Documente Academic
Documente Profesional
Documente Cultură
Achieving High Resolution at High Beam Energy High-Resolution Imaging at Low Voltage Contrast Formation in Low Voltage Images Radiation Damage at Low Beam Energies SEM Observation of Non-Conducting Samples at Low Voltage The Effects of the Choice of Electron Energy STEM-in-SEM: High Resolution for the Special Case of Thin Specimens Variable-Pressure and Environmental SEM The Effect of the Gas on Charging Imaging in the VPSEM and ESEM Gaseous SE Detector (GSED)
One disadvantage is that X-ray analysis can only give the K lines for the very lightest elements and only L & M lines for heavier elements, due to the lower energy of the electrons. Contamination is a serious limitation to low-voltage microcopy because the range of the signals is so short that the contamination may dominate the image as it builds up. The rate of contamination build-up can be reduced by minimizing the electron dose to the specimen with the following procedures: Avoiding high magnification except when essential. Focusing and stigmatizing the image in an area remote from the one that is to be used to record an image. Never using spot or reduced-area raster modes.
Comparison of Interaction Volumes: High Acceleration Voltage vs. Low Acceleration Voltage
Stopping Power Variation as a Function of Energy for Electron Traveling through Carbon
Fig. 1 Typically between 30 and 100eV corresponding to the energy at which the stopping power reaches its maximum value.
The following figure shows the calculated electron ranges as a function of incident energy for four elements.
Fig. 2 It can be seen that the range decreases rapidly with energy falling from a micrometer or more at 10keV to only tens of nanometers for energies below 1keV. Note that while there are significant differences in the range at higher energies for the various targets, at the lowest energies all materials have approximately the same range.
Schematic representation of the spatial distribution of SE1 and SE2 electrons at (a) high energy (20keV) and (b) low energy (1keV)
At high beam energies the total electron yield (+) is less than unity, so charge is being injected into the specimen, which therefore acquires a negative charge. Negatively charged regions appear bright in the SE image. This is because the negative charge repels secondary electrons from the surface, decreases the landing energy of the incident beam (so raising the SE yield) and increases the collection field between the sample and the detector (so enhancing the detection efficiency). At low energies the total electron yield (+) may becomes greater than unity, so that charge is being removed from the specimen which then acquires a positive charge (since initially it was electrically neutral). Positively charged areas can recollect their own SE emission, and so they appear dark in the SE image. As a special case, if (+) equals unity then there is a dynamic charge balance and no net charging occurs even though the sample is not conducting. This will clearly be the optimum choice of operating energy in most cases. Because of the rise and subsequent fall of SE production with accelerating voltage, there are two incident beam energies at which this condition might be met, a lower value E1 typically between 50 and 150eV and a higher value E2 which is typically between 0.5 and 3 keV. The fact that E2 is in the few keV range of materials of technological importance such as semiconductors, polymers, and ceramics, has been an important factor in the development of low voltage SEM techniques.
Monte Carlo simulation of the electron trajectories in the specimen with respect to the variation in beam interaction volume with energy plotted at a constant scale of magnification shown below:
This confirms that while scanning microscopy at conventional energies (10-30keV) can observe the bulk of a specimen, operation at low beam voltages restricts the interaction to the near-surface regions of the material.
Fig. 4 Images of a 20nm thick carbon film stretched over a copper grid at (a) 20keV and (b) 1keV. The image shows a 200 thick carbon film stretched over a copper mesh grid obtained at 20keV. The film is almost invisible because the majority of the incident electrons traverse the carbon without producing any interaction. But at 1keV, Fig. (b), all of the electron interaction volume is contained within the thickness of the carbon, and the resultant SE and BSE emissions from the film make it appear solid and opaque.
The strategy of lowering the beam energy to improve the lateral spatial resolution by restricting the range of the beam electrons is useful for surface imaging. As incident beam energy decreases and the BSE2 and SE2 ranges approach the BSE1 and SE1 ranges, the image sampling depth is also greatly restricted. Because we are often dealing with an unknown surface chemistry/structure both in depth and laterally, the best approach may be to consider a systematic imaging study of the material as a function of beam energy. By comparing a series of images to search for changes, surface features such as discontinuous surface layers can often be revealed. When the same area is compared during a beam energy image series, care must be taken not to confuse contamination deposited during the earlier stages of the image series with the true specimen information. Minimum dose microscopy principles should be followed, with specimen exposure only occurring during image collection. Since the low-beam energy images are the most sensitive to prior contamination, these images should be recorded first in a beam energy sequence, finishing with the high-beam-energy images.
