Sunteți pe pagina 1din 104

Calibration and

Electrical Safety of
Medical Equipment
Dr Fadhl Al-Akwaa
fadlwork@gmail.com
www.Fadhl-alakwa.weebly.co
m
Please contact Dr Fadhl to use this material

Why Test & Calibration


What you cannot
measure you cannot
control
Please contact Dr Fadhl to use this material

As components age and equipment


undergoes
changes in temperature or humidity or
sustains
mechanical stress, performance gradually
degrades.
This is called drift.
When this happens your test results
become unreliable and both design and
performance quality suffer.
While drift cannot be eliminated, it can be
detected and either corrected or
Please contact Dr Fadhl to use this material
compensated for through the process of

Definitions
Calibration: process of comparing an unknown against
a reference standard within defined limits, accuracies
and Uncertainties
Verification: process of comparing an unknown against
a reference standard at usually one data point

Please contact Dr Fadhl to use this material

Requirements of Test &


Calibration service
Written Program
Routine calibration or verification at suitable intervals
Control of inspection, measuring and test equipment.
Calibration procedures including specific directions and limits for
accuracy and precision
Deviation or discrepancies should be investigated
Traceable Calibration Standards
Calibration records
Visible Calibration status
Please contact Dr Fadhl to use this material

What to TEST for?


Performance Testing
Safety Testing

Please contact Dr Fadhl to use this material

When to test
On newly acquired equipment prior
to being accepted for use
During routine planned preventative
maintenance.
After repairs have been carried out
on equipment.

Need for Medical Equipment Testing


Medical device incidents resulting in
patient injury and death
Ensure that the equipment is performing
to the expected standards of accuracy,
reliability, free of hysteresis and linear (as
designed).
Safe and effective devices need to be
available for
patient care
Downtime costs money
Regulations, accreditation requirements
Please contact Dr Fadhl to use this material
and standards.

Why do we do electrical
safety?
Ensure patient safety
Protect against macroshock
Protect against microshock

Test for electrical internal breakdown /


damage to power cord, AC mains feed,
etc.
Meet codes & standards
AAMI, IEC, UL, NFPA, etc.

Protect against legal liability


In case of a patient incident

International Electrotechnical
Commission
The International Electrotechnical
Commission[1] (IEC) is a non-profit,
non-governmental international
standards organization that prepares
and publishes International Standards
for all electrical, electronic and related
technologies collectively known as
"electrotechnology".

International Electrotechnical
Commission

IEC standards cover a vast range of


technologies from power generation,
transmission and distribution to home
appliances and office equipment,
semiconductors, fibre optics, batteries, solar
energy, nanotechnology and marine energy as
well as many others.
The IEC also manages three global conformity
assessment systems that certify whether
equipment, system or components conform to
its International Standards.

International Electrotechnical
Commission

Today, the IEC is the world's leading


international organization in its field, and its
standards are adopted as national standards by
its members. The work is done by some 10 000
electrical and electronics experts from industry,
government, academia, test labs and others
with an interest in the subject.
They also first proposed a system of standards,
the Giorgi System, which ultimately became
the SI, or Systme International dunits (in
English, the International System of Units).

IEC, ISO, ITU, IEEE


The IEC cooperates closely with the
International Organization for Standardization (ISO)
and the
International Telecommunication Union (ITU). In
addition, it works with several major standards
development organizations, including the IEEE with
which it signed a cooperation agreement in 2002,
which was amended in 2008 to include joint
development work.
Other standards developed in cooperation between
IEC and ISO are assigned numbers in the 80000
series, such as IEC 82045-1.

List of IEC standards


IEC standards have numbers in the
range 6000079999 and their titles
take a form such as IEC 60417:
Graphical symbols for use on
equipment. The numbers of older IEC
standards were converted in 1997 by
adding 60000, for example IEC 27
became IEC 60027.

List of IEC standards

IEC 60027 Letter symbols to be used in electrical technology...


