Documente Academic
Documente Profesional
Documente Cultură
THIS DOCUMENT IS A
WORK IN PROGRESS MAY
CHANGE OFTEN. - DON
Table of Contents
Starting and Stopping the ICP-MS...................................................................................... 2
INSTALL TUBING ON PERISTALTIC PUMP........................................................... 4
RESOLVING PROBLEMS.................................................................................................7
PLASMA LIGHTS, BUT DOES NOT STAY ON FOR SOMETIME..........................9
SHUTTING DOWN..................................................................................................... 10
SHUTTING DOWN THE VACUUM PUMP FOR ICP-MS........................................... 11
MAKING A NITRIC ACID DILUTION.......................................................................... 12
MIXING A 10 ppb TUNE A DILUTION......................................................................... 14
CLEANING THE CONES WITH ULTRASOUND.........................................................16
TAKING THE COVER OFF OF THE INSTRUMENT...................................................17
OTHER GENERAL TIPS AND GUIDELINES ..............................................................18
1
STARTING AND STOPPING THE ICP-MS
TURN ON THE ARGON
1. Turn the Argon gas valve on. The pressure on the secondary lower
pressure regulator will need to maintain at ≥85 psi to light the plasma.
Note: To check whether the pressure is correct, turn off the argon to the
regulator if it is connected. Loosen the nut holding the hose to the regulator
to relieve any left over pressure in the regulator; tighten after releasing gas.
Turn the knob to “Decrease” several turns. Turn the gas on again and adjust
to the 80 psi or so pressure by turning the knob in the direction of
“Increase”. If the tank has adequate pressure, it should be ready for use.
Figure 1: Use the side labeled GAS on the liquid argon tank to attach the
regulator. Use a CGA 580 adapter.
2
TURN ON THE CHILLER
2. Turn on the chiller. The chiller is set for a water temperature of 20°C.
If operating properly, the chiller should reach its target temperature in
less than 3 minutes. Otherwise, when trying to ignite the plasma, an
error message will appear.
Again wait until the chiller reaches a temperature of 20°C before trying to
ignite the plasma.
The chiller requires that enough water in the tank for operations. Never put
DI water into the chiller or this will hasten corrosion. Mix the DI water with
a 50/50 mix of tap water and DI Water. Fill the tank to the point indicated.
3
INSTALL TUBING ON PERISTALTIC PUMP
3. Tighten the tubing on the peristaltic pump on the ICP-MS. The tubing
sections around the pump should be replaced after 40 hours of use.
This is a peristaltic pump, so make sure to align the tubing so that the
liquid from the samples will uptake by the pump.
4
• Release pressure on pump when not in use.
• Pump and capillary tubing can be a source of contamination.
• Pump tubing is a consumable-keep a large supply on hand.
NOTE: For most of our purposes, there are two types of tubing. One is
Black-White/Orange-White and another is Black-White/Black-White. The
latter is 1-mm diameter tubing and this should go after the nebulizer since it
has a large diameter. The tubing is PVC material and that is fine for most of
the acid/water solutions. If using the HPLC and organic compounds are
used, check the tubing for compatibility with organic materials.
Make sure that the tubing connections stay together, otherwise any exposure
to air from an unconnected tubing setup will kill the plasma.
Before sampling, pay attention to the nebulizer pressure on the status bar; it
should be between 1.5 mbar to 2.0 mbar. Too little pressure and there is leak
somewhere; too much and the nebulizer may be plugged.
If running no more samples for the day, always make sure to relieve the
tension on the tubing.
5
COLLECT EXCESS SAMPLE
4. Make sure to have container ready for the excess sample and DI water
rinse. The container may be laboratory glassware or any container that
is made of HDPE (high density poly ethylene). Even a small amount
of diluted nitric acid will lower the pH to around 2 or below.
Neutralize the acid and water solution with ordinary before
pouring down the drain.
5. Turn on auto-sampler.
Once these items are in place, start Plasma Lab and press “ON” to ignite the
plasma.
Figure 4: Ice build up on a tank of argon. The hose popped off of the
regulator and the valve on the tank was not closed. The argon vented at low
pressure for several hours and because of the rapid escape of gas, the ice
formed.
6
RESOLVING PROBLEMS
First, check that the gas is at the required 90-lbs or so. Remove the air pipe
from the ICP-MS and check whether there is any flow or suction through the
air pipe. Make sure the chiller is at operating temperature. Check the tubing
connected to the nebulizer and make sure there are no leaks.
7
It is possible under certain circumstances, for the plasma to not light because
the cone hole is too big. The increased hole size will not allow the plasma to
light because the proper vacuum is not obtainable.
Another trick that helps, and this from Thermo tech support, is to disconnect
the multi-pin cable on the backside of the instrument near the hose to the
output vent. Disconnect this cable and re-connect leaving it disconnected for
several seconds; repeat about 3 to 5 times and try starting plasma again.
Pages 130 to 131 from the X Series ICP-MS Getting Started Guide start-
up guide has a flowchart for troubleshooting plasma start issues.
The Load can range from 0 to 255, but should be greater than 90. The Tune
value can range from 0 to 255, but should be AROUND 200.
8
PLASMA LIGHTS, BUT DOES NOT STAY ON FOR SOMETIME
Sometimes the plasma will run for a few minutes and then turn off. First,
make sure that the all the tubing to the nebulizer and from pump is
connected correctly. Any substantial amount of air will interfere with the
plasma staying on for a period of time.
