1. Objective
The objectives of this instrumentation laboratory are: To review the operation of instruments in the lab Gain practical experience in using o DC Power Supply o Digital Multi-Meter (DMM) o Function Generator o Oscilloscope
Various voltages and currents are measured for DC and AC circuits. Different AC signals are analyzed with an oscilloscope. Error in measurements and accuracy of equipment are reviewed.
2. Background
Let us quickly review the basic operation of different instruments in the lab. This is simply a repetition of what you learned in NE-241 NE-242 labs.
The dashed lines in figure 1 indicate that all parts are contained within one physical unit. In this case, the resistance is a part of the meter and there is no way to separate it from the meter. An ideal meter has a resistance of 0 or depending upon the type of meter (current or voltage) and the meter model (series or parallel). To measure voltage electronically, a voltmeter uses a known, stable, resistor. All voltage meters draw current, because of that resistor. A typical digital volt meter (DVM) has a resistance of 10 M while some older analog meters may be as low as 3.2 k.
2
Because a current meter has resistance it will cause a voltage drop and may affect the operation of the circuit. The actual amount of resistance depends upon the sensitivity of the meter. For example, the Tektronix 4020 digital multi-meter has a resistance of 2 on the 20 mA range when it is used as an ammeter.
too much. Remember that every time you change the range, the shunt resistance changes. Since the shunt resistance, in a normal meter, can vary from milliohms to kilo-ohms, using the wrong current range can easily affect your circuit. In comparison, a voltmeters resistance will typically only vary from 100 K to 10 M. Digital Meters For all ac and dc voltage ranges, the DMM has a fixed input resistance which is typically 10M. On the current range, the meter will usually have a 200 mV drop at the maximum current, but the resistance is different for each range. Digital meters are usually more accurate, since they remove parallax and other mechanical errors. However, if you do not know how to select the range properly, it is very easy to obtain errors as large as any analog meter. This happens quite often on the current ranges.
A DMM has two sources of systematic error. One part of the error is due to tolerances in resistors. It is expressed as a percentage of the input and therefore varies with the magnitude of the voltage being measured. The second part of the error is due to the digital conversion process for monitoring. It is fixed for a given meter range and is expressed as a percentage of that range. For example, consider a Tektronix 4020 with a reading of 1.34545V on a 2 V DC range. The meter displays a maximum of 1.99999 V on this range. The smallest change on the meter display is 0.00001 V. Then the errors are (based on specifications from Appendix A), percent error of reading: 0.015% * 1.34545 V = 201.82 V percent error of range: 0.003% * 2 V = 60 V Therefore, V = 1.34545 V 0.26182 mV or V = 1.34545 V 0.019%.
Now consider the same measurement on the 200 V range. The reading would have been 1.345V and, percent error of reading: 0.015% * 1.345 V = 201.75 V Percent error of range: 0.003% * 200V = 6 mV. Therefore, V = 1.345 V 6.20175 mV or V = 1.345 V 0.461%. This shows that the most accurate measurements are made on the lowest meter range that will accommodate the measured value. If the meter were to read 16.3245 Vrms while measuring an 8 kHz triangle wave on a 20.00 Vrms scale; how large is the error? The meter measures True RMS and we can trust a reading done on a triangular wave. On the 45Hz 20kHz AC range the error is 0.2% of reading + 0.05% of range. Then the errors are, percent of reading error: 0.2% * 16.3245 V = 32.65 mV (rms) percent of range error: 0.05% * 20V = 10 mV (rms) Therefore V = 16.3245 VRMS 42.65 mVRMS or V = 16.3245 VRMS 0.261%. Table 1 shows the accuracy for various measurements made with the Tecktronix 4020 model. Important note: Use the values under the 1 year column as the default accuracy values. The complete accuracy tables can be found in the Appendix.
