Sunteți pe pagina 1din 59

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY


L 22: Optimization of Protection in Dental Radiology

IAEA
International Atomic Energy Agency

Introduction
Dental radiology makes use of specific types
of equipment, needed for different purposes. Frequent exposures though each with low dose involve a risk for the practitioner and for the patient Background: general principles of x-ray diagnostic imaging

IAEA

22: Optimization of Protection in Dental Radiology

Topics

Dental X-ray equipment Radiation protection in dental


radiology

Quality control for dental


equipment

IAEA

22: Optimization of Protection in Dental Radiology

Overview

To be able to apply the principle of


radiation protection to dental radiology system including design and Quality Control.

IAEA

22: Optimization of Protection in Dental Radiology

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Part 22: Optimization of protection in dental radiology


Topic 1: Dental x-ray equipment

IAEA
International Atomic Energy Agency

Types of units
Intra-Oral units
Standard dental tube uses an intra-oral image receptor has extra-oral x-ray tube Panoramic (OPG) Cephalometric ( Ceph)

IAEA

22: Optimization of Protection in Dental Radiology

Intra-Oral Dental X-Ray Equipment

IAEA

22: Optimization of Protection in Dental Radiology

Modern Dental X-Ray Unit

IAEA

22: Optimization of Protection in Dental Radiology

Panoramic X-Ray Equipment

IAEA

22: Optimization of Protection in Dental Radiology

Cephalometric X-Ray Equipment

IAEA

22: Optimization of Protection in Dental Radiology

10

X-Ray Tube
stationary Anode avoid overheating tube duty cycle:
typical: 1:30 intaroral 1:10 OPG 420 mAs/hr intraoral

IAEA

22: Optimization of Protection in Dental Radiology

11

Tube Head

IAEA

22: Optimization of Protection in Dental Radiology

12

Generator Circuit

IAEA

22: Optimization of Protection in Dental Radiology

13

Generators & Pre-Heat


Medium frequency - stable waveform Single phase (SP) - pulsed Pre-Heat: separate circuit for heating
filament Single Phase units without a pre-heat circuit initial pulses of variable kV

IAEA

22: Optimization of Protection in Dental Radiology

14

Collimator
1. Lead Collimator with central hole

2. Spacer Tube

IAEA

22: Optimization of Protection in Dental Radiology

15

Applicator Cones

Good
IAEA

Bad
22: Optimization of Protection in Dental Radiology

Bad
16

Cephalometric Holder

IAEA

22: Optimization of Protection in Dental Radiology

17

Intra-Oral Dental X-Ray Equipment (technical data)


Exposure time
Tube Focal spot size

from 60 ms to 2.5 s
Min. 50 kV, ~7mA 1 mm

Inherent filtration
Focus-skin distance Irradiated field

~2 mm Al equivalent
20 cm 28 cm2 with round section, 6 cm diameter collimator

IAEA

22: Optimization of Protection in Dental Radiology

18

Panoramic X-Ray Equipment (technical data)

Focal spot kV mA Exposure time

0.5 mm 60 - 80 kV in 2 kV steps 4 - 10 mA steps 4, 5, 6, 8, 10 12 s (standard projections) 0.16 - 3.2 s (cephalometric projections)

Flat panoramic cassette 15x30 cm (Lanex Regular screens))

IAEA

22: Optimization of Protection in Dental Radiology

19

Image Receptors in Dental Radiology


Intraoral Radiology Small films (2 x 3 or 3 x 4 cm) in light-tight envelopes (no screen) Digital intraoral sensors - compared with category E film, the radiation dose is reduced by 60%. Panoramic Radiology and Cephalometry Film-screen combination Digital sensors - compared with filmscreen sensitivity class 200, the radiation dose is reduced by 50-70%.
IAEA
22: Optimization of Protection in Dental Radiology
20

Dental Radiology Film Types


Sensitivity class D Very good spatial resolution Typical delivered dose: about 0.5 mGy Typical exposure times: 0.3 - 0.7 s Sensitivity class E Good spatial resolution Typical delivered dose: about 0.25 mGy Typical exposure times: 0.1 - 0.3 s
IAEA
22: Optimization of Protection in Dental Radiology
21

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Part 22: Optimization of Protection in Dental Radiology


Topic 2: Radiation Protection in Dental Radiology

IAEA
International Atomic Energy Agency

Radiation Protection in Dental Radiology


Facts
Very frequent examination (about 25% of all the radiological examinations) Delivered doses may differ of a factor 2 or 3. (entrance doses between 0.5 and 150 mGy)

Image Quality often very low


Organs at risk: parathyroid, thyroid, larynx, parotid glands
IAEA
22: Optimization of Protection in Dental Radiology
23

Radiation Protection in Dental Radiology


Technical hints to reduce patient doses Quality Control of Film Processing

Keep under control time and temperature of the developing process. Do not use oxydized chemicals

Do not adjust development time by viewing the film


IAEA
22: Optimization of Protection in Dental Radiology
24

RP in Dental Radiology
Technical hints to reduce patient doses Privilege the use of sensitive films Sensitivity class D Very good spatial resolution Typical delivered dose: about 0.5 mGy Typical exposure times: 0.3 - 0.7 s Sensitivity class E Good spatial resolution Typical delivered dose: about 0.25 mGy Typical exposure times: 0.1 - 0.3 s
IAEA
22: Optimization of Protection in Dental Radiology
25

Radiation Protection in Dental Radiology

Technical hints to reduce patient doses Lead apron and collar


Useful when the path of primary beam intercepts the protected organs (downward bite-twin projection).

