Documente Academic
Documente Profesional
Documente Cultură
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to
this laboratory to perform the following calibrations1:
I.
Chemical Quantities
Parameter/Equipment
CMC2 ()
Range
Comments
Conductivity3
Measure
84 S
1413 S
0.51 S
0.51 S
Comparison to standard
solutions
pH Measure3
0.02 pH unit
Comparison to standard
solutions
II. Dimensional
Parameter/Equipment
CMC2 ()
Range
Comments
Micrometers Length
and Flatness3
Resolution:
0.001 mm
(0 to 25) mm
4 m
Page 1 of 15
III.
Parameter/Equipment
DC Voltage Generate
Range
CMC2, 4, 5 ()
Comments
(0 to 329.9999) mV
(0 to 3.299999) V
(0 to 32.99999) V
(33 to 329.9999) V
(330 to 1020) V
3.7 V + 58 V/V
5.8 V + 58 V/V
57 V + 58 V/V
0.6 mV + 64 V/V
0.76 mV + 67 V/V
Fluke 5500A
220 mV
2.2 V
11 V
22 V
220 V
1100 V
Fluke 5720A
DC Voltage Measure3,
Fixed Points
100 mV
1V
10 V
100 V
1000 V
0.0012 mV + 11 V/V
0.0012 mV + 9.5 V/V
0.0026 mV + 9.5 V/V
0.036 mV + 12 V/V
0.12 mV + 1.2 V/V
DC Current Generate3
(0 to 3.29999) mA
(0 to 32.9999) mA
(0 to 329.999) mA
(0 to 2.19999) A
(0 to 11) A
(0 to 220.000) A
(0 to 2.20000) mA
(0 to 22.0000) mA
(0 to 220.000) mA
(0 to 2.200000) A
Fluke 5500A
100 nA
1 A
10 A
100 A
1 mA
10 mA
100 mA
1A
0.047 nA + 36 V/V
0.0001 A + 24 V/V
0.00012 A + 24 V/V
0.00095 A + 24 V/V
0.0000059 mA + 24 V/V
0.000059 mA + 24 V/V
0.001 mA + 42 V/V
0.000012 A + 0.13 mV/V
Fixed Points
DC Current Measure3
Fixed Points
Fluke 5720A
Page 2 of 15
Parameter/Range
Frequency
CMC2, 4, 5 ()
Comments
AC Voltage Measure
(0 to 100) mV
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
0.0035 mV + 0.036 %
0.0013 mV + 0.024 %
0.0013 mV + 0.036 %
0.0013 mV + 0.12 %
0.0013 mV + 0.6 %
0.0024 mV + 4.8 %
100 mV to 10 V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(1 to 2) MHz
0.000048 V + 43 V/V
0.000024 V + 83 V/V
0.000024 V + 0.017 %
0.000024 V + 0.036 %
0.000024 V + 0.095 %
0.00012 V + 0.36 %
0.00012 V + 1.2 %
0.0012 V + 1.8 %
(10 to 100) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
0.0047 V + 0.024 %
0.0024 V + 0.024 %
0.0024 V + 0.024 %
0.0024 V + 0.042 %
0.0024 V + 0.15 %
0.012 V + 0.48 %
0.012 V + 1.8 %
(100 to 1000) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.05 V + 0.048 %
0.028 V + 0.048 %
0.024 V + 0.071 %
0.024 V + 0.15 %
0.024 V + 0.36 %
(0 to 32.999) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.023 mV + 0.41 %
0.023 mV + 0.18 %
0.023 mV + 0.24 %
0.024 mV + 0.29 %
0.039 mV + 0.41 %
0.072 mV + 1.2 %
(33 to 329.999) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
0.058 mV + 0.29 %
0.024 mV + 0.058 %
0.024 mV + 0.12 %
0.047 mV + 0.19 %
AC Voltage Generate
Fluke 5500A
Page 3 of 15
Frequency
CMC2, 4 ()
(33 to 329.999) mV
0.2 mV + 0.28 %
0.39 mV + 0.81 %
(330 to 3.29999) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.29 mV + 0.18 %
