Sunteți pe pagina 1din 8

Analytica Chimica Acta 587 (2007) 89–96

Headspace–mass spectrometry determination of benzene, toluene


and the mixture of ethylbenzene and xylene isomers in
soil samples using chemometrics
F.A. Esteve-Turrillas, S. Armenta, S. Garrigues, A. Pastor, M. de la Guardia ∗
Department of Analytical Chemistry, Universitat de València, Edifici Jeroni Muñoz, 50th Dr. Moliner, 46100 Burjassot, València, Spain
Received 5 October 2006; received in revised form 5 January 2007; accepted 15 January 2007
Available online 21 January 2007

Abstract
A simple and fast method has been developed for the determination of benzene, toluene and the mixture of ethylbenzene and xylene isomers
(BTEX) in soils. Samples were introduced in 10 mL standard glass vials of a headspace (HS) autosampler together with 150 ␮L of 2,6,10,14-
tetramethylpentadecane, heated at 90 ◦ C for 10 min and introduced in the mass spectrometer by using a transfer line heated at 250 ◦ C as interface.
The volatile fraction of samples was directly introduced into the source of the mass spectrometer which was scanned from m/z 75 to 110. A partial
least squares (PLS) multivariate calibration approach based on a classical 33 calibration model was build with mixtures of benzene, toluene and
o-xylene in 2,6,10,14-tetramethylpentadecane for BTEX determination. Results obtained for BTEX analysis by HS–MS in different types of soil
samples were comparables to those obtained by the reference HS–GC–MS procedure. So, the developed procedure allowed a fast identification
and prediction of BTEX present in the samples without a prior chromatographic separation.
© 2007 Elsevier B.V. All rights reserved.

Keywords: Benzene; Toluene; Ethylbenzene; Xylene isomers; Headspace-mass spectrometry; Chemometric analysis

1. Introduction together with the chronic toxicity associated with the aromatic
ring present in their structure, make BTEX display a high
Volatile organic compounds (VOCs) are chemical products pollution potential. These aromatic hydrocarbons are widely
based on carbon chains or rings, also containing hydrogen with distributed in the environment due to natural sources; which
a vapour pressure greater than 0.27 kPa at 25 ◦ C excluding include superior plant wax, algae and plankton, and anthro-
methane. Measurement of VOCs in different kind of samples pogenic sources of hydrocarbons; that comprise domestic and
has received significant attention over the past few years because industrial wastes, biomass and wood burning, incomplete fuel
of direct and indirect impacts of individual VOCs on human oil combustion and urban runoff [7].
health and ecosystem. Many VOCs are either known or sus- Analytical methods of reference for the determination of
pected carcinogens and some have toxic effects [1–3]. The 1990 BTEX residues in environmental samples are based on gas
US Environmental Protection Agency (EPA) Clean Air Act [4] chromatography–mass spectrometry (GC–MS), using either
includes 189 hazardous air pollutants that are mostly VOCs, and static [8] or dynamic [9] headspace (HS) as sample introduction
it has been evidenced that these compounds also play a critical modules. On the other hand, some other techniques like liquid
role in formation of tropospheric ozone [5,6]. chromatography [10] or near infrared [11] have been already
BTEX [benzene, toluene, ethylbenzene and the three xylene employed.
isomers (ortho, meta and para)] are a subclass of VOCs with In recent years, several novel techniques of sample intro-
boiling points between 80 and 150 ◦ C. Their solubility in water, duction and sample pre-treatment for GC analysis, like
purge-and-trap [12], membrane extraction [13], solid-phase
microextraction [14] and single-drop microextraction [15]
∗ Corresponding author. Tel.: +34 963544838; fax: +34 963544838. have been developed for the determination of BTEX and other
E-mail address: miguel.delaguardia@uv.es (M. de la Guardia). volatile compounds in different samples. On the other hand,

