Documente Academic
Documente Profesional
Documente Cultură
IAEA
International Atomic Energy Agency
Introduction
Dental radiology makes use of specific types
of equipment, needed for different purposes. Frequent exposures though each with low dose involve a risk for the practitioner and for the patient Background: general principles of x-ray diagnostic imaging
IAEA
Topics
IAEA
Overview
IAEA
IAEA
International Atomic Energy Agency
Types of units
Intra-Oral units
Standard dental tube uses an intra-oral image receptor has extra-oral x-ray tube Panoramic (OPG) Cephalometric ( Ceph)
IAEA
IAEA
IAEA
IAEA
IAEA
10
X-Ray Tube
stationary Anode avoid overheating tube duty cycle:
typical: 1:30 intaroral 1:10 OPG 420 mAs/hr intraoral
IAEA
11
Tube Head
IAEA
12
Generator Circuit
IAEA
13
IAEA
14
Collimator
1. Lead Collimator with central hole
2. Spacer Tube
IAEA
15
Applicator Cones
Good
IAEA
Bad
22: Optimization of Protection in Dental Radiology
Bad
16
Cephalometric Holder
IAEA
17
from 60 ms to 2.5 s
Min. 50 kV, ~7mA 1 mm
Inherent filtration
Focus-skin distance Irradiated field
~2 mm Al equivalent
20 cm 28 cm2 with round section, 6 cm diameter collimator
IAEA
18
IAEA
19
IAEA
International Atomic Energy Agency
Keep under control time and temperature of the developing process. Do not use oxydized chemicals
RP in Dental Radiology
Technical hints to reduce patient doses Privilege the use of sensitive films Sensitivity class D Very good spatial resolution Typical delivered dose: about 0.5 mGy Typical exposure times: 0.3 - 0.7 s Sensitivity class E Good spatial resolution Typical delivered dose: about 0.25 mGy Typical exposure times: 0.1 - 0.3 s
IAEA
22: Optimization of Protection in Dental Radiology
25
IAEA
26
Panoramic examination
Image quality not as good as in intra-oral
films
IAEA
International Atomic Energy Agency
Why Dental QC ?
Widespread use of dental units Lack of QC history on most units Dental practitioners working in the primary
health care sector do not have the continuous medical physics support available in a hospital-based diagnostic imaging department
IAEA
22: Optimization of Protection in Dental Radiology
29
What Tests ?
Collimation
Dose Evaluation
Kilovoltage (kVp)
Leakage Radiation
Exposure Time
Half Value Layer
IAEA
30
31
IAEA
32
Beam size/collimation <60 mm diameter 3 yearly (intra-oral) <150 x 10 mm at cassette (panoramic) Dose at cone tip 50 kV: <5.0 mGy 70 kV: <2.5 mGy (E speed film) <75 mGy mm 1-3 yearly
1-3 yearly
IAEA
33
Dental QC Methods
Unit
Intra-Oral Receptors (I/O) Test Method
as for Radiology QC
as for Radiology QC
34
Test Equipment
Dosimeter
small & large volume
chambers 2mm wide detector for OPG
kVp meter
measure kVp
average Programmable delay ~100 ms Range: 50 to 120 kV
Timer
triggering at 75%
peak kV
Aluminum filters
4 x 1mm Grade 1100
IAEA
35
Collimation
Expose film or fluorescent screen Measure x-ray field image
Unit Film position Limits of X Ray field
Standard
OPG
Ceph
IAEA
IAEA
37
IAEA
38
Kilovoltage
Accuracy: Set kVp meter to ~100 msec delay Observe kVp waveform at 70 kV if poss. Limit: measured kVp within 5% of set value Reproducibility: Take 5 repeat exposures Limit: coefficient of variation 2%
IAEA
39
Dose Evaluation
Skin dose from I/O units: place cone 10 mm from dosimeter set maxillary molar/ bitewing setting
Should be (65-70 kVp):
2-3 mGy for molar view < 5 mGy for any view
IAEA
40
IAEA
41
IAEA
42
IAEA
43
Timer Reproducibility
Standard I/O units & Ceph units
IAEA
44
HVL Set-up
IAEA
46
IAEA
48
60 70 80 90
IAEA
49
Leakage Radiation
Cover collimator with 2 mm lead Set large ionization chamber (~180 cc)
or solid state detector at known distance d cm from focus Expose on maximum kVp for 0.5 - 1 sec (but be careful!) Scale up the reading to 1 hour
IAEA
50
Leakage Radiation
51
Leakage Measurement
In normal radiography, leakage
measurements are normalised to the maximum continuous rated tube current R:
Leakage = Dose (for P mAs at 100 cm) x (R x 3600/P)
Leakage Measurement
IAEA
53
Calculation of Leakage
Measured leakage at 100 cm = 2mGy/hr Duty cycle = 1: 20 (i.e. must wait ~ 20 times
the exposure time between exposures to allow tube cooling) Thus actual leakage = 2 / 20 mGy/hr = 0.1 mGy/hr i.e. acceptable
IAEA
54
Dental Phantom
IAEA
56
Dental Phantom
IAEA
57
Summary
General and specific types of x-ray equipment
(and image receptors) are reviewed: (panoramic and cephalometric), with technical data about operating conditions