IEC 60034 Rotating electrical machinery
IEC 60038 IEC Standard Voltages
IEC 60044 Instrument transformers
IEC 60050 International Electrotechnical Vocabulary
IEC 60062 Marking codes for resistors and capacitors
IEC 60063 Preferred number series for resistors and capacitors
IEC 60065 Audio, video and similar electronic apparatus Safety requirements
IEC 60068 Environmental Testing
IEC 60071 Insulation Co-ordination
IEC 60073 Basic Safety principles for man-machine interface,
marking and identification
http://en.wikipedia.org/wiki/List_of_IEC_standards

List of IEC standards


IEC 60601 Medical Electrical Equipment
IEC 62304 Medical Device Software Software Life Cycle Processes
IEC 62366 Medical devices
Application of usability engineering to
medical devices
IEC 62464 Magnetic resonance
equipment for medical imaging
http://en.wikipedia.org/wiki/List_of_IEC_standards

IEC 60601-x-xx
the IEC 60601-1-xx series of collateral
standards for MEDICAL ELECTRICAL
EQUIPMENT;
the IEC 60601-2-xx series of particular
standards for particular types of MEDICAL
ELECTRICAL EQUIPMENT; and
the IEC 60601-3-xx series of performance
standards for particular types of MEDICAL
ELECTRICAL EQUIPMENT.

IEC 60601-x-xx
IEC 60601-1-2, Medical electrical
equipment Part 1-2: General
requirements for safety Collateral
standard: Electromagnetic compatibility
Requirements and tests
IEC 60601-1-3, Medical electrical
equipment Part 1: General requirements
for safety 3. Collateral standard: General
requirements for radiation protection in
diagnostic X-ray equipment

IEC 60601-x-xx
IEC 60601-1-6, Medical electrical equipment
Part 1-6: General requirements for safety
Collateral standard: Usability
IEC 60601-1-8, Medical electrical equipment
Part 1-8: General requirements for safety
Collateral standard: General requirements,
tests and guidance for alarm systems in
medical electrical equipment and medical
electrical systems

Physiological Effects of
Electricity

The human body can easily detect macroshock and


violent reactions occur to high current flow level in
the body
Below 1 ma (1,000 a), it is often much more
difficult to detect the presence of a shock hazard

Classes and types of medical


electrical equipment
Equipment Class{I,II,III} method of
protection against electric shock
Equipment Type{B,BF,CF} degree of
protection

Classes and types of


medical electrical
equipment
All electrical equipment is categorised
into classes according to the method of
protection against electric shock that is
used. For mains powered electrical
equipment there are usually two levels of
protection used, called "basic" and
"supplementary" protection. The
supplementary protection is intended to
come into play in the event of failure of
the basic protection.

Class I
Class I equipment has a protective earth. The basic
means of protection is the insulation between live
parts and exposed conductive parts such as the
metal enclosure.
In the event of a fault that would otherwise cause
an exposed conductive part to become live, the
supplementary protection (i.e. the protective earth)
comes into effect. A large fault current flows from
the mains part to earth via the protective earth
conductor, which causes a protective device
(usually a fuse) in the mains circuit to disconnect
the equipment from the supply.

Class I

Class I
term referring to electrical
equipment in which protection
against electric shock does not rely
on BASIC INSULATION only, but which
includes an additional safety
precaution in that means are
provided for ACCESSIBLE PARTS of
metal or internal parts of metal to be
PROTECTIVELY EARTHED

CLASS II
term referring to electrical equipment
in which protection against electric
shock does not rely on BASIC
INSULATION only, but in which
additional safety precautions such as
DOUBLE INSULATION or REINFORCED
INSULATION are provided, there being
no provision for protective earthing or
reliance upon installation conditions

Class II

Class III equipment


Class III equipment is defined in some
equipment standards as that in which
protection against electric shock relies on
the fact that no voltages higher than
safety extra low voltage (SELV) are
present. SELV is defined in turn in the
relevant standard as a voltage not
exceeding 25V ac or 60V dc. In practice
such equipment is either battery operated
or supplied by a SELV transformer.