0003-2670/$ – see front matter © 2007 Elsevier B.V. All rights reserved.
doi:10.1016/j.aca.2007.01.036
90 F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96

it has been done a big effort to improve the BTEX extraction Stock standard solutions employed to build the PLS calibra-
by using the accelerated solvent extraction (ASE) in order to tion models were prepared in 2,6,10,14-tetramethylpentadecane
achieve an efficient removal of analytes from different matrices at a concentration of 5.0 mL L−1 and stored in amber glass-
[16,17] prior to chromatographic analysis. stoppered bottles at 4 ◦ C. Standard solutions containing BTEX
However, the use of chromatography means long time of anal- at individual concentration of 5, 25 and 50 ␮L L−1 of
ysis and, because of that, sample screening methodologies to benzene, toluene and xylene were also prepared in 2,6,10,14-
detect BTEX residues like the combination of headspace and tetramethylpentadecane by appropriate dilution of the stock
mass spectrometry detection have been explored providing a ones. PLS calibration standards were prepared at three different
yes/no response in order to reduce the amount of samples to levels by adding 50 ␮L of the different dilutions, that corre-
be analyzed [18]. The, HS–MS technique has been successfully sponds to 0.25, 1.25 and 2.5 nL of benzene, toluene and o-xylene,
applied in qualitative applications such as the detection of hydro- inside glass headspace standard vials.
carbons in environmental matrices [19] and for classification of Spiked samples were prepared by adding 50 ␮L of each stan-
the geographical origin of olive oils [20]. dard solution at 12.5, 25 and 37.5 ␮L L−1 level, that correspond
Multivariate calibration is a very useful tool for extract- to 0.25, 0.625 and 1.875 nL of considered compounds, to 1 g of
ing chemical information from instrumental signals. The most soil.
commonly used multivariate methods for chemical analysis are Real soil samples were taken from the vicinity of petrol pump
partial least squares (PLS) regression and principal component stations in the area of Valencia and were directly heated in HS
regression (PCR), where factors that relate to variation in the glass vials after adding 150 ␮L of solvent.
response measurements are regressed against the properties of
interest. 3. Apparatus
Y = Tq × f
3.1. The extraction–injection module
where T is a matrix of latent variables and q contains the regres-
sion coefficients for the columns in T. The PLS factors (T) must A static headspace autosampler Thermo Finnigan model
model most of the variation, both in the X- and Y-variable and HS 2000 (Waltham, MA, USA) equipped with standard glass
also attain fully adequate predictive ability. Ideally, each factor vials of an internal volume of 10 mL (23 mm × 46 mm) was
added to the model would describe variation relevant for predict- employed. Once equilibrium was reached between the matrix
ing property values. To find the adequate number of PLS factors and the gaseous phase, 1 mL vapour sample was injected into
to retain in the calibration model there are different methods. the system.
One of the most employed is that which minimizes the predicted
residual error sum of squares (PRESS) of the model [21].
3.2. The transfer line
In summary, the relationships between analytical signals and
data to be determined are developed from the training set and
The transfer line ensures coupling of the extraction–injection
then applied to the set of unknowns. For unknown samples, the
module to the mass spectrometer. This line must be designed to
concentrations of various constituents can often be predicted
maintain a high-quality vacuum in the spectrometer source while
with a good accuracy.
allowing rapid transfer of the extracted molecules between the
The aim of this study has been the development of a simple
two modules. A Scharlab fused silica tube of 30 m length with
and quick method capable to identify and to quantify BTEX
an internal diameter of 0.32 mm was employed. The temperature
in soil samples using HS–MS combined with a partial least
of this tube was held at 250 ◦ C to prevent condensation of the
squares (PLS) treatment of spectral data obtained, without any
considered compounds.
chromatography separation.
A Hewlett Packard HP5MS column (Palo Alto, CA, USA)
(30 m × 0.32 mm i.d., 0.25 ␮m film thickness) was used to
2. Experimental
obtain the reference data by chromatography.
2.1. Chemicals and reagents
3.3. The mass spectrometer
Toluene (99.8%, v/v), benzene (99.7%, v/v), hexane (96%,
v/v), iso-octane (99.8%, v/v) and petroleum ether (multisol- A Thermo Finnigan Trace gas chromatograph, equipped
vent grade) were supplied by Scharlau (Barcelona, Spain) and with a Thermo Finnigan ion trap mass spectrometer detector
the isomers of xylene, ortho, meta and para (99.0%, v/v) Polaris Q was used. Ionization of the molecules introduced
and ethylbenzene (99%, v/v) were provided by Fluka (Buchs, into the source of the mass detector was carried out by elec-
Switzerland). 2,6,10,14-Tetramethylpentadecane (98%, v/v) tron ionization at 70 eV and an helium flow of 0.3 mL min−1
used to prepare the standard solutions an added to the samples to was used as damping gas. Data acquisition in full scan mode
compensate the background was purchased from Sigma–Aldrich was carried out over a mass range of m/z 75–110. The detec-
(Madrid, Spain). Methyl tert-buthyl ether (MTBE) (99%, v/v), tor temperature was fixed at 250 ◦ C. Ion trap tests and mass
used in the evaluation of interferents, was provided by Merck calibration were weekly performed with perfluorotributylamine
(Darmstadt, Germany). (PFTBA).
F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96 91

