Sunteți pe pagina 1din 9

IVC FILTERS: A REVIEW AND WHATS NEW

Coordinator: Michael S. Stecker, MD, FSIR FACULTY


Meghal Anatani, MD Christoph Binkert, MD, FSIR Chadi Chahin, MD Albert Chun, MD Anthony Daniele, DMD, MD John Gemery, MD Bryan Hankins, MD Sanjeeva Kalva, MD Matthew Schenker, MD David Trost, MD, FSIR

Pulmonary embolism (PE) occurs with an incidence of approximately 600,000 cases annually and it has been estimated that it causes 150,000 to 200,000 deaths per year. Without treatment, PE carries a 30% mortality rate, which decreases to 8% after treatment with systemic anticoagulation. Although systemic anticoagulation remains the cornerstone of venous thromboembolism treatment, not all patients are candidates for this therapy, some fail the therapy, and some patients on anticoagulation suffer complications from the treatment. Fortunately, inferior vena cava (IVC) filtration is available for these selected patients as an adjunctive treatment for venous thromboembolism, or as an effective prophylactic measure in selected high-risk patients. Indications for IVC Filtration Classic Indications Presence of venous thromboembolic disease (pulmonary embolus or IVC, iliac, or femoropopliteal deep venous thrombosis (DVT)) combined with one of the following: o Contraindication to anticoagulation Bleeding complication of anticoagulation Known recent hemorrhage Recent major trauma or surgery Hemorrhagic stroke Ataxia or increased risk for falls Thrombocytopenia Heparin-associated thrombocythemia thrombosis syndrome Guaiac-positive stools Known central nervous system neoplasm, aneurysm, or vascular malformation o Complication of anticoagulation o Failure of anticoagulation Recurrent PE despite adequate anticoagulation Inability to achieve therapeutic systemic anticoagulation

Other Well-Accepted Indications Massive pulmonary embolism with residual DVT in a patient at risk for further PE Free-floating iliofemoral or IVC thrombus Severe cardiopulmonary disease and DVT (e.g. cor pulmonale with pulmonary hypertension) Poor compliance with anticoagulant medications

Extended Indications Prophylactic IVC filter placement in selected, high-risk patients without documented PE or DVT: o Severe trauma victims with closed head injury, spinal cord injury, and/or multiple long bone or pelvic fractures o Patients who are immobilized or subjected to prolonged intensive care o Preoperatively in patients who have multiple risk factors for venous thromboembolism Protection during iliofemoral DVT thrombolysis

It must be emphasized that the prophylactic placement of IVC filters is for prevention of clinically significant PE, not for prophylaxis against the development of DVT. Suprarenal Placement Indications Renal vein thrombosis IVC thrombosis extending up to or above the level of the renal veins Renal cell carcinoma with renal vein or IVC involvement Thrombus extending above a previously placed infrarenal filter PE after gonadal vein thrombosis Anatomic variants, such as a duplicated IVC and low insertion of the renal veins Pregnant women or women of childbearing age - this is controversial!!

Contraindications to IVC Filtration Absolute Contraindications (rare) Complete, chronic thrombosis of the IVC Inability to gain central venous access

Relative Contraindications Severe, uncorrectable coagulopathy Bacteremia and/or sepsis Pediatric and young adult patients (long-term consequences of filters are not completely known)

Success and Efficacy of IVC Filtration Technical Success IVC filter placement technical success should be at least 97% in experienced hands Rate below that should generate a quality assurance review to ascertain the cause(s) of the high failure rate

Filter Efficacy Primary indicator is the recurrent PE rate All available IVC filters have fairly comparable recurrent symptomatic PE rates, ranging between 2 and 5% True incidence of recurrent PE is probably higher, since most asymptomatic PE go undiagnosed

Since the introduction of the Mobin-Uddin umbrella in 1969 and the Greenfield filter in 1973, there have been innumerable observational and retrospective studies showing that IVC filters do reduce the PE rate compared to historical controls. A prospective, randomized trial of anticoagulated patients with proximal deep venous thrombosis by Decousus et al. compared IVC filters to a control group. This study showed a