If battery operated equipment is capable of being


operated when connected to the mains (for
example, for battery charging) then it must be
safety tested as either class I or class II equipment.
Similarly, equipment powered from a SELV
transformer should be tested in conjunction with
the transformer as class I or class II equipment as
appropriate.
It is interesting to note that the current IEC
standards relating to safety of medical electrical
equipment do not recognise Class III equipment
since limitation of voltage is not deemed sufficient
to ensure safety of the patient. All medical
electrical equipment that is capable of mains
connection must be classified as class I or class II.
Medical electrical equipment having no mains
connection is simply referred to as "internally
powered".

Equipments Type

different pieces of medical electrical equipment {APPLIED PARTS}


have different areas of application and therefore different electrical safety
requirements. For example, it would not be necessary to make a
particular piece medical electrical equipment safe enough for direct
cardiac connection if there is no possibility of this situation arising.

Normative Reference Page


371
Current density and electrically
induced ventricular fibrillation.
Medical Instrumentation, JanuaryFebruary 1973, Vol. 7, No. 1.
WATSON, AB. and WRIGHT, JS.,
Electrical thresholds for ventricular
fibrillation in man. Medical Journal of
Australia, June 16, 1973.

Terminology and
definitions
http://www.601help.com/Disclaimer/g
lossary.html#ProtectiveEarthTerminal

Terminology and
definitions

L1 Hot
L2
Neutral
Earth
Ground
Mains Line
Voltage
Applied Parts
Patient Leads
Enclosure/Case
Chassis
Protective Earth
Ground Wire
Earth Leakage Current Leakage in Ground
Wire

Terminology and
definitions
Enclosure Leakage
Chassis Leakage
Patient Leakage
Lead Leakage
Patient Auxiliary
Leakage between
Patient Leads
Mains on Applied Parts
Lead Isolation
Insulation Resistance Dielectric Strength
or Insulation Resistance between Hot and
Neutral to Ground
Earth Resistance Ground Wire Resistance

R.M.S and Peak to Peak


Vrms is the value indicated by the vast majority of AC voltmeters.

The RMS value of an alternating voltage or current is the same as


the level of direct voltage or current that would be needed to produce
the same effect in an equal load.
For example, 1 V applied across a 1 resistor produces 1 W of
heat. A 1 Vrms square wave applied across a 1 resistor also
produces 1 W of heat. That 1 Vrms square wave has a peak voltage of
1 V, and a peak-to-peak voltage of 2 V.

Calculate RMS

0.707 Vpk

RMS is a sort of average and peak is the top level


A peak is an instant reading - RMS is an "average"
reading.
RMS means to take the root of the mean and
square it.
RMS value is the DC equivalent value to an AC
stream.
The RMS value of an alternating voltage or current
is the same as the level of direct voltage or
current that would be needed to produce the
same effect in an equal load.

Crest Factor

The Crest Factor is equal to the peak


amplitude of a waveform divided by the
RMS value.
Electrical engineering for describing
the quality of an AC power waveform

Applied Part

No applied part

table

Parts that contact PATIENTS


Applied Part
A part of the equipment which in normal use:
necessarily comes into physical contact with the patient for the equipment
to perform its function; or can be brought into contact with the patient; or
needs to be touched by the patient

Accessible Part
Part of equipment which can be touched
without the use of a tool.
EXAMPLE 1 Illuminated push-buttons
EXAMPLE 2 Indicator lamps
EXAMPLE 3 Recorder pens
EXAMPLE 4 Parts of plug-in modules
EXAMPLE 5 Batteries

Leakage currents
Current that is not functional.
several different leakage currents are
defined according to the paths that the
currents take.
Earth Leakage Current
Enclosure Leakage Current
Patient Leakage Current
Patient auxiliary current

Causes of Leakage
currents
If any conductor is raised to a potential
above that of earth, some current is bound
to flow from that conductor to earth.
The amount of current that flows depends
on: 1- the voltage on the conductor.
2- the capacitive reactance between the
conductor and earth.
3-the resistance between the conductor and
earth.