Table 2. Crest factor for different waveforms Waveform Square wave Sine wave Triangular wave White noise Crest Factor 1.0 1.414 1.732 3.0 to 4.0
The Tektronix 4020 meters only measure the correct rms value if the Crest Factor is under 3.0. So if, for a sinewave, you measure the RMS voltage and measure the peak voltage on an oscilloscope (that can handle the waveform without introducing more errors), you can see if the meter is reading the RMS value properly. The Crest Factor is more difficult to calculate for rectangular waveforms. A good instrumentation book or instrument manual will explain this. An rms reading meter measures the root of the mean of the square of the signal (proportional to the power) or:
VRMS
1 T 2 v (t )dt T 0
(=
V peak to peak 2 2
The purpose of an rms equivalent value of a waveform, is to simplify power calculations by providing an equivalent DC voltage, that would develop the same average power in a resistive load. If a voltage waveform is not described, but you know that it is equivalent to 10 Vrms, then you will have 100W of average power dissipated in a 1 load (i.e. PAVERAGE = VRMS *VRMS /R). If you did not know the rms voltage, the only way to calculate the power would be to do an integration, from an approximation of the wave-shape as seen on an oscilloscope. This only works for resistive loads because rms value is a magnitude and does not consider phase information. A meter that reads an average rectified value for a waveform of period T, indicates a value
V AVE
1 T v(t ) dt T 0
Therefore meter scales are calibrated to read 1.11 * Vaverage, and since 1.11 * 0.637 = 0.707, the reading is the rms value for a sine-wave and a sine-wave only. Notes: 1) As 0.637 VPK is the average value of full-wave rectified sinusoids only, a rectifying averaging meter is only accurate for sinusoidal waveforms. The multiplier 1.11, known as the form factor, is specific to a pure sinusoidal waveform. Averaging AC meters are therefore, only accurate for sinusoidal waveforms. 2) The above discussion assumes that the meter uses a full-wave rectifier. A second type, called a halfwave rectifier, exists and would read a different voltage because it only responds to the positive parts of a waveform. Because of this, a half-wave rectifier can give very strange measurements if the waveform is not symmetrical about 0V.
7
Table 3 shows how some meters would respond to common signals of 1 V (peak) or 2 V (peak-to-peak). The DC average value will usually be 0V for a sine-wave since sine-wave generators usually have no DC voltage while square or triangular wave usually start from 0V and go to some positive or negative voltage (single ended signal generator).
Table 3. Comparing the reading of different meters for a 2 VP-P signal Waveform 1 VPK (2 Vpeak-to-peak) Square Sine Triangular DC average voltage 1 or 0 0 0.5 or 0 AC rectified average voltage * 1.11 (A) 1.11 0.707 0.555 RMS Volts (B) 1.00 0.707 0.577 % difference (A), (B) +11.0% +0% -3.8%
Power Supply Connections: Most power supplies are floating. This means that neither the positive, nor negative outputs are connected to ground or any common point. This means that you can connect the power supply in any way, even several hundred volts above ground. A non-floating power supply has one lead grounded or connected to a common point. This means that there is only one lead which can be connected anywhere in a
8
circuit since the common or grounded one cannot be moved. These power supplies may be referred to as single ended. The word single means that you can only have a positive or negative voltage from the supply. If the supply is double ended, you can obtain either a positive or negative voltage from it. For the Agilent E3620A power supply, the two outputs are floated and therefore can be connected to each other to build a power supply with positive and negative output voltages. Here is a simplified block diagram of the power supplies in the lab:
Figure 4. Function generator model Signal generators, like DC power supplies, can have different properties. They may have one of the outputs connected to ground or have a circuit common point. This is a single-ended generator as opposed to a floating one, where neither of the outputs is connected to a common or ground. The generator may be unipolar, where the voltage is always of one polarity, or it may be bipolar, where the voltage is both positive and negative at different times. Generally, sine-wave generators are bipolar and square-wave generators are unipolar. Unipolar generators are further broken down into positive and negative. A positive, unipolar, square-wave generator produces squarewaves that start at 0 V, go up to some positive peak, and then return to 0 V in each cycle.
Signal generators have imperfections. The waveforms are not pure. There can be noise or distortion in the output. Noise is a random signal and distortion is a change in or distortion of a pure signal. An ideal sine-wave generator should produce v(t) = sin(t), while a distorted signal could be v(t) = sin(t) + 0.1 * sin(3t) . The frequency or amplitude of a generator output voltage will be stable. The user is expected to keep this in mind. Each generator has an amplitude and frequency control. The controls generally work in ranges so you will usually see a range control that limits the minimum and maximum frequency or amplitude. With time, and as the instrument warms up, the frequency and amplitude of the signal generator may drift slightly from starting points.