significant benefit of filters at 12-day follow-up, with the control group experiencing a more than four fold increase in PE rate compared to the filter group (4.8% without filter vs. 1.1% with filter). This difference was even greater when only patients with PE at enrollment were considered (8.6% without filter vs. 1.1% with filter). However, there was no significant difference in mortality between these two groups. In addition, at 2-year follow-up, there was no significant difference between the two groups with respect to PE rate, but the filter group did experience more recurrent deep venous thrombosis than the control group (20.8% and 11.6%, respectively). These findings persisted in a subsequent paper reporting on 8-year follow-up of these patients. There was a significantly lower rate of symptomatic PE in the filter group compared with the control group (6.2% vs. 15.1%, p=0.008) while there was a slightly higher rate of DVT (35.7% vs. 27.5%, p=0.042). Complications of IVC Filtration Vary among the specific filters (given occurrence rates are approximate) Most complications are minor, but occasionally can be very serious Thromboembolic Events o Recurrent PE 0.5-6% o IVC thrombosis 2-30% Most partial IVC thromboses are diagnosed incidentally, remain asymptomatic, and might be better interpreted as evidence of efficient embolus trapping by the filter Complete IVC thrombosis can result in phlegmasia cerulea dolens, and can sometimes be treated with venous thrombolysis. o Recurrent DVT 20.8% o Access site thrombosis 2-28% Filter Movement o Migration 0-18% o Embolization to the right heart or pulmonary arteries 2-5% o May be precipitated by entrapped guidewires o May be treated using Interventional Radiology transcatheter techniques IVC Penetration o Rate may be as high as 41% o Clinically significant penetration is probably a rare event o May cause retroperitoneal hematoma or perforation into the aorta or gastrointestinal tract Filter Fracture 2-10% o Usually a late occurring event o Detected on plain radiographs

Special Circumstances Superior Vena Caval Filtration Experience with filters in the superior vena cava (SVC) is limited Appears to be a rare, but definite, risk of PE from upper extremity venous thromboembolism Larger studies of SVC filters suggest that they are safe, feasible, and efficacious Is an off-label use for all devices Shorter devices generally fit better Must be cautious of caval penetration of hooks as pericardial reflection may come up high and hemopericardium with tamponade is a risk Must remember to use jugular insertion kit when placing via a femoral approach and vice versa to have filter apex correctly oriented

Bed-side Vena Caval Filter Placement Numerous reports have been published demonstrating the feasibility, safety, and efficacy Major advantage is that filter may be placed without moving critically-ill patients out of the intensive care setting Imaging modalities described for this technique include portable fluoroscopy and digital subtraction angiography, transabdominal ultrasonography, and intravascular ultrasound (IVUS)

Optional Filters Temporary or retrievable filters are better classified as optional filters, since they have the option to be permanent or temporary devices Avoid long-term complications of IVC filtration Attractive alternative for patients with a time-limited need for IVC filtration, for example: o Severely injured trauma patients at high risk for pulmonary thromboembolism o Patients with venous thromboembolism and a temporary contraindication to anticoagulation, who subsequently may be able to undergo anticoagulation Currently available devices FDA-approved for retrievability: o G2 and G2 Express Filters (Bard Peripheral Vascular) o Gnther Tulip and Celect Filters (Cook Medical) o OptEase Filter (Cordis Endovascular) All of these devices are designed for retrieval from a jugular vein approach except the OptEase, which must be retrieved from the femoral approach Gnther Tulip and Recovery (predecessor of the G2) filters have both been retrieved after prolonged implantation times without repositioning

Alternative Retrieval Techniques Failed filter retrieval is most commonly due to an inability to grasp the hook or apex with a retrieval device. Authors have published many different methods to help free tilted or incorporated devices. Tip-Deflecting Wire Technique Hagspiel KD, et al. Difficult retrieval of a Recovery IVC filter. JVIR 2004; 15:645-646. A tip-deflecting wire is used through the central lumen of a recovery Cone device and used to displace the apex of the filter away from the vena cava wall in order to allow capture of the apex with the cone itself. Guidewire Shearing of Neointima Contractor S, et al. Retrieval of a tilted Gunther Tulip filter with the superior hook embedded in the caval wall. JVIR 2007; 18:1455-1456. Snare guidewire to form a loop between filter apex and caval wall and apply traction to shear the embedded tip off the caval wall, allowing access to the hook Balloon Displacement Technique Ullman JM. Technique for snaring an inaccessible Gunther Tulip filter. JVIR 2006; 17:1067-1068. An angioplasty balloon is passed below the filter, inflated, and pulled up into the cone of the filter to displace the apex from the vena cava wall to allow grasping with a snare. Curved Sheath Technique Yamagami T, et al. Successful retrieval of a Gunther Tulip vena cava filter with the assistance of a curved sheath introducer. JVIR 2005; 16:1760-1761. A curved sheath is used, via a femoral approach, to push the vena cava wall opposite an inaccessible hook in order to move the hook away from the wall, which is then snared from above.

