Documente Academic
Documente Profesional
Documente Cultură
Quadrupoles
Mass Separation in space Lower sensitivity Full Scan High sensitivity SIM and SRM Offers Only MS or MS/MS
LCQ Instrumentation
Classic
Duo
3
Deca
Duo/Deca Comparisons
LCQ DUO
400um capillary 2 octopoles One rotary pump
LCQ DECA
500um capillary 1 square quadrupole & 1 octopole 2 rotary pumps
Advantage/XP Comparisons
LC Pump
ESI
Ion optics
Detector
Syringe Pump
Quadrupole refers to the shape of the ion confining field inside the trap and not the shape of or number of electrodes in the trap.
7
10
API
CI Ionization
EI
Soft
No Fragments
11
Hard
Fragments
58
100
EI
% Relative Intensity
50
0 100
20
40
60
80
100
120
140
160
180
CI
50
0 0
20
40
60
80
100
120
140
160
180
MW = 165Th
12
m/z
Ionization Techniques
200,000
ESI
15,000 1,000
APCI
Molecular Weight
TSP GC
Non Polar
FAB
PBI
Polar
13
What is API?
Atmospheric Pressure Ionization
Source Types:
1. 2.
Electrospray Ionisation (ES) Solution phase process (for the most part). APCI (Atmospheric Pressure Chemical Ionization) - Gasphase process.
Ionize the analyte (APCI) or transport ion in solution to the gas phase. Desolvate sample flow for introduction into mass spectrometer. Baffle the first vacuum region of the MS from atmospheric pressure in the source. Pump away neutrals and opposite charged ions which would otherwise interfere with the analysis of the desired polarity.
Source Purpose:
1. 2. 3. 4.
14
Peek insulator
Chemistry Considerations
ESI:
Ions formed by solution chemistry Good for Thermally labile analytes Good for Polar analytes Good for Large Molecules (Proteins / Peptides)
APCI:
Ions formed by gas phase chemistry Good for Volatile / Thermally Stable Good for Non-polar analytes Good for Small Molecules (Steroids)
17
1-5 Positions
(Left: Picture represents a
ElectrosprayBasic Layout
Heated Capillary ESI Needle Solvent evaporation and ion release +/- 5 kV +
+
++ ++ ++ ++
+
+
+ + + +
+ + + +
++ + +
++ + +
+ + + + ++
++ ++ ++ + +
+ + + + +
+ ++
+ ++
+
+
+
+
++ + +
++ + +
+ ++ +
+ ++
++ ++
+ + ++
+ +++ + +
+
+
+
+
+
+
Taylor Cone
19
20
LC Flow Rates
ESI:
3 L/min - 1mL/minute Optimal Flow Rate: 200 L/min Generally, higher flow rates require higher heated capillary temperatures and higher gas flow rates.
APCI:
200 L/min - 2mL/minute. Optimal Flow Rate: 500 L/min Generally, higher flow rates require more sheath and auxiliary gas, but do not require higher heated capillary temperatures.
21
Column 2
2.0 0.2 5
1.0 0.05 21
Capillary
< 10 l / min
22
LC Additives
Acids Do not use inorganic acids (may cause source corrosion) Formic and acetic acid are recommended Bases Do not use alkali metal bases (may cause source corrosion) Ammonium hydroxide is recommended Surfactants (surface active agents) Detergents and other surface active agents may suppress ionization Trifluoroacetic Acid (TFA) May enhance chromatographic resolution, but causes ion suppression in both negative and positive ion mode Isopropyl Alcohol May Enhance Negative Ion Formation
23
Buffers (pH)
Avoid using non-volatile HPLC additives such as: Alkali Metal Phosphates Borates Citrates Keep Buffer concentrations below 20 mM using volatile salts such as ammonium acetate. When using buffers, more frequent cleaning of the heated capillary and API stack will be necessary
24
Proton Donors Proton Acceptors Chromatographic Separation Negative ion formation Buffers
26
Solvent System
200000
300000
400000
500000
API Stack
Removal tool
Heated capillary
29
Ion Optics
Intermultipole Lens First multipole Lens Second multipole Lens
IONS IN
IONS OUT
Analyzer Mount
31
33
Vacuum System
Every mass analyzer must operate under vacuum in order to minimize both ion/molecule and molecule/ molecule collisions. At atmospheric pressure, the mean free path of a typical ion is only ca. 52 nm and at 1 mTorr, it is 40 m. Without vacuum, the ions produced in the source wont make it to the detector. The LCQ vacuum is maintained by a both rotary and turbomolecular pumps
34
760 torr
1.3 torr
1.7x10-3 torr
3.5x10-3 torr He
220 L/sec
Trapping- all scans Isolation- SIM and MSn Excitation- MSn Ejection- all scans
36
Without Helium + + + +
With Helium
He
collision
He
He He He
37
38
39
Traps injected ions by removing kinetic energy Damps ion motion to center of trap
Result...