3.4. HS–GC–MS reference procedure employing their characteristic ions. However, ethylbenzene and
xylenes cannot be differentiated from each other because they
One gram of soil sample was accurately weighted inside present the same characteristic ions and only could be quantified
a 10 mL standard glass vial and 150 ␮L of 2,6,10,14- after an appropriate separation.
tetramethylpentadecane were added. After that, the vial was Additionally, in Fig. 1, it can be seen the mass spectra of
capped hermetically and heated in the headspace autosampler 2,6,10,14-tetramethylpentadecane. The specific ions generated
at 90 ◦ C for 10 min with shaking. The syringe temperature was by this molecule at m/z 83–85 and 97–99 do not overlap with the
selected at 100 ◦ C and 1 mL injection volume was introduced characteristic ions of the analytes and, thus this compound could
into the system in split mode (1:10), employing a constant be used as internal reference for the determination of BTEX
flow of 1 mL min−1 helium as carrier. The oven temperature compounds.
program started from 40 ◦ C, held for 10 min, increased at a On the other hand, it is clear from Fig. 1 that ethylbenzene and
rate of 20 ◦ C min−1 up to 200 ◦ C and finally held for 2 min. xylenes could be determined by MS without any interference of
The transfer line and source temperature were fixed at 300 and benzene nor toluene.
250 ◦ C, respectively.
4.2. HS–GC–MS recording of a BTEX mixture
3.5. Headspace–mass spectrometry procedure (HS–MS)
A conventional chromatographic separation of a mixture of
Preparation of sample and headspace conditions were kept as 1 nL of each one of BTEX compounds, using an HP-5 (5%-
for the reference procedure. In order to measure the patterns of phenyl)-methylpolysiloxane column and full scan acquisition
volatile compounds in soil samples without any chromatography mode with selected ions at m/z 77, 78, 91, 92 and 106, is shown
separation (HS–MS methodology), the column was replaced by in Fig. 2. As it can be seen in the figure, BTEX compounds
a silica tube transfer line and the oven temperature was pro- are well separated in a run time of 11 min, except meta- and
grammed to keep a constant temperature of 250 ◦ C. para-xylene that coelute. Thus, five peaks were obtained at 2.4,
4.1, 7.8, 8.4 and 10.0 min, corresponding to benzene, toluene,
3.6. Chemometric analysis ethylbenzene, (m,p)-xylenes and o-xylene, respectively.

For instrumental, measurement control and data acquisition 4.3. HS–MS determination of benzene, toluene and
it was employed the Xcalibur from Thermo Finnigan (Waltham, o-xylene
MA, USA). Data were exported in excel format and the PLS cali-
bration models were developed by means of TurboQuant Analyst The main objective of this study was to establish a fast screen-
6.0 software developed by Thermo Nicolet Corp. (Madison, WI, ing procedure based on the decrease of the run time and the
USA). increase of the sample throughput for BTEX determination in
The proposed PLS model was built using 27 standards for soils. So, the chromatography column was changed by a silica
calibration (using a classical 33 design) and it was validated empty tube as transfer line, in order to obtain one unresolved
by means of a set of 10 spiked samples with known amounts peak and to use the mass spectra of this peak for BTEX quan-
of BTEX. As it has been aforementioned, the calibration stan- tification.
dards correspond to different mixtures of benzene, toluene and In a first attempt, a silica tube of 5 m length and 0.32 mm
o-xylene at three different concentration levels (33 ). High level i.d. was assayed, but the vacuum of the mass spectrometer was
was fixed at 2.5 nL, medium value was evaluated at 1.25 nL and not acceptable, so a longer column of 30 m length and 0.32 mm
low concentration level was fixed at 0.25 nL. Concentration of i.d. was used as transfer line and the vacuum problems were
the aforementioned components in spiked samples was predicted avoided.
employing the PLS calibration using the information of the mass Using the aforementioned interface, a standard mixture of
spectra between m/z 75 and 110. 1 nL of benzene, toluene and o-xylene was measured by HS–MS
To build and select PLS models, the optimum number of fac- in full scan mode in a mass range of m/z 75–110. Fig. 3 shows the
tors was chosen to minimize the PRESS, based on the criterion obtained MS time recording for selected ions of m/z 77, 78, 91,
of Haaland and Thomas [22]. 92 and 106. Inset of Fig. 3 shows the corresponding mass spec-
trum of this wide peak from 1 to 4 min. Mass spectra obtained as
4. Results and discussion described for mixtures of benzene, toluene and o-xylene were
used for multivariate calibration based on partial least squares
4.1. MS spectra of BTEX compounds (PLS) to predict the concentration of these compounds in soil
samples.
One nanoliter benzene, toluene, ethylbenzene and each
xylene isomer standards were measured by HS–GC–MS, in 4.4. PLS models for HS–MS determination of benzene,
order to obtain the retention time and mass spectra of each com- toluene and o-xylene
pound. Mass spectra are shown in Fig. 1 in which it can be seen
the specific ions generated at m/z 77 and 78 for benzene, m/z 91 A classical design of 33 standards for the three analytes con-
and 92 for toluene and finally m/z 91 and 106 for ethylbenzene sidered (benzene, toluene and o-xylene), at three concentration
and the three xylenes. Benzene and toluene can be determined levels was used. Table 1 shows the analytical features of the
92 F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96