Loop Through Filter Techniques Kuo WT, et al. Retrieval of trapped Gunther Tulip inferior vena cava filters: snare-over-guide wire loop technique. JVIR 2006; 17:1845-1849. Rubenstein L, et al. Loop-snare technique for difficult inferior vena cava filter retrievals. JVIR 2007; 18:1315-1318. A wire and/or angiographic catheter is passed between the legs of the filter, forming a loop around the filter apex. Bronchoscopic Forceps Technique Stavropoulos SW, et al. Embedded inferior vena cava filter removal: use of endobronchial forceps. JVIR 2008; 19:1297-1301. Large sheath and rigid bronchoscopic forceps are used to dissect the embedded apex free which is then grasped Retrieval Cone for Gunther Tulip and OptEase Filters Miller MJ, et al. Retrieval of the Gunther Tulip and OptEase filters using the Recovery Cone removal system. JVIR 2006; 17:899-902. The Recovery Cone device may be used to grasp and retrieve other IVC filters. Internal Jugular Vein Occlusion Van Ha TG, et al. Removal of Gunther Tulip vena cava filter through a femoral vein approach. JVIR 2005; 16:391-394. Snare initially placed via a femoral route up past filter apex, opened to encircle entire filter, closed to collapse the legs, and pull filter down into the ipsilateral iliac vein; the hook is then grasped with a second snare via the contralateral femoral approach and the device removed. Available FDA-Approved Filters (see Table) Permanent Filters Stainless Steel Greenfield Filter and Titanium Greenfield Filter Boston Scientific Corporation Descendents of the original Kimray-Greenfield Filter introduced over 30 years ago Cone-shaped design allows for 70 to 80% of the volume of the device to be filled with clot without significant reduction in caval blood flow Titanium version is MRI-compatible Stainless version has an alternating hook arrangement and is an over-the-wire device, which facilitates insertion up the left iliac venous system and improved centering within the IVC Gianturco-Roehm Birds Nest Filter Cook Medical Unique design unlike most other filters with two V-shaped struts supporting a random tangle of very fine stainless steel wire filaments, introduced in 1984 Ability to be placed into large diameter IVCs, up to 40 mm Severe magnetic susceptibility artifacts on MRI Placement is more technically complex than other filters Simon-Nitinol Filter Bard Peripheral Vascular First filter made of nitinol (alloy of nickel and titanium which has a thermo-mechanical memory), available since 1990 Very flexible at room temperature, but conforms to its original design at body temperature Allowed for first low-profile insertion sheath (9 Fr outer diameter) Was first filter approved for antecubital insertion Provides for two levels of filtration, but insertion technique requires a two-stage deployment Causes only minimal MRI artifact

VenaTech LGM Filter and VenaTech LP B. Braun Medical Conical designs with longitudinal anchoring side rails designed to better center the device to decrease tilting Eight-metal proprietary alloy with low ferromagnetism that does not cause significant artifact on MRI Same device can be placed from jugular or femoral vein approach LP version has a lower profile insertion system and has eight wires formed in a conventional conical configuration with additional side-rail wires LP version approved for IVC diameters up to 35 mm TrapEase Filter Cordis Endovascular Newest non-retrievable design with a symmetric trapezoidal shape, providing dual-level filtration Nitinol alloy is MRI-compatible Same device can be placed from jugular femoral or arm vein approach Approved for IVC diameters up to 30 mm Optional Filters Gnther Tulip Filter and Celect Filter Cook Medical First of the currently available retrievable designs, Gnther Tulip used in Europe since 1992 Made from conichrome, a non-ferromagnetic, MRI-compatible alloy Deployment is distinctly different for jugular and femoral approaches Approved for retrieval via a jugular approach using a hook at the apex Celect filter is a revised version, substituting 8 arms for the original upper looped wires to allow better centering and longer implantation intervals OptEase Filter Cordis Endovascular Very similar to TrapEase Filter, but retrievable due to only one set of securing barbs and a caudal hook Approved for retrieval up to 23 days post placement Only filter that must be retrieved from a femoral approach Recovery Filter, G2 Filter, G2 Express Filter Bard Peripheral Vascular Original Recovery Filter no longer available Two levels of filtration similar to Simon-Nitinol Filter Nitinol alloy is MRI-compatible G2 Filter is redesigned version with longer arms with hands articulated inward, rounder angle of the arms adjacent to apex, resting diameter increased to 40 mm, and more robust hooks at the ends of the six legs All versions can be retrieved via jugular approach with a Retrieval Cone, which consists of nine metal claws covered with urethane material G2 Express adds a hook on the filters head to also allow retrieval with a conventional snare Extended implantations have occurred, with filters successfully removed up to four years post insertion

