i-CHROMA
Reader
Operation Manual
TABLE OF CONTENTS
English
I Quick Manuals E.1
IV Specifications E.4
Quick Manual
I
II
III
Rev. 10 E.1
Section II
II
Introduction and Principle of Operation
Introduction
Reader and for the sample testing procedures. Refer to the test procedure of
of analytes in blood or urine. The assays and the instrument are for in-vitro
laser as the excitation light source. The emitted light from the fluorescence
dye is collected and converted into an electrical signal. The signal is closely
related to the amount of fluorescing dye molecules present on the spot under
examination.
After a buffer-mixed sample is applied to the Test Device, the Test Device is
can accept Test Devices that are designed specifically for use with this
instrument.
Rev. 10 E.2.1
II
Introduction
i-CHROMATM Reader is equipped with a built-in Test Device
holder and does not require an external holder to place the Test
Device in. The Test Device holder will appear through the
The main power switch, located on the left side of the instrument,
with its own power supply. These two power supplies have
Rev. 10 E.2.2
Section III
III Contents of System
Unpacking
Contents
After opening the box, compare the contents with the list below.
Contents
i-CHROMATM Reader set
i-CHROMATM Reader 1
Operation manual 1
Power cable for Reader 1
Connection cable 1
Printer set (Optional)
Thermal printer 1
Manual for printer 1
Printer paper (roll) 1
Pipette set (Optional)
Pipette (10, 20, 30, 75, 100ul) 1
Capillary tube (5ul) 1
System Check Chip set
Cartridge 1
ID chip 1
Rev. 10 E.3
Section IV
Specifications IV
Physical Description
Specifications
Dimensions 250 (L) x 185 (W) x 80 (H) mm
Weight 2 kg
Power supply 100-240V AC, 50/60Hz, 0.5-1.3A
Data output: On-board screen / Printer
Environmental Set-up
Temperature 15oC ~30oC
Humidity 10 ~ 80%
Location Dry, clean, flat, horizontal surface away from
direct sunlight and mechanical vibration.
Optical Description
Light source Laser diode, 2.5 mW, 637nm
Detector Silicon photo diode
Other
Driver Motor 12V
Interface RS-232 serial (I/O) port
Printer Thermal
Display LCD (16x4 character)
Key pad 5 function keys
Rev. 10 E.4
Section V
Initial Installation Procedure
V
Power requirements
Installation
Location
To set-up i-CHROMATM Reader for the first time, remove the contents from
the packing materials. Inspect the contents for any physical damages and any
missing components compared to the suggested list of contents.
Rev. 10 E.5.1
4. Turn on the power switch on the left side of i-CHROMATM Reader.
V
Installation
Reader
5. i-CHROMATM Reader will execute a series of self-checking routine.
Upon successful completion, the display monitor will show the initial
screen as shown below. (The actual contents of the display could be differ,
depending on the specification and customer requirements.)
MENU 22D
13:13:00
* Test Mode
Setting Mode
7. At the end of the run, remove the Test Device from i-CHROMATM
Reader. Also, make sure the sample holder is retracted before turning
the power off.
Caution: Keep the cables, cords and wirings for the power adapter
and the printer away the front and the top surface of i-CHROMATM
Reader. These could render it hazardous to connect/disconnect the
power adapter (and the printer).
Rev. 10 E.5.2
Section VI
Printer Installation
VI i-CHROMATM Reader comes with an optional printer that preconfigured
to work. For proper handling and installation of the printer, refer to the
Installation
instruction manual accompanying the printer. With the printer ready for
Printer
Rev. 10 E.6
Section VII
Operation
The operation of i-CHROMATM Reader is performed by five primary
functions each of which can be represented by corresponding key pad on the
front panel. The software in i-CHROMATM Reader is designed concisely
VII
and user-friendly to be used in a point-of-care or a central laboratory setting.
Each of these functions is addressed in this section.
Operation
Function Keys
The “cursor” in these descriptions stands for the “>>” notation.
This function moves the sample carrier in and out of the main
In/Out body of i-CHROMATM Reader. This key is also used to finish
the run: the sample holder stays extended at the end of the run
and this key retracts the holder back into the main body.
Rev. 10 E.7.1
Procedure for System Check
The following steps describe recommended procedures for a test in i-CH
VII ROMATM Reader with a System Control Device.
Operation
Rev. 10 E.7.2
Prepare a System Control Device
Insert Control
Cartridge And then, you should do a test
Select to Start Insert a System Control Device into the holder
VII
after a sample loading .
Press “Select” key to start.
Operation
System Check i-CHROMATM Reader will scan the System
Reading
Progressing
Control Device automatically.
Rev. 10 E.7.3
Procedure for Test Mode
The following steps describe recommended procedures for a test in i-CH
VII ROMATM Reader with a specific Test Device. You can use only whole
blood for test (No serum or commercial controls)
The actual display may differ from the ones shown below, depending on
Operation
Rev. 10 E.7.4
Prepare a Test Device and apply sample mixture
Insert hsCRP
Cartridge onto the Test Device.
And then, you should do a test
Select to Start VII
Insert a Test Device into the holder after a sample
loading .
Press “Select” key to start.
Operation
hsCRP Test
Reaction
Progressing
05 : 00
Rev. 10 E.7.5
Procedure for Time Setting and
Recall Previous Results
VII MENU 22D
13:13:00
Test Mode Choose “Setting Mode” and “Time Set” or
* Setting Mode “Recall Result” by moving cursor with “Up” or
Operation
MENU 22D
“Down” key. Press “Select” key to move next stage
13:13:00 after you chose a specific mode.
Recall Result
* Time Set
Rev. 10 E.7.6
Caution in handling the cartridge
• The test sample should be applied to the “B” well, as shown below.
VII
A B
Operation
• When inserting the Test Device into the holder, orient the Test
• Push the Test Device all the way into the holder until the Test Device
comes to a stop. Use only the push of your thumb or the index finger.
mechanical failure.
Rev. 10 E.7.7
Section VIII
Warnings, Precautions and Limitations
system.
Warnings
Rev. 10 E.8
Section X
Trouble Shooting
Symptoms users can take corrective measures
Rev. 10 E.10.1
Symptoms Probable Causes Recommended
Measures
Message: Technical problem with the Retest using fresh
“Error 25~28” Test Device sample or device.
Message: Optical failure Call customer service.
“Error S1 or S2”
Unusual sample or X
Message: incorrect storage of Retest using fresh
“Value Error 1-10” reagents sample or device.
Shooting
Test Device
Trouble
“Error” and repeat the test.
Message: Remove the Test
“Caution Barcode Device and try again.
Error” Test Device alignment is Repeat with a new Test
off. Device .
Message:
“Lot Number Error” Call customer service.
Message: Does not match the Test Check the Test Device
“ID Chip Lot Device lot # with ID Chip lot # and ID Chip #.
Number Mismatch” #
Message: Does not match the Test Check the Test Device
“ID Chip Cartridge Device Item with ID Chip Item and ID Chip Item.
Mismatch” Item
Rev. 10 E.10.2