Fig. 1. Mass spectra of benzene, toluene, ethylbenzene, xylene and 2,6,10,14-tetramethylpentadecane obtained in the full scan mode for an individual amount of
1 nL of each one.

PLS model developed for the different compounds under study. As it can be seen, in the best conditions, using all the infor-
It has been indicated the number of factors required for predic- mation of the spectral range, without any baseline correction nor
tion of each analyte concentration with the minimum predicted spectra pre-treatment, RMSEC of 12.1, 24.2 and 34.6 and mean
residual error sum of squares (PRESS), the coefficient of deter- relative validation errors of 4.0, 5.4 and 5.9% were obtained for
mination, R2 of the calibration model, the root mean square error benzene, toluene and xylene, respectively.
of calibration (RMSEC) and the mean absolute validation errors, The principal component spectra (PCS) were obtained in
established for ten spiked soil samples previously analyzed by order to explain the contribution of factors on the analysis of
the reference procedure and not included in the calibration set. each volatile organic compound (see Fig. 4). On comparing
F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96 93

Fig. 2. Typical HS–GC–MS chromatogram obtained for a BTEX mixture of 1 nL of each compound measured in full scan mode at m/z 77, 78, 91, 92 and 106
selected ions.

Fig. 3. HS–MS time recording of a benzene, toluene and o-xylene mixture of 1 nL of each compound. Inset: Mass spectrum obtained for the recording between 1
and 4 min.

these signals with spectra in Fig. 1 it can be seen that the most exist in spectra [21]. Fig. 5 shows the evolution of both, the
discriminant original variables correspond to the m/z ratios 85, percentage of explained variance and that of the percentage of
77, 78, 91, 92 and 106, which are the major mass fragments cumulative variance for benzene determination as a function of
for the 2,6,10,14-tetramethylpentadecane, used as internal ref- the number of PCs, that four components are enough for a good
erence, and BTEX compounds, and that the signals are well modelization.
modelized with the PCS from 1 to 5 being all other variables Calibration models obtained in different sessions provided
characteristics of the background noise. different detector responses in the whole m/z range used and thus,
On the other hand, the general occurrence of positive and in the following studies the calibration model was established
negative peaks in PLS PC plots is due to the sign of corre- from measurements made in the same session than samples.
lation between the component to be determined and the PLS There are several circumstances that can introduce variations
factor. The negative correlation between a PLS factor and an in the new spectra that had not been included in the calibra-
analyte will provide a negative peak at its specific m/z. The load- tion step. The presence of these unmodeled variations in the
ings cannot be interpreted as spectra. Due to the orthogonality new data can lead to strongly biased predictions from multivari-
forces, some loadings could have negative values that cannot ate calibration models. The situation arises from changes in the

Table 1
Statistics for the PLS-MS calibration and prediction

Region Baseline Spectra pre-treatment Analyte Factorsa R2b RMSECc Mean accuracy errord (nL)