2.8 Oscilloscope
Oscilloscope Operation (Y vs. X mode): The main function of an oscilloscope is to measure time varying voltages. The measured voltage is amplified and used to dynamically graph the time-varying electrical signal which is shown as a trace on the oscilloscope display. This trace can be measured with the aid of calibrated lines on the display. There are three main controls for this process: horizontal, vertical, and trigger. The vertical position of the trace is proportional to the voltage measured at the input, and its exact position is determined by the vertical controls which determine the amplitude per division setting. There are typically 8 vertical divisions on a display. The amplitude per division setting determines the maximum voltage swing that can be displayed as well as the volts per division seen. The input signal can be AC or DC coupled. AC coupling involves adding a series capacitor. This has the effect of blocking (removing) the dc bias and low frequency components of a signal. DC coupling does not have this capacitor and therefore allows measurements down to 0 Hz. AC coupling is useful when you are trying to measure a small AC voltage that is on-top of a large dc voltage. A typical example is trying to measure the noise of a DC power supply. The horizontal position of the trace is controlled by the trigger and horizontal controls. When activated the trigger circuitry causes the trace to sweep from the left of the screen to the right. When the trace reaches the right of the display it will automatically reset so the trace will start rendering at the left side of the display the next time its triggered. The time it takes to traverse the display is controlled by the horizontal controls. Here the user can change the time per division settings. There are typically 10 horizontal divisions on the display. The time per division setting determines the total time shown on the display was well as the time per division. The trigger controls determines when the trace will start its sweep across the display. This is controlled by a reference voltage level (trigger level) which is set by the user. Once in input voltage reaches this trigger level the trigger circuitry is activated and the trace will start to sweep across the display. Note: Once the trigger is activated it cannot be retriggered until the trace has finished its sweep, even if the input voltage crosses the trigger level during the sweep. At any time the position of the trace is determined by the combination of the voltage (vertical) and time (horizontal) conditions.
10
To avoid a lengthy document, no more info is given on oscilloscope here. If you need more detailed review on oscilloscope, please go to your lab resources for NE-241.
3. Pre Lab
Question 1. For the dual-trace oscilloscope display as shown in the figure 9, calculate: a) b) c) The RMS value of the sinusoidal waveforms displayed. The frequency of the waveforms. The phase angle between the two waveforms.
11
Question 3. What would the accuracy of the 5 digit DMM be if it measured 1.3454V (DC) on the 20V range?
12
4. In Lab Procedure
Since DMM and oscilloscope are the main measurement devices in the lab, we will review these items in detail in following sections.
Oscilloscope
The front panel of Agilent DSO-X 2012A oscilloscope is shown in figure 11.
To review the operation of each section, we will examine each button/knob individually for each section. Put the Agilent function generator to 1 kHz sine wave, 4 V peak-to-peak with no offset and connect it to a 50 load (use the decade box) and to channel-1, see Figure 11.5.
13
Oscilloscope 50 channel-1
Figure 11.5 Circuit for 1kHz sine wave, 4 Vpp connected to 50 load
Press the Autoscale button. You should have gotten a screen similar to figure 12.
Please note that you are NOT allowed to use this Autoscale button during all sessions of the NE-344 lab! This one time use is only to have consistent result at the beginning. In the following sections, examine the function of each button or knob on the graph you see on your oscilloscopes display. Make sure to undo any changes you apply before going to next sub-section.
14
Vertical inputs/controls Measure the peak-to-peak amplitude of the signal using the oscilloscope. Change the Volts/div button and read the value on the display. Fill out the table 4.
So, as you see, changing the Volts/div does only change the amplifier/attenuator ratio, and in any case we read 4 V peak-to-peak for the incoming signal. Rotate the vertical position knob. As you see it moves the signal up or down, and the amount you move is written on display during the move. Press the button 1 several times. As you see, it toggles between on and off state for the channel-1 signal. Once you press this button, you also get the menu for channel-1. In this menu, you have these controls: Coupling (AC, DC, or Ground). AC eliminates the DC component of the incoming signal. In other words, shows only AC part of the signal. DC shows both AC and DC parts. Ground shows a ground signal (as if the input is grounded). BW Limit. Limits the band-width of the oscilloscope to 20 or 25 MHz. Signals with higher frequencies cannot be shown on the display. Vernier. Allows having smaller steps between different Volts/div values. Invert. Show the signal as if it was inverted. Probe. Allows using probes with different attenuation ratios. Make sure it is set to 1:1.
