Sunteți pe pagina 1din 2

1 GHz Spectrum Analyzer HM5510

HM5510
Unmodulated RF Signal

R requency Range 150 kHz1 GHz F R Amplitude Measurement Range -100+10 dBm R hase synchronous, Direct Digital Frequency Synthesis (DDS) P R Resolution Bandwidths (RBW): 20 kHz and 500 kHz
Amplitude-modulated RF Signal

R Keypad for Frequency and Amplitude Setting R Analog Signal Processing and Display R Test Signal Output

All data valid at 23 C after 30 minutes warm-up.

1 GHz Spectrum Analyzer HM5510

Frequency Characteristics 0.15 MHz1.05 GHz Frequency Range: 5 ppm Stability: 1 ppm/year Aging: 1 kHz (6-digit in readout) Frequency Resolution: 01.05 GHz Center Frequency Range: TCXO with DDS LO Frequency Generation: (Digital Frequency Synthesis) Zero-Span and 11000 MHz Span Setting Range: (125 Sequence) Marker: Frequency Resolution: 1 kHz, 6-digit, Amplitude Resolution: 0.5 dB, 3-digit Resolution Bandwidths 500 kHz and 20 kHz (RBW) @ 3 dB: 4 kHz Video filter (VBW): 20 ms Sweep Time: Amplitude Characteristics (Marker Related) 150 kHz1 GHz -100+10 dBm Measurement Range: 10 dB/div. Scaling: 80 dB (10 dB/div.) Display Range: Amplitude Frequency Response (at 10 dB Attn., Zero Span and RBW 500 kHz, Signal -20 dBm): 3 dB 8 x 10 division Display (CRT): logarithmic Amplitude Scale: dBm Display Units: 2 Lines x 20 Characters, Parameter Display (LCD): Center Frequency, Span, Marker Frequency, Reference Level, Marker Level 040 dB (10 dB increments) Input Attenuator Range: Tolerance of input attenuator: 2 dB relative to 10 dB position Max. Input Level (continuous) 1040 dB attenuation: +20 dBm (0.1 W) 0 dB attenuation: +10 dBm 25 V Max. DC Voltage: -50+10 dBm Max. Reference Level: Reference Level Accuracy rel. to 500 MHz, 10 dB Attn., Zero Span and RBW 500 kHz: 2 dB approx. -100 dBm (RBW 20 kHz) Min. Average Noise Level: typical >75 dBc (2 Signals: 200 MHz, Intermodulation Ratio 203 MHz, -3 dB below Reference Level) (3rd Order): Harmonic Distortion Ratio typical >75 dBc (200 MHz, Reference Level) (2nd harm.): Bandwidth Dependent Amplitude Error rel. to RBW 500 kHz and Zero Span: 1 dB

Inputs/Outputs Measurement Input: Input Impedance: VSWR: (Attn. 10 dB) Supply Voltage for Probes (HZ530): Audio output (phone): Test Signal output: Frequency: Level Functions Keyboard Input: Rotary Encoder Input:

N-socket 50 typ. 1.5:1 6 Vdc 3.5 mm jack N-socket, output Impedance 50 10 MHz 0 dBm (3 dB) Center Frequency, Reference Level, Center Frequency, Reference Level, Testsignal output Level, Marker, Intensity (CRT), Contrast (LCD) D14-363GY, 8 x 10 div. with internal graticule approx. 2 kV adjustable on front panel 105253 V, 50/60 Hz 10 %, CAT II approx. 31 W at 230 V/50 Hz Safety class I (EN61010-1) +5+40 C -20+70 C 580 % (non condensing) 285 x 125 x 380 mm, with adjustable, lockable tilt handle approx. 5.6 kg

General information CRT: Acceleration Voltage: Trace Rotation: Power Supply: Power Consumption: Safety class: Operating temperature: Storage temperature: Rel. humidity: Dimensions (W x H x D): Weight:

Accessories supplied: Line Cord, Operators Manual, 2x HZ21 Adapter Plug (N-plug with BNC socket) Recommended accessories: HZ20 Adapter, BNC to 4 mm banana HZ33 Test cable 50 , BNC/BNC, 0.5 m HZ34 Test cable 50 , BNC/BNC, 1 m HZ43 19-Rackmount Kit 3RU HZ520 Antenna HZ525 50 -Termination, N plug HZ530 Near Field Probe Set for EMI Diagnosis HZ560 Transient Limiter HZ575 Converter 75 to 50 HZO30 Active probe 1 GHz (0.9 pF, 1 M, including many accessories)

HM5510E/121010 C&E Subject to change without notice HAMEG Instruments GmbH DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08

HAMEG Instruments GmbH Industriestr. 6 D-63533 Mainhausen Tel +49 (0) 6182 8000 Fax +49 (0) 6182 800100 www.hameg.com info@hameg.com

S-ar putea să vă placă și