109-76 None None Benzene 4 0.9994 12.1 0.04


Toluene 4 0.998 24.2 0.05
Xylene 5 0.996 34.6 0.06
a The number of factors were chosen in order to obtain the minimum predicted residual error sum of squares.
b R2 is the coefficient of determination of the regression line between predicted and actual values of analyte concentration in the calibration set.
c Root mean square calibration error for a cross-validation.
d Mean accuracy error (%) found from the comparison of data obtained for the validation set of samples with results obtained by HS–GC–MS.
94 F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96

Table 2
Effect of the use of o-xylene in the calibration set for PLS HS–MS determination
of ethylbenzene, different xylene isomers and a mixture of all of them in the
presence of benzene and toluene
Compound Amount added (nL) %Recovery

o-Xylene 1 101 ± 6
m-Xylene 1 102 ± 9
p-Xylene 1 101 ± 11
Ethylbenzene 1 93 ± 8
Mixa 1 99 ± 12

Results are the average of three independent measurements ± the corresponding


standard deviation.
a Mix: 0.25 nL of o-xylene, m-xylene, p-xylene and ethylbenzene.

ent, also, in the case of two different sessions, due to changes in


the environment of the instrument or in its response.
A solution to the calibration transfer problem is to apply
chemometric techniques to correct for instrumental and envi-
ronmental differences, thereby making the model transferable
and avoiding full recalibration. Various methods for calibration
transfer exist in the literature, and most have been discussed in
recent reviews [23,24].

4.5. Effect of the use of different xylene isomers in the


calibration set

Ethylbenzene and the three xylene isomers present exactly


the same mass spectra in the m/z range monitored (see Fig. 1).
So, it seems impossible to do an individual determination of
all of them by HS–MS without any chromatography separation.
Due to this, the effect of the use of o-xylene in the calibration
set to determine the aforementioned compounds as total xylene
has been evaluated.
Fig. 4. Principal component spectra diagnostic applied to verify the effect of
the number of factors on the signals of benzene, toluene and o-xylene. Vertical
One nanoliter of ethylbenzene and the different xylene iso-
lines represent the characteristic m/z for each BTEX compound. Note: Benzene mers and a mixture of 0.25 nL of each one of the aforementioned
(B), toluene (T), ethylbenzene (E) and xylene (X). compounds were measured in the presence of 1 nL of both, ben-
zene and toluene. Table 2 shows the mean recovery data found
instrumental response function. These changes can occur when for the considered compounds in the assayed solutions expressed
the data to be predicted are collected on an instrument differ- as o-xylene. As it can be seen, the recovery of the different
ent from that used to build the model or simply in the same
instrument but in different measurement sessions. Although the Table 3
spectra may have the same basic shape or profile, the measured Effect of increasing amounts from 0.25 to 25 nL of different interferents on the
intensity values from two instruments will generally be differ- recovery of benzene, toluene and xylene at 1.25 nL level by PLS HS–MS
Interferent Compound %Recovery

Petroleum ether Benzene 105 ± 19


Toluene 104 ± 24
Xylene 107 ± 18
n-Hexane Benzene 91 ± 14
Toluene 98 ± 19
Xylene 106 ± 21
iso-Octane Benzene 104 ± 16
Toluene 105 ± 9
Xylene 96 ± 14
MTBEa Benzene 94 ± 13
Toluene 94 ± 14
Xylene 100 ± 9
Fig. 5. Effect of the number of PCs used in the PLS model on the explained a
variance and cumulative variance for benzene determination. MTBE: methyl tert-buthyl ether.
F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96 95

Table 4
BTEX concentration in soil samples obtained by both, reference HS–GC–MS and HS–MS developed procedures
Sample Soil concentration (nL g−1 ± s, n = 3)

HS–GC–MS HS–MS
 
Benzene Toluene E. Benzene (m,p)-Xylene o-Xylene E. Benzene + Xylenes Benzene Toluene E. Benzene + Xylenes