EARTH LEAKAGE
CURRENT
current flowing
from the MAINS
PART through or
across the
insulation into the
PROTECTIVE EARTH
CONDUCTOR

EARTH LEAKAGE
CURRENT
Under normal conditions, a person who is in contact
with the earthed metal enclosure of the equipment and
with another earthed object would suffer no adverse
effects even if a fairly large earth leakage current were
to flow. This is because the impedance to earth from
the enclosure is much lower through the protective
earth conductor than it is through the person. However,
if the protective earth conductor becomes open
circuited, then the situation changes. Now, if the
impedance between the transformer primary and the
enclosure is of the same order of magnitude as the
impedance between the enclosure and earth through
the person, a shock hazard exists.

EARTH LEAKAGE
CURRENT Measurement

Measurement of earth
leakage current

Enclosure leakage
current /
touch
current
LEAKAGE CURRENT
flowing from the
ENCLOSURE to
earth or to another
part of the
ENCLOSURE
through a
conductor other
than the protective
earth conductor.

Enclosure leakage
current/
touch current

Measurement of enclosure
leakage current

Patient leakage current


Patient leakage current
is the leakage current
that flows through a
patient connected to an
applied part or parts.
It can either flow from
the applied parts via
the patient to earth or
from an external source
of high potential via the
patient and the applied
parts to earth.

Patient leakage current

Measurement of patient
leakage current

Measurement of patient
leakage current

Patient auxiliary current


The patient auxiliary
current is defined as
the current that
normally flows
between parts of the
applied part through
the patient, which is
not intended to
produce a
physiological effect

Patient auxiliary current

Measurement of patient
auxiliary current.

Mains on applied parts

Protective Earth
Continuity
The resistance of the
protective earth conductor
is measured between the
earth pin on the mains plug
and a protectively earthed
point on the equipment
enclosure (see figure 6).
The reading should not
normally exceed 0.2 at
any such point. The test is
obviously only applicable
to class I equipment.

Protective Earth
Continuity
In IEC60601, the test is
conducted using a 50Hz
current between 10A and 25A
for a period of at least 5
seconds. Although this is a
type test, some medical
equipment safety testers
mimic this method. Damage
to equipment can occur if
high currents are passed to
points that are not
protectively earthed, for
example, functional earths.

Applicable to Class I, all types


Limit: 0.2
DB9801 recommended?: Yes, at 1A
or less.
HEI 95 recommended?: Yes, at 1A or
less. Notes: Ensure probe is on a
protectively earthed point

Insulation Tests Class I


HEI 95 and DB9801
recommended that for
class I equipment the
insulation resistance be
measured at the mains
plug between the live and
neutral pins connected
together and the earth
pin. Whereas HEI 95
recommended using a
500V DC insulation tester,
DB 9801 recommended
the use of 350V DC as the
test voltage.

Applicable to Class I, all types


Limits: Not less than 50M
DB9801 recommended?: Yes
HEI 95 recommended?: Yes
Notes: Equipment containing mineral
insulated heaters may give values
down to 1M. Check equipment is
switched on.

Insulation Tests Class II


HEI 95 further
recommended for class
II equipment that the
insulation resistance be
measured between all
applied parts connected
together and any
accessible conductive
parts of the equipment.
The value should not
normally be less than
50M (see figure 10).

Leakage current
summary
The following table summarises the
leakage current limits (in mA) specified by
IEC60601-1 (second edition) for the most
commonly performed tests. Most
equipment currently in use in hospitals
today is likely to have been designed to
conform to this standard, but note that the
allowable values of earth leakage current
have been increased in the third edition of
the standard as discussed above.

