HO. = LENGTH ZN FEET

2026 CABLE

USE VITR

BT-SET 4B AND 4C

DASB NO. = LEMGTH IN FEET

Of4A

REV. 4/4/90

Wind Speed Sensor


014A
The Model 014A Wind Speed Sensor is an accurate, durable and economical anemometer suitable for a wide range of wind study applications. It is designed for long term unattended operation in most meteorological environments. 15 feet. For greater sensitivity, the optional Lexan plastic cup assembly may be specified, which has a distance constant of less than 5 feet.

Construction
The construction of the sensor reflects the requirement for reliability and durability. Only the best corrosion resistant materials, such as stainless steels and anodized aluminum are used. The Model 014A sensor uses a quickconnect sensor cable. Cable length may extend hundreds of feet without affecting measurement performance. Accuracy, reliability and economy make the Model 014A Wind Speed Sensor an ideal choice for most applications.

Features
s Range to 100 mph s Low starting threshold s Broad temperature

operating range
s Accuracy of 1.5% s Stainless steel and

aluminum construction

Operation
For maximum operational reliability, the sensor uses a sealed magnetic read switch. This switch produces a series of contact closures at a rate proportional to wind speed. With its pulsed output, the Model 014A lends itself to applications involving both digital and analog measurement systems. The pulsed signal may be converted to standardized analog voltage and/or current output by use of translator electronics. Direct connection to a datalogger is also possible. The robust aluminum cup assembly normally supplied has a distance constant of less than

Specifications
Range: Starting Threshold: Accuracy: Distance Constant Standard: Fast Response: Operating Range: Weight: Contact Rating: Mounting: Ordering Information Standard Model: Fast Response Model: Cable: 0-100 mph 1.0 mph 0.25 mph or 1.5% FS Less than 15 ft (Metal Cup) Less than 5 ft (Lexan Cup) -50C to +70C 11 oz (.31 kg) 10 mA maximum PN 191 Crossarm 014A (Metal Cup) 014A-1 (Lexan Cup) PN 1805-xx (xx=length in ft)

Met One Instruments, Inc.

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com metone@metone.com
014A- REV. 9/1/97

Service Report
014A Wind Speed Sensor
Met One Instruments, Inc. 1600 NW Washington Blvd. Grants Pass, OR 97526
Ph. 541-471-7111 Fax 541-471-7116

E-mail: service@metone.com

Parts & Service Description


REPAIR AND FIELD MAINTENANCE
A complete repair and service department is maintained at our factory in Grants Pass, Oregon, and at our regional service facility in Dallas, Texas. Most repairs can be made in the field by a simple substitution of components. Some repairs do require recalibration following the replacement of faulty components. If calibration standards are not available, the instrument can be returned to the service facility. The turn around time is generally less than five (5) working days.

RETURN AUTHORIZATION
Prior to any product return, please contact the factory at (541) 471-7111 or e-mail service@metone.com for a return authorization number (RA).

BASIC SERVICE(S)
Ordering No. SVC0014 Description 014 WS Sensor Std. Repair Including: Replace bearings & reed switch. Check starting threshold & full scale output. Check cups against NIST standard As found Sensor Test List Price $ 70

SVC0100

$ 15

REPLACEABLE PARTS
Note: Replacement of some of the following parts will require re-calibration of the sensor. Those replacement items that require re-calibration of the sensor are indicated with an * . Instructions are provided in the Product Users Manual.

REPLACEABLE/SPARE PARTS LIST


Ref. (Dwg. Attached) 1 2 3* 4* 5 6* 7* 9 10* 13* 19* 21* 23 No Reference Part No./ Ordering No. 1685-2 1685-4 1685-7 1715 1812 1898 860001 1812-1 2844 970062 860250 601250 601645 2402 Description WS Support Bearing Mount Collar Magnet Assy Cup Assembly (Alum) Bearing Shaft Cup Assembly, Lexan Switch Reed Terminal HH Smith Spacer, .250 ID, .375 OD, .010 THK Set Screw 4-40X 1/8 Set Screw 8-32 X 3/8, SST, Brass Tip (2 required) Rebuild Kit, assy, incl. Dwg ref. 2,3,4,6,7,10,13,19,21 List Price $ 35 $ 19 $7 $ 16 $ 84 $ 16 $5 $ 84 $4 $3 $3 $ 0.25 $ 0.25 $ 95

4/2/2003

Rev D

ASSEMBLY/PARTS REFERENCE DRAWING

4/2/2003

Rev D

1600 Washington Blvd. Grants Pass, Oregon 97526


Telephone 541-471-71 11 Facsimile 541 -471 -71 16

Met One
Instruments

Regional Sales & Service 3206 Main St., Suite 106 Rowlett, Texas 75088 Telephone 972-412-4715 Facsimile 972-41 2-4716

024 WIND DIRECTION SENSOR OPERATION MANUAL

1 .1

The Met One 024A V'dind Direction Sensor uses a lightweight, air-foil vane and a potentiometer to produce an output that varies proportional to wind direction. The Sensor Cable has a quick-connect connector with vinyl-jacketed, shielded cable. Cable length is given in -XX feet on each cable part number. An 1806-XX cable is used with transiators having terminal strip connections, and an 1809-XX cabie is used with translators having circular MS type connectors.

1 .2
i

TAELE 1-4 Model 0 2 4 4 Wind Direction Sensor Specifications

Performance Azimuth Threshold Accuracy Damping Ratio

Characleristics Electrical 0-356"* Mechanical 0-360" 1.0 mph ?53 Standard 0.25 Optional 0.4

Sand,Dust,Fungus MIL-E-5272 Salt Spray MIL-E-12934 Temperature Range Delay Distance Electrical Characteristics Outp5t Signal Physical Characteristics Weight Finish Mounting Fixtures Cabling
1.5 pounds Anodized Use with 191 Crossarm 3-Conductor Shielded Type Cable, xx is cable length in feet

5 ft.

Varying resistance 0-1 0 K ohms

024A Revised 6/94

3 .O
3.1

OPERATIONAL CHECK-OUT AND CALIBWTION


024 Wind Direction Sensor Check-Out

A.

Rotating the vane in a clockwise direction as viewed from above will increase the output up to the 360" point and it will start over 0". The 024A wind direction sensor should be inspected periodically for physical damage to the vane assembly and cable connections. lnspect all vane assembly parts to be sure that they are securely fastened. Inspect the sensor connector and mating cable connector for corrosion.

B.

4.0

MAINTENANCE General Maintenance Schedule*


6-12 Month Intervals:

4.1

A.

inspect sensor for proper operaiion per Section 3.0.

24-36 Month I ~ t e r v a l s :

A.

Factory replacement of potentiometer per Section 4. Recommended complete factory overhaul of sensor,

B.

4.3

Remove the sensor from the ciossarm and remove the vane assembly. Replacement of the potentiometer will require reaiignment with respect to 180 degrees. Use the iollowing procedure to replace and realign the potentiometer. Loosen the three set screws which hold the potentiometer in the sensor housing. Pull the potentiometer up and out of the housing. Remove the three wires from the potentiometer assembly. Note the color code of the wires with respect to the pins on the potentiometer. (See the 024A Assy. Dwg.) Solder the wires to the new potentiometer and install the potentiometer in the sensor housing.

A)

B)
C)

024A Revised 6/94

I I
D

REDRAWN PER E.O. WIRE COLORS

1131

REFERENCE COUIBON CLR OR WIT

&-

SICNAL SHIELD ITEM


1

PART NO. 500351

DESCRIPTION

CONNECTOR. tl PIN. FEMALE

SENSOR END

TRANSLATOR END 980075

CABLE. 3 WIRE. SHIELDED SLEEVINC. P/4". SLEEVINC. I/B". SLEEVING. 1/4". SllRIHK SllRlNK CLEAR SHRINK

A DA@

IDENTIFY CABLE 18" FROM EACB END. NUMBER P LENGTH IN FEET.

MET ONE INSTRUMENTS


ASSY, CABLE, 024 WIND DIRECTION SENSOR

Wind Direction Sensor


024A
The Model 024A Wind Direc-tion Sensor is an accurate, durable and economical sensor suitable for a wide range of wind study applications. It is designed for long term unattended operation in most meteorological environments.

Features
s Range to 100 mph s Low starting threshold s Broad temperature

operating range
s Built-in alignment and

calibration feature
s Accuracy of 5 s Stainless steel and

aluminum construction

Operation
The sensor incorporates a precision wire-wound potentiometer for accurate resolution of wind direction. The potentiometer is directly coupled to the vane assembly. Variations in wind direction produce a corresponding varying voltage, which lends itself to both digital and analog measurement systems. Accuracy, reliability and economy make the Model 024A Wind Direction Sensor an ideal choice for most applications.

Specification
Range: Starting Threshold: Accuracy: Delay Distance: Damping Ratio Standard: Fast Response: Potentiometer Sand, Dust, Fungus: Salt Spray: Electrical Range: Operating Range: Weight: Mounting: Ordering Information Standard Model: Fast Response Model: Cable: 0 - 360 1.0 mph 5 Less than 5 ft 0.25 (Metal Vane) 0.4 (Foam Vane) MIL-E-5272 MIL-E-12934 0 - 360 0 - 540 with approp. translator -50C to +70C 1 lb 2 oz PN 191 Crossarm 024A (Metal Vane) 024A-1 (Lexan Vane) PN 1806-xx (xx=length in ft)

Construction
The construction of the sensor reflects the requirement for reliability and durability. Only the best corrosion resistant materials, such as stainless steels and anodized aluminum are used. The potentiometer meets stringent military specifications for sand, dust, salt spray and fungus resistance. The Model 024A sensor uses a quick-connect sensor cable. Cable length may extend hundreds of feet without affecting measurement performance.

Met One Instruments, Inc.

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com metone@metone.com
024A- REV. 8-1-2000

Service Report
024A Wind Direction Sensor
Met One Instruments, Inc.
1600 NW Washington Blvd. Grants Pass, OR 97526
Ph. 541-471-7111 Fax 541-471-7116

E-mail: service@metone.com

Parts & Service Description


REPAIR AND FIELD MAINTENANCE
A complete repair and service department is maintained at our factory in Grants Pass, Oregon, and at our regional service facility in Dallas, Texas. Most repairs can be made in the field by a simple substitution of components. Some repairs do require recalibration following replacement of faulty components. If calibration standards are not available, the instrument can be returned to the service facility. The turn around time is generally less than five (5) working days.

RETURN AUTHORIZATION
Prior to returning any product, please contact the factory at (541)471-7111 or e-mail service@metone.com for a return authorization number (RA).

BASIC SERVICE(S)
Ordering No. SVC0039 Description Basic Sensor Service Including: Inspection and performance test. Test starting threshold, NIST calibration, and linearity. As found Sensor Test List Price $ 50

SVC0100

$ 15

ADDITIONAL SERVICES Replacement of potentiometer and other components is performed on an as-needed basis. Replacement parts are billed based on the following parts price list (below).

REPLACEABLE PARTS
Note: Replacement of some of the following parts will require re-calibration of the sensor. Those replacement items that require recalibration of the sensor are indicated with an * Instructions are provided in the Product Users Manual.