IAEA

22: Optimization of Protection in Dental Radiology

26

Radiation Protection in Dental Radiology

Panoramic examination
Image quality not as good as in intra-oral
films

Important global information

Relatively low dose


IAEA

(one panoramic examination 35 intra-oral films)


22: Optimization of Protection in Dental Radiology
27

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Part 22: Optimization of Protection in Dental Radiology


Topic 3: Quality Control for Dental Equipment

IAEA
International Atomic Energy Agency

Why Dental QC ?
Widespread use of dental units Lack of QC history on most units Dental practitioners working in the primary
health care sector do not have the continuous medical physics support available in a hospital-based diagnostic imaging department
IAEA
22: Optimization of Protection in Dental Radiology
29

What Tests ?

Collimation
Dose Evaluation

Kilovoltage (kVp)
Leakage Radiation

Exposure Time
Half Value Layer

IAEA

22: Optimization of Protection in Dental Radiology

30

Quality Control for Dental Equipment


The recommended tests are consequently divided into: those simple tests which can be performed by dental practice staff those more complex tests which can be carried out by medical physicists.
IAEA

22: Optimization of Protection in Dental Radiology

31

Quality Control for Dental Equipment


Tests which can be performed by dental practice staff
Physical parameter Image Quality Tolerances 10% reference values Frequency Quarterly

Developer temperature and condition of processing solutions


Processing

Specified by the film manufacturer

Every time processing solutions are used

Base+Fog: >0.2 OD Speed and Contrast >0.15 OD about baseline

Every time processing solutions are changed

IAEA

22: Optimization of Protection in Dental Radiology

32

Quality Control for Dental Equipment


Tests performed by medical physicists
Physical parameter Tube voltage Tolerances >50 kV and error <10% Frequency 3 yearly

Beam size/collimation <60 mm diameter 3 yearly (intra-oral) <150 x 10 mm at cassette (panoramic) Dose at cone tip 50 kV: <5.0 mGy 70 kV: <2.5 mGy (E speed film) <75 mGy mm 1-3 yearly

Dose-width product for panoramic film

1-3 yearly

IAEA

22: Optimization of Protection in Dental Radiology

33

Dental QC Methods
Unit
Intra-Oral Receptors (I/O) Test Method
as for Radiology QC

Cephalometric (Ceph) Panoramic (OPG )


IAEA

as for Radiology QC

where possible: immobilise unit remove slit collimator

22: Optimization of Protection in Dental Radiology

34

Test Equipment
Dosimeter
small & large volume
chambers 2mm wide detector for OPG

kVp meter
measure kVp
average Programmable delay ~100 ms Range: 50 to 120 kV

Timer
triggering at 75%
peak kV

Aluminum filters
4 x 1mm Grade 1100

IAEA

22: Optimization of Protection in Dental Radiology

35

Collimation
Expose film or fluorescent screen Measure x-ray field image
Unit Film position Limits of X Ray field

Standard
OPG

End of Cone 60 mm diameter


Slit, in film cassette (2 films) At cassette Equal slit images Vertical image film height 1 % of FFD on all margins
36

Ceph
IAEA

22: Optimization of Protection in Dental Radiology

Collimator Light Intensity


Ceph units: Place external detector 1m from focus Measure illuminance in lux ( lumen/m2) Read each quadrant Limit: >100 lux at 1m

IAEA

22: Optimization of Protection in Dental Radiology

37

Dead man Switch



timer at 50 cm from focus set low kV, mA, long time start exposure release switch during exposure
Require exposure cut-out when switch is released. Check exposure time is less than set time

IAEA

22: Optimization of Protection in Dental Radiology

38

Kilovoltage
Accuracy: Set kVp meter to ~100 msec delay Observe kVp waveform at 70 kV if poss. Limit: measured kVp within 5% of set value Reproducibility: Take 5 repeat exposures Limit: coefficient of variation 2%

IAEA

22: Optimization of Protection in Dental Radiology

39

Dose Evaluation
Skin dose from I/O units: place cone 10 mm from dosimeter set maxillary molar/ bitewing setting
Should be (65-70 kVp):
2-3 mGy for molar view < 5 mGy for any view

IAEA

22: Optimization of Protection in Dental Radiology

40

Output Reproducibility (1)


