Documente Academic
Documente Profesional
Documente Cultură
CT SCAN
History
The
Prototype of CT scanner
EMI scanner
Tomosynthesis
Simple
Generations
generation
configuration
detector
beam
first
Translate -rotate
1-2
Pencil thin
2.5min
second
Translate -rotate
3-52
Narrow fan
10sec
Third
Rotate- rotate
256-1000
Wide fan
0.5sec
fourth
Rotate- fixed
600-4800
Wide fan
1sec
fifth
Electron beam
1284
Wide fan
electron beam
33ns
Third
Electron Beam CT
Electron
Instead
Helical or Spiral CT
Helical,
In
Multislice CT
The
Dual Source CT
Siemens
introduced a CT model with dual Xray tube and dual array of 64 slice detectors,
at the 2005 Radiological Society of North
America (RSNA) medical meeting.
Dual sources increase the temporal
resolution by reducing the rotation angle
required to acquire a complete image, thus
permitting cardiac studies without the use of
heart rate lowering medication, as well as
permitting imaging of the heart in systole.
The use of two x-ray units makes possible
the use of dual energy imaging.
Diagnostic use
Since
Advantages
First
In
Process
CT
scan
illustration
X-ray
contrast
Windowing
Windowing
Artifacts
Although
CT is a relatively accurate
test, it is liable to produce artifacts,
such as the following:
Aliasing Artifact or Streaks
These appear as dark lines which
radiate away from sharp corners.
It occurs because it is impossible for
the scanner to 'sample' or take
enough projections of the object,
which is usually metallic. It can also
occur when aninsufficient penetration
of the x-ray occurs.
Ring Artifact
Probably the most
common
mechanical artifact, the image of one
or many 'rings' appears within an
image. This is due to a detector fault.
Noise Artifact
This appears as gaining on the image
and is caused by a low signal to noise
ratio. This occurs more commonly
when a thin slice thickness is used. It
can also occur when the kV or mA is
too low.
Motion Artifact
This is seen as blurring
which is caused by patient
movement. This is not so
much a problem these
days with faster scanning
times in the use of MDCT.
Beam Hardening
This can give a 'cupped
appearance'. It occurs
when there is more
attenuation in the center
of the object than around
the edge. This is easily
corrected by filtration .
The principle
Multiplanar reconstruction
Multiplanar
MPR
3D rendering techniques
Surface
rendering
A threshold value of radiodensity is chosen
by the operator (e.g. a level that corresponds
to bone). A threshold level is set, using edge
detection image processing algorithms.
From this, a 3-dimensional model can be
constructed and displayed on screen.
Multiple models can be constructed from
various different thresholds, allowing
different colors to represent each anatomical
component such as bone, muscle, and
cartilage.
However, the interior structure of each element is
not visible in this mode of operation
Volume
rendering
Surface rendering is limited in that it will
only display surfaces which meet a threshold
density, and will only display the surface that
is closest to the imaginary viewer.
In volume rendering, transparency and
colors are used to allow a better
representation of the volume to be shown in
a single image - e.g. the bones of the pelvis
could be displayed as semi-transparent, so
that even at an oblique angle, one part of
the image does not conceal another.
3D rendering software
Some
examples of CT 3D surface
rendering software include
Mimics, 3D doctor, Amira....etc
Some examples of CT 3D volume
rendering software include
3D doctor, ScanDoc-3D....etc
Image segmentation
Segmentation
(image processing)
Where different structures have
similar radiodensity, it can become
impossible to separate them simply by
adjusting volume rendering
parameters. The solution is called
segmentation, a manual or automatic
procedure that can remove the
unwanted structures from the image.
reconstructed in 3D
Conclusions.
This
Thank
you