REPLACEABLE/SPARE PARTS LIST (Aluminum Vane, Model 024A)


Ref. (Dwg. Attached A B C D E A-E F G -L K M Part No./ Ordering No. 2089 2088 1685-10* 1814-1 1057 2105* 2017* 860015 601625 601680 601070 2748 Description Aluminum Vane Vane Arm For Aluminum Vane Hub Counterweight Arm for Aluminum Tail Counterweight Aluminum Vane Assembly, Complete Potentiometer Assembly Shoulder Screw 8-32 x Set Screw 8-32 x 3/8 Set Screw 2-56 x 1/8 Set Screw Label List Price $ 20 $ 13 $ 30 $5 $ 17 $ 95 $ 130 $4 $1 $1 $1 No Charge

5-14-2001

REV A

REPLACEABLE/SPARE PARTS LIST (Foam Vane, Model 024A-1)


Ref. (Dwg. Attached A,B C D E A-E F G -L K M Part No./ Ordering No. 1286 1685-10* 1814-2 1057 2106* 2017* 860015 601625 601680 601070 2748 Description Foam Vane & Arm Assembly Hub Counterweight Arm for Foam Tail Counterweight Foam Vane Assembly, Complete Potentiometer Assembly Shoulder Screw 8-32 x Set Screw 8-32 x 3/8 Set Screw 2-56 x 1/8 Set Screw Label List Price $ 62 $ 30 $5 $ 17 $ 95 $ 130 $4 $1 $1 $1 No Charge

ASSEMBLY/PARTS REFERENCE DRAWING

5-14-2001

REV A

Barometric Pressure Sensors


090D 091
Barometric Pressure Sensors convert absolute atmospheric pressure into a linear, proportional voltage, which may be used in any meteorological program.

Features
s Compact size s Weatherproof enclosure s Remote output s Permanent calibration s Robust construction These sensors are inherently stable devices that do not require periodic service or routine recalibration.

090D 091

Operation
The enclosure houses a solidstate pressure transducer, with linearization and amplification electronics. The Model 090D is housed in a heavy duty fiberglass enclosure, suitable for harsh and severe environments. A hose barb is provided to enable the connection of a 1/4 sampling tube to the outside environment. The Model 091 is contained within a small polycarbonate enclosure, and may be mounted outside or inside a building or other enclosure. Small inlet holes allow the atmospheric pressure access to the sensing element. The standard range of the 090D/ 091 is 26-32 in. Hg,* suitable for elevations sea level to 1500 ft. Other ranges are available.

Specifications
Performance Resolution: Temp. Operating Range: Temp. Compensated Range: Accuracy: Electrical Characteristics Power Requirement: Sensor Output: Physical Characteristics 090D 091 Weight: Dimensions: Weight: Dimensions: 2 lbs, 5 oz (1.05 Kg) 5.5 x 5 x 7.5 in (14 x 12 x 19 cm) 8.8 oz. (250 g) 2.1 x 3.2 x 5 in (5.4 x 8.3 x 13 cm) 11 mA @ 12 VDC, Typical 0-1 VDC, Standard 0-5 VDC, Optional Infinite -40C to 65C -18C to 65C 0.04 in Hg (1.35 mbar) or 0.125% FS

Ordering Information Specify elevation Specify output voltage #1169-xx (xx = length in feet) Specify length in feet

Cable:

Specifications subject to change without notice. *Conversions: 1 in. Hg = 3.3864 kPa, 1 in. Hg = 33.864 mbar, 1 in. Hg = 25.4 mm/Hg

Met One Instruments, Inc.

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com
090D/091 - 7/25/97

Service Report
090D Barometric Pressure Sensor
Met One Instruments, Inc. 1600 NW Washington Blvd. Grants Pass, OR 97526
Ph. 541-471-7111 Fax 541-471-7116

E-mail: service@metone.com

Parts & Service Description


REPAIR AND FIELD MAINTENANCE
A complete repair and service department is maintained at our factory in Grants Pass, Oregon, and at our regional service facility in Dallas, Texas. Most repairs can be made in the field by a simple substitution of components. Some repairs do require recalibration following the replacement of faulty components. If calibration standards are not available, the instrument can be returned to the service facility. The turn around time is generally less than five (5) working days.

RETURN AUTHORIZATION
Prior to any product return, please contact the factory at (541)471-7111 or e-mail service@metone.com for a return authorization number (RA).

BASIC SERVICE(S)
Ordering No. SVC0057 Description 090 Standard Repair Including: Calibration to NIST standard List Price $ 60

REPLACEABLE PARTS
Note: Replacement of some of the following parts will require re-calibration of the sensor. Those replacement itmes that require re-calibration of the sensor are indicated with an * . Instructions are provided in the Product Users Manual.

REPLACEABLE/SPARE PARTS LIST


Part No./ Ordering No. 1931-1 2696 2540* 960200 580281 720050 971106 580285 480510 600146 Description Case Mounting Plate Circuit Card Assy ( specify range ) Tygon Tubing, ID x 3/8 OD x 4 Elbow, O-Ring Seal Nylon Washer Bulkhead Fitting, Barb to Barb Strain Relief U-Bolts w/ clamp List Price $ 67 $ 9.45 $ 265 $ 0.35 $ 2.40 $ 0.15 $ 1.86 $ 4.74 $ 4.11 $ 3.06

ASSEMBLY/PARTS REFERENCE DRAWING (Next Page)

9-10-2001

Rev A

9-10-2001

Rev A

elative Hermi ity

W/

Platinum Tern erature Sensor

Qseralion Manual

083C SERIES RELATIVE HUMlDlTY/TEMPERATURESENSORS

1.0 1. I

GENERAL INFORMATION I h e 083C Sensor contains an extremely accurate and sensitive relative humidity sensor which responds to the full range of 0-100% humidity. Response is linear with small hysteresis and negligible temperature dependence. The sensor is designed to be housed in a radiation shield when used outdoors. Certain models also contain a high-accuracy linearized air temperature sensor, permitting simultaneous measurement of relative humidity and temperature. The Sensor Cable uses quick-connect connectors with vinyl-jacketed shielded cable. Cable length is given in -XX feet on each cable part number. The cable part number depends on which radiation sheld the sensor is mounted in. The 083C-17-1 sensor is mounted in the 071 Radiation Shield and uses the 1873-XX cable. The 083C-17-35 is mounted in the 0738 or 075B Radiation Shield and uses a 2348-XX cable. Both of these cables are documented at the end of this manual section. The 083C-17-6 sensor mounts in the 0768 Radiation Shield and uses a 2144-XX cable, which is documented in the 076B manual section. Table 1.I Model 083 Relative Humidity Sensor Specifications Sensing Element Thin-film capacitor

1.2

Temperature Range Response Time

-40F to 175" F 10 to 20 seconds at 68F 90% of final RH value

Accuracy Hysteresis

Better than +3% For 0% to 100% to 0% excursion less than +I %


+0.07% per 1C

Temperature Coefficient Output Input Power

- 1V

at full scale (most models)

12V DC

+ 2V, 12 ma (most models)

Table 1.2 Temperature Sensor Specifications for 083C-17

Range Ice Point Resistance, RO Temperature Coefficient Time Constant

-SO0 to +50C (most models)

1000R +4S2 (F.4%) (standard) .00375R/L2/"C


10 sec.

2.0

INSTALLATION If sensor comes mounted in a radiation shield, refer to radiation shield manual section for mounting details. Sensors not furnished in a radiation shield should be mounted in a representative location having good air flow and shaded from sunlight or other heat radiation sources that would affect measurement of relative humidity. OPERATIONAL CHECK-OUT AND CALIBRATION Relative Humiditv Measurement Relative Humidity Sensor Check-out 1.
2.

2.1

3.0

3.1

3.2

To verify correct wiring and as a rough test of sensor operation, blow on the sensor. The relative humidity will rise to a higher level. The Relative Humidity Sensor has been calibrated at the factory and will not change unless it is damaged. To check for proper operation of the sensor it is advised that the output signal be checked against a local weather service facility. Exact correlation is not to be expected due to atmospheric and geographical variations.

3.3

Temperature Sensor
1.

Compare actual readings with precision mercury thermometer. As an alternative, measure sensor resistance with a Lo Current Digital Ohm Meter and compare readings of temperature vs resistance. The temperature sensor changes resistance in accordance with the equation:

2.

In applying this formula the actual resistance at ice point is used for calculations, i.e. if the sensor is exactly 1,000 ohms then the equation is

Refer to calibration certificate for exact reading at ice point. 4.0


4.1

MAINTENANCEAND TROUBLE SHOOTING General Maintenance Schedule*


6

12 Month Intervals:

A.
B.

Inspect sensors for proper operation per Section 3.0. Clean Relative Humidity sensor element per Section 4.2A.

*Schedule is based on averaae to a 4.2

m environments.

083 Relative Humidity Sensor Maintenance and Calibration

Warning:

The sensor can be miscalibrated or permanently damaged through improper acts. Do not attempt a repair or calibration if you are unsure of the procedure. Do not touch if you do not know how. This instrument should operate for an extended period of time with a minimum of care or maintenance. If parts or maintenance assistance are required, contact Met One Instruments Inc. Obtain shipping instructions before returning any unit.

A.

Maintenance Cleaning the Sensor Element. Unscrew the filter. Dust and other particles may be removed by gently blowing on the sensor chip. DO NOT USE COMPRESSED AIR. After dusting, the sensor element may be wiped clean with a soft brush dipped in distilled water. DO NOT USE DETERGENTS. DO NOT APPLY POWER TO THE SENSOR WHEN CLEANING, and do not re-connect power to the sensor until the element has dried. CAUTION: NEVER TOUCH THE SENSOR CHIP WITH BARE HANDS.

The life of the sensor is related to the environment in which it operates. In a pure air and water vapor surrounding, the sensor element will have an indefinite life. The presence of chemical pollutants in the environment may corrode the materials of the sensor chip. The polymer material is resistant to most chemical attacks, but the metal electrodes, are sensitive to corrosion effects, particularly when a DC voltage is applied to the sensor. The most harmful pollutant has been sulphur dioxide absorption in small soot particles. When such particles fall on the thin metal electrode, they may, if water condensation is present, form traces of sulphuric acid to corrode the surface of the sensor. For these reasons, a careful cleaning as described in the preceding paragraph is recommended whenever the sensor has been exposed to corrosive pollutants. Also, a periodic cleaning every two weeks with an atomizer of distilled water, thoroughly washing the chip clean, may remove harmful particles before they can damage the sensor. Be sure that no power is applied when washing the chip and that power remains off until after the chip has dried. Replacement of Sensor Element. If the sensor element has been damaged, it can be easily replaced. Disconnect power to the probe. Unscrew the sintered bronze filter. Un-solder the old chip and solder a new one in its place. The sensor chip is very delicate, so observe the following precautions. DO NOT TOUCH THE CHIP WITH BARE HANDS. Handle the chip only by gripping its lead with pliers. When soldering, hold the lead with the pliers to prevent the heat from the soldering operation from damaging the chip. Do not bump the chip when reinstating the protective grid.

3.
B.

After replacement of the sensor element, the humidity probe must be recalibrated.

Humiditv Probe Calibratio~.


1.