1 0.20 ± 0.10 1.5 ± 0.2 1.10 ± 0.08 1.68 ± 0.11 1.31 ± 0.10 4.09 –a 1.5 ± 0.2 4.2 ± 0.3
2 –a 2.50 ± 0.15 0.91 ± 0.07 –a –a 0.91 –a 2.4 ± 0.3 1.4 ± 0.2
3 –a 3.1 ± 0.2 –a –a –a –a –a 3.0 ± 0.4 –a
4 –a 1.72 ± 0.11 –a –a –a –a –a 1.5 ± 0.3 –a
5 –a 3.3 ± 0.2 0.20 ± 0.01 0.50 ± 0.03 0.53 ± 0.02 1.23 –a 3.4 ± 0.2 1.4 ± 0.2
6 –a 0.86 ± 0.10 1.4 ± 0.2 2.1 ± 0.10 1.16 ± 0.12 4.66 –a 0.5 ± 0.2 4.8 ± 0.2
7 –a –a 0.71 ± 0.08 1.81 ± 0.10 0.62 ± 0.03 3.14 –a –a 2.90 ± 0.10
8 –a 4.2 ± 0.3 –a –a –a –a –a 4.8 ± 0.3 –a
9 –a –a –a 1.52 ± 0.09 2.2 ± 0.2 3.72 –a –a 3.4 ± 0.4
10 –a –a 6.8 ± 0.4 –a –a 6.80 –a –a 7.0 ± 0.2
a Less than LOD (estimated in 0.1 nL g−1 ).

xylenes is around 100%. So, being acceptable, the use of o- The regression between concentrations found by the recom-
xylene in the calibration set for the prediction of ethylbenzene, mended procedure and those obtained by HS–GC–MS provide
m- and p-xylenes. equations CHS–MS = (−0.15 ± 0.15) + (1.08 ± 0.09)CHS–GC–MS
with a coefficient of determination R2 = 0.98 (n = 10) and
CHS–MS = (0.06 ± 0.11) + (1.00 ± 0.03)CHS–GC–MS with a coef-
4.6. Study of interferences ficient of determination R2 = 0.991 (n = 10) for toluene and
xylenes, respectively. In those equations, the intercept and slope
The main contamination sources of BTEX in soils include values were statistically comparables to 0 and 1, respectively,
the massive use of petroleum and its derivatives, and that of for a probability level of 95% (being texp lower in all cases
solvents. So, it was evaluated the interference of hydrocarbons than ttheoretical = 1.734 (n = 18)) thus indicating that the devel-
like petroleum ether, n-hexane or iso-octane together with that oped procedure provides an accuracy comparable with that of
of methyl tert-buthyl ether, a common used additive of gasoline. the chromatography method.
In order to assure the accuracy of the developed procedure, The developed procedure allowed the individual identi-
it was evaluated the interference of increasing amounts of the fication and quantification of benzene and toluene and the
considered compounds from 0.25 to 25 nL for a fixed amount of determination of the sum of ethylbenzene and xylenes. As it
benzene, toluene and xylene of 1.25 nL of each one. can be seen, benzene was undetected in actual samples, proba-
We can notice that whatever the interferent studied and the bly as a result of the easy evaporation of this compound, as well
concentration of that, the mean recovery values stay around as its low concentration in petroleum derivatives (less than 5%,
100%, as it can be seen in Table 3. So, the presence of these com- v/v). Toluene and xylenes were found in the major part of soil
pounds does not interfere on BTEX determination by HS–MS, samples analyzed, at concentrations higher than those of other
for interferent/analyte ratios from 0.2 to 20. analytes. These results are consistent with the high concentration
of these compounds in petroleum derivatives (mainly gasoline).
4.7. Real sample analysis
5. Conclusions
In a first attempt, 1 g of soil samples taken near petrol sta-
tions in the area of Valencia were introduced inside headspace The present work describes the development of an HS–MS
vials and measured by HS–MS, being the obtained results for method for the determination of aromatic hydrocarbons (BTEX)
benzene, toluene and xylene unacceptables. in soil matrices by using a PLS multivariate calibration approach.
The main difference between soil samples and standards was Considering the results, the HS–MS methodology could be
the presence of the solvent and, taking into consideration the applied for the differentiation and classification of samples, and
effect of the solvent in the PLS model which is well explained also for the quantitative determination of BTEX in soil samples.
by the PC1 (see Fig. 4), it was decided to add 2,6,10,14- In fact, benzene, toluene and the mixture of ethylbenzene and
tetramethylpentadecane to the solid samples. xylenes could be determined with an excellent accuracy. The dis-
A 150 ␮L of solvent added to the soil samples avoided the advantage of the need of a multivariate calibration technique for
mentioned problem and results found for the PLS HS–MS deter- data treatment is compensated by the time saving (no chromatog-
mination of toluene and ethylbenzene plus xylenes shown in raphy separation step is required). Results are comparables with
Table 4 indicate the excellent comparability with data found by those obtained by HS–GC–MS and the sampling throughput is
HS–GC–MS. Benzene was detected only in a single sample by considerably improved, from 5 samples per hour by HS–GC–MS
GC–MS at a level of 0.2 ± 0.1 nL g−1 . to 12 samples per hour by the developed procedure.
96 F.A. Esteve-Turrillas et al. / Analytica Chimica Acta 587 (2007) 89–96