Current and Upcoming IVC Filters


Name G2 G2 Express Simon Nitinol Titanium Greenfield Over-the-Wire Greenfield Vena Tech LGM Vena Tech LP Gianturco-Roehm Bird's Nest Gunther Tulip Celect TrapEase OptEase SafeFlo Option Manufacturer Bard Peripheral Bard Peripheral Bard Peripheral Boston Scientific Boston Scientific B.Braun B.Braun Cook Medical Cook Medical Cook Medical Cordis Cordis Rafael Medical Technologies Rex Medical FDA Approved
(permanent/retrieval)

Insertion Routes Fem, Jug/Subcl Fem, Jug/Subcl Fem, Jug/Subcl, Antecub Fem, Jug Fem, Jug Fem, Jug Fem, Jug Fem, Jug Fem, Jug Fem, Jug

Delivery Size 10 Fr ID / 12 Fr OD (J/S); 7 Fr ID / 9 Fr OD (Fem) 10 Fr ID / 12 Fr OD (J/S); 7 Fr ID / 9 Fr OD (Fem) 7 Fr ID / 9 Fr OD 12 Fr ID / 15 Fr OD 12 Fr ID / 15 Fr OD 10 Fr ID / 13 Fr OD 7 Fr ID / 9 Fr OD 12 Fr ID / 14 Fr OD

Composition Nitinol Nitinol Nitinol

Max Caval Diameter 28 mm 28 mm 28 mm

Filter Height MRI Compatible 41 mm 41 mm 38 mm 49 mm 49 mm 38 mm 43 mm 50-80 mm 50 mm 48 mm 50 mm 54 mm Yes Yes Yes Yes Yes (artifacts) Yes Yes Yes (artifacts) Yes Yes Yes Yes Yes 51 mm Yes

Retrievable Design, Route Yes, Jugular Yes, Jugular No No No No No No Yes, Jugular Yes, Jugular No Yes, Femoral

Notes 10 Fr ID Recovery Cone 10 Fr ID Recovery Cone or conventional snare

Yes (2005/2008) Yes (2008/2008) Yes (1990) Yes (1989) Yes (1995) Yes (1989) Yes (2001) Yes (1982) Yes (2000/2003) Yes (2007/2008) Yes (2000) Yes (2002/2004) No* No**

Beta III Titanium 28 mm 316L Stainless Steel Phynox Phynox 304L Stainless Steel 28 mm 28 mm 28 mm 40 mm 30 mm 30 mm 30 mm 30 mm
16-19mm 19-22mm 22-25mm

7 Fr ID / 9 Fr OD (Jug); Conichrome 8.5 Fr ID / 10 Fr OD (Fem) 7 Fr ID / 9 Fr OD (Jug); Conichrome 8.5 Fr ID / 10 Fr OD (Fem) Nitinol Nitinol Nitinol Nitinol

11 Fr Retrieval Set 11 Fr Retrieval Set

Fem, Jug, Antecub 6 Fr ID / 8 Fr OD Fem, Jug, Antecub 6 Fr ID / 8 Fr OD Fem, Jug Fem, Jug 6 Fr ID / 8 Fr OD 5 Fr ID / 6 Fr OD