Before attempting to recalibrate the probe, make sure that the translator module containing the signal conditioning electronics is still properly calibrated. If 0% and 100% do not produce corresponding readings on the indicator, recalibrate the translator module. The calibration method described in this instruction manual is based on the constant water vapor pressure over saturated salt solutions and constant temperature. Materials used for the calibration are Lithium Chloride (LiCI) and Sodium Chloride (NaCI). The former creates a humidity of approximately 13% and the latter approximately 76% in 68F (20C) ambient temperature. Both of these chemical agents are available from drug stores, pharmacies, and other chemical suppliers. To guarantee accurate calibration, the salts must be of high purity.

2.

TEST EQUIPMENT REQUIRED:


2 Calibration Bottles:

HM-111-CG-L and HM-111-CB-N Lithium Chloride Salts, Reagent Grade 1 Sodium Chloride, Reagent Grade 1 Thermometer to measure ambient temperature Distilled water
3.

Pre~arationsfor Calibration Cithil~mChloride: 13% RH


1. 2.

3.
4.

5.

Open the bottle which has the label LiCI. Pour Lithium Chloride into the bottle up to the level marked by the line on the bottle. Slowly pour distilled water over the chloride until all the chloride is thoroughly wet and a clear layer of water can be noticed above the chloride. Close the bottle tightly and check that the rubber plug is tight in its sleeve on the cork. Let the bottle cool down to ambient temperature, so that the liquid and the atmosphere above the liquid in the bottle has the same temperature as the room atmosphere. The relative humidity value is than about 13% RH. For accurate work, let the cooling go on for 24 hours.

Sodium Chloride: 76% RH


1. 2.

3.
4.

Open the other bottle marked NaCI. Pour sodium chloride into the bottle up to the level marked by the line on the bottle. Pour distilled water over the chloride until a 1 cm. thick layer of water is above the chloride. Close the bottle tightly and wait for 24 hours. After that time, the relative humidity should be approximately 97%.

The calibration bottles can be used for up to one year without changing fresh chemicals. The bottles should be stored in a place with constant temperature, so as to have them ready for use with just a short preparation time. It is not necessary to shake the bottle with salt solution before use. Care should be taken to see that there are no droplets of salt solution inside the mouth piece of the bottle. This might affect the accuracy of the calibration. Do not get any salt solution on the sensor element directly.

TABLE 4-1 Calibration Tables LITHIUM CHLORIDE Ambient Temperature O C Calibration Value % RH
10 14.3 15 13.8
20 25

30 12.8

35 12.7

40 12.6

13.4

13.0

SODIUM CHLORIDE Ambient Temperature O C Calibration Value % RH


10 15
20 25

30

35

40

75.2

75.3

75.5

75.8

75.6

75.5

75.4

C.

CALIBRATION FOR LOW HUMIDITY (13% RH)


1.

Unscrew the sintered bronze filter. Do not bump the sensor element while removing the grid. Take the lithium chloride (LiCI) bottle, pull out its rubber plug, and push the sensor probe in its place in the cork's sleeve. The sleeve fitted with a safety flange prohibits the probe from falling through. Read the ambient room temperature. Note the humidity percentage from the lithium chloride calibration table, which corresponds to the temperature in question. After 1 hour, read the humidity value, If the reading differs from the table value, adjust R15, zero adjust. After use, close the bottle tightly with the rubber plug.
t

2.

3.
4.

5.
6.

D.

CALIBRATION FOR HIGH HUMIDITY (76%) Repeat the calibration procedure as described above, but now using the sodium chloride.

E.

Repeat steps C and D until no further adjustments are required.

5 1 O O X I RELATIVE HUMIDiTY CALIBRATOR

1.I

Suitable for ell probes. Calibration by means of iithium chloride end sodium chicride saturated salt solutions. Bottles for salt solutions in metal box providing stable temperature. Due to the minimal space o air above the salt solution no ventilation is f needed. Solid construction. Humidity and Temperature scale for each salt solution printed on the box lid. A thermometer situated between the salt bottles in the box provides a very reliable calibration.

Calibrztor

before setting up

INSTRUCTIONS FOR USE PREPARATIONSFOR CALIBRATION Open the lid of the metal box by removing the screws. Pick up the rubber plugs from the box. Cut the perforated holes in the plastic lids with a knife. Preparing the saturated salt solution: a. Pour distilled water in the salt bottle labelled LiCl up to the lower edge of the black guide line.

Solution Level

\ Level Water
Salt Solution Bottle

b.

c.

Add lithium chloride carefully in small quantities until the solution reaches the level of the upper edge of the guide line simultaneously stirring the solution which is getting quite warm. The other solution sodium chloride NaCl is made in the same way. When NaCl is dissolving the solution is cooling down.
t

Place the plastic lids on the bottles and put the plugs tightly into the calibration hole.

Close the metal box by fastening the lid with the screws.
Let the calibrator stay closed for 24 hours to let the temperature differences stabilize. The salts will hold good in the bottles for about one year.

1
+12 VOLT POWER
'

EC IT I SR I PC

PROD REL EO 1191

ITEM

PART NO. 500351 480500 600193

DESCRIPTION

W Y 1 I
8

1
2

1 1
I

1 1
I
I

CONNECTOR. 6 PIN. FEMALE CLAMP LUG. SPADE. I6 CABLE. 5 COND.. SLEEVINC. 1 4 . /" SHIELDED SHRINK

SENSOR END

TRANSLATOR END

900075 980510 980000

7
8

SLEEVING. 1 8 . SHRINK /" WIRE. 22 AWC. IWT/BRN SLEEVINC. 1 4 . CLEAR SHRINK /"

IAIR I
I

6"

AR /

SOLDER CUP VIEW

IDENTIFY CABLE 18" FROM EACH END. DASH NUMBER = LENGTH IN FEET,

I
I

MET ONE INSTRUMENTS


ASSY. CABLE, RELATIVE

HUMIDITY AND TEMPERATURE


sE FH M.

lDYGN]. 234-8

Pi7
A

M
DOCUMENT 090D-9800

1600 Washington Blvd. Grants Pass, Oregon 97526 Telephone 541-471-7111


Facsimile 541-471-71 16

Regional Sales & Service 3206 Main St., Suite 106 Rowlett, Texas 75088
Telephone 972-412-4715 Facsimile 972-412-4716

Met One Instruments

Barometric Pressure Sensor Model 090D Operation Manual


1.0
1 .I

GENERAL INFORMATION
090D Barometric Pressure Sensor uses an active solid-state device to sense barometric pressure. Self-contained electronics provide a regulated voltage to the solid state sensor and amplification for the signal output.
A 1169-XX Sensor Cable is a 4-conductor shielded, vinyl jacketed cable. Length is given in -XX feet on each cable part number label.

1.2

TABLE 1-1 Model 0900-26132-1 Pressure Senscr S~ecifications

Performance
Calibrated Range Calibrated Operating Range Operating temperature range Resolution Accuracy Accuracy Output
26-32" (standard)* -18C to +50C -40C to +50C Infinite 20.04 in Hg (1-1.35 mb) or f0.125?& FS 0-1V DC (standard)*

*Refer to model number of sensor. Example: 090D -26132 -1


Basic Mod # Range ("Hg) Output Voltage (In this example, the sensor output is 0-lv ior a range of 25 to 32"

/ -

Hg)

Electrical

Characteristics
11 m a @ 12VDC 0-1 VDC Standard 0-5 VDC Optional

Power Requirement Sensor Output

Physical Characteristics
Weight Dimensions

2 Ibs. 5 oz. (1.05 Kg) 5.5" x 5" x 7.5" (14x12~19cm)

4.0

MAIMENANCE AND TROUBLESHOOTING


General Maintenance Schedule. Inspect pressure inlet port occasionaily to insure it is free of obstruction. A. other periodic maintenance or calibration is required. 8. lnspect sensor for proper operation per Section 3.1 No

4.1

4.2

09CD Pressure Sensor Maintenance. The pressure sensor is an inherently stable


device that does not require periodic service or recalibration. Should service or recalibration become necessary, the sensor must be returned to the factory. Always inspect Model 090D Pressure Sensor to make sure that inlet port is clean and free from obstructions.

RODFL 0 9 0 D BAROM~TKIC PR5+.Rk CABLE CCNb4EC-T I ON5

5ENWR

(See Section 2.2)

1600 Washington Blvd. Grants Pass, Oregon 97526


Telephone 503-471-71 11 Facsimile 553-471-7116

Regional Sales & Service 3206 Main St., Suite 106 Rowlett, Texas 75088
Telephone 214-412-4747 Facsimile 214-412-4716

Met One

Instruments

Model 370C/372C Tipping Bucket Rain Gauge is an accurate, sensitive and lowmaintenance sensor designed to measure rainfall on a continuous basis. Water does not collect in the sensor, but is drained each time an internal bucket fills with 0.01 inch of rainfall (standard calibration). At this time, a switch closure pulse is also sent to the translator module or datalogger for counting. The sensor is calibrated at shipment and requires no adjustments after mounting. Sensor Cable is a vinyl-jacketed 2-conductor shielded cable connecting to the sensor via an internal terminal strip. Cable length is designated in -xx feet on each cable part number label. Table 1-1 Model 370@/372C Rainfall Sensor Soecifications

Calibration (std) Calibration (options) Calibration (Model 3726) Accuracy Switch Type Mounting

.01" Rain per switch c~osure 0.2rnn-1, 0.25rnm, O.5rnm

+ I % at 1" to 3" per hour aC "" 7 ,


Magnet & Reed

3 Pads for I / 4 " bolts ~n 9-21/32?" (9.66") circle diameter

Dimensions

13-114" high, 8" diameter not including mounting pads 6.7 lbs13 kg (10 Ibs shipping wlcables)

Weight, less cables

INSTALLATION Chmse a site where the height of any nearby trees or other objects above the sensor is no more than about twice their distance from the sensor. A uniform surrounding of objects (such as an orchard) is beneficial as a wind break. Nonuniform surroundings (such as a nearby building) creates turbulence which affects accuracy. Using washers for shims, mount the sensor level on a platform or tower, using the built-in level as an aid; Three 1/4" diameter boits are used to mount the unit on a 92 4 1 3 2 V o l t circle.

2.3

Remove shipping restraint (This may be tape, rubber band, or similar ifern) from sensor bucket and verify that bucket RIGVeS freely and that all adjusting screws are tight. Connect the signal cable lugs to the terminal strip i f not connected already. Polarity is not important.

2.5
b

Wepiace cover on sensor, tightening screws at base. Route cable to transaator or atalogger. If this rain gauge is part of a Met One Instruments weather station, refer to System Interconnect Diagram for connection of signal cable to translator or datalogger.

2.6

3.1

The sensor is factory calibrated; recalibration is not required unless damage has occurred or the adjustment screws have loosened. To check or recalibrate, peeorm the following steps: a b. Check to be sure the sensor is level.

Wet the mechanism and tipping bucket assembly. Using a graduated cylinder, slowly pour the measured quantity of water through the inner funnel to the tipping bucket, which should then tip, Repeat for the alternate bucket. If both buckets tip when filled with the measured quafitity of water, the sensor is properly calibrated. If they do not, recalibrate as follows:
Release the lock nuts on the cup adjustments. Move the adjustment screws down to a position that would place the bucket far out o calibration. f 3. Allow the measured quantity of water to enter the bucket. 4 . Turn the cup adjustment screw up until the bucket assembly tips, Tighten the lock nut. 5. Repeat atsteps 3 and 4 for the opposite bucket. easure the quantity of water necessary to tip each bucket several times to ensure proper calibration.
1. 2,

3.2

After installation and calibration (if necessary), replace the cover on the gauge.

