Documente Academic
Documente Profesional
Documente Cultură
The
Abuse
Thomas
of Radiographic
T. Thompson, M.D.
Tubes
Introduction
With spective diagnosis there pitals
that
Causes
this fall of probased groups pressure and (DRGs), on hosit is likely will share
of Radiographic Failure
failure always and simply of heat tube by of an related x-ray
Tube
ufacturers now recognize major problem and place is scribing on the nately proper generator many warmup console; technologists
this decals
as a de-
on almost
tube pro-
will
radiology
be great costs;
life can
recommended
departments
generated tubes
(2)
in their expenses.
play,
are
damaged a cold
the fila-
Filament
Filament major (mA) filament,
Evaporation
evaporation is another as current x-ray tube to a before (In some
of power is
(1) overheating
heating
an unduly
extended
pen-
the resources available to to extend the useful life of tubes. 95 percent to the damaged error. The
include: (2) anode
is heated
returned been
errors
of
Overheating
Warming important useful erated target step
a Cold
up an x-ray in
Anode
tube is an the
on the
filament
between
ex-
of comprolonging
a standby In this
erosion,
is boosted In any
and (4) burned out or For the most part, are controllacontrol, could signifithe useful life of the
life of the tube. If heat is genin a cold target too quickly, the will not expand
it
of damage
uniformly,
will crack.
and
This
continues
across
a high
between
voltage
the
to taking and
a Such
pottery placing
dish, if
cathode
celerates
and
the
anode.
electrons
voltage
the
acgap,
across
but
causes generating
makes
them
the
in temperature Dr. Thompson is Professor of Radiology at the University of Miami, and Director of Radiology of the University of Miami Hospital and Clinics.
is very
Heating
filament duces
over what
a prolonged is known as
pro-
filament
on a cold
Volume
5, Number
September
1985
RadioGraphics
397
Thompson
Radiographic
tubes
Eventualwill arcing
tage
exceeds
the kilowatts
allowed
for
circuit
times faster than standard thus, provides for more pation of heat), the wear x-ray tube experienced leads life. Control may matic. whenever be manual High of a high or, more speed exposure to
speed (and, rapid dissifactor on the that This tube rotor autois used exceed
between the cathode, lope and the anode. bution from sures heated to filament
the design of the generator against instantaneous overis more of the likely tube, to result,
Damage
delaying or withholding expoafter the filament has been and the rotor has been brought
however,
from
primarily lumbosacral
exceeding
the
capacity
up to speed.
Most filament
evapora-
in two types of spine studies and only way to protect the heat capacity of a lumbosacral the in-service spine edu-
rotation factors
tion problems can through the in-service the cases, place control rotation control. switches seconds) is activated is made. This equipment
operators.
certain limits; in the manual mode, conversion to high speed rotation is made by activating a button or toggle switch. Again, unfortunately, the rotor control is often left on high speed, even when it is unnecessary; this eventually the tube.
is used, it
of filament heating and anode independent of exposure Single provide between delay position a delay the time provides exposure (about 0.8 the control the exposure time for
cation of technologists, assuming exposure techniques are satisfactory. There are two ways to protect against exceeding the heat capacity of the tube in the course of a tomographic examination. One, if two technologists work together to perform tomography, be made If a single entire longer the exposures faster than technologist will invariably the tube can cool. the
leads
should
to destruction
of
If a high
prevent
this occurrence.
The
control
issue
and
of high
bearing
speed
deterioration
rotor
the filament to be heated and for the rotor to come up to speed, but it is impossible with such a switch to voluntanily extend heating. Filament more stations rapidly are used. the time evaporation if high Another of filament occurs disadvan-
performs
raises another problem. The operation of an x-ray tube at standard anode rotation speed is essentially vibrationless. speed from
frequencies
milliamperage
tage to the constant use of high mA stations is that the focal spot enlarges, thereby degrading spatial (1). resolution in the radiograph
serves to protect the tube. Two, a time delay switch may be installed on the generator that will prevent exposures from being made too rapidly to permit adequate anode cooling.
there
the for
periods
of the tube. be quickly ac(very difficult anodes) and high speed to friction pro-
of High
Speed
Rotation
celerated
to high
with The question rotation of the x-ray perplexing letter in Graphics of high speed tube anode is a Grays Radioquickly standard duced greater current higher problem, tube rently erated pable
30,000
large
brought speed.
heat
he stated
that
the Mayo
by the bearings at speeds than 9-10,000 rpm prevents tubes from being operated at speeds. Toshiba without in field To overcome has developed this a
to them. of exratings,
Clinic was ordering high speed tubes for single phase equipment. This offers some in shortening operation, proximately (rpm);
Whereas,
advantages tube
but
may
result tube
life. In normal
bearings that is curtrials. This tube is opflux and is caapproximately clinical may have trial merit is
at high high
9-10,000 rotation
successful,
398
RadioGraphics
September
1985
Volume
5, Number
Radiographic
tubes
Thompson
in CT
scanning, forms
digital
radiography sequence
of rapid
may one
exposures portant
will
result.
This
is an im-
area
CT tubes.
termine from the manufacturer what tubes are available for the equipment in question. hanger particular
100
Choice
X-ray various size, and a tube lecting heat target tubes angle, capacity.
of Tubes
are available of focal target The useful with spot life of
without increasing the area of the focal spot as it is projected on the plane of the film. As the target angle steepens, however, the area that can be covered by the beam decreases.
In some tube.
cases,
the
tube a
combinations
Hence,
target
steep
film focal
diameter,
coverage
receptor spots. can
at a standard
source-image
do large
target spot
angle is adequate
applications.
A small
The heat capacity also be increased of the tube by increasing
investment
in an exposure
for a particular
The
facturers
National
Electrical
XR5-1974) tolerances
Association
the
target
diameter
diameter
of the
over
target.
As the
the heat area that is incan
counter is money well spent, and the next time a tube is replaced, insist on a tube report from the tube manufacturer.
distributed creasing be
specifies Focal
how
the amount
spots cannot
the nominal
Larger take
speed
Summary To avoid
placement, the specific select
focal spots of less than 0.8 millimeters, the tolerance is plus 50 percent; of 0.8 through of greater cent. spot 1.5 mm, than actually smaller plus 40 percent; 1.5 mm, measure plus and 30 per0.9 spot, mm
unnecessary
a tube
tube
designed at hand,
refor and
than do smaller targets; and this may become a limiting factor in some radiographic applications.
application
Hence,
may The
a nominal
are all kinds of devices on designed to detect or calof heat in the tube, designed of heat the heat the tube capaciat 70 life tube
observe meticu-
and
dards.
acceptable
the focal
the amount
better token,
smaller the
resolution
in the
and there are other devices to facilitate the dissipation from sensors operate ty,
can
lously the warmup routine prescribed by the manufacturer-never make a diagnostic exposure with a cold tube;
(2) do not delay
once and
the the
radiographic
the tube.
In many
has
the smaller
(5) consult
target,
but
been brought up to speed; the tube heat rating chart extended study and take care the specified speed limits anode and rotause high
kilowatt
rating
of the x-ray
tube.
To
percent
capacity,
not to exceed
( 4) never
tion
unnecessarily.
References
1. Thompson U. Selecting medical x-ray film. Applied Radiology 1974; 3:47-50.
Gray JE. Letter to the Editor. Ra-
2.
dioGraphics
1983; 3(2):324.
Volume
5, Number
September
1985
RadioGraphics
399