0.082 mV + 0.035 %
0.076 mV + 0.093 %
0.36 mV + 0.17 %
2 mV + 0.28 %
3.9 mV + 0.58 %
(3.3. to 32.9999) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
2.9 mV + 0.18 %
0.83 V + 0.046 %
3.1 mV + 0.093 %
5.9 mV + 0.22 %
20 mV + 0.28 %
(33 to 329.999) V
45 Hz to 1 kHz
(1 to 10) kHz
(10 to 20) kHz
9.6 mV + 0.058 %
0.018 V + 0.093 %
0.039 V + 0.11 %
(330 to 1020) V
45 Hz to 1 kHz
0.096 V + 0.058 %
(1 to 2.2) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300 ) kHz
5.8 V + 0.034 %
5.1 V + 0.012 %
4.9 V + 98 V/V
4.9 V + 0.025 %
6.7 V + 0.067 %
13 V + 0.15 %
(2.2 to 22) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
4.9 V + 0.03 %
4.9 V + 0.011 %
4.8 V + 96 V/V
4.8 V + 0.024 %
6 V + 0.06 %
12 V + 0.13 %
24 V + 0.17 %
24 V + 0.32 %
Parameter/Range
Comments
AC Voltage Generate
(cont)
Fluke 5500A
Page 4 of 15
Parameter/Range
CMC2, 4 ()
Frequency
Comments
AC Voltage Generate
(cont)
(22 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
14 V + 0.029 %
9.4 V + 0.011 %
9.2 V + 92 VV
8.8 V + 0.024 %
20 V + 0.055 %
24 V + 0.11 %
30 V + 0.17 %
53 V + 0.32 %
220 mV to 2.2 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.054 mV + 0.3 %
0.019 mV + 0.11 %
0.01 mV + 55 V/V
0.014 mV + 88 V/V
0.038 mV + 0.014 %
0.096 mV + 0.05 %
0.24 mV + 0.12 %
0.35 mV + 0.21 %
(2.2 to 220) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
0.54 mV + 0.03 %
0.2 mV + 0.011 %
0.067 mV + 53 V/V
0.13 mV + 91 V/V
0.26 mV + 0.02 %
0.72 mV + 0.033 %
2.4 mV + 0.12 %
3.8 mV + 0.18 %
(22 to 220) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
5.4 mV + 0.03 %
1.8 mV + 0.011 %
0.85 mV + 59 V/V
1.4 mV + 99 V/V
3.1 mV + 0.018 %
19 mV + 0.011 %
(220 to 1100) V
(15 to 50) Hz
50 Hz to 1 kHz
20 mV + 0.036 %
5.5 mV + 82 V/V
0.3 mV to 3.5 V
(10 to 30) Hz
30 Hz to 120 kHz
0.27 %
0.14 %
Fluke 5500A
Page 5 of 15
Parameter/Range
Frequency
CMC2, 4, 5 ()
Comments
AC Voltage Generate
(cont)
(0.3 to 1.1) mV
0.46 %
0.62 %
0.80 %
2.4 %
(0.3 to 3) mV
0.22 %
0.37 %
0.55 %
1.4 %
3 mV to 3.5 V
0.12 %
0.21 %
0.38 %
0.86 %
(0 to 100) A
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 1 kHz
0.024 A + 0.48 %
0.024 A + 0.18 %
0.024 A + 0.071 %
0.024 A + 0.071 %
(1 to 100) mA
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.024 mA + 0.48 %
0.024 mA + 0.18 %
0.024 mA + 0.071 %
0.024 mA + 0.036 %
0.024 mA + 0.071 %
0.047 mA + 0.48 %
0.18 mA + 0.65 %
100 mA to 1 A
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
0.00024 A + 0.48 %
0.00024 A + 0.19 %
0.00024 A + 0.095 %
0.00024 A + 0.12 %
0.00024 A + 0.35 %
0.00047 A + 1.2 %
AC Current Measure
Page 6 of 15
Parameter/Range
Frequency
CMC2, 4, 9 ()
Comments
AC Current Generate
(9 to 220) A
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
25 nA + 0.036 %
16 nA + 0.024 %