Calibration Quantiirites

TIP Caiibration
0.01" (standard) 0.2mm 0.25mm 0.5mm (klodef 3726)

Water Quantity 8.24 milliliters 6.49 milliliters 8.1 I milliliters 46.2 millili'rers

4.1

General

aintenance Schedule*:

At six month intervals, perform the foliowing steps:

a b.
c.

Clean sensor funnel and buckets. Do NOT lubricate the pivots, as any lubricant may attract dust end dirtand cause
wear of the jewel bearings. Verify that buckets move freely and that translator card or databgger regisie7s 0.01'' (or as calibrated) for each bucket tip.
to adverse environments.

"Based on

Model 370@/3726 Rain Gauge Parts List


Part No. Assembly, Tip Bucket (.OIn, .2mm, .25mrn) Assembly, Tip Bucket (0.5mm, Model 372) Pin, Pivot Barrier, Strip 3 pos. Nut, Crown, Nylon #8-32 Screen, Base Screen, Primary Top Clamp, Liquid-Tight Reed Switch Cartridge Adjustable Magnet Bracket Lightning Protection Diode Standard Cable Assembly Cable Assembly (for use with Autornei) Screen, Secondary Assembly, HousingJFunnel 8" Fmt

FIGURE 1

Precipitation Gauges
300 Series
The accurate measurement of rain and snow precipitation remains one of the most basic elements of meteorology. To enable accurate measurement of precipitation in all environments, Met One Instruments provides a series of instruments incorporating a tipping bucket mechanism. The tipping bucket design allows accurate, repeatable measurements, requires no regular operator maintenance, and is economical and proven in operation.

Features
s Jeweled bearings s Teflon coated bucket s Reed switch s Self-emptying s Corrosion resistant materials s Quality construction

Each model in the series is optimized to meet a particular site and sampling requirement.

The Model 370 Rain Gauge

Operation
A dual-chambered tipping bucket assembly is located below the collection funnel. When a precise amount of precipitation has been collected in one side of the bucket, gravity tips the assembly and activates a reed switch. A momentary electrical contact closure through the switch is provided for each increment of rainfall. The sample is discharged through the base of the gauge. For environments that can typically expect a significant amount of frozen precipitation, internal sensor heaters are available. The heating elements are thermostatically controlled to melt and measure the water content of snow and frozen rain, but to avoid evaporative loss. the sample. The bucket pivots are precision machined and fitted with jeweled bearings to reduce wear and friction. The funnel is powder coated aluminum and has two screens for preventing leaves and other debris from entering or clogging the gauge. A circular bubble-level and adjustable feet facilitate proper mounting of the unit. Major components are finished in catalyzed polyurethane paint, with a color and texture chosen to allow the sensor to blend into the environment.

Construction
The heavy machined aluminum base provides a stable platform for the tipping assembly. The bucket is made from stainless steel and is Teflon coated to prevent retention of

Met One Instruments, Inc.

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com metone@metone.com
Precipitation Gauges - REV. 9/1/97

Specifications
Rain Gauges
Model Number 370 372 380 382 Funnel Diameter 8 in (20.3 cm) 8 in (20.3 cm) 12 in (30.5 cm) 12 in (30.5 cm) Standard Calibration 0.01 in 0.5 mm 0.01 in 0.1 mm Optional Calibration 0.2 or 0.25 mm N/A 0.2 or 0.25 mm N/A

Rain and Snow Gauges


Model Number 375 376 377 379 385 386 387 389 Funnel Diameter 8 in (20.3 cm) 8 in (20.3 cm) 8 in (20.3 cm) 8 in (20.3 cm) 12 in (30.5 cm) 12 in (30.5 cm) 12 in (30.5 cm) 12 in (30.5 cm) Standard Calibration 0.01 in 0.01 in 0.5 mm 0.5 mm 0.01 in 0.01 in 0.1 mm 0.1 mm Funnel Heater Base Heater Funnel Base -50C to +50C at 0.5/hr at 1 to 3/hr Type Rating 8 in Gauges 12 in Gauges 8 in Rain Gauges 12 in Rain Gauges 8 in Rain/Snow 12 in Rain/Snow 8 in Rain Gauges 12 in Rain Gauges 8 in Rain/Snow 12 in Rain/Snow Optional Calibration 0.2 or 0.25 mm 0.2 or 0.25 mm N/A N/A 0.2 or 0.25 mm 0.2 or 0.25 mm N/A N/A Heater Voltage 115 VAC, 315 watts 220 VAC, 315 watts 115 VAC, 315 watts 220 VAC, 315 watts 115 VAC, 315 watts 220 VAC, 315 watts 115 VAC, 315 watts 220 VAC, 315 watts

Thermostat Set Point: Heaters : Operating Temperature: Accuracy: Switch: Height: Weight: (not including cabling)

40F (4.4C) 40F (4.4C) 240 watts, composite with built-in thermostat 75 watts, composite with built-in thermostat 0.5% 1.0% Reed 10 mA, 28 VDC 18 in (46 cm) 20.5 in (52 cm) 6 lbs (2.7 kg) 7.5 lbs (3.4 kg) 6.5 lbs (3 kg) 11.5 lbs (5.2 kg) 8.5 lbs (3.9 kg) 10 lbs (4.5 kg) 9 lbs (4 kg) 14 lbs (6.4 kg)

Shipping Weight: (not including cabling)

Finish:

White gloss/beige textured powder coat and clear anodized aluminum

Ordering Information
Specify Model Number, calibration factor, cable length(s), and accessories Cable: Signal PN 1566-xx (xx=length in feet) Power (as required) PN 2517-xx (xx=length in feet)

Accessories
Model 820440 Wind Screen: The improved Alter-design screen is constructed of 32 free-swinging, separated leaves. It can greatly improve the accuracy of the precipitation catch by reducing local turbulence.

1600 Washington Blvd. Grants Pass, Oregon 97526


Telephone 541-471-7111 Facsimile 541-471-7116

Regional Sales & Service 3206 Main St,, Suite 106 Rowlett, Texas 75088
Telephone 214-412-4747 Facsimile 21 4-412-4716
0

Met One Instruments

096-1 SOLAR WADlATlOM SENSOR

GENERAL INFORMATION
The Model 096-1 Solar Radiation Sensor is an accurate and sensitive sensor using a LICOR sensing element and designed for the continous measurement of solar radiation. Typically, the sensor is mounted with the 1259 Mounting Plate. Spectral Response. The relative spectral response of the silicon photodiode does not extend uniiormly o w the full solar radiation range, A. typical respcnse ciirve is presentee' In Figure 1 . I . .The response is very low at 0.4 pm and then iccreases nearly linear to a cutoff near 1 .I pm. Chani;es in the spectral distribution of the incident lighr, coupled with the non-uniform spectral respoiise, can cause errors in the photodiode output. Hull3 shows that in the 0.4 to 0.7 pm range, the spectral distribution of sun plus sky radiation on a horizonial sirrfsse is remarkably constant even when clear and overcasi days are ccmpared. However, Gates' indicates that the major change in spectral distribution of solar radiation occurs in the near infared where water vapor absorption takes place on cloudy days. Data collected at low solar elevations can show sig~ificant error because of aitered spectral distribution which changes in atmospheric Iransmission. This is a small part oi the daiiy total so the possible observed error usually has an insignificant effect on daily integrations. The area under the spectral irradiance curve of the source is dirzctiy porportional to the energy received by a horizontal surfsce. Under specific but typic4 conditions, energy received on a completeiy overcast day has beer, estimated to be 11.3% of that received on a clear day, Wheri both spectral distributions are weighied according te a typical responsa curve of a silicon photodiode, the response on this cloudy day is 12.6% Therefore, errors incurred under different sky conditions, due to the spectral response of the photodiode, will be smail. The field tests of Federer and Tanner1 and Kerr, Thurteil and Tanner4 confirm this conclusion. Calibration. The 096-1 Pyranometer has been calibrated against an Eppley Precision Spectral Pyramonmeter (PSP) of which the calibration is periodically confirmed. The calibration was performed under daylight conditions by a computer sampling of instantaneous readings from the Eppley and LI-COR gyranometers. Instantaneous readings were taken continuously for 10 minutes and then averaged. Sequential ten minute averaging periods were run from sunup to sundown for 3-4 days. These ten minute averages were then evaluated and used to compute an average calibration constant. The uncertainty of calibration is f 5%. Table 1-1 Model 096-1 Solar Radiation Sensor Specifications Calibration Calibrated against an Eppley Pyranometer under natural daylight clear conditions. Absolute accuracy under these conditions is k5%. All sensors are calibrated to wiihin 1% of each other. *

096-1 SR Rev. 7/93

Sensitivity Linearity Stability Response Time (1 0-9096j Ternpersture D~pendence Cosine Correciion Azimuth Error Sensor Case

Typically 80 microamp/1000 watts rn-2. Ivlaximum deviation of 1% to 3000 watts rn-2. Less than 2% change over a 1 yesr period.

k .15% per

O C

rnaxirnlim

Cosine corrected up to 82"angle of incidence. Less than 1 % over 366" at 45" elevation. Weather-proof anodized aluminum case with diffuser and stainless steol hardware, Precision level suppiied.
A 1289 Mounting Plete is provided,

Mounting
2.0

INSTALLATION
096-1 Solar Radialion Sensor lnsiallation

2.1

A.

Typically, ?he sensor is mounted to the 1289 Mounting Plate. Refer to Figure 2-1. Using a 1552 Mounting Cianp, the radiation sensor and mounting plsie may be directly mounted to Met One Instrumenis' Model 191 Mcunling Arm, or similar. For proper operation it is necessary that the sensor be lwei. Level the sensor using the three outer screws. Lock into piace using three inner screws. The sensor is supplied with an sliached 1832 cable. The white or red wire is the + signal. The black wire is the common.

B.

C.

3.0

096-1 SOLAR RADIATION SENSOR CHECK-OUT

3.1

Model 096-1 Solar Radiation Sensor has been calibrated at the factory. It will not change unless it is damaged. To check for proper operation of the sensor, expose the sensor to bright sunlight and check datalogger or translator for reasonable output, and then completely cover the sensor with a black tape and check for an output of near zero.

3.2

When this sensor is used with a Met One Instruments translator, the trans!atcr calibration is matched to the individual sensor. If used with dataloggers or other recording devices the use of a terminaiing resistor is required. Terminating Resistor
A.

3,3

The resistor is used to convert ths current output signal to a voltege output signal, ar,d is required when the scnsor is connected to millivolt recording dsvices.
A precision 100-0 crl50C res'stor has been supplied with the senscr io allow

E.

for the correct in~erfacebetwaen !he sensor and millivoll recording e!ec;ronics (no1 reauired when the 096-1 is iised with the hilet One Irisiiumeqts Traris!a?or circuit cards). Place this resistor in parallel with the signs1 le&s from the sensor. The outpur signal is then equai to: Mv=I*R Where: M = Output microvolts v I = Output signal in microamps R = Resis;ance in ohms of terminating resistor

.-

4.I

General Itlaintenance Schedule*


A

lcspect sensor for proper operation as per Section 3.9. Clean sensor element mcnthiy usicg ciean rag or tissue.

