Mechanical Pressure Measurement

Differential Pressure Gauges Model A2G-10, air2guide P


WIKADatasheetA2G-10

Applications
Fordry,clean,non-aggressivegases,air Monitoringofblowersandventilators Filtermonitoringviadifferentialpressure Positiveoverpressureincleanrooms Positiveandnegativepressureinisolationchambers

Special Features
Zeroadjustmentinfront Redmarkpointerstandard Simplemountinganddismounting Twopartconstruction(measuringelementandcase) Optionalversionwithadditionalelectricaloutputsignals (air2guideP+E,seedatasheetA2G-15) Differential Pressure Gauge air2guide P Model A2G-10

Description
Design PerEN837-3andHVACregulations Nominal size 4 Accuracy 3.0%offullspan Ranges 0/0.2InWCupto0/50InWC Compound ranges Startingfrom-0.1/+0.1InWC (otherpressureunitsanddualscalesareavailable) Pressure limitation Steady: fullscalevalue Fluctuating: fullscalevalue Overpressure Safety 80InWC Max working pressure (static pressure) 80InWC Temperature limits Ambient: -22F...176F Media: 3F...122F Temperature effect Whenthetemperatureofthemeasuringsystemdeviates fromthereferencetemperature(+68F): max.0.5%/18Foffullscalevalue Ingress protection NEMA3(IP54perEN60529)

WIKA Datasheet A2G -10 03/2011

Page 1 of 2
R

Standard version
Process connection Plastic Lowermount(LM)orbackmountG1/8female forhoseconnector Suppliedwithstraightorangledhosebarbadaptors 1/8/3/16(standard).1/4hosebarb adaptorsavailable Separating diaphragm Siliconerubber Movement Frictionlessmagnetictransmission Dial Aluminium,scaleover90o. Pointer and mark pointer Plastic Housing Blackthermoplastic,highimpactresistant Window Polycarbonate
2.64

Dimensions (inches)
Standard version
5.2 0.15

Surface mount, lower connection

3.27

3.54

1.57

4.65(3x4.7)

0.59 2.36-3.35 1.2 0.04-2.76 0.94 0.15

Panel mount, back connection

Ingressprotection,NEMA4(IP65) Internationallyrecognizedtraceablecalibration certificate MeasuringhosePVCorsilicone,innerdiameter 4or6mm,length25mPVCorsiliconetube1/8 or1/4insidediameter,80feetlong 2ductconnectors Outputsignal4...20mA,2-wire,powersupply DC14Voroutputsignal0...10V,3-wire,power supplyDC15V(seedatasheetA2G-15) Withoutcase

5.2

1)Mountingforwallthicknessfrom0.04to2.76

Ordering information
Model / Connection location / Wall mount or panel mount / Inner dimensions of the threaded pressure connection / Scale range / Options
Page 2 of 2
Ordering information Pressuregaugemodel/Nominalsize/Scalerange/Sizeofconnection/Optionalextrasrequired Specificationsanddimensionsgiveninthisleafletrepresentthestateofengineeringatthetimeofprinting. Modificationsmaytakeplaceandmaterialsspecifiedmaybereplacedbyotherswithoutpriornotice.

WIKA Datasheet A2G-10 03/2011


R

WIKA Instrument Corporation 1000WiegandBoulevard Lawrenceville,GA30043 Tel(770)513-8200Toll-free1-888-WIKA-USA Fax(770)338-5118 E-Mail info@wika.com www.wika.com

4.37

Options

4.65(3x4.7)