12 nA + 0.017 %
18 nA + 0.042 %
95nA + 0.16 %
(0.22 to 2.2) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.073 A + 0.037 %
0.068 A + 0.025 %
0.054 A + 0.019 %
0.17 A + 0.032 %
0.95 A + 0.16 %
(2.2 to 22) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.67 A + 0.032 %
0.57 A + 0.02 %
0.46 A + 0.016 %
0.73 A + 0.027 %
5.9 A + 0.13 %
(22 to 220) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
6.1 A + 0.032 %
5 A + 0.02 %
3.4 A + 0.017 %
12 A + 0.14 %
(0.22 to 2.2) A
20 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.044 mA + 0.031 %
0.11 mA + 0.053 %
0.29 mA + 0.83 %
(0.029 to 0.32999) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.18 A + 0.29 %
0.17 A + 0.15 %
0.29 A + 0.15 %
0.18 A + 0.47 %
0.18 A + 1.5 %
(0.33 to 3.2999) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.39 A + 0.23 %
0.38 A + 0.12 %
0.38 A + 0.12 %
0.38 A + 0.23 %
0.37 A + 0.70 %
(33 to 329.99) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
4.3 A + 0.23 %
3.8 A + 0.12 %
3.8 A + 0.11%
3.7 A + 0.24 %
35 A + 0.70 %
Fluke 5720A
Page 7 of 15
Parameter/Range
Frequency
CMC2, 4, 9 ()
Comments
AC Current Generate
(cont)
(0.33 to 2.19999) A
(10 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
0.37 mA + 0.23 %
0.37 mA + 0.12 %
0.36 mA + 0.87 %
(2.2 to 11) A
(45 to 65) Hz
(65 to 500) Hz
500 Hz to 1 kHz
2.7 mA + 0.068 %
2.5 mA + 0.12 %
2.4 mA + 0.39 %
Parameter/Equipment
Range
CMC2, 4, 5 ()
Fluke 5500A
Comments
Resistance Measure
10
100
1 k
10 k
100 k
1 M
10 M
100 M
1 G
0.000081 + 17 /
0.00059 + 15 /
0.00000061 k + 12 /
0.0000062 k + 12 /
0.000061 k + 12 /
0.0000024 M + 18 /
0.00012 M + 60 /
0.0012 M + 0.06 %
0.000012 G + 0.6 %
Resistance Generate3
(0 to 10.99)
(11 to 32.999)
(33 to 109.999)
(110 to 329.999)
(0.330 to 1.09999) k
(1.1 to 3.29999) k
(3.3 to 10.9999) k
(11 to 32.99999) k
(33 to 109.999) k
(110 to 329.999) k
(0.33 to 1.09999) M
(1.1 to 3.29999) M
(3.3 to 10.9999) M
(11 to 32.9999) M
(33 to 109.999) M
(110 to 330) M
9.4 m + 0.014 %
17 m + 0.014 %
17 m + 0.011 %
17 m + 0.011 %
70 m + 0.011 %
77 m + 0.011 %
0.71 + 0.011 %
0.7 + 0.011 %
7 + 0.013 %
7 + 0.014 %
64 + 0.018 %
64 + 0.018 %
640 + 0.070 %
630 + 0.12 %
6.4 k + 0.58 %
19 k + 0.58 %
Fluke 5500A
Page 8 of 15
Parameter/Equipment
Range
CMC2, 5, 9 ()
Comments
Resistance Generate,
Fixed Points3
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
0.00012
0.00022
0.00028
0.00058
0.0013
0.0024
0.00001 k
0.00002 k
0.00001 k
0.00019 k
0.0013 k
0.0025 k
0.000024 M
0.000047 M
0.00047 M
0.0013 M
0.012 M
Fluke 5720A
Capacitance Generate
(0.33 to 0.4999) nF
(0.5 to 1.0999) nF
(1.1 to 3.2999) nF
(3.3 to 10.999) nF
(11 to 32.999) nF
(33 to 109.99) nF
(110 to 329.99) nF
(0.33 to 1.0999) F
(1.1 to 3.2999) F
(3.3 to 10.999) F
(11 to 32.999) F
(33 to 109.99) F
(110 to 329.99) F