E.

*Schedule is based on average to adverse environments.

REFEFEITES
1. Federer, C.A., and C.B. Tanner,lSSS. A sirnp!e inlegrating pyranometer for measuring daiiy solar

radiarion. J. Geophys, Res. 70, 2301-2306. 2. Gares, D.M.,1965. Radiant energy, its receipt a i d disposal. Meteor. Monogr.,6,No.28, 1-26. 3. Hull,J.N.,1951. Spectral distribuiion of radia4ion from sun and sky, Trans.lllun.Eng. Scc.(London),l9:21-28. 4 . Kerr, J.P.,G.W Thurtell, and C.B. Tanner,1967. An integraiing pyranometer for ciima~ological observer stations and mescoscale networks, Journal of Applied MeteoroIogy,6,688-06Sd.
.r

*.

096-1 SR Rev. 7/93

FIGURE 2-1;

. . . . . . . .1:.; . .: .: . : .I / ::::::::: ::

........

: i : ; ; ; ; : j ; ; ; _....; j / ;;j
.....

I . . . . . . . . .

...... a . . . . . . . . . . . ::::::::: . . ....... ...... ................

~
L ,

:::::::::i::::::::;iii:::::::~::::::;;Y,::::::::l::::::: ...! . . . . . . . . ................ . . . - . . . ....... .-.,.......... ,....+,.+ .....,...


I . .

,,,.....
8 , : . . I . . .

, * , . . I . , .

.-

............... +., '


l l

:....4 1 . . ........................... .....,. ......... i . . . : ...................................... ................................. +::;I;; i : l . " . ' . .................... ..,.,- ...*...,, ".."'.. ".'..... " ..." ""'..*, ;i:;;:i;; ......................................... .B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

;...;i..

.I....,.i I . . , . . . . .

I!!;::;,:

.................................. .............. ................ *


I . . .

Solar Radiation Sensors


Solar energy is a significant element in large-scale atmospheric motion, and as a result, it has an important place in meteorology. It is directly related to atmospheric stability, and is used in determining stability classes for pollution studies. Met One Instruments supplies solar sensors to meet virtually any monitoring requirement.

Model 095 Pyranometer


The Model 095 Pyranometer incorporates a multi-junction differential thermopile and a precision ground optical glass hemisphere which is transparent to wavelengths 0.285 to 2.80 microns. It is used for high precision, broad band width measurements of incident solar radiation.

The Model 095 Pyranometer

Features
s Differential thermopile

detector
s High accuracy, broad

bandwidth
s Temperature compensated s Rapid response time s Built-in leveling devices

The detector element is of wirebound-plated construction with black and white segments. When exposed to solar radiation, the differing absorptivity of the black and white surfaces develops a temperature differential. The thermopile then produces a voltage proportional to the solar radiation. Built-in thermistor circuitry is incorporated to eliminate the effects of ambient temperature.

The single hemispherical optical glass dome has a waterproof seal, but can easily be removed for repairs. The case is cast aluminum, painted white, and is fitted with a desiccator, circular level, and leveling screws. A mounting base with a vertical 3/4 IPS pipe simplifies field installation of the sensor. This rugged instrument is capable of withstanding mechanical vibrations of up to 20 Gs.

Met One Instruments,

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com metone@metone.com
Solar Radiation - REV. 9/1/97

Model 394 Pyranometer


The precision spectral pyranometer is used to measure the total sun and sky radiation, or a defined wavelength band. It utilizes a circular, multi-junction thermopile protected by a glass hemispheres. The selection of the outer hemisphere determines the spectral range of the sensor.

Features
s Extremely high accuracy s Circular multi-junction

thermopile detector
s Full or partial band

sensitivity
s Temperature compensated s Rapid response time s Built-in leveling devices

The Model 394 Pyranometer

The detector element is a circular wirebound multi-junction thermopile. The junctions are copper-constantan plated. Built-in thermistor circuitry is incorporated to eliminate the effects of ambient temperature. The hemispherical optical glass domes have a waterproof seal, but can easily be removed for repairs. The case is cast bronze, enameled white, and includes a guard disk. The sensor is fitted with a desiccator, circular level, and leveling screws. A mounting base with a vertical 3/4 IPS pipe simplifies field installation of the sensor.

Model 096 Pyranometer


The Model 096 Pyranometer uses a silicon solar cell mounted in a cosine corrected miniature head. It provides an economical solution to the measurement of solar radiation for use in meteorological, agricultural and hydrological studies.

Features
s Silicon cell sensing

element
s Extremely rapid response s Fully cosine-corrected s Error less than 5% s Built-in leveling devices

The Model 096 Pyranometer

The spectral response of the sensor is .4 to 1.1 microns (400 to 1100 nanometers). This does not cover the full range of the solar spectrum, but the error introduced is less than 5% under most conditions of natural daylight. The cable is directly attached to the sensor without connectors. The sensor is furnished with a mounting plate, which contains a circular level and adjusting screws. The supplied base includes a vertical 3/4 IPS pipe to facilitate mounting and field installation.

Model 097 Net Radiometer


The Model 097 Net Radiometer is a high output device designed to measure the sum of all incoming radiation less the outgoing radiation. This is the energy retained by the surface for heating soil and air, plant growth, and water evaporation. Incoming radiation consists of direct beam and diffuse solar radiation and longwave sky radiation. Outgoing radiation is the reflected radiation and terrestrial longwave radiation.

The Model 097 Pyranometer

Features
s 62-junction thermopile

detector
s High output/low resistance s Leveling ball joint with built-

in level
s Balanced long- and short-

wave calibration coefficients s Desiccant and purge points in support arm s No power required s Linear calibration s Non-pressurized, easy replacement of windshields

The spectral response of the Model 097 is 250 to 60,000 microns (.25 to 60 nanometers). The sensor is extremely sensitive to levelness, and a built-in circular level is provided to facilitate proper installation. The mounting arm of the sensor contains the desiccant supply. Field maintenance is simplified by the design which incorporates heavyduty polyethylene hemispherical windshields. The windshields are easily replaced and require no pressurization. The cable is directly attached to the sensor without connectors. The sensor is furnished with a supporting arm and mounting plate, enabling the sensor to install on any horizontal or vertical pipe having a diameter less than 1.5 (38 mm).

Specifications
Model 095 Spectral Response: Calibration: Sensitivity:** Impedance: Temperature Depend.: Linearity: Response Time: Cosine Response: Dimensions: Weight: Mounting: Cable: Spectral Response: Calibration: Sensitivity:** Impedance: Temperature Depend.: Linearity: Response Time: Cosine Response: Dimensions: Weight: Mounting: Cable: Spectral Response: Calibration: Sensitivity:** Impedance: Temperature Depend.: Linearity: Response Time: Cosine Response: Dimensions: Weight: Mounting: Cable: Spectral Response: Calibration: Sensitivity:** Impedance: Response Time: Dimensions: Weight: Mounting: Cable: Nanometers: 285 - 2800 Microns:0.28 - 2.8 Integrating hemisphere approx. 1 cal cm-2 min-1, at 25C 11 mV/kwatt meter -2, approx. 350 ohms, approx. 1.5% constancy from -20C to +40C 1% from 0 to 1400 watts meter-2 5 second (1/e signal) 2% from normalization 0-70 zenith angle; 5% 70-80 zenith angle 5.75 in (14.6 cm) diam. x 21 in (53.4 cm) h, including mount 4 lbs (1.8 kg), including mount Leveling plate and mounting base included. Requires PN 1552 fitting or similar device. PN 1138-xx (xx = length in feet) Nanometers: 285 - 2800 (clear)* Microns: 0.28 - 2.8 (clear) Integrating hemisphere approx. 1 cal cm-2 min-1, at 25C 9 mV/kwatt meter -2, approx. 650 ohms, approx. 1% constancy from -20C to +40C 5% from 0 to 2800 watts meter-2 1 second (1/e signal) 1% from normalization 0-70 zenith angle; 3% 70-80 zenith angle 5.75 in (14.6 cm) diam. x 23 in (58.4 cm) h, including mount 9 lbs (4 kg), including mount Leveling plate and mounting base included. Requires PN 1552 fitting or similar device. PN 1138-xx (xx = length in feet) Nanometers: 400 - 1100 Microns: 0.4 - 1.1 Against Eppley B&W under natural daylight 8 mV/kwatt meter -2 with 100 ohm load, approx. 100 ohms load (dependent upon sensor sensitivity) .15%/degree C, max. 1% from 0 to 3000 watts meter-2 10 microseconds (10% - 90%) Corrected up to 82 incident angle. Azimuth error less than 1% over 360 at 45 elevation 3 in (7.6 cm) diam. x 19 in (48.2 cm) h, including mount 1.2 lbs (.54 kg), including mount Leveling plate and mounting base included. Requires PN 1552 fitting or similar device. PN 1832-xx (xx = length in feet) Nanometers: 250 - 60000 Microns: 0.25 - 60 Against transfer standard, compared to a tungsten-halide light source 75 mV/kwatt meter -2, approx. 4 ohms 10.5 seconds 2.8 in (7.1 cm) w x 2.5 in (6.35 cm) h x 37 in (94 cm) 3 lbs (1.3 kg), including mount Mounting plate for support arm included PN 2437-xx (xx = length in feet)

Model 394

Model 096

Model 097

* Contact factory for other ranges. ** Sensitivity varies among sensors of the same type. A calibration certificate is supplied with each sensor.

Service Report
096-1 SOLAR RADIATION SENSOR
Met One Instruments, Inc. 1600 NW Washington Blvd. Grants Pass, OR 97526
Ph. 541-471-7111 Fax 541-471-7116

E-mail: service@metone.com

Parts & Service Description


REPAIR AND FIELD MAINTENANCE
A complete repair and service department is maintained at our factory in Grants Pass, Oregon, and at our regional service facility in Dallas, Texas. This sensor must be returned to the factory for service and re-certification. Turnaround time for this sensor is generally six to eight weeks.

RETURN AUTHORIZATION
Prior to any product return, please contact the factory at (541)471-7111 or e-mail service@metone.com for a return authorization number (RA).

BASIC SERVICE(S)
Ordering No. SVC0059 Description Calibration of 096-1 List Price Consult Factory

REVISION DATE: March 2002

Carefully unpack all components and compare to the packing list. Notify NovaLynx Corporation immediately concerning any discrepancy. Inspect equipment to detect any damage that may have occurred during shipment. In the event of damage, any claim for loss must be filed immediately with the carrier by the consignee. Damages to equipment sent via Parcel Post or UPS require the consignee to contact NovaLynx Corporation for instructions.

If equipment is to be returned to the factory for any reason, call NovaLynx between 8:00 a.m. and 4:00 p.m. Pacific Time to request a Return Authorization Number (RA#). Include with the returned equipment a description of the problem and the name, address, and daytime phone number of the sender. Carefully pack the equipment to prevent damage or additional damage during the return shipment. Call NovaLynx for packing instructions in the case of delicate or sensitive items. If packing facilities are not available take the equipment to the nearest Post Office, UPS, or other freight service and obtain assistance with the packaging. Please write the RA# on the outside of the box.