Increase in resolution and sensitivity
40
Helium Effect
Helium flowing into trap
S#:1 RT:0.00 AV: SM:7G NL:2.50E7 1 T:+ p Full ms Relative Abundance 100 80 60 40 20 0 514 516 518 520 522 m/z 524 526 528 525.3 524.3 S#:23-32 RT:0.71-1.00AV: 0 SM:7G NL:5.61E7 1 T:+ p Full ms Relative Abundance 100 80 60 40 20 195.15 0 500 1000 m/z 1500 2000 524.26 1522.04 1322.06 1222.14 1122.21 1022.09 1621.97 1721.89 1821.95 1921.88
520
522 m/z
524
526
528
41
V qz = k ( m / e)
43
45
pre-defined time (10 ms) -Allows software to determine optimum ion injection time
46
No AGC Spectrum of Ultramark 1621, Caffeine, MRFA Calibration Mixture space charging
47
48
~ 1500 Ions
100 524.4
~ 3000 Ions
100 524.5
~ 6000 Ions
100 524.8
Relative Abundance
Relative Abundance
Relative Abundance
Relative Abundance
80 60 40 20 0 522 525.7
526.7
0 m/z
530
m/z
530
m/z
530
m/z
530
Good Resolution
49
Poor Resolution
50
Unscaled TIC
400
500
600
Injection Time
Scaled TIC
100
200
400
500
600
51
Isolation of Ions
qz
0.0 0.908
q z 2 2
52
Isolation Waveforms
~ m/z 200
q axis
.908
500 Hz
16 msec
q axis
.908
53
Intensity
Stages of Analysis
54
MS/MS Parameters
Intensity
0.0
1.0
2.0
3.0
4.0
77001-1285 970219
qz
0.0
0.225
0.908
1/4
qz
1/3
0.0
0.30
0.908
qz
1/2
0.0
56
0.45
0.908
1. Collect
3. Fragment
2. Isolate
4. Eject
57
For Scans: All By: Ring Electrode Method: Alternating RF frequency (760 kHz) at a set amplitude along with He dampening gas traps and cools the ions to the center of the trap.
58
For Scans: SIM, MSn By: Endcap Electrodes Method: a) Tailored waveform applied to all ions in the trap except ion of interest b) Thus, only ions of interest remain in the trap.
59
60
By: Ring Electrode Method: Ramp ring RF power to increase the q values of all ions in desired scan range, low mass to high mass. (i.e. Mass Selective Scanning) Also, ramp the RF amplitude on the endcap electrodes to consolidate the ions to a group (Resonance Ejection)
61
Tune Page
convection gauge pressure < 1.5 torr?
62
Ultramark stock: Measure 10 l of Ultramark 1621, and dissolve it in 10 ml acetonitrile ESI calibration solution Into a clean vial pipette 100 l of caffeine stock, 5 l of MRFA stock and 2.5 ml of Ultramark stock. Add 50 l of glacial acetic acid and 2.34 ml 50:50 methanol:water
63
64
TSQ 15, TSQ 45, TSQ 46, TSQ 70, TSQ 700, TSQ 7000, TSQ, TSQ Quantum
65
Smaller because
8degrees 25 cm quad 25 cm quad
25 cm quad
Quantum
25 cm quad 90 degrees
66
Forms Pure Quadrupolar Fields Reduces Fringing Effects Significantly Improves Resolution Improves Transmission Improves Peak Shapes
67
+ +
The ion is transmitted along the quadrupole in a stable trajectory Rf field. The ion does not have a stable trajectory and is ejected from the quadrupole.
68
HYPERQUAD
T r a n s m % i s s i o n
ROUND RODS
70
Resolution
6000
4000 2000
Quantum, API 4000, Ultima at 0.7 FWHM at m/z 1000 R = 1428 R is relatively flat across m/z
m/z
71
72
0.3 FWHM
0.2 FWHM
0.1 FWHM
73
2.5e6
2.1e6
1.9e6
74
75
ms/amu
+
+ + +
To detector
76
77
5 sec
78
5 sec
1 sec
1 sec
79
5 sec
+ + +
80
+ + +
81
So, for full scan MS, a trap will give better sensitivity because there are more ions representing each m/z arriving at the detector for each scan.
+ + +
82
Accounting for pre-scan/interscan timing, the trap produces ca. 300 times (750 ms/2.5 ms) for the collection of each m/z compared to a quadrupole (i.e 2 orders of magnitude).
