Preventive Maintenance Suven Life Sciences Limited

Preventive maintenance for HPLC


Instrument ID No._________________

Instrument ID No:____________ Installation Date :____________

Date:____________ Sign:____________

Parameters Before performing preventive maintenance


Value Description Pump seal delivered (In L) Pluger usage (In months) Check valve in & Out Usage (In months) Injector Needle seal Used (in No. of Injections) Needle used (In No. of Injections) Injector Rotor Seal (In No of Ratations) Pressure variation during injector valve switch over (In bar/Kgf) Detector Lamp Used (In hours) Lamp Intensity (In mV/Counts) Back pressure produced by flow cell (In bar/kgf)

PM done by:

Verified by

Page 2 of 5

Instrument ID No:____________ Installation Date :____________

Date:____________ Sign:____________

Preventive Maintenance Schedule Preventive maintance program has been Performed according to the below mentioned procedure. If any abnormalities, deviations found during preventive maintenance, same should be mentioned in Comments section. Preventive maintenance period in mentioned in years. The corresponding number should be rounded according to the Year of preventive maintenance program. Gray colored boxes meant for Not to perform. If period cell is clear then we have to clean and verify the certain parts. If black circle appears then part should be replaced. Test Name 1. Cleaning of entire flow lines.
Replace the Solvents in all the channels with water. Purge the system thoroughly. Replace the column with union with 3ml/min and equal ration for all channel flush for 30 minutes. Repeat this procedure with IPA and again with water.

Preventive Maintenance period 1 2 3 5

Comments

2. Cleaning of Suction filters.


If the filters were glass then soak in 10%Nitic acid overnight & wash with water thouroghly. Other wise sonicate all the suction filters with organic solvent for an hour.

3. Verification of Inline fileter.


With 1ml/Min flow check back pressure developing by inline filter if exceeds 5 bar clean/replace the same

4. Check valve in and Out verification


Open and sonicate the cartridge of inletvalve and total outlet valve in Organic solvent then in water.

5. Plunger seal and plunger cleaning


Open pump head clean the plunger thoroughly with IPA and check for scratches. If appear replace the plungers otherwise clean and reassemble.

6. Cleaning of Needle & Needle seat/seal.


Move the needle up open the needle seat or needle if number of injections less than 1000 clean the both by sonication else, replace the needle seal.

7. Adjusting the needle position.


Inject from blank vial and verify the moments of needle if found any misalignment adjust the same.

8. Cleaning/replacing the high pressure Rotor seal.


Open and clean the High pressure rotor with IPA If found any scratches on Rotor replace the same.

9. Lubrication of pneumatic/Robotic arms.


Clean the arms to remove earlier lubrication and apply low viscocity lubricating oil/gease which is applicable.

10. Flow cell cleaning.


Open the two quartz lenses and clean with IPA & Ethanol thoroughly. If found any scratches on leanses replace the same.

PM done by:

Verified by

Page 3 of 5

Instrument ID No:____________ Installation Date :____________ Test Name 11. D2 lamp replace ment.
Lamp should be replaced when light intensity falls below the predefined limit or after 2 years.

Date:____________ Sign:____________ Preventive Maintenance period 1 2 3 5 Comments

12. Checking hard disk.


Check the hard disk for bad sectors using scan disk command. If any bad sector apprears replace the hard disk.

For Test procedures refer to Maintenance chapter in User manual of respective instrument. For cleaning, adjusting and verification task Use the mark and the parts replaced should be mentioned by the mark Any deviations and explanations is written on . Comments section. Performance verification has to be made after complition of Preventive maintenance and the same should be documented as performance verification report. Mention the performance verification report number and details in below table. PV Done by PV performed date PV document Number. Document location

PM done by:

Verified by

Page 4 of 5

Parameters After performing preventive maintenance


Value Description Pump seal delivered (In L) Pluger usage (In months) Check valve in & Out Usage (In months) Injector Needle seal Used (in No. of Injections) Needle used (In No. of Injections) Injector Rotor Seal (In No of Ratations) Pressure variation during injector valve switch over (In bar/Kgf) Detector Lamp Used (In hours) Lamp Intensity (In mV/Counts) Back pressure produced by flow cell (In bar/kgf)

PM done by:

Verified by

Page 5 of 5