To understand the function of Coupling, give some positive or negative (like +1 V or -1 V) offset to the signal using the function generator, and at the same time look at the oscilloscope while using DC coupling and AC coupling separately. Are you able to measure the offset added to the signal? Put the Volts/div back to 1 V, and bring back the vertical position to zero. Horizontal controls Measure the period of the signal using the oscilloscope. Change the Time/div knob and fill out the table 5. Table 5. Changing Time/div knob # of horizontal divisions Time/div for one period 200 S 5 100 S 500 S Time/div x # of horizontal divisions (period of the signal) 200 S x 5 = 1000 S = 1 ms
15
Horizontal position knob controls the position of time-reference for the signal. Once you rotate this knob, pay attention to the small triangle at the top of screen display. Once you press the Main button, you get the Horizontal menu. In this menu, you have these controls: Main. This is the main mode of operation for an oscilloscope. It is the mode in which oscilloscope is showing the voltage versus time for both channels. Delayed. Not covered in this lab exercise. X-Y. In this mode, the time dependence of incoming signals will not be shown! Instead, channel-2 signal (Y) will be shown versus channel-1 (X) signal. So, both axes will have voltage scales. This mode is used for example when you want to see the I-V characteristic of a diode. Roll. Used for low frequency inputs. It shows the changes to input in real time. Put the frequency of function generator to 2 Hz to see the effect of this button. Vernier. Allows having smaller steps between different Time/div values. Time Reference (Left, Centre, and Right). Changes the reference point for time axis.
To examine the Roll function, put the oscilloscope to this mode and change the frequency of the function generator to 1 Hz. Your signal should be rolling on the display. To have a better understanding of the Roll/Main/X-Y modes, wire the circuit in figure 13 with a 10 k resistor and a 10 nF capacitor. Put the function generator on 1 kHz sine wave, 1 V P-P with no offset. Without going through circuit theory to justify circuits behavior, briefly describe what you see on oscilloscopes display in Main and X-Y mode. Put the frequency to 1 Hz, and compare the three modes of Roll, Main, and X-Y. In 1 Hz, the capacitor works as an open circuit (like if there is no capacitor in the circuit at all), and therefore the output is the same as input.
Figure 13. An RC circuit Put the Time/div back to 200 S, and bring back the horizontal position to centre.
16
Run controls There are only two buttons in this section. Run/Stop. You can freeze the oscilloscopes operation and bring it to a stopped mode, once this button is pressed and illuminated red. If you press the button again, it goes back to run mode again. Single. Not covered in this lab exercise.
One application for this Stop button is when you have a continuously changing signal, and you want to study one pattern for example. Once the oscilloscope is in Stop mode, change something on the function generator (like frequency or amplitude of the signal). Do you see it on the oscilloscopes display?
Trigger controls This is the most important section in an oscilloscope in order to capture a signal effectively on the screen display. Level. This knob is simply controlling the trigger level. If this level is within the range of maximum to minimum of the incoming signal, the oscilloscope is able to trigger and capture the incoming signal. Otherwise, a stationary signal cannot be achieved on display. Edge (Rising and Falling). It determines whether the rising edge or falling edge of the signal should be shown right at the reference point of time. In the Edge Trigger Menu, you can also select the source of trigger signal to be channel-1, channel-2, or the External trigger input. Mode/Coupling. By pressing this button, you get the Trigger Mode and Coupling Menu using the softkeys. Here is the function of button: (no description is required for your report regarding the Mode/Coupling section) o Mode (Auto, Auto-Level, and Normal). For simplicity, use the Auto or Auto-level modes in this lab. o Coupling (DC, AC, LF-Reject). Similar to Coupling in Vertical control section, determines the coupling applied for the signal used for trigger. If the incoming signal has a large DC component with a weak AC part mounted on it, it may be helpful if AC coupling is used. In most cases, DC coupling is fine for us. LF-Reject (low frequency reject) is used when strong low-frequency, like 60 Hz, signals present in the incoming signal, and make it difficult for the trigger system to do its function. o Noise-Reject. Is used when the incoming signal is noisy. o HF-Reject. Is used when the incoming signal has unwanted high-frequency signals added to it. o Holdoff. Not covered in this lab exercise. o External. Used for the external trigger input only. Pulse-Width. Not covered in this lab exercise. Pattern. Not covered in this lab exercise. More. Not covered in this lab exercise.