(330 to 1100) F
0.012 nF + 0.50 %
0.012 nF + 0.57 %
0.012 nF + 0.59 %
0.012 nF + 0.58 %
0.12 nF + 0.30%
0.12 nF + 0.30 %
0.35 nF + 0.30 %
1.2 nF + 0.30 %
3.5 nF + 0.41 %
0.011 F + 0.41 %
0.035 F + 0.47 %
0.11 F + 0.59 %
0.35 F + 0.82 %
0.27 F + 1.19 %
Fluke 5500A
Page 9 of 15
Parameter/Equipment
Range
CMC2, 9 ()
Comments
Electrical Calibration
of Thermocouple
Indicating Devices
Measure and Generate
Type E
(-250 to -100) C
(-100 to -25) C
(-25 to 350) C
(350 to 650) C
(650 to 1000) C
0.58 C
0.19 C
0.17 C
0.19 C
0.26 C
Type J
(-210 to -100) C
(-100 to -30) C
(-30 to 150) C
(150 to 760) C
(760 to 1200) C
0.31 C
0.27 C + 0.09 %
0.16 C
0.25 C
0.28 C
Type K
(-200 to -100) C
(-100 to -25) C
(-25 to 120) C
(120 to 1000) C
(100 to 1372) C
0.41 C
0.26 C
0.24 C
0.3 C
0.47 C
Type R
(0 to 250) C
(250 to 400) C
(400 to 1000) C
(1000 to 1767) C
0.66 C
0.43 C
0.38 C
0.46 C
Type T
(-250 to -150) C
(-150 to 0) C
(0 to 120) C
(120 to 400) C
0.73 C
0.32 C
0.24 C
0.17 C
Oscilloscope
Fluke 5500A
1 mV to 6.6 V(pk-pk)
1 mV to 130 V(pk-pk)
0.32 % + 60 V
0.3 % + 60 V
5 mV to 5.5 V
2 % + 110 V
Amplitude
3.5 % + 300 V
4 % + 300 V
Page 10 of 15
Parameter/Equipment
CMC2, 5 ()
Range
Comments
Oscilloscope (cont)
Flatness referenced to
50 kHz reference
0.64 % + 110 V
0.60 % + 170 V
5 s to 50 ms
20 ms to 100 ns
Rise Time
300 ps
2.4 ns + 0.1 %
Frequency
CMC2, 9 ()
Comments
Attention Measure
Coaxial, 10 dB Step
(0 to 70) dB
(71 to 139) dB
(0 to 70) dB
(71 to 139) dB
DC to 100 kHz
Agilent
N5531S/OPT518
Amplitude
Modulation
Generate
Rate: 6 MHz to 50
GHz, 0 % to 100 %
HP 83650B
Page 11 of 15
Frequency
CMC2, 5, 6 ()
Rate: 50 Hz to 10 kHz,
5 % to 99 %
50 Hz to 100 kHz,
20 % to 99 %
10 MHz to 3 GHz
50 Hz to 100 kHz,
5 % to 20 %
10 MHz to 3 GHz
(3 to 26.5) GHz,
5 % to 20 %
(3 to 26.5) GHz
(3 to 26.5) GHz,
20 % to 99 %
(3 to 26.5) GHz
Rate: 20 Hz to 10 kHz
Dev: 20 Hz to 40 kHz
peak
0.0057 kHz + 18 Hz
0.0066 kHz + 12 Hz
0.0051 kHz + 18 Hz
0.0061 kHz + 12 Hz
0.004 kHz + 30 Hz
0.0064 kHz + 12 Hz
0.0038 kHz + 44 Hz
0.0076 kHz + 12 Hz
Parameter/Range
Comments
Amplitude Modulation
Measure
Agilent N5531S
Frequency Modulation
Measure
Agilent N5531S
Page 12 of 15
Parameter/Range
Frequency
CMC2, 9 ()
Comments
Frequency Modulation
Generate
Rate:
1 kHz rate
Max. Dev. 2 MHz
Max. Dev. 4 MHz
Max. Dev. 8 MHz
Max. Dev. 16 MHz
Max. Dev. 32 MHz
Max. Dev. 64 MHz
Max. Dev. 128 MHz
24 Hz + 0.04 MHz/MHz
HP 83640B
0.012 rad + 1 %
HP 8648C
0.0071 rad + 1 %
0.0071 rad + 1 %
0.0073 rad + 1 %
Agilent N5531S
0.065 %
Agilent N5531S
50 MHz
0.008 mW
HP 478A
Phase Modulation
Generate
1 kHz rate
Max. Dev. (0 to 20) rad
Max. Dev. (0 to 40) rad
Max. Dev. (0 to 80) rad
Max. Dev. (0 to 160) rad
Max. Dev. (0 to 320) rad
Max. Dev. (0 to 640) rad
Max. Dev. (0 to 1280) rad
Phase Modulation
Measure
>0.7 rad Dev.
