WBC
Defend the body against infectious disease and foreign materials
Granulocyte
Blood
Blood Cell
Leukocyte
Erythrocyte
Platelet
Lymph
Baso
Mono
Eosino Neutro
Lymphocyte
Mid-sized Cell
Neutrophilic Granulocyte
Blood
Blood Cell
Leukocyte
Erythrocyte
Platelet
Blood
Blood Cell
Leukocyte
Erythrocyte
Platelet
Coagulation or clotting
Reagents
WBC, RBC and PLT are counted and sorted by the Electrical Impedance Method, just Coulter Principle. This method is based on the measurement of changes in electrical impedance produced by a particle passing through an aperture.
1.2 Principle
Electrical Impedance Method
Reagents
1.2 Principle
Electrical Impedance Method
Reagents
The amplitude of each pulse is proportional to the size of each blood cell.
Version: 1.3, Update Date: 2008-2-1
Reagents
1.2 Principle
Electrical Impedance Method
Reagents
Reagents
1.2 Principle
Electrical Impedance Method
Reagents
Volumetric Metering
Colorimetric Method
Reagents
This analyzer uses a volumetric metering unit to control the count cycle and to ensure that a precise volume of sample is analyzed.
Upper sensor Upper sensor
Lower sensor
Lower sensor
Upper sensor
Upper sensor
Lower sensor
Lower sensor
3 Counting start when the liquid1.3, surface pass the upper Version: Update Date: 2008-2-1 sensor.
4 Counting end when the liquid surface pass the lower sensor
Reagents
HGB is determined by the colorimetric method. Lyse will release HGB when RBC was broken down and react with HGB to generate a mixture for HGB measurement. The mixture has the biggest absorbance at 540nm.
I0
It
Reagents
Light source
Absorbing container
Detector
LED light
Bath
Photo sensor
Reagents
Reagents are one part of the system. Lyse Break down RBC; Release HGB and react with HGB to generate a mixture for HGB measurement; Help the differential of WBC. Diluent Dilute the blood sample; Provide a conductive environment for counting and an environment which is similar to the plasma to maintain the size of blood cell; Clean the tube and bath.
Reagents
Rinse Clean the bath and metering tube. E-Z cleanser An enzyme-based cleaning solution used to clean the tube and bath.
Probe cleanser An alkaline cleaning solution used to clean the tube and bath.
Fluctuating
Discriminator 1 30~85fL
WBC
2, 3 85~125fL
Fixed Discriminator
4 400fL
[fL]
Lymph
Mid
(BasoMonoEosino)
Neutro
No differential result
No differential result
No differential result because WBC count result is less than a certain value (WBC < 0.5)
No differential result
No differential result because the peak of WBC histogram lies in the middle of the histogram and thus cannot identify the type of peak cells
Version: 1.3, Update Date: 2008-2-1