Try Level knob and Rising/Falling edges to see how they affect the signal on display. To try the source of trigger in the Edge Trigger Menu, we need to have another signal source to oscilloscope. Connect the BK Precision function generator to channel-2. Using the Coarse/Fine frequency control knobs try to keep the frequency to 1000 2 Hz. Put the amplitude to minimum and waveform to triangular. Make sure only three buttons are pushed in: Power (red), 1k for frequency, and triangular waveform (all other buttons should not be pushed in).
17
Put the channel-2 to 1 V/div. Now if you toggle the source of trigger between channel-1 and channel-2, you will see only the signal used for trigger stationary, and the other one rolling slowly because its frequency is slightly off compared to the other function generator. Waveform There are two buttons in this section. - Acquire. Controls the way signal is acquired to the oscilloscope. It has the following options using softkeys: o Normal. This is the mode usually used for most applications. o Peak Detect. Not covered in this lab exercise. o Averaging. Used to reduce random noise associated with the signal to create a smoother waveform on the display. The depth of the averaging function can be controlled by the control knob. o Realtime. Not covered in this lab exercise. - Display. Controls the way signal is displayed. It has the following options: o Persistent. Gives memory to the screen display pixels. In other words, keeps a faint track of the previously shown signals. Change the frequency/amplitude of the function generator to see the effect of this button. o Clear Display. Clears patterns created by the persistence mode. o Grid. Controls the intensity of the grid shown on the display. o Vectors. Not covered in this lab exercise. Try the Averaging, Persistence, and Clear Display modes on the oscilloscope. Measure This section has two buttons. Cursors. Creates horizontal and vertical cursors used for measurement. Here are the options using the softkeys: o Mode (Normal, Binary, and Hex). For analog measurements, we only use Normal mode. o Source (1, 2, and Math). Selects the channel to work on. o X & Y. Gives the option to move/select X1, X2, Y1, and Y2 axes. Quick measurement. Using this tool one can do many different automated measurements using the internal functions of the oscilloscope. Here are the options available using softkeys: o Source (1, 2, and Math). Selects the channel to work on. o Clear measurement. Deletes posted numbers on the display. o Functions. Such as Frequency, Period, Peak-to-peak, Maximum, Minimum, Rise time, Fall time, Duty cycle, RMS, + Width, - Width, Average, Amplitude, Top, Base, Overshoot, Pre-shoot, X at max.
It should be noted that the Quick Measurement function performs the task for that part of the signal which is seen on the oscilloscopes screen display. Measure the frequency and peak-to-peak amplitude of the voltage signal provided by the function generator. Do these values match with each other and also with what you read on function generator?
18
Digital-Multi-Meter (DMM)
A DMM can be used to measure resistance (), DC voltage, DC current, AC voltage, and AC current. Let us review DC measurement separate from AC measurement. DC measurement Wire the circuit shown in figure 14 on the breadboard. Put the DC power supply on 10 V.
Fill out the table 6. Use 20 V range for voltmeter and 20 mA range for ammeter. Remember, Voltmeter across the component, and Ammeter in series with the component under measurement!
Table 6. DC measurement errors Item V1 V2 I1 I2 I3 AC measurement Replace the DC power supply in the above circuit with the Agilent function generator. But to keep the AC circuit identical with DC circuit, put another 100 resistor parallel with the 100 resistor! Why? (Remember the function generator has an internal resistor of 50 ) Put the function generator to sine wave, 100 Hz, 10 Vp-p with no offset. Fill out the table 7. Use 20 V range for voltmeter and 20 mA range for ammeter. Remember, the Tektronix DMMs in the lab are true RMS measuring equipment! Note that the voltage you measure across the 50 resistor is function generator is in this case, and voltage across the Calculated value (A) Measured value (B) Error =
19
Table 7. AC measurement errors Item V1 V2 I1 I2 I3 Calculated value (A) Measured value (B) Error =
20
21
22