Leakage current
summary
The following table summarises the
leakage current limits (in mA)
specified by IEC60601-1 (second
edition)

http://www.ebme.co.uk/arts/safety/part6.htm

Limitation of voltage, current or


energy(87)

Electrical Safety Tests

Available electrical safety tests include:


Mains Voltage
Dual Lead Voltage
Dual Lead Leakage
Current Consumption
Insulation Resistance
Protective Earth Resistance
Earth Leakage Current
Enclosure Leakage Current
Patient Leakage Current
Mains on Applied Part Leakage
Patient Auxiliary Current
Accessible Voltage
Accessible Leakage
Equivalent Device Leakage
Equivalent Patient Leakage

How to measure resistance?

Proper grounding is the best defense


against macroshock & microshock!
The 2-terminal method is
less accurate due to the
effects of test lead
resistance, especially with
long leads and low resistance
value measurements. The
much preferred 4-terminal
Kelvin technique negates
the effects of test lead

An Introduction to Safety
Analyzer

Electrical Shock
Hazard
common experience due to electric

A
shock
Associated with equipments
Electric current can flow through the
human body either
Accidentally or Intentionally

Other reasons of electric shock include

Careless use of electricity


Usage of faulty cords and appliances
Lack of concept/Faulty design
Relied upon life support devices (Pace
maker/ respirators)

An Introduction to Electric
Shock Hazard

Electrical Shock Hazard


Use of medical equipments
in conjunction with other
instruments
and
equipments
Environmental conditions
Patient/Operator
not
realizing potential hazards
2
situations
account
hazards from electric shock
Gross shock
Micro - current shock

An Introduction to Electric
Shock Hazard

Gross shock
Experienced
by
the
subject by an accidental
contact
with
electric
wiring at any point on
the surface of the body
Current flows through
the body of the subject
(ex. from arm to arm)
Body acts as a volume
conductor at the mains
frequency
Degree of simulation
varies from individual to
individual

An Introduction to Electric
Shock Hazard

Micro - current shock


Current
passes
directly
through the heart wall
Thresholds of sensation of
electric currents
differ
widely
Greater % of current may
flow via the arterial system
directly through the heart
Requires much less currents
to
produce
ventricular
fibrillation
EX. Catheter laboratory or
operating room where patient
connected to catheter in the
heart

Here patients have very little


resistance
to
electric
currents

Physiological Effects of
Electricity

Tissue Resistance

Skin - 5000 ohms/cm2


Blood - 100 ohms/cm2
Muscle - 200-400 ohms/cm2
Fat - 2000-3000 ohms/cm2
Bone - 3000+ ohms/cm2
Current goes to the path of least
resistance

An Introduction to Electric
Shock Hazard

Leakage Current
Inherent flow of non functional
current from live electric parts
of instrument to accessible
metal parts
Usually flow through 3rd wire
connection to ground
Magnitude of leakage current is
determined by the value of the
capacitance present
Originates due to capacitive
coupling
from
the
power
transformer primary to other
parts of the transformer (or
instruments)

An Introduction to Electric
Shock Hazard

Types of Leakage
Current

Enclosure leakage current: Current


flows in normal condition from the
enclosure (or part of enclosure)
through a person in contact with an
accessible part of enclosure to earth
(or another part of the enclosure)
Earth leakage current: Current flows
in normal condition to earth from main
parts of apparatus via earth conductor
Patient leakage current: Current
flows through patient from or to
applied parts of the patient circuits

An Introduction to Electric
Shock Hazard

Effects of Electric Current on


Human Body

Threshold of perception of
electric shock is about 1mA
Tingling sensation is felt
when
contacted
with
electrified object through
intact of skin
As magnitude of alternating
current is increased
Tingling sensation leads to
contraction of muscles
Muscular
contraction
increases
Finally value of current is
reached where grip of current
cannot be released

An Introduction to Electric
Shock Hazard

Effects of Electric Current on


Human Body

Let -go- current: the max. current at which the


subject is still capable of releasing a conductor by
using muscles
Here individual can withstand with no serious effects
Average let go current for males- 16mA, females-10.5mA,
approx. 9-6 mA for both.