Standard I/O units & Ceph units: Dosimeter position:
I/O units ~10 mm from cone Ceph units: 75 cm from focus
or other recommended distance

Five repeat exposures


Limit: coefficient of variation 5%

IAEA

22: Optimization of Protection in Dental Radiology

41

Output Reproducibility (2)


Optional Method for OPG units: Align detector on film cassette slit Measure dose rate Take 5 repeat exposures Limit: coefficient of variation 5%

IAEA

22: Optimization of Protection in Dental Radiology

42

Exposure Time Accuracy


Standard I/O units & Ceph units: Set timer to trigger at 75% peak kV Test times in the normal working range
Limit: 10 % error for I/O units 5% error for all other units

IAEA

22: Optimization of Protection in Dental Radiology

43

Timer Reproducibility
Standard I/O units & Ceph units

Place timer in beam 5 repeat exposures


Limit: coefficient of variation 5%

IAEA

22: Optimization of Protection in Dental Radiology

44

Half Value Layer (HVL)


Standard I/O units: Position cone facing down Place dosimeter at ~ 40 cm from focus Position Al filters near end of cone Measure dose
measure with no added filters with 2,3,4 mm Al added, then again with no filters

Plot on semi-log paper and find HVL


IAEA
22: Optimization of Protection in Dental Radiology
45

HVL Set-up

IAEA

22: Optimization of Protection in Dental Radiology

46

HVL: Ceph & OPG Units


Position Al filters on collimator
Cephalometric units Position dosimeter at 75 cm from focus OPG Units Position dosimeter on film cassette slit Measure dose rate, dose for fixed exposure time, or dose for full scan NB Test kVp accuracy before measuring HVL
IAEA
47

22: Optimization of Protection in Dental Radiology

OPG Quality Control (kVp/HVL measurement)

IAEA

22: Optimization of Protection in Dental Radiology

48

HVL- Minimum Values


kVp HVL (mm Al)
Intraoral 1.5 1.5 2.3 2.5 Ceph/OPG 1.8 2.1 2.3 2.5

60 70 80 90

IAEA

22: Optimization of Protection in Dental Radiology

49

Leakage Radiation
Cover collimator with 2 mm lead Set large ionization chamber (~180 cc)
or solid state detector at known distance d cm from focus Expose on maximum kVp for 0.5 - 1 sec (but be careful!) Scale up the reading to 1 hour

IAEA

22: Optimization of Protection in Dental Radiology

50

Leakage Radiation

Choose the most likely points for


leakage where collimator joins to housing, lateral to
the tube, joints in housing etc.

Normalise measurement to 1 m from


focus with inverse square law:
Dose1m = dosed x (distanced/100)2
IAEA

22: Optimization of Protection in Dental Radiology

51

Leakage Measurement
In normal radiography, leakage
measurements are normalised to the maximum continuous rated tube current R:
Leakage = Dose (for P mAs at 100 cm) x (R x 3600/P)

BUT...a dental unit has no continuous operation rating


IAEA
22: Optimization of Protection in Dental Radiology
52

Leakage Measurement

Instead we normalize the dose reading


to tube duty cycle:

Leakage Limit at 1 m from focus:


0.25 mGy/hour 1.0 mGy/hour - Standard I/O units - Ceph & OPG units

IAEA

22: Optimization of Protection in Dental Radiology

53

Calculation of Leakage
Measured leakage at 100 cm = 2mGy/hr Duty cycle = 1: 20 (i.e. must wait ~ 20 times
the exposure time between exposures to allow tube cooling) Thus actual leakage = 2 / 20 mGy/hr = 0.1 mGy/hr i.e. acceptable
IAEA

22: Optimization of Protection in Dental Radiology

54

Radiation Protection in Dental Radiology


Quality Control of Film Processing

Keep under control time and


temperature of the developing process.

Do not use oxidized chemicals Regularly check processing with


phantom
IAEA
22: Optimization of Protection in Dental Radiology
55

Dental Phantom

IAEA

22: Optimization of Protection in Dental Radiology

56

Dental Phantom

IAEA

22: Optimization of Protection in Dental Radiology

57

Where to Get More Information


Code of practice for protection of persons against
ionising radiations arising from medical and dental use, DHSS, HMSO, London, 1964 Guidance notes for the protection of persons against ionising radiations arising from medical and dental use, DHSS, HMSO, London, 1988 Radiation protection and quality assurance in dental radiology. Radiation protection 81. European Commission.(1995) CG-89-95-971-ENC
IAEA
22: Optimization of Protection in Dental Radiology
58

Summary
General and specific types of x-ray equipment
(and image receptors) are reviewed: (panoramic and cephalometric), with technical data about operating conditions

Although doses are generally low, the high


frequency of examinations requires radiation protection (for the practitioner) in dental radiology

Some tests are detailed for Quality control of


dental equipment.
IAEA
22: Optimization of Protection in Dental Radiology
59

S-ar putea să vă placă și