NovaLynx Corporation warrants that its products are free from defects in material and workmanship under normal use and service for a period of one year from the date of shipment from the factory. NovaLynx Corporation's obligations under this warranty are limited to, at NovaLynx's option: (i) replacing; or (ii) repairing; any product determined to be defective. In no case shall NovaLynx Corporation's liability exceed product's original purchase price. This warranty does not apply to any equipment that has been repaired or altered, except by NovaLynx Corporation, or that has been subjected to misuse, negligence, or accident. It is expressly agreed that this warranty will be in lieu of all warranties of fitness and in lieu of the warranty of merchantability.

NovaLynx Corporation 4055 Grass Valley Highway, Suite 102 Auburn, CA 95602 Phone: (530) 823-7185 Fax: (530) 823-8997 E-mail: nova@novaIynx.com Web Site: www.novaly nx.corn

Copyright O 1998 by NovaLynx Corporation

EQUIPMENT CONFIGURATION AND IDENTIFICATION

. . . . . . . . . . . . . . . . . . . . . . . iii

1.0

INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 GaugeDesign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.3 Use of Gauge with Evaporation Pan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.4 Evaporation Gauge Tester . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.5 Automatic Evaporation Pan Refill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 iNSTALLATlON and Quick Set Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 3.2 Site Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3.3 Leveling the Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3.4 Testing for Leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3.5 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 OPERATiON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.1 General Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.2 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.3 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.4 Float . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.5 Natural Influences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.6 Evaporation Pan Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Drawings And Data Logger Programming
!

2.0 3.0

4.0 5.0 6.0

7.0

iii

Evaporation Gauge Tester ........... Optional 1.4 To aid the user in the set-up, testing, and reading measurements of the evaporation gauge, NovaLynx has developed a hand-held tester, Model 255-1 10. The tester connects to the evaporation gauge output signal terminal block and converts the sensor's resistance into a Direct Reading of the evaporation pan water level in inches. The tester features an easy-to-read LCD and a mode selection switch. The tester is self-contained and battery-powered, Refer to the operator's manual for the tester for additional information.

1.5

Auto

........ Optional

Evaporation gauges are often used with data loggers to collect and store the data at the gauge location until the data is to be transmitted or transferred to the operator's main computer. Novalynx has created an automatic evaporation pan refill device, Model 255-620. If a water supply line or water tank is available at the site, 255-620 will turn on the water, until the pan is refilled to a level of 8 inches. This can be set for daily, weekly, or every few days. The refill occurs at the time that is Set. NovaLynx recommends using a time such as midnight for the refill time, Refer to the automatic refill manual for more information.

0' of 3-Conductor, 24 AWG, Shielded r: 3-pin terminal block, standard (opposite end ~ l cable terminated to meet f oring equipment interface requirements) g), Stainless Steel ater input port: 112" NPT coupling, female ase dimensions: 16" triangle with leveling screws Total resolution: 0.03" (0.76 mm)

(20 340, +lo " Gap ) esistance: I,000 ohms, 21 0% rature: -40 to +140 F(-40 to +60GC)
echanical range: 0 to 10" (0 - 254 mm) Electrical range: 0 to 9.44 inches (0- 240 mm)

mation for
V
T

Evaporation Gauge should be and unpacked, leveled and cb,tnected to the Evaporation Pan. A Digital VoltKIhm Meter D.V.M. should be Used, but is Not mandatory.

esistance Between The White & Black Wires If You Do Not Have a D.V.M. See Step 6
otate the Sprocket to Read between 40 to 80 ohms.

3. Attach the Float I Chain Asse onto the Sprocket See Dwg. ( Float hanging on Right Side as viewed from the Front )

.'-, --

Sprocket

esting on the
ou%d read bed:
5. if the reading Not between 40 to 80 ohms Then, Lift the chain off the sprocket & rotate it as needed one or two gear tooth positions. Replace the chain and repeat step 3 8 4 ( Chain
7

Float
I
I
I

6----

Back of thk sprbcket

4
float Resting on the Bottom

6. SET- UP If You DO NOT Have a D.V.M.. A White Line is painted on the Back Side of the Sprocket, When Viewed from the Back Side, Rotate the sprocket, so that the White Line is' pointing at 1 O'clock or 30" ( see dwg. ) 7. Attach the Float 1 Chain Assembly onto the Sprocket as Shown. With the Float hanging on Left Side as viewed from the back The float should be Resting on the Bottom .

** Attach the 3 conductor cable to your Data Logger or measuring Device. Refer to the Manual for a more detailed Set-lip Procedure.

Carefully unpack ail of the evaporation gauge. Remove the top cover. The fifty foot cable, and float with chain and counter-weight: are shipped inside the bottom of the gauge housing. Remove the cable assembly from the housing. The float cannot be removed, without removing the PotiGear Assembly housing Use caution when removing packing materials and parts from inside the gauge. Take care to avoid hitting the potentiometer sprocket wheel.

The Novalynx Evaporation Gauge water reservoir is physically connected to the pan by using a 112" diameter pipe. Threaded couplings are provided on both the pan and the gauge. Typically, 112" rigid water pipe is used to connect the gauge to the pan. Flexible tubing may be used, provided it does not deteriorate in outdoor weather conditions. The gauge should be placed far enough away from the pan and on the North side to avoid casting any shadows or reflections inside or onto the sides of the pan. Shadows will affect the evaporation process. Both the evaporation pan and the gauge need to be as level as possible in order to maximize the amount of water that can be poured into and measured in the pan. A level pan will provide uniform exposure of the water to the atmosphere, eliminating uneven depths of water. The site should be /eve1and free of nearby obstructions that can cast shadows or reflect sunlight onto the evaporation pan. The pan should be placed upon a wooden platform over soil typical of the area. Level the platform before installing the pan. Place the pan so that the water pipe fitting faces the evaporation gauge. A second platform for mounting the gauge may be needed in order to place the bottom of the gauge at the same elevation as the pan.

3.3

Levelin
\

Level the gauge by adjusting the three leveling screws located on the triangular base. With the top cover removed, place a carpenter's levei across the opening and check the level. Adjust the screws until the air bubble in the level is centered. After leveling the gauge look down inside the stillwell to make certain that the float and chain hang down straight and are centered in the gauge. The float must be free to move up and down without making contact with the sides of the gauge.

3.4

Testing for Leaks

After connecting the gauge to the pan and all electrical connections are in place, fill the evaporation pan with the desired amount of water and carefully check all of the joints for leaks. Using Teflon tape or plumbers pipe joint compound at each threaded coupling will help prevent leaks.

3.5

iring

Whenever the gauge is to be used with other equipment, the output signal wiring may vary depending upon the way that the gauge is to be used. The signal and power wires are connected into the gauge through a three-pin terminal block. The terminal block is mounted onto the side of the potentiometer housing. To access the terminal block, the cover must be removed from the gauge. Four screws hold the cover in place. The signal cable is terminated with three spade lugs. The spade lugs are attached to the screws of the terminal block. Refer to the gauge schematic to obtain wiring details. The opposite end of the cable will generally not have a connector attached to it unless one has been ordered for a specific purpose, The ends of the wires will be stripped and tinned. Refer to the table presented below for wiring details.

Power Input

5 vdc (typical)
0-5 vdc (wiper )

RED WHITE BLACK

TB PIN 1 - T O P TB PIN 2

Evap Signal Power Ground

GND

- MIDDLE TB PIN 3 - BOTTOM

The evaporation gauge potentiometer exhibits a varying resistance in response to the motion of the float. Applying a voitage across the potentiometer allows monitoring of the gauge by a data logger or any other electronic voltage sensing instrument. The output signal of the potentiometer can be configured to give an increasing or a decreasing voltage with regards to the changing level of the water. For most systems provided by NovaLynx, the signal is configured so that a decrease in the water level is represented by a decrease in the gauge voltage. The Gauge Will Never Read below 1.4" This is when the float physically hits the Bottom of the Gauge, and there is still 1.4" of water present. Normal Range is Typically is 3" to 8.5 or 9" , above 9" wind could blow water over the edges, below 3" the Low level of water absorbs more heat from the sun & Bottom Pan Reflections,causing Increased Evaporation

Note: that during winter months when the water may freeze, NovaLynx recommends draining the pan and storing the pan and the evaporation gauge indoors.

Proper calibration of the evaporation gauge is critical to the accuracy of the data as well as to the correct operation of the gauge. Calibration of the gauge is required upon initial installation of the gauge and whenever the gauge is removed from its platform or from inside the evaporation pan. The evaporation gauge must be calibrated in order to set the operating range of the potentiometer and to determine the zero point of the float motion. For data logging systems, the slope and intercept data will be measured or calculated from the calibration of the gauge. After the gauge has been calibrated, the operator needs only to keep track of the amount of water added back into the evaporation pan. With the gauge calibrated, the gauge may be used to measure the amount of water added into the pan each week. The following procedure is used to properly calibrate the evaporation gauge. Record the calibration values whenever possible in order to check the gauge calibration during the year. Upon completion of the gauge and pan installation, fill the evaporation pan with approximately eight inches of water. The water height can be checked by using a ruler or a tape measure. Check the evaporation gauge to ensure that the water has filled the stillwell to the correct height inside the gauge. Disconnect the signal cable from the evaporation gauge. Use a digital ohmmeter to and measure the resistance of the pot. across terminal block pins (2) middle (+) (3) bottom (-). Rotate the sprocket and you will notice the resistance value changes from 0 to 1000 ohms. Rotate the sprocket slowly until the reading jumps to a floating condition or reads 0. This is the dead band it has a mechanical range of about 20". Some pots may be exactly I K ohm, but most will vary by 5 15 ohms There is a white kine marked on the back side of the sprocket. When this line is Vertical the pot shouie be in the 20" Gap Area. Check to make certain that the float is hanging down near the center of the evaporation gauge. As viewe face sf the sprocket wheel, sprocket wheel with the counter-weight to the left side of the wheel. A White Line is painted on the Back of the Sprocket, When Viewed from the Back side the Line should be at the 1 O'Clock or 30" when the Float is resting on the bottom of the gauge. Move the float by hand down toward the bottom of the gauge. As soon as the float contacts the bottom surface of the gauge, hold the float in position and check to see that the chain is straight and tight. The chain must not be moving nor pulled downward other than by the tension exerted by the float. Measure the resistance of the potentiometer.