Check to make sure that the silver screen is touching the grounding pin on
the teflon torch holder (inside the cage).
Check the Event Log under the instrument guy icon (the person wearing a
lab coat) to see what errors may have occurred.
Sometimes, after several attempts, the problem will work itself out and the
plasma will stay on for the duration of the available argon.
9
SHUTTING DOWN
Once the samples are run, turn off plasma, loosen tubing from peristaltic
pump, turn off the chiller, turn off the argon gas. Leave the desktop
computer on and the vacuum pump on.
Figure : After using, loosen guides from tubing and loosen the tubing.
10
SHUTTING DOWN THE VACUUM PUMP FOR ICP-MS
There are times when it is necessary to totally shutdown the ICP-MS. The
only acceptable times for this is when maintenance is required on the
vacuum pump like changing the oil or to replace the detector that is located
inside the vacuum chamber.
After the maintenance is completed or the power is back on, return the pump
to a vacuum state.
11
MAKING A NITRIC ACID DILUTION
Safety Note: When pouring a strong acid, wear laboratory nitrile or latex
gloves, goggles, and a lab coat or apron. Pour under a fume hood!
V1 C1 = V2 C2
V1 = 0.0294-L
V1 = 29.4-mL
To mix, pour water in a 1-L volumetric flask until it is about ¾ full. Pour 30-
mL of the nitric acid into a clean beaker (do this under the fume hood and
wear goggles, nitrile or latex gloves and a lab coat or apron).
12
Slowly mix the DI water into the flask and mix contents; a bit of heat will be
released. Mix and swish contents until the line is reached on the volumetric
flask.
This is the base for mixing calibration standards and diluting prepared
samples. It is recommended to prepare much since it will be used in other
sample preparations.
The HNO3 dilution does not need to be poured into other containers under a
fume hood.
13
MIXING A 10 ppb TUNE A DILUTION
A 10 ppb Tune A solution is required for the calibration of the ICP-MS and
obtaining the performance reports. The Tune A solution can be mixed to
other concentration values either higher or lower depending upon the
samples for analysis.
How many millilieters of Tune A at 10 mg/L (10, 000) are needed to prepare
1 L (1000 mL) of 10 ppb Tune A?
Ci Vi = Cf Vf
After rearranging and dividing out the ppb units, the final result is:
Vi = 1 mL
Take 1 L volumetric flask and put the 1 mL first; file to the line with the 1%
to 3% nitric acid solution. Repeat the mathematic and procedure for other
concentrations of Tune A or other necessary standards. Since Tune A is used
to autotune and for the daily performance report, 1L is not too much.
14
Application note: If using an Eppendorf, make sure to practice the manual
actions of taking a sample. The best method for checking whether the
pipette and one's technique is good by taking some DI water and pipette 1
mL; weigh this 1-mL of DI water on a 3 to 4 digit scale. If weighs close to
1 gram, then the equipment and technique are good. Remember that for DI
water, 1 mL = 1 gram = 1 cubic centimeter.
Ci Vi = Cf Vf
Use this formula and procedure for making other concentrations of standard.
Reference:
Calculating Concentration
http://chemistry.about.com/library/weekly/aa081003a.htm
15
CLEANING THE CONES WITH ULTRASOUND
After some hours of sample analysis, the cones at the entry point of mass
spectrometer will become dirty.
Follow the instructions in the manual for removing the cones from the ICP-
MS.
Fill the ultrasonic unit pictured below with DI water to about an 1 inch to 1.5
inches of the top. Find a small beaker to put the cones in with some DI water
filling the beaker. Do each cone separately for about 15 minutes.
Rinse the cones a couple of times with DI water while in the beaker. Try to
minimize contact with the cones unless wearing gloves. Set the cones out to
dry. After drying, reinstall the cones.
After cleaning the cones or installing new ones, run some 300 ppm Ca+
conditioning solution to prepare cones. The Ca+ in a plasma deposits a
ceramic surface.
16
TAKING THE COVER OFF OF THE INSTRUMENT
If turning off the breakers to reset the instrument, leave them off for at least
30 seconds before turning back on. This should allow for most of the
electronics to discharge.
The breaker to the far right says “Electronics Embedded PC Rotary Pump
Turbo Pump”. This breaker also controls the roughing pump on sitting on
the floor beneath the table. If it does not come back on after resetting the
breaker, press the RED button on the board just to the right. The GRAY
button stops the roughing pump.
17
OTHER GENERAL TIPS AND GUIDELINES
3. Auto-tunes are only necessary after replacing the cone and torch, but
only do an autotune if it fails the performance reports.
8. Calibration: (2) points are enough. A blank and an upper limit. Could
be validated by preparing a standard in between those values.
10.If the elements are unknown, run sample with out the internal
standards and see what is in the sample. Then adjust or choose the
procedure depending upon elements that you are looking for.
18
11.Pay attention to interferences when setting up experiment. To reduce
interferences, choose an isotope that is less abundant.
12.As the argon gas decreases in the tank, Kr (Krypton) does not bail off
as fast and that may generate interferences that require a correction for
certain analysis (like with arsenic and selenium).
13.If the plasma is flickering, this may indicate the flow is too high. Call
maintenance to have Siemens adjust the flow rate within an acceptable
range
19