>0.6 rad Dev.
>1.2 rad Dev.
Distortion Measure
Rate: 20 Hz to 250 kHz
(0.01 to 100) %
Power Measure,
1 mW reference
Page 13 of 15
V. Mechanical
Parameter/Equipment
Scales, Fixed Points
Range
20 mg
50 mg
100 mg
200 mg
500 mg
1g
2g
5g
10 g
20 g
50 g
100 g
CMC2 ()
0.2 mg
0.2 mg
0.2 mg
0.2 mg
0.2 mg
0.2 mg
0.2 mg
0.3 mg
0.3 mg
0.3 mg
0.4 mg
0.6 mg
Comments
Class 1 and F weights
Range
CMC2 ()
Comments
Frequency Measuring
Equipment
1 Hz to 20.999 MHz
0.01 nHz/Hz
HP 3325A w/58503A
10 MHz to 40 GHz
0.01 nHz/Hz
HP 83640B w/58503A
Frequency Measure
1 Hz to 225 MHz
0.01 nHz/Hz
HP 53131A w/ 58503A
0.01 nHz/Hz
Agilent E4440A
w/58503A
____________________________________________
1
This laboratory offers commercial calibration service and field calibration service.
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a
laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration
and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 %
level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a
specific calibration performed by the laboratory may be greater than the CMC due to the behavior of
the customers device and to influences from the circumstances of the specific calibration.
Page 14 of 15
Field calibration service is available for this calibration and this laboratory meets A2LA R104 General
Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer's site can normally be
expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects
such as the environment at the place of calibration and for other possible adverse effects such as those
caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty
introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own,
could result in the actual measurement uncertainty achievable on a customers site being larger than the
CMC.
The measurands stated are generated with the Fluke 5500A, Fluke 5700A series of instruments. This
capability is suitable for the calibration of the devices intended to measure the stated measurand in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a
fraction of the reading plus a fixed floor specification.
The measurands stated are measured with the HP 3458A series of instruments. This capability is
suitable for the calibration of the devices intended to generate the measurand in the ranges indicated.
CMCs are expressed as either a specific value that covers the full range or as a combination of the
fraction of the reading/output plus a range specification.
In the statement of CMC, L is the numerical value of the nominal length of the device measured in
inches. In the statement of CMC, R is the numerical value of the resolution of the device in
microinches.
Pitch diameter is measured by the three-wire method. D is the numerical value of the
nominal diameter of the device, measured in inches. AM is amplitude modulation.
Where ranges are not specified, the CMC stated is for the cardinal points only.
"Supermicrometer" is a registered trade mark with a last listed owner of Pratt & Whitney Measurement
Systems, Inc., Connecticut U.S.A.
In the statement of CMC, percentages are percentage of reading, unless otherwise defined.
Page 15 of 15
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General Requirements for
the Competence of Testing and Calibration Laboratories. This laboratory also meets the requirements of ANSI/NCSL Z540-1-1994 and any
additional program requirements in the field of calibration. This accreditation demonstrates technical competence for a defined scope and the
operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqu dated 8 January 2009).
Presented this 3rd day of January 2012.
_______________________
President & CEO
For the Accreditation Council
Certificate Number 3239.01
Valid to December 31, 2013
For the calibrations to which this accreditation applies, please refer to the laboratorys Calibration Scope of Accreditation.