The potentiometer should be at its point of low resistance for the range of motion of the pot. The resistance should be above zero ohms but may be as large as 150 ohms. If the value appears to be floating or within the deadband of the potentiometer the sprocket wheel must be adjusted with relation to the chain. The potentiometer must not cross into or go beyond the deadband as the float approaches and touches the bottom of the gauge. To adjust the sprocket, allow the float to return to the top of the water. Lift the chain off the sprocket and rotate the sprocket counter-clock-wise, one or two gear tooth positions. Replace the chain and repeat step 4.4. Repeat this step (4.5) until the chain is properly positioned on the sprocket. Observe the ohmmeter to detect any crossover of the deadband as the float is moved downward. Record the resistance of this Bottom or ZERO position of the float. If the gauge is connected into the monitoring equipment and power has been applied, measure the voltage at this position as well as the resistance, Measure the position of the center of the float with respect to the bottom of the evaporation pan for pan mounted gauges. For gauges mounted outside the pan, measure the water level inside the gauge to the center of the float. Usually, the center of the float is aligned with the center of the threaded pipe fitting on the housing when the float touches the bottom of the gauge. Move the float back to the top of the water and let go of the float. The center of the float should now be even with the surface of the water. Measure the resistance and voltage of the potentiometer at this position. Also measure the water level at this paint with respec: to the L30t:m of the pan. The resistance of the potentiometer should be high at this point and approaching the 1000 ohm end of the pot. If the chain and sprocket have been properly set, the potentiometer should not move past the 1000 ohm end of the element and stop in the deadband. If it appears that the pot is in its deadband at the top of the water level, then either the sprocket and chain must be adjusted further or else there is too much water in the pan and some water must be drained ..... Typically this Setting should be offset sligthly to avoid going into the Gap Area. When the Pan is Full to the Brim, the Resistance should be between 950 and 995 ohms When the Pan is Empty or < 1%'' Deep .... The resistance should be between 20 to 90 ohms. At this point the slope and intercept information can be generated and the operating curve of the instrument can be calculated. The operating curve can then be verified by setting the float to several points of water level and by comparing the actual output voltage to the calculated voltage. Remember the Gap is 20" therfore 340"/360 = 0.9444..... With a 5.00 volt excitation applied, A 5 Volt reading on the The formulas used are: Wiper means 9.44" Not 10.0"

m = (Y2 - Y,)

(X2 - X,)

where Y is inches of water, X is ohms or volts, rn is the slope of the line and b is the zero offset. For a range of 10 inches and a potentiometer resistance of 1000 ohms the calculation reveals that b = 0, and rn = 0.00944 For a range of 10 inches and a voltage range of 5 Vdc, b = 0 and
n = ;

1.88.

If the float range of motion happens to be 1.5 to 8.0 inches, then the output voltage can be calculated as 0.75 and 3.776 volts, respectively. Use the actual values measured in the preceding steps, 4.2, 4.4, 4.6, and 4.7, with the above formulas to determine the operating line of the gauge. The line should be linear allowing calculation of intermediate outputs for known positions of the float.

Always disconnect the reportinglrecording device from its power source before making any changes to the wiring connections. If possible, troubleshoot the gauge immediately whenever any of the following conditions are observed: severe weather has recently occurred; the gauge does not appear to operate normally or exhibits a marked change in performance; the data is missing or appears to be incorrect; the gauge has been dropped or damaged; water has damaged the wiring or electronic components; water has been allowed to dry out totally in the evaporation pan.Remembera 90" Rotation on the sprocket should indicate a change ot 2.50 inches on your reportinglrecording device, 180" would be a 5.00 inch Change.

If the gauge does not register correctly, first check the power connections. Check the voltage with a voltmeter. Be sure the reportinglrecording device has been powered up correctly. If the reportinglrecording device uses batteries for it's primary source of power, check the batteries to be sure they have sufficient voltage and that they are securely in place. Check the battery terminals to ensure that they are clean and provide solid contact.
6.

Cables

Check the sensor cable connections both at the gauge and at the control unit; cable shorts or opens can cause loss of data. If a connection is found to be loose, reattach the wire and check to see if the problem has been corrected. Check for damage to the cable insulation. Replace the cable if it appears that the jacket has been worn or cut open.

Inspect the float and chain to make certain that they have not become entangled. Whenever the evaporation pan has been allowed to empty or dry out completely, the float may tip sideways at the bottom of the gauge and the chain may become kinked at the mechanical connector. Test the motion of the float to ensure that it moves smoothly and freely.

A number of naturally occurring events can influence the evaporation gauge reading and will appear as errors or as a gauge malfunction. These natural events include animals the water, birds bathing in the water, leaves and debris fallin into the pan, ed rainfall or sno during the night, thermal expansion of the high level winds, lanre water and metal parts.

Additional errors in the evaporation gauge's measurement may be related to maintenance of the evaporation pan. The pan must be kept free of algae. Plants must not be allowed to grow up and over the edges of the pan. Dirt and dust must not be allowed to accumulate inside the pan. The growth of algae in the evaporation pan can be discouraged by the addition of small amounts of copper sulphate lo the water, however, algae already present must be removed by a thorough cleaning of the pan. Most of these problems are alleviated by regular rinsincj sf the pan. In severe cases, spa chemicals may be used to combat algae, however, the addition of chemicals to the water will influence the evaporation ing months when freezing conditions are likely, ernp , clean, and store pan should be stored indoors. If it must be left in the nced enclosure, it should be turned bottom side up and secured to the platform with stout rope.

** SAMPLE INPUT INSTRUCTIONS **


( For Campbell Data Loggers CR10 & 1O ) X

For Evaporation Gauge 255-1 00 ( scaled in inches, range 0.00 to 9.44" ) -

With 2500 mv Excitation NOTE: When the Float is touching the bottom of the Evap. Gauge, the chain linkage should be set for a output signal that is slightly positive. ie. 50 to 250 mv The Evaporation is Normally Calculated as Daily Totals. Example: Start 7.65" after 24 hrs. 6.87" Daily Evaporation = 7.65-6.87 = 0.78 inches per day The Exact Physical Level in the Pan Is Not Needed for Evaporation Measurements. The Difference or Change from on time to another is all that is Necessary

Excite-Delay (SE) (P4) 1: I Reps 2: 5 i 2500 mV Slow Range : SE Channel 3: 1 4: 01 Excite all reps w1Exchan 1 5: 2 Delay (units 0.01 sec) 6: 2500 mV Excitation 7: 2 Loc [ Evap-inch ] 8:.00377 Mult 9:0.0 Offset WIRING CONNECTIONS: Red ........Connect to E l or E2 or E3 Black ....... Connect to AG ( Analog Ground ) White ...... Connect to SE Channel (the one that has been programed ) Shield ..... Bare Wire ...Connect to Ground ( if necessary, not required for general operation )

SAMPLE SENSOR SET- UP for the WS-I6 Modular Weather Station (c) NovaLynx Corporation Inc. 2002

Sensor Types

1. Generic 2. Wind Speed 3. Wind Direction 4. Temperature 5. Humidity 6. Pressure 7. Precipitation 8. Solar Radiation 9. Evaporation 10. Soil Moisture 11. Water Level
Select type [ I

- 1I ] or quit [Q]: 9

Select model [I - 11 or quit [Q]: 1 Sensor Input Channel Pulse Counter PI Pulse Counter P2 Pulse Counter P3 Analog lnput A0 Analog lnput A1 Analog Input A2 Analog lnput A3 Analog lnput A4 Analog lnput A5 10. Analog lnput A6 11. Analog lnput A7 12. Analog lnput A8 13. Analog lnput A9 1. 2. 3. 4. 5. 6. 7. 8. 9.

( fypical input channel)

Select channel [ - 131 or quit [Q]: 4 I


*******************************************x****************************************-**********x*******************~***~*

GENERIC Data Logger I RTU 1 PLC ISignal Conditioner ( Assuming 5.00 Vdc Excitation Voltage )
The Evaporation Gauge can be set up One of 2 ways 1. Direct Reading of Water level in Pan ..... Example : 1.44" to 9.44" = 2. Direct Reading , with a positive Offset of Water level in Pan Example : 2.50' to 10.00" No 2 is the preferred Method, Because the Gauge will not go into the Gap Area Evap. Measurements are Measured as a Change over time Example : . I I 6 inches per Day . The Exact Water Level in the Pan is Not Important, Only the difference in Change
V w = Wiper Voltage V+ = Excitation Voltage L = level of water inches

L=

Vw V+

x 0.944

COVER ASSEMBLY PN 10000067

# I 0 NYLON SPACER PN 7 2 3 1 0 8 i 0 ( X r ) COVER SCPEW 10-32 x 1 / 2


ME? weSiER i C c 3

PN 721BE033

(XC)

\
MOUNTING SCREW / 10-32 X 3/8 FAN HEAD PN 72121032 ( X 2 ) $10 SPLIT LOCK WASHER PN 72340210 (X2)

.
/

/
/
/

POT & GEAR HOUSING ASSEMBLY PN 10030330

STRAIN SELIEF PN 41500305 CABLE ASSEMBLY PN 103000247

--- - 3N

GAUGE

N_D. BASE

Cr 0
G

I-Q

0DL

-t

w>

-wpcn = o z cn-, @ - a ui -

.-Lz
0

ogw-

40 0 0 0 ~ 0 s7- 0 ' 0 .
A

0 x 2

04 ni---Jo i w LLIOOZ

cl

Evaporation 255-100 Analog Output Evaporation Gauge


The Model 255-100 Analog Output Evaporation Gauge is used to determine the evaporation rate by measuring the changing water level in an evaporation pan. A standard National Weather Service Class A Evaporation Pan is recommended. The instrument consists of a float, pulley, and counterweight attached to a precision 1000-ohm potentiometer mounted through a gear assembly in a weatherproof housing. The triangular base plate is equipped with three leveling screws and a bubble level. The potentiometer produces a resistance output proportional to the position of the float which can be monitored on site using a data logger or a strip chart recorder, or monitored remotely by telemetry equipment. The gauge can be placed directly in the pan, or connected to the pan by using the 255-100P/F stainless steel pipe and fittings.

Specifications
Potentiometer: Accuracy: 0.25% Rotation: Continuous Resistance: 1000 ohms Operating temperatures: -40 C to 60 C Linearity: 0.25% Range: 0-10" System accuracy: Gauge with pan & pipes 0.25% over 10" range Float: 4" diameter Height: 27-1/2" Diameter: 8" Base: 16" triangle with leveling screws Water input port: 1/2" NP coupling, female Cable: 3 conductor, 50 included Weight/Shipping: 7-1/2 lbs/15 lbs (Box 15" x 15" x 30")

255-100 Analog Output Evaporation Gauge

Ordering Information
255-100 255-100P/F 255-100POT Analog Output Evaporation Gauge, 50' cable Stainless Steel Connecting Pipe and Fittings for attaching gauge to pan, 2' pipe + 4' pipe Replacement Potentiometer

255-110 Evaporation Gauge Tester


The Model 255-110 Evaporation Gauge Tester works with the 255-100 Evaporation Gauge to provide a direct reading in inches, corresponding to the output from the evaporation gauge. The tester is used to set-up the evaporation gauge during initial installation and can be used to provide a direct reading of daily evaporation loss.
255-110 Evaporation Gauge Tester

Specifications
Range: 0 to 10.00 inches Resolution: 0.01 inches Temperature range: 0 to 50C (32 to 122F) Humidity range: Non-condensing Accuracy: Better than 0.25% of reading Display: 3-1/2 digit, 7 segment 0.5" high LCD Batteries: 9 V display, 1.5 V tester Size: 5" H x 3" W x 2" D Weight/shipping: 12 oz/2 lbs

Ordering Information
255-110 255-100MS Evaporation Gauge Tester 3-Pin MS Connector Option

NovaLynx Corporation PO Box 240 Grass Valley CA 95945 Phone: (530) 823-7185 Fax: (530) 823-8997 USA Toll Free: 1-800-321-3577

www.novalynx.com 139

1600 Washington Blvd.

Grants Pass,Oregon 97526


Telephone 541 47%-7111

Instruments

Model 455A 1456A Data Logger Smart Keyboard Display

1.0

GENERAL 1NFORMATION

The Met One Instruments dats logger is configured with the 'Smart Keyboa-d Disp!i?yl'on the front panel of the Model 455A. It allows the local operator to examine various locations in the logger and view currerA values of the output channels.
,

With the Sma-t Keyboard Dispiay (SKD), in fieid programming of ihe data iooger can be performed. It slso sllows for the operatsr to examine the contents of the data logger encjne CR-lox.