Acknowledgements [8] Z. Wang, K. Li, M. Fingas, L. Sigouin, L. Menard, J. Chromatogr. A 971


(2002) 173–184.
[9] E. Martınez, S. Lacorte, I. Llobet, P. Viana, D. Barcelo, J. Chromatogr. A
The authors acknowledge the financial support of the Ministe-
959 (2002) 181–190.
rio de Educación y Ciencia (Project CTQ2005-05604, FEDER) [10] Y. Yang, A.D. Jones, C.D. Eaton, Anal. Chem. 71 (1999) 3808.
and Direcció General d’Investigació i Transferència Tecnològica [11] J. Buerck, S. Roth, K. Kraemer, M. Scholz, N. Klaas, J. Hazard. Mater. 83
de la Generalitat Valenciana (Project 017609). F.A. Esteve- (2001) 11–28.
Turrillas and S. Armenta also acknowledge the V segles [12] M. Rosell, S. Lacorte, A. Ginebreda, D. Barceló, J. Chromatogr. A 995
(2003) 171–184.
(Universitat de València) and FPU grant (Ministerio de Edu-
[13] B. Hauser, P. Popp, J. Chromatogr. A 909 (2001) 3–12.
cación y Ciencia (Ref. AP2002-1874)), respectively. [14] J. Ji, C. Deng, W. Shen, X. Zhang, Talanta 69 (2006) 894–899.
[15] J. Cacho, V. Ferreira, P. Fernandez, Anal. Chim. Acta 264 (1992) 311–317.
References [16] K. Li, M. Landriault, M. Fingas, M. Llompart, J. Hazard. Mater. 102 (2003)
93–104.
[1] K. Saarela, T. Tirkkonen, J. Laine-Ylijoki, J. Jurvelin, M.J. Nieuwenhui- [17] H. Giergielewicz-Mozajska, L. Da browski, J. Namiesnik, Crit. Rev. Anal.
jsen, M. Jantunen, Atmos. Environ. 37 (2003) 5563–5575. Chem. 31 (2001) 149–165.
[2] C.Y. Chan, L.Y. Chan, X.M. Wang, Y.M. Liu, S.C. Lee, S.C. Zou, G.Y. [18] F. Peña, S. Cárdenas, M. Gallego, M. Valcárcel, J. Chromatogr. A 1052
Sheng, J.M. Fu, Atmos. Environ. 36 (2002) 2039–2047. (2004) 137–143.
[3] K. Na, Y.P. Kim, Atmos. Environ. 35 (2001) 2603–2614. [19] J.L. Perez-Pavon, A. Guerrero-Peña, C. Garcia-Pinto, B. Moreno-Cordero,
[4] Amendments to 1990 Clean Air Act—List of 189 Hazardous Air Pollutants, J. Chromatogr. A 1047 (2004) 101–109.
US Environmental Protection Agency, 1994. [20] C. Cerrato-Oliveros, R. Boggia, M. Casale, C. Armanino, M. Forina, J.
[5] R. Atkinson, Atmos. Environ. 34 (2000) 2063–2101. Chromatogr. A 1076 (2005) 7–15.
[6] B.J. Finlayson-Pitts, J.N. Pitts, J. Air Waste Manage. Assoc. 43 (1993) [21] P. Geladi, Spectrochim. Acta B 58 (2003) 767–782.
1091–1100. [22] D.M. Haaland, E.V. Thomas, Anal. Chem. 60 (1988) 1193–1202.
[7] A.L. Hinwood, H.N. Berko, D. Farrar, I.E. Galbally, I.A. Weeks, Chemo- [23] X.L. Chu, H.F. Yuan, W.Z. Lu, Spectrosc. Spec. Anal. 21 (2001) 881–885.
sphere 63 (2006) 421–429. [24] T. Fearn, J. NIRS 9 (2001) 229–244.

S-ar putea să vă placă și