+ + +
83
But
What if I wanted to pass (filter) only one m/z ion to the detector (i.e. SIM or SRM) then I could spend more time on that ion
+ + +
84
+ + +
85
Interscan delay (ISD) is the time taken to return all system voltages to the start values and reach a stable state
86
In Addition
While the beam instrument is continuously detecting one particular m/z a trap builds a curve from an average over each collection time and the points are least frequent at the most important region for quantitation (the take off).
87
For Example
SRM of 5 pg Alprazolam with LCQ Deca gives a %RSD of 6.11 while SRM of 750 fg Alprazolam with TSQ 7000 gives a %RSD of 1.87. Signal to noise ratio (S/N) are similar (20:1 and 28:1 respectively).
88
%RSD 6.11
%RSD 1.87
89
In general, faster scanning produces more points across a chromatographic peak, hence better precision and lower LOQs. In fact, for an ion-trap, scan speed refers only to the time taken to scan ions from the trap during mass analysis. Scan speed does not refer to the total analytical cycle. In an ion trap device, an MSn analytical scan comprises at least four events:
90
1- AGC Pre-scan 2- Ion Injection (usually the rate-determining step) 3- Isolation and activation of the parent ion within the trap 4- Scanning the ions out of the trap (mass analysis)
91
MS Scan Function
Ion isolation Ion Injection Ion Activation Mass Analysis
AGC Prescan
92
Analytical Scan
Scan Terminology
Prescan
Mass Analysis
Prescan
Mass Analysis
Prescan
Mass Analysis
1st Microscan
3rd Microscan
Save Data
93
94
Quantitation m/z 500 Isol /Activ //Download scanning 135-510 Isol /Activ Download
60 msec 375 amu @ 13,000 amu/sec 60 msec 375 amu @ 13,000 amu/sec Isol /Activ //Download time80 Injection Isol /Activ Download time80 Injection Pre-scan 60 msec Pre-scan 60 msec 375 amu @ 13,000 amu/sectime Total scan time 375 amu @ 13,000 amu/sec 30 Total scan 30 Isol/Activ/ Download Isol/Activ/ Download time80Scans / 10 sec500 peak Injection time80Scans / 10 sec500 peak wide Injection wide 375 amu @ 5500 amu/sec time 375 amu @ 5500 amu/sec time Total scan 70 670 Total scan 70 670 Injection time Scans / 10 sec500 peak Injection time Scans / 10 sec500 peak wide 15 wide 15 Total scan time Total scan time Scans //10 sec wide peak Scans 10 sec wide peak
95
Pre-scan Pre-scan
710 710 14 14
The only significant way to increase the sampling rate across the peak is to reduce the injection time which can be achieved in two ways: 1- Set a lower max injection time which reduces the number of ions in the trap, hence sensitivity 2- Increase the efficiency of the source and lenses to improve the transmission of ions; I.e filling the trap to the same level in a shorter period of time.
96
Software selects most intense ion from scan event 1 as precursor ion for ms2 experiment in scan event 2, provide that its intensity is above a user selected threshold
m/z
m/z
97
MS
407
MS/MS
377
Threshold
231 365
MS
407 452
m/z
206 255
MS/MS
377
m/z
m/z 98
m/z
SRM Sensitivity Neutral Loss Scan Mode Parent (Precursor) Scan Mode
Major Strengths of the LCQ Deca XP Plus
MSn Scan Mode Full Scan MS/MS Sensitivity Consecutive Reaction Monitoring (CRM)
99
What are neutral loss scans ? Both Q1 and Q3 are scanned together Q3 is offset by the neutral loss under investigation The precursor ions collide with Argon gas in Q2 to create fragment ions Only those compounds which give a fragment having that specific loss are detected Since both Q1 and Q3 are scanning, neutral loss scan mode is slower than any other mode
100
H2 N N H2 N N
- m/z 84
H2 N N H2 N N
- m/z 84
N
NH2 N H2 N N
NH2
- m/z 84
H2 N N
HO N
- m/z 84
HO N
101
Precursor ion scans also known as parent ion scans Q1 is scanned Q3 is set to allow only a fragment ion of one m/z to pass; (Q3 fixed) ions collide with Argon gas in Q2 to create fragment or product ions Only those compounds which give that specific fragment ion are detected
102
m/z 192
HO N
m/z 192
N N
m/z 84
H 2N N H 2N N
m/z 268
m/z 238
103
104
Quadrupoles
Mass Separation in space Lower sensitivity Full Scan High sensitivity SIM and SRM Offers Only MS or MS/MS
105