Yes*, Same site as * CE market approval, not available in placement US

32 mm

Yes**, Jugular

** Currently under IDE investigation

REFERENCES 1. Grassi CJ, Swan TL, Cardella JF, et al. Quality improvement guidelines for percutaneous permanent inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol 2003; 14:S271-S275. 2. Kaufman JA, Kinney TB, Streiff MB, et al. Guidelines for the use of retrievable and convertible vena cava filters: report from the Society of Interventional Radiology multidisciplinary consensus conference. J Vasc Interv Radiol 2006; 17:449-459. 3. Participants in the vena caval filter consensus conference. Recommended reporting standards for vena caval filter placement and patient follow-up. J Vasc Interv Radiol 2003; 14:S427-S432. 4. Millward SF, Grassi CJ, Kinney TB, et al. Reporting standards for inferior vena caval filter placement and patient follow-up: supplement for temporary and retrievable/optional filters. J Vasc Interv Radiol 2005; 16:441-443. 5. Kinney TB. Update on inferior vena cava filters. J Vasc Interv Radiol 2003; 14:425-440. 6. Millward SF, Oliva VL, Bell SD, et al. Gnther tulip retrievable vena cava filter: results from the registry of the Canadian Interventional Radiology Association. J Vasc Interv Radiol 2001; 12:10531058. 7. Chaudry G, Padua HM, Alomari AI. The use of inferior vena cava filters in young children. J Vasc Interv Radiol 2008; 19:1103-1106. 8. Reed RA, Teitelbaum GP, Stanley P, et al. The use of inferior vena cava filters in pediatric patients for pulmonary embolus prophylaxis. Cardiovasc Intervent Radiol 1996; 19:401-405. 9. Cahn MD, Rohrer MJ, Martella MB, et al. Long-term follow-up of Greenfield inferior vena cava filter placement in children. J Vasc Surg 2001; 34:820-825. 10. Decousus H, Leizorovicz A, Parent F, et al. A clinical trial or vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. N Engl J Med 1998; 338:409415. 11. The PREPIC Study Group. Eight-year follow-up of patients with permanent vena cava filters in the prevention of pulmonary embolism: the PREPIC randomized study. Circulation 2005; 112:416-422. 12. Asch MR. Initial experience in humans with a new retrievable inferior vena cava filter. Radiology 2002; 225:835-844. 13. Ferris EJ, McCowan TC, Carver DK, McFarland DR. Percutaneous inferior vena cava filters: follow-up of seven designs in 320 patients. Radiology 1993; 188:851-856. 14. Becker DM, Philbrick JT, Selby JB. Inferior vena cava filters: indications, safety, effectiveness. Arch Intern Med 1992; 152:1985-1994. 15. Athanasoulis CA. Complications of vena cava filters. Radiology 1993; 188:614-615. 16. Spence LD, Gironta MG, Malde HM, Mickolick CT, Geisinger MA, Dolmatch BL. Acute upper extremity deep venous thrombosis: safety and effectiveness of superior vena caval filters. Radiology 1999; 210:53-58. 17. Ascher E, Hingorani A, Tsemekhin B, et al. Lessons learned from a 6-year clinical experience with superior vena cava Greenfield filters. J Vasc Surg 2000; 32:881-887.

18. Kalva SP, Chlapoutaki C, Wicky S, Greenfield AJ, Waltman AC, Athanasoulis CA. Suprarenal inferior vena cava filters: a 20-year single-center experience. J Vasc Interv Radiol 2008; 19:1041-1047. 19. Rogers FB, Strindberg G, Shackford SR, et al. Five-year follow-up of prophylactic vena cava filters in high-risk trauma patients. Arch Surg 1998; 133:406-412. 20. Pasquale M, Fabian TC, and the EAST Ad Hoc Committee on Practice Management Guideline Development. Practice management guidelines for trauma from the Eastern Association for the Surgery of Trauma. J Trauma 1998; 44:941-956. 21. Ray CE, Kaufman JA. Complications of inferior vena cava filters. Abdom Imaging 1996; 21:368-374. 22. Wojtowycz MM, Stoehr T, Crummy AB, et al. The Birds Nest inferior vena cava filter: review of a single-center experience. J Vasc Interv Radiol 1997; 8:171-179. 23. Morris CS, Rogers FB, Najarian KE, Bhave AD, Shackford SR. Current trends in vena caval filtration with the introduction of a retrievable filter at a level I trauma center. J Trauma 2004; 57:32-6. 24. Nunn CR, Neuzil D, Naslund T, et al. Cost-effective method for bedside insertion of vena caval filters in trauma patients. J Trauma 1997; 43:752-8. 25. Rosenthal D, Wellons ED, Levitt AB, Shuler FW, OConner RE, Henderson VJ. Role of prophylactic temporary inferior vena cava filters placed at the ICU bedside under intravascular ultrasound guidance in patients with multiple trauma. J Vasc Surg 2004; 40:958-64. 26. Binkert CA, Sasadeusz K, Stavropoulos SW. Retrievability of the Recovery vena cava filter after dwell times longer than 180 days. J Vasc Interv Radiol 2006; 17:299-302. 27. Rosenthal D, Wellons ED, Hancock SM, Burkett AB. Retrievability of the Gunther Tulip vena cava filter after dwell times longer the 180 days in patients with multiple trauma. J Endovasc Ther 2007; 14:406-410.

S-ar putea să vă placă și