2.0

FUNCTION

The Model 4554 is facisry programmed to record dsia from the various sensors. providing information in the format and repoiring values required by the customer. An internal battery will maintain the intelligence of the system during brief periods of power failure. The Smart Key values are stored in non-volatile memory, and will not be lost. Instantaneous measurements, stored values, time, date, and the internai program signature is easily displayed by ths operator. 3.0
INSTALLATION

lflstaiiation is sasy. since the SKD functton has already been installed at the factory. There are a few items, which will effect the operation of the system, and should be noted.
The fecboy set password ts enter the

KD functions i - 1 2 3 4 6 6 7' 8 D s

B.

CAUTION: When the SKI3 d~spfay x t i v e , communications to the central data is system is shut off If communications to the central is lost, check to be sure that the display on the MicroMst 455A is not operatio~al.

4.0

KEY FUNCTION DESCRIPTIONS

A.

DISPLAY POWER

This Key controis the dispiay. The data logger functions continue, whether or not the dispiay is operating. Use this key when you want to view inputs, recorded dsta, or charrge programming. Turning off the power to the display also puts the SK3 into "sleepJ'mads, conserving power. When the SKD is first powered up, it will dispby "SELECT FUNCTIGN". This is your indication that the system is ready for you t@ d further. a

When pressed, this key will display the currently setdata logger time. Press again t display 6 the current date. The key wifj continue to toggie between time and dale. No modifications to the time or date may be made using this k e y 4 is "read onlyn. It may be depressed a4 any time to check time andlor date.

G.

OTHER CONTROLS 1. Keyboard Reset

Should the SMD "Lock-Up" for any reason; pressing the Keyboard Reset button, located behind the front panel, may reset it. The button will be found on the upper [eft side of the interface board.
OPE: This will reset the Keyboard only. Should the data logger "iock", if. will be necessery to disconnect both primary and backup power to the Ioggei: This would eniajl unplugging the power supply and d:'sconnecting the backup battery. This will render the logger ''sfupid': and it will have io be reprogrammed.
2.

Display Contrast

i o enhance viewing of the LCD display, the contrsst may be adjusted. The Contras: Control is located on the interface board just above the sensor connectbn terminals. A srnzdl ( 0 or $ #1) Fi-iiliips screwdriver is requirpd 10 change the contrast adjustment.

E.

CHANGING THE PASSWORD Put thi: SKD in the Smart Mode, scroll to CHANGE PASSWORD, and press the key. Kev(s) ?or&
D
Dis~lav Shows

"iY'

CHANGE PASSWOaD

ENTER PASSWORD
VEF?!fY PASSWORD

Enter new password, then "D" Re-enter new password, then


1

"D"

MODIFY? ? = Y 2 = N
SELECT FUNCTION

The password is now changed and you can select another function.

F.

EDITING INPUT LOCATIONS The input labels a-e used in the SCAN iNPiiTS function. First, put the SKD in the Smart Mode. scroll to INPUT LOCATIONS, and press D (as described pieviously).

Dzia logger
Ii~put Location O'l thru 28 Dispky Status +- = Displayed Elank=Not Displayed

Measursment

Units Wind Speed Wind Direciion Air Temp Delta Temp Re1 Humidity Rainfall Earo Press Solar Rad 1 min. avg, WS

WS WB AT DT

=
= = = =

MPH, KPfi7m/s DEG DEG DEG


% rn, mrn

RH

RN = BP = SF? = WSCAL=

"Hg, kPa, mb Wm2, kWm, Urn (same as WS)

The UP and DOWN arrow keys will scroll the display through all 28 lnput Locations. Press the "C" key to edit (Change) the location label being displayed. The "A" key will cause the "+" (Display Status) symbol to appear or disappear, resulting in that lnput Location being displayed (or not) when using SCAN INPUTS. Use the " -+ " and " c " (side arrows) keys to move the cursor through the display.

The Measurement name (up to 5 characters) and Units name (up to 3 chaiadeisj can be changed by using the " ? " and " $ " (up and down arrow keys). The f ~ l set of i ASCli characters is aiM!able.

When you are finished editing an lnput Location, press the "D (Done) key. YOUcan
then scroll to another input Location.

H.

EDITING THE DATALOGGER PROGRAM WITH THE SKD

Kevs
DISPLAY POWER

Dis~lav Shows

SELECT FUNCTION

ENTER PASSWORD

READY
The data logger is now ready to accept keypad commands for the CRIOX. Consul: tbe programinins sections of this manual for commands and instructions, (The logger-programmmg manual is not always provided ss pari of this manual, unless requested as part of project If you are editing the data logger program, refer to the program prhtout in the program section of this manuai, ~nciuded with your system.

Integrated Data Acquisition Units


455 456 457
The central brain of the modern weather station is the data acquisition unit, or DAU. It is here that the data collected by the sensors are recorded, manipulated, and stored for future use and analysis. The DAUs offered by Met One Instruments, Inc. offer the highest degree of integration and ease of use available today. They are designed to provide the environmental engineer with:
s A simplified design and

procurement procedure
s A system of guaranteed

matched components s The benefits of Met One Instruments field expertise s Assurance of compliance A built-in datalogger, scaled directly to the recorded measurements, provides an easy to read, permanent record of the measured meteorological variables. The system includes the datalogger, the interface board, and panel-mounted Smart Keyboard/ Display. Provision is made for the addition of an optional Storage Module for extended memory capability. An RS-232 dataport is available for direct communications with a computer or terminal. A modem may be added to facilitate remote access of the datalogger for programming or data retrieval. Password protection and automatic shutoff of the display provide additional data security. Standard unit provides 12 analog and 2 pulsed signals, expandable with additional ana-

SKB - The Smart Keyboard/Display contains Smart Keys which offer shortcut communications with the datalogger.

log or digital inputs or outputs. The motherboard includes lightning protection diodes, and provides a charging function for a backup battery (optional with the Model 457). Display Power turns on the keyboard display and prepares the unit for further operator interaction. Time/Date will return the datalogger time. Press again to display datalogger date. Scan Inputs key will display the data going into the system in channel #1. By using the Arrow keys, other data channels can be viewed. Scan AVGS key will return the data most recently written to datalogger memory. The Arrow keys allow viewing all the elements of the array.

Verify Program key will display the program signature, a number unique to the specific program. This feature alerts the operator if a change has been made to the datalogger program. Program key, which is password protected, allows access to the dataloggers native programming functions and/or the SKB functions. Standard equipment includes datalogger, computer interface, lightning protection and memory backup battery (optional with the Model 457), direct sensor hookup, graphic user interface, and enclosure. Differences between model numbers relate to features and packaging. They are functionally equivalent.

Met One Instruments,

Corporate Sales & Service: 1600 Washington Blvd., Grants Pass, OR 97526, Phone (541) 471-7111, Fax (541) 471-7116 Distribution & Service: 3206 Main Street, Suite 106, Rowlett, TX 75088, Phone (972) 412-4747, Fax (972) 412-4716 http://www.metone.com
455/456/457 - REV. 9/1/97

Model 455
This full-featured device provides virtually everything needed for a complete and total weather data system. All hardware is mounted within a high quality compression molded fiberglass enclosure. The enclosure is designed specifically for harsh manufacturing conditions and hostile environments, and will outlast steel or stainless steel in a corrosive or marine environment. The enclosure meets and exceeds most NEMA grade specifications. Door hinges are molded, with stainless steel hinge pins. Cable entry is through a port in the enclosure bottom, fitted with a sealing clamp. The enclosure contains a latch with integral tumbler lock for security.

Model 457
The Model 457 is competitively priced to allow any sized operation to take advantage of the functionality and ease of use of the Met One Instruments system. All hardware is mounted within an economical, attractive grey fiberglass enclosure. The enclosure features stainless steel hinge pins and latches with padlock hasps, and meets NEMA type 4 classification. Cable entry is through a port in the enclosure bottom, fitted with a sealing clamp. The backup battery and lightning protection are optional with this model.

Model 456
This unit offers all the features of the Model 455, but is designed for mounting in a standard 19 equipment rack. The panel height is 7, and the minimum depth required is 15. Terminal blocks are provided for direct sensor cable connections. A metal chassis cover or benchtop enclosure is available to allow this model to function as a benchtop unit.

Options
To optimize the unit to your specific application and requirements, a broad range of optional configurations and accessory components are available. A complete range of meteorological sensors may be added to the system, including, but not limited to, wind speed and direction, temperature, relative humidity, solar radiation, and precipitation. Instrument carrying cases are available for all sensors. Expanded memory storage modules are available for applications requiring longterm unattended operation. (The standard internal memory will allow recording hourly averages of 15 inputs for 130 days.) Communication options include conventional or shorthaul modem, interfacing cables, and data manipulation software. A backlit display will allow easy viewing in poorly lighted environments. Due to the power required, this option is only available on AC powered systems. A multi-channel graphic recorder may be interfaced to the DAU, allowing for hard-copy backup of important data. Please consult your Met One Instruments sales specialist for additional information on these options and accessories. These units are designed for easy installation and maintenance by non-technical personnel. As with all equipment, technicians and engineers are only a phone call away.

Specifications
Enclosure Construction Color/Finish Chassis Enclosure Security Supplied Hardware

Model 455

Model 456

Model 457

Compn molded fiberglass Beige, textured N/A Locking latch w/key Tower mount kit

19-inch rack panel Beige, textured enamel PN 5787 N/A

Molded fiberglass Grey, smooth N/A Padlockable latch Tower mount kit

Dimensions HxWxD, in (cm) Weight, lbs (kg) Shipping Weight, lbs (kg) Inputs Analog Pulsed Computations Viewable Elements Lightning Protection

16x12x11 (40x30x28) 11 (5) 15 (6.8) 12 (2) 8-bit or (1) 16-bit 28 20 Standard

7x19x13 (18x48x33) 8 (3.6) 12 (5.5) 12 (2) 8-bit or (1) 16-bit 28 20 Standard MicroMet Plus

15x13x7 (39x34x18) 12 (5.5) 16 (7.3) 12 (2) 8-bit or (1) 16-bit 28 20 Assy #2671 MicroMet Plus

Software Sensor Basic Software

MicroMet Plus

Options
Power Standard (Specify) Memory Backup Battery Solar Power System Connections Water-tight Glands Bottom Port with Clamp Terminal Strips Bulkhead Connectors Display LCD, 1 line x 16 char. Backlit (AC only) 115 VAC 230 VAC, 12 VDC STD Consult Factory Optional Standard Standard Optional 115 VAC 230 VAC, 12 VDC STD Consult Factory N/A N/A Standard N/A 115 VAC 230 VAC, 12 VDC PN 2477 Consult Factory N/A Standard Standard N/A Standard N/A

Standard Standard PN 2456 PN 2456 Original order only - cannot retrofit backlit display

Accessories
Expanded Memory 96K Datapoints Communications Conventional Modems Rapid Modems Sensors Accepts all standard meteorological sensors Accepts all standard meteorological sensors Accepts all standard meteorological sensors PN 550088 Yes Yes PN 550088 Yes Yes PN 2605 (Incl. 550088) Yes Yes

S-ar putea să vă placă și