Documente Academic
Documente Profesional
Documente Cultură
Resolution: 1m
LONG STROKE TYPE: GS-1830 Suitable for in-line use
GS-5050A/5100A
GS-6713 Resolution: 10m
10 m RESOLUTION
GS-5051A/5101A GS-6730 Resolution: 10m
1 m RESOLUTION
GS-6813 Resolution: 1m
All models are dust and splash proof conforming to the protection class IP64 of environmental resistance.
Compatible with all of our gauge counters.
Can be connected to a PLC* using 2-phase square wave output signal.
*PLC: Programmable Logic Controller
Range of models available
Suitable for production line in use
High Density design
Simple Mounting
A choice of output format
Specification
Digital Linear Gauge GS-1713/1730/1813/1830
Mechanical specification
Item GS-1713 GS-1730 GS-1813 GS-1830
Measurement range 13mm 30mm 13mm 30mm
Resolution 10m 1m
Accuracy (+20C) 3m 2m 3m
Maximum spindle velocity*1 1 (4) m/s 0.3 (1.2) m/s
Measurement force*2 (downward) 1.7N or less 2.0N or less 1.7N or less 2.0N or less
Number of sliding times (proven in our endurance test) 6.5 million or more
Protection class (excluding connector section) IP64
*1 When used with Ono Sokki's Gauge Counter. The value within parentheses ( ) is the maximum spindle velocity used with the DG-4140/4190/5100.
*2 An attachment may not return completely if it is facing upwards.
Electrical specification
OUTPUT:
The gage head processes internally detected signals and outputs square-wave signals as shown below. These
operating signals, which are square waves having
phase difference of 90 degrees, are equivalent to RS-422A signals, allowing for the independent use of the gage
head. However, certain models (LGD and LGS), do not output square-wave signals but generate Digimatic
code (SPC) output in order to identify the measurement position.
Phase-A
Phase-B
Mount GS-1700/1800/6700/6800 Series Linear Gauge Sensor to Mounting by squeezing stem
the xing device so that the spindle moving direction coin- cides Recommended tightening torque: 0.4 to 0.5Nm
with the length (displacement) direction of the object to be
measured (see the gure below).
Mounting using the xing holes on the main unit.
When the sensor is mounted, there is a gap of approximately 0.2mm
between the sensor top and the mounting surface.
Never try to eliminate this gap by pressing with brackets, etc
When squeezing the stem, note the following:
After mounting, check movement of the spindle. If the spindle
does not move smoothly, it is tightened excessively. Retighten the stem Mounting by squeezing stem using collar
with less torque. Notethat, with too little torque, the stem is not secured
disabling accurate measurement (do not x the stem by directly pushing Recommended screw : 4
with a screw). Recommended tightening torque : 0.4 to 0.5Nm
If the spindle moves smoothly after tightening the stem, it is
tightened appropriately.
The mounting surface and base bottom of the sensor are made parallel
and rectangular with the stem, respectively. Rectangular mounting can
be accomplished easier by forming L-shaped lower end of the xing jig.
Mount a rigid sensor xing jig so that the spindle is rectangular with the
object to be measured.
The recommended mounting angle is 0.15 (squareness 260m/100mm)
or less.
The linear gauge sensor obliquely mounted will cause a lateral force to Recommended collar size
be applied to the spindle, resulting in and material
failures. Material: Brass
MEASUREMENT:
Procedures for measuring with GS-1700/1800/6700/6800 Series Linear
Gauge Sensor are as described below.
Connect a gauge counter to the linear gauge sensor.
Holding the gauge head with ngers, slowly move the spindle for
measurement (move the spindle close to the measurement object until
the gap is 1mm or less and then let the spindle fall).
The spindle stopper inside the sensor is made of rubber. Therefore, it
should not be used as the reference point for measurement.
Use the point where the spindle is pushed in 0.2mm or more as the
reference point.
APPLICATIONS:
Incorporation into auto-measurement machine:
Gage heads (LGF with EH counter) can be incorporated Mobile terminal chassis multipoint measurement:
into the auto- measurement unit for outside diameter Multiple point measurement of mobile terminal parts is
and/or height measurement. performed with Linear Gage LGF-0510L-B and EV counter
EV-16P.