Hold-on-type: A current level higher than let go


current, the subject looses ability to control his own
muscle action and unable to release grip on the
conductor
Such currents are very painful and hard to bear
Physical injury is caused b currents in range of 20-100mA

A very high currents (6 amperes and above) lead to


Temporary respiratory paralysis
Serious burns

An Introduction to Electric
Shock Hazard

Recommendations of
IEC

International Electrotechnical
Commission (IEC)
Continuous
medical
equipment current should
not exceed 100 uA
Should
be
with
in
a
frequency range of 0 to 1kHz
In
abnormal
situations
recommended max. current
is 500 uA
Should be in a frequency
upto 1 kHz
Above 1 kHz max increases
is
proportionally
with
frequency

An Introduction to Electric
Shock Hazard

Precautions
Use apparatus or appliances
with three wire power cord
Provide isolated input circuits on
monitoring equipment
Have periodic checks of ground
wire continuity
No other equipment to be
connected
when
patient
monitoring
equipment
is
connected
Clearly mark functional controls
Take care of adapter plugs that
do not ensure proper grounding
circuit
Direct operating instructions to
the operators
Maintaining voltage differences

Macroshock and
Microshock

History
The expansion of technology was unregulated, and
unexplained deaths in hospitals were attributed
by
some to electrical shocks
Studies showed that electrical shock risks were the
greatest when the patient had conductors internal
to the body
If a conductive catheter is placed in the heart, 100
micro amps at 60 Hz can cause fibrillation of heart
and death

Automated Electrical Safety


Analyzer
601PRO Series XL Fluke
Biomedical

601PRO Series XL

Standard Features
The most advanced Electrical Safety
Analyzer on the market
EN60601-1, EN601010-1, and AAMI
& ESI test loads (user selectable)
into one device
The One-Touch-Testing user interface
Allows user to perform rapid tests on
various medical devices
Multiple enclosure-leakage points
Multiple patient-applied-part types

601PRO Series XL

Standard Features
Power ON/OFF delay
DC only for patient- and auxiliary-leakage
tests
User-programmable test sequences
Offers manual, auto, step, and computercontrol mode operations
ASCII data transfer
Memory for up to 1000 device-information
records
Conducts electrical safety testing in
accordance with IEC 601-1, VDE 751, VDE
701, HEI 95, IEC 1010, AAMI, and AS/NZS
3551 requirements

601PRO Series XL

Standard Features
Flags
failures,
and
simulates
performance,
ECG,
and
arrhythmia,
waveforms.
Results
automatically
analyzed and saved in
non-volatile memory
Accepts device information
that is input using an
External keyboard,
Integrated keypad,
Barcode keyboard wedge

Optional Feature
Onboard thermal printing

601PRO Series XL

Specifications
Voltage
Insulation
Resistance
Current
Consumptio
n

Range:

0 to 300 V True RMS (single and dual lead)

Accuracy:

DC - 100 Hz 1.5 % of reading 1 LSD

Range:

0.5 to 400.0 M

Accuracy:

5 % of reading 2 LSD

Range:

0 to 15 A ac True RMS

Accuracy:

5 % of reading 2 LSD

Mains on
Applied
Applied Part Voltage:
Protective
Earth
Resistance

Supply
Voltage

110 % of mains voltage

Accuracy:

2 % of reading 6 A

Range:

0.000 to 2.999

Accuracy:

5 % of reading 4 m (1 A, 10 A, and 25 A test


currents) (Refer to Operators Manual for
additional specs qualifying the effects on accuracy
of variations in load inductance and phase angle.)
90 to 265 Vac, auto switching

601PRO Series XL

Specifications
IEC601-1 and
AAMI Leakage
Currents

ECG Simulation
and
Performance
Testing

Range:

0 to 8000 A True RMS

Accuracy:

(per IEC601-1 or AAMI filter),


-DC - 1 kHz 1 % of -reading 1
A
-1 to 100 kHz 2 % of reading 1
A
- 100 kHz to 1 MHz 5 % of
reading 1 A

DC-Only Frequency
Response:

DC - 5 Hz (approx)

ECG Complex:

30, 60, 120, 180, 240 BPM

Performance

Dimensions

Pulse:
600 to 700 s rise
and fall time
Sine Waves:

30, 60 BPM, 63 ms pulse width

Square Wave:

0.125, 2.000 Hz (50 % duty cycle)

Triangle Wave:

2 Hz, 2 mV

10, 40, 50, 60, 100 Hz

601PRO Series XL

Available electrical
safety tests

Mains Voltage
Dual Lead Voltage
Dual Lead Leakage
Current
Consumption
Insulation Resistance
Protective Earth
Resistance
Earth Leakage
Current
Enclosure Leakage
Current
Patient Leakage
Current

Mains on Applied
Part Leakage
Patient Auxiliary
Current
Accessible Voltage
Accessible
Leakage
Equivalent Device
Leakage
Equivalent Patient
Leakage

601PRO Series XL

Accessories

Probe/Safety Lead, Red - 1


Probe/Safety Lead, Black - 1
Adapter, Banana/Alligator - 5
Operators Manual - 1
Large Clamp, Red - 1
Warranty Card - 1
Printer Paper Roll (original) - 1
Printer Paper Roll (new style) - 1

601PRO Series XL

Optional
Accessories

Carry Case
RS232 Cable (9M-9F)
Printer Cable
Barcode, Keyboard, Wedge
Adapter, Banana, ECG
Keyboard English
Powercord Set Australian
Powercord Set Schuko
Powercord Set US 120 V
Powercord Set UK

601PRO Series XL

System
Characteristics

Keys
grouped
by
color
and
functionality
Red keys -used to access menu
options
Include previous key, the four SOFT
KEYS, and the enter key

Black keys -gain access to additional


functions
Include the esc/stop key, the view
present settings key, the print header
key, and the print data key.

601PRO Series XL

Setting Up the 601PRO


1.
2.
3.
4.

Using Factory Default Settings


Selecting the Test Standard
Selecting the Printer Output
Selecting the RS232 Baud
Rate
5. Activating the Beeper
6. Setting the Time and Date

601PRO Series XL

Setting Up the
601PRO
7. Configuring the Enclosure Leakage
for the Auto mode Sequence
8. Selecting Language Options
9. Selecting the DC Option
10.Selecting the Auto/Step Tests:
Controlled Power Sequences or
601CE Conventional Test
11.Sequences enabling Stop on Failure
12.Configuring for Device Records or
Templates

601PRO Series XL

Manual Mode
1. Connecting the Device Under
Test
2. The Power-Up Sequence
3. Selecting the Test Standard
4. Selecting the Class/Type
5. Saving Standard, Class, Type and
Test Current
6. Using View Present Settings
7. Manual Operation

601PRO Series XL

Auto/Step Modes
1. Selecting Auto or Step Mode
Testing
2. Executing Auto and Step Mode
Tests
3. Creating/Editing
a
Device
Record or Template

601PRO Series XL

Test Records
1. Sending Test Results from the
601PRO to the Host
2. Computer
3. Test Data Record: Serial Output
4. Printing Test Records
5. Deleting Test Records

601PRO Series XL

Device Records and


Templates
1. Connecting the 601PRO and the Host
Computer
2. Sending Device Information Records
from the 601PRO to the Host Computer
3. Receiving Device Information Records
from the Host computer
4. Device Information Record: Definition
of Fields
5. Device Information Record Format
6. Deleting Device Records and Templates

601PRO Series XL

Testing Devices
1. Permanently Wired Devices
2. Portable Devices
3. Portable Devices in Isolated Power
Systems
4. Testing
Three-Phase
Portable
Devices
5. Testing Conductive Surfaces
6. Detachable Power Supply Cable
7. Battery-Powered Equipment

601PRO Series XL

Standards and
Principles
1. Accessing System Setup
2. Selecting the Test Standard
3. Referring to Test Limits for the
Selected Standard

S-ar putea să vă placă și