Documente Academic
Documente Profesional
Documente Cultură
and Measurements
N.V. Raghavendra
L. Krishnamurthy
Measurement Systems
Threshold
If the input to the instrument is gradually increased from zero, a
minimum value of that input is required to detect the output.
Loading Effects
Any measuring instrument generally consists of different elements
that are used for sensing, conditioning, or transmitting purposes.
Ideally, when such elements are introduced into the measuring
system, there should not be any distortion in the original signal.
However, in practice, whenever any such element is introduced into
the system, some amount of distortion occurs in the original signal,
making an ideal measurement impossible.
Time delay type In this type, the measuring system begins to respond after a
dead time to the applied input.
Dead time is defined as the time required by the measuring system to begin
its response to a change in the quantity to be measured.
Dead time simply transfers the response of the system along the time scale,
thereby causing a dynamic error.
This type of measurement lag can be ignored as they are very small and are
of the order of a fraction of a second.
If the variation in the measured quantity occurs at a faster rate, the dead
time will have an adverse effect on the performance of the system.
© Oxford University Press 2013. All rights reserved.
Resolution of Measuring Instruments
Fidelity: It is defined as the degree to which a measurement system
indicates the changes in the measured quantity without any dynamic
error.
The device used for detecting the input signal is known as a transducer
or sensor.
Table 12.1 gives some of the examples for the three stages of a
generalized measurement system.