In Fig. 5.11, the images show a similar array of particles and are complementary, in that bright areas in the conventional E-T image, which show areas of strong scattering, appear dark in the STEM image because these scattered electrons are lost to the transmitted signal. The big difference is that the conventional E-T image in (a) is noisier than the STEM-in-SEM mode image.
For very fine particles, the poor mans STEM-in-SEM detector can produce some outstanding images when operated in conjunction with the through the lens (TTL) detector. Fig. 5.12a shows the SEM TTL and Fig. 5.12b shows the STEM in SEM of fine particles attached to lacey carbon film with a beam energy of 25 keV. The internal structure as well as the surface structure can be simultaneously observed.
Variable-Pressure and Environmental SEM Variable-pressure SEM (VPSEM) or environmental SEM (ESEM) is the instrument that can be operated while its specimen chamber contains a gas or a vapor at a pressure in the range of approximately 200 to 2500Pa (2-20 torr). In this case, the SEM may not offer a secondary electron detector for use when there is gas in the chamber, relying only on BSE detection for imaging. Typically these instruments use a relatively modest electron-optical system and a basic thermionic gun because the performance of the machine is ultimately limited by the electron-gas interaction. VPSEMs and ESEMs are also usually designed to operate optimally in the energy range 10-30keV rather than at the lower energy of 1-5 keV favored by conventional high-vacuum machines.
The difference between the vacuum system of conventional SEM and VPSEM and ESEM is that a pressure-limiting aperture (PLA) or differential aperture is usually positioned at the bottom of the objective lens so that the distance that electrons have to travel through the gas to reach the sample is no more than the working distance. Some instruments, however, mount this aperture on an extension tube beneath the lens so that the gas path length can be much smaller than the working distance of the lens. In either case, the aperture is typically a few hundred micrometers in diameter, a figure which is sufficiently small to allow a pressure differential of 100 to 1000 times to be maintained between the specimen chamber and the column. The isolation of the column from the specimen chamber can be improved by reducing the diameter of the PLA, but this leads to a problem when the instrument is used for low-magnification operation because the scanned rays then interact the edge of the aperture, causing vignetting of the image. Depending on the working distance, the lowest useful magnification is therefore typically 50x or more, unless the instrument is specially modified.
FESEM Images of CdS Nanowires Show the Effect of the Incident Electron Beam Energy (a) (a) (b)
In addition, the probability of an SE, produced at some depth z beneath the surface, escaping from the specimen is proportional to exp(-z/) where is an escape depth and is typically 3 to 6 nm for most elements. With further decrease or increase of the incident energy around the optimum incident energy range (1keV for silver), the yield of SE begins to fall again because of the decrease in the stopping power of the incident electron and the increase in the escape depth, respectively. In general, the yield of SE reaches a maximum value of the order of unity or greater when the electron range RB is about 2.5 . Where is an escape depth.
The variation in the backscattering yield for a low, a medium, and a high atomic number specimen is shown on the next page. Although the backscattering yield is almost constant for energies above about 10keV, it may show significant variations between 1 and 10 keV. The yield rises for a light element such as carbon or silicon, stays approximately constant for a transitional element, such as copper, and falls for heavier elements such as gold. At energies around 1 keV the effective backscattering coefficients of all materials are thus approximately in the range of 0.2 to 0.3. For energies below 1keV, the backscattering yields are not well-known, but significant variations with energy are very likely.
The Relationships Between SE and BSE Images for High Incident Energy and Low Incident Energy
The usual relationships between SE and BSE images may be reversed for low voltage operation. At typical SEM energies (2030keV) the contrast information in the SE image comes from a region about 4 beneath the surface, i.e., about 15 20 nm in extent. By comparison, the average BSE electron is emerging from a depth of about 0.2 RB, i.e., about 0.25m or more. The SE, therefore, carry surface information while the BSE can be said to image the bulk of the sample. At low voltages, however, while the depth of information for the SE is still 1520nm because the energy and distribution of the SE does not change with incident beam energy, the corresponding depth for the backscattered signal of 0.2RB may now only be 3 to 5nm because of the fall in electron range. Thus at low energies SE are emitted from the whole interaction volume while the BSE only emit from a thin surface region and it